1 | Users Manual | Users Manual | 4.79 MiB | February 06 2011 |
User Manual User Manual Copyright 2007 IntroMedic Co., Ltd. MM1100-U-1105 Version 1.00 Date : 2011-05-11 Warranty Trademarks MiroCam, MiroView, IntroMedic, and the associated logos are the registered trademarks or trademarks of IntroMedic Co., Ltd. IntroMedic Co., Ltd. July 2007. Except as required by applicable copyright laws; any use of the IntroMedic trademarks, or any reprinting, reproduction, modification, referencing and translations of the User Manual, without the prior written approval of IntroMedic Co., Ltd. is strictly prohibited. Warranty Every effort has been made to ensure the information contained in this User Manual is accurate, and is believed to be correct at time of printing. IntroMedic reserves the right to change any content contained with this User Manual without prior notice. IntroMedic Co., Ltd. warrants the product against defects in material and workmanship for a period of twelve (12) months from the date of sale, unless different local regulations apply. IntroMedic Co., Ltd. will repair or replace products that are ascertained by IntroMedic to have defects during the warranty period. IntroMedic Co., Ltd. is not liable for the defects occurred by misuse, careless handling, unauthorized modifications or erroneous use, or any use that is non-compliant with instructions detailed within this User Manual. This includes use of the product in non-appropriate locations or conditions. Any other warranties are neither represented here nor recognized by implication. To validate the warranty, please complete product registration with the local authorized IntroMedic distributor. Page 5 Warranty Exclusive warranty service The warranty service provided hereby is applicable exclusively to the purchaser of the product. IntroMedic will only warranty the product for purposes and usage as defined in this User Manual. Any usage not heeding the warnings, cautions and recommended usages as defined in this manual will nullify the warranty. Support For warranty or repair service please contact the local authorized IntroMedic distributor. For customer service or support please contact your point of purchase or IntroMedic Co., Ltd. Service agreements are only applicable to products of IntroMedic Co., Ltd. IntroMedic Customer Service TEL: 82-2-801-9300 FAX: 82-2-801-9330 http://www.intromedic.com E-mail: helpdesk@intromedic.com Safety Non-compliance with the users manual, unauthorized modifications of the product or replacement of parts, and/or opening of the product casing is prohibited and may be hazardous. Declaration on translation of labeling Based on buyer's request, IntroMedic will provide the labeling, such as ID label, Instruction for use and other promotional materials, which is translated into the national language(s) of European countries. And it will be evaluated by the local language expert, and will be confirmed by the native speaker of the local language. Page 6 CONTENTS 1. SAFETY INFORMATION 1.1 Warnings 1.2 Symbols for Safety 1.3 Function Symbols 1.4 Notes for Safe Use 2. SYSTEM OVERVIEW Intended Purpose 2.1 2.2 Observable Diseases 2.3 Product Lifetime 2.4 Product Specification Contents 3 3 4 5 9 3 3 4 6 7 3. PERFORMING CAPSULE ENDOSCOPY 3 3 3 4 6 7 16 17 3.1 General Comments 3.2 Safety Warnings 3.3 Examination Preparation 3.4 Patient Preparation 3.5 Sensor Placement & Capsule Ingestion 3.6 Uploading Images 3.7 Receiver Management Post Procedure Page 7 Contents 4. USING MIROVIEW 4.1 MiroView H/W Specifications 4.2 Overview MiroView Server 4.3 Overview - MiroView Operator 4.4 Overview - MiroView Client 4.5 Overview - MiroView Viewer 4.6 Performing Capsule Endoscopy 4.7 How to import MiroView 1.0 files 5. USING MIROVIEW RTV 5.1 Overview 5.2 Using MiroView RTV 6. USING WIRELESS MIROVIEW RTV 6.1 Overview 6.2 Using Wireless MiroView RTV 3 5 8 10 18 31 32 52 3 3 6 3 3 6 Page 8 Safety Information 1 Safety Information Chapter 1 1- 2 Chapter 1 Safety Information 1. SAFETY INFORMATION 1.1 Warnings together with IEC 60601-1, the Collateral Standard MiroCam has been manufactured to conform with the International Standard for Medical Electrical Equipment: General Requirements for Safety for Electromagnetic Compatibility Requirement and Tests IEC 60601-1-2. MiroCam has been manufactured to conform to the electric shock, fire and mechanical hazard standards as defined in CAN/CSA C22.2 NO.601.1. Based on request of the buyer, IntroMedic will provide the labeling, such as ID labels, and the User Manual in the national language(s) of European countries. Translated documents will be evaluated by a local language expert, and will be confirmed by a native speaker of the respective national language. Safety Symbols: The User Manual incorporates various safety symbols to ensure safe and correct use of the product and to prevent any personal injury or property damage. These symbols are defined in the following table:
WARNING WARNING indicates a potential hazard that, if not avoided, could result in serious personal injury or damage to the product. CAUTION CAUTION indicates a potential hazard that, if not avoided, could result in minor personal injury or damage to the product. NOTE NOTE does not indicate potential hazards as in Caution or Warning, but contains the installation, operation or maintenance of the product. information regarding important 1-3 Safety Information Chapter 1 1.2 Symbols for Safety This section describes a set of symbols that the IEC (International Electrotechnical Commission) has established for medical electronic equipment to classify connections and warnings of any potential hazards. IEC 348: Notice for the user to pay special attention to the following details IEC 878-02-03: Indicates that this is classified into Type BF equipment EN 980: Denotes Date of Manufacture EN 980: Denotes Address of Manufacture IEC60601, ANNEX D: Denotes ON status of main power switch IEC60601, ANNEX D: Denotes OFF status of main power switch EN 980: Denotes serial number IEC 417-5031: Denotes DC (Direct Current) IEC 417-5032: Denotes AC (Alternating Current) Denotes Ampere, the unit of current Denotes Volt, the unit of Voltage Denote Hertz, the unit of Frequency IEC 417-5021: Denotes potential equalization terminal Single Use Only Use by date 1- 4 SN A V Hz Chapter 1 Safety Information 1.3 Function Symbols 1.3.1 MiroView Function Symbols following The MiroView software. table describes symbols or icons used in the Symbol Description Symbol Description Connect to the receiver and open receiver control screen. Open the patient data screen. Open screen to review for a specific patient Place a landmark in an image being reviewed. Show images in full screen Show images in the Single Screen Show images in the Quad Screen. Show images in the Map View Play the selected images only. View images via Express Mode function. Open the report screen to create a patient capsule endoscopy report. Open screen to export
(save externally) selected image data for a specific patient. Select and save an image being reviewed. Open the Color Enhancement Window Time Display Show images in the Dual Screen. Show images in the Range Mode. Play all images. SGIB - Play the images captured via Suspected GI Bleeding function. 1-5 Safety Information Chapter 1 Symbol Description Symbol Description Move to the previous image. Play images in reverse order. Play Move to the next captured image. Open the Positioning Window Expand/Shrink Capture Box Show Circumference Markers Add a Line on the image Color Palette Redo the action Move to the previous captured image. Stop playback of images. Play images in sequential order. Move to the next image. Open the Comparison Window Statistic Box Add a Circle on the image Delete selected Circle /
Line Undo the action Check selected captured images 1- 6 Chapter 1 Safety Information Symbol Description Symbol Description Uncheck selected captured images Review Export Review Export Captured Images with EXMIF Export Segments as Image / Video / MiroVideo Select the area as a Segment Print Save as PDF Open PACS function
(Registered User Only) Export Captured Images as Image / Video /
MiroVideo Export Case as Video /
MiroVideo Deselect the area from Segment 1-7 Safety Information Chapter 1 1.3.2 MiroCam Receiver Function Symbols Symbol Description SIG INI BAT Indicates status of signal from capsule Green : Signal is being received from capsule Yellow : Signal is not being received from capsule Initialization status of Receiver Unit Green : Receiver is initialized Yellow : Receiver is not initialized Battery Status Green : Fully charged Yellow : Not charged 1- 8 Chapter 1 Safety Information 1.4 Notes for Safe Use n n n Follow the safety instructions included in this User Manual and clinical precautions advised by the medical professional. The manufacturer is not liable for harm or damage caused by improper, unauthorized, unprofessional or inexpert use of the device and/or product. IntroMedic Co., Ltd. for physical harm or equipment problems caused by the users careless operation or mismanagement of the device and/or product. is NOT responsible n Users MUST have read and understood the User Manual. ONLY trained and qualified medical professionals or authorized representatives of IntroMedic Co., Ltd. may operate the system. n User Manual must ALWAYS be with the equipment. This is the USERS RESPONSIBILITY. n CAUTION:
foreign substances fluids, disinfecting cleanser and such substances may harm the equipment, and should not enter the equipment. including water, cleaning n ONLY authorized personnel may perform repairs. Never attempt to open covers, panels or casings. n DO NOT crease, bend, fold or twist the data cables. Take care to guard n them against mechanical stress (e.g. wheels or heels)!
