SULFILOGGERUSERMANUAL
SulLoggerUserManual COPYRIGHT2019-UNISENSE A/S REV.2019-7-5, VALID FOR SERIAL NUMBERS FROMS/N:1000000
TABLEOFCONTENTS Warrantyandliability
. Warranty . Recyclingprogram . Generalinformation
. Safety . Support Support,ordering,andcontactinformation . Overview Maintenance Cleaning . Calibration . Accessories Listofaccessories . UsingtheSulLoggerApp . UsingthePowerComBox . Specications
. Operatingprinciplesofthesensor . Measuringindifferentmedia-unitconversions . SulLoggersensors . Knowninterferingsubstances . SulLoggersensorsewerinstallations-tipsandtricks Troubleshooting Complianceinformation
. 1 1 1 2 2 5 5 6 7 7 7 13 13 13 14 16 16 17 17 18 20 22 25 iii
FCCPart15,ClassA . Appendices A:Installationguide 25 26 26
WARRANTYANDLIABILITY This product is designed for industrial use. Not for use in hu-
mandiagnosticortherapeuticprocedures. Thisproductmustbe handledwithcareandonlybytrainedpersonnel. UnisenseA/S recommendsuserstoreadthedocumentationverycarefully. WARRANTY The SulLogger is covered by a limited warranty according to theproductdatasheet. Thewarrantydoesnotincluderepairorreplacementnecessitated byaccident,neglect,misuse,unauthorizedrepair,ormodication of the product. In no event will Unisense A/S be liable for any direct, indirect, consequential or incidental damages, including lostprots,orforanyclaimbyanythirdparty,arisingoutofthe use,theresultsofuse,ortheinabilitytousethisproduct. Unisense assumes no liability for damages on instruments de-
signedforindooruseduetounintendedelduseorexposureto dust,humidityorcorrosiveenvironments. RECYCLING PROGRAM Unisense a/s is committed to help our customers to recycle re-
sponsibly. Morethan80%oftheproductcanbereused,soweask youtorequestapick-uplabelbywritingtosullogger@unisense.com. 1
GENERALINFORMATION Thank you for choosing the Unisense SulLogger sensor - the onlyhydrogensulde(H2S)sensoronthemarketenabling:
DirectmeasurementofH2Sconcentrationsineitherliquid orgasphase Easyplugnplayinstallation Operationformeasuringcampaignsinharshenvironments suchassewernetworks Easycleaning,maintenanceandcalibration SAFETY Pleasereadthismanualcarefullybeforeunpacking,installingor operatingthisequipment. AlthoughtheoperationoftheSulLog-
gersensorisstraightforward,itismandatorytobefamiliarwith thecontentofthismanual. Failuretodosomayresultinserious injurytooperatorsorotherpersons,ordamagetoequipment. Do notuseorinstalltheSulLoggersensorandauxiliaryequipment inanyotherwaythanspeciedinthismanual. Specialattention mustbepaidtoalldangerandcautionstatements. DonotusetheSulLoggersensoroutsidethespeciedelectri-
cal,mechanicalandthermalparametersasspeciedinthedata sheet. During the installation please wear personal protective equipment(PPE)accordingtothesafetyregulationoftheinstalla-
tionsite. Hazardstatementnomenclature 2
Indicates a potentially or imminently hazardous situation which,ifnotavoided,couldresultindeathorseriousinjury.
DANGER
! CAUTION Indicatesapotentiallyhazardoussituationthatmayresultin minorormoderateinjury.
IMPORTANTNOTE Informationthatrequiresspecialemphasis. Hazardousareas-SulLoggersensorEXversion PaycloseattentiontootherEXequipmentusedinconjunc-
tionwiththeSulLoggersensor. Observe the specications in the certicate as well as na-
tionalandlocalregulations. Ensurethatallpersonnelaresuitablyqualied. AlwaysuseapprovedEXbarriersforinstallation.
