PCTEST ENGINEERING LABORATORY, INC. 18855 Adams Ct, Morgan Hill, CA 95037 USA Tel. +1.410.290.6652 / Fax +1.410.290.6654 http://www.pctest.com Certification of Calibration D835V2 SN: 4d180 Object Calibration procedure(s) Extended Calibration date:
Description:
Calibration Equipment used:
Manufacturer Agilent Agilent Amplifier Research Anritsu Anritsu Anritsu Control Company Control Company Keysight Keysight Technologies Mini-Circuits Mini-Circuits Narda Pasternack Pasternack SPEAG SPEAG SPEAG Model 8753ES N5182A 15S1G6 ML2495A MA2411B MA2411B 4040 4352 772D 85033E BW-N20W5+
NLP-2950+
4772-3 PE2208-6 PE5011-1 DAKS-3.5 EX3DV4 DAE4 Procedure for Calibration Extension for SAR Dipoles. May 16, 2019 SAR Validation Dipole at 835 MHz. Description S-Parameter Network Analyzer MXG Vector Signal Generator Amplifier Power Meter Pulse Power Sensor Pulse Power Sensor Temperature / Humidity Monitor Ultra Long Stem Thermometer Dual Directional Coupler Standard Mechanical Calibration Kit (DC to 9GHz, 3.5mm) 6/4/2018 DC to 18 GHz Precision Fixed 20 dB Attenuator Low Pass Filter DC to 2700 MHz Attenuator (3dB) Bidirectional Coupler Torque Wrench Portable DAK SAR Probe Dasy Data Acquisition Electronics CBT CBT CBT CBT 7/19/2017 9/11/2018 7/20/2018 7/10/2018 CBT 10/21/2018 10/30/2018 11/20/2018 2/28/2018 2/28/2018 CBT Serial Number Cal Date Cal Interval Cal Due 10/2/2019 10/2/2018 US39170118 6/15/2019 MY47420837 6/15/2018 Annual Annual N/A Annual Annual Annual Biennial Biennial N/A Annual N/A N/A N/A N/A Biennial Annual Annual Annual CBT 10/21/2019 10/30/2019 11/20/2019 2/28/2020 2/28/2020 CBT 6/4/2019 343972 941001 1207470 1339007 150761911 170330160 MY52180215 MY53401181 CBT CBT CBT CBT 7/19/2019 9/11/2019 7/20/2019 7/10/2019 N/A N/A 9406 N/A N/A 1045 7416 1402 Measurement Uncertainty = 23% (k=2) Name Function Signature Calibrated By:
Parker Jones Team Lead Engineer Approved By:
Kaitlin OKeefe Senior Technical Manager Object:
D835V2 SN: 4d180 Date Issued:
05/16/2019 Page 1 of 4 DIPOLE CALIBRATION EXTENSION Per KDB 865664 D01, calibration intervals of up to three years may be considered for reference dipoles when it is demonstrated that the SAR target, impedance and return loss of a dipole have remained stable according to the following requirements:
1. The measured SAR does not deviate more than 10% from the target on the calibration certificate. 2. The return-loss does not deviate more than 20% from the previous measurement and meets the required 20dB minimum return-loss requirement. 3. The measurement of real or imaginary parts of impedance does not deviate more than 5 from the previous measurement. The following dipole was checked to pass the above 3 requirements to have 2-year calibration period from the calibration date:
Certificate SAR Target Head (10g) W/kg @ 23.0 dBm 1.244 Certificate SAR Target Body (10g) W/kg @ 23.0 dBm 1.262 Measured Head SAR
(10g) W/kg @
23.0 dBm Deviation 10g
(%) Certificate Impedance Head (Ohm) Real Measured Impedance Head (Ohm) Real Difference
(Ohm) Real Certificate Impedance Head (Ohm) Imaginary Measured Impedance Head (Ohm) Imaginary Difference
(Ohm) Imaginary Certificate Return Loss Head (dB) Measured Return Loss Head (dB) Deviation (%) PASS/FAIL 1.3 4.50%
50.7 49.8 0.9
-5.1
-6.5 1.4
-25.9
