Hao-Hao Intelligent Pregnant Preparation Instrument TM Product Name: Comper Smart Fertility Tracker Product Model: BBTX COMPER TM Content Basic Function and Working Principle Composition Safety Precautions Application Method Instructions for use Disinfection function Charging function Prompt function Voice Prompt Cleaning and maintenance Problems and Troubleshooting Storage and Shipping Condition Basic Parameters and Specifications The Registration Certificate Number and Production License Number Service Agencies and Information 1 Basic Function and Working Principle Dear customers, Thank you for purchasing Comper Intelligent Pregnant Preparation Instrument TM (Product Name: Medical electronic thermometer, hereinafter referred to as Pregnant Preparation Instrument). For appropriate application of the product, please read the operating manual before using. Keep it safely after reading for reference. Illustrations in the Manual are effect pictures. Instrument Basic function of the product Pregnant Preparation is a mouth-contained basal body temperature monitoring device for women. Body temperature can be accurately measured by electronic sensors, satisfying the demand of daily basal body temperature monitoring of women, and is satisfied the requirement of daily monitoring by laypersons, nursing staff and care givers in obstetrical department in hospital, clinics and pregnant women at home. Application scope: monitor the ovulation cycle Working principle A temperature sensor outputs electric signals and then electric signals will be converted into digital signals that can be identified by internal integrated circuit. Temperature value will be obtained by the processing and computing of digital signals by internal chips. 2 Composition 1 4 2 3 5 6 Manual Qualification certificate Parts list:
1. One pregnant preparation instrument
(thermometer) 2. One disinfection charging box 3. One USB charging cable 4. One manual 5. One qualification certificate 6. Hao-Hao Health APP 3 3 4 7 4 8 particles 5. Disinfection charging box interface 6. Charging indicator 7. Contact-based charging connector 8. UV disinfection lamp 1 2 6 5 Instruction of parts:
1. Pregnant preparation instrument indicator lamp 2. Heat probe 3. Contact-based charging interface 4. The anti-skid Safety Precautions In the Manual, all possible harms or damages caused by operation not according to instructions have been indicated and explained in accordance to the following classification. Please be aware and understanding that, for all death, injury or damage to goods due to operations not according to instructions specified in the Manual or deviating from the designed purpose of the product, the Company shall not bear any responsibility. Refer to instruction manual/ booklet Prohibition It represents absolutely not allowed There is death or injury risk caused by operation that is out of the performance of the product or appropriate. Warning Represent matters that require special attention Represent matters that require special attention, which may cause death or serious disease even in normal operation. Caution Represent general considerations in using If misused, it may cause personal injury or property damage Prohibition Do not use the product in flammable environment to protect from fire or explosion. Do not dismount, dismantle or repair the product voluntarily.
* There is possibility of causing serious body injury, measurement error and damages to the product. 5 Warning:
Keep the product out of the reach of infants and young children, and do not let it use the product alone.
* If misused, there is the possibility of causing damages When the electronic device is limited in hospital or aircraft cabin, please do not turn on the product. Prohibition:
It is dangerous to conduct self-diagnosis or self-treatment in accordance to measurement results, please follow the guidance of a doctor
* Self-diagnosis may cause possibility of deterioration of diseases This product is an oral dedicated thermometer, and is not allowed to use in any parts other than the mouth (especially the parts of the ear) to measure body temperature.
* It is unable to correctly measure the temperature, and there is possibility of causing harm to the ears Do not share this product with others
* There is possibility of causing a bacterial infection or transmission of the virus. Do not damage the battery, heat the battery or throw it in the fire
* There is possibility of a battery explosion. 6 Do not vigorously bite the product.
