Instruction Manual Product Description External probe socket in models with external battery, follow up the +/- sign on PCB. Bluetooth Data Logger Introduction Congratulations on your purchase of this bluetooth datalogger. This datalogger is probe port START/STOP (ON/OFF) key:
LCD (Liquid Crystal Display). After installing the batteries, press key to power on the logger. LCD displays for 2 seconds. Press again to turn off. After set by App, if the logger is programmed as Key start, press the Start/Stop button for 5 a. Measured temperature or humidity or seconds to start logging. if the logger is pressure. LCD updates and interexchange every programmed as Key stop, press the Start/Stop 5 seconds. button again for 5 seconds to stop logging. In the designed for monitoring temperature, humidity b. REC flashes every second when logger is in logging mode, short press key wont power off and pressure subject to quality control logging mode. While logger is programmed with the logger. requirement. start delay, REC displays but doesnt flash MAX/MIN key:
Temperature, humidity and pressure during the standby status. Press key to check the MAX or MIN data of all measurements are saved throughout the entire c. T1 is air temperature, T2 is external probe logged data. duration of the logging period. This datalogger is equipped with Bluetooth temperature Magnet mounting d. H or L displays when measuring set limit value 3 magnets on the rear side to fix the logger on communication for programming purpose. A free is exceed. ventilation shaft or iron rack. download Bluetooth iOS and Android APP is LED indicator:
e. MX or MI display when checking MAX or MIN Hanger required. The measurement report output a PDF Low/High: Red LED blinks every 5 seconds value from logged data. file and a csv file generated by APP. when the measuring set limit value is exceeded f. Low battery icon flashes on display every Operation Read through the instruction manual before during record. second when battery level is too low to NOTE:
using this logger and APP. The logger is OK: When the logging function is started, the accurately measure and log data. 1. Smart phone with Bluetooth is required. calibrated before shipment. green LED blinks once every 5 seconds when g.The real time value is broadcasted to APP iOS above 10 or Android above 5.0 are also a there is no alarm occurs. every 5 seconds. must to run programming APP smoothly. NTC thermistor for temperature measuring Battery cover, operated by 1 pc CR2 battery. 2. Programming APP can be download for and capacitive humidity sensor for humidity Use coin and follow up the direction indicator to free. Please search for Nice Logger in measuring. open and close the cover. While inserting the App Store (iOS) or in Play Store (Android) . Step1 Configuring the data logger interval should be 10 sec. 20sec. 30sec. and so till time is up. It is possible to make configuration changes at anytime before the logging function is started. on Start & Stop Mode D. Memory full or on Date/Time: Means user program a desire time point to stop logging. Once the logger is started, configuration Select how you would like to Start and Stop the Before reaching the stop time, if the memory is changes cannot be made unless stopping the logger. There are 2 choices for start logging and full, the logging stop as well. logging first. If the logger is locked with a Password, the Password is required to make configuration changes. Main Operation Functions are Here!
4 choices for stop logging, no matter which Start delay method is picked up, user can always stop Select the start delay from 0 min to 24 hours. For logging from APP. example: If the delay is 10 minutes, and the
- Press ON/OFF to power on the logger. Start choices:
sampling rate is 20 minutes, the real time to log
- Turn ON the Bluetooth of smart device & APP. A.By key start: Means pressing key to start the first temperature measurement is 10 minutes APP will auto scan available and recognized loggers nearby.3 loggers can be displayed in logging after set by App after the START button is pressed. All B.By on time start: Means user program a measurements after the first measurement will one page. Click gray dot to switch to other page. desire time point to start logging after set by be at a 20 minute (or selected) interval. Press the desired logger to link for further programming. Rate, start/stop, App. alarm setting are Stop choices:
The increment of the start delay is 10 minutes, so the start delay should be 10 min. 20 min. 30 here!
