BA01697D/06/EN/01.17 71370929 Valid as of version 01.00.zz (Device firmware) Products Solutions Services Operating Instructions Picomag IO-Link Electromagnetic flowmeter Picomag IO-Link Make sure the document is stored in a safe place such that it is always available when working on or with the device. To avoid danger to individuals or the facility, read the "Basic safety instructions" section carefully, as well as all other safety instructions in the document that are specific to working procedures. The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these instructions. 2 Endress+Hauser Picomag IO-Link Table of contents 1 1.1 1.2 1.3 1.4 About this document . 4 Document function . 4 Symbols used . 4 Safety symbols . 4 1.2.1 Electrical symbols . 4 1.2.2 Communication symbols . 4 1.2.3 Symbols for certain types of 1.2.4 information . 4 Symbols in graphics . 5 1.2.5 Documentation . 5 Registered trademarks . 5 Basic safety instructions . 6 2 Requirements for the personnel . 6 2.1 Designated use . 6 2.2 2.3 Workplace safety . 6 Operational safety . 7 2.4 Product safety . 7 2.5 IT security . 7 2.6 Device-specific IT security . 7 2.7 2.7.1 Access via the SmartBlue App . 7 2.7.2 Access via Bluetooth wireless technology . 7 3 3.1 3.2 4 4.1 4.2 4.3 Incoming acceptance and product identification . 8 Incoming acceptance . 8 Product identification . 8 Symbols on measuring device . 9 3.2.1 Storage and transport . 10 Storage conditions . 10 Transporting the product . 10 Packaging disposal . 10 5 5.1 Installation . 11 Installation conditions . 11 5.1.1 Mounting position . 11 5.2 Mounting the measuring device . 11 6 6.1 6.2 6.3 7 7.1 Electrical connection . 13 Connection conditions . 13 Requirements for connecting cables . 13 6.1.1 Pin assignment, device plug . 13 6.1.2 Connecting the measuring device . 15 Post-connection check . 16 Operation options . 17 Access to the operating menu via the SmartBlue App . 17 Table of contents 8 8.1 8.2 9 9.1 9.2 System integration . 18 Overview of device description files . 18 Device master file . 18 Commissioning . 19 Switching ON the measuring device . 19 Configuring the measuring device . 19 Configuring the display . 20 9.2.1 Configuring system units . 21 9.2.2 Setting the installation direction and 9.2.3 measurement . 21 Configuring the IO modules . 22 Simulation . 26 9.2.4 9.2.5 10 Operation . 27 10.1 Reading measured values . 27 10.2 Configuring Bluetooth . 28 10.3 Administration . 28 Diagnostics and troubleshooting . 29 11 11.1 General troubleshooting . 29 11.2 Diagnostic information on local display . 30 11.2.1 Diagnostic message . 30 11.3 Overview of diagnostic events . 31 11.4 Pending diagnostic events . 32 11.5 Device information . 32 11.6 Firmware history . 32 12 Accessories . 33 Technical data . 34 13 13.1 Input . 34 13.2 Output . 34 13.3 Power supply . 34 13.4 Performance characteristics . 34 13.5 Installation . 34 13.6 Environment . 34 13.7 Process . 35 13.8 Mechanical construction . 35 13.9 Operability . 36 13.10 Certificates and approvals . 36 14 Appendix . 38 14.1 Radio approvals . 38 14.1.1 Europe . 38 14.1.2 Canada and USA . 38 14.1.3 Other countries . 38 Index . 39 Endress+Hauser 3 About this document Picomag IO-Link 1 About this document Document function 1.1 These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal. 1.2 Symbols used 1.2.1 Safety symbols Symbol Meaning DANGER!
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury. NOTE!
This symbol contains information on procedures and other facts which do not result in personal injury. 1.2.2 Electrical symbols Symbol Meaning Direct current Alternating current Direct and alternating current Ground connection A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system. 1.2.3 Communication symbols Symbol Meaning Bluetooth Wireless data transmission between devices over a short distance. 1.2.4 Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted. Preferred Procedures, processes or actions that are preferred. 4 Endress+Hauser DANGERWARNINGCAUTIONNOTICE Picomag IO-Link About this document Symbol Meaning Forbidden Procedures, processes or actions that are forbidden. Tip Indicates additional information. Reference to documentation Reference to page Reference to graphic Notice or individual step to be observed
,
, Series of steps Result of a step 1.2.5 Symbols in graphics Symbol 1, 2, 3, ... A, B, C, ... Meaning Item numbers Views 1.3 Documentation For an overview of the scope of the associated Technical Documentation, refer to the following:
The W@M Device Viewer: Enter the serial number of the measuring device
(www.endress.com/deviceviewer) The Endress+Hauser Operations App: Enter the serial number of the measuring device or scan the 2-D matrix code on the measuring device. 1.4 Registered trademarks Is a registered trademark. It may only be used in conjunction with products and services by members of the IO-Link Community or by non-members who hold an appropriate license. For more detailed information on the use of IO-Link, please refer to the rules of the IO-
Link Community at: www.io.link.com. Bluetooth wireless technology The Bluetooth word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of such marks by Endress+Hauser is under license. Apple Apple, the Apple logo, iPhone, and iPod touch are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc. Android Android, Google Play and the Google Play logo are trademarks of Google Inc. Endress+Hauser 5 A1.2.3. Basic safety instructions Picomag IO-Link 2 Basic safety instructions and task. Requirements for the personnel 2.1 The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements:
Trained, qualified specialists must have a relevant qualification for this specific function Are authorized by the plant owner/operator. Are familiar with federal/national regulations. Before starting work, read and understand the instructions in the manual and supplementary documentation as well as the certificates (depending on the application). Follow instructions and comply with basic conditions. The operating personnel must fulfill the following requirements:
Are instructed and authorized according to the requirements of the task by the facility's owner-operator. Follow the instructions in this manual. Designated use 2.2 Application and media The measuring device described in these Brief Operating Instructions is intended only for flow measurement of liquids with a minimum conductivity of 20 S/cm. To ensure that the measuring device remains in proper condition for the operation time:
Use the measuring device only for media against which the process-wetted materials are adequately resistant. Incorrect use Non-designated use can compromise safety. The manufacturer is not liable for damage caused by improper or non-designated use. LWARNING Danger of breakage due to corrosive or abrasive fluids!