DANGER Chemicalandgashandlingsafety Handlinghazardousgasessuchashydrogensuldeorchemi-
calsamplescontainingdissolvedsuldescanbedangerous. Usersofthisproductareadvisedtofamiliarizethemselves withsafetyprocedures,useofpersonalprotectiveequipment 3
(PPE)andthecorrectuseofgasesand/orchemicals,andto carefullyreadallrelevantSafetyDataSheets. InstallationandcalibrationoftheSulLoggersensormayinclude theriskofcontactwithhazardousgasesorliquids. Itisespecially importantregardingtheSulLoggersensorforhydrogensulde. Hydrogensuldeisahighlytoxicgasandspecialprecautionsare neededtoavoidtheriskofinjuryordeath. Furthermore, the installation of the SulLogger sensor in ap-
plicationssuchassewernetworksorotherfacilitiescontaining rawwastewaterincludesariskofcontactwitharangeofother chemicalandbiologicalthreats,andanyoperatorperformingin-
stallationorserviceontheSulLoggersensorinsuchenviron-
ments should therefore be trained in the environmental safety andneedofPPEforthegivenapplication. Thismanualdoesnot cover any such training. The use of the SulLogger sensor for specic applications may involve risks that are not mentioned here. 4
SUPPORT SUPPORT, ORDERING, AND CONTACT INFORMATION Ifyouwishtoorderadditionalproductsorifyouencounterany problemsandneedscientic/technicalassistance,pleasedonot hesitatetocontactoursalesandsupportteam. Wewillrespond toyourinquirywithinoneworkingday. E-mail: sullogger@unisense.com UnisenseA/S Tueager1 DK-8200AarhusN,Denmark Tel: +4589449500 Fax: +4589449549 Furtherdocumentationandsupportisavailableatourwebsite sullogger.com 5
OVERVIEW TheSulLoggersensorisavailableinanumberofdifferentcon-
gurations. SomeoftheseareavailableinATEXcertiedversions. Pleasepaycloseattentiontoyourproducttypeduringthereading ofthismanualandtheinstallationofthesensor. ATEXcertied versionsofthesensorarelabeledSulLoggerX1andstandard versionsofthesensorarelabeledSulLoggerS1sensor. Ifyour productisanATEXcertiedproduct(X1series)itisveryimpor-
tant that you carefully follow the instructions relating to ATEX specic installation requirements. Guidance for installation of theSulLoggersensorcanbefoundinAppendixA.Thisguideis relevantforsensorswithserialnumbersstartingfrom1000000. Foroldersensorspleasendtherelevantguideonourhomepage https://sullogger.com 6
MAINTENANCE TheSulLoggersensorhasnoactivefoulinghandling,andthe sensor head therefore needs periodic cleaning. The necessary cleaningfrequencydependsverymuchonthetypeofwastewater andontheturbulencearoundthesensor. Stagnantwatergives idealconditionsforbiofouling,andwhenchoosingamountingpo-
sitionitisthereforeadvisabletolookforapositionwithrelatively highow/turbulence. For new installations it is recommended to clean and calibrate thesensoronetimeperweektoensurestableoperations. When moreexperiencehasbeengainedontheactualfoulingrateatthe installationsite,thismaintenancefrequencymaybechanged. CLEANING TheSulLoggersensorsdurabledesignmakescleaningeasy. Removeanyaccumulatedrags/bersandlargerdebrisfrom thesensor. Wipeoffthesensorheadwithacloth. Takespecialcareto ensurethattheinletslit(s)atthefrontofthesensorheadis completelyclean. Ifneededusewateroramildsoaptocleanthesensorhead. This is especially important if a greasy layer has build up on the sensor or if a layer of biolm can be directly seen. Ethanolcanalsobeused. CALIBRATION TheSulLoggersensorcomesfactorycalibratedandisassuch ready to use immediately as received. However, calibration in 7
theeldcanbeagoodwaytoensurethatthesensorkeepsthe absoluteaccuracyafterprolongeddeployment. Initiallyitisrec-
ommended to calibrate the sensor before installation and sub-
sequently once every month, but this can be subject to specic installationconditions. Thecalibrationprocedureissimpleandhasnodemandforaccess toSCADAvaluesorconnectionofcomputer/laptoporotherdata viewingdevices. ThecalibrationroutinereliesontheSulLogger sensors internal detection capability of the calibration gas and on the presence of a magnetic switch which is activated by the calibrationcap. Anumberofdifferentindicatorlightmodesgives signal to the operator of what to do during the calibration, as described below. The position of the indicator light is shown in Figure1. Figure1: Close-upviewoftheSulLoggersensorshowingthe positionoftheLEDindicatorlightusedforsignalingduringthe calibrationprocess(seebelow). TocalibratetheSulLoggersensor,acalibrationgaswithacon-
centrationof1000ppm H2Sisneeded. Thiscanbeorderedsepa-
ratelyfromUnisensetogetherwithacalibrationkitthatfacilitates theactivationofthecalibrationmodeinthesensorandensuresa relevantowofgastothesensorduringcalibration. Refertothe 8
tableonpage13formoredetailsontheorderingofcalibration equipment. Figure2showshowthecalibrationkitandgasbottle is mounted on the SulLogger sensor. The reduction valve on thegasbottleensuresthatthecorrectowofgasisexposedto thesensor,andthevalveshouldthereforebeopenedfullywhena calibrationisongoing. Figure2: Mountingofthecalibrationcapandgasbottle OverviewoflightindicationsontheSulLoggersensor TheindicatorlightontheSulLoggersensorisconstantlygiving astatusofthesensorandcanassuchbeusedtocheckthatthe sensor is on. Before calibration is initiated it is important to be familiarizedwiththevariousstatussignalsthatthesensorusesas indicationsandcallsforoperatoractions. Thesearesummarized inthetablebelow. ThecalibrationprocedureoftheSulLoggersensorconsistsof thefollowingsteps:
1. Initially,checkthatthesensorisonandisinnormalopera-
tionmodeshowingGreenashesontheindicatorlightwith oneashevery4seconds. (slowblinking) 9
Explanation Lightindication Green,1per4sec Sensorisoperatingnormally Green,1persec Calibrationcapmounted Green,3persec Gassignaldetected Greenandred,1persec Minimumcalibration Greenandred,2persec Fullcalibration. Closeandtakeoffcap Green,15times Calibrationsucceeded Red,15times Calibrationfailed 2. Connectthegasbottletothecalibrationcapwiththesup-
pliedowregulatorandgastube 3. Mountthecalibrationcaponthesensorandmakesureto align the arrow of the calibration cap with the position of theindicatorlight(thisactivatesthemagneticswitchinthe sensor). Therebythecalibrationmodeinthesensorisacti-
vatedandthelightindicatornowswitchestogreenashes withoneasheverysecond(mediumblinking) 4. OpentheowregulatoronthegasbottletoallowH2Sgas to reach the sensor. After a short time the light indicator should switch to green ashes at 3 ashes every second
(fastblinking) 5. Afterapproximately3minutesaminimumcalibrationhas beenachievedandthecalibrationprocedurecanbetermi-
nated. Thisisindicatedbyanalternatinggreenandredlight signalwithoneasheverysecond. Iftimeallowswerecom-
mendtocontinuethecalibrationuntilafullcalibrationhas beencompleted(seebelow) 6. A few minutes after the minimum calibration the ashing rateswitchestofasteralternationsbetweenredandgreen lightat2ashespersecond. Thisindicatesthatafullcali-
10
brationhasbeencompletedanditistimetoterminatethe calibrationprocedure 7. Wheneitheraminimumorafullcalibrationhasbeencom-
pletedtheregulatorvalveisclosedonthegasbottleandthe calibrationcapshouldberemoved. Paycloseattentionto theindicatorlightwhenremovingthecalibrationcapsince the result of the calibration is immediately shown on the indicatorlightwhenthemagneticswitchisturnedoff. The indicatorlightshows15fastashes(5persecond)ofeither green light or red light. If the ashes are green, the cali-
brationhasbeencompletedandstoredsuccessfully. Itthe ashesarered,thecalibrationhasfailedduringtheprocess andthecalibrationproceduremustberestarted.
