all | frequencies |
|
exhibits | applications |
---|---|---|---|---|
manual |
app s | submitted / available | |||||||
---|---|---|---|---|---|---|---|---|
1 2 |
|
User Manual | Users Manual | 2.14 MiB | ||||
1 2 | Cover Letter(s) | |||||||
1 2 | Operational Description | |||||||
1 2 | Attestation Statements | |||||||
1 2 | Cover Letter(s) | |||||||
1 2 | Cover Letter(s) | |||||||
1 2 | External Photos | |||||||
1 2 | ID Label/Location Info | |||||||
1 2 | Internal Photos | |||||||
1 2 | Cover Letter(s) | |||||||
1 2 | RF Exposure Info | |||||||
1 2 | Test Report | |||||||
1 2 | Test Report | |||||||
1 2 | Test Setup Photos | |||||||
1 2 | Test Setup Photos | |||||||
1 2 | Test Setup Photos |
1 2 | User Manual | Users Manual | 2.14 MiB |
Radical-7 Operator's Manual The Radical-7 operating instructions provide the necessary information for proper operation of all models of the Radical-7 Pulse CO-Oximeter system. There may be information provided in this manual that is not relevant for your system. General knowledge of pulse oximetry and an understanding of the features and functions of the Radical-7 are prerequisites for its proper use. Do not operate the Radical-7 without completely reading and understanding the instructions in this manual. NOTICE:
Purchase or possession of this instrument does not carry any express or implied license to use this instrument with replacement parts which would, alone or in combination with this instrument, fall within the scope of one of the patents relating to this instrument. Caution Federal law restricts this device to sale by or on the order of a physician. Wireless Radio FCC ID: VKF-RAD7CA, IC: 7362A-RAD7CA Masimo Corporation 40 Parker Irvine, CA 92618, USA Tel.: 949-297-7000 Fax.: 949-297-7001 www.masimo.com EU authorized representative for Masimo Corporation:
MDSS GmbH Schiffgraben 41 D-30175 Hannover, Germany Medical electrical equipment with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1/CAN/CSA C22.2 No. 601.1 Patents: www.masimo.com/patents.htm 2012 Masimo Corporation
, Adaptive Probe Off Detection, APOD, Discrete Saturation Transform, DST, FastSat, FST, Masimo, Pulse Oximeter, PVI, rainbow, SatShare, SET, Signal Extraction Technology, Signal IQ, SpCO, SpHb, SpMet are federally registered trademarks of Masimo Corporation. Adaptive Threshold Alarm, In Vivo Adjustment, Pleth Variability Index, Radical-7, Rainbow Acoustic Monitoring, rainbow Resposable, RDS, RRa, RRp,SafetyNet, SpOC are trademarks of Masimo Corporation. All other trademarks and registered trademarks are property of their respective owners. The use of the trademarks PATIENT SAFETYNET and PSN is under license from University HealthSystem Consortium. www.masimo.com 1 Masimo Contents About this Manual ----------------------------------------------------------------------------------------------- 7 Product Description, Indications for Use, Contraindications, and Features ------------------------9 Product Description -----------------------------------------------------------------------------------------9 Indications for Use ---------------------------------------------------------------------------------------- 10 Contraindications ------------------------------------------------------------------------------------------ 10 Safety Information, Warnings, and Cautions -------------------------------------------------------------11 Parameter Related Safety Information, Warnings, and Cautions --------------------------------11 Device Related Safety Information, Warnings, and Cautions ------------------------------------ 15 Electrical Safety Information, Warnings, and Cautions ------------------------------------------- 18 Alarm Related Safety Information, Warnings, and Cautions------------------------------------- 20 Sensor Related Safety Information, Warnings, and Cautions------------------------------------ 21 Chapter 1: Technology Overview --------------------------------------------------------------------------- 23 Signal Extraction Technology (SET) ------------------------------------------------------------------- 23 rainbow Pulse CO-Oximetry Technology -------------------------------------------------------------- 26 rainbow Acoustic Monitoring (RAM) Technology--------------------------------------------------- 30 Chapter 2: Radical-7 Descriptions ------------------------------------------------------------------------- 33 General System Description ----------------------------------------------------------------------------- 33 Functionality of the Radical-7 --------------------------------------------------------------------------- 34 Handheld ---------------------------------------------------------------------------------------------------- 35 Standalone -------------------------------------------------------------------------------------------------- 37 Monitor Interface ------------------------------------------------------------------------------------------ 40 Chapter 3: Setup ----------------------------------------------------------------------------------------------- 41 Unpacking and Inspection ------------------------------------------------------------------------------- 41 Docking Station Power Requirements----------------------------------------------------------------- 