Size:480*297mm Fingertip Pulse oximeter XM-103 Document No.: JDXM-0304-052 Version: Z General Description CAUTION:
Federal (U.S.) Law restricts this device to sale by or on the order of a physician. Oxygen binds to hemoglobin in red blood cells when moving through the lungs. It is transported throughout the body as arterial blood. A pulse oximeter uses two frequencies of light (red and infrared) to determine the percentage(%)of hemoglobin in the blood that is saturated with oxygen. The percentage is called blood oxygen saturation, or SpO2. A pulse oximeter also measures and displays the pulse rate at the same time it measures the SpO2 level. The oximeter is for prescription use or prescription home use. Safety Notice 1. Before use,carefully read the manual. 2. Do not use the pulse oximeter:
-if you are allergic to rubber products.
-if the device or finger is damp.
-during MRI or CT scan.
-while taking a blood pressure measurement on the arm.
-nail polish, dirty, coating fingers and false nails applied fingers.
-fingers with anatomical changes, edemas, scars or burns.
-Too big finger: the width of finger is over than 20mm and the thickness is over than 15mm.
-Too small finger: the width of finger is less than 10mm and the thickness is less than 5mm.
-Minors under 18 years old.
-The environmental light changes strongly.
-near flammable or explosive gas mixtures. 3. Extended use may cause pain for people with circulatory disorders. Do not use the pulse oximeter for longer than two hours on one finger. 4. Measurements are for your information only - they are no substitute for a medical examination. 5. Check the pulse oximeter regularly before use to ensure that there is no visible damage and the batteries are still sufficiently charged. In case of doubt, do not use the device and contact customer services or authorized retailer. 6. Do not use any additional parts that are not recommended by the manufacturer. 7. Any circumstances do not open or repair the device by yourself. Failure to comply will result in voiding of the warranty. For repairs, please contact customer services or authorized retailer. 8. Do not look directly inside the housing during the measurement. The red light and the invisible infrared light in the pulse oximeter are harmful to your eyes. 9. This device is not intended for use by people (including children) with restricted physical, sensory or mental skills or a lack of experience or a lack of knowledge, unless they are supervised by a person who has responsibility for their safety or they receive instructions from this person on how to use the device. Children should be supervised around the device to ensure they do not play with it. 10. If the unit has been stored at temperatures below 0 , leave it in a warm place for about two hours before using it. This device conforms to IEC6061-1, IEC60601-1-2, IEC60601-1-
11. If the unit has been stored at temperatures above 40 , leave it in a 11, ISO 80601-2-61, IEC 62304, 47 FCC part 15C, ANSI C63.27 . cool place for about two hours before using it. Measurement Principle PRINCIPLE of the oximeter is as follows: The pulse oximeter works by applying a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light 12. Neither of the displays for the pulse wave and pulse bar allows the strength of the pulse or circulation to be evaluated at the measurement site. Rather, they are exclusively used to display the current visual signal variation at the measurement site and do not enable diagnostics for the pulse. 13. Operation of the fingertip pulse oximeter may be affected by the use of source is 660nm, which is red light the other is 905nm, which is an electrosurgical unit (ESU). infrared-red light skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. Diagram of Operation Principle 14. Follow local ordinances and recycling instructions regarding disposal or recycling or the device and device components, including batteries. 15.This equipment complies with IEC 60601-1-2:2014 for electromagnetic compatibility for medical electrical equipment and systems. In healthcare center or other environment, their radio transmission equipment and electromagnetic interference may affect the performance of the oximeter. 16.The oximeter contains radio communication function, it may affect other electronic medical equipment, so it should not be used close to or stacked with other equipment. 1.Red and infrared-ray emission tube 2.Red and infrared-ray receipt tube 18. When the signal is not stable, the reading may inaccurate. Please do not reference. 17. This equipment is not intended for use during patient transport outside Display the healthcare facility. Features To Use Cleaning and Maintenance 1.Non-observance of the following instructions can lead to incorrect or failed measurements
-There must not be any nail polish, artificial nails or other cosmetics on the finger to be measured.
- Ensure that the finger nail on the finger to be measured is short enough that the fingertip covers the sensor element in the housing.
- Keep your hand, finger and body steady during the measurement.
- In cases of carbon monoxide poisoning, the pulse oximeter will display a measurement value that is too high.
- To avoid incorrect results, there should not be any strong light sources( e.g. fluorescent lamps or direct sunlight ) in the immediate vicinity of the pulse oximeter.
- Protect the pulse oximeter from dust, shocks, moisture, explosive materials.