! CAUTION Ifyoursensorisinstalledusingaowcell,youmustinspect theO-ringatthetipofthesensorwhenitisremovedfrom theowcell. IfO-ringiswornordamageditmustbereplaced byanew. ReplacementO-ringsandO-ringgreasehavebeen deliveredtogetherwiththeowcell.
IMPORTANTNOTE Thefteenshortgreenashesatthenalstageofcalibration istheonlyindicationvisibletotheoperatorofasuccessful calibration. However, if in doubt, we recommend that the calibrationvalidityischeckedsubsequenttocalibrationby exposingtheSulLoggersensortothecalibrationgasand checking the reading at the SCADA system or through the SulLoggerApp(TheSulLogger sensoronly). Pleasemake sure to turn the calibration cap so that the arrow is NOT 11
aligned with the light indication signal when checking the calibrationinordertoavoidthatanewcalibrationprocedure isinitiated. 12
ACCESSORIES HereisalistofaccessoriesavailablefortheSulLoggersensor. PleasecontactUnisensetoordertheseitems. LIST OF ACCESSORIES Unisenseitemno. Description S00021 S00022 S00023 C00005 C00004 S00032 S00020 C00055 C00002 C00003 C00056 S00016 S00017 S00049 S00035 SulLoggerX1/S1extensionpipekit RS-232X1barrierkit Power/4-20mAX1barrierkit H2Scalibrationgas,1000ppm Gasowregulator CalibrationCap CalibrationCap/Flowcell,threaded. 10bar 30mmpipestud 10meterRS-232cable,single-ended 10meterPower/4-20mAcable,single-ended Transportbox(pelicase) PowerComBoxS1EU PowerComBoxS1US PowerComS1cable,10m SulLoggerX1/S1Calibrationkit(allincl.) USING THESULFILOGGERAPP Onlyrelevantforsensorversion The SulLogger sensor comes with a free App which can be installedonWindowscomputers. The SulLogger App can be downloaded and installed from the SulLoggerwebsitehttps://sullogger.com/products TheSulLoggerAppenables:
13
1. VericationofthesensorandbatterystatusoftheSulLog-
gersensor. 2. Easy access to the data stored in the SulLogger sensor device. Thedatacanbeexportedincsv-leformat. 3. ViewingoflivedatavaluesfromtheSulLoggersensor. TodownloaddatafromtheSulLoggersensor,simplyconnect the RS232 cable to a USB port on a laptop with the SulLogger App installed. Open the SulLogger App and press the "Down-
load data" button. The data is stored as a csv-le in the folder
"UnisenseData"inyourlocaldocumentsfolder. USING THEPOWERCOMBOX Thecombinedpowerandcommunicationbox(PowerComBox)is abatteryboxusedinconnectionwiththeSulLogger sensor. It enablestheuseoftheSulLoggersensorinareaswithouta4-20 mAconnection. ThePowerComS1cableisusedtoconnectthe SulLogger sensortothePowerComBox. Thesensorcableis connectedtothesensorconnector(leftsideconnectorongure 3)onthePowerComBox. Data stored on the SulLogger sensor is downloaded using a USBcable. ConnectaUSBcabletotheUSBport(rightsidecon-
nector on gure 3) and open the SulLogger App on your com-
puter. For further information on how to download the sensor data,seetheSulLoggerAppsectioninthismanual. Thebatteryhasalifetimeofabout2months. Whenrecharging thebattery,usethepowercableincludedwiththePowerComBox. Thebatteryisrechargedafterabout6hours. TheSulLogger sensorcannotbeconnectedtothePowerComBoxwhilecharging. 14
Figure3: ConnectionsonthePowerComBox. Thesensorcableis connectedtothePowerComBoxusingthesensorconnectionto theleft. DataaredownloadedfromtheSulLoggersensortoa computerusingtheUSBportontherightside. 15
SPECIFICATIONS OPERATING PRINCIPLES OF THE SENSOR Electrochemicalmeasurementsofhydrogensulde TheSulLoggersensorcontainsmultipleindividualsensingel-
ementswhichareallbasedonUnisensecoremicrosensortechnol-
ogy. Forfurtherinformationonthetechnology,visit:www.sullogger.com TheelectrochemicalmicrosensorsintheSulLoggersensormea-
suresdissolvedH2Swhichisoneofthethreeformsofdissolved sulde in equilibrium in the water phase, H2S, HS-, and S2-. For allpracticalpurposesinsewerapplications,onlythersttwoof theseareofinterest. The pKa valuefordeprotonationofH2Sis around pH = 7,whichisapproximatelythesameastypicaldomes-
ticwastewater. Forthisreason,wastewaterwilltypicallycontain bothH2Sand HS,andaseparatemeasurementofthe pH-value ofthewastewaterisnecessarytocalculatethetotalamountof dissolvedsuldesinthewastewater. PleasecontactUnisenseif youneedguidanceonhowtondtheconcentrationoftotaldis-
solvedsuldefromtheSulLoggersensorsignal(onlydissolved H2S).Formanypurposeshowever,thedissolvedH2Smeasuredby theSulLoggersensorisindeedtheparameterofinterest,since theconcentrationofdissolvedH2Sisdirectlylinkedtotheemis-
