all | frequencies |
|
|
exhibits | applications |
---|---|---|---|---|---|
manuals |
app s | submitted / available | |||||||
---|---|---|---|---|---|---|---|---|
1 |
|
User Manual 20180228 v1 - SiXCO EXHIBIT 9-1 Install Guide | Users Manual | 1.10 MiB | May 03 2018 / January 09 2018 | |||
1 |
|
User Manual 20180228 v1 - SiXCO EXHIBIT 9-2 Manual | Users Manual | 1.23 MiB | May 03 2018 / January 09 2018 | |||
1 | Test Report | May 03 2018 | ||||||
1 | Cover Letter(s) | May 03 2018 | ||||||
1 | Cover Letter(s) | May 03 2018 | ||||||
1 | External Photos | May 03 2018 / January 09 2018 | ||||||
1 | Cover Letter(s) | May 03 2018 | ||||||
1 | Internal Photos | May 03 2018 / January 09 2018 | ||||||
1 | Test Report | May 03 2018 | ||||||
1 | RF Exposure Info | May 03 2018 | ||||||
1 | ID Label/Location Info | May 03 2018 | ||||||
1 | Test Setup Photos | May 03 2018 / January 09 2018 |
1 | User Manual 20180228 v1 - SiXCO EXHIBIT 9-1 Install Guide | Users Manual | 1.10 MiB | May 03 2018 / January 09 2018 |
SiXCO Carbon Monoxide (CO) Detector / Monoxyde de carbone (CO) Quick Installation Guide This device is intended for use with Honeywell controls that support SiX series devices including the SiXCO. Check the control for compatibility with this device. This device occupies one control panel zone. See Honeywell Installation Guide (P/N 800-23974 https://mywebtech.honeywell.com/) and the Limitations of Carbon Monoxide Alarm Systems Insert (P/N 800-15144 12/15) and Limitation of Carbon Monoxide Detectors (P/N K14670V1 4/08 Rev. A). FEATURES 10-year life Carbon Monoxide (CO) sensor CO Detector End-of-Life reporting (detector needs replacing) CO Maintenance Reporting Low Battery Detection One Go / All Go: all CO detectors in the system programmed as one-go-all-go will sound on alarm. POWER UP Remove battery pull tab. Be sure batteries are seated properly to avoid a low battery condition after 15 seconds. Upon power up, Green LED Blinks every 2 secs / Sounder is Silent. ENROLLMENT IMPORTANT NOTES:
The Carbon Monoxide (CO) has a unique MAC ID number (serial number) and is used for enrollment. Once enrolled, the device cannot be used with another system until it is removed from the current controller. See the 24-
hour Enrollment Deletion section below or the controllers instructions. Enroll the Device Services (CO) 1. Remove the detector from the base plate (tamper switch must be faulted when enrolling). 2. Put the control panel in Zone Programming mode. For Lyric Controller: Security >
installer code > Program >
(scroll down) > SiX Programming 3. Pull the battery tab to activate enrollment process (be sure batteries are seated properly to avoid a false low battery condition). If the battery tab has already been removed and the unit is powered up, press the Test switch for less than 3 seconds to activate the enrollment process. Guide d'installation rapide This device is intended for use with Honeywell controls that support SiX series devices including the SiXCO. Check the control for compatibility with this device. This device occupies one control panel zone. Voir les instructions d'installation compltes (800-23974 -
https://mywebtech.honeywell.com/) et les limites de cet insert de systme d'alarme au monoxyde de carbone (800-
15144FC 12/15) CARACTRISTIQUES Signalement de la fin de vie du dtecteur de CO (dtecteur devant tre remplac) 10-year life Carbon Monoxide (CO) sensor CO Maintenance Reporting Dtection de pile faible Une seule fois pour tout : Tous les dtecteurs de fume du systme programms comme Une seule fois pour tout retentiront. POWER UP Remove battery pull tabs. Be sure batteries are seated properly to avoid a low battery condition after 15 seconds. Upon power up, Green LED Blinks every 2 secs / Sounder is Silent. ENRLEMENT IMPORTANT NOTES:
The Carbon Monoxide (CO) has a unique MAC ID number (serial number) and is used for enrollment. Once enrolled, the device cannot be used with another system until it is removed from the current controller. See the 24-hour Enrollment Deletion section below or the controllers instructions. Enroll the Device Services (CO) 1. Remove the detector from the base plate (tamper switch must be faulted when enrolling). 2. Put the control panel in Zone Programming mode. For Lyric Controller: Security >
installer code > Program >
(scroll down) > SiX Programming 3. Pull the battery tab to activate enrollment process (be sure batteries are seated properly to avoid a false low battery condition). If the battery tab has already been removed and the unit is powered up, press Test switch for less than 3 seconds to activate the enrollment process. 4. The device attempts enrollment (green LED flashes about 8 seconds). If successful, the 4. The device attempts enrollment (green LED flashes about 8 seconds). If successful, the green LED lights steady for 3 seconds and the detector chirps indicating a successful enrollment. green LED lights steady for 3 seconds and the detector chirps indicating a successful enrollment. If the LED does not light steady, enrollment failed. Retry by pressing Test switch for less than 3 seconds.Supervision: Detector supervision is factory set for 1 minute and is not programmable. If the LED does not light steady, enrollment failed. Retry by pressing Test switch for less than 3 seconds.Supervision: Detector supervision is factory set for 1 minute and is not programmable. 5. To assign zone descriptors for each of the zones and/or change device type or zone response type, select the zone then select Edit and make the desired changes. Press Save when done. 5. To assign zone descriptors for each of the zones and/or change device type or zone response type, select the zone then select Edit and make the desired changes. Press Save when done. 24-Hour Enrollment Deletion Use this procedure within 24 hours of initial enrollment to delete this device from the controller. This allows reenrollment in a specific controller in case enrollment in that controller failed (i.e., the device inadvertently enrolled in a different controller). This resets the device to factory default settings and removes the communication link data in the controller in which it is enrolled, allowing the device to be reenrolled in the desired controller. 