-25.2 2.60%
PASS Measured Body SAR
(10g) W/kg @
23.0 dBm Deviation 10g
(%) Certificate Impedance Body (Ohm) Real Measured Impedance Body (Ohm) Real Difference
(Ohm) Real Certificate Impedance Body (Ohm) Imaginary Measured Impedance Body (Ohm) Imaginary Difference
(Ohm) Imaginary Certificate Return Loss Body (dB) Measured Return Loss Body (dB) Deviation (%) PASS/FAIL 1.34 6.18%
47.2 47.3 0.1
-8.2
-6.3 1.9
-21.1
-23.1
-9.20%
PASS Certificate SAR Target Head (1g) W/kg @ 23.0 dBm 1.92 Certificate SAR Target Body (1g) W/kg @ 23.0 dBm 1.918 Calibration Date Extension Date Certificate Electrical Delay (ns) 5/18/2018 5/16/2019 1.396 Calibration Date Extension Date Certificate Electrical Delay (ns) 5/18/2018 5/16/2019 1.396 Measured Head SAR (1g) W/kg @ 23.0 Deviation 1g
(%) dBm 1.99 3.65%
Measured Body SAR (1g) W/kg @ 23.0 Deviation 1g
(%) dBm 2.05 6.88%
Object:
D835V2 SN: 4d180 Date Issued:
05/16/2019 Page 2 of 4 Impedance & Return-Loss Measurement Plot for Head TSL Object:
D835V2 SN: 4d180 Date Issued:
05/16/2019 Page 3 of 4 Impedance & Return-Loss Measurement Plot for Body TSL Object:
D835V2 SN: 4d180 Date Issued:
05/16/2019 Page 4 of 4 PCTEST ENGINEERING LABORATORY, INC. 18855 Adams Ct, Morgan Hill, CA 95037 USA Tel. +1.410.290.6652 / Fax +1.410.290.6654 http://www.pctest.com Certification of Calibration Object Calibration procedure(s) Extended Calibration date:
Description:
Calibration Equipment used:
D1750V2 SN: 1104 Procedure for Calibration Extension for SAR Dipoles. September 07, 2018 SAR Validation Dipole at 1750 MHz. Measurement Uncertainty = 23% (k=2) Name Function Signature Calibrated By:
Sangmin Cha Team Lead Engineer Approved By:
Kaitlin OKeefe Senior Technical Manager Object:
D1750V2 SN: 1104 Date Issued:
09/07/2018 Page 1 of 4 Manufacturer Model Description Cal Date Cal Interval Cal Due Serial Number Agilent 8753ES S-Parameter Network Analyzer 9/14/2017Annual 9/14/2018US39170118 Agilent N5182A MXG Vector Signal Generator 3/19/2018Annual 3/19/2019US46240505 Amplifier Research 15S1G6 Amplifier CBTN/ACBT343972Anritsu ML2496A Power Meter 10/9/2017Annual 10/9/20181138001Anritsu MA2411B Pulse Power Sensor 11/15/2017Annual 11/15/20181339007Anritsu MA2411B Pulse Power Sensor 11/22/2017Annual 11/22/20181339008Control Company 4040Temperature / Humidity Monitor 2/28/2018Biennial 2/28/2020150761911Control Company 4352Ultra Long Stem Thermometer 2/14/2017Biennial 2/14/2019170112507Keysight 772D Dual Directional Coupler CBTN/ACBTMY52180215 Keysight Technologies 85033E Standard Mechanical Calibration Kit (DC to 9GHz, 3.5mm) 6/4/2018Annual 6/4/2019MY53401181 Mini-Circuits BW-N20W5+ DC to 18 GHz Precision Fixed 20 dB Attenuator CBTN/ACBTN/A Mini-Circuits NLP-2950+ Low Pass Filter DC to 2700 MHz CBTN/ACBTN/A Narda 4772-3 Attenuator (3dB) CBTN/ACBT9406Pasternack PE2209-10 Bidirectional Coupler CBTN/ACBTN/A Pasternack PE5011-1 Torque Wrench 7/19/2017Biennial 7/19/2019N/A SPEAG DAK-3.5 Dielectric Assessment Kit 5/15/2018Annual 5/15/20191070SPEAG ES3DV3 SAR Probe 9/18/2017Annual 9/18/20183287SPEAG DAE4 Data Acquisition Electronics 1/26/2018Annual 1/26/20191533 DIPOLE CALIBRATION EXTENSION Per KDB 865664 D01, calibration intervals of up to three years may be considered for reference dipoles when it is demonstrated that the SAR target, impedance and return loss of a dipole have remained stable according to the following requirements:
1. The measured SAR does not deviate more than 10% from the target on the calibration certificate. 2. The return-loss does not deviate more than 20% from the previous measurement and meets the required 20dB minimum return-loss requirement. 3. The measurement of real or imaginary parts of impedance does not deviate more than 5 from the previous measurement. The following dipole was checked to pass the above 3 requirements to have 2-year calibration period from the calibration date:
Object:
D1750V2 SN: 1104 Date Issued:
09/07/2018 Page 2 of 4 Calibration DateExtension DateCertificate Electrical Delay (ns)Certificate SAR Target Head (1g) W/kg @ 20.0 dBmMeasured Head SAR (1g) W/kg @ 20.0 dBmDeviation 1g (%)Certificate SAR Target Head (10g) W/kg @ 20.0 dBmMeasured Head SAR (10g) W/kg @ 20.0 dBmDeviation 10g (%)Certificate Impedance Head (Ohm) RealMeasured Impedance Head (Ohm) RealDifference (Ohm) RealCertificate Impedance Head (Ohm) ImaginaryMeasured Impedance Head (Ohm) ImaginaryDifference (Ohm) ImaginaryCertificate Return Loss Head (dB)Measured Return Loss Head (dB)Deviation (%)PASS/FAIL9/7/20179/7/20181.2173.643.62-0.55%1.921.941.04%50.849.81-0.2-0.30.1-41.5-35.913.50%PASSCalibration DateExtension DateCertificate Electrical Delay (ns)Certificate SAR Target Body (1g) W/kg @ 20.0 dBmMeasured Body SAR (1g) W/kg @ 20.0 dBmDeviation 1g (%)Certificate SAR Target Body (10g) W/kg @ 20.0 dBmMeasured Body SAR (10g) W/kg @ 20.0 dBmDeviation 10g (%)Certificate Impedance Body (Ohm) RealMeasured Impedance Body (Ohm) RealDifference (Ohm) RealCertificate Impedance Body (Ohm) ImaginaryMeasured Impedance Body (Ohm) ImaginaryDifference (Ohm) ImaginaryCertificate Return Loss Body (dB)Measured Return Loss Body (dB)Deviation (%)PASS/FAIL9/7/20179/7/20181.2173.663.844.92%1.962.075.61%46.52745.41.1-0.69-1.60.9-28.7-25.810.10%PASS Impedance & Return-Loss Measurement Plot for Head TSL Object:
D1750V2 SN: 1104 Date Issued:
09/07/2018 Page 3 of 4 Impedance & Return-Loss Measurement Plot for Body TSL Object:
D1750V2 SN: 1104 Date Issued:
09/07/2018 Page 4 of 4 PCTEST ENGINEERING LABORATORY, INC. 18855 Adams Ct, Morgan Hill, CA 95037 USA Tel. +1.410.290.6652 / Fax +1.410.290.6654 http://www.pctest.com Certification of Calibration D1750V2 SN: 1092 Object Calibration procedure(s) Extended Calibration date:
Description:
Calibration Equipment used:
Manufacturer Agilent Agilent Amplifier Research Anritsu Anritsu Anritsu Control Company Control Company Keysight Keysight Technologies Mini-Circuits Mini-Circuits Narda Pasternack Pasternack SPEAG SPEAG SPEAG Model 8753ES N5182A 15S1G6 ML2495A MA2411B MA2411B 4040 4352 772D 85033E BW-N20W5+
NLP-2950+
4772-3 PE2209-10 PE5011-1 DAKS-3.5 EX3DV4 DAE4 Procedure for Calibration Extension for SAR Dipoles. May 15, 2019 SAR Validation Dipole at 1750 MHz. Description S-Parameter Network Analyzer MXG Vector Signal Generator Amplifier Power Meter Pulse Power Sensor Pulse Power Sensor Temperature / Humidity Monitor Ultra Long Stem Thermometer Dual Directional Coupler Standard Mechanical Calibration Kit (DC to 9GHz, 3.5mm) 6/4/2018 DC to 18 GHz Precision Fixed 20 dB Attenuator Low Pass Filter DC to 2700 MHz