Verify the compatibility of the process fluid with the sensor material. Ensure the resistance of all fluid-wetted materials in the process. Keep within the specified pressure and temperature range. Residual risks LWARNING The electronics and the medium may cause the surfaces to heat up. This presents a burn hazard!
For elevated fluid temperatures, ensure protection against contact to prevent burns. 2.3 Workplace safety For work on and with the device:
Wear the required personal protective equipment according to federal/national regulations. For welding work on the piping:
Do not ground the welding unit via the measuring device. 6 Endress+Hauser Picomag IO-Link Basic safety instructions Operational safety 2.4 Risk of injury!
Operate the device in proper technical condition and fail-safe condition only. The operator is responsible for interference-free operation of the device. Product safety 2.5 This measuring device is designed in accordance with good engineering practice to meet state-of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate. It meets general safety standards and legal requirements. It also complies with the EU directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device. IT security 2.6 We only provide a warranty if the device is installed and used as described in the Operating Instructions. The device is equipped with security mechanisms to protect it against any inadvertent changes to the device settings. IT security measures in line with operators' security standards and designed to provide additional protection for the device and device data transfer must be implemented by the operators themselves. 2.7 Device-specific IT security 2.7.1 Access via the SmartBlue App Two access levels (user roles) are defined for the device: the Operator user role and the Maintenance user role. The Maintenance user role is the default setting. If a user-specific access code is not defined (in the Set access code parameter), the default setting 0000 continues to apply and the Maintenance user role is automatically enabled. The device's configuration data are not write-protected and can be edited at all times. If a user-specific access code has been defined (in the Set access code parameter), all the parameters are write-protected and the device is accessed with the Operator user role. The previously defined access code must first be entered again before the Maintenance user role is enabled and all the parameters can be write-accessed. 2.7.2 Access via Bluetooth wireless technology Signal transmission via Bluetooth wireless technology uses a cryptographic technique tested by the Fraunhofer Institute The device is not visible via Bluetooth wireless technology without the SmartBlue App. Only one point-to-point connection is established between a sensor and a smartphone or tablet. The Bluetooth wireless technology interface can be disabled via SmartBlue. Endress+Hauser 7 Incoming acceptance and product identification Picomag IO-Link 3 Incoming acceptance and product identification 3.1 Incoming acceptance A0028673 A0028673 A0028673 A0028673 Are the order codes on the delivery note (1) and the product sticker (2) identical?
Are the goods undamaged?
Do the nameplate data match the ordering information on the delivery note?
Is the CD-ROM with the Technical Documentation
(depends on device version) and documents present?
If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. Depending on the device version, the CD-ROM might not be part of the delivery!
The Technical Documentation is available via the Internet or via the Endress+Hauser Operations App, see the "Product identification" section . Product identification 3.2 The following options are available for identification of the measuring device:
The device label Order code with breakdown of the device features on the delivery note Enter the serial number on the device label in W@M Device Viewer
(www.endress.com/deviceviewer): all the information about the measuring device is displayed. Enter the serial number on the device label into the Endress+Hauser Operations App or scan the 2-D matrix code (QR code) on the measuring device with the Endress+Hauser Operations App: all the information about the measuring device is displayed. 8 Endress+Hauser 1212Order code:Ser. no.:Ext. ord. cd.:iiDate:Picomag IO-Link Incoming acceptance and product identification 3.2.1 Symbols on measuring device Symbol Meaning WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Reference to documentation Refers to the corresponding device documentation. Endress+Hauser 9 Storage and transport Picomag IO-Link 4 Storage and transport 4.1 Storage conditions Observe the following notes for storage:
Store in the original packaging to ensure protection from shock. Store in a dry place. Do not store outdoors. Storage temperature 34 Transporting the product 4.2 Transport the measuring device to the measuring point in the original packaging. Do not remove protective covers or caps installed on process connections. They prevent mechanical damage to the sealing surfaces and contamination in the measuring tube. Packaging disposal 4.3 All packaging materials are environmentally friendly and 100% recyclable:
Carton in accordance with European Packaging Directive 94/62EC; recyclability is confirmed by the affixed RESY symbol. 10 Endress+Hauser Picomag IO-Link Installation 5 Installation 5.1 Installation conditions 5.1.1 Mounting position Mounting location Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the next pipe elbow: h 2 DN Inlet and outlet runs Observe the following inlet and outlet runs to comply with accuracy specifications:
A0032998 A0032997 Installation dimensions: information on the dimensions and installed lengths of the device 35 5.2 Mounting the measuring device LWARNING Burn hazard!