The sensor pads, receiver, data cables, and capsules must not be exposed to mechanical shock (e.g. by dropping). Any damage caused that way will void the product warranty. n CAUTION: Damage/injuries to the sensor pad or data cable may repaired items MUST be cause a safety hazard. Damaged IMMEDIATELY. n DO NOT handle fluids in the vicinity of the system. n When using a cart purchased elsewhere, ensure the brake or latch guard is in use to prevent the wheels from rolling. 1-9 Safety Information Chapter 1 n DO NOT USE in moist or damp places. n DO NOT operate the equipment with wet hands. n Avoid using environments. the equipment in extreme temperatures or humid n DO NOT keep the equipment or carry out the procedure in places such as areas exposed to direct sunlight, vicinity of heaters, vicinity of chemical materials or gases, areas moist/damp or dusty, or poorly ventilated areas. n DO NOT disassemble or open the equipment without permission. This will invalidate the warranty. n DO NOT carry out the procedure in areas with high vibrations or in environments where high electro-magnetic waves are generated. n DO NOT pull out the power cord by grabbing the cable. When disconnecting the power cord, grasp the plug, and pull out. This prevents short-circuits, disconnection, or cord damage. n CAUTION: Verify that the power voltage supplied from the power receptacle matches with the voltage the system requires. Check Voltage and Frequency on the AC/DC adaptor. n CAUTION: Turn off the power switch before connecting the sensor pad. n DO NOT discard sensor pads, cables and connectors with general waste. Discard separately as industrial or medical waste. n CAUTION: Discard the battery according to the regulations of industrial waste. DO NOT discard with general waste. n DO NOT carry out the procedure simultaneously with other procedures using medical products or equipment. n DO NOT use for purposes other than medical treatment. n DO NOT connect the USB cable to the receiver while the receivers data cable and sensor pads are still connected n DO NOT charge the receiver while the receivers data cable and sensor pads are still connected. 1- 10 Chapter 1 Safety Information n Connect USB cable to receiver only after mounting it on charger. n DO NOT install any other programs onto the workstation utilized for review and diagnosis of patient image data (i.e computers with the MiroView software). The capsule is disposable and should not be reused. The capsule and sensor pads are medical waste, and should be disposed of according to local regulations or WEEE directive on waste disposal. n n n Only use the capsule, receiver, data cables and sensor pads in the medical environment condition. n All products connected with the MiroCam Endoscope system must be compliant with requirements of IEC60950-1 or UL certifications. n Check the outer surface of the endoscope for rough surfaces, sharp edges, or protrusions which may cause a safety hazard. n Please dont try to cut the label. 1-11 Safety Information Chapter 1 1.4.1 Environmental Conditions for Operation n Temperature n Relative humidity n Atmospheric pressure
: +0 - +40
: 45% - 75%
: 700hPa to 1060hPa WARNING DO NOT operate power the equipment in stations, X-ray the vicinity of generators, devices, and broadcasting stations where high levels of electro-magnetic waves are generated. The electro-magnetic waves can cause equipment malfunctions. CAUTION If the equipment has been brought from a cold environment (stock room, airfreight) into a warm room, initial activation should take place after a few hours, to allow temperature adjustment and balance and evaporation of condensed humidity. for in WARNING DO NOT operate the equipment in the vicinity of heat sources, strong electric or magnetic fields (close to a transformer), or near instruments generating high-
frequency signals. WARNING Do not use MiroCam alongside or together with medical devices or procedures involving electrical currents. Do not use MiroCam with h.f. surgical equipment. It may result in burns at the site of the electrodes and possible damage to the capsule and receiver. Do not use the unit in close radius (within 1 m) of short wave or microwave therapy equipment. It may produce instability in the captured image. WARNING This device is a Class B device according to EN60601-1-2 standards. This equipment can cause radio interference in residential areas. In this case, the owner (or operator) can be held responsible to take appropriate measures or take proper measures for compensation. 1- 12 Chapter 1 Safety Information 1.4.2 Safety Precaution CAUTION
- Make sure the environment is without interference from electromagnetic fields.
- Make sure the environment is without noise and
- DO NOT carry out the procedure while using other equipments, devices or products.
- The instruction for use of the sensor pads MUST be vibration. observed.
- DO NOT use on patients with pacemakers or defibrillators. CAUTION DO NOT use the capsule if the package is unsealed. n DO NOT reuse a used capsule. n To prevent unexpected accidents like fire or explosion, do not use any product near or in the presence of inflammable or ignitable substances. n DO NOT disassemble the equipment case nor open the cover. In case service is required, please contact IntroMedic customer support or local point of sale immediately. n Only the accessories authorized and designed by IntroMedic Co., Ltd. should be used with this equipment. Faults resulting from the usage of unapproved or inappropriate accessories are not guaranteed against. This equipment may have an effect on other products or be effected by other products. Follow your doctors instructions and abide by the guidelines in the User Manual. n n n DO NOT try to upload the data while the data cables are still connected to the receiver. n DO NOT charge the rechargeable battery in the receiver while the data cable and sensor pads are still connected to the receiver. 1-13 Safety Information Chapter 1 n Stay away from high frequency radiation sites (such as high voltage, radar, installation power plants, MRI, CT or electric blankets etc.) during your capsule endoscope procedure. (It may result in serious side effects requiring an emergency operation.) In case of any symptoms of abdominal pain, vomiting, fever, heart trouble, dizziness or seizure during or after the capsule endoscope procedure, please notify your doctor. n n Always check the connection between the receiver and the data cable. n Always check that the battery in the receiver is fully charged before use. n DO NOT use the capsule if the package is unsealed. n After ingesting the capsule, always check whether the capsule has been e xcreted. n n Prior to undergoing the capsule endoscopy procedure, patients with diabetes must be informed via a medical professional regarding appropriate medication & dosage. For more accurate data and better analysis, patients can have solid food for lunch, but must have liquid food for dinner on the day prior to the procedure. The patient must fast for 11 hours prior to ingesting the capsule. If recommended by the physician, the patient may take a laxative prior to the procedure. n n DO NOT smoke for at least 12 hours prior to the capsule endoscope procedure. n DO NOT apply body lotion before the procedure. n DO NOT bite the capsule. n Avoid excessive physical activity during the capsule endoscope procedure. 4 hours after ingesting the capsule, patients may have liquid food. It to drink mineral water (non-colored is recommended liquid) n n 1- 14 Chapter 1 Safety Information frequently, so that it may raise the possibility to take much better quality of the image. n When undergoing the capsule endoscope procedure, DO NOT make physical contact with another person undergoing the same procedure. n During operation of the receiver, DO NOT touch the receiver, or get the receiver wet. n Only use the provided batteries, and never remove the battery from the receiver during the procedure. n During upload of the data recorded in the receiver to the workstation, avoid disconnecting the USB. This may damage the patients data. n Always confirm that the USB is connected by checking the Receiver screen on the MiroView software. n Always check the AC Power range before use the workstation. n DO NOT touch AC Power code with wet hands. n DO NOT open the receiver bag or touch receiver outside of the hospital. This device is intended for patient over the age of 18. n WARNING The Capsule takes a picture for 11 hours and gets naturally excreted in about 24 hours under normal conditions. If the capsule has not been excreted from the patient within 72 hours, patient should contact the physician. After examining, the physician may need to perform a surgical operation or treatment to remove the capsule. WARNING Before moving the system, always make sure to disconnect the monitor from the main system, and then safely move the main system and monitor separately. Connect the main system and monitor only after the hardware is fully installed, secure and stable. 1-15 Safety Information Chapter 1 1.4.3 Cleaning and Maintenance n System and accessories
- All products should be cleanly maintained. For cleaning, rub them lightly with a soft cloth wet with warm water at least once a week. Do not use organic solvents such as lacquer, thinner, ethylene and oxide because they can damage the equipment. Be careful that foreign substances do not enter the main system when cleaning.
- ALWAYS operate the equipment under sanitary environmental conditions. DO NOT use heat or gas for disinfection of the capsule. n Service Document If required, or upon request, the local IntroMedic Distributor (authorized IntroMedic Representative) may provide block diagrams, lists of spare parts, descriptions, adjustment instructions or other related information which may help qualified technical personnel in repairing specified parts of the equipment which have been defined repairable by IntroMedic Co., Ltd. n Moving the Equipment
- CAUTION when moving equipment.
- WARNING: Excessive impact/shock causes internal damage.
-
If wiring is connected/disconnected when moving, check the exact wiring status after moving. If damage to the equipment is discovered after moving, immediately contact IntroMedic or local Distributor.