* May cause product breakdown. Please contact your physician if the condition of swallowing occurs. Do not place the product 22cm within the scope closely proximity to other medical instruments (such as heart pacemakers) Do not look directly at disinfection lamp
* There is possibility of causing damage to eyes Caution:
Make sure the positive and negative side of prepare pregnant preparation instrument when putting into the disinfection box
* Incorrect placement may cause possibility of damages to product. Do not use the product closely to strong electromagnetic fields, electromagnetic wave and magnetic environment
* There is possibility of measurement errors or damage to the product Avoid severe shock, falling from the height or stomping after putting into the disinfection box
* There is possibility of causing damage to the product. Do not make the charging part getting wet
* There is possibility of causing damage to the product. 7 Safety Precautions Dismantlement is prohibited. Do not place the product into boiling water for sterilization
* There is possibility of causing damage to product breakdown or product failure. Do not dismount, dismantle or repair the product voluntarily.
* There is possibility of causing serious body injury, measurement error and damages to the product Do not take out the magnet from disinfection charging box.
* Magnet damage may cause possibility of damage to product. At the end of product life, please deal with the product according to the relevant local environmental protection laws and regulations. Scope of application and contraindications Application scope: monitor the ovulation cycle Contraindications: Not found 8 Application Method Pregnant preparation instrument should be used with APP. 1. Download by scanning QR code. 2. Download from the application store: search Hao-Hao Health Note:
Piggyback Bluetooth 4.0 and above module IOS7.0 and above operating system Piggyback Bluetooth 4.0 and above module Android 4.3 and above operating system 9 Instructions for Use Note: The instrument should be charged for at least 3 hours in first use. 1. Open the Bluetooth of mobile phone. 2. Open the Hao-Hao Health APP and create an account. 3. Select device of Hao-Hao Health Intelligent Pregnant Preparation Instrument, and conduct subsequent operations according to APP tips. For questions, please contact customer service. 9 Disinfection function When putting the pregnant preparation instrument back into disinfection charging box after measurement, pregnant preparation instrument will automatically shut down. Please connect with the power and close the disinfection charging box, the function of disinfection will turns on automatically. 10 Charging Function 1. Put the pregnant preparation instrument in the disinfection charging box. 11 2. Insert the USB charging cable into the charging interface of disinfection charging box. 12 3. Lamp displays orange light when it is in charging. Lamp displays white light when charging is completed. 13 Prompt function Charging indicator lamp Long white light Long orange light Pregnant preparation instrument indicator lamp Long white light & one time of beep Flicker white light Long white light & three times of beep Light turned down Orange light Blue light White light Charging is completed. In charging The main body turns on. Successful connection of Blueteeth. No measurement/temperature is too low or too high, or abnormal measurement Electric quantity exhaustion Period of easy pregnancy Period of non-easy pregnancy No measurement/results 15 Voice Prompt 1 x 1. One time of beep after measurement represents completion of measurement. 2. Three times of beep after measurement represent abnormal measurement temperature. 3 x 16 Cleaning and maintenance Please keep the cleaning of pregnant preparation instrument. Wipe the heat detector probe with a dry, soft cloth If there is a lot of dirt on the heat detector probe, please wipe the probe with a wringed cloth with neutral detergent or water, and then use a soft cloth to wipe it After connecting with the power, put the instrument into the disinfection charging box and automatic disinfection of ultraviolet ray will begin and last about 5 minutes. Please comply with the following precautions to avoid breakdown or failure Do not remove dirt by using volatile agents, diluent or gasoline, etc. The disinfection charging box is not waterproof; please avoid water flowing into the box. Do not immerse the heat detector probe in alcohol for a long time. Do not use hot water to conduct sterilization of pregnant preparation instrument