If medium temperatures or ambient temperatures exceed 50 C, areas of the housing can heat to over 65 C. Safeguard the housing so that it cannot be touched accidentally. Endress+Hauser 11 h3 DN1 DN Installation Picomag IO-Link Pipe Seal (not supplied) Adapter: available adapters 33 Seal (included in delivery) 1 2 3 4 5 Measuring device connection A0033002 12 Endress+Hauser 1123445532 Picomag IO-Link Electrical connection 6 Electrical connection 6.1 Connection conditions Requirements for connecting cables 6.1.1 National regulations and standards apply. Connecting cable Conductor cross-section Temperature range Degree of protection M12 1 A-coded At least 0.12 mm2 (AWG26) 10 to +80 C (+14 to +176 F) IP65/67 6.1.2 Pin assignment, device plug A0033004 A B Socket Connector Pin Assignment 1 2 3 4 L+
Output 2 L-
Output 1 Description Supply voltage + (18 to 30 VDC/max. 3 W) Output 2, can be configured independently of output 1 Supply voltage -
Output 1, can be configured independently of output 2 Endress+Hauser 13 A21431234+-EABBAL+L-OUT2OUT1 Electrical connection Picomag IO-Link Switch/pulse output configuration version Configurable switching behavior:
pnp 1) npn 2) A B L+
L-
Socket Connector Supply voltage +
Supply voltage -
A B L+
L-
Socket Connector Supply voltage +
Supply voltage -
A0033005 A0033006 The load is switched to low side L-. The maximum load current is 250 mA. The output is overload-resistant. The load is switched to high side L+. The maximum load current is 250 mA. The output is overload-
resistant. 1) 2) positive negative positive (high side switch) negative positive negative (low side switch) Current output configuration version 1 A B L+
L-
Current output, active, 4 to 20 mA Socket Connector Supply voltage +
Supply voltage -
The current flows from the output to L-. The maximum load may not exceed 500 . A bigger load distorts the output signal. Voltage output configuration version A0033007 A0033007 2 A B L+
L-
Voltage output, active, 2 to 10 V Socket Connector Supply voltage +
Supply voltage -
14 Endress+Hauser 1234BAL+L-1234BAL+L-1234BAL+L-1234BAL+L-Picomag IO-Link Electrical connection The voltage from the output applies to L-. The load must be at least 500 . The output is overload-resistant. Status input configuration version 15 V (switch-on threshold) 5 V (switch-off threshold) A0033008 3 A B L+
L-
Status input Socket Connector Supply voltage +
Supply voltage -
Internal resistance: 7.5 IO-Link configuration version Option only available for output 1 in the Output 1 22 submenu The measuring device has a Type 2 IO-Link communication interface with a second IO function on pin 2 and with a baud rate of 38 400. This requires an IO-Link-enabled module
(IO-Link Master) for operation. The IO-Link communication interface allows direct access to the process and diagnostics data. Connecting the measuring device 6.2 NOTICE The measuring device may only be installed by properly trained technicians. Comply with national and international regulations regarding the installation of electrotechnical systems. Power supply according to EN 50178, SELV, PELV or Class 2. 1. De-energize the system. 2. Connect the measuring device via the connector. 3. In the case of non-grounded pipes:
The device must be grounded using the ground terminal accessory. A0033003 15 Endress+Hauser 1234BAL+L-==1 Electrical connection Picomag IO-Link 6.3 Post-connection check Are cables or the device undamaged (visual inspection)?
Do the cables have adequate strain relief?
Is the connector connected correctly?
Does the supply voltage match the specifications on the measuring device?
Is the pin assignment of the connector correct?
Is the potential equalization established correctly?
16 Endress+Hauser Picomag IO-Link Operation options 7 Operation options Access to the operating menu via the SmartBlue App 7.1 The device can be operated and configured via the SmartBlue App. In this case, the connection is established via the Bluetooth wireless technology interface. Supported functions Device selection in Live List and access to the device (login) Configuration of the device Access to measured values, device status and diagnostics information The SmartBlue App is available for free download for Android devices (Google Playstore) and iOS devices (iTunes Apple Shop) : Endress+Hauser SmartBlue Directly to the app with the QR code:
A0033202 System requirements Devices with iOS:
iPhone 4S or higher, from iOS9.0 iPad2 or higher, from iOS9.0 iPod Touch 5th generation or higher, from iOS9.0 Devices with Android:
Android 4.4 KitKat or higher Download the SmartBlue App:
1. Install and start the SmartBlue App. A Live List shows all the devices available. The list displays the devices with the configured tag name. The default setting for the tag name is EH_DMA_XYZZ (XYZZ = the last 7 digits of the device serial number). 2. Select the device from the Live List. The Login dialog box opens. Logging in:
3. Enter the user name: admin. 4. Enter the initial password: serial number of the device. 5. Confirm your entry. The main menu opens. Navigate through the various items of information about the device: swipe the screen to the side. Endress+Hauser 17 System integration Picomag IO-Link System integration 8 The measuring device has an IO-Link communication interface. The IO-Link interface allows direct access to process and diagnostics data and enables the user to configure the measuring device on the fly. Properties:
IO-Link Specification: Version 1.1 IO-Link Smart Sensor Profile 2nd Edition SIO mode: yes Speed: COM2 (38.4 kBaud) Minimum cycle time: depends on device and is defined by a number of factors including Process data width: depends on device and is defined by a number of factors including the process data to be transmitted. 10 ms the process data to be transmitted. 80 bit IO-Link data storage: yes Block configuration: no More information on IO-Link is available at www.io-link.com 8.1 Overview of device description files Current version data for the device Firmware version 01.00.zz On the title page of the Operating instructions On the device label Parameter Firmware version System Device info Firmware version Release date of firmware version 09.2017 Profile version 1.1 Smart Sensor Profile
---
---
Device master file 8.2 In order to integrate field devices into a digital communication system, the IO-Link system needs a description of the device parameters, such as output data, input data, data format, data volume and supported transmission rate. These data are available in the device master file (IODD 1)) which is provided to the IO-Link Master via generic modules when the communication system is commissioned. The IODD can be downloaded as follows:
Endress+Hauser: www.endress.com IODDfinder: ioddfinder.io-link.com 1) IO Device Description 18 Endress+Hauser Picomag IO-Link Commissioning 9 Commissioning Switching ON the measuring device 9.1 Once the supply voltage has been switched on, the measuring device adopts the normal mode after a maximum of 5 s. During the start-up phase, the outputs are in the same state as the measuring device in the switched-off state. 9.2 Configuring the measuring device Overview of the operating menu Application Measured values System units Process Sensor Output 1 Output 2 Device info Display Bluetooth configuration Administration Actual diagnostics Simulation 1 27 21 21 22 22 32 20 28 28 32 26 System Diagnostics Endress+Hauser 19 Commissioning Picomag IO-Link Configuring the display 9.2.1 The Display submenu contains all the parameters that can be configured for the configuration of the onsite display. Navigation Menu: "System" Display Parameter overview with brief description Parameter Format display Description Select how measured values are shown on the display. Selection/input Display value 1st line + display value 2nd line:
Volume flow + temperature Volume flow + totalizer Temperature + totalizer Factory setting Volume flow +
temperature Rotation display Select local display rotation. Auto (automatic) The display rotates automatically depending on the installation position Auto 0 Can be read in the horizontal installation position with flow from left to right 90 Can be read in the vertical installation position with flow from bottom to top A0033013 180 Can be read in the horizontal installation position with flow from right to left A0033014 270 Can be read in the vertical installation position with flow from top to bottom A0033015 Backlight Set the intensity of the backlighting. 0 to 100 %
A0033016 50 %
20 Endress+Hauser AAAA Picomag IO-Link Commissioning Configuring system units 9.2.2 In the System units submenu, you can configure the units of all measured values. Navigation Menu: "Application" System units Parameter overview with brief description Parameter Description Options Factory setting Volume flow unit Select volume flow unit. Volume unit Select volume unit. Temperature unit Select temperature unit. Totalizer unit Select totalizer unit. l/s, m/h, l/min gal/min (us) ml, l, m fl. oz (us), gal (us) C F l, m 1 000 l, 1 000 m3 fl. oz (us), gal (us) 1 000 gal (us) l/min ml C m Setting the installation direction and measurement 9.2.3 The Sensor submenu contains parameters for specific settings of the measuring device. Navigation Menu: "Application" Sensor Parameter overview with brief description Parameter Description Installation direction Select the installation direction. Selection/input Flow in arrow direction (forwards) Positive flow measurement in the direction of the arrow. Flow against arrow direction (backwards) Positive flow measurement in the opposite direction of the arrow. On value Enter the on value for low flow cut off. Positive floating point number A flow measured value that is less than the value of the on value forces the display to zero. In the event of plant downtime, this prevents the totalizer from continuing to totalize even though there is no flow. Damping Enter the time constant for damping the flow measured value. 0 to 10 s Factory setting Flow in arrow direction (forwards) Depends on the nominal diameter:
DN 15 ("):
0.4 l/min (0.1 gal/min) DN 20 ("):
0.75 l/min (0.2 gal/min) DN 25 (1"):
1.2 l/min (0.3 gal/min) DN 50 (2"):
5.0 l/min (1.3 gal/min) 0 s Endress+Hauser 21 Commissioning Picomag IO-Link Configuring the IO modules 9.2.4 The measuring device has two signal inputs or signal outputs that can be configured independently of one another:
Current output 22 Pulse output 22 Switch output 23 Voltage output 24 Status input 25 Navigation Menu: "Application" Output 1 Menu: "Application" Output 2 Parameter overview with brief description Parameter Description Options Factory setting I/O module 1 type Select the operating mode of output 1. I/O module 2 type Select the operating mode of output 2. IO-Link Switch output Pulse output Current output Switch output Voltage output Digital input IO-Link Off Pulse output Current output Switch output Voltage output Digital input Off Configuring the current output The Current output submenu contains all the parameters that must be configured for the configuration of the current output. The output is used to output process variables by analog means in the form of a 4-20 mA current. Navigation Menu: "Application" Output 1 Current output Menu: "Application" Output 2 Current output Parameter overview with brief description Parameter Description Selection/input Factory setting Assign current output Select process variable for current output. 4 mA value 20 mA value Enter 4 mA value. Enter 20 mA value. Off Volume flow Temperature Volume flow Floating point number with sign 0 l/min Floating point number with sign Depends on the nominal diameter:
DN 15 ("): 25 l/min (6.6 gal/min) DN 20 ("): 50 l/min (13.2 gal/min) DN 25 (1"): 100 l/min (26.4 gal/min) DN 50 (2"): 750 l/min (198.1 gal/min) Configuring the pulse output The Pulse output submenu contains all the parameters that must be configured for the configuration of the pulse output. Navigation 22 Endress+Hauser Picomag IO-Link Commissioning Menu: "Application" Output 1 Pulse output Menu: "Application" Output 2 Pulse output Parameter overview with brief description Parameter Description Entry Factory setting Value per pulse Enter the value for the pulse output. Floating point number with sign Depends on the nominal diameter:
DN 15 ("): 0.5 l/min (0.1 gal/min) DN 20 ("): 1.0 l/min (0.3 gal/min) DN 25 (1"): 2.0 l/min (0.5 gal/min) DN 50 (2"): 10.0 l/min (2.6 gal/min) The current pulse repetition frequency is calculated from the current flow and the configured pulse value:
Pulse repetition frequency = flow/pulse value Example Flow: 300 l/min (79.25 gal/min) Pulse value: 0.001 l Pulse repetition frequency = 5 000 Pulse/s The maximum pulse repetition frequency is 10 kHz. The Pulse output only outputs positive flow components in the set installation direction. Negative flow components are ignored and not balanced. Configuring the switch output The Switch output submenu contains all the parameters that must be configured for the configuration of the switch output. Navigation Menu: "Application" Output 1 Switch output Menu: "Application" Output 2 Switch output Parameter overview with brief description Parameter Description Polarity Select the switching behavior. Switch output function Selection/input NPN (low-side-switch) Switches load to low side to L-