24-Hour Enrollment Deletion Use this procedure within 24 hours of initial enrollment to delete this device from the controller. This allows reenrollment in a specific controller in case enrollment in that controller failed (i.e., the device inadvertently enrolled in a different controller). This resets the device to factory default settings and removes the communication link data in the controller in which it is enrolled, allowing the device to be reenrolled in the desired controller. NOTE: Pressing and holding the test button (while powering the detector) will arm the detector to perform a reset to factory defaults. NOTE: Pressing and holding test button (while powering the detector) will arm the detector to perform a reset to factory defaults. TO SILENCE THE DETECTOR Press the test switch. The detector sounds again after 5 minutes if the alarm condition continues. POUR FAIRE TAIRE LE DTECTEUR Appuyez sur le bouton de test. Le dtecteur sonne de nouveau aprs 5 minutes si la condition d'alarme est toujours prsente. NOTE: The detector can only be silenced once per event. NOTE: The detector can only be silenced once per event. CEILING / PLAFOND WALL / MUR MOUNTING If the device is not in its final mounting location when enrolled, conduct Go/No Go tests (see controllers instructions) to verify adequate signal strength. Adjust the device location and orientation as necessary. FUNCTIONAL TEST Product must be tested at least once each year TEST FONCTIONNEL Le fonctionnement du produit doit tre vrifi au moins une fois par anne PERIODIC CLEANING NOTE Notify the proper authorities that the CO detector is undergoing maintenance and the system will temporarily be out of service. LED Indication & Sounder during Test and Trouble /
Confirmation par DEL et sonore durant le test et la maintenance Status LED (Top) / Estado del LED Blink Red once every sec /
Clignotement rouge une fois par seconde Blink Green once a sec /
clignotement en vert par seconde Sounder / Sonador Temp-4 Tempo. -4 Silent /
Silencieux Blink Red every 10 secs /
clignotement en rouge toutes les 10 secs Temp-4 Tempo. -4 MODE/ MODO CO Test 1ab / Test CO RealTest Functional CO Test1ab Waiting for gas entry / Test fonctionnel de CO RealTest 2ab En attente d'une entre de gaz RealTest Functional CO Test -
Successful gas entry /
Test fonctionnel de CO RealTest 2ab Entre de gaz effectue avec succs Low Battery / Pile faible CO Trouble/
Dfectuosit du dt. de CO Blink Amber once every 45 sec /
Clignotement ambre toutes les 45 secs Chirp every 45 seconds after 7 days / met des piements toutes les 45 secs aprs 7 jours Silent /
Silencieux Double Blink Amber every 5 secs / Clignotement ambre toutes les 3 secs Double Blink Amber every 3 secs / Clignotement ambre toutes les 3 secs Double Blink Amber every 3 secs / Clignotement ambre toutes les 3 secs CO End of Life2- First 29 days /
Fin de vie du dt. de CO4-
Premier 29 jours CO End of Life2- After 30 days /
Chirp every 45 secs/
Fin de vie du dt. de CO4- aprs met des piements 30 jours toutes les 45 secs 1a Test activated by pressing and releasing the test switch; Detector is functioning Silent /
Silencieux properly (within proper sensitivity). See Functional Test section for Functional Test activation and procedure. / Test activ par le bouton de test; Le dtecteur fonctionne adquatement (avec la sensibilit approprie). If Test mode is activated and the LED and Sounder do not function, check for maintenance or trouble conditions. / Si le mode de test est activ et que la DEL et l'avertisseur sonore ne fonctionnement pas, vrifiez s'il y a des conditions de maintenance ou de dfectuosit. Starts chirping after 30 days, continues until the batteries are replaced or die / Le dtecteur commence mettre des ppiements 30 jours aprs sur une base continue jusqu' ce que la pile soit remplace ou morte. 1b 2 instructions du contrleur. NOTE: For programming, see the Controllers instructions. NOTE: Pour la programmation des alarmes de fume ou de CO et Tamper, consultez les Approval Listings / Approbations Homologations:
FCC/IC Listed to UL 2075, and CSA 6.19 MAR2001 (R2016) / Conforme aux normes UL 2075, ou CSA 6.19 MAR2001 (R2016). Other Standards / Autres normes: RoHS FEDERAL COMMUNICATIONS COMMISSION & INDUSTRY CANADA STATEMENTS The user shall not make any changes or modifications to the equipment unless authorized by the Installation Instructions or User's Manual. Unauthorized changes or modifications could void the user's authority to operate the equipment. FCC / IC STATEMENT This device complies with Part 15 of the FCC Rules, and Industry Canadas license-exempt RSSs. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation. Cet appareil est conforme la partie 15 des rgles de la FCC et exempt de licence RSS dIndustrie Canada. Son fonctionnement est soumis aux conditions suivantes: (1) Cet appareil ne doit pas causer dinterfrences nuisibles. (2) Cet appareil doit accepter toute interfrence reue y compris les interfrences causant une rception indsirable. Agile Reference #800-23974 1/18 Rev A I56-0620-000 MONTAGE Si lappareil nest pas dans son emplacement de montage final lorsquil est enrl, effectuez des tests Tout ou rien (voir les instructions du panneau de commande) pour vrifier que la puissance dusignal est adquate. Au besoin, modifiez l'emplacement ou l'orientation du dispositif. REPLACE BATTERIES /
REMPLACEMENT DES PILES NETTOYAGE PRIODIQUE NOTE :
Avisez les personnes comptentes qu'un dtecteur de CO est hors service pour cause d'entretien et que, par consquent, le systme sera temporairement hors service. local.