(over 50 C water). dry. 17 Problems and Troubleshooting Problem/failure Possible reason Long white light & three times of beep No flicker of white light after startup The measured value is too high due to activity or eating before measurement. The measured value is too low due to incorrect measurement position. The measured value is too low due to open mouth measurement. Artificial interruption of the measurement. The measured value is too high due to a long time placement in hot environment. Bluetooth is not connected with the phone. No indicator lamp is turned on after Electric quantity exhaustion taking out of the disinfection charging box. 17 Adopted measures Please conduct measurement in bed immediately after waking up in the morning. Please conduct measurement in correct measurement position. Please close the lips during the measurement and do not breathe by mouth. Please conduct measurement according to correct instructions. Please use the product in the environment of +5 C to +40C Please check whether the Hao-Hao Health APP is in operation. Please check whether your phone Bluetooth is turned on in operation Please charge. 18 Storage and Shipping Condition Special storage Do not store the product in the following environments, or otherwise it will cause a breakdown or failure Humid environment High temperature, direct sunlight, dusty environment, or salt environment Environment with chemicals or corrosive gas Storage environment Temperature: -10C to +40C Humidity: 85%
Ventilated room without corrosive gases Transportation environment Temperature: -40C to + 55C Humidity: 90% RH Atmospheric pressure: 50 to 101.3 kPa Method: air, sea, rail transport or coach transport 19 Basic Parameters and Specifications Model: BBTX Power of the product Battery: Rechargeable lithium battery (3.7 V) Battery life: continuous use of 12 hours, more than 300 times of charge-discharge Charging power supply: Micro USB (DC 5.0 V; 100 mA)
(This product should be charged with the charging power supply of medical products comply with GB 9706.1/IEC 60601-1 for two MOPP insulation system. The charging power supply should be checked or replaced regularly.) Category of product Electric shock protection type: internal power supply Electric shock protection type: BF product Level of anti-liquid: IPX6 (main body of pregnant preparation instrument) Anti-level liquid feed: IPX6 (prepared pregnant instrument body) The degree of safety of combustible gas: not suitable for using in environment with flammable gas Operating mode: Single measurement Size of product: 95 x 23 x 12 mm Shape of disinfection box: 117 x 44 x 30 mm Net weight: Disinfection charging box of 69 g; pregnant preparation instrument of 14 g Communication: Ultra-low-power consumption Bluetooth 4.0 Compliance:
IEC 60601-1 20 Working environment Temperature: + 5C to + 40C Humidity: 85%
Atmospheric pressure: 70 to 101.3 kPa Product performance Sensor: Rapid response thermistor Measurement: High-speed precision Found Measuring range: 35.00C to 42.00C Accuracy: 35.00C to 39.00C, 0.05C 39.1C to 42.0C, 0.1C Measurement accuracy: 0.01 C (displayed on the Hao-Hao Health APP) Life: 4 years 21 The Registration Certificate Number and Production License Number Product registration certificate number:
Product production license number:
Service Agencies and Information Production company name: Comper (Beijing) Technology Co., Ltd. Manufacturer address: Floor 3 Unit 712, 713 Building 1, No. 3 Yongchang North Road, Economic and Technological Development Zone, Beijing Manufacturer Tel.: +86 01057480968 National hotline400 900 1020 Manual version: BT-M-01-15 Rev.02 Date of formulation / revision:
September 2015 22 23 COMPER TM How the best is done Copyright Copyright is owned by Comper (Beijing) Technology Co., Ltd., the Company shall enjoy and reserve all exclusive authority of copyright. No part of the Manual shall be increased or decreased, excerpted, reproduced, altered or imitated without the prior written consent of the Company. For the content of this Manual, the Company is not responsible for any explicit or implied warranty or guarantee liability. The content change caused by product improvements and technology updates could occur at any time without prior notice. The power of interpretation is vested by the Company. Federal Communications Commission (FCC) Statement 15.21 You are cautioned that changes or modifications not expressly approved by the part responsible for compliance could void the users authority to operate the equipment. 15.105(b) This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
-Consult the dealer or an experienced radio/TV technician for help. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
1) this device may not cause interference and 2) this device must accept any interference, including interference that may cause undesired operation of the device. FCC RF Radiation Exposure Statement:
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. End users must follow the specific operating instructions for satisfying RF exposure compliance. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.