sionofgasphaseH2Swhichgivesrisetothewell-knownodour andcorrosionproblemsinthesewer. Measuringtheconcentra-
tionofH2Sintheliquidinsteadofinthegasphasegivesamore reliablesignalsincethemeasurementisindependentofthedaily variationsinventilationconditionsintheheadspaceofthesewer, anditisindependentoftheexactmountingpositionofthesen-
sor(intheliquid,thesuldeconcentrationisfairlyhomogeneous, whereas very large differences in gas phase concentration has 16
beenseenfromtoptobottomofsewerheadspaces). Electrochemical sensors are sensitive to temperature changes, and for this reason each transducer in the sensor heads is com-
pensated for temperature effects with reference to an internal temperature measurement in the SulLogger sensors sensor head. MEASURING IN DIFFERENT MEDIA- UNIT CONVERSIONS TheelectrochemicalmicrosensorsintheSulLoggersensorsen-
sorheadmeasuresthepartialpressureofH2S.Inthegasphase, this can be converted to a ppm-reading if the total barometric pressureisknown,andintheliquidphasethemeasuredvaluecan beconvertedtoaconcentrationmeasuredineitherMormg/L byusingHenrysconstantandthemolarweight. Eachsensorisorderedwithaspecicationofmediumtype,and thesignaloutputfromtheSulLoggersensorischosenasppm for gas phase sensor heads and Smg/L for liquid phase sensor heads,the"S"infrontofmg/Lmeaningthatonlythesulfurfrac-
tionoftheweightiscountedinthespeciedmeasurement. The ppm values for gas phase sensors is calculated by using a standardbarometricpressureof1atm=101325Pa,andthemg/L valuesforliquidphasesensorsiscalculatedbyusingatempera-
turecorrectedvalueofHenrysconstant. Pleasenotethatother dissolvedsubstancesandhighsalinityvaluesmayalsoinuence theconversiontomg/Lunits. Formoreinformationonthis,please contactUnisense. SULFILOGGER SENSORS AnoverviewoftheSulLoggersensortypesandtheirfeatures canbefoundinthetablebelow. 17
SulLogger sensor Environment Range Maxpressure Inputpower Outputdata Mounting S1-1010-
5mgL Wastewater 0-5mg/L 10m Battery Loggeddata Freehangingor Wallmounted S1/X1-1020-
5mgL Wastewater 0-5mg/L 10m 20to30V RS-232or 4-20mA Freehangingor Wallmounted S1/X1-1120-
5mgLor 1000ppm Pressurized wastewater orgases 0-5mg/L 3barg 20to30V RS-232or 4-20mA In-lineor Flowcell X1-1220-
1000ppm Pressurized gases 0-1000ppm 10barg 20to30V RS-232or 4-20mA In-lineor Flowcell
+/-5%offullrange
<25seconds 1%offullrange 0-100%
-20 Cto+60 C(-4 Fto+140 F) 0 Cto+40 C(32 Fto+104 F) Technicalspecicationscommonforallsensors:
Accuracy Responsetime(t90) Detectionlimit Operatinghumidityrange Designtemperaturerange Operatingtemperaturerange Mechanicalspecicationscommonforallsensors:
Length Diameter Weight Material Warranty StainlessSteel-EN1.4404(316L) 240mm 0.85kg 3months 48.3mm/1pipe KNOWN INTERFERING SUBSTANCES Belowisalistofchemicalsthathasbeentestedagainstthesensor. Thetableshowswhereinterferenceshasbeenfoundandtowhich extendtheyinterfere. Thislistisalsoavailableonline:
www.sullogger.com/technology 18
Interferenceinliquid(%)2 Interferenceingas(%)1 Gas Methane Carbondioxide Nitrogen Oxygen Air Nitrousoxide Ammonia Hydrogen Carbonmonoxide Dimethylsulde Methylmercaptan Ethylmercaptan Sulfurdioxide Formula CH4 CO2 N2 O2 O2,N2,Ar N2O NH3 H2 CO
(CH3)2S CH3SH C2H6S SO2 0 0 0 0 0 0 0 96 77 18 44 14 1 1 Givenassignalfortheinterferingspeciesin%ofH2Ssignalat equalpartialpressures. 2 Givenassignalfortheinterferingspeciesin%ofH2Ssignalat equalmolarconcentrations. 0 0 0 0 0 0 0 0.8 0.6 18 174 13 40 19
SULFILOGGERSENSORSEWER INSTALLATIONS-TIPSANDTRICKS When installing the SulLogger sensor for monitoring hydro-
gensuldeinsewernetworksthefollowingnotesmayserveas guidelinesforamonitoringcampaign:
Hydrogen Sulde (H2S) is created under oxygen free con-
ditionsinthesewerwhichtypicallyoccursinForceMains
(pumpedsewerlines). TheconcentrationofdissolvedH2S increaseswhentheresidencetimeand/ortemperatureof thewastewaterincrease. H2Sisreleasedattheendofthe pipe,wherethewastewaterisagainincontactwithair(head space). Thiscanbeasimpleoverowtogravitationalsewer lines,apumpsumpforasubsequentForceMainspumpline, abuffertankorsimilar. The easiest and recommended way of deploying the Sul-