(-10-70C) SPECIFICATIONS / SPCIFICATIONS:
Battery / Pile:
2 x 3V, CR123A Lithium; Panasonic CR123A or DURACELL DL123, DL123A Sensitivity / Sensibilit :
Meets UL2034/UL2075 CO sensitivity requirements Audible Signal: 85dBA / Signal audible: 85 dBA Dimensions: Thick 1.65 in (4.19 cm); Diameter 5.30in (13.462cm) Weight: 7.7 oz; 218.3 g / Poids : 7,7 oz; 218.3 g Tamper / Interrupteurs antisabotage: Wall Tamper/Wall Tamper Operating Temperature / Temprature de fonctionnement : 32 122 F / (0 50 C) Storage Temperature / Plage de temprature de stockage : 14-158F Relative Humidity / Humidit de fonctionnement :
20-95% RH (Agency Compliance 93% max.), non-condensing NOTE: Batteries available at your local retailer or distribution center. NOTE : Batteries disponibles chez votre revendeur ou votre centre de distribution Caution!
The batteries used in this device may present a fire or chemical burn hazard if mistreated. Do not recharge, disassemble, heat above 212F (100C) or dispose of in fire. Use two Panasonic CR123A or DURACELL DL123, DL 123A Lithium batteries. Use of other batteries may present a risk of fire or explosion. Keep used batteries away from children. Dispose of used batteries properly. Remove old batteries. Wait 10 seconds and then replace with two new batteries. To avoid a low battery indication when installing new batteries, both batteries must be installed within 15 seconds of installing the first one. Any low battery condition that may have occurred should clear when the back plate is installed. Mise en garde!
Les piles utilises dans ce dispositif peuvent prsenter des risques d'incendie ou de brlure chimique si elles sont mal traites. Ne rechargez pas, ne dsassemblez pas, ne faites pas chauffer et ne jetez pas au feu les piles plus de 100 C (212 F). Utilisez des dos piles au lithium Panasonic CR123A ou DURACELL DL123, DL 123A. L'utilisation d'autres types de piles peut prsenter des risques de feu ou d'explosion. Gardez les vieilles piles hors de la porte des enfants. Dbarrassez-vous des batteries utilises de faon approprie. Retirez les vieilles piles. Attendez 10 secondes et remplacer-les par dos nouvelles piles. Pour viter une indication de piles faibles lorsque vous installez de nouvelles piles, les dos
(2) piles doivent tre installes en de de 15 secondes aprs l'installation de la premire. Toute condition de pilles faibles qui aurait pu se produire devrait tre rtablie une fois la plaque arrire installe. REFER TO THE INSTALLATION INSTRUCTIONS FOR THE CONTROL WITH WHICH THIS DEVICE IS USED, FOR DETAILS REGARDING LIMITATIONS OF THE ENTIRE ALARM SYSTEM. POUR LES LIMITES DU SYSTME DALARME AU COMPLET, REPORTEZ-VOUS AU GUIDE DINSTALLATION DU PANNEAU DE COMMANDE. Support / Pour de lassistance en ligne, visitez : https://mywebtech.honeywell.com/
U.S. warranty / U.S. garantie: www.honeywell.com/security/hsc/resources/wa For patent info, see: / Pour des informations sur les brevets, voir :
www.honeywell.com/patents MyWebTech Warranty / Garantie Patents / brevets 2018 Honeywell International Inc. Honeywell and is a registered trademark of Honeywell International Inc. All other trademarks are the properties of their respective owners. All rights reserved.
1 | User Manual 20180228 v1 - SiXCO EXHIBIT 9-2 Manual | Users Manual | 1.23 MiB | May 03 2018 / January 09 2018 |
SiXCO Wireless Carbon Monoxide (CO) Detector/ Wireless Carbon Monoxide (CO) Detector Installation Instructions This device is intended for use with Honeywell control panels that support SiX series devices. Before installing detectors, please thoroughly read these installation instructions. Instructions dinstallation This device is intended for use with Honeywell control panels that support SiX series devices. Before installing detectors, please thoroughly read these installation instructions. Features 10-year life CO Detector End-of-Life reporting (detector needs replacing) Carbon Monoxide detector maintenance reporting Low Battery Detection Multi-color status LED Features 10-year life CO Detector End-of-Life reporting Carbon Monoxide detector maintenance reporting Low Battery Detection Multi-color status LED POWER UP Remove battery pull tab. Be sure batteries are seated properly to avoid a low battery condition after 15 seconds. Upon power up, Green LED Blinks every 2 secs / Sounder is Silent. ENROLLMENT This device can be enrolled and programmed either before or after installation. 1. Set the Lyric Controller in Programming Mode and go to Zone Programming. Select Add New and then Serial Number. 2. Remove the battery tabs to activate the device and begin the enrollment process. 3. The Green LED flashes rapidly during enrollment (up to about 20 seconds).