-
1- 16 Overview & Intended Usage 2 Overview & Intended Usage Chapter 2 2- 2 Chapter 2 Overview & Intended Usage 2. SYSTEM OVERVIEW MiroCam is an orally ingested capsule endoscope designed to capture and wirelessly transmit images of the lining of the small intestine. Captured images are viewed via the MiroView software for diagnosis of diseases related to the small intestine. The MiroCam capsule endoscope system consists of the MiroCam capsule, MiroCam receiver, MiroView software, and accessories. 2.1 Intended Purpose MiroCam is intended for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule
(containing a light source, camera, transmitter, and battery), an electrode array, a receiving/recording unit, a data storage device, computer software to process the images (MiroView), and accessories. 2-3 Overview & Intended Usage Chapter 2 2.1.1 Contraindications n Not intended for patients with known or suspected gastrointestinal tract obstructions, perforations, strictures or fistular. n Not intended for patients who have difficulty ingesting food or pills
(dysphagia). n Not intended for patients who have difficulty communicating. n Not intended for patients with indigestion, or slow gastric emptying. n Not intended for patients who may be affected by electromagnetic radiation, such as pregnant women, infants, and patients with heart disease or epilepsy. n Not intended for patients with diverticulosis in the Small Bowel. n Not intended for patients who are recommended against having the procedure by a Physician. n Not intended for patients with pacemakers and/or defibrillators. n This device is intended for patient over the age of 18. 2.2 Observable Diseases 2.2.1 Obscure gastrointestinal bleeding (OGIB) n Angioectasia n Meckel diverticulum n NSAIDs 2.2.2 Crohns Disease 2- 4 Chapter 2 Overview & Intended Usage 2.2.3 Small Bowel Tumors Familial adenomatous polyposis n Polyps n Peutz-Jeghers Syndrome n n Lymphoma n Carcinoid n Gastro-Intestinal Stromal Tumor (GIST) n n Hemangioma Lipoma 2.2.4 Inflammation Intestinal Tuberculosis Typhoid Fever n n n Angeitis n Amyloidosis n Eosinophilic enteritis 2.2.5 Celiac Disease 2.2.6 Behcets Disease 2.2.7 Cowden Disease 2.2.8 Cronkhite-Canada Syndrom 2-5 Overview & Intended Usage Chapter 2 2.3 Product Lifetime 2.3.1 Maximum Storage Time of MiroCam Capsule n MC1000-C : 1 Year n MC1000-F : 1 Year n MC1000-G : 1 Year 2.3.2 MiroCam Receiver, MR1000 n MiroCam Receiver Battery : 4.62 Year 2- 6 Chapter 2 Overview & Intended Usage 2.4 Product Specification 2.4.1 MiroCam Capsule n Component Component Model Name Quantity Capsule MC1000 1 Piece (Optional) MC1000-F MC1000-G Sensor Pads MC1000-S 10 Piece n Specification Item Specification 11 X 24 mm 3.4 0.05g Capsule Unit Size Weight
(MC1000, MC1000-F, MC1000-G) Frame Rate 3 frame/sec Field of View 150 degrees Battery Life 12 hours Pixels Lighting 102,400 6 LEDs Communication Method Human Communication Body 2-7 Overview & Intended Usage Chapter 2 2.4.2 MiroCam Receiver n Component Component Model Name Quantity Receiver Unit Battery Pack Data Cable Receiver Bag System Carrying Case Battery Charger Adaptor Power Code Measuring Tape USB Cable MR1100 MR1100-B MR1000-D MR1000-G MR1000-A MR1000-C MR1000-T MR1000-P MR1000-M MR1000-U 1 piece 2 piece 1 piece 1 piece 1 piece 1 piece 1 piece 1 piece 1 piece 1 piece n Specification Item Specification Receiver Unit Battery Life 12 Hours
(MR1100) Size Weight 83.5 X 138 X 28 mm 350g(including Battery) 2- 8 Chapter 2 Overview & Intended Usage Item Specification Battery Pack Type Lithium-Ion storage cell
(MR1100-B) Capacity Voltage 10400mA 3.6V Recharging Time Approx. 4 hours Data Cable
(MR1000-D) Size Channel Length 73.8 X 84.3 X 21.4mm 9 90cm Receiver Bag Size 100 X 190 X 55mm
(MR1000-G) System Carrying Case(MR1000-A) Size 480 X 350 X 100mm Battery Charger Power Source Adaptor
(MR1000-C) Adaptor
(MR1000-T) Power Code
(MR1000-P) Size Weight Weight 135 X 75 X 105mm 285g 231g Power Rating AC100~240V/ 50~60Hz Length Measuring Tape Length
(MR1000-M) 1.5m 1.5m 2-9 Overview & Intended Usage Chapter 2 2.4.3 MiroCam Software n Component Component Model Name Version MiroView Workstation MiroView 2.5 MiroView RTV MiroView RTV 1.0 Wireless MiroView RTV Wireless MiroView RTV 1.0 2- 10 3 Performing Capsule Endoscopy Chapter 3 Performing Capsule Endoscopy 3- 2 Performing Capsule Endoscopy Chapter 3 3. PERFORMING CAPSULE ENDOSCOPY 3.1 General Comments n n This chapter describes how to operate and control MiroCam capsule endoscope system. Installation and initial operation must be done by an Authorized IntroMedic Service Technician. 3.2 Safety Warnings WARNING Before turning on the system, check the environment CAUTION CAUTION For the system has been brought condition for MiroCam capsule endoscope system. In case from cold environment (stock room or airfreight) into a warm room, you should for temperature balance and removing of condensation humidity (Risk of leakage current). it after some hours first switch on the electrical safety, make sure that peripheral equipments (external PC, Monitor, printer etc) should be connected to wall socket of independent power supply with a perfect ground. Particularly make sure whether or not those peripheral equipments should be authorized for medical usage. CAUTION DO NOT touch the dome of the capsule. CAUTION CAUTION if SIG indicator of the Receiver turns yellow after ingesting the capsule, immediately contact the physician. If SIG indictor of the Receiver does not turn green when the patient holds the capsule by the gold bands, immediately stop the procedure. If images of the small bowel are not captured during the procedure, ensure the capsule has been excreted by the patient and perform procedure again. NOTE 3-3 Chapter 3 Performing Capsule Endoscopy 3.3 Examination Preparation The following steps (3.3.1 to 3.3.4) explain the process to initialize the receiver and register patient data for the capsule endoscopy procedure. For more information see section 4.2. 3.3.1 Start System n n Click n Turn on power of MiroCam workstation. icon to start the MiroView 2.0 Operator. Insert the ID and password and click icon. WARNING Before starting the system, check the power rating for the workstation. Initialize the Receiver 3.3.2 Turn on the power of the Receiver unit. n Connect the smaller connector of USB Cable to Receiver unit. n Connect normal connector of USB cable to workstation. n n Click n Click n After completing the required fields, click the save the patient information onto the Receiver. icon in MiroView 2.0 Operator. icon in MiroView 2.0 Operator. button to 3- 4 Performing Capsule Endoscopy Chapter 3 Note Before execute MiroView 2.0 Operator, connect USB WARNING cable to MiroCam Receiver unit. If MiroCam receiver unit is initialized, any patient data stored on the device will be deleted permanently. 3.3.3 Receiver Preparation n Connect battery to the Receiver unit.
-
Insert pins on back side of the Receiver unit in the slots on the battery.
- Press battery to receiver until clasp locks battery in place. Turn on power switch of receiver unit. n n Check battery indicator (BAT) of upper side of receiver unit. If battery is fully charged, light will be green. If light is not green, charge the battery. n Check initialization indicator (INI) of upper side of receiver unit. Prior to beginning a procedure, this light must be green. n Connect data cable to receiver unit. Indicator Description SIG INI BAT Indicates status of signal from capsule Green : Signal is being received from capsule Yellow : Signal is not being received from capsule Initialization status of Receiver Unit Green : Receiver is initialized Yellow : Receiver is not initialized Battery Status Green : Fully charged Yellow : Not charged WARNING DO NOT do begin procedure when receiver units battery indicator is not green. This indicates battery is not fully charge or out of order, and patient data may be lost. 3-5 Chapter 3 Performing Capsule Endoscopy 3.4 Patient Preparation Provide the patient with the following instructions to prepare for the capsule endoscope procedure. 3.4.1 Patient Preparation n One day prior to the capsule endoscope procedure
-
Lunch: Patient can have a normal mean around noon, followed by a liquid diet as instructed by the physician.
- The patient should fast for at least 12 hours prior to the procedure. Patient may only drink water (no food or other beverages such as milk or coffee).
- The patient should not take any medications two hours before ingesting the capsule endoscope.
- Male patients may need to shave area, as advised by physician or nurse.
- Patient should stop taking iron supplements 1 week before the capsule endoscope procedure, and should not ingest any medicine at least 2 hours prior to the procedure.
- Diabetic patients need to follow any changes to insulin dosage as prescribed by the physician.
- Physician is recommended to prescribe patient a laxative such as PEG or Sodium Phosphate. Patient should ingest laxative 12 hours prior to procedure.
- Physician is recommended to prescribe patient an anti-foaming agent (such as simethicone), to reduce bubbles in the GI tract. This should be ingested after the laxative.
- The patient should abstain from smoking 12 hours before the procedure.
- On the day of the capsule endoscope test, the patient should wear comfortable and loose cloths. One-piece clothing should not be worn.
- Prior to the procedure, do not apply any lotions or perfumes. n On the day of capsule endoscope procedure
- After arriving in a hospital at the appointed time, the patient should submit an examination permission form and check in. n After ingesting capsule endoscope
- The patient should refrain from taking any food or drink two hours after ingesting the MiroCam capsule endoscope. Water is permitted after two hours. 3- 6 Performing Capsule Endoscopy Chapter 3
- The patient can start with light food after 4 hours. When the examination is complete, the patient can start having normal meals. In case of any abdominal pain, nausea or vomiting after ingesting the MiroCam capsule, the patient should contact physician or nurse immediately.
-
3.5 Sensor Placement & Capsule Ingestion Following is the detailed procedure required to correctly administer the capsule endoscope procedure. 3-7 Chapter 3 Performing Capsule Endoscopy 3.5.1 Connecting Sensor Pads After attaching sensor pads to the data cables, attach the sensor pad correctly to each area as shown below. R 1 5 2 3 4 8 6 7 Xiphoid process Line Pelvic Line 3- 8 Performing Capsule Endoscopy Chapter 3 3.5.2 1 5 4 8 Xiphoid process Line Pelvic Line n Attach sensor pads according to the number on the data cable.
-
Locate the center between the umbilical and the xiphoid process. From this center point, attach the sensor to the distal point on the right flank of the patients body.
- Place the sensor pad on the xiphoid process line, directly below the right clavicle.