LoggersensorforH2Smonitoringinsewernetworksis toinstallthesensorwithaconnectingpipeprotectingthe cables. Thepipemustbesecuredtoeitherthesewerwallor atthetop/inletscreen. ThedissolvedH2Sisreleasedtogasphaseduringtherst stretchofanaeratedsewer. Theexactdistanceofrelease dependsonmanydifferentparameterssuchasturbulence andventilationconditions,butingeneralitisadvisableto measure as close to the end-of-pipe of the Force Main as possibletogetarealisticmeasureoftheinitialH2Sconcen-
tration. TheSulLoggerS1/X1-1120-5mgLsensorcanbemounted ataccesspointsdirectlyontheForceMain. Thiscaneither 20
be done by attaching the sensors threaded tting to the pressure pipe or through a dedicated load-lock system at thechosenpositiononthepipe. Notehowever,thatliquid velocities are often high in pressurized pipes, and hence thereisanincreasedriskofsensordamageassociatedwith thistypeofmeasurements. IngeneralitisrecommendedthattheSulLoggersensor ismountedinawaywhereitiseitheralwayssubmergedor alwaysoutofthewater(intheheadspace). Interpretation ofdatacanbedifcultiftheSulLoggersensorisintermit-
tently submerged. Furthermore, a mounting at a position wheretheSulLoggersensorgetsinandoutofthewater often leads to an increased risk of biofouling and scaling, whichsignicantlylimitsthetimewithreliabledatainaeld deployment. 21
TROUBLESHOOTING Problem: The SulLogger App does not recognize the SulLog-
gersensorwhenitisconnected(version1010only). Solution: TrychargingthebatteryofthePowerComBox,ortry disconnectingthewirefromthePowerComBoxtothesensorand reconnectitsubsequently. Problem: TheSulLoggersensorisshowingvaluesclosetozero all the time even though grab sample analysis of the medium showssignicantsuldeconcentrations. Solution: TheinletoftheSulLoggersensormybeblockedby fouling. Tryrinsingthesensorandgentlywipingthefrontwitha wetcloth,takingcarenottoapplyforcearoundtheinletsofthe sensingelements. Ifthemeasurementsareperformed inliquid phase(e.g. inwastewater),checkthe pH-valueoftheliquid. Ifthe mediumisalkaline(above pH-8)thenthedissolvedsuldeispri-
marilyonthe HS-form,whichisnotdetectedbytheSulLogger sensor. For slightly alkaline pH-values (pH-7-8), the pH-effect can be corrected for, but for higher pH-values, we recommend usingtheSulLoggersensorinalow-volumesidestreamreac-
tor,wherethewastewaterisacidiedtoensurethatalldissolved suldeisontheH2S-form. Problem: TheSulLoggersensordoesnotmeasurecorrectval-
uesaccordingtoourstandardtests. Solution: When judging the accuracy of the SulLogger sen-
sor we always recommend to apply a gas phase measurement withgasfromapre-mixedgasbottlecontainingaknownH2Scon-
centration. Please note that the accuracy (grade) of gas bottle 22
concentrationsvary. Theaccuracyshouldbestatedonthebottle orcanbeprovidedbythevendor. Ifjudgingtheaccuracyofthe SulLoggersensorbasedonliquidphasemeasurementsthere are various ways to do this, all of which are prone to higher de-
greesofuncertaintythanthegasphasetesting: Oneoptionisto prepare a standard (stock solution) mixture with a known con-
centrationofH2S.Experienceshowsthatthisisnotaneasytask andthattheconcentrationshouldcontinuouslybecheckedwith laboratoryequipmentsuchasspectrophotometricmethods. For moreinformationonthis,pleasecontactUnisense. Anotherop-
tionistodograbsampleanalysisofwastewaterandcomparethe resultsofasuldeanalysis(e.g. Clinesmethod)withthemeasure-
mentvaluesoftheSulLoggersensor. Pleasebeaware,though, thatstandardsuldeanalysesofwastewatergrabsamplesmea-
surestheTotalSuldeandnotonlytheTotalDissolvedSulde. For a comparison between the SulLogger sensor measurements andgrabsamplevaluesitisthusneededtolteroutanyparticu-
latesuldeofthesamples,anditisnecessarytodoa pH-based calculation of the total dissolved sulde from the SulLogger sensorsignal. Furthermore,itshouldbenotedthatcorrectsam-
plingtechniquesandsamplepreservationshouldbeconducted withutmostcaretoavoidsignicantinuencesonthemeasured suldeconcentrationsinthegrabsamples. Forfurtherinforma-
tion and guidance on procedures for liquid phase comparisons, pleasecontactUnisense. Problem: EventhoughtheSulLoggersensorhasjustbeencali-
brated,themeasurementsareshowingincorrectvalues. Solution: When calibrating the SulLogger sensor it is impor-
tant that the procedure described in the Maintenance chapter above is followed. Please refer to the relevant chapter, and if 23
this is still giving problems, contact Unisense. Please also note thateventhoughthesignaloftheSulLoggersensoristemper-