*NOTE: Enrollment time varies depending on the signal strength between the device and the controller. The detector sends its unique MAC ID (Serial Number) and Services information to the controller. The controller registers the device and displays the transmitter data on screen. 4. Enrollment is confirmed when the Green LED is ON for 3 seconds and the detector chirps. 5. Press SAVE. POWER UP Remove battery pull tab. Be sure batteries are seated properly to avoid a low battery condition after 15 seconds. Upon power up, Green LED Blinks every 2 secs /
Sounder is Silent. ENRLEMENT Le dispositif peut tre enrl et programm avant ou aprs l'installation. 1. Rglez le contrleur Lyric en mode de programmation et accdez la programmation des zones. Slectionnez Add New (ajouter nouveau) et ensuite Serial Number (numro de srie). 2. Retirez la languette qui recouvre la pile pour activer le dispositif et commencer la procdure d'enrlement. 3. La DEL verte clignote durant l'enrlement
(pendant jusqu' environ 20 secondes*). NOTE:
Le dispositif transmet au contrleur son ID La dure de l'enrlement varie selon la puissance du signal entre le dispositif et le contrleur. MAC (numro de srie) et l'information pour les services. Le contrleur enregistre le dispositif et affiche les donnes du transmetteur sur l'cran. 6. If the detector is not successfully enrolled during the enrollment period, the LED turns off and the device powers down. Pressing and holding the test button, while powering the detector, will arm the detector to perform a reset to factory defaults. The device can also be manually enrolled. See the Controllers Programming Guide for details. 4. L'enrlement est confirm lorsque la DEL verte est allume durant 3 secondes et que le dispositif met un ppiement . 5. Appuyez sur GARDER (Enregistrer). 6. Si le dispositif n'est pas enrl avec succs durant la priode d'enrlement, la DEL s'teint et le dispositif se met hors tension. Pressing and holding the test button, while powering the detector, will arm the detector to perform a reset to factory defaults. Le dispositif peut galement tre enrl manuellement. Pour les dtails, reportez-vous au Guide de programmation du contrleur. 24-Hour Enrollment Deletion and Default If the device is enrolled in a panel different than the intended panel, and you are unable to delete it from the unintended panel, reset default the device to factory default setting:
This procedure is available for 24 hours after enrollment with a panel and the device remains powered (battery installed). Remove power from the detector. Press the CO Test Button. While holding the test button, insert the batteries. Continue holding the test button for 10 seconds and release. The status LED should start blinking green rapidly to confirm its deleted enrollment. If the device is enrolled in a panel different than the intended panel, and you are unable to delete it from the unintended panel, reset default the device to factory default setting:
This procedure is available for 24 hours after enrollment with a panel and the device remains powered (battery installed). Remove power from the detector. Press the CO Test Button. While holding the test button, insert the batteries. Continue holding the test button for 10 seconds and release. The status LED should start blinking green rapidly to confirm its deleted enrollment. 2 MOUNTING After enrolling and before mounting permanently, conduct Go/No Go test (see controllers instructions) to verify adequate signal strength. Adjust the device location as necessary. 1. Remove the CO with a counter-clockwise motion. 2. Using two supplied screws and anchors, mount the MONTAGE After enrolling and before mounting permanently, conduct Go/No Go test (see controllers instructions) to verify adequate signal strength. Adjust the device location as necessary. 1. Remove the CO with a counter-clockwise motion. 2. Using two supplied screws and anchors, mount the 3. Attach the CO detector to the mounting base with a 3. Attach the CO detector to the mounting base with a 4. Test each detector as described in the Testing section. 5. Confirm all desired signals have been received by the 4. Test each detector as described in the Testing section. 5. Confirm all desired signals have been received by the base. clockwise motion. Central Station. base. clockwise motion. Central Station. NOTE:
NFPA 72 recommends the installation of detectors only after completing construction or any other dust producing activity. NOTE:
NFPA 72 recommends the installation of detectors only after completing construction or any other dust producing activity. Ceiling Mount / Ceiling Mount Wall Mount / Wall Mount 3 TESTING Test communications between the detector and the control panel. The detector has one test button (for testing CO). The detector may also be functionally tested using canned CO. If the detector fails the test method, the detector should be replaced. NOTE: Testing the detector will activate the alarm and Before testing, notify the proper authorities to avoid any false alarms. CO Test (Alarm Test) Press and hold the CO Test button for 1 to 2 seconds. The control panel should display and sound a CO alarm
(all programmed CO detector loops are sent). send a signal to the panel. TESTING Test communications between the detector and the control panel. The detector mode has one test button
(for testing CO) The detector may also be functionally tested using canned CO. If the detector fails the test method, the detector should be replaced. NOTE: Testing the detector will activate the alarm and Before testing, notify the proper authorities to avoid any false alarms. CO Test (Alarm Test) send a signal to the panel. Press and hold the CO Test button for 1 to 2 seconds. The control panel should display and sound a CO alarm (all programmed CO detector loops are sent). CO System Test (Functional CO Test) Press and hold the CO Test button for 3-5 seconds to enter the functional gas test mode. See Functional Gas Test section below. Functional Gas Test Solo C6 brand canned CO may be used to verify the detectors ability to sense CO by utilizing the RealTest feature as follows:
1. Remove cover by rotating counter clockwise. 2. Press and hold the CO test button for 3 to 5 seconds. The green LED will start blinking once per second indicating the detector is in RealTest mode. (If the detector will not go into RealTest mode, the CO sensor may be in fault or at end-of- life.) CO System Test (Functional CO Test) Press and hold the CO Test button for 3-5 seconds to enter the functional gas test mode. See Functional Gas Test section below. Functional Gas Test Solo C6 brand canned CO may be used to verify the detectors ability to sense CO by utilizing the RealTest feature as follows:
1. Remove cover by rotating counter clockwise. 2. Press and hold the CO test button for 3 to 5 seconds. The green LED will start blinking once per second indicating the detector is in RealTest mode. (If the detector will not go into RealTest mode, the CO sensor may be in fault or at end-of- life.) 3. While the green LED is blinking once per second, spray a small amount of canned CO directly into the CO gas entry port. 4. Upon successful gas entry and if functioning properly, the detector will go into CO alarm and send an alarm to the control panel. 3. While the green LED is blinking once per second, spray a small amount of canned CO directly into the CO gas entry port. 4. Upon successful gas entry and if functioning properly, the detector will go into CO alarm and send an alarm to the control panel. 5. The CO test will automatically clear when the CO clears 5. The CO test will automatically clear when the CO from the sensor or in 30 seconds if no CO was introduced. 6. Reattach the cover to the detector. clears from the sensor or in 30 seconds if no CO was introduced. 6. Reattach the cover to the detector. 4 HUSH FEATURE / ALARM SILENCE If required, the audible alarm for CO conditions can be silenced for 5 minutes by pushing the Test/Hush button. In addition, low chirping can be silenced for 12 hours when the Test/Hush button is pressed. During a CO alarm, if carbon monoxide is still present after the 5-minute hush period, the alarm will sound. CO SENSOR END-of-LIFE FEATURE When the CO sensor has passed end-of-life, a trouble signal is sent to the controller (if programmed). This indicates that the CO sensor inside the detector must be replaced. If unresolved for 30 days, the detector will chirp every 45 seconds. The typical life of the CO sensor is ten years from the date of manufacture. It is recommended to periodically check the Replace by date located on the label on the back of the detector head. HUSH FEATURE / ALARM SILENCE If required, the audible alarm for CO conditions can be silenced for 5 minutes by pushing the Test/Hush button. In addition, low chirping can be silenced for 12 hours when the Test/Hush button is pressed. During a CO alarm, if carbon monoxide is still present after the 5-minute hush period, the alarm will sound. CO SENSOR END-of-LIFE FEATURE When the CO sensor has passed end-of-life, a trouble signal is sent to the controller (if programmed). This indicates that the CO sensor inside the detector must be replaced. If unresolved for 30 days, the detector will chirp every 45 seconds. The typical life of the CO sensor is ten years from the date of manufacture. It is recommended to periodically check the Replace by date located on the label on the back of the detector head. Carbon Monoxide Detector: Events and Their ID Codes Event CO Alarm / CO Alarm CO Test / CO Test Low Battery / Low Battery Detector Supervision /
Detector Supervision Detector End-of-Life/Trouble /
Detector End-of-Life/Trouble Alpha Keypad CO Alarm CO Alarm Lo Bat CO Trouble CO Trouble CS Report CO Alarm (CID 162) / CO Alarm (CID 162) CO alarm (CID 162) / CO Alarm (CID 162) RF low-battery (CID 384) / RF Low-Battery (CID 384) RF sensor supervision (CID 381) / RF sensor supervision (CID 381) Sensor Trouble - End-of Life (CID 380) / Sensor Trouble
- End-of Life (CID 380) Tamper / Tamper Disarmed .......... CO Trouble Armed ................. CO Alarm RF Sensor Tamper (CID 383) / RF Sensor Tamper TESTING SIGNAL STRENGTH Perform this test in accordance with NFPA 72 inspection, testing and maintenance requirements to determine a strong communication path with the control panel. 1. Activate the wireless systems GO/NO GO TEST mode. 2. Press the detectors TEST button () for 1-2 seconds. The detector should immediately transmit an alarm signal to the control panel. The built-in horn will start to sound about 2.5 seconds after pressing the button. 3. The wireless systems keypad should emit at least three beeps when the alarm transmission is received and display the transmitting detectors zone number. 4. When the console has received the test signal, the horn will stop and a few seconds later the detectors zone number will clear from the console display. 5. If the console does not respond as noted, and if this is an initial installation, try moving the detector to another location that provides proper reception. Also, be sure that the detector has been enrolled by the controller (see Enrollment section). Then, repeat the test. 6. Turn off the systems TEST mode (typically security code + OFF). TESTING SIGNAL STRENGTH Perform this test in accordance with NFPA 72 inspection, testing and maintenance requirements to determine a strong communication path with the control panel. 1. Activate the wireless systems GO/NO GO TEST mode. 2. Press the detectors TEST button () for 1-2 seconds. The detector should immediately transmit an alarm signal to the control panel. The built-in horn will start to sound about 2.5 seconds after pressing the button. 3. The wireless systems keypad should emit at least three beeps when the alarm transmission is received and display the transmitting detectors zone number. 4. When the console has received the test signal, the horn will stop and a few seconds later the detectors zone number will clear from the console display. 5. If the console does not respond as noted, and if this is an initial installation, try moving the detector to another location that provides proper reception. Also, be sure that the detector has been enrolled by the controller (see Enrollment section). Then, repeat the test. 6. Turn off the systems TEST mode (typically security code + OFF). 5 REPLACING THE BATTERIES Remove old batteries. Wait 10 seconds and then replace with two new batteries. To avoid a low battery indication when installing new batteries, both batteries must be installed within 15 seconds of installing the first one. Any low battery condition that may have occurred should clear when the base plate is installed. REPLACING THE BATTERIES Remove old batteries. Wait 10 seconds and then replace with two new batteries. To avoid a low battery indication when installing new batteries, both batteries must be installed within 15 seconds of installing the first one. Any low battery condition that may have occurred should clear when the base plate is installed. CAUTION!
The batteries used in this device may present a fire or chemical burn hazard if mistreated. Do not recharge, disassemble, heat above 100C (212F) or dispose of in fire. Use only Panasonic CR123A OR DURACELL DL123, DL123A Lithium batteries. Use of other batteries may present a risk of fire or explosion. Keep used batteries away from children. Dispose of used batteries properly. CAUTION!
The batteries used in this device may present a fire or chemical burn hazard if mistreated. Do not recharge, disassemble, heat above 100C (212F) or dispose of in fire. Use only Panasonic CR123A OR DURACELL DL123, DL123A Lithium batteries. Use of other batteries may present a risk of fire or explosion. Keep used batteries away from children. Dispose of used batteries properly. BATTERY REPLACEMENT / BATTERY REPLACEMENT will be temporarily out of service. CLEANING NOTE: Notify the proper authorities when the system IMPORTANT:
This detector must be tested and maintained regularly following NFPA-72 requirements. The detector should be cleaned at least once a year. 1. Remove the detector from the base plate by turning counterclockwise. 2. Clean the outside casing with a cloth. Ensure that the holes on the front of the detector are not blocked with dirt and dust. Canned air can be used to remove any dust or debris. 3. Reattach the detector to the base plate by rotating clockwise. 4. Test the detector to insure it is fully functional.
