DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China UV Visible Spectrophotometer SP-UV1000 SCI-UV1000 SP-V1000 SCI-V1000 SP-V1100 SCI-V1100 SP-UV1100 SCI-UV1100 SP-UV3101 SCI-3101UV SP-UV2102 SCI-2102UV SP-UV3102 SCI-3102UV SP-UV2101 SCI-2101UV SP-XUV5102 SCI-5102XUV SP-XUV5101 SCI-5101XUV User Manual DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Table of Contents Introduction ......................................................................................................................................... 3 I. Safety ......................................................................................................................................... 3 II. About the device ....................................................................................................................... 3 III. Description of related symbols ............................................................................................... 3 Chapter 1. Overview ......................................................................................................................... 2 I. Instrument rationale and structure ............................................................................................. 2 II. Key technical indicators ........................................................................................................... 3 III. Key functions .......................................................................................................................... 3 Chapter 2. Instrument Installation ....................................................................................................... 4 I. Environmental requirements ...................................................................................................... 4 II. Check supply voltage ............................................................................................................... 5 III. Installation ............................................................................................................................... 5 Chapter 3. Instrument Description .................................................................................................... 6 I. Instrument description ............................................................................................................ 6 II. Operating panel ..................................................................................................................... 7 III. Key description ....................................................................................................................... 7 Chapter 4. Instrument Use ................................................................................................................ 7 I. Software system ......................................................................................................................... 7 II. Basic operations ....................................................................................................................... 8 1. Measurement mode selection ........................................................................................... 8 2. Wavelength Setting .......................................................................................................... 8 3. Parameter Setting ............................................................................................................. 8 4. Delete the measurement results and stored data ............................................................... 9 5. Calibrating 100%T/0Abs.................................................................................................. 9 6. Measuring samples ........................................................................................................... 9 7. Measurement results printing ........................................................................................... 9 III. Preparations before measurement ......................................................................................... 10 1. Power-ON self-test ......................................................................................................... 10 2. Preheating ....................................................................................................................... 10 3. Confirm colorimeter cell ................................................................................................ 11 IV. Measurement ......................................................................................................................... 11 1. Photometric measurement .............................................................................................. 11 2. Quantitative measurement .............................................................................................. 13 V. System Setting ................................................................................................................... 18 1. Measurement mode ........................................................................................................ 19 2. Turn ON and OFF the deuterium lamp ........................................................................... 19 3. Turn ON and OFF the tungsten lamp ............................................................................. 19 4. Dark Current Calibration ................................................................................................ 20 5. Wavelength calibration ................................................................................................... 20 6. Version information ........................................................................................................ 21 7. Export data to USB ......................................................................................................... 21 8. Date and Time Setting .................................................................................................... 21 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Introduction I. Safety The instrument design complies with the safety specifications set forth by the National Standard of the Peoples Republic of China GB4793.1-2007/IEC61010-