ature compensated, the accuracy of the sensor is highest if the calibrationisdoneatatemperatureclosetothetemperatureof subsequentmeasurements. 24
COMPLIANCEINFORMATION FCCPART15,CLASSA This device complies with part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may notcauseharmfulinterference,and(2)thisdevicemustaccept anyinterferencereceived,includinginterferencethatmaycause undesiredoperation. Changesormodicationstothisunitnotexpresslyapprovedby thepartyresponsibleforcompliancecouldvoidtheusersauthor-
itytooperatetheequipment. Note: Thisequipmenthasbeentestedandfoundtocomplywith thelimitsforaClassAdigitaldevice,pursuanttopart15ofthe FCCRules. Theselimitsaredesignedtoprovidereasonablepro-
tectionagainstharmfulinterferencewhentheequipmentisop-
eratedinacommercialenvironment. Thisequipmentgenerates, uses,andcanradiateradiofrequencyenergyand,ifnotinstalled andusedinaccordancewiththeinstructionmanual,maycause harmfulinterferencetoradiocommunications. Operationofthis equipment in a residential area is likely to cause harmful inter-
ference in which case the user will be required to correct the interferenceathisownexpense. F00001_FCC_ComplianceInformation_100.pdf 25
A:INSTALLATIONGUIDE 26
SulLoggerInstallationmanualRevision1.00(ValidfromMay2019)INTRODUCTIONThisguidecoversthestepsrequiredtoin-stalltheSulLoggersensor,includingtheelectricalandmechanicalinstallationpro-cedures.Forfurtherinformationabouttheuseandmaintenanceoftheproduct,wekindlyrefertothelatestUserManualavail-ableatwww.SulLogger.com.Warrantyandlegalinformationaredescribedinthetermsandconditionsofyourorderconr-mationandalsoavailableuponrequest.SAFETYPleasereadthismanualcarefullybeforeun-packing,installingoroperatingthedevice.AlthoughtheoperationoftheSulLoggersensorisstraightforward,itismandatorytobefamiliarwiththecontentofthismanual.Failuretodosomayresultinseriousinjurytooperatorsorotherpersons,ordamagetoequipment.DonotuseorinstalltheSul-Loggersensorandauxiliaryequipmentinanyotherwaythanspeciedinthisman-ual.Specialattentionmustbepaidtoalldangerandcautionstatements.DonotusetheSulLoggersensoroutsidethespeci-edelectrical,mechanicalandthermalpa-rametersasspeciedinthedatasheet.Dur-ingtheinstallation,pleasewearpersonalprotectiveequipment(PPE)accordingtothesafetyregulationoftheinstallationsite.HazardstatementnomenclatureDANGERIndicatesapotentiallyorimminentlyhazardoussituationwhich,ifnotavoided,couldresultindeathorseri-ousinjury.!CAUTIONIndicatesapotentiallyhazardoussitu-ationthatmayresultinminorormod-erateinjury.IMPORTANTNOTEInformationthatrequiresspecialem-phasis.Hazardousareas-SulLoggersensorEXversionPaycloseattentiontootherEXequip-mentusedinconjunctionwiththeSul-Loggersensor.Observethespecicationsinthecerti-cateaswellasnationalandlocalregula-tions.Ensurethatallpersonnelaresuitablyqualied.AlwaysuseapprovedEXbarriersforin-stallation.ChemicalandgashandlingsafetyDANGERHandlinghazardousgasessuchashy-drogensuldeorchemicalsamplescontainingdissolvedsuldescanbedangerous.Usersofthisproductareadvisedtofamiliarizethemselveswithsafetyprocedures,useofpersonalpro-tectiveequipment(PPE)andthecor-rectuseofgasesand/orchemicals,andtocarefullyreadallrelevantSafetyDataSheets.InstallationandcalibrationoftheSulLog-gersensormayincludetheriskofcon-tactwithhazardousgasesorliquids.Hydro-gensuldeisahighlytoxicgasandspecialprecautionsareneededtoavoidtheriskofinjuryordeath.Furthermore,theinstalla-tionoftheSulLoggersensorinapplica-tionssuchassewernetworksorotherfacil-itiescontainingrawwastewaterincludesariskofcontactwitharangeofotherchem-icalandbiologicalthreats,andanyopera-torperforminginstallationorserviceontheSulLoggersensorinsuchenvironmentsshouldthereforebetrainedintheenviron-mentalsafetyandneedofPPEforthegivenapplication.Thismanualdoesnotcoveranysuchtraining.TheuseoftheSulLoggersensorforspecicapplicationsmayinvolverisksthatarenotmentionedhere.OVERVIEWTheSulLoggersensorisavailableindif-ferentcongurationsinanATEXcertiedversion(X1sensors)andanon-ATEXcerti-edversion(S1sensors).IfyourproductisanATEXcertiedproduct,itiscrucialthatyoucarefullyfollowtheinstructionsre-latingtoATEXspecicinstallationrequire-ments.EU-ATEXcerticationTheSulLoggerX1sensorhasthefollow-ingmarking:ELECTRICALINSTALLATIONBarriersforSulLoggerX1installationWhenusingtheSulLoggersensor,aZenerbarrierorEXisolatorMUSTalwaysbeused.WerecommendusingbarrierstestedandapprovedbyUnisense.Hazardous