-
- Place the sensor pad on the xiphoid process line, directly below the left clavicle. Locate the center between the umbilical and the xiphoid process. From this center point, attach the sensor to the distal point on the left flank of the patients body.
- Place the sensor pad on the right pelvic line, 2 cm behind
- Place the sensor on the right inguinal line, 2 cm to the outside of
- Place the sensor on the right inguinal line, 2 cm to the outside of sensor #1. sensor #2. sensor #3. 3-9 Chapter 3 Performing Capsule Endoscopy
- Place the sensor pad on the right pelvic line, 2 cm behind sensor #4.
- Place the sensor directly below the right clavicle. CAUTION The numbering of the data cables and sensor pad must match the specified area. Please follow the numbering sequence appropriately. CAUTION To prevent entanglement of the data cables during procedure, organize cables and place appropriately in cable bag. WARNING Correct image signal may not be obtained if a sensor pad is not in direct, immediate contact with the skin. Ensure the sensor pads are in direct contact with the skin. Shaving may be required for male patients. The pictures below show the sensor pads attached in the correct locations. 3- 10 Performing Capsule Endoscopy Chapter 3 3.5.3 Removing Capsule from Packaging While holding the MiroCam upside down, peel off the sterilized packaging cover. 3.5.4 Removing the Case Turn the package over, the MiroCam case will slip out of the packaging. While holding the MiroCam case, open the lid. 3.5.5 Removing the Capsule Grip capsule with right hand, and remove from cast. Do not touch the dome of the capsule. 3-11 Chapter 3 Performing Capsule Endoscopy n Make sure that the capsule packaging is not damaged in any way. When removed capsule from case, the capsule should be flashing. CAUTION Do not use a capsule endoscope when the packaging or the capsule is damaged. n Check if the front of the capsule is flashing when it is removed from the case n After washing hands, patients should grip the gold section of the capsule between the thumb and the forefinger of the both hands. NOTE After turning the receiver power on, if a signal is not received by the receiver within 50 seconds (i.e. the SIG indicator does not turn green), the SIG indicator will flash orange and the receiver will make a beep beep warning sound. This will occur until a signal is received or the power is turned off. In this case, you must begin the procedure immediately or turn the receiver off until ready to begin the procedure. 3- 12 Performing Capsule Endoscopy Chapter 3 3.5.6 Checking Operation of the Receiver and Signal Reception After the patient holds the capsule as detailed above, the signal indicator on the receiver is green and signal indication melody is generated.. If signal indicator is yellow, there is no signal being received from the capsule. n Switch on the receiver. n Check if the battery indicator (BAT) at the top of the receiver shows that the battery is fully charged. The battery indicator on the receiver should be green. n Check if the initialize indicator (INI) at the top of the receiver shows that initialization is complete. The INI indicator on the receiver should be green. n Connect the data cable to the MiroCam receiver. n After patient grasps capsule by opposing gold bands (as detailed in picture above, check the SIG indicator on the receiver is green and signal indication melody is generate. CAUTION If the battery indicator on the receiver shows that the battery is not fully charged, do not start the endoscopy procedure. The battery may be depleted during the procedure and images may not be saved properly. CAUTION Please be noticed to confirm the shape and the direction of the connector when you connect to MiroCam Receiver, since it may cause cable damage problem. the data cable 3-13 Chapter 3 Performing Capsule Endoscopy 3.5.7 Ingesting the Capsule After checking if the signal indicator on the receiver is green, ingest the capsule with water. CAUTION Do NOT bite the capsule while ingesting. 3- 14 Performing Capsule Endoscopy Chapter 3 3.5.8 Checking and Wearing the Receiver n After ingesting the capsule, make sure that the SIG indicator is green. n Place the receiver in the bag. n Organize the contents of the bag so that the cables are not entangled. n Adjust cables, and place excess in cable bag. n Adjust the bag strap as needed. n Capsule endoscopy procedure will be complete after 11 hours. CAUTION Patient should not touch receiver during the procedure. This CAUTION During the procedure the SIG indicator should be green. may lead to loss of image data. In this case, check data cable to ensure it is securely inserted and check the sensor pads to ensure these are fixed to the body. 3-15 Chapter 3 Performing Capsule Endoscopy 3.6 Uploading Images Once the capsule endoscopy procedure is complete, upload image data from the receiver to the workstation and start diagnosis of the data using MiroView 2.0 Software. 3.6.1 Transferring and Checking image data n Switch on the workstation. n Click the Operator. icon on the desktop to launch MiroView 2.0 n When the login window appears, enter the user ID and password, and then click the button. n Connect the smaller end of the USB cable to the receiver. n Connect the other end of the USB cable to the workstation. n Click the n When the image data in the MiroCam receiver is successfully button in MiroView 2.0 Operator. uploaded, execute MiroView 2.0 client to review and report. WARNING For more information on using MiroView, please refer to Chapter 4, Using MiroView 2.0. 3- 16 Performing Capsule Endoscopy Chapter 3 3.7 Receiver Management Post Procedure 3.7.1 Cleaning n Remove data cable from receiver unit. n Capsule and sensor pads should not be reused. n To clean the MiroCam Receiver unit and Workstation, dampen a soft cloth with warm water or with a commercial, nonabrasive cleaner (or a mild soap or detergent solution) and wipe the exterior surface lightly. Do not allow any liquids to come in contact with the power connector, cable connector or switches. Do not allow any liquids to penetrate connectors or openings in the system cover. Cleaning of the system and accessories should be conducted at least once a week. CAUTION Do not immerse the MiroCam capsule endoscope system or its accessories in liquid or clean with caustic or abrasive cleaners. Do not spray or pour any liquid on the system or its accessories. Do not use the organic solvents such as a lacquer, thinner, ethylene and oxide because they can damage the equipment. When cleaning the data cables, do not use the cleaning tools such as brush and sandpaper. 3.7.2 Charging Battery n Battery must be recharged after every procedure. n Battery can be individually recharged or with receiver unit. n Battery must be recharged in battery charger. n in charger, If automatically consecutively charge (not simultaneous). two batteries are inserted the batteries will 3-17 Chapter 3 Performing Capsule Endoscopy n Connect adapter DC cable to battery charger. n Connect AC power code to adaptor. n Connect AC power code to AC consent or multi-tap. n Battery charger indicator lights:
n Battery is charging when green light is blinking n Battery is fully charged when light is solid blue n Battery is not-charging if light is yellow n Note: second battery in charger will automatically begin charging once first battery is fully charged. 3- 18 Performing Capsule Endoscopy Chapter 3 Waiting Charging Charging Charged WARNING If the indicator on the battery charger is blinking yellow there is a bad connection. Please disconnect the battery from the battery charger and try again. WARNING Only use batteries provided by IntroMedic. Usage of inappropriate batteries can cause serious damage to the receiver. WARNING Recharge the MiroCam receiver batteries with the adaptor and the battery charger only from IntroMedic. Improper adaptors and battery chargers cause serious damage. 3-19 Chapter 3 Performing Capsule Endoscopy WARNING Try not to contact the metallic terminal board of batteries with conductive objects or human body. It can cause an electric shock or damage and/or breakage of the batteries. WARNING Do not use the batteries with visible surface damage on. Usage of damaged batteries can result in leakage of contents of the batteries. fire. WARNING To avoid the risk of explosion, keep batteries away from WARNING Treat the batteries carefully from serious impact or shock. WARNING Do not expose the batteries to liquids. This can cause This can cause serious damage to the batteries. serious damage. 3- 20 4 Using MiroView Chapter 4 Using MiroView 4- 2 Using MiroView Chapter 4 4. USING MIROVIEW MiroView is a computer software program that aids in diagnosis of disease of the small bowel via displaying images obtained from the MiroCam capsule. MiroView version 2.5 is an update to the MiroView version 1.0. The updated functionalities and user interface further enhance the efficiency and effectiveness of the MiroCam capsule endoscopy procedure. Following is a list of new & updated features:
- User Interface v MiroView enables adjustable resolution, to appropriately fit to the v MiroView has been designed with a dark background to reduce eye v MiroView provides intuitive and unified icons so that the user can screen size. strain. adapt quickly.