PNP (high-side-switch) Switches load to high side to L+
Diagnostic behavior The output switches when an event with the status signal F occurs Off On The switch output is permanently switched off (open, non-conductive). The switch output is permanently switched on (closed, conductive). Indicates if a specified limit value has been reached for the process variable. Indicates if a specified limit value has been reached for the process variable. Limit volume flow Limit temperature Range volume flow Range temperature Empty pipe detection Output switches on when empty pipe detection is active. Switch-on value Enter the measured value for the switch-on value. Floating point number with sign Switch-off value Enter the measured value for the switch-off value. Floating point number with sign Endress+Hauser Factory setting PNP (high-side-switch) Diagnostic behavior Country-specific:
1 000 m/h 1 000 gal/min (us) Country-specific:
1 000 m/h 1 000 gal/min (us) 23 Commissioning Picomag IO-Link Hysteresis switched on Hysteresis switched off A0034163 A0034164 1.1 Input variables 1.2 Switch output A B C Range switched on Switch-on point Switch-off value Hysteresis 2.1 Input variables 2.2 Switch output A B C Range switched off Switch-on point Switch-off value Hysteresis A0034165 A0034166 3.1 Input variables 3.2 Switch output A B D Window On-value (lower range limit) Off-value (upper range limit) 4.1 Input variables 4.2 Switch output A B D Window On-value (lower range limit) Off-value (upper range limit) Configuring the voltage output The Voltage output submenu contains all the parameters that must be configured for the configuration of the voltage output. Navigation Menu: "Application" Output 1 Voltage output Menu: "Application" Output 2 Voltage output Parameter overview with brief description Parameter Description Selection/input Factory setting Assign voltage output Select process variable for voltage output. 2 V value 10 V value Enter the lower-range value. Enter the upper-range value. Off Volume flow Temperature Volume flow Floating point number with sign 0 l/min Floating point number with sign Depends on the nominal diameter:
DN 15 ("): 25 l/min (6.6 gal/min) DN 20 ("): 50 l/min (13.2 gal/min) DN 25 (1"): 100 l/min (26.4 gal/min) DN 50 (2"): 750 l/min (198.1 gal/min) 24 Endress+Hauser 10ABtC1.11.2Q / TQ / T10ABtC2.12.2Q / T10AtDD3.13.2BBAQ / T10ABtDD4.14.2AB Picomag IO-Link Commissioning Unidirectional flow measurement (Q) A B Q Lower-range value = 0 Upper-range value Flow A0032995 Current I or voltage U are interpolated linearly between the lower-range value (A) and upper-range value (B). The output range ends at 20.5 mA or 10.5 V. Bidirectional flow measurement (Q) or temperature measurement (T) A B Q Lower value Upper-range value Flow A0032996 Current I or voltage U are interpolated linearly between the lower-range value (A) and upper-range value (B). Rather than having a hard upper and lower limit, the output range ends at 20.5 mA or 10.5 V at the top end, and at 3.8 mA or 1.9 V at the bottom end. Configuring the status input The Digital input submenu contains all the parameters that must be configured for the configuration of the digital input. The input is used to control an action with an external voltage signal. The minimum pulse duration is 100 ms. Navigation Menu: "Application" Output 1 Digital input Endress+Hauser 25 A = 0BQmA / V201016812846420BQmA / V20101688442A126 Commissioning Picomag IO-Link Menu: "Application" Output 2 Digital input Parameter overview with brief description Parameter Description Options Active level Select the switching behavior of the digital input. Assign status input Select process variable for status input. High Low Input reacts to high level Input reacts to low level Off Reset totalizer Resets the totalizer Flow override Flow measured value = 0 Does not affect the temperature measurement Factory setting High Reset totalizer Simulation 9.2.5 The Simulation submenu enables you to simulate, without a real flow situation, different process variables in the process and the device alarm behavior and to verify downstream signal chains (switching of valves or closed-control loops). Navigation Menu: "Diagnostics" Simulation 1 Parameter overview with brief description Parameter Description Selection/input Simulation process variable Activate the simulation of process variables. Volume flow value Temperature value Enter the value for volume flow simulation. Enter the value for temperature simulation. Off On Simulation is deactivated. Simulation is activated. Deactivate the simulation again once the test has Positive floating point number been performed. Positive floating point number Factory setting Off 26 Endress+Hauser Picomag IO-Link Operation 10 Operation 10.1 Reading measured values You can read all measured values using the Measured values submenu. Navigation Menu: "Application" Measured values Parameter overview with brief description Parameter Description Volume flow Displays the volume flow currently measured. Temperature Displays the temperature currently measured. Totalizer Displays the current totalizer counter value. The totalizer continuously totalizes only positive flow measured values in the set flow direction. Negative components are not counted. Reset totalizer Reset the totalizer. Display/options Factory setting Floating point number with sign Floating point number with sign Floating point number with sign Cancel The totalizer is not reset. Reset + totalize The totalizer is reset. Cancel Endress+Hauser 27 Operation Picomag IO-Link 10.2 Configuring Bluetooth The Bluetooth configuration submenu contains all the parameters to configure the Bluetooth connection. Navigation Menu: "System" Bluetooth configuration Parameter overview with brief description Parameter Bluetooth Description Options Enable or disable the Bluetooth wireless technology interface. If the interface is disabled, it can only be re-enabled by tapping the device. Disable Enable Disable the interface. The connection to the measuring device is torn down. Enable the interface. The connection to the measuring device is established. Factory setting Enable 10.3 Administration The Administration submenu contains all the parameters that can be used for the administration of the device. Navigation Menu: "System" Administration Parameter overview with brief description Description Entry/selection/display Factory setting Parameter Set access code Enter access code Enter a user-specific access code