areaSafe areaApparatus, e.g. SCADAZener barrier4-20 mA / powerEX isolatororRS-232Zener barrierEX isolatororConnectionTheSulLoggersensorhastwoconnec-tors:A4-20mApowerconnector(5-pinM12male)andanRS-232connector(5-pinM12female).WerecommendusingM12plugsinstainlesssteel(EN1.4404/316L).ThesecablescanbeprovidedbyUnisense.Pleasenotethatproductspoweredbyabatteryboxmayrequirespecialattention.PleaserefertothelatestUserManual.PortparametersPleasenotethatthecolorslistedhereonlyapplyforcablessuppliedbyUnisense.TheSulLoggerisdesignedfor4-20mAloop-power.However,thisconnectorcanalsobeusedonlyforDCsupply.Power/4-20mAPinColorConnection1BrownPositive(+12to24v)2WhiteNotconnected3BlueNegative(-)4BlackNotconnected5GrayNotconnectedCommunication(RS-232)PinColorConnection1BrownTXD2WhiteNotconnected3BlueRXD4BlackGND5GrayNotconnectedGroundingTheSulLoggersensorshouldalwaysbegrounded.However,national,regionalorlocalregulationmayapply.ForuseinATEX-regulatedareas,seeEN/IEC60079-14.UnisenserecommendsthattheSulLog-gersensorisalwaysgroundedusingtheM6threadatthebackofthedeviceforasecuregroundingconnection.DANGERUsingaSulLoggersensorfromtheX1-serieswithoutgroundingwillvoidtheATEXcerticate.
MECHANICALINSTALLATIONTheSulLoggersensorcanbeinstalledinfourdifferentways:a)Freehanging(liquids)b)Fixed(liquids)c)Inline(gases)d)Fixed(gases)FLOWa)b)c)d)xxa)Freehanging(liquids)Thismountingoptionisidealforlowowreservoirsandchannelsandgenerallysuit-ableforshortmeasurementcampaigns.TheSulLoggersensorisinstalledattheendofanextensionpipe,attachedtoanan-chorpointusingasteelwireandsuspendeddownintothesewage.Theextensionpipeensuresthatthecablesandconnectorsareprotectedfromfouling.Toperformthisinstallation,youmustensure:Thatthecirculationinthereservoirissuf-cienttoensuregoodmixing.Thatthesensortipcannottouchthewallsandbottomofthereservoir,espe-ciallyduringhighows.Ifthereisariskofthesensorhittingthewalls,thenyoushouldperformthe"xed(liquids)"instal-lationinstead.Thatyouhaveasuitableanchorpointavailabletoattachthesteelwireto.Thatthelengthofthesteelwireallowsthesensortoremainintheliquidatalltimeswhilethebackoftheextensionpipeatthesametimeremainsoutoftheliquid.b)Fixed(liquids)Thismountingoptionisidealforpermanentinstallationsinlowandhighowreservoirsandchannels.Itissimilartothe"Freehang-ing(liquids)"installationinmostregardsbuthere,theextensionpipe(andsensor)isxated-nothanginginthesteelwire.Unisensedoesnotsupplybracketsoranyothermaterialsforthisinstallationmethod.c)Inline(gases)DirectlyonthemainpipeTheSulLoggersensorcanbeinstalledonapipestuballowingmeasurementinaclosedsystem.Beforeinstallingthepipestuba30mmholemustbedrilledinthemainpipe.ThepipestubissuppliedbyUnisenseandmustbeweldedontothemainpipe.Useaspanner(size36)toscrewtheSul-Loggersensorontothepipestub.BewareofthepressureratingoftheSul-Loggersensorandensurethatthepressurepipesystemdoesnotexceedthepressurelimit.BypassowusingaowcellAportionoftheowcanbedi-rected/bypassedintoaowcellonwhichtheSulLoggersensorisinstalled.Beforetheinstallation,pleaseensurethatthepressureandtemperaturerat-ingsofthebypassedowarewithinthespecicationsoftheSulLoggersensor.TheowcellissuppliedbyUnisense.Useaspanner(size36)toscrewtheSul-Loggersensorontotheowcell.d)Fixed(gases)TheSulLoggersensorcanbemountedinpumpwells,reservoirsetc.abovewa-terlevelforgasphasemeasurements.TheouterdiameteroftheSulLoggersen-sorresemblesa1.5"pipeandcanthusbemountedusingastandard1.5"pipering.ExtensionpipeinstallationForfreehangingandxedmeasurementsinliquids,anextensionpipemustbeat-tachedtotheSulLoggersensor.Thisg-ureshowsthestepsneededtoperformthisinstallation:UnscrewthethreehexscrewsonthebackoftheSulLoggersensorandre-movetheouterring.Feedthecablesthroughtheextensionpipeandattachthemtotheconnectorsonthesensor.Alignthesensorwiththescrewholesoftheextensionpipeandscrewinthethreehexscrewstosecurethesensortotheextensionpipe.TemperatureratingThesensorisdesignedtowithstandtemper-aturesfrom-20Cto+60C.DANGERExposingaSulLoggersensorfromtheX1-seriesforambienttempera-tureslowerthan-20Corhigherthan+60CwillvoidtheATEXcerticate.Thesensor,bothS1andX1series,performsoptimallybetween0Cand+40C.DisassemblyWedonotrecommendyoutodisassembletheSulLoggersensor.IMPORTANTNOTEDisassemblyofthesensorwillvoidthewarrantyandrenderthesensornon-functional.DANGERDisassemblyofaSulLoggersensorfromtheX1-serieswillvoidtheATEXcerticate.SUPPORTUnisenseA/STueager1DK-8200AarhusNDenmarkE-mail:SulLogger@unisense.comTel:+4589449500Fax:+4589449549Forfurtherdocumentation,thelatestUserManualandotherresources,pleasevisittheSulLoggersensorwebsite:www.SulLogger.com