1:2001 (i.e., Safety Requirements for Measuring, Control and Lab Electrical Equipment: Part 1) and the National Standard of the Peoples Republic of China GB9706.1-2007/IEC60601-1:21988) (i.e., Medical Electrical Equipment: Part 1). II. About the device SP-UV1000/SP-V1000 serial model visible spectrophotometer is an electronic laboratory testing equipment. Instrument use mode:
Intermittent use Presence or absence of overvoltage (current): No Class of pollution:
2 III. Description of related symbols General Warning, danger Warning, electricity hazard Warning, hot surface Mandatory, Connect an earth terminal to the ground Fuse DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Crossed out recycle bin, this device is to be recycled by the designated electronic equipment disposal department or vendor after use. Chapter 1. Overview is characterized by wide SP-UV1000/SP-V1000 visible spectrophotometer wavelength range, high sensitivity, powerful functions, ease of use, simple structure and acceptable appearance, especially the application of large-screen LCD, high-
precision A/D converter and nonvolatile memory, which makes the instrument uniquely advantageous among the products of the same grade and widely applicable textile, chemical, pharmaceutical, biochemical, metallurgical, materials, environmental protection, medical assay and education industries, as one of the important quality control instruments in the analytical testing industries and an essential instrument in conventional labs. industry, light I. Instrument rationale and structure The rationale of spectrophotometric analysis involves using the selective absorption of light of different wavelengths by samples to perform qualitative and quantitative analysis of the samples and judging the structure and chemical composition of substances through analysis of absorption spectrum. This instrument works on the principle of relative measurement, i.e., selecting a certain solvent (distilled water, air or test sample) as the reference solution and setting its transmittance (i.e., transmission rate T) as 100%, while the transmittance of the testing sample is derived with reference solution as reference standard. The changes in transmittance (transmission rate T) will have a certain functional relation with the concentration of the test sample and it complies with the Lambert-Beer's Law within a certain range. T=I/Io A=KCL=-log I/Io Wherein T transmittance (transmission rate) C solution concentration K absorption coefficient of solution L length of liquid layer in light path 2 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China I intensity of light irradiated on the photoelectric converter through tested sample Io intensity of light irradiated on the photoelectric converter through reference test sample. SP-UV1000/SP-V1000 ultraviolet/visible spectrophotometer is a domestically leading new-generation UV-VIS spectrophotometer. These are developed based on this rationale, in conjunction with the modern precision optics, the latest microelectronics and other high-end new technologies. II. Key technical indicators SP-UV1000/SP-V1000 serial model key technical indicators are as follows:
Item UV Visible Spectrophotometer Wavelength range Photometric range Spectrum bandwidth Wavelength accuracy Wavelength repeatability Photometric accuracy Photometric repeatability Stray light Baseline drift Display Data output 325~1000nm
-0.3~3Abs 4nm (2nm alternative) 1nm 0.2nm 0.5%T(0~100%T) 0.2%T(0~100%T) 0.2%T@360nm 0.002A/h@500nm (preheated for 1hour) 12864 dot-matrix graphic LCD USB, parallel port Overall dimensions 490mm360mm210mm Weight III. Key functions 12kg SP-UV1000/SP-V1000 serial model visible spectrophotometer has three major functions. Photometric measurement 3 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China The measurement modes are switchable (absorbance, transmission rate and energy) and the measurement results (up to 200 groups) can be saved in the instruments memory and printed. Quantitative measurement Standard curve can be established using either coefficient method or standard sample calibration method, and the regression equation (up to 200 groups) and measurement results (up to 200 groups) can be saved in the instruments memory or printed. System application Turn ON and OFF deuterium and tungsten lamps, calibrate dark current, calibrate wavelength, export data and select measurement mode and restore factory settings. Chapter 2. Instrument Installation I. Environmental requirements SP-UV1000/SP-V1000 serial model visible spectrophotometer is designed in response to the effects of environmental factors on the instrument. To ensure normal operation of the instrument and achieve an extended service life, check installation environment condition before installation. The environmental requirements that ensure normal operation of the instrument are as follows:
Avoid high-temperature and high-humidity environments The instrument shall be installed away from high-temperature and high-humidity environments. The instrument shall be used at 16~35 and 45-80% RH. Keep the instrument away from outside magnetic field interferences Ensure to place the instrument away from magnetic/ electric field and high-
frequency waves emitting electronic devices as much as possible. Keep the instrument away from corrosive gases Please dont install the instrument in the presence of corrosive gases such as chlorine, hydrochloric acid gas, hydrogen sulfide gas and sulfurous acid gas in air significantly exceeds the applicable standards. 4 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Instrument must be placed on a steady surface Place the instrument on a steady and vibration free surface. The instruments fan should be sufficient to allow it vent smoothly. Supply voltage The power supply for normal operation of the instrument is AC 22022V@
501Hz or 11011V@601 Hz. Power supply must be well grounded The instrument should preferably use a standalone power socket and the power supply should be well grounded in order to prevent abnormal operation of the instrument. Regulated power supply When the local voltage is unstable, please equip the instrument with regulated power supply. Instrument should be protected from direct sunlight. Maintain dust-free environment. II. Check supply voltage Ensure to check the supply voltage before installation, in order to avoid damages to the instrument. Supply voltage varies greatly from one country to another. The input power supply to any devices sold by DLAB in mainland China is 220V/50Hz. III. Installation The installation of SP-UV1000/SP-V1000 serial model visible spectrophotometer is very simple. Please follow the following steps:
Step 1. Check the contents of package Open the package and carefully and examine its contents against the packing list. Please contact us or our dealer immediately if any item inside the package is found missing or damaged. Step 2. Place the instrument Take out the instrument and gently place it on a steady surface. 5 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Step 3. Install the printer (an optional accessory) Check and ensure that the power switch on the printer is OFF and connect the data cable of the printer in the printer port of the instrument. Step 4. Install the power cable Ensure to switch power OFF the instrument. Insert one end of the power cable into the power port and the other end into the power socket of the instrument. Step 5. Connect the power supply Ensure whether all connecting points are well-connected and switch power ON the instrument. The instrument is now ready for normal use after self-test. Chapter 3. Instrument Description I. Instrument description Sample chamber cover Sample rack Operating panel Pull rod Cooling fan Louver cover Printer port USB(B) for software USB interface Power socket 11 Power switch 6 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China II. Operating panel The operating panel of UV-1000/V-1000 serial model visible spectrophotometers is shown as follows:
LCD display Keypad III. Key description Set the parameters Set the wavelength Calibrate 100%T/0Abs Print the data Function key, promoting functions according to display screen Roll down the menu options or turn datasheet pages Roll up the menu options or turn datasheet pages Chapter 4. Instrument Use I. Software system The instruments software system is shown as follows:
7 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Main menu Photometric measurement Measurement Printing Quantitative measurement Establish a curve Coefficient method Set the testing wavelength Invoke a curve Clear a curve Measurement Printing Standard sample method System setting Measurement mode Transmittance Input the measurement coefficient K Input the measurement coefficient B Set the testing wavelength Calibrate blank Set the number of standard samples Absorbance Set the standard sample Energy Deuterium lamp Tungsten lamp Dark current calibration Wavelength calibration Version information II. Basic operations 1. Measurement mode selection On the main interface, press the corresponding function key to enter the corresponding measurement mode. 2. Wavelength Setting On the key pad, press to enter the wavelength setting interface, press or to change the entered wavelength value and press
(left) to proceed to the set wavelength value and automatically calibrate 100%T/0Abs. 3. Parameter Setting 8 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Press key to enter the parameter setting interface, press or to select the settings or input parameters and press
(left) to confirm. 4. Delete the measurement results and stored data If the local memory is full and you want to clear all saved data, you can long press the button 5. Calibrating 100%T/0Abs to delete all data in the interface of date view. Place the reference into the light path and press 100%T/0Abs. key to calibrate 6. Measuring samples Place the sample into the light path and press
(left) to measure the sample. 7. Measurement results printing On the measurement interface, short press to enter print setting, input the data number you want to print, press
(left) to print the corresponding test results.
<PRINT>
<OK>
Instrument host and Bluetooth printer online tutorial:
1) Turn on the printer switch, the green light continues to flash;
2) Open the instrument host, after the instrument self-test is completed, it can automatically connect with the Bluetooth printer. 3) If the connection is successful, the green indicator light of the printer will be long bright. 4) The operating instrument can be printed!
The specific operation process of the Bluetooth printer can be viewed in its user manual!!
9 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China III. Preparations before measurement 1. Power-ON self-test Ensure that the light path is clear and close the sample chamber cover. Turn ON the power switch on the instrument to start self-test. The start window is like this:
Self-test It can self-check the following items: color filter, lamp switch, detector, deuterium lamp, tungsten lamp, wavelength calibration and dark current calibration. Self-test error If any item has error during self-checking, it will beep and show the error item. You can press any key to continue the process. After the self-test is done, you can go to the system application for further inspections and calibrations. If errors continue, please contact the service team. 2. Preheating After self-test, the instrument enters into the preheating state. For accurate measurement, preheating time needs to be 20-30 minutes. 10 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China 3. Confirm colorimeter cell Before moving the sample into the colorimeter cell, confirm the cell is clean and free of residues. If the wavelength to be tested is less than 400nm, please use quartz cell. IV. Measurement 1. Photometric measurement Photometric measurement is used to test the absorbance or transmittance at a fixed wavelength. Step 1. Enter the photometric measurement interface After the self-checking and preheating, the device will go to the main interface of photometric measurement automatically as follows. You can select the working mode by pressing key and
(left) to enter the working mode interface. Then press or to choose working mode, like absorbance, transmittance or energy. Then confirm and press
(right) to back to main interface.
<SET>
<Cancel>
<OK>
<Cancel>
On the main interface, press
(left) to select Basic
(left)
(right) 11 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Step 2. Set the wavelength to be tested Press to enter the wavelength setting interface and or to input the wavelength value, and press value.
