Device specifications Device specifications CAUTION y The maximum sound pressure level is measured based on the distance between the position where the device operator stands during normal operation and any position which is one meter from the device and has the maximum sound pressure level. y Because the temperature and humidity fluctuations influence the accuracy of the experiment results, it is recommended that you install an air conditioning system and a humidifier or dehumidifier in the laboratory to maintain the temperature and humidity. Table 40 Sequencer specifications Item Description Laser classification of the device Class 1 laser product Dimensions 1656 mm (W) 1815 mm (H) 903 mm (D)
(65.2 inches 71.5 inches 35.6 inches ) Net weight Approximately 765 kg (1687 lb) Touch screen Type Size LCD 20 inches Power Resolution 1920 1080 pixels Voltage 200 V to 240 V~
Frequency 50/60 Hz Rated power 3000 VA Overvoltage category II Cable min.10AWG Maximum sound pressure level 75 dBA Degrees of protection provided by IPX0 enclosures (IP Code) 173 AppendixFor research use only Device specifications 174 Item Description Accompanying items Refer to the packing list Temperature 19 C to 25 C (66 F to 77 F) Relative humidity 30% RH to 80% RH, non-condensing Atmospheric pressure Altitude 80 kPa to 106 kPa 2000 m Operating environment requirements Pollution degree 2 Indoor use Because the temperature and humidity fluctuations influence the accuracy of the experiment results, we recommend that you install an air conditioning system and a humidifier or dehumidifier in the laboratory to maintain the temperature and humidity. Transportation/
Storage environment requirements Temperature
-20 C to 50 C (-4 F to 122 F) Relative humidity 15% RH to 85% RH, non-condensing Atmospheric pressure 80 kPa to 106 kPa Table 41 DNB loader specifications Item Description Dimension 430 mm (W) 750 mm (H) 780 mm (D)
(17 inches 30 inches 31 inches) Net weight Approximately 81 kg (179 lb) Touch screen monitor y Size: approximately 13.3 inches y Type: LCD touch screen y Resolution: 1280 600 pixels y Voltage: 100 V - 240 V~
y Frequency: 50/60 Hz Power y Rated power: 600 VA y Overvoltage category: II y Cable: min. 16 AWG AppendixFor research use only Device specifications Item Description Fuse specification F10AL250V Maximum sound pressure level 75 dB(A) D e g r e e s o f p r o t e c t i o n provided by enclosures (IP IPX0 Code) y Temperature: 19 C to 25 C (66 F to 77 F) y R e l a t i v e h u m i d i t y : 3 0 % R H t o 8 0 % R H , n o n -
condensing y Atmospheric pressure: 80 kPa to 106 kPa O p e r a t i n g e n v i r o n m e n t requirements y Pollution degree: 2 y Indoor use Because the temperature and humidity fluctuations influence the accuracy of the loading results, we recommend that you install an air conditioning system and a dehumidifier in the laboratory to maintain appropriate temperature and humidity. Storage/transportation environment requirements y Temperature: -20 C to 50 C (-4 F to 122 F) y Relative humidity: 15% RH to 85% RH, non-condensing y Atmospheric pressure: 80 kPa to 106 kPa Accompanying items Refer to the packing list. 175 AppendixFor research use only---This page is intentionally left blank.---Compliance information Compliance information The device complies with the following standards:
Item Standard Electromagnetic IEC 61326-1 Compatibility
(EMC) Electrical equipment for measurement, control and laboratory use EMC requirements Part 1: General requirements Safety requirements y UL 61610-1/CAS C22.2 No.61010-1-12 S a f e t y r e q u i r e m e n t s f o r e l e c t r i c a l e q u i p m e n t f o r measurement, control, and laboratory use-Part 1: General requirements y UL 61610-2-081/CSA C22.2 No. 61010-2-081 S a f e t y r e q u i r e m e n t s f o r e l e c t r i c a l e q u i p m e n t f o r measurement, control and laboratory use - Part 2-081:
Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes y UL 61010-2-010/CSA C22.2 No. 61010-2-010 S a f e t y r e q u i r e m e n t s f o r e l e c t r i c a l e q u i p m e n t f o r measurement, control and laboratory use - Part 2-010:
Particular requirements for laboratory equipment for the heating of materials y IEC 60825-1 Safety of laser product part 1: equipment classification and requirements 177 AppendixFor research use only---This page is intentionally left blank.---Research use only Research use only Complete Genomics has labeled the product solely for research use only and specified RS in the model name which means it should not be used for clinical diagnosis. Please refer to FDA Guidance, Distribution of In Vitro Diagnostic Products Labeled for Research Use Only or Investigational Use Only (Nov. 2013)
(available at: https://www.fda.gov/media/87374/download). If you have any question, please contact Complete Genomics at +1 (888) 811-9644. 179 AppendixFor research use only---This page is intentionally left blank.---Manufacturer information Manufacturer information Manufacturer Complete Genomics, Inc. Address 2904 Orchard Parkway San Jose, CA 95134 Technical support Complete Genomics, Inc. Technical support E-mail US-TechSupport@mgi-tech.com Customer service telephone
+1 (888) 811-9644 Website www.completegenomics.com 181 AppendixFor research use only---This page is intentionally left blank.---Order information Order information Catalog number Model Name Version Recommended brand 940-000838-00 FCL PE100 940-000836-00 FCL PE150 DNBSEQ-T7RS High-throughput Sequencing Set V1.0 CG DNBSEQ-T7RS High-throughput Sequencing Set V1.0 CG 940-000840-00 stLFR FCL PE100 940-000871-00 940-000872-00 940-000873-00
DNBSEQ-T7RS High-throughput Sequencing Set Sequencer Cleaning Cartridge Cleaning Cartridge DNB Load Plate (no Reagent) 900-000697-00 DL-T7RS DNB Loader DL-T7RS V1.0
CG CG CG CG CG 183 AppendixFor research use only---This page is intentionally left blank.---Acronyms and abbreviations Acronyms and abbreviations Item BCS BIC cPAS CPU Description Basecall server Basecall Information Content Combinatorial Probe-anchor Synthesis Central Processing Unit DL-T7RS DNB Loader for DNBSEQ-T7 DNA DNB Deoxyribonucleic Acid DNA Nanoball dsDNA double-stranded DNA EMC ESR FAQ FCC FCL FIT FOV FPGA GPU IC ID ISW LAN LT MDA PE QC Electromagnetic Compatibility Effective Spots Rate Frequently Asked Questions Federal Communications Commission Flow Cell Large Least square fit to the DNB intensities in 4 color space to represent the overall quality of the clusters Field of View Field-programmable Gate Array Graphics Processing Unit Interference-Causing Identification Instrument Control Software Local Area Network Laser temperature Multiple Displacement Amplification Pair-end sequencing Quality Control 185 AppendixFor research use only Acronyms and abbreviations Item QR RCR RFID RHO RNA SBC SNR Description Quick Response Rolling Circle Replication Radio Frequency Identification Rho (), intensity of raw signals Ribonucleic Acid Single Board Computer Signal to Noise Ratio ssDNA single-stranded DNA stLFR single-tube Long Fragment Read TV UDI UMI UPS USB VGA WES WGS Television Unique Dual Index Unique Molecular Identifier Uninterruptible Power Supply Universal Serial Bus Video Graphics Array Whole Exome Sequencing Whole Genome Sequencing ZLIMS ZTRON laboratory information management system 186 AppendixFor research use only Index Index C ChipProductivity(%) 98 D DNB Loading 69 E Effective spot rate 99 ESR(%) 99 F Flow cell 16 flow cell drive control button 87 flow cell loading interface 38, 75 H Host power button 15 L Lag/Runon 99 Log interface 24 M Main interface 20 187 AppendixFor research use only P Ports 18 Power switch 18 S sensor status indicators 21 Status indicator 14 T TotalReads(M) 98 W Wash 116 Index 188 AppendixFor research use only Part No.: H-020-000589-00