UNISENSEUNISENSE, DENMARKwww.unisense.com info@unisense.com
1 | External Photos | External Photos | 273.57 KiB | June 11 2019 |
External Photos - Label placement File Name: F00001_FCC_External_photos_100.pdf
1 | Internal Photos | Internal Photos | 385.63 KiB | June 11 2019 |
Internal Photos - PCB Internal photo Inner core File Name: F00001_FCC_Internal_photos_100.pdf Assembly photos outer core inner core - sensorhead File Name: F00001_FCC_Internal_photos_100.pdf
frequency | equipment class | purpose | ||
---|---|---|---|---|
1 | 2019-11-06 | JAD - Part 15 Class A Digital Device | Original Equipment |
app s | Applicant Information | |||||
---|---|---|---|---|---|---|
1 | Effective |
2019-11-06
|
||||
1 | Applicant's complete, legal business name |
Unisense a/s
|
||||
1 | FCC Registration Number (FRN) |
0028625234
|
||||
1 | Physical Address |
Tueager 1
|
||||
1 |
Aarhus N, N/A
|
|||||
1 |
Denmark
|
|||||
app s | TCB Information | |||||
1 | TCB Application Email Address |
T******@intertek.com
|
||||
1 | TCB Scope |
A1: Low Power Transmitters below 1 GHz (except Spread Spectrum), Unintentional Radiators, EAS (Part 11) & Consumer ISM devices
|
||||
app s | FCC ID | |||||
1 | Grantee Code |
2ATWP
|
||||
1 | Equipment Product Code |
SULFILOGGER1
|
||||
app s | Person at the applicant's address to receive grant or for contact | |||||
1 | Name |
S**** P****
|
||||
1 | Telephone Number |
+4542********
|
||||
1 | Fax Number |
89449********
|
||||
1 |
s******@unisense.com
|
|||||
app s | Technical Contact | |||||
1 | Firm Name |
Unisense a/s
|
||||
1 | Name |
C**** S******** M******
|
||||
1 | Physical Address |
Denmark
|
||||
1 |
c******@unisense.com
|
|||||
app s | Non Technical Contact | |||||
n/a | ||||||
app s | Confidentiality (long or short term) | |||||
1 | Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | No | ||||
1 | Long-Term Confidentiality Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | No | ||||
if no date is supplied, the release date will be set to 45 calendar days past the date of grant. | ||||||
app s | Cognitive Radio & Software Defined Radio, Class, etc | |||||
1 | Is this application for software defined/cognitive radio authorization? | No | ||||
1 | Equipment Class | JAD - Part 15 Class A Digital Device | ||||
1 | Description of product as it is marketed: (NOTE: This text will appear below the equipment class on the grant) | Gas Sensing Device | ||||
1 | Related OET KnowledgeDataBase Inquiry: Is there a KDB inquiry associated with this application? | No | ||||
1 | Modular Equipment Type | Does not apply | ||||
1 | Purpose / Application is for | Original Equipment | ||||
1 | Composite Equipment: Is the equipment in this application a composite device subject to an additional equipment authorization? | No | ||||
1 | Related Equipment: Is the equipment in this application part of a system that operates with, or is marketed with, another device that requires an equipment authorization? | No | ||||
1 | Is there an equipment authorization waiver associated with this application? | No | ||||
1 | If there is an equipment authorization waiver associated with this application, has the associated waiver been approved and all information uploaded? | No | ||||
app s | Test Firm Name and Contact Information | |||||
1 | Firm Name |
Intertek Testing Services NA, Inc.
|
||||
1 | Name |
J******** K****
|
||||
1 | Telephone Number |
651-7********
|
||||
1 | Fax Number |
651 7********
|
||||
1 |
j******@intertek.com
|
|||||
Equipment Specifications | |||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Line | Rule Parts | Grant Notes | Lower Frequency | Upper Frequency | Power Output | Tolerance | Emission Designator | Microprocessor Number | |||||||||||||||||||||||||||||||||
1 | 1 | 15B |
some individual PII (Personally Identifiable Information) available on the public forms may be redacted, original source may include additional details
This product uses the FCC Data API but is not endorsed or certified by the FCC