- Reporting Tool
- Express View
- Range View
- Network Support
- Account Management
- Multi-Review / Report Generation Support
- Multi-Receiver Management Support The workstation where the MiroView is installed properly has icons for four applications on the Desktop folder: MiroView Client, MiroView Operator, MiroView Server, and MiroView Viewer. However this manual will not explain how to use the MiroView Viewer. For additional information for MiroView Viewer, please refer to the MiroView Viewer Instruction Manual. To get the installation procedure for MiroView, refer to the MiroCam Service Manual. 4-3 Chapter 4 Using MiroView MiroView Client MiroView Operator MiroView Server MiroView Viewer from program generating This is used for reviewing the images the capsule MiroCam and a report. The patient MiroView Server should be started prior to starting the MiroView Client. program is This for managing used the receivers
(initialize, upload etc.) and managing user The accounts. MiroView Server started should be prior starting MiroView Operator. to The server stores the patient data files. Start the MiroView Server to starting MiroView 2.0 or MiroView Operator. prior is used The MiroView Viewer for reviewing the images exported the MiroView Software;
EXMIF JPEG images. from or Functions:
- DB, Mass storage
- Network connectivity Functions:
- Study Information
- Image Information
- Show Findings /
Diagnosis Functions:
- Input Patient Information
- Receiver Initialization
- Upload Data from Receiver
- Data Backup
- Account Management
- Network connectivity Functions:
- List Mode
- Review Mode
- Report Mode
- Export Mode
- Image Atlas
- SGIB
(Suspected GI Bleeding)
- Positioning Window
- Network connectivity
- Multi-Review
- Express View
- Range View
- Map View
- GI Dictionary
- Finding / Diagnosis Information 4- 4 Using MiroView Chapter 4 4.1 MiroView H/W Specifications 4.1.1 MiroView Server n Minimum Hardware Requirements
-
OS: Windows Vista Business
-
CPU: > Core 2 Duo E6300
- Memory: > DDR II 4GByte
-
-
-
- Monitor Resolution: 1024x768 Video Card: Windows Compatible HDD: SATA 500GB ODD: DVD+-RW n Recommended Hardware Requirements OS: Windows Vista Business
-
-
CPU: > Core 2 Duo E6300
- Memory: > DDR II 4GByte
-
-
-
- Monitor Resolution: 1680x1050 Video Card: Windows Compatible HDD: SATAII RAID 1TB ODD: DVD+-RW 4.1.2 MiroView Client n Minimum Hardware Requirements OS: Windows XP, Vista or Windows7 CPU: Pentium4 2 Ghz
-
-
- Memory: 1GB
-
-
-
- Monitor Resolution: 1024x768
-
Sound Card / Mike / Speaker Video Card: Geforce 7600GT 128MB HDD: SATA 160GB ODD: DVD+-RW n Recommended Hardware Requirements
-
OS: Windows XP, Vista or Windows7 4-5 Chapter 4 Using MiroView
-
CPU: > Core 2 Duo E6300
- Memory: > DDR II 2GByte
-
-
-
- Monitor Resolution: 1680x1050
-
Video Card: > Geforce 7600GT 128MB HDD: SATA II 500 GB ODD: DVD+-RW Sound Card / Mike / Speaker 4.1.3 MiroView Operator n Minimum Hardware Requirements OS: Windows XP, Vista or Windows7 CPU: Pentium4 2 Ghz
-
-
- Memory: 1GB
-
-
-
- Monitor Resolution: 1024x768 Video Card: Windows Compatible HDD: SATA 160GB ODD: DVD+-RW n Recommended Hardware Requirements OS: Windows XP, Vista or Windows7 CPU: > Core 2 Duo E6300
-
-
- Memory: 1GB
-
-
-
- Monitor Resolution: 1024x768 Video Card: Windows Compatible HDD: SATA 160GB ODD: DVD+-RW 4.1.4 MiroView Viewer n Hardware Requirements OS: Windows XP, Vista or Windows7 CPU: Pentium4 2Ghz
-
-
- Memory: 1GB
-
-
-
- Monitor Resolution: 1024x768 Video Card: Windows Compatible HDD: SATA 160GB ODD: DVD+-RW 4- 6 Using MiroView Chapter 4 CAUTION MiroView cannot be viewed on a monitor with a resolution under 1024x768. NOTE The GUI is best viewed on a wide monitor of 19 or larger. 4-7 Chapter 4 Using MiroView 4.2 Overview MiroView Server The MiroView Server software is intended to be used as a system administrative managing tool. The MiroView Server manages file systems and database systems. the workstation, MiroView Server will start When powering up automatically. If the following window appears, please set up the configuration for the first run. n Set the port number for the Network. If unknown, please leave this field as 2000. This field should be managed by your Network Administrator in your hospital. n Set the password for Administrator account. Additional accounts can be made via Account Manager function, after logging on the MiroView Operator as Administrator. n Confirm the password. n Type the Hospital Name. 4- 8 Using MiroView Chapter 4 n n Type the Address that should appear on the Capsule Endoscope Report. Type the Telephone Number that should appear on the Capsule Endoscope Report. n After setting up the Configuration Windows, MiroView Server window above will appear. Click the icon to hide the window. 4-9 Chapter 4 Using MiroView 4.3 Overview - MiroView Operator MiroView Operator software is intended to be used for administrative purposes, including account and receiver management. 4.3.1 Start MiroView Operator n Double click the icon to run the MiroView Operator. When executed for the first time, the following window will appear. Input 127.0.0.1 for the IP Address and 2000 for the Port. n n Direct the Local File Path to the caching folder and choose the button to access preferred Language settings and press MiroView Operator. n After MiroView Operator has launched, the Log-in window will appear. 4- 10 Using MiroView Chapter 4 n Input Administrator for ID and password. Then press the button. CAUTION ID and password are encrypted. Do not forget the ID and password. CAUTION Before MiroView Operator has launched, you must create the password for the administrator account. When running the MiroView Server for first time, the user must create the password for the administrator account. NOTE You can create new and edit existing accounts by using the Account Management tool on MiroView Operator. 4-11 Chapter 4 Using MiroView The following subsections describe the main interfaces and functions of the MiroView Operator. 4.3.2 List Mode Opens List Mode 1 List Mode 2 Receiver Mode Opens Receiver Mode 3 Print Report 4 Recycle Bin 5 Export 6 Search 7 Show all Print out the selected Report Open Recycle Bin Window Export the selected Report Search Cases and/or Reviews by keyword Refresh and Show all the Cases and Review on the List 4- 12 Using MiroView Chapter 4 4.3.3 Receiver Mode Screen enables administration and status display for upto four receivers. Upload images in the receiver to the server Input Patients Information to the selected receiver Delete all data prepare the receiver for the next patient. Shows the status of the connected receiver for each of the four ports Shows selected receiver the receiver and/or the patients information from the for 1 Upload 2 Input Information Initialization 3 Patients 4 Receiver Port 5 Detail View 4-13 Chapter 4 Using MiroView 4.3.4 Receiver Initialization Click the button to initialize the selected receiver. 4- 14 Using MiroView Chapter 4 Inputting Patient Information 4.3.5 button to enter the patient information. After Click the completing the required fields, click the button to save the patient information onto the Receiver. The fields in the patient information are editable and additional fields can be added. This can be accessed via Tools Preferences Receiver Set Add /
Modify.
* indicates required fields. 4-15 Chapter 4 Using MiroView 4.3.6 Data Upload After the capsule endoscopy procedure, upload the data from the Receiver to the MiroCam system via MiroView Operator. Click the button to upload the data from the Receiver. in the image data After the MiroCam Receiver has been successfully uploaded, enter the List Mode and check if the new Case has been created.
- Double click the
- Select the newly uploaded Case and enter the Review Mode.
- Play the case to check the uploaded image data. icon to run MiroView Client. CAUTION If you fail to upload the data from the Receiver, disconnect and reconnect the USB cable from the workstation, then turnoff the MiroCam Receiver and turn it back on. Try uploading again. If the problem persists, please contact IntroMedic. NOTE If the user uploads a patient data which has been uploaded before, an error message Data has been previously uploaded. Would you like to upload again? will appear. NOTE Upload progress will be displayed. The user can stop the process while uploading by clicking the button. 4- 16 Using MiroView Chapter 4 4.3.7 User Account Management MiroView Operator enables management of multiple users &
Groups. Users can be organized into Groups to readily control access. The following screen can be accessed by Tools Account Manager. 4-17 Chapter 4 Using MiroView Add an account. Click displayed below. button to show the window as ID: User ID Password: Password Name: User Name Type: User type.
- Super User: can access all the reviews in the System
- User: can access all the reviews in the Group of the Users
- Operator: can log in via MiroView Operator S/W only. Administrator account can log in via both MiroView Client and MiroView Operator. Group: Set of users in the same Clinic or Laboratory. 4.4 Overview - MiroView Client 4.4.1 Start MiroView Client n Double Click the button to run MiroView Client. n Enter the User ID and password and click the button. 4- 18 Using MiroView Chapter 4 4.4.2 List Mode n List 11 12 13 Open List Mode Open Receiver Mode Open Report Mode Open Export Mode Print selected Report Add a Review for the Selected Case Delete selected Case 1 List Mode 2 Review Mode 3 Report Mode 4 Export Mode 5 Print Report 6 Add a Review 7 Delete Case 8 Delete Review Delete selected Review 9 Recycle Bin 10 Search 11 Show all Open Recycle Bin Search Cases and/or Reviews by key word Refresh and Show all the Cases/Reviews on the List A rectangular area to show selected item Show Detail Information for selected Case or Review 12 Selection Bar 13 Detail View 4-19 Chapter 4 Using MiroView 4.4.3 Review Mode Select a case to review on the List Mode and click the enter the Review Mode. n Capture Box Mode icon to 28 25 26 27 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 List Mode Review Mode Report Mode Export Mode Positioning Window Single View Open List Mode Open Review Mode Open Report Mode Open Export Mode Current position of the capsule in 2D Change to Single View Dual View Tri View Range View Normal View Change to Dual View Change to Quad View Change to Range View Shows all Capsule images taken by the 1 2 3 4 5 6 7 8 9 10 4- 20 Using MiroView Chapter 4 11 Express View 12 SGIB 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 Previous Image Previous Capture Play Backward Pause / Resume Play Forward Next Capture Next Image Capture Landmark Image Enhancement Time Display Zoom View Comparison Window Thumbnail Extension Capture Box images reducing Actively removes images based upon similarity; effectively the number of images for review. This mode is recommended to review the patient file. Shows that have high possibility of bleeding and / or blood. This mode may assist in finding areas of suspected bleeding; however, it is NOT 100% reliable, and should only be used as an adjunctive tool to assist in reviewing the images. Go to Previous Image Go to Previous Capture Play Backward Pause / Resume Play Forward Go to the Next Capture Go to the Next Image Image capture button. Set landmarks for transition points in the GI tract. Alter images characteristics (ie. color, saturation etc.) to assist in identifying abnormalities. Displays capsule from procedure start. Fit image to full screen Displays the image selected from the Image Atlas. Extends the Capture Box to display all captured images. Toggle the Capture Box to Thumbnail and Finding Summary Box time duration of the the 4-21 Chapter 4 Using MiroView n Finding Summary Box 28 Finding Summary Box 28 Toggle the Capture Box to Thumbnail and Finding Summary Box Initially, a captured image will be classified as a generic capture. A user may classify a finding through the reporting tool Auto-Finding Checker, which will then appear in the appropriate category in the Finding Summary Box (if it is one of the major categories). 4- 22 Using MiroView Chapter 4 n Mode Control
- Zoom View Click the icon to change to Zoom View mode NOTE Play Slide Control is available on the Zoom View mode to set playing direction and speed. NOTE Move the cursor away from the bottom of the screen and the Play Control will appear. 4-23 Chapter 4 Using MiroView
- Viewing Modes Select one of the Mode. icons for preferred Viewing Single View Dual View Quad View Range View NOTE Range View has been designed to assist in readily finding transition points in the GI tract. The display will show 9 images before and 9 images after the current position. The 9 images are defaulted to 10 minutes apart, resulting in an overview of 90 minutes prior to, and 90 minutes after the current position. The number of images displayed is set at 9, and is not configurable. The 90 minute duration is configurable and can be adjusted by clicking and dragging the mouse icon on the time boundary. 4- 24 Using MiroView Chapter 4 Map View This view has been designed to assist in findings by showing all the images on the same page like a wall paper. You can move the cursor on the designated the original image in detail. to see image Map View The icons will change to playing Mode, as detailed below. Normal Mode Express Mode SGIB Mode 4-25 Chapter 4 Using MiroView n Activated Windows
- Positioning Window Displays the location of where the capsule took the current image.