A. Memory loop till key stop: Means pressing min. and so on key to stop logging. When the memory is full Alarm types before user pressing stop key, the new data will Regardless of the Alarm Type, if the red LED is overwrite the old one till user pressing key to triggered, it wont stop even the reading return to Available Loggers are listed there. Configure end. normal range or logging is stopped. To stop red B. Memory full or key stop: Means pressing key alarm LED, logging must be stopped and then 3 loggers per page In Configure, you can program logger with to stop logging. When the memory is full, the connect logger to APP to read out data or desired way to start/stop logging and things logging will stop even user didnt press key to re-config. about alarm. Sampling rate stop. There are two types of alarm types, single and C. Memory loop till Date/Time: Means user cumulative. However, while users dont choose
- User programmable parameters are grouped Select the sampling interval you need from 10 program a desire time point to stop logging. any parameter in threshold value setting, it will as two sections in APP. One is Configure and seconds to 2 hours, the increment of the When the memory is full but not yet reaching be treated as NO alarm. the other one is Logger ID. sampling interval is 10 seconds, so the sampling stop time, the new data will overwrite the old one Single event: an alarm is triggered immediately when the measured value exceeds the alarm For example: Temp. is marked, High is 8 and for temperature and inHg for pressure.
-REC will appear and flash on LCD to indicate threshold. Low is 2 , it means that an alarm condition will Time zone the logging is activated. Measured parameters Cumulative: an alarm is not triggered when the occur below 2 , or above 8 . The Before programming the logger the user must update and interexchange on LCD every 5 measured value exceeds the alarm threshold, programmable alarm limit of each parameter is assure that the smart phone is set for the correct seconds. but only once the overall average value during limited to one decimal. time zone. The logger will auto synchronize to
- If Logger is programmed with Start delay, the alarm delay duration exceeds the alarm Once all the programming is done, press "Save"
the time zone of the smart phone, when Save is REC will appear (not flashing) to show the threshold. While choosing cumulative, alarm to confirm the setting and then you may continue pressed. Time zone changes over the transit logging is started and in standby status. delay time is required to input. The adjustable to link another logger to program next one. distance are not adjusted in the logged data.
-During the logging, the green LED will blink alarm delay interval for a cumulative alarm type can be 5 minutes to 12 hours, the increment of Logger ID the cumulative interval is 5 minutes, so the start In Logger ID, you can program logger with delay should be 5 min. 10 min. 15 min. and so specific identification to distinguish it from other on Alarm Threshold logger. Password Select the parameter and input threshold values. The Password function is default OFF. The user every 5 seconds if there is no alarm occurred. If any, it turns into flashing red LED.
-If the red LED is triggered, it wont stop even the reading return to normal range or logging is stopped. To stop red alarm LED, logging must be stopped and then connect logger to APP to read out data or re-config. Password, logger name &
unit are programmed here!
While users dont choose any parameter in may enable it to prevent unauthorized
-Press MAX/MIN key any time to review the threshold value setting, it will be treated as NO reprogramming on App. A Password may have maximum and minimum data stored in memory. alarm. up to 4 numeric characters (0, 1, 2, .9) If you didnt tick any parameter here, it will be treated as NO alarm. Device Name Once all the programming is done, press "Save"
Step3 Download data A user defined name, or descriptor, can be input to confirm the setting and then you may continue
- This APP and logger provides you an under Device Name. It will be displayed on the to link another logger to program next one. advanced design to review logged data graph report as the Title, with a maximum of 7 Step2 Start logging before logging task is ended. It means you can characters.
- Depended on what kind of start mode you pick readout the logged data anytime without Unit of Measurement (UoM) up, you may ress "START" key for 5 seconds to stopping the logging. Select the unit that will be displayed on the start the logging. Logger programmed with
-Of course you may still choose to stop the report and LCD. The selections are Metric or starting time will automatically start at desired logging task first before downloading data. Imperial. In Metric, it is Celsius for temperature time. Or, you may start logging any time from Depended on what kind of stop mode you pick and hpa for pressure. In Imperial, it is Fahrenheit APP. up in configure, you may press "STOP" key for 5 seconds to stop the logging. Logger programmed with stop time will automatically stop at desired time. Or, you can stop logging anytime from APP.