to restrict write access to parameters. Max. 4-digit character string comprising numbers, letters and special characters Enter the access code. Restrict write access to parameters in order to protect the device configuration from unauthorized modification. Max. 4-digit character string comprising numbers, letters and special characters Device reset Reset the entire device configuration or some of the configuration to a defined state. Access status tooling Displays the access status. Cancel To factory defaults Restart device Operator Maintenance 0000 0000 Cancel Maintenance 28 Endress+Hauser Picomag IO-Link Diagnostics and troubleshooting 11 Diagnostics and troubleshooting 11.1 General troubleshooting For local display Error Possible causes Solution Local display dark and no output signals Supply voltage does not match the value indicated on the nameplate. Apply the correct supply voltage 34. The polarity of the supply voltage is wrong. Correct the polarity. The connecting cables are not connected correctly. Check the cable connection and correct if necessary. For output signals Error Possible causes Solution Device shows correct value on local display, but signal output is incorrect, though in the valid range. Device measures incorrectly. Configuration error Check and correct the parameter configuration. Configuration error or device is operated outside the application. 1. Check and correct parameter configuration. 2. Observe limit values specified in the "Technical Data". For access Error Possible causes Solution No connection established via Bluetooth Bluetooth communication is disabled No communication with device via SmartBlue App No Bluetooth connection 1. Check whether the Bluetooth logo is visible on the local display or not. 2. Re-enable Bluetooth communication by tapping the device. Enable Bluetooth function on smartphone or tablet. The device is already connected with another smartphone/tablet. Login via SmartBlue App not possible Device cannot be operated via SmartBlue App Device is being put into operation for the first time Enter initial password (device serial number) and change. Incorrect password entered Enter correct password. Password forgotten Contact Endress+Hauser Service. Endress+Hauser 29 Diagnostics and troubleshooting Picomag IO-Link 11.2 Diagnostic information on local display 11.2.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. Diagnostic message Alarm Function check Warning A0033011 A0033010 A0033009 If two or more diagnostic events are pending simultaneously, only the message of the diagnostic event with the highest priority is shown. Status signals The status signals provide information on the state and reliability of the device by categorizing the cause of the diagnostic information (diagnostic event). The status signals are categorized according to VDI/VDE 2650 and NAMUR Recommendation NE 107: F = Failure, C = Function Check, S = Out of Specification, M
= Maintenance Required Symbol Meaning Failure An operating error has occurred. The measured value is no longer valid. Function check The device is in simulation mode. Out of specification The device is being operated:
Outside its technical specification limits (e.g. outside the process temperature range) Outside of the configuration carried out by the user (e.g. maximum flow in parameter 20 mA value) Diagnostic behavior Diagnostic message Meaning Alarm Measurement is interrupted. Signal outputs and totalizers assume the defined alarm condition. A diagnostic message is generated. Function check Process measured values are simulated to test the outputs/wiring. Warning Measurement is resumed. Measuring operation with limited accuracy The signal outputs and totalizers are not affected. A diagnostic message is generated. 30 Endress+Hauser XxxxxxxF180TEMP. CIRC. FAIL.XxxxxxxC446I/O 1 OVERLOADXxxxxxxS962EMPT Y PIPE???Picomag IO-Link Diagnostics and troubleshooting Diagnostics behavior of outputs Output Switch output Pulse output Totalizer Current output Voltage output IO-Link Diagnostic behavior Setting for reporting events with the status signal F Switch output is switched on if an event occurs No further response to events with other status signals Pulse output stops if events with the status signal F occur No further response to events with other status signals Totalizing stops if events with the status signal F occur No further response to events with other status signals 3.5 mA output to report events with the status signal F No further response to events with other status signals 1.75 V output to report events with the status signal F No further response to events with other status signals All events reported to the Master Events read and processed further by the Master 11.3 Overview of diagnostic events Diagnostic event Event text Reason Remedial measures Status signal [ex-
factory]
181 180 201 283 446 447 485 453 441 444 443 442 445 448 962 Coil. circ. fail. Coil/frequency failure Coil current PWM outside tolerance range Replace the measuring device. Temp. circ. fail. Device fail. Temperature sensor open circuit/short-
circuit Replace the measuring device. No communication to ADC/Nordic/BMA Replace the measuring device. Memory fail. CRC failure Reset to factory settings. Overload at output 1 Increase load impedance. Overload at output 2 Increase load impedance. I/O 1 overload I/O 2 overload Simulation act. Measured value simulation active (via remote configuration) Flow override Flow override active (via auxiliary input) I-Out 1 range I-output 1 at range limit Adjust parameter or process. U-Out 1 range P-Out 1 range U-output 1 at range limit P-output 1 at range limit Adjust parameter or process. Adjust parameter or process. I-Out 2 range I-output 2 at range limit Adjust parameter or process. U-Out 2 range P-Out 2 range Empty pipe U-output 2 at range limit P-output 2 at range limit Pipe is completely or partially empty Adjust parameter or process. Adjust parameter or process. Adjust the process. F F F F C C C C S S S S S S S Endress+Hauser 31 Diagnostics and troubleshooting Diagnostic event 834 841 Event text Reason Remedial measures Temperat. range Flow range Medium temperature outside the permitted range Flow rate outside the permitted range Adjust the process. Adjust the process. S S 11.4 Pending diagnostic events Navigation Menu: "Diagnostics" Diagnostics Picomag IO-Link Status signal [ex-
factory]