(See Testing section). system will be temporarily out of service. CLEANING NOTE: Notify the proper authorities when the IMPORTANT:
This detector must be tested and maintained regularly following NFPA-72 requirements. The detector should be cleaned at least once a year. 1. Remove the detector from the base plate by turning counterclockwise. Clean the outside casing with a cloth. Ensure that the holes on the front of the detector are not blocked with dirt and dust. Canned air can be used to remove any dust or debris. 2. Reattach the detector to the base plate by rotating clockwise. 3. Test the detector to insure it is fully functional. (See Testing section). Notify the proper authorities and Central Station when the system is back in service. Notify the proper authorities and Central Station when the system is back in service. 6 MAINTENANCE Do not paint, and do not use cleaning agents, bleach or polish the detector. NOTE: Before performing any maintenance on the MAINTENANCE Do not paint, and do not use cleaning agents, bleach or polish the detector. NOTE: Before performing any maintenance on the detector, notify the proper authorities and Central Station that maintenance is being performed and the system will be temporarily out of service. Disable the zone or system undergoing maintenance to prevent any unwanted alarms. Power must be removed from the detector before performing maintenance of any kind. The SiXCO detector reports maintenance issues to the control panel and communicates them visually and audibly per the table below. detector, notify the proper authorities and Central Station that maintenance is being performed and the system will be temporarily out of service. Disable the zone or system undergoing maintenance to prevent any unwanted alarms. Power must be removed from the detector before performing maintenance of any kind. The SiXCO detector reports maintenance issues to the control panel and communicates them visually and audibly per the table below. TROUBLE FEATURE When the sensor (supervision) is in a trouble condition
(such as CO sensor end-of-life), the detector will send a trouble signal to the control panel. Depending on the issue, the detector must then be serviced or replaced. NOTE: CO detectors are not to be used with detector guards unless the combination is evaluated and found suitable for that purpose. TROUBLE FEATURE When the sensor (supervision) is in a trouble condition
(such as CO sensor end-of-life), the detector will send a trouble signal to the control panel. Depending on the issue, the detector must then be serviced or replaced. NOTE: CO detectors are not to be used with detector guards unless the combination is evaluated and found suitable for that purpose. 7 LED INDICATORS / LED INDICATORS The SiXCO has a multi-color top LED:
Green/Green ...................... Supervisory indication; blinks during power on, reset, and during normal operation /
Supervisory indication; blinks during power on, reset, and during normal operation Amber / Amber ................. Signal maintenance and trouble events / Signal maintenance and trouble events Red ......................................... CO Alarm condition / CO alarm condition MODE RealTest Functional CO gas entry test Waiting for gas entry / RealTest Functional CO gas entry test Waiting for gas entry RealTest Functional CO gas entry test -Upon successful gas entry / RealTest Functional CO gas entry test -Upon successful gas entry Low Battery / Low Battery CO Trouble / CO Trouble CO End of Life First 29 days / CO End of Life First 29 days CO End of Life after 30 days / CO End of Life after 30 days Normal (Standby) / Normal (Standby) Status LED (Top) Blink Green once per second /
Blink Green once per second Sounder Silent / Silent Blink Red once every 10 secs / Blink Red once every 10 secs Modified Temp-4 / Modified Temp-4 Blink Amber every 45 secs / Blink Amber every 45 secs Double Blink Amber every 5 secs / Double Blink Amber every 5 secs Double Blink Amber every 3 secs / Double Blink Amber every 3 secs Double Blink Amber every 3 secs / Double Blink Amber every 3 secs Single Blink Green every 10 secs / Single Blink Green every 10 secs Chirp every 45 secs after 7 days / Chirp every 45 secs after 7 days Silent / Silent Silent / Silent Chirp every 45 secs / Chirp every 45 secs Silent / Silent Many cases of reported carbon monoxide poisoning indicate that while victims are aware that they do not feel well, they become so disoriented that they are unable to save themselves by either exiting the building or calling for assistance. Also, young children, elderly and pets may be the first to be affected. Many cases of reported carbon monoxide poisoning indicate that while victims are aware that they do not feel well, they become so disoriented that they are unable to save themselves by either exiting the building or calling for assistance. Also, young children, elderly and pets may be the first to be affected. SYMPTOMS OF CARBON MONOXIDE POISONING Carbon monoxide bonds to the hemoglobin in the blood and reduces the amount of oxygen being circulated in the body. The following symptoms are examples taken from NFPA 720; they represent approximate values for healthy adults. SYMPTOMS OF CARBON MONOXIDE POISONING Carbon monoxide bonds to the hemoglobin in the blood and reduces the amount of oxygen being circulated in the body. The following symptoms are examples taken from NFPA 720; they represent approximate values for healthy adults. 8 LIMITED LIFE OF CO SENSOR This detector is manufactured with a long-life electrochemical carbon monoxide sensor. Over time the sensor will lose sensitivity and will need to be replaced. The life span of the CO sensor is approximately ten years from the date of manufacture. Periodically check the detectors replacement date. Remove the detector head and refer to the replace by sticker placed on the underneath side of the detector. The sticker will indicate the date the detector should be replaced. Reminder: This detector is also equipped with a feature that will signal the panel once the CO sensor has passed the end of its useful life. If this occurs, it is time to replace the detector. What to do if the detector goes into CO alarm:
If the detector goes into CO alarm (4 beeps), immediately move to a spot where fresh air is available, preferably outdoors, where the air is safe and call your security service provider. Tell your provider the detector alarm status, and that you require professional assistance in ridding your home of the carbon monoxide. This detector is NOT:
A substitute for the proper servicing of fuel-burning appliances or the sweeping of chimneys. To be used on an intermittent basis or as a portable alarm for the spillage of combustion products from fuel-burning appliances or chimneys. Carbon monoxide gas is a highly poisonous gas which is released when fuels are burnt. It is invisible, has no smell and is therefore is impossible to detect with the human senses. Under normal conditions in a room where fuel burning appliances are well maintained and correctly ventilated, the amount of carbon monoxide released into the room by appliances should not be dangerous. LIMITED LIFE OF CO SENSOR This detector is manufactured with a long-life electrochemical carbon monoxide sensor. Over time the sensor will lose sensitivity and will need to be replaced. The life span of the CO sensor is approximately ten years from the date of manufacture. Periodically check the detectors replacement date. Remove the detector head and refer to the replace by sticker placed on the underneath side of the detector. The sticker will indicate the date the detector should be replaced. Reminder: This detector is also equipped with a feature that will signal the panel once the CO sensor has passed the end of its useful life. If this occurs, it is time to replace the detector. What to do if the detector goes into CO alarm:
If the detector goes into CO alarm (4 beeps), immediately move to a spot where fresh air is available, preferably outdoors, where the air is safe and call your security service provider. Tell your provider the detector alarm status, and that you require professional assistance in ridding your home of the carbon monoxide. This detector is NOT:
A substitute for the proper servicing of fuel-burning appliances or the sweeping of chimneys. To be used on an intermittent basis or as a portable alarm for the spillage of combustion products from fuel-burning appliances or chimneys. Carbon monoxide gas is a highly poisonous gas which is released when fuels are burnt. It is invisible, has no smell and is therefore is impossible to detect with the human senses. Under normal conditions in a room where fuel burning appliances are well maintained and correctly ventilated, the amount of carbon monoxide released into the room by appliances should not be dangerous. Concentration (ppm CO) /
Concentration (ppm CO) 200 400 800 Symptoms / Symptoms Mild Headache after 2-3 hours of exposure / Mild Headache after 2-3 hours of exposure Headache and nausea after 1-2 hours of exposure / Headache and nausea after 1-2 hours of exposure Headache, nausea, and dizziness after 45 minutes of expo- sure; collapse and unconsciousness after 2 hours of exposure / Headache, nausea, and dizziness after 45 minutes of expo- sure; collapse and unconsciousness after2 hours of exposure 9 CO ALARM ACTIVATION Per UL standard 2075, the SiXCO detector has been tested to the sensitivity limits defined in UL standard 2034. CO ALARM ACTIVATION Per UL standard 2075, the SiXCO detector has been tested to the sensitivity limits defined in UL standard 2034. CO Alarm Thresholds / CO Alarm Thresholds No alarm within 30 days / No alarm within 30 days 60 - 240 10 - 50 4 - 15 Parts per Million / Parts per Million Detector Response Time (Minutes) / Detector Response Time (Minutes) 30 +/- 3ppm 70 +/- 5ppm 150 +/- 5ppm 400 +/- 10ppm SPECIFICATIONS / SPCIFICATIONS Electrical Specifications Voltage / Voltage .......................................................................................................... 3 volts DC Battery Type / Pile Type ............................................................................................ CR123A OR DURACELL DL123, DL123A lithium Battery Manufacturer / Pile Manufacturer ..................................................... Panasonic, DURACELL Number of Batteries / Number of Batteries ................................................... 2 Sensitivity / Sensitivity .............................................................................................. Meets UL2034/UL2075 CO sensitivity requirements Audible Signal / Audible Signal ............................................................................. 85dBA Physical Specifications Diameter / Diameter .................................................................................................. 5.3 inches (13.46cm) Diameter; 1.65 inches (4.19cm) Thick Weight / Weight ............................................................................................................ 7.7oz. (21g) Operating Temperature / Temprature de Fonctionnement ........... 32 122 F / 0 50 C Storage Temperature / Storage Temperature .............................................. -10 - 70 C (14 - 158 F) Operating Humidity / Operating Humidity ..................................................... 20-95% RH Tamper/Tamper ........................................................................................................... Wall Tamper/Wall Tamper REFER TO THE INSTALLATION INSTRUCTIONS FOR THE CONTROL WITH WHICH THIS DEVICE IS USED, FOR DETAILS REGARDING LIMITATIONS OF THE ENTIRE ALARM SYSTEM. POUR LES LIMITES DU SYSTME DALARME AU COMPLET, REPORTEZ-VOUS AU GUIDE DINSTALLATION DU PANNEAU DE COMMANDE. 10 RF EXPOSURE Warning The antenna(s) used for this device must be installed to provide a separation distance of at least 7.8 inches (20 cm) from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter except in accordance with FCC multi-transmitter product procedures. Mise en Garde Exposition aux Frquences Radio: La/les antenne(s) utilise(s) pour cet metteur doit/doivent tre installe(s) une distance de sparation d'au moins 20 cm (7,8 pouces) de toute personne et ne pas tre situe(s) ni fonctionner paralllement tout autre transmetteur ou antenne, except en conformit avec les procdures de produit multi transmetteur FCC et ISED. FEDERAL COMMUNICATIONS COMMISSION (FCC) & INDUSTRY CANADA (IC) STATEMENTS The user shall not make any changes or modifications to the equipment unless authorized by the Installation Instructions or User's Manual. Unauthorized changes or modifications could void the user's authority to operate the equipment. FCC / IC STATEMENT This device complies with Part 15 of the FCC Rules, and Industry Canadas license-exempt RSSs. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation. Cet appareil est conforme la partie 15 des rgles de la FCC et exempt de licence RSS dIndustrie Canada. Son fonctionnement est soumis aux conditions suivantes: (1) Cet appareil ne doit pas causer dinterfrences nuisibles. (2) Cet appareil doit accepter toute interfrence reue y compris les interfrences causant une rception indsirable. Approval Listings / Approbations homologations FCC/ICC ETL Listed to UL2075 and CSA 6.19 MAR2001 (R2016) Other Standards RoHS 11 SUPPORT, WARRANTY, & PATENT INFORMATION For online support information, please go to:
https://mywebtech.honeywell.com/
Pour de lassistance en ligne, visitez :
https://mywebtech.honeywell.com/
For the latest warranty information, go to:
www.honeywell.com/security/hsc/resources/wa Pour les dernires informations de garantie, s'il vous plat aller :
www.honeywell.com/security/hsc/resources/wa For patent information, see www.honeywell.com/patents Pour des informations sur les brevets, voir www.honeywell.com/patents MyWebTech Warranty Patents 800-23974l 800-23974 1/18 Rev A 2017 Honeywell International Inc. Honeywell and is a registered trademark of Honeywell International Inc. All other trademarks are the properties of their respective owners. All rights reserved. Honeywell est une marque dpose de Honeywell International Inc. Toutes les autres marques de commerce appartiennent leurs propritaires respectifs. Tous droits rservs.