- Comparison Window Displays the image selected from the Image Atlas.
- Image Enhancement Window Alter images characteristics (ie. color, saturation etc.) to assist in identifying abnormalities. 4- 26 Using MiroView Chapter 4 4.4.4 Report Mode Select the desired review for the Case from the List Mode and click the icon to move to the Report Mode. n Report Mode 1 List Mode 2 Review Mode 3 Report Mode 4 Export Mode 5 Scale 6 Drawing Tools 7 Image Enhancement 11 12 13 Open List Mode Open Review Mode Open Report Mode Open Export Mode Opens graduated scale for display on the image Enables users to include annotations for the image. Alter saturation etc.) abnormalities. images characteristics in
(ie. color, identifying to assist 4-27 Chapter 4 Using MiroView 8 GI Dictionary Provides the frequently-used-words to search. 9 Text Comments Write comments for the Capture on the main for Referral, findings, Summary &
display and the report.
(Image Comments, Reason Procedure Recommendation) info &
10 Audio Comment Record voice commenst 11 Input Word 12 Add Word 13 Delete Word
(The user can record min 3 sec ~ max 300 sec for each Capture) Input the selected word into the text editor. Add the typed word into the Dictionary. Delete selected word from the Dictionary. NOTE Additional tabs in the report can be created or modified via Tools -> Preferences -> Report 4- 28 Using MiroView Chapter 4 4.4.5 Export Mode 11 12 13 15 14 1 List Mode 2 Review Mode 3 Report Mode 4 Export Mode 5 Check Captured Image 6 Uncheck Captured Image 7 Print 8 Preview 9 Save as PDF 10 Export Review Open List Mode Open Review Mode Open Report Mode Open Export Mode Select Captured image and add to the list to print/export. Deselect Captured image and remove from the candidate list to print/export. Print out the report for the Review. Preview the report Save the report as a PDF file. Export Review data only without Video. Review data is the own format of MiroView, which 4-29 Chapter 4 Using MiroView 11 Send to PACS 12 Export EXMIF 13 Export 14 Export Segment 15 Export Entire contains Captured Images and Comments by the user. Send selected captured images to PACS system via DICOM Sender Export Captured Images as EXMIF
(eXchangeable Medical Information Format) Export Captured Images as Image, Video, or MiroVideo. MiroVideo is the own video format for MiroView. Export segment as Video, MiroVideo Export entire file as Video, MiroVideo 4- 30 Using MiroView Chapter 4 4.5 Overview - MiroView Viewer 4.5.1 Start MiroView Viewer n Double Click the button to run MiroView Viewer. Open a Image File (EXMIF-Exchangeable Medical Image Format data or JPEG) via File -> Open function. Image Display 1 2 Title 3 Description 4 Study Information 5 Image Information 6 Findings 7 Diagnosis Information), Patient Image Display Area (EXMIF or JPEG) Title for the image in the Image Display Description for the image in the Image Display General Information (Software / Manufacturer /
Hospital Information
(Doctor, Patient Body Shape) and Study Information (Doctor, Receiver, Capsule and Procedure Information) for the image in the Image Display Creator Information and/or Finding Comments for the image in the Image Display Findings information the creator captured and commented before for the image in the Image Display. Diagnosis information the creator captured and commented on for the image in the Image Display. 4-31 Chapter 4 Using MiroView 4.6 Performing Capsule Endoscopy 4.6.1 Initializing Receiver n Connect the Receiver to the workstation via USB port. n Run the MiroView Operator by double clicking the icon on the desktop or from the Windows Start-up menu. Log in. n Enter the Receiver mode and select the Receiver port to initialize. 4- 32 Using MiroView Chapter 4 n Click the icon. The following message will appear. n Click Yes to start the initializing process. n After the initializing process is complete, the following message will appear. 4-33 Chapter 4 Using MiroView n Click OK and the Patient Information window will appear. Input the patient information for each field. n Click Check Name button to input the patient name and then click the Check Doctor Name button to input the doctor name.
(The fields marked with * is required.) n Click to save the Patient Information to the Receiver. The Receiver is now ready for the procedure. NOTE The fields in the patient information are editable, and additional fields can be added. This can be accessed via Tools Preferences Receiver Set Add / Modify. 4- 34 Using MiroView Chapter 4 4.6.2 Upload Data (MiroView Operator) n Connect the Receiver to the workstation via USB port. n Run the MiroView 2.0 Operator by double clicking the icon on the desktop or from the Windows Start-up menu. Log in. n After switching the Receiver on, enter the Receiver mode and select the proper Receiver port to upload the patient file. 4-35 Chapter 4 Using MiroView n Click the button to start the Upload. n After completing the Upload, the following message will appear. 4.6.3 Review (MiroView Client) n Run MiroView Client by double clicking the icon on the desktop or from the Windows Start-up menu. Log in.
*If you want to run the MiroView Client as Local Mode, click Local button on the Log-in screen. 4- 36 Using MiroView Chapter 4 n Select the Patient Video Data on the List Mode
- Select a patient video data to review from the List Mode and click the
- Check to see if another user is reviewing the patient file on the Current User column in the List Mode. icon to enter the Review Mode.
- If the patient file is being reviewed by another user, a message will appear. 4-37 Chapter 4 Using MiroView n Review the Patient Video on the Review Mode NOTE For efficient video playback the video is cached locally in the server. During this buffering process, the video may be reviewed. Click the Range View icon to readily set landmarks. 4- 38 Using MiroView Chapter 4
- Use the Range View and Normal View to find all the landmarks. NOTE Click the landmark icon and select landmark type:
Esophagus, Gastric, Duodenum, IleoCecal, and Cecal.
- After marking the necessary landmarks, play the video via the Express mode and capture abnormalities and / or suspected disease pathologies in the small bowel. 4-39 Chapter 4 Using MiroView NOTE The user can press Enter key to capture the image.
- To identify abnormalities and/or disease pathology more easily, use the Image Enhancement function to adjust preferred image settings. Click the icon to open the Image Enhancement window. 4- 40 Using MiroView Chapter 4 Composing the Report on the Report Mode
- After capturing images of suspected disease in the Review Mode pathology, click the icon to enter the Report Mode.
- Use the auto complete dictionary slide and choose available disease pathology from the list.
- Completing the report:
1. Complete image comments for captured images (as required) 2. Complete the following tabs: Reason for referral; Procedure Info & Findings; and Summary & Recommendations 3. After completing the report, enter the Export Mode to print out the complete report and/or export MiroView 2.0 data. 4-41 Chapter 4 Using MiroView NOTE The GI Dictionary tab and findings tab allow the user to adjust the findings and the level of severity. The user can use the GUI easily. CAUTION Do not detach and/or attach the Microphone from/to the workstation while recording the audio comment. CAUTION Sound Card Driver must be installed prior to using the Recording function. CAUTION Prior to recording the comments, close all the the Sound Card such as applications using music/video player. 4- 42 Using MiroView Chapter 4 n Print report, create file and export the Review on the Export Mode.
- Select the images to select and click button to add the selected captured images to the candidate list to export. (You can remove the selected images from the candidate list using the button) 4-43 Chapter 4 Using MiroView
- Click the button to use export function. A window will appear like below.