- Excessive patient movement. 1.Simple to operate and convenient to carry. 2.Small volume, light weight and low power consumption. 3.Displays SpO2, PR, Pulse bar, and waveform. 4.Level 1-5 adjustable brightness. 5.5 display modes. 6.A low voltage warning will be indicated in visual window when battery voltage is so low that normal operation of the oximeter might be influenced. 7.When it shows Finger out, the pulse oximeter will power off automatically in 10 seconds. 8.Beep. 9.When the buzzer and reminder function are turned on, the numbers on the screen will flash when the reminder occurs, and the buzzer will beep. 10.Bluetooth function. 2.The following situations may cause inaccurate measurements
- Significant levels of dysfunctional hemoglobin (such as carbonyl -
Unit Operation hemoglobin or methemoglobin).
- Venous pulsations.
- Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter,or intravascular line.
- The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia.
- The patient is in cardiac arrest or is in shock.
- Weak pulse quality (low perfusion).
- Low hemoglobin. Battery Installation Slide battery cover off as indicate by arrow. Install 2 new AAA alkaline batteries according to polarity. Close battery Cover. CAUTION: Please make sure your finger size is appropriate
(fingertip width is about 10~20 mm, thickness is about 5~15 mm) CAUTION: This device cannot be used in strong radiation environment. It can only be used after binding. CAUTION: This device cannot be used with other medical devices or non-
medical devices. CAUTION:
When placing your fingers, ensure your fingers can completely cover the LED transparent window in the finger clamp compartment. 1. As shown in the figure, squeeze the clip of the pulse oximeter, fully insert your finger into the finger clip compartment, and then loosen the clip 2.Press the power button one time on front panel to turn the pulse oximeter on. 3.Keep your hands still for the reading. Do not shake your finger during the test. It is recommended that you do not move your body while taking a reading. 4. Read the data from the display screen. 5.To select your desired display brightness, press and hold the power button during operaion untill the brightness level changes. 6.To choose among the various display formats, press the power button briefly during operation. 7.If you remove the oximeter from your finger, it will shut off after about 10 seconds. 6)Do not try to recharge batteries not intended to be recharged; they password to login. c)Create a new user account on the APP or use existed user name and Problem cause Solution The fingertip pulse oximeter is a portable non-invasive, spot-check, Reversed batteries may cause damage to the device. oxygen saturation of arterial hemoglobin and pulse rate of adult at 2)Use only the size and type of batteries specified. Note:
1)Be sure to follow the correct polarity when installing the batteries. home, and hospital (including clinical use in internist/surgery, anesthesia etc). Unit Illustration 3)Do not mix different types of batteries together or old batteries with fresh ones. Always replace batteries as a simultaneous set. 4)Replace the batteries in a timely manner when low voltage lamp is lighted. 5)If the batteries in the device are depleted or the device will not be used for a long period of time, remove the batteries to damage or injury from possible battery leakage. Contents
- 1 x XM103 pulse oximeter
- 1 x Owners Manual
- 1 x Retaining strap
- 1 x Storage Bag
- 2 x 1.5v AAA batteries Monitor Unit Power Button Finger Opening can overheat and rupture. 7)Do not dispose of batteries in fire, batteries may explode or leak. 8)Keep batteries away from children and pets. Batteries may be harmful of swallowed. Should a child or pet swallow a battery, seek medical assistance immediately. 9)Please follow the law of the local government to deal with used batteries.. Slot for retaining strap 8.Using Bluetooth communication a)Download and install the"JOYTECH Healthcare" app from your smartphones app store. Recommended App Store: recommend
"Google store" for Android users, and recommend "App store" for IOS users. b)Open the App on your phone. If requested, you should enable Bluetooth on your phone. You can enable Bluetooth under the Settings menu on your smart phone. d) Attention First, turn on the oximeter and then click "not connected" on the APP. When the APP scans the Bluetooth of the oximeter, the icon and the name of the oximeter will be displayed on the APP. At this time, click "Pair this device" on the APP, when
"connected" is displayed on the APP, the connection is successful. Note: To realize the connection between the designated mobile phone and the designated oximeter, it is necessary to ensure that all oximeters except the designated oximeter are turned off. e)When your oximeter is connected successfully to your smart phone, The data transfer will begin automatically. The APP will display the received data from oximeter immediately without delay.Note: The monitor requires a smart device with: Android 5.0 or later, IOS9.0 or later. 9. APP Introduction As shown in the figure, this is an app icon. The APP interface can display blood oxygen value, pulse rate value, pulse waveform and historical data curve, historical data is sotred in the APP,When the user pulls out the finger from the oximeter, the APP will record the data once The "Home" page of the APP can display Last 3 historical dataand the "Curve" page of the APP can display a graph with all of historical data of any date in the past. The oimeter can only send the data to APP, mutual control is not supported. When the Bluetooth connection is succcessful between the oximeter and the APP , the APP cannot actively disconnect.The Bluetooth connection can only be disconnected when the oximeter is turned off or the Bluetooth of the mobile phone is turned off. Only when the original oximeter is disconnected, the APP can be paired with the new oximeter 1.Please use medical alcohol to clean the silicone touching the finger inside of oximeter with a soft cloth dampened with 70%