frequency | equipment class | purpose | ||
---|---|---|---|---|
1 | 2018-03-05 | 2405 ~ 2475 | DTS - Digital Transmission System | Original Equipment |
app s | Applicant Information | |||||
---|---|---|---|---|---|---|
1 | Effective |
2018-03-05
|
||||
1 | Applicant's complete, legal business name |
Ademco Inc.
|
||||
1 | FCC Registration Number (FRN) |
0004071130
|
||||
1 | Physical Address |
2 Corporate Center Drive
|
||||
1 |
9040
|
|||||
1 |
Melville, New York 11747
|
|||||
1 |
United States
|
|||||
app s | TCB Information | |||||
1 | TCB Application Email Address |
c******@ccsemc.com
|
||||
1 | TCB Scope |
A4: UNII devices & low power transmitters using spread spectrum techniques
|
||||
app s | FCC ID | |||||
1 | Grantee Code |
CFS
|
||||
1 | Equipment Product Code |
8DL6CO
|
||||
app s | Person at the applicant's address to receive grant or for contact | |||||
1 | Name |
M**** A********
|
||||
1 | Title |
RF Compliance Engineer
|
||||
1 | Telephone Number |
516-5********
|
||||
1 | Fax Number |
516-5********
|
||||
1 |
m******@honeywell.com
|
|||||
app s | Technical Contact | |||||
1 | Firm Name |
Honeywell International Inc
|
||||
1 | Name |
M******** A********
|
||||
1 | Physical Address |
2 Corporate Center Dr
|
||||
1 |
Melville, New York 11747
|
|||||
1 |
United States
|
|||||
1 | Telephone Number |
516-5********
|
||||
1 |
M******@honeywell.com
|
|||||
app s | Non Technical Contact | |||||
1 | Firm Name |
Honeywell International Inc
|
||||
1 | Name |
M****** A********
|
||||
1 | Physical Address |
2 Corporate Center Dr
|
||||
1 |
Melville, New York 11747
|
|||||
1 |
United States
|
|||||
1 | Telephone Number |
516-5********
|
||||
1 |
M******@honeywell.com
|
|||||
app s | Confidentiality (long or short term) | |||||
1 | Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | Yes | ||||
1 | Long-Term Confidentiality Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | Yes | ||||
1 | If so, specify the short-term confidentiality release date (MM/DD/YYYY format) | 09/01/2018 | ||||
if no date is supplied, the release date will be set to 45 calendar days past the date of grant. | ||||||
app s | Cognitive Radio & Software Defined Radio, Class, etc | |||||
1 | Is this application for software defined/cognitive radio authorization? | No | ||||
1 | Equipment Class | DTS - Digital Transmission System | ||||
1 | Description of product as it is marketed: (NOTE: This text will appear below the equipment class on the grant) | Carbon Monoxide Detector | ||||
1 | Related OET KnowledgeDataBase Inquiry: Is there a KDB inquiry associated with this application? | No | ||||
1 | Modular Equipment Type | Does not apply | ||||
1 | Purpose / Application is for | Original Equipment | ||||
1 | Composite Equipment: Is the equipment in this application a composite device subject to an additional equipment authorization? | No | ||||
1 | Related Equipment: Is the equipment in this application part of a system that operates with, or is marketed with, another device that requires an equipment authorization? | No | ||||
1 | Grant Comments | Power Output listed is conducted. The antenna(s) used for this transmitter must not be co-located with any other transmitters except in accordance with FCC multi-transmitter product procedures. End-users and installers must be provided with transmitter operating conditions for satisfying RF exposure compliance. | ||||
1 | Is there an equipment authorization waiver associated with this application? | No | ||||
1 | If there is an equipment authorization waiver associated with this application, has the associated waiver been approved and all information uploaded? | No | ||||
app s | Test Firm Name and Contact Information | |||||
1 | Firm Name |
Ademco Inc.
|
||||
1 | Name |
M******** A******
|
||||
1 | Telephone Number |
516-5********
|
||||
1 | Fax Number |
000-0********
|
||||
1 |
m******@resideo.com
|
|||||
Equipment Specifications | |||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Line | Rule Parts | Grant Notes | Lower Frequency | Upper Frequency | Power Output | Tolerance | Emission Designator | Microprocessor Number | |||||||||||||||||||||||||||||||||
1 | 1 | 15C | 2405.00000000 | 2475.00000000 | 0.0700000 |
some individual PII (Personally Identifiable Information) available on the public forms may be redacted, original source may include additional details
This product uses the FCC Data API but is not endorsed or certified by the FCC