- After stopping the logging (or without stopping it), choose Readout to download the logged data from logger. Step4 Report generation
-Go to Files function of APP and input the serial number of the logger which you want to generate the report.
-Click the displayed simplified graph to check the detail statistic data.
- Choose function Convert to PDF or Convert to csv to generate the report in preferred format.
- The created csv report contains all data shown in PDF report except the graph.
- The created report can be mailed out to share data. Click to generate report in PDF or csv. Can be saved or send out via email Input s/n here to Model Technical data search read out data Click graph to see details Model 1 : Temperature logger Model 2: Temperature + Ext. temp. logger Model 3: Temperature + Humidity logger Model 4: Temperature +Humidity +Barometric logger Temp. & Ext. By NTC thermistor, measurement -30.0~70.0oC (-22.0~158.0oF) Temp. resolution 0.1oC (0.1oF) Temp. accuracy +/-0.5oC Humidity 0.1~99.9%rH Humidity res. 0.1%rH Humidity accu. +/-3% at 25 oC(10-90%rH), others
+/-5%
Barometric 300~1100hpa; 8.9~32.5inHg Baro res. 1hpa;0.1inHg Baro accu.(hpa) +/-3(0~60 oC), +/-5 (-20~0 oC), the Error Code
- - - - . -
Why?
In model 2 only. External probe is unplugged Solution Check the contact between probe and logger E02 Why?
Measured value is below specified range Solution Put the logger in specified range E03 Why?
Measured value is above specified range Solution Put the logger in specified range E04 Why?
In humidity display of model 3 & 4 only, means temperature is in error mode Solution Solve the error issue of temp. sensor E11 rest is not specified. Why?
In humidity display of model 3 &4 only, Logging Type Multiple use Sampling points Model 1: 48000 points Model 2: 24000 Temp. +24000 ext. Temp. Model 3: 24000 Temp. +24000 Humidity means humidity calibration is fail Solution Re-calibrate the humidity E31 Why?
Microprocessor is fail Solution Contact distributor for after service Model 4: 16000 Temp.+ 16000 Humidity+ 16000 Baro. E33 Battery Life 3 months Operating temp. -30~70oC (Logging status); room temp.(PC status) Operating RH% Humidity < 80%
Storage temp. -40~85oC Storage RH% Humidity <90%
Weight ~90g Battery 1PC 3.0V CR2 Sampling interval 10 second to 2 hours Start delay 0 mins to 24 hours Alarm range Temp: -30.0~70.0oC Humidity:1~99%rH Baro:300~1100hpa Alarm delay 0, 5, 10, 15720 minutes Alarm type Single, Cumulative, Disable Operation keys 2 Keys, Start/Stop & MX/Mn LED indicator REC, High /Low alarm Protection class Model 1: IP67 Model 2: IP65 Model 3: IP65, not include sensor Model 4: IP65, not include sensor. Directives EN12830 Operating System iOS and Android Why?
In humidity display of model 3&4 only, means humidity measuring is fail Solution Contact distributor for after service Why red LED is flashing but recorded data is OK Why?
Alarm detection works every 5 seconds but data logging are possibly programmed as every 2 hours. In this condition, if alarm happens between two logging points, it might lead the recorded data is OK but red LED is flashing, especially if the alarm mode is set as Single. Solution To program the alarm function as cumulative can reduce this situation FCC Caution: Any changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate this equipment.This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. RF exposure warning The equipment complies with FCC RF exposure limits set forth for an uncontrolled environment. The equipment must not be co-located or operating in conjunction with any other antenna or transmitter. to the residential to provide FEDERAL COMMUNICATIONS COMMISSION INTERFERENCE STATEMENT This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed reasonable protection against harmful interference in a installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance instructions, may cause with harmful interference radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this harmful equipment interference television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
-Consult the dealer or an experienced radio/ TV technician for help. does cause to radio or