Parameter overview with brief description Parameter Prerequisite Description Actual diagnostic A diagnostic event has occurred. Displays the current diagnostic event along with the diagnostic information. If two or more messages occur simultaneously, the message with the highest priority is shown on the display. Display Symbol for diagnostic behavior, diagnostic code and short message. 11.5 Device information The Device info submenu contains all parameters that display different information for device identification. Navigation Menu: "System" Device info Parameter overview with brief description Parameter Device name Device tag Description Display Displays the name of the measuring device. Picomag Shows name of measuring point. Max. 32 characters, such as letters, numbers or special characters (e.g. @,
%, /). Serial number Displays the serial number of the measuring device. Max. 11-digit character string comprising letters and numbers. Firmware version Displays the device firmware version installed. Character string in the format xx.yy.zz Extended order code Displays the extended order code. Character string composed of letters, numbers and certain punctuation marks (e.g. /). 11.6 Firmware history Release date Firmware version Firmware changes 09.2017 01.00.zz Original firmware Documentation type Documentation Operating Instructions BA01697D/06/EN/01.17 32 Endress+Hauser Picomag IO-Link Accessories 12 Accessories Various accessories, which can be ordered with the device or subsequently from Endress
+Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com. Endress+Hauser 33 Technical data Picomag IO-Link 13 Technical data 13.1 Input Measured values Volume flow Temperature Totalizer Measuring range (upper-range value E) DN 15 ("): 25 l/min (6.6 gal/min) DN 20 ("): 50 l/min (13.2 gal/min) DN 25 (1"): 100 l/min (26.4 gal/min) DN 50 (2"):750 l/min (198.1 gal/min) 13.2 Output Output Current output Voltage output Digital output (SmartBlue App) Signal on alarm 13.3 Power supply Max. load 500 The load may not be any greater 500 The load resistance may not any smaller Bluetooth wireless technology The device has a Bluetooth wireless technology interface and can be operated and configured via the SmartBlue App. The range under reference conditions is 10 m (33 ft) Incorrect operation by unauthorized persons is prevented by means of encrypted communication and password encryption. The Bluetooth wireless technology interface can be deactivated. Status signal (as per NAMUR Recommendation NE 107) Plain text display with remedial action Supply voltage range Power consumption 18 to 30 VDC (SELV, PELV, Class 2) Max. 3 W (excluding outputs IO1 and IO2) 13.4 Performance characteristics 2 % o.r. 0.5 % o.f.s. 0.2 % o.r. (95% confidence interval, measuring time 30 s) The response time depends on the configuration (damping). 2.5 C 0.5 C (95% confidence interval, measuring time 30 s) Volume flow measurement Maximum measured error Repeatability Response time Medium temperature measurement Maximum measured error Repeatability Installation 13.5 11 13.6 Environment Ambient temperature range Storage temperature 10 to +60 C (+14 to +140 F) 25 to +85 C (13 to +185 F) 34 Endress+Hauser Picomag IO-Link Technical data Degree of protection Shock resistance Vibration resistance Electromagnetic compatibility (EMC) IP65/67 20 g (11 ms) as per IEC/EN60068-2-27 Acceleration up to 5 g (10 to 2 000 Hz) as per IEC/EN60068-2-6 In accordance with IEC/EN61326 and/or IEC/EN55011 (Class A) 13.7 Process Medium temperature range Medium properties Pressure 10 to +70 C (+14 to +158 F) Liquid, conductivity > 20 S/cm Max. 16 barrel 13.8 Mechanical construction A0033012 Dimensions in SI units DN
[mm]
15, 20, 25 A
[mm]
110 B
[mm]
73 C
[mm]
40.5 D
[mm]
69.5 E
[mm]
M12 1 F
[mm]
43 50 200 113 80 120 M12 1 58 Dimensions in US units DN
[in]
, , 1 A
[in]
4.33 B
[in]
2.87 C
[in]
1.59 D
[in]
2.74 E
[in]
M12 1 2 7.87 4.45 3.15 4.72 M12 1 Weight in SI units F
[in]
1.69 2.28 G
[mm]
"
"
1"
2"
G
[in]
"
"
1"
2"
H
[mm]
56 I
[mm]
56 86 86 H
[in]
2.20 I
[in]
2.20 3.39 3.39 DN
[mm]
15 20 25 50 Weight
[kg]
0.34 0.35 0.36 1.55 Endress+Hauser 35 BACDEFGHI==Technical data Picomag IO-Link Weight in US units DN
[in]
"
"
1"
2"
Materials Component Measuring tube Electrodes, temperature sensor Process connection Housing Seal Display window 13.9 Operability Weight
[lbs]
0.75 0.77 0.79 3.42 Material PEEK 1.4404/316L 1.4404/316L 1.4404/316L FKM Polycarbonate Onsite display The device has an onsite display:
A0032991 Tag name (configurable) 1 2 Measured variable 1 (configurable), with sign 3 Measured variable 2 (configurable), with sign 4 5 6 Active Bluetooth connection Active I/O-Link connection Device status Display element A maximum of 2 readings from the 3 measured variables (volume flow, temperature, totalizer) can be displayed Remote operation Via Bluetooth wireless technology Digital communication Via IO-Link 13.10 Certificates and approvals 36 Endress+Hauser 28.5623.6l / minCXxxxxxx123456 Picomag IO-Link CE mark Technical data The measuring system is in conformity with the statutory requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. Radio approval The measuring device has radio approval. For detailed information on the radio approval, see the Appendix 38 Pressure Equipment Directive Devices not bearing this marking (PED) are designed and manufactured according to good engineering practice. They meet the requirements of Art. 4, Par. 3 of the Pressure Equipment Directive 2014/68/EU. The range of application is indicated in tables 6 to 9 in Annex II of the Pressure Equipment Directive 2014/68/EC. CULUS listing The measuring device is UL-listed. Endress+Hauser 37 Appendix Picomag IO-Link 14 Appendix 14.1 Radio approvals 14.1.1 Europe This device meets the requirements of the Telecommunications Directive RED 2014/53/EU:
EN 300 328 V2.1.1 EN 301 489-1 V1.9.2 EN 301 489-17 V2.2.1 EN 62311: 2008 14.1.2 Canada and USA English This device complies with Part 15 of the FCC Rules and with Industry Canada licence-
exempt RSS standard(s). Operation is subject to the following two conditions:
This device may not cause harmful interference, and This device must accept any interference received, including interference that may cause undesired operation. Changes or modifications made to this equipment not expressly approved by Endress+Hauser Flowtec AG may void the user's authorization to operate this equipment. Franais Le prsent appareil est conforme aux CNR d'industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorise aux deux conditions suivantes :