(left) to proceed to the set wavelength
(left)
<OK>
The valid WL range is 200-1000nm, if invalid number is entered, it will beep three times and go back to the main interface of photometric measurement. Step 3. Calibrate 100%T/0Abs Place the reference into the light path and wait until the stability indicator shows up, then press to calibrate 100%T/0Abs. Note: remember to put in the reference sample before doing blanking, otherwise the test results will not be accurate. Step 4. Measure the samples Place the sample into the light path, wait until the stability indicator shows up, and press
(test) again to have the test results, and the readings will be displayed in the data list. Repeat this operation to complete the sample measurements
(left) to enter measurement interface, then press Each page can show three data and you can see other pages via pressing or
. In the measurement interface, you can set the wavelength too, but need to do the zero again, otherwise the results will not be accurate. Step 5. Print the data 12 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China On the measurement interface, short press to enter print setting, input the data number you want to print, press
(left) to print the corresponding test results.
<PRINT>
<OK>
Step 6. Delete the data If the local memory is full and you want to clear all saved data, you can long press the button to delete all data in the interface of date view. 2. Quantitative measurement You can create the curve via standard samples with known concentration and use this curve to do quantitative measurement for similar samples. If the K and B are known in advance, you can also set up the curve via coefficient method. Step 1. Enter the quantitative measurement interface On the main interface, press
(right) to select quantitative measurement.
(right)
(right) Step 2. Establish the standard curve or load the stored standard curve There are two ways to establish standard curve available for users to choose, depending upon the actual situation. Coefficient method:
This method is a simple way of quantitative measurement, it is realized by a formula with the absorbance value tested by the spectrophotometer. 13 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Coefficient Set Wavelength Test Data Enter K Enter B Zero with reference Measurement Print Data Delete Data 1) Select the coefficient method. Press select coefficient method. or to select create curve and press or to
<OK>
<Cancel>
2) Set the wavelength to be measured. Press or to enter the measured wavelength value and press
(left) to confirm and back to previous interface.
<Keypad>
<OK>
3) Set the measurement coefficient K. Press or and back to previous interface. to enter the K value and press
(left) to confirm
<Keypad>
<OK>
4) Set the measurement coefficient B. 14 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China or Press and back to previous interface. to enter the B value and press
(left) to confirm
<Keypad>
<OK>
Continue to press
(left) to confirm and complete the coefficient method setting. Now instrument displays the newly established curve. Then you can do the measurements. until Keypad become OK, press or
(left) Test Note: before do the tests, make sure do the blanking with your reference sample first, otherwise the results will not be accurate. Standard Sample Method Std. Sample Create Curve Load Curve Set WL Input Sample NO Create Curve Show Curve Test/save data Print Data Delete Data Select Curve Show Curve Test/Save data Print Data Delete Data 15 Remove Curve Select Curve Confirm Delete Curve DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China 1) Select the standard sample method. Press or to select create curve, press press or to select "standard Curve" and press
(left) to enter standard sample method.
<OK>
<Cancel>
2) Build new standard curve. Enter the wavelength and number of samples (up to 9), press
(left) to confirm and enter the next interface. Press or to input the concentration of each sample via keypad. Then press to do blanking, after that, press as follows.
(left) read the absorbance value of each sample. The details are
<Enter WL>
<Sam No.>
<Enter Conc>
<Test>
<OK>
<Read Abs>
The standard curves established after successive completion of calibration of all standard samples will be automatically saved in the instruments memory. 16 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China Note: the number range of the samples is 1-9. If invalid number is entered, it will beep three time and back to setting interface. Invoke the stored standard curve:
The newly built curves will be saved to the local memory automatically and uses can invoke the directly. It can save up to 50 curves and the latest curve is at the end of the list. 1) Select the saved curve. Press or to select Load Curve, press or to select the corresponding curve equation and then press
(left) to confirm
<OK>
<Cancel>
2) Enter the sample measurement interface Press
(left) to enter the measurement interface after invoking a curve. 3) Calibrate 100%T/0Abs. Place the reference into the light path and press 4) Measure the sample using the standard curve. to calibrate 100%T/0Abs. Place the sample into the light path and press
(left) to measure it. The results will be displayed in the data list. Repeat this operation to complete measurement of all samples. 5) Print the data 17 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China On the measurement interface, short press to enter print setting, input the data number you want to print, press
(left) to print the corresponding test results. 6) Delete the data If the local memory is full and you want to clear all saved data, you can long press the button to delete all data in the interface of date view. Remove curve:
This is used to delete the saved curves. You can go to this menu and select the curve needs to be removed, and press
(left) to delete the curves.