41 Setting Up the Docking Station ------------------------------------------------------------------------ 42 Initial Battery Charging ---------------------------------------------------------------------------------- 42 Setting Up for Philips, Agilent, or HP VueLink ------------------------------------------------------ 42 Setting Up for SpaceLabs Flexport--------------------------------------------------------------------- 43 Setting Up and Using SatShare ------------------------------------------------------------------------ 44 Chapter 4: Operation------------------------------------------------------------------------------------------ 45 Using the Touchscreen and Buttons ------------------------------------------------------------------ 45 www.masimo.com 3 Masimo Radical-7 Contents Using Screenlock ------------------------------------------------------------------------------------------- 46 Using the Home Button ---------------------------------------------------------------------------------- 46 Navigating the Radical-7 --------------------------------------------------------------------------------- 47 About the Display View ----------------------------------------------------------------------------------- 48 Accessing the Main Menu -------------------------------------------------------------------------------- 55 Navigating the Main Menu ------------------------------------------------------------------------------ 56 Parameter Settings ---------------------------------------------------------------------------------------- 58 Sounds -------------------------------------------------------------------------------------------------------- 73 Device Settings --------------------------------------------------------------------------------------------- 74 Trends -------------------------------------------------------------------------------------------------------- 81 About --------------------------------------------------------------------------------------------------------- 87 Chapter 5: Configurations ----------------------------------------------------------------------------------- 89 About Profiles and Default Settings ------------------------------------------------------------------- 89 Changing Profiles Settings ----------------------------------------------------------------------------- 90 About Patient Types --------------------------------------------------------------------------------------- 92 Powering Off the Radical-7------------------------------------------------------------------------------- 93 Chapter 6: Alarms and Messages -------------------------------------------------------------------------- 95 About Alarms ----------------------------------------------------------------------------------------------- 95 Silencing the Alarms -------------------------------------------------------------------------------------- 95 Adaptive Threshold Alarm (ATA) Feature ------------------------------------------------------------ 97 3D Alarms---------------------------------------------------------------------------------------------------- 98 Messages --------------------------------------------------------------------------------------------------- 100 Chapter 7: Troubleshooting---------------------------------------------------------------------------------107 Troubleshooting Measurements -----------------------------------------------------------------------107 Troubleshooting the Radical-7 ------------------------------------------------------------------------ 109 Chapter 8: Specifications ----------------------------------------------------------------------------------- 111 Performance ------------------------------------------------------------------------------------------------ 111 Accuracy ----------------------------------------------------------------------------------------------------- 111 Resolution -------------------------------------------------------------------------------------------------- 113 Electrical ---------------------------------------------------------------------------------------------------- 113 Environmental --------------------------------------------------------------------------------------------- 114 Physical Characteristics --------------------------------------------------------------------------------- 114 Trending----------------------------------------------------------------------------------------------------- 114 Alarms ------------------------------------------------------------------------------------------------------- 115 www.masimo.com 4 Masimo Radical-7 Contents Display Indicators----------------------------------------------------------------------------------------- 116 Compliance ------------------------------------------------------------------------------------------------- 117 Output Interface ------------------------------------------------------------------------------------------ 118 Wireless Radio (If Installed)---------------------------------------------------------------------------- 118 Serial Interface Specifications ------------------------------------------------------------------------- 118 Serial Interface Setup ------------------------------------------------------------------------------------ 119 Analog Output and Nurse Call Specifications ----------------------------------------------------- 120 Symbols ----------------------------------------------------------------------------------------------------- 122 ISO Country Codes (FCC and EU) ----------------------------------------------------------------------123 Citations ---------------------------------------------------------------------------------------------------- 125 Chapter 9: Service and Maintenance -------------------------------------------------------------------- 127 Cleaning----------------------------------------------------------------------------------------------------- 127 Battery Operation and Maintenance ----------------------------------------------------------------- 127 Performance Verification------------------------------------------------------------------------------- 130 Repair Policy -----------------------------------------------------------------------------------------------132 Return Procedure -----------------------------------------------------------------------------------------132 Contacting Masimo-------------------------------------------------------------------------------------- 133 Appendix: Best Practices for Comparisons to Reference Measurements ------------------------ 137 Best Practices Checklist for Continuous SpHb Comparisons ----------------------------------- 137 Best Practices Checklist for SpCO Comparisons -------------------------------------------------- 140 Best Practices Checklist for Acoustic Respiration Rate Comparisons------------------------- 142 Index ------------------------------------------------------------------------------------------------------------- 145 www.masimo.com 5 Masimo About this Manual This manual explains how to set up and use the Radical-7 Pulse CO-Oximeter. Important safety information relating to general use of the Radical-7 appears in this manual. Read and follow any warnings, cautions, and notes presented throughout this manual. The following are explanations of warnings, cautions, and notes. A warning is given when actions may result in a serious outcome (for example, injury, serious adverse affect, death) to the patient or user. The following is an example of a warning:
Warning This is a sample of a warning statement. A caution is given when any special care is to be exercised by the patient or user to avoid injury to the patient, damage to this instrument or damage to other property. The following is an example of a caution:
Caution This is a sample of caution statement. A note is given when additional general information is applicable. The following is an example of a note:
Note: This is a sample of a note. www.masimo.com 7 Masimo Product Description, Indications for Use, Contraindications, and Features The following chapter contains the Radical-7 product description, key features and benefits, indications for use, contraindications, and safety information, including cautions, warnings, and notes. Product Description The Radical-7 is a noninvasive monitor that measures arterial oxygen saturation (SpO2), pulse rate, and perfusion index (PI), along with optional measurements of hemoglobin
(SpHb), carboxyhemoglobin (SpCO), methemoglobin (SpMet), Pleth Variability Index (PVI), Acoustic Respiration Rate (RRa), and Pleth Respiration Rate (RRp) . The Radical-7 can be used as either a Handheld or a Standalone monitor. The Radical-7 features a touchscreen Liquid Crystal Display (LCD) that continuously displays numeric values for all parameters. The Radical-7 provides graphical displays for plethysmographic waveform, respiratory waveform, Signal Identification and Quality Indicator (Signal IQ). The Radical-7 can also be used to interface with a multi-parameter patient monitor to send Masimo SET pulse oximetry information to that monitor for display. The Radical-7 has an embedded 802.11 wireless radio that can be used for connectivity. Key Features The following features are available for the Radical-7. Some features are optional:
Masimo SET is clinically proven to satisfy all sensitivity and specificity requirements for pulse oximeter technology. Rainbow technology uses 7+ wavelengths of light to continuously and noninvasively measure carboxyhemoglobin (SpCO), methemoglobin (SpMet), and total hemoglobin (SpHb), as well as providing a more reliable probe-off detection. Total oxygen content (SpOC) provides a calculated measurement of the amount of oxygen in arterial blood, which may provide useful information about oxygen both dissolved in plasma and combined with hemoglobin. Perfusion Index (PI) with trending capability indicates arterial pulse signal strength and may be used as a diagnostic tool during low perfusion. Pleth Variability Index (PVI) may show changes that reflect physiologic factors such as vascular tone, circulating blood volume, and intrathoracic pressure excursions. Respiration rate can be determined by the acoustic (RRa) or plethysmographic waveform (RRp) www.masimo.com 9 Masimo Radical-7 Product Description, Indications for Use, Contraindications, and Features Signal IQ waveform for signal identification and quality indication during excessive