Operation Manual BM1000 Pulse Oximeter BERRYMED LIMITED Version: 1.0 Product Description BM1000 Fingertip Pulse Oximeter is an important and common device to check oxygen saturation
(SpO2) and pulse rate. Its a small, compact, simple, reliable and durable physiological monitoring device. Include the mainboard, OLED display and dry batteries. It is suitable for monitoring adults and children. It can be used in the hospitals operation room, ICU, clinic section office, out-patient department, sickroom and emergency treatment. It can also be used in the recovery and health care organizations, th e community m edical treatments, th e o xygen bars, t he fam ily n ursing, the physical care in sports (you can use it before or after the sport, but it is not recommended to use it during the sport). Measurement Principle An experience formula of data process is established taking use of Lambert Beer Law according to Spectrum A bsorption Characteristics of hemoglobin ( Hb) and O xyhemoglobin ( HbO2) in glow and near-infrared zo nes. O peration principle of th e i nstrument is Ph otoelectric O xyhemoglobin Inspection Technology is ad opted in a ccordance wi th Capacity Pu lse Sc anning an d R ecording Technology, so that two beams of dif ferent wavelength of lights (660nm glow and 940nm near infrared light) can be focused onto human nail tip through perspective clamp finger-type sensor. Then measured signal can be obtained by a photosensitive element, information acquired through which w ill be show n o n tw o gro ups of LED s th rough pr ocess in e lectronic c ircuits a nd microprocessor. Arterial oxygen s aturation is m easured by a m ethod called p ulse oximetry. It i s a continuous, non-invasive method based on the different spectra absorption of hemoglobin and oxyhemoglobin
(called spectrophotometer principle). It measures how much light, sent from light sources on the other side. Diagram of Operation Principle Safety Information
The person who uses the pulse oximeter must receive adequate training before use.
The pulse oximeter is intended only as adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms. It is not intended as a device used for treatment purposes.
When using the pulse oximeter together with the electrical surgery equipment, the user should pay attention to and guarantee safety of the patient being measured.
EXPLOSION HAZARD: Do not use the pulse oximeter in the presence of flammable anesthetics, explosive substances, vapors or liquids.
Make sure not to use the pulse oximeter during MRI (magnetic resonance imaging) scanning or CT (Computed Tomography) environment because induced current could potentially cause burns.
The pulse oximeter is without alarm function. Continuous monitoring for a long time is not suitable.
No modification of this product is allowed. Maintenance should be operated by professional maintenance personnel who are approved by manufacturers.
Please shut off the power before clean the pulse oximeter. Never permit high-pressure and high-temperature disinfection of the device. Never use cleaning agents/disinfectants other than the recommended.
The product is commonly seal product. Keep its surface dry and clean, and prevent any liquid from infiltrating it.
The pulse oximeter is precision and fragile. Avoid pressure, knock, strong vibration or other mechanical damage. Hold it carefully and lightly. If it is not in use, it should be appropriately placed.
For disposal of pulse oximeter and accessories, follow local regulations or your hospitals policy regarding disposal of such pulse oximeter and accessories. Do not dispose randomly.
Use AAA alkaline batteries. Do not use carbon or poor quality batteries. Remove the
batteries if the product is not to be used for a long time. If patient is an intended operator, you must read the operation manual carefully and understand deeply or consult with the doctor and manufacturer before using. If you have any discomfort in use, please stop using immediately and go to the hospital.
A functional tester cant be used to assess the accuracy.
The patient is an intended operator and can perform the maintenance the equipment. Product Feature 1. Simple and convenient usage of product, simple one-touch operation. 2. Small volume, light weight, convenient to carry. 3. Lower consumption, original two AAA batteries can continuously work for 24 hours. 4. Low voltage reminder shows in screen when theres low battery, may influence the normal working. 5. The machine will automatically power off when theres no signal generated. 6. Daily maintenance and calibration is unnecessary. 7. Communication c an be real ized between the pr oduct and m obile phone w ith i ts w ireless Bluetooth. Display Introduction Figure 1 Battery Installation 1. Hold the product in one hand with the front panel facing the palm. Put the other hands big finger on the battery cabinet lids press sign, press downwards and push the lid open at the same time. The battery cabinet is opened as shown in Figure 2. Install batteries into the slots per the + and - symbols as shown in Figure 3. Cover the lid onto the cabinet and push it upwards to make it close well. 2.
The positive and negative electrodes of batteries should be installed correctly. Otherwise the device will be damaged.