- Select the File Type to export and change the desired file location on the Selected Folder and click the A window will appear with a progress bar. button. 4- 44 Using MiroView Chapter 4
- - The user can use the button for EXMIF, the for Segment Export (Video / Miro Video) and the for Export Case function. button button NOTE Select the segments on the Timebar to export. If button to export no segment is selected, the segments would be disabled. NOTE MiroVideo includes Reviews and selected Segments so the contents and edit from Open Case / Review file, Import Case / Review file main menu. the user can confirm that 4-45 Chapter 4 Using MiroView n Print Report and save the report as PDF on the Report Preview icon to print reports and/or click the
- Click the save the report as PDF.
- Show report preview on the screen. icon to
- The user can print out the report and save the report to PDF file in this screen.
- If the user selects the 6 images or 9 images per page mode and tries to print out, the message If you proceed with the current Layout mode, image comments may be omitted and not printed. Please select another Layout Mode to include all comments in the report. will appear. 4- 46 Using MiroView Chapter 4 4.6.4 Image Atlas n Atlas images are searchable by Capsule Endoscopy Reference Structured Terminology (CEST), Endoscopy 2005; 37: 951 - 959
(February 16, 2002 C Given Imaging Ltd.) 11 12 13 14 15 16 1 Atlas 2 My Atlas 3 Atlas + My Atlas Show Atlas Images and My Atlas Images at Default Atlas Images Atlas Images configured by the user 4 Import Atlas Images from file Atlas Images to file(s) 5 Export once. Import Atlas Images from file; BMP, JPEG, GIF, EMF, WMF, TIFF and PNG file format. Export Atlas Images to file; BMP, JPEG, GIF, EMF, WMF, TIFF and PNG file format. 6 Edit Information Edit the Information for the selected Atlas 7 Export Atlas Image(s) to E-
Image. Export Atlas selected Atlas Image(s) and sends it / them to another doctor via E-mail. 4-47 Chapter 4 Using MiroView mail 8 Delete Atlas Image(s) Image(s) 9 Print Atlas 10 Save as PDF 11 Show/Hide Marks 12 Search 13 Comparison 14 Window Browser Tab(Findings /
Diagnosis) 15 Recent Images Delete the selected Atlas Image(s) Print the selected Atlas image(s) Save the selected Atlas Image(s) as PDF Show / Hide Marks on the Atlas Image(s) Search the Atlas Image Show the Comparison Window Area where to select the Category of the Atlas Image to see. Images that the user saw on the Comparison Window 17 18 19 4- 48 Using MiroView Chapter 4 17 Image Information 18 Findings /
Diagnosis 19 Atlas Thumbnail Area(2) Show the Information for the Atlas Image Show the Findings / Diagnosis Information Search the Atlas Image 20 20 Comparison Window Show the Current Image and Comparison Image at once. 4-49 Chapter 4 Using MiroView n Import Atlas Images from file(s) MiroView allows the user to import image files for personal purpose.
- Click the to import. icon to open the Open Browser to look for the files 4- 50 Using MiroView Chapter 4
- Select the file to import and click Open
- Complete the Title, Description and Image Information fields and click OK. 4-51 Chapter 4 Using MiroView 4.7 How to import MiroView 1.0 files 4.7.1 Preparing files for import n Run the MiroView 1.0 n Export files by clicking Export for MiroView Express function on MiroView 1.0 Before using the Export to MiroView Express function, you must insert DVD media to the DVD writer or connect the external hard disk or USB flash memory to the workstation via USB port. 1. Select a patient file from the list to export to MiroView Express. 2. Execute Export to MiroView Express on the main menu 4- 52 Using MiroView Chapter 4 3. If the form, Save As appears like below, select the location to export and fill in the File name field and click Save. 4. The window appears like below and if the function ends successfully you will see the message, Exporting has been completed. 4-53 Chapter 4 Using MiroView 4.7.2 Import MiroView 1.0 files n Run the MiroView 2.0 Operator or MiroView 2.0 Client. n Execute the Import Case / Review File -> Import Case in the List menu 4- 54 Using MiroView Chapter 4 n Select MiroView 1.0 files from the file extension drop-down box and look for the file to upgrade n Look for the File with extension .MiroCam and click Open 4-55 Chapter 4 Using MiroView n Task Manager will appear and will proceed with importation of the file. If the Doctor name on the file is not registered on the MiroView 2.0 Server, Doctor in charge registration process will start. 4- 56 Using MiroView Chapter 4 n If the Reviewer name on the file is not registrered on the MiroView 2.0 Server, Reviewer registration process will start. n Complete necessary fields and click OK. 4-57 Chapter 4 Using MiroView n Progress bar will appear. n Upgrading MiroView 1.0 file to MiroView 2.0 file will be completed. 4- 58 Using MiroView RTV 5 Chapter 5 Using MiroView RTV 5- 2 Using Wireless MiroView RTV Chapter 5 5. USING MIROVIEW RTV MiroView RTV, the application software for the MiroCam System Real-
time Viewer, consist of a data transmission module that directly receives the image data from the MiroCam Receiver and an output module for image display. 5.1 Overview MiroView RTV software can be installed on any mobile platform. Click the corresponding icon on the desktop to launch the MiroView RTV. 5.1.1 Basic Screen Layout Name / ID / Gender / Age Start / Pause / Resume Main Display Time Display Display patient name, ID, gender and age Change MiroView RTV operation status between start, pause and resume. Patient Information Operation Main Display Display captured image. Time Display Button Indicates procedure time. the MiroCam capsule endoscope 5-3 Chapter 5 Using MiroView RTV 5.1.2 Basic Mobile PC Layout USB Port DC-Jack Power Switch Mobile PC Power on switch. USB Port for connect with MiroCam Receiver Connector for Mobile PC Charger WARNING Only recharge the mobile PC batteries when MiroCam Receiver does not connect to the mobile PC. WARNING For safe use of mobile PC, please refer to mobile PC User Manual. 5- 4 Using Wireless MiroView RTV Chapter 5 5.1.3 Recommended Hardware Specification n Operating System : Windows XP Professional n CPU : A110 800Mhz n Memory : DDR II 1GByte(667MHz) n Display Adaptor : Windows Compatible n Hard Disk : Over 1GB n Monitor Resolution : 1024x600 n Monitor Contrast : Windows Compatible n In compliance with IEC60950-1 or IEC60601-1 5-5 Chapter 5 Using MiroView RTV 5.2 Using MiroView RTV 5.2.1 Before Using MiroView RTV The MiroView RTV software is pre-install on mobile PC n WARNING Do not use MiroView RTV software and mobile PC before upgrading the MiroCam Receiver. WARNING Do not upgrade MiroCam Receiver during the MiroCam capsule endoscopy procedure, it may result in loss if image data. n Connect the data cable for the Real-time Viewer to the MiroCam Receiver. NOTE The data cable for MiroView RTV has no cover for the USB port on the MiroCam Receiver. n Fallow the MiroCam capsule endoscopy procedure. NOTE For more information on the MiroCam capsule endoscope procedure, please refer to Chapter 3, Performing Capsule Endoscopy. 5- 6 Using Wireless MiroView RTV Chapter 5 5.2.2 Starting MiroView RTV n Switch on the mobile PC. n Click the n Connect the small side of the USB cable to the MiroCam icon on the desktop to launch MiroView RTV. Receiver. n Connect the large side of the USB cable to the Mobile PC. n Click the n Click the n Click the n Click the button to start image display button to stop image display button to restart image display button on the top-right to close the MiroView RTV. NOTE Start, Pause, and Resume use the same button. The function (Start, Pause or Resume) changes based upon the status of MiroView RTV software. CAUTION MiroView RTV is not software for to be used for diagnostic purposes. Use the MiroView software for viewing the patient images for diagnostic purposes. CAUTION Do not disconnect the MiroCam Receiver while the RTV is in use. If the MiroCam Receiver is disconnected during use, the software may need to be restarted. NOTE If an error message appears while using MiroView RTV, reconnect the USB cable between MiroCam Receiver and Mobile PC 5-7 Chapter 5 Using MiroView RTV 5.2.3 After Using MiroView RTV n Disconnect USB cable from MiroCam Receiver and mobile PC. n Continue MiroCam capsule endoscope procedure. 5- 8 6 Using Wireless MiroView RTV Chapter 6 Using Wireless MiroView RTV 5- 2 Using Wireless MiroView RTV Chapter 5 6. USING WIRELESS MIROVIEW RTV Wireless MiroView RTV, the application software for the MiroCam System Real-time Viewer, consist of a data transmission module that directly receives the image data from the MiroCam Receiver and an output module for image display. Wireless MiroView RTV receive image data via Wi-Fi channel. 6.1 Overview Wireless MiroView RTV software can be installed on I-OS 3 or I-OS 4 running on I-Pad / I-Phone / I-Pod Touch. Click the corresponding icon on the desktop to launch the Wireless MiroView RTV. 6.1.1 Basic Screen Layout I-Phone / I-Pod Touch ID / Name / Gender / Age Receiver Scan Main Display Time Display 5-3 Chapter 6 Using Wireless MiroView RTV Display patient name, ID, gender and age Patient Receiver Information Scan Button Show the available Receiver Access Points so that the user can choose to see the patient image. Main Display Display captured image. Time Display Indicates procedure time. the MiroCam capsule endoscope 6.1.2 Basic Screen Layout I-Pad ID / Name / Gender / Age Receiver Scan Main Display Time Display NOTE If turning the I-Pad upright as Portrait View, Receiver List will disappear for the Main Display to be enlarged. 5- 4 Using Wireless MiroView RTV Chapter 5 Patient Receiver Information Scan Button Display patient name, ID, gender and age Show the available Receiver Access Points so that the user can choose to see the patient image. Main Display Display captured image. Time Display Indicates procedure time. the MiroCam capsule endoscope 6.1.3 Recommended Hardware Platform n n n I-Phone 3G, 3GS, 4G I-Pod Touch 3G I-Pad 5-5 Chapter 6 Using Wireless MiroView RTV 6.2 Using Wireless MiroView RTV 6.2.1 Network configuration for Wireless MiroView RTV The Wireless MiroView RTV software is pre-installed on I-OS 3 /