isopropyl alcohol. Also clean the being tested finger using alcohol before and after each test. 2.Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device allow the oximeter to dry thoroughly before reuse. 3.The fingertip pulse oximeter requires no routine calibration or maintenance other than replacement of batteries. 4.The use life of the device is 3 years when it is used for 10 measurements every day and 15 minutes per one measurement. Stop using and contact local service center if one of the following cases occurs:
An error in the Possible Problems and solutions is displayed on screen. The oximeter cannot be powered on in any case and not the reasons of battery. There is a crack on the oximeter or damage on the display resulting readings cannot be identified; the spring is invalid or the key is unresponsive or unavailable. 5.Cleaning and Disinfecting Cleaning procedures:
a)Place the oximeter on a clean table and wipe the entire surface and upper and lower finger pads for 3 times by a clean soft cloth dipped in 70% isopropanol;
b) Wait for 1 minute to fully dry the oximeter. Disinfection procedures:
a) Dip a clean soft cloth in 70% isopropanol, and clip it by the finger pads for at least 3 minute;
b) Take out the soft cloth, wait for 1 minute to fully dry the oximeter. CAUTION: Never use EtO or formaldehyde for disinfection. 6.The devise is provided as non-sterile, please clean and disinfect it according to the instructions before each use. Troubleshooting Guide Batteries are epleted d Replace the batteries Monitor do not display Batteries not inserted correctly Reinsert the batteries. If after reinserting the batteries correctly There are still no measurement values displayed, contact customer service Insufficient circulation in the measurement finger Observe the Important Guidelines Measurements are erratic Finger,hand or body is moving Keep your finger, hand and body still during the measurement Cardiac arrhythmia Seek medical attention Finger is not inserted correctly Retry by inserting the finger Measurements can not be shown normally Patients SpO2 value is too low to be measured There is excessive illumination; Or, Try some more times. If you can make sure no problem exist in the product, please go to a hospital timely for exact diagnosis. System Settings With power off, press the power button about 5 seconds to actuate system setting. Setting available for Tips, Beep, Language,Default, SpO2 Tips Lo PR Tips Hi,PR Tips Lo and EXIT. Long press to enter the specific value setting, short press to switch among the setting items.
, Settings Tips Tips Settings Tips Tips Tips The sensor with specific monitor has been validated and tested for compliance with EN ISO 60601-2-61:2011 and FDA Guidance--
Pulse Oximeters - Premarket Notification Submissions
[510(k)s]. In the clinical recruitment, the data is obtained from a controlled, induced hypoxia study in healthy adult volunteers. A total of 12 subjects including 6 females and 6 males were recruited from the healthy adult volunteers aged from 21 to 40 without smoker. In the clinical evaluation, 288 paired data of the 12 adults were validated for XM-103 Fig.1: Linear regression fit (X axis is SaO2, Y axis is the difference of SpO2-SaO2) Fig. 2: Linear regression fit (X axis is SaO2, Y axis is SpO2 for the subject device) a) When the Oximeter is placed on the patients finger or simulated finger, the SpO2 values and PR values can be displayed normally. b) Measurement accuracies:
* Clinical accuracy of SpO2 (Arms): in the range of 70%-100%, 2%;
* Clinical accuracy of pulse rate (Arms): in the range of 30~100bpm, 2bpm;
100~240bpm,2%. The oximeter might not be porperly placed within the smart devices tranmission range and is too far from the smart device If there are no causes of data transmission interference found near the oximeter, move the oximeter with 16ft.
(5m) of the smart device ang try again Connection failure/ Data is not being transmitted The oximeter did not pair successfully to the smart device Try to pair the devices once again The application on the smart device is not ready Check the application then try sending the data again FCC Information Note:
The functional tester cannot be used to assess the accuracy of the oximeter. The test methods used to establish the SpO2 accuracy is clinical testing. The oximeter used to measure the arterial haemoglobin oxygen saturation levels and these levels are to be compared to the levels Caution: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user authority to operate the equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
determined from arterial blood sampling with a CO-oximeter.
(1) this device may not cause harmful interference, and 1.ISO 80601-2-61, medical electrical equipment - part 2-61: particular requirements for the basic safety and essential performance of pulse oximeter equipment. 2.AAMI/ANSI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and a2:2010/(r)2012 (consolidated text) medical electrical equipment -- part 1: general requirements for basic safety and essential performance. 3.AAMI/ANSI/IEC 60601-1-2, medical electrical equipment -- part 1-2:
general requirements for basic safety and essential performance --
collateral standard: electromagnetic disturbances -- requirements and tests (General II (ES/EMC)). 4.IEC 60601-1-11, medical electrical equipment -- part 1-11: general requirements for basic safety and essential performance - collateral
(2) this device must accept any interference received, including interference that may cause undesired operation.