When install or remove batteries, please follow the correct operation sequence to operate. Otherwise the battery compartment will be damaged. Figure 2 Figure 3 Hang lace Installation 1. Thread thinner end of the hang lace through the hanging hole. 2. Thread thicker end of the lace through the threaded end before pulling it tightly. Directions for use 1. Press Clips press sign in the Figure 1 and open the clip. Let the testees finger put into the rubber cushions of the clip, make sure the finger is in the right position as shown in Figure 4, and then clip the finger. 2. Press the POWER/FUNCTION" switch but ton on the front pa nel to turn on the pro duct. Using first finger, middle finger or ring finger when doing test. Dont shank the finger and keep the testee at case during the process. The readings will be displayed on the OLED screen a moment later as shown in Figure 5. Display screen data refresh time for one second. Figure 4 Figure 5 NOTE:
Dont put the product on extremities with arterial catheter or venous syringe.
Cover the sensor with opaque material under the condition of strong light. Failure to do so will result in inaccurate measurement.
Try to keep the patient still (specially the arm) and avoid the measured site suffering excessive motion.
Dont use the product to measure patients whose pulse rate is lower than 30bpm, which may cause incorrect results.
Make sure no contamination or scar exists in the size where the product is placed. Otherwise, the measured result may be incorrect because the signal received by the sensor is affected.
When used on different patients, the product is prone to crossed contamination, which should be prevented and controlled by the user. Disinfection is recommended before using the product on other patients.
Prolonged use or the patients condition may require changing the sensor site periodically. Change sensor site and check skin integrity, circulatory status, and correct alignment at least even 2 hours.
Dont perform SpO2 monitoring and NIBP measurements on the same arm simultaneously. Obstruction of blood flow during NIBP measurements may adversely affect the reading of the SpO2 value.
Testees fingernail cant be too long. Otherwise the finger cant be inserted into the sensor to a suitable depth and the SpO2 measurements may be inaccurate.
Make sure to place the product on the finger in a correct direction. The LED part of the sensor should be at the backside of the patient hand and photodetector part at the inside. Make sure to insert the finger to suitable depth into the sensor so that the fingernail is just opposite to the light emitted from the sensor.
The highest temperature of sensor contacts with patients skin dont be allowed more than 41.
Shock, anemia, hypothermia and the application of vasoconstriction drug may decrease arteria blood flow to an unmeasurable level.
Pigment, or deep color (for example: nail polish, artificial nails, dye or pigmented cream) may cause inaccurate measurements. Function Description a. When the data has been displayed on the screen, change the display direction by pressing the POWER/FUNCTION button again. (as shown in Figure 6) b. When th e product is powered o n, long press the POWER/FUNCTION but ton, Bluetooth function will be started. The Bluetooth indicator light on the top of display will flicker. (as shown in Figure 7) c. The product will a utomatically be powered off when no fi nger is in the device for longer than 10 seconds. And switch to another display mode. (as shown in Figure 8) Figure 6 Figure 7 Figure 8 Bluetooth Communication Function a. Open the mobile phone, and double-click SpO2 software icon following interface. (as shown in Figure 9)
. And enter the b. Click the A dd new users s and enter th he following interface. (as s shown in Fi igure 10) Figure 9 c. er name, click Input the use in Figure 11)
) k the Detail Figure 10 s and jump the following g information n table. (as sh hown Figure 11 Input person Figure 12) nal informatio on, then click k Save and enter the foll lowing interf face. (as show wn in Figure 12 Click berry and enter th he following i interface. (as shown in Fig gure 13) d. e. Figure 13 f. Click
, search equ uipment and j jump the foll lowing interfa face. (as show wn in Figure 1 14) Figure 14 g. Wait to sear shown in Fig rch out the e gure 15) equipment, c l lickBerryMe ed and ente er the follo w wing in terface e. (as Figure 15 h. ClickPair, Figure 16) to p air wi th hBerryMed a nd e nter t the follow ing g test interf a ace. (a s sho w wn in Figure 16 i. If you wan t following wi t to e xit the indow. And th software, c l hen clickOK lick escape k K, exit the so key of your oftware. (as sh mobile pho hown in Figu ne a nd jum p ure 17) p the Figure 17 Cleaning a and Disinfe ection otton or soft c he water with dry. nts include: et yde( 2%) solu ructed above. ith cot ton or oduct with co ng, wipe off th oduct to air d aning Clean the pro After cleanin Allow the pr nfection recommende ed disinfectan glutaraldehy ropanol70%, oduct as instr Clean the pro Disinfect th e e prod uct wi