<OK>
<Cancel>
V. System Setting Users may set the system configuration of the instrument in the system application depending upon actual use. On the main interface, press setting parameter interface. key to enter the 18 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China 1. Measurement mode Press or to select Test Mode, press
(left) to enter, press or to select absorbance, transmittance or energy and then press
(left) to confirm. 2. Turn ON and OFF the deuterium lamp When measuring with testing wavelength within the range of 340-
1000nm, users can turn OFF the deuterium lamp to extend the service life of deuterium lamp. Press or to select D2 Lamp, press
(left) to enter, press or to select ON or Off and then press
(left) to confirm.
<OK>
<Cancel>
3. Turn ON and OFF the tungsten lamp 19 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China When measuring with testing wavelength within the range of 200-
339nm, users can turn OFF the tungsten lamp to extend the service life of tungsten lamp. Press or to select D2 Lamp, press
(left) to enter, press or to select ON or Off and then press
(left) to confirm.
<OK>
<Cancel>
4. Dark Current Calibration When the working environment of the instrument changes, users may recalibrate the dark current after the instrument preheating and before sample measurement. or Press
(left) to calibrate. to select Dark Current, press The sample chamber cover must be closed throughout the entire calibration process. Return after completion.
<OK>
<Cancel>
5. Wavelength calibration Sudden power failures while the instrument is in use, abnormal power OFF and attenuation of light source energy will all result in wavelength deviation and deviation of measurement results, where wavelength re-
calibration is required. Make sure that the light path is clear and the sample chamber cover is closed. Press or to select Reset WL and press
(left) to calibrate. The sample chamber cover may not be opened throughout the calibration process. Return after completion.
<OK>
<Cancel>
20 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China 6. Version information Press version information interface and then press any key to return. to select version, press or
(left) to view the
<OK>
<Cancel>
7. Export data to USB You can transfer the saved data on the spectrophotometer to the USB. Press or to select Export, press
(left) to transfer data.
<OK>
<Cancel>
Before export the data, you must insert the USB to the USB port on the back of the device, otherwise it will give error information. 8. Date and Time Setting Press menu and set the correct date and time. or to select Time Setting, press
(left) to enter the
<OK>
<Cancel>
21 DLAB Scientific Co., Ltd. YuAn Road 31, Airport Economic Core Zone,Shunyi District,Beijing101318,P.R.China FCC Statement This device complies with part 15 of the FCC rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interferenc e that may cause undesired operation. Changes or modifications not expressly approved by the party responsible for compliance could void the users authorit y to operate the equipment. NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to par t 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a resid ential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used i n accordance with the instructions, may cause harmful interference to radio communications. However, there is no guar antee that interference will not occur in a particular installation. If this equipment does cause harmful interference to ra dio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try t o correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. Important Note:
Radiation Exposure Statement This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance 20cm between the radiator and your body. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. ISED Statement English: This device complies with Industry Canada licenseexempt RSS standard(s). Operation is subject to the following two conditions: (1) This device may not cause interference, and (2) This device m ust accept any interference, including interference that may cause undesired operation of the device. The digital apparatus complies with Canadian CAN ICES3 (B)/NMB3(B). French: Le prsentappareilestconforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitationestautorise aux deux conditions suivantes: (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareildoit accepter tout brouillageradiolectriquesubi, mmesi le brouillageest susc eptible d'encompromettre le fonctionnement. This radio transmitter has been approved by Industry Canada to operate with the antenna types listed with the maximum permissible gain indicated. Antenna types not included in this list, having a gain greater than the maximum gain indicated for that type, are strictly prohibited for use with this device. Le prsent metteur radioa t approuv par Industrie Canada pour fonctionner avec les types d'antenne numrs ci-
dessous et ayant un gain admissible maximal. Les types d'antenne non inclus dans cette liste, et dont le gain est suprieur au gain maximal indiqu, sont strictement interdits pour l'exploitation de l'metteur. Radiation Exposure Statement This equipment complies with Canada radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance 20cm between the radiator & your body. Dclaration d'exposition aux radiations Cet quipement est conforme Canada limites d'exposition aux radiations dans un environnement non contrl. Cet quipement doit tre install et utilis distance minimum de 20cm entre le radiateur et votre corps. 22