*Note:The device has been evaluated to meet general RF exposure requirement.The device can be used in portable exposure condition without restriction. This equipment has been tested and found to comply with the limits for a Class B digital device,pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radiofrequency energy. If this equipment does cause harmful interference to radio or television reception,which can be determined by turning the equipment off and on, the user is encouraged to try and correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna. standard: requirements for medical electrical equipment and medical
- Increase the distance between the equipment and the receiver. Model Display SpO2 Pulse Rate Power supply Power Consumption Weight XM-103 OLED display Display Range 0%~99%
electrical systems used in the home healthcare environment. Measurement Range 70%~100%
Correct disposal of this product. Accuracy 70%~100% 2%
0%~69% no definition
(Waste electrical & electronic equipment) This marking shown on the product indicates that it should not be Resolution 1%
disposed with other household waste at the end of its life. To prevent Display Range 0~240bpm Measurement Range 30~240bpm Accuracy Resolution 30~100bpm, 100~240bpm,2%
2bpm;
1bpm potential harm to the environment or to human health, please separate this product from other types of wastes and recycle it responsibly. When disposing this type of product, contact the retailer where product was purchased or contact your local government office for details regarding how this item can be disposed in an environmentally safe recycling center. Business users should contact their supplier and check the terms and 2x1.5vAAA batteries conditions of the purchasing agreement. This product should not be mixed
<60mA Approx.50g with other commercial wastes for disposal. This product is free of hazardous materials. Dimensions Approx.60mm*32mm*31.4mm Operating Environment Humidity 15%~93%RH Temperature 5 ~40 Icon Explanation
- Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help. Warranty The Fingertip Pulse oximeter is guaranteed for 2-year from the date of purchase. If the Fingertip Pulse oximeter does not function properly due to defective components or poor workmanship, we will repair or replace it freely. The warranty does not cover damages to your Fingertip Pulse oximeter due to improper handling. Please contact local retailer for details. Contact Information JOYTECH Healthcare Co., Ltd. No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100 Zhejiang, China Please contact us on:
Email: info@sejoy.com Telephone: +86-571-81957767 Fax: +86-571-81957750 Temperature
-20~55 Storage Environment Humidity 15%~93%RH Pressure 700hPa~1060hPa Ingress Protection Rating IP22 Classification Internal Powered Equipment Type BF Frequency range 2.4GHz(2400~2483.5MHz) Modulation GFSK BT Version 5.0 BT Signal mode Transmit output power 3dBm @ room temperature Bluetooth communication Rx sensitivity 97dBm @ 1Mbps mode Supply voltage 1.8V - 3.6V Current consumption 2uA @Sleep Mode 0.7uA @OFF Mode 5mA @TX mode(0dBm) 5mA @RX mode(0dBm) Probe LED Specifications Wavelength Power Consumption RED IR Approx. 660nm Approx.3.2mW Approx. 905nm Approx.2.4mW The Date UPDATE period Less than 12s Type BF applied part.
Attention Items Descriptions IP22 Protected against dripping water. SpO2%
Oxygen saturation PR bpm Pulse rate (BPM) Low power indication SN Serial No. Clinical SpO2 accuracy (Arms) (70-80%) Clinical SpO2 accuracy (Arms) (80-90%) Clinical SpO2 accuracy (Arms) (90-100%) Sterile Default settings Reuse 2%
2%
2%
Non-sterile None Disinfect for repeated use Alarm inhibit
(Note: This device does not provide any alarm function) Storage temperature and relative humidity Date of Manufacture 0123 European union approval Conformity to WEEE Directive Bluetooth Follow instruction for use Authorized representative in the European community Manufacturers information The Bluetooth Smart word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by JOYTECH Healthcare Co.,Ltd. 1.Instructions for the frequency of inspection of the application site for skin integrity Before each finger is inserted into the oximeter probe, the integrity of the skin should be visually checked to ensure that the skin is free from injury and other conditions. 2.Instructions for the frequency of sensor relocation There is no need to replace the blood oxygen sensor within the service life of the product. 3.Use during exercise and weak perfusion:
(1) DO NOT move your finger, arm and body during the measurement. Movement, including talking, coughing, or sneezing, during measurement, can affect the accuracy of the measurement results. MR unsafe Prescription use
(2) The reading should NOT be considered reliable and accurate in the condition of low perfusion during measurement. Pressure 700hPa~1060hPa Symbol Definition Symbol Definition Probe Accuracy of Oximeter Statement of Essential performance