. disinfectants After disinfe ection, be s ur with water. moistened w Allow the sen nsor to air dr ry. ed with water cloth moistene r. h a soft cloth. thanol70%, ution disinfec tants. ximeter gertip pulse o g lace operation ma anual The sta ndard config guration 1pc 1pc 1pc C Clea 1. 2. 3. Disi The isop 1. 2. 3. 4. P Fing Han The T 1. 2. r soft cloth m moistened w with on e of t the recomm e ended re to wipe o f ff the disinfe ectant left on n the product with a soft cloth Packing Li st Technical S Specificat tions Display mod SpO2:
de: OLED Measurement range: 35~100%
Accuracy: 2%80%~100%; 3%70%~79%
3. Pulse Rate:
Measurement range: 25~250bpm Accuracy: 2bpm 4. Electrical specifications:
Working voltage: D.C.2.2 V~D.C.3.4V Battery Type: Two 1.5V AAA alkaline batteries Power consumption: smaller than 50mA 5. Product specifications:
Size: 58 (H) 34 (W) 30(D) mm Weight: 50g (include two AAA batteries) 6. Environment requirements:
Temperature:
Operation: +5~+40 Transportation and storage: -10~+50 Humidity:
Operation: 15%~80%noncondensing Transportation and storage: 10%~90%noncondensing Atmospheric pressure:
Operation: 860hPa~1060hPa Transportation and storage: 700hPa~1060hPa Troubleshooting Possible reason solution is n ot properly 1. Please the finger pr operly and try Trouble The SpO2 and P R cant be displayed d normally an the v alue disappeared. 1. The fi nger positioned. 2. The p atients SpO2 is to o l ow to be detected. 3. Bluetooth signal is interrupted. again. 2. Try ag ain; Go to a ho spital fo r a diagnosis if you are sure the device works all right. 3. Check t he Bluetooth co nnection and reconnect. The SpO2 and P R display instable. 1. The finger is not placed inside 1. Place t he f inger properly and try enough. again. 2. The fi nger is sh aking o r the 2. Let the testee keep calm. The device be powered on. cant The scre suddenly off. en is testee is moving. 1. The batt eries are drai ned or almost drained. 2. The ba tteries are not i nserted properly. 3. The devices malfunction. 1. The product is a utomatically powered off when n o si gnal is 1. Change batteries. 2. Reinstall batteries. 3. Please contact the supplier. 1. Normal. 2. Replace the batteries. detected longer than seconds. 10 2. Power quantity of the ba tteries is exhausted. Symbol Meaning Symbol Meaning CAUTIOUS! Please refer to the operation manual. Type BF Equipment. The product does not contain alarm function. When the end-user wishes to discard this product, it must be sent to separate collection facilities for recovery and recycling. Information of manufacture, including name and address. Date of manufacture. European Union for approval. Serial Number. Batch Code. Type Number. The European Union authorized. IP21 The product is protected against harmful effects of dripping water per IEC 60529. Shanghai Berry Electronic Tech Co., Ltd Suit C, 1st Floor, 7th Building, N0.1188 Lianhang Road, Minhang District, Shanghai, China 201112 TEL: 0086-21-5853 1958 FAX: 0086-21-5853 0468 www.berry-med.com Kingsmead Service limited (UK) 145-157 St John Street, London, EC1V 4PY (UK) TEL 044-20-7193 9159 F AX: 044-20-7193 9159 www.kingsmead-service.com If you need additional information, please contact with the company. FCC warning statements:
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. Caution: Any changes or modif cations to this device not explicitly approved by manufacturer could void your authority to operate this equipment. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. IC warning statements:
-English Warning Statement:
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device."
The digital apparatus complies with Canadian CAN ICES-3 (B)/NMB-3(B).
-French Warning Statement:
Le prsent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorise aux deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radiolectrique subi, mme si le brouillage est susceptible d'en compromettre le fonctionnement."
Cet appareil numrique de la classe B est conforme la norme NMB-003 du Canada. This equipment complies with IC radiation exposure limits set forth for an uncontrolled environment and meets RSS-102 of the IC radio frequency (RF) Exposure rules. This equipment has very low levels of RF energy that are deemed to comply without testing of specifc absorption ratio (SAR).Cet quipement est conforme aux limites dexposition aux rayonnements nonces pour un environnement non contrl et respecte les rgles dexposition aux frquences radiolectriques (RF) CNR-102 de lIC. Cet quipement met une nergie RF trs faible qui est considre conforme sans valuation du dbit dabsorption spcifque
(DAS).