Attenuator (3dB) Bidirectional Coupler Torque Wrench Portable DAK SAR Probe Dasy Data Acquisition Electronics CBT CBT CBT CBT 7/19/2017 9/11/2018 7/20/2018 7/10/2018 CBT 10/21/2018 10/30/2018 11/20/2018 2/28/2018 2/28/2018 CBT Serial Number Cal Date Cal Interval Cal Due 10/2/2019 10/2/2018 US39170118 6/15/2019 MY47420837 6/15/2018 Annual Annual N/A Annual Annual Annual Biennial Biennial N/A Annual N/A N/A N/A N/A Biennial Annual Annual Annual CBT 10/21/2019 10/30/2019 11/20/2019 2/28/2020 2/28/2020 CBT 6/4/2019 343972 941001 1207470 1339007 150761911 170330160 MY52180215 MY53401181 CBT CBT CBT CBT 7/19/2019 9/11/2019 7/20/2019 7/10/2019 N/A N/A 9406 N/A N/A 1045 7416 1402 Measurement Uncertainty = 23% (k=2) Name Function Signature Calibrated By:
Parker Jones Team Lead Engineer Approved By:
Kaitlin OKeefe Senior Technical Manager Object:
D1750V2 SN: 1092 Date Issued:
05/15/2019 Page 1 of 4 DIPOLE CALIBRATION EXTENSION Per KDB 865664 D01, calibration intervals of up to three years may be considered for reference dipoles when it is demonstrated that the SAR target, impedance and return loss of a dipole have remained stable according to the following requirements:
1. The measured SAR does not deviate more than 10% from the target on the calibration certificate. 2. The return-loss does not deviate more than 20% from the previous measurement and meets the required 20dB minimum return-loss requirement. 3. The measurement of real or imaginary parts of impedance does not deviate more than 5 from the previous measurement. The following dipole was checked to pass the above 3 requirements to have 2-year calibration period from the calibration date:
Certificate SAR Target Head (10g) W/kg @ 20.0 dBm 1.9 Certificate SAR Target Body (10g) W/kg @ 20.0 dBm 1.94 Measured Head SAR
(10g) W/kg @
20.0 dBm Deviation 10g
(%) Certificate Impedance Head (Ohm) Real Measured Impedance Head (Ohm) Real Difference
(Ohm) Real Certificate Impedance Head (Ohm) Imaginary Measured Impedance Head (Ohm) Imaginary Difference
(Ohm) Imaginary Certificate Return Loss Head (dB) Measured Return Loss Head (dB) Deviation (%) PASS/FAIL 1.89
-0.53%
49.2 48.4 0.8
-1
-0.1 0.9
-37.8
-38.2
-1.00%
PASS Measured Body SAR
(10g) W/kg @
20.0 dBm Deviation 10g
(%) Certificate Impedance Body (Ohm) Real Measured Impedance Body (Ohm) Real Difference
(Ohm) Real Certificate Impedance Body (Ohm) Imaginary Measured Impedance Body (Ohm) Imaginary Difference
(Ohm) Imaginary Certificate Return Loss Body (dB) Measured Return Loss Body (dB) Deviation (%) PASS/FAIL 1.91
-1.55%
44.8 44.2 0.6
-0.6 1.1 1.7
-25.2
-24.2 3.90%
PASS Certificate SAR Target Head (1g) W/kg @ 20.0 dBm 3.61 Certificate SAR Target Body (1g) W/kg @ 20.0 dBm 3.64 Calibration Date Extension Date Certificate Electrical Delay (ns) 5/15/2018 5/15/2019 1.217 Calibration Date Extension Date Certificate Electrical Delay (ns) 5/15/2018 5/15/2019 1.217 Measured Head SAR (1g) W/kg @ 20.0 Deviation 1g
(%) dBm 3.59
-0.55%
Measured Body SAR (1g) W/kg @ 20.0 Deviation 1g
(%) dBm 3.62
-0.55%
Object:
D1750V2 SN: 1092 Date Issued:
05/15/2019 Page 2 of 4 Impedance & Return-Loss Measurement Plot for Head TSL Page 3 of 4 Object:
D1750V2 SN: 1092 Date Issued:
05/15/2019 Impedance & Return-Loss Measurement Plot for Body TSL Object:
D1750V2 SN: 1092 Date Issued:
05/15/2019 Page 4 of 4 PCTEST ENGINEERING LABORATORY, INC. 18855 Adams Ct, Morgan Hill, CA 95037 USA Tel. +1.410.290.6652 / Fax +1.410.290.6654 http://www.pctest.com Certification of Calibration Object Calibration procedure(s) Extended Calibration date:
Description:
Calibration Equipment used:
D1900V2 SN: 5d181 Procedure for Calibration Extension for SAR Dipoles. September 07, 2018 SAR Validation Dipole at 1900 MHz. Measurement Uncertainty = 23% (k=2) Name Function Signature Calibrated By:
Sangmin Cha Team Lead Engineer Approved By:
Kaitlin OKeefe Senior Technical Manager Object:
D1900V2 SN: 5d181 Date Issued:
09/07/2018 Page 1 of 4 Manufacturer Model Description Cal Date Cal Interval Cal Due Serial Number Agilent 8753ES S-Parameter Network Analyzer 9/14/2017Annual 9/14/2018US39170118 Agilent N5182A MXG Vector Signal Generator 3/19/2018Annual 3/19/2019US46240505 Amplifier Research 15S1G6 Amplifier CBTN/ACBT343972Anritsu ML2496A Power Meter 10/9/2017Annual 10/9/20181138001Anritsu MA2411B Pulse Power Sensor 11/15/2017Annual 11/15/20181339007Anritsu MA2411B Pulse Power Sensor 11/22/2017Annual 11/22/20181339008Control Company 4040Temperature / Humidity Monitor 2/28/2018Biennial 2/28/2020150761911Control Company 4352Ultra Long Stem Thermometer 2/14/2017Biennial 2/14/2019170112507Keysight 772D Dual Directional Coupler CBTN/ACBTMY52180215 Keysight Technologies 85033E Standard Mechanical Calibration Kit (DC to 9GHz, 3.5mm) 6/4/2018Annual 6/4/2019MY53401181 Mini-Circuits BW-N20W5+ DC to 18 GHz Precision Fixed 20 dB Attenuator CBTN/ACBTN/A Mini-Circuits NLP-2950+ Low Pass Filter DC to 2700 MHz CBTN/ACBTN/A Narda 4772-3 Attenuator (3dB) CBTN/ACBT9406Pasternack PE2209-10 Bidirectional Coupler CBTN/ACBTN/A Pasternack PE5011-1 Torque Wrench 7/19/2017Biennial 7/19/2019N/A SPEAG DAK-3.5 Dielectric Assessment Kit 5/15/2018Annual 5/15/20191070SPEAG ES3DV3 SAR Probe 9/18/2017Annual 9/18/20183287SPEAG DAE4 Data Acquisition Electronics 1/26/2018Annual 1/26/20191533SPEAG ES3DV3 SAR Probe 3/13/2018Annual 3/13/20193131SPEAG DAE4 Dasy Data Acquisition Electronics 3/7/2018Annual 3/7/2019604 DIPOLE CALIBRATION EXTENSION Per KDB 865664 D01, calibration intervals of up to three years may be considered for reference dipoles when it is demonstrated that the SAR target, impedance and return loss of a dipole have remained stable according to the following requirements:
1. The measured SAR does not deviate more than 10% from the target on the calibration certificate. 2. The return-loss does not deviate more than 20% from the previous measurement and meets the required 20dB minimum return-loss requirement. 3. The measurement of real or imaginary parts of impedance does not deviate more than 5 from the previous measurement. The following dipole was checked to pass the above 3 requirements to have 2-year calibration period from the calibration date:
Object:
D1900V2 SN: 5d181 Date Issued:
09/07/2018 Page 2 of 4 Calibration DateExtension DateCertificate Electrical Delay (ns)Certificate SAR Target Head (1g) W/kg @ 20.0 dBmMeasured Head SAR (1g) W/kg @ 20.0 dBmDeviation 1g (%)Certificate SAR Target Head (10g) W/kg @ 20.0 dBmMeasured Head SAR (10g) W/kg @ 20.0 dBmDeviation 10g (%)Certificate Impedance Head (Ohm) RealMeasured Impedance Head (Ohm) RealDifference (Ohm) RealCertificate Impedance Head (Ohm) ImaginaryMeasured Impedance Head (Ohm) ImaginaryDifference (Ohm) ImaginaryCertificate Return Loss Head (dB)Measured Return Loss Head (dB)Deviation (%)PASS/FAIL9/7/20179/7/20181.23.953.991.01%2.062.070.49%53.749.64.14.62.62-24.9-28.9-16.10%PASSCalibration DateExtension DateCertificate Electrical Delay (ns)Certificate SAR Target Body (1g) W/kg @ 20.0 dBmMeasured Body SAR (1g) W/kg @ 20.0 dBmDeviation 1g (%)Certificate SAR Target Body (10g) W/kg @ 20.0 dBmMeasured Body SAR (10g) W/kg @ 20.0 dBmDeviation 10g (%)Certificate Impedance Body (Ohm) RealMeasured Impedance Body (Ohm) RealDifference (Ohm) RealCertificate Impedance Body (Ohm) ImaginaryMeasured Impedance Body (Ohm) ImaginaryDifference (Ohm) ImaginaryCertificate Return Loss Body (dB)Measured Return Loss Body (dB)Deviation (%)PASS/FAIL9/7/20179/7/20181.23.954.134.56%2.092.142.39%49.245.83.45.62.13.5-24.8-24.70.40%PASS Impedance & Return-Loss Measurement Plot for Head TSL Object:
D1900V2 SN: 5d181 Date Issued:
09/07/2018 Page 3 of 4 Impedance & Return-Loss Measurement Plot for Body TSL Object:
D1900V2 SN: 5d181 Date Issued:
09/07/2018 Page 4 of 4 PCTEST ENGINEERING LABORATORY, INC. 18855 Adams Ct, Morgan Hill, CA 95037 USA Tel. +1.410.290.6652 / Fax +1.410.290.6654 http://www.pctest.com Certification of Calibration D1900V2 SN: 5d026 Object Calibration procedure(s) Extended Calibration date:
Description:
Calibration Equipment used:
Manufacturer Agilent Agilent Amplifier Research Anritsu Anritsu Anritsu Control Company Control Company Keysight Keysight Technologies Mini-Circuits Mini-Circuits Narda Pasternack Pasternack SPEAG SPEAG SPEAG Model 8753ES N5182A 15S1G6 ML2495A MA2411B MA2411B 4040 4352 772D 85033E BW-N20W5+
NLP-2950+
4772-3 PE2209-10 PE5011-1 DAKS-3.5 EX3DV4 DAE4 Procedure for Calibration Extension for SAR Dipoles. May 14, 2019 SAR Validation Dipole at 1900 MHz. Description S-Parameter Network Analyzer MXG Vector Signal Generator Amplifier Power Meter Pulse Power Sensor Pulse Power Sensor Temperature / Humidity Monitor Ultra Long Stem Thermometer Dual Directional Coupler Standard Mechanical Calibration Kit (DC to 9GHz, 3.5mm) 6/4/2018 DC to 18 GHz Precision Fixed 20 dB Attenuator Low Pass Filter DC to 2700 MHz Attenuator (3dB) Bidirectional Coupler Torque Wrench Portable DAK SAR Probe Dasy Data Acquisition Electronics CBT CBT CBT CBT 7/19/2017 9/11/2018 7/20/2018 7/10/2018 CBT 10/21/2018 10/30/2018 11/20/2018 2/28/2018 2/28/2018 CBT Serial Number Cal Date Cal Interval Cal Due 10/2/2019 10/2/2018 US39170118 6/15/2019 MY47420837 6/15/2018 Annual Annual N/A Annual Annual Annual Biennial Biennial N/A Annual N/A N/A N/A N/A Biennial Annual Annual Annual CBT 10/21/2019 10/30/2019 11/20/2019 2/28/2020 2/28/2020 CBT 6/4/2019 343972 941001 1207470 1339007 150761911 170330160 MY52180215 MY53401181 CBT CBT CBT CBT 7/19/2019 9/11/2019 7/20/2019 7/10/2019 N/A N/A 9406 N/A N/A 1045 7416 1402 Measurement Uncertainty = 23% (k=2) Name Function Signature Calibrated By:
Parker Jones Team Lead Engineer Approved By:
Kaitlin OKeefe Senior Technical Manager Object:
D1900V2 SN: 5d026 Date Issued:
05/14/2019 Page 1 of 4 DIPOLE CALIBRATION EXTENSION Per KDB 865664 D01, calibration intervals of up to three years may be considered for reference dipoles when it is demonstrated that the SAR target, impedance and return loss of a dipole have remained stable according to the following requirements:
1. The measured SAR does not deviate more than 10% from the target on the calibration certificate. 2. The return-loss does not deviate more than 20% from the previous measurement and meets the required 20dB minimum return-loss requirement. 3. The measurement of real or imaginary parts of impedance does not deviate more than 5 from the previous measurement. The following dipole was checked to pass the above 3 requirements to have 2-year calibration period from the calibration date:
Certificate SAR Target Head (10g) W/kg @ 20.0 dBm 2.11 Certificate SAR Target Body (10g) W/kg @ 20.0 dBm 2.12 Measured Head SAR
(10g) W/kg @
20.0 dBm Deviation 10g
(%) Certificate Impedance Head (Ohm) Real Measured Impedance Head (Ohm) Real Difference
(Ohm) Real Certificate Impedance Head (Ohm) Imaginary Measured Impedance Head (Ohm) Imaginary Difference
(Ohm) Imaginary Certificate Return Loss Head (dB) Measured Return Loss Head (dB) Deviation (%) PASS/FAIL 2.05
-2.84%
53 54.2 1.2 8 4.1 3.9
-21.6
-25.8
-19.30%
PASS Measured Body SAR
(10g) W/kg @
20.0 dBm Deviation 10g
(%) Certificate Impedance Body (Ohm) Real Measured Impedance Body (Ohm) Real Difference
(Ohm) Real Certificate Impedance Body (Ohm) Imaginary Measured Impedance Body (Ohm) Imaginary Difference
(Ohm) Imaginary Certificate Return Loss Body (dB) Measured Return Loss Body (dB) Deviation (%) PASS/FAIL 2.04
-3.77%
47.1 48.4 1.3 7.4 3.5 3.9
-21.8
-25.1
-15.10%
PASS Certificate SAR Target Head (1g) W/kg @ 20.0 dBm 4.02 Certificate SAR Target Body (1g) W/kg @ 20.0 dBm 3.99 Calibration Date Extension Date Certificate Electrical Delay (ns) 5/14/2018 5/14/2019 1.199 Calibration Date Extension Date Certificate Electrical Delay (ns) 5/14/2018 5/14/2019 1.199 Measured Head SAR (1g) W/kg @ 20.0 Deviation 1g
(%) dBm 3.98
-1.00%
Measured Body SAR (1g) W/kg @ 20.0 Deviation 1g
(%) dBm 3.97
-0.50%
Object:
D1900V2 SN: 5d026 Date Issued:
05/14/2019 Page 2 of 4 Impedance & Return-Loss Measurement Plot for Head TSL Page 3 of 4 Object:
D1900V2 SN: 5d026 Date Issued:
05/14/2019 Impedance & Return-Loss Measurement Plot for Body TSL Page 4 of 4 Object:
D1900V2 SN: 5d026 Date Issued:
05/14/2019 PCTEST ENGINEERING LABORATORY, INC. 18855 Adams Ct, Morgan Hill, CA 95037 USA Tel. +1.410.290.6652 / Fax +1.410.290.6654 http://www.pctest.com Certification of Calibration Object Calibration procedure(s) Extended Calibration date:
Description:
Calibration Equipment used:
D2600V2 SN: 1069 Procedure for Calibration Extension for SAR Dipoles. September 10, 2018 SAR Validation Dipole at 2600 MHz. Measurement Uncertainty = 23% (k=2) Name Function Signature Calibrated By:
Sangmin Cha Team Lead Engineer Approved By:
Kaitlin OKeefe Senior Technical Manager Object:
D2600V2 SN: 1069 Date Issued:
09/10/2018 Page 1 of 4 Manufacturer Model Description Cal Date Cal Interval Cal Due Serial Number Agilent 8753ES S-Parameter Network Analyzer 9/14/2017Annual 9/14/2018US39170118 Agilent N5182A MXG Vector Signal Generator 3/19/2018Annual 3/19/2019US46240505 Amplifier Research 15S1G6 Amplifier CBTN/ACBT343972Anritsu ML2496A Power Meter 10/9/2017Annual 10/9/20181138001Anritsu MA2411B Pulse Power Sensor 11/15/2017Annual 11/15/20181339007Anritsu MA2411B Pulse Power Sensor 11/22/2017Annual 11/22/20181339008Control Company 4040Temperature / Humidity Monitor 2/28/2018Biennial 2/28/2020150761911Control Company 4352Ultra Long Stem Thermometer 2/14/2017Biennial 2/14/2019170112507Keysight 772D Dual Directional Coupler CBTN/ACBTMY52180215 Keysight Technologies 85033E Standard Mechanical Calibration Kit (DC to 9GHz, 3.5mm) 6/4/2018Annual 6/4/2019MY53401181 Mini-Circuits BW-N20W5+ DC to 18 GHz Precision Fixed 20 dB Attenuator CBTN/ACBTN/A Mini-Circuits NLP-2950+ Low Pass Filter DC to 2700 MHz CBTN/ACBTN/A Narda 4772-3 Attenuator (3dB) CBTN/ACBT9406Pasternack PE2209-10 Bidirectional Coupler CBTN/ACBTN/A Pasternack PE5011-1 Torque Wrench 7/19/2017Biennial 7/19/2019N/A SPEAG DAK-3.5 Dielectric Assessment Kit 5/15/2018Annual 5/15/20191070SPEAG ES3DV3 SAR Probe 9/18/2017Annual 9/18/20183287SPEAG DAE4 Data Acquisition Electronics 1/26/2018Annual 1/26/20191533SPEAG ES3DV3 SAR Probe 4/12/2018Annual 4/12/20193275SPEAG DAE4 Dasy Data Acquisition Electronics 4/12/2018Annual 4/12/2019501 DIPOLE CALIBRATION EXTENSION Per KDB 865664 D01, calibration intervals of up to three years may be considered for reference dipoles when it is demonstrated that the SAR target, impedance and return loss of a dipole have remained stable according to the following requirements:
1. The measured SAR does not deviate more than 10% from the target on the calibration certificate. 2. The return-loss does not deviate more than 20% from the previous measurement and meets the required 20dB minimum return-loss requirement. 3. The measurement of real or imaginary parts of impedance does not deviate more than 5 from the previous measurement. The following dipole was checked to pass the above 3 requirements to have 2-year calibration period from the calibration date:
Object:
D2600V2 SN: 1069 Date Issued:
09/10/2018 Page 2 of 4 Calibration DateExtension DateCertificate Electrical Delay (ns)Certificate SAR Target Head (1g) W/kg @ 20.0 dBmMeasured Head SAR (1g) W/kg @ 20.0 dBmDeviation 1g (%)Certificate SAR Target Head (10g) W/kg @ 20.0 dBmMeasured Head SAR (10g) W/kg @ 20.0 dBmDeviation 10g (%)Certificate Impedance Head (Ohm) RealMeasured Impedance Head (Ohm) RealDifference (Ohm) RealCertificate Impedance Head (Ohm) ImaginaryMeasured Impedance Head (Ohm) ImaginaryDifference (Ohm) ImaginaryCertificate Return Loss Head (dB)Measured Return Loss Head (dB)Deviation (%)PASS/FAIL9/11/20179/10/20181.1525.695.52-2.99%2.542.51-1.18%49.148.90.2-6.1-4.91.2-24.1-25.8-7.10%PASSCalibration DateExtension DateCertificate Electrical Delay (ns)Certificate SAR Target Body (1g) W/kg @ 20.0 dBmMeasured Body SAR (1g) W/kg @ 20.0 dBmDeviation 1g (%)Certificate SAR Target Body (10g) W/kg @ 20.0 dBmMeasured Body SAR (10g) W/kg @ 20.0 dBmDeviation 10g (%)Certificate Impedance Body (Ohm) RealMeasured Impedance Body (Ohm) RealDifference (Ohm) RealCertificate Impedance Body (Ohm) ImaginaryMeasured Impedance Body (Ohm) ImaginaryDifference (Ohm) ImaginaryCertificate Return Loss Body (dB)Measured Return Loss Body (dB)Deviation (%)PASS/FAIL9/11/20179/10/20181.1525.535.28-4.52%2.482.35-5.24%46.244.71.5-4.7-8.23.5-24.1-21.311.60%PASS