I-OS 4 running on I-Phone, I-Pod Touch or I-Pad. n NOTE When the Wireless MiroView RTV run first, network configuration for I-Phone and I-Pad is required. n n Fallow the MiroCam capsule endoscopy procedure. Turn on the Wi-Fi switch on the MiroCam Receiver System during the Capsule Procedure. NOTE For more information on the MiroCam capsule endoscope procedure, please refer to Chapter 3, Performing Capsule Endoscopy. 5- 6 Using Wireless MiroView RTV Chapter 5 n Click the icon on the desktop to launch MiroView RTV. n Click the button on the right-top to show the available MiroCam Receiver Access Points 5-7 Chapter 6 Using Wireless MiroView RTV n If no available Receiver Access Points are shown on the list, the image will appear like below. Please make sure that the Wifi-
Swtich on the Receiver is turned on. n If the MiroCam Receiver Access Points are shown on the list. Click the desirable AP (Access Point) to show the patients image. 5- 8 Using Wireless MiroView RTV Chapter 5 n At the first time to connect to the AP, IP setting is required as below steps:
l Step 1. Press the Home Button on the I-Phone/I-Pad/I-Pod touch to close the Wireless MiroView RTV Software. l Step 2. Run the Settings for the I-OS. 5-9 Chapter 6 Using Wireless MiroView RTV l Step3. Set the IP for the desirable AP. (Setup as Static; IP Address : 192.168.3.200, Subnet Mask: 255.255.255.0) 5- 10 Using Wireless MiroView RTV Chapter 5 n After completing the IP setting, close the setting window by clicking the home button on the I-Phone/I-Pad/I-Pod touch and go back to the desktop. 5-11 Chapter 6 Using Wireless MiroView RTV 6.2.2 Running MiroView RTV n n Fallow the MiroCam capsule endoscopy procedure. Turn on the Wi-Fi switch on the MiroCam Receiver System during the Capsule Procedure. NOTE For more information on the MiroCam capsule endoscope procedure, please refer to Chapter 3, Performing Capsule Endoscopy. 5- 12 Using Wireless MiroView RTV Chapter 5 n Click the RTV. icon on the desktop to launch Wireless MiroView The Splash window will appear for a while and will disappear. n 5-13 Chapter 6 Using Wireless MiroView RTV n After the splash window disappears, the real time play will be started automatically. (The number of receiver is one). If there are two or more Receiver AP is available, the Receiver AP List window will be shown. Choose one to continue as below:
n After reviewing the real time play is completed, click the home button to close. 5- 14 Using Wireless MiroView RTV Chapter 5 n To terminate the Software completely, double click the home button. The menu will pop up like below. n Press the Wireless MiroView RTV icon for a while. The will appear. mark 5-15 Chapter 6 Using Wireless MiroView RTV n Click the icon, then the Wireless MiroView RTV will be terminated completely. CAUTION Wireless MiroView RTV is not to be used for diagnostic purposes. Use the MiroView software for viewing the patient images for diagnostic purposes. 6.2.3 After Using Wireless MiroView RTV Turn off the Wi-Fi switch on the MiroCam receiver system. n n Continue MiroCam capsule endoscope procedure. 5- 16 A Appendix Case Form Report A- 2 Case Form Report Trial ID :
Version No. : 1.0 : 30/07/2007 CASE REPORT FORM Subject No.:
Subject Initials :
Center No.:
:
:
:
:
:
:
Principal Investigator Center Name Address Telephone No. Fax No. E-mail CONFIDENTIAL A-3 Trial ID :
Case Form Report Version No. : 1.0 :
30/07/2007 Screening Subject No.:
Subject Initials :
Center No.:
Date of Informed consent signed D M M Y Y Y Y Inclusion Criteria Yes No N/A 1. Informed consent form voluntarily signed 2. Male / Female subject of age 19 years~70 years. If there is one or more mark of No, you shall be contradicted for this clinical test. CONFIDENTIAL A- 4 Case Form Report Trial ID :
Version No. : 1.0 : 30/07/2007 Screening Subject No.:
Subject Initials :
Center No.:
Date of visit :
D M M Y Y Y Y Exclusion Criteria Yes No N/A 1. Do you have a gastrointestinal disorder
(suspected malignantlesions or gastrointestinal obstructive disorders, perforation, stenosis or fistula)?
2. Do you have a cardiovascular disorder (dysrhythemia, trouble with cardiac conduction system, hypertension(high blood pressure) or ischemic heart disease)?
3. Have you been diagnosed of cardiac insufficiency, angina pectoris or cardiac arrest (heart attack) in recent 6 months?
bearing potential?
pacemaker or defibrillator?
4. Are you with (or implanted) medical devices such as cardiac 8. Are you an infant or epileptic? Are you easily affected by 5. Do you have a swallowing disorder?
6. Do you have delayed gastric emptying or serious gastric disorder?
7. Are you a pregnant or nursing woman, or a woman of child 9. Do you have several small bowel diverticulums ?
10. Did you fail to obtain a voluntary agreement?
11. Are you judged unqualified for this test by doctors or nurses?
If there is one or more mark of Yes, you shall be contradicted for this clinical test. CONFIDENTIAL electronic waves?
A-5 Case Form Report A- 6 EC Representative SeDa S.p.A 20090 Trezzano sul Naviglio(MI), Via Toistoi, 7 Italy Key Contact: Roberta Cimolato Office: +39-02-484241 Mobile: +39-02-48424290 Email: roberta.cimolato@sedaitaly.it Disposal of Old Electrical &
Electronic Equipment Applicable in the European Union and other European countries with separate collection systems IntroMedic Co.,Ltd. Suite 1104, E&C Venture Dream Tower 6-Cha 197-28 Guro-Dong, Guro-Gu, Seoul, KOREA 152-719 Tel : +82-2-801-9300 Fax : +82-2-801-9330 http://www.intromedic.com e-mail : help@intromedic.com 0297
frequency | equipment class | purpose | ||
---|---|---|---|---|
1 | 2011-06-02 | JBP - Part 15 Class B Computing Device Peripheral | Original Equipment |
app s | Applicant Information | |||||
---|---|---|---|---|---|---|
1 | Effective |
2011-06-02
|
||||
1 | Applicant's complete, legal business name |
Intromedic Co., Ltd.
|
||||
1 | FCC Registration Number (FRN) |
0016453136
|
||||
1 | Physical Address |
Suite 1104, E&C Venture Dream Tower 6-Cha, 197-28
|
||||
1 |
Seoul, N/A
|
|||||
1 |
South Korea
|
|||||
app s | TCB Information | |||||
1 | TCB Application Email Address |
c******@ccsemc.com
|
||||
1 | TCB Scope |
A1: Low Power Transmitters below 1 GHz (except Spread Spectrum), Unintentional Radiators, EAS (Part 11) & Consumer ISM devices
|
||||
app s | FCC ID | |||||
1 | Grantee Code |
VAX
|
||||
1 | Equipment Product Code |
INTROMEDIC3
|
||||
app s | Person at the applicant's address to receive grant or for contact | |||||
1 | Name |
J****** L******
|
||||
1 | Title |
Manager
|
||||
1 | Telephone Number |
82-2-********
|
||||
1 | Fax Number |
82-2-********
|
||||
1 |
j******@intromedic.com
|
|||||
app s | Technical Contact | |||||
n/a | ||||||
app s | Non Technical Contact | |||||
n/a | ||||||
app s | Confidentiality (long or short term) | |||||
1 | Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | No | ||||
1 | Long-Term Confidentiality Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | No | ||||
if no date is supplied, the release date will be set to 45 calendar days past the date of grant. | ||||||
app s | Cognitive Radio & Software Defined Radio, Class, etc | |||||
1 | Is this application for software defined/cognitive radio authorization? | No | ||||
1 | Equipment Class | JBP - Part 15 Class B Computing Device Peripheral | ||||
1 | Description of product as it is marketed: (NOTE: This text will appear below the equipment class on the grant) | MiroCam Capsule Endoscope System | ||||
1 | Related OET KnowledgeDataBase Inquiry: Is there a KDB inquiry associated with this application? | No | ||||
1 | Modular Equipment Type | Does not apply | ||||
1 | Purpose / Application is for | Original Equipment | ||||
1 | Composite Equipment: Is the equipment in this application a composite device subject to an additional equipment authorization? | No | ||||
1 | Related Equipment: Is the equipment in this application part of a system that operates with, or is marketed with, another device that requires an equipment authorization? | No | ||||
1 | Is there an equipment authorization waiver associated with this application? | No | ||||
1 | If there is an equipment authorization waiver associated with this application, has the associated waiver been approved and all information uploaded? | No | ||||
app s | Test Firm Name and Contact Information | |||||
1 | Firm Name |
EMC Compliance Ltd.
|
||||
1 | Name |
H******** Y******
|
||||
1 | Telephone Number |
82-31********
|
||||
1 | Fax Number |
82-31********
|
||||
1 |
y******@emc2000.co.kr
|
|||||
Equipment Specifications | |||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|