L'appareil ne doit pas produire de brouillage, et L'utilisateur de l'appareil doit accepter tout brouillage radiolectrique subi, mme si le brouillage est susceptible d'en compromettre le fonctionnement. Les changements ou modifications apportes cet appareil non expressment approuve par Endress+Hauser Flowtec AG peut annuler l'autorisation de l'utilisateur d'oprer cet appareil. 14.1.3 Other countries Other national approvals are available on request. 38 Endress+Hauser Picomag IO-Link Index A About this document . 4 Accessories . 33 Appendix . 38 Application Media . 6 B Basic safety instructions . 6 Bidirectional flow measurement (Q) or temperature measurement (T) . 25 C CE mark . 7, 37 Commissioning . 19 Configuring system units . 21 Configuring the current output . 22 Configuring the display . 20 Configuring the IO modules . 22 Configuring the measuring device . 19 Configuring the pulse output . 22 Configuring the status input . 25 Configuring the voltage output . 24 Connecting the measuring device . 15 Connection conditions . 13 Current output configuration version . 14 Current version data for the device . 18 D Declaration of Conformity . 7 Designated use . 6 Device information . 32 Device master file . 18 Diagnostic behavior . 30 Diagnostic information on local display . 30 Diagnostic message . 30 Diagnostics and troubleshooting . 29 Dimensions in SI units . 35 Dimensions in US units . 35 Document function . 4 Documentation . 5 E Electrical connection . 13 F Field of application Residual risks . 6 Firmware history . 32 I Identifying the measuring device . 8 Incoming acceptance . 8 Inlet and outlet runs . 11 Inspection Received goods . 8 Installation . 11 Index Installation conditions . 11 IO-Link configuration version . 15 M Materials . 36 Mounting location . 11 Mounting position . 11 Mounting the measuring device . 11 O Operation . 27 Operation options . 17 SmartBlue App . 17 Operational safety . 7 Overview of device description files . 18 Overview of diagnostic events . 31 Overview of the operating menu . 19 P Packaging disposal . 10 Pending diagnostic events . 32 Pin assignment, device plug . 13 Post-connection check . 16 Pressure Equipment Directive . 37 Product safety . 7 R Radio approval . 37 Radio approvals . 38 Reading measured values . 27 Registered trademarks . 5 Requirements for connecting cables . 13 Requirements for personnel . 6 S Setting the installation direction and measurement . 21 Status input configuration version . 15 Status signals . 30 Storage conditions . 10 Storage temperature . 10 Switch/pulse output configuration version . 14 Switching ON the measuring device . 19 Symbols used . 4 System integration . 18 T Technical data . 34 Certificates and approvals . 36 Environment . 34 Input . 34 Installation . 34 Mechanical construction . 35 Operability . 36 Output . 34 Performance characteristics . 34 Power supply . 34 Process . 35 Endress+Hauser 39 Index Picomag IO-Link Temperature range Tools Storage temperature . 10 Transport . 10 Transporting the measuring device . 10 Troubleshooting General . 29 U Unidirectional flow measurement (Q) . 25 Using measuring device Borderline cases . 6 Incorrect use . 6 V Voltage output configuration version . 14 W W@M Device Viewer . 8 Weight Transport (notes) . 10 Weight in SI units . 35 Weight in US units . 36 Workplace safety . 6 40 Endress+Hauser www.addresses.endress.com
frequency | equipment class | purpose | ||
---|---|---|---|---|
1 | 2017-09-27 | 2402 ~ 2480 | DTS - Digital Transmission System | Original Equipment |
app s | Applicant Information | |||||
---|---|---|---|---|---|---|
1 | Effective |
2017-09-27
|
||||
1 | Applicant's complete, legal business name |
Endress&Hauser Flowtec AG
|
||||
1 | FCC Registration Number (FRN) |
0025620105
|
||||
1 | Physical Address |
Kaegenstrasse 7
|
||||
1 |
Reinach, N/A 4153
|
|||||
1 |
Switzerland
|
|||||
app s | TCB Information | |||||
1 | TCB Application Email Address |
b******@phoenix-testlab.de
|
||||
1 | TCB Scope |
A4: UNII devices & low power transmitters using spread spectrum techniques
|
||||
app s | FCC ID | |||||
1 | Grantee Code |
2AIMC
|
||||
1 | Equipment Product Code |
DMA
|
||||
app s | Person at the applicant's address to receive grant or for contact | |||||
1 | Name |
E****** H******
|
||||
1 | Telephone Number |
+4161******** Extension:
|
||||
1 | Fax Number |
+4161********
|
||||
1 |
E******@flowtec.endress.com
|
|||||
app s | Technical Contact | |||||
n/a | ||||||
app s | Non Technical Contact | |||||
n/a | ||||||
app s | Confidentiality (long or short term) | |||||
1 | Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | Yes | ||||
1 | Long-Term Confidentiality Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | No | ||||
if no date is supplied, the release date will be set to 45 calendar days past the date of grant. | ||||||
app s | Cognitive Radio & Software Defined Radio, Class, etc | |||||
1 | Is this application for software defined/cognitive radio authorization? | No | ||||
1 | Equipment Class | DTS - Digital Transmission System | ||||
1 | Description of product as it is marketed: (NOTE: This text will appear below the equipment class on the grant) | Flowmeter with Bluetooth LE | ||||
1 | Related OET KnowledgeDataBase Inquiry: Is there a KDB inquiry associated with this application? | No | ||||
1 | Modular Equipment Type | Does not apply | ||||
1 | Purpose / Application is for | Original Equipment | ||||
1 | Composite Equipment: Is the equipment in this application a composite device subject to an additional equipment authorization? | No | ||||
1 | Related Equipment: Is the equipment in this application part of a system that operates with, or is marketed with, another device that requires an equipment authorization? | No | ||||
1 | Grant Comments | Output power listed is max peak conducted. | ||||
1 | Is there an equipment authorization waiver associated with this application? | No | ||||
1 | If there is an equipment authorization waiver associated with this application, has the associated waiver been approved and all information uploaded? | No | ||||
app s | Test Firm Name and Contact Information | |||||
1 | Firm Name |
PHOENIX TESTLAB GmbH
|
||||
1 | Name |
H**** B****
|
||||
1 | Telephone Number |
49-52********
|
||||
1 | Fax Number |
49-52********
|
||||
1 |
o******@phoenix-testlab.de
|
|||||
Equipment Specifications | |||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Line | Rule Parts | Grant Notes | Lower Frequency | Upper Frequency | Power Output | Tolerance | Emission Designator | Microprocessor Number | |||||||||||||||||||||||||||||||||
1 | 1 | 15C | 2402.00000000 | 2480.00000000 | 0.0005000 |
some individual PII (Personally Identifiable Information) available on the public forms may be redacted, original source may include additional details
This product uses the FCC Data API but is not endorsed or certified by the FCC