motion and low signal to noise situations. FastSat tracks rapid changes in arterial O2. Variable pitch provides tonal variance for every 1% change in saturation. SatShare interface allows transfer of SpO2 and pulse rate to an existing multi-parameter monitor and allows for the reading of SpCO, SpMet, SpHb, and SpOC on the monitor. Automatic screen rotation provides upright display for vertical or horizontal monitor positioning. Multi-gesture touchscreen interface. Detachable portable Handheld for patient transport. Desat Index Alarm may help clinicians to detect an increasing quantity of smaller desaturations that may precede declining respiratory status. Perfusion Index (PI) Delta Alarm alerts clinicians to possible changes in perfusion, often a reliable indicator of illness severity. Advanced Threshold Alarms to dynamically adjust alarm thresholds that are tailored specifically to the patient. Remote alarm interface. Indications for Use The Masimo Radical-7 and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical-7 and accessories have been validated and are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical-7 and accessories are indicated to provide the continuous noninvasive monitoring data obtained from the Masimo rainbow SET Radical 7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate to multi-parameter devices for the display of those devices. Contraindications The Radical-7 is not intended for use as an apnea monitor. www.masimo.com 10 Masimo Safety Information, Warnings, and Cautions The following section lists warnings, caution, notes, and safety information. The Radical-7 is designed to minimize the possibility of hazards from errors in the software program by following sound Engineering Design Processes, Risk Analysis and Software Validation. The Radical-7 is to be operated by qualified personnel only. The manual, accessories, directions for use, all precautionary information, and specifications should be read before use. Always use the Radical-7 precisely in accordance with the directions in this manual, including finger selection, finger alignment in the sensor, and subject behavior during testing. Failure to follow all of the directions in this manual could lead to inaccurate measurements. Caution For SpHb, the Radical-7 should be considered an early warning device. Blood samples should be analyzed by laboratory instruments prior to clinical decision making to completely understand the patients condition. Caution Variation in hemoglobin measurements may be profound and may be affected by sample type, body positioning, as well as other physiological conditions. As with most hemoglobin tests, Radical-7 test results should be scrutinized in light of a specific patients condition. Any results exhibiting inconsistency with the patients clinical status should be repeated and/or supplemented with additional test data. Parameter Related Safety Information, Warnings, and Cautions This section contains parameter related safety information. Caution If patient hypoxemia is indicated, blood samples should be analyzed by laboratory devices to completely understand the patients condition. Caution Confirm offset values(s) periodically as the difference between the displayed parameter value and the laboratory reference value may vary over time. www.masimo.com 11 Masimo Radical-7 Safety Information, Warnings, and Cautions Caution Do not use In Vivo Adjustment if the monitor displays a Low SpHb SIQ message. Caution If the Low Perfusion message is frequently displayed, find a better perfused monitoring site. In the interim, assess the patient and, if indicated, verify oxygenation status through other means. Caution Changing the SpHb Cal, the date and time of the system clock, or the trend period clears the data in the trend memory. Note: Interfering Substances: Dyes, or any substance containing dyes, that change usual blood pigmentation may cause erroneous readings. Note: Inaccurate SpCO and SpMet readings can be caused by:
Levels of methemoglobin approximately 1.5% or above Intravascular dyes such as indocyanine green or methylene blue Abnormal hemoglobin levels Low arterial perfusion Low arterial oxygen saturation levels including altitude induced hypoxemia Elevated total bilirubin levels Motion artifact SpCO readings may not be provided if SpO2 readings are less than 90%
SpCO readings may not be provided if SpMet readings are greater than 2%
Note: SpO2, SpCO, SpMet, and SpHb are empirically calibrated in healthy adult volunteers with normal levels of carboxyhemoglobin (COHb) and methemoglobin (MetHb). Note: The Radical-7 cannot measure elevated levels of COHb or MetHb. www.masimo.com 12 Masimo Radical-7 Safety Information, Warnings, and Cautions Note: Inaccurate SpO2 readings may be caused by:
Elevated levels of COHb and MetHb For increased COHb: COHb levels above normal tend to increase the level of SpO2. The level of increase is approximately equal to the amount of COHb that is present. Note: High levels of COHb may occur with a seemingly normal SpO2. When elevated levels of COHb are suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed. For increased MetHb: the SpO2 may be decreased by levels of MetHb of up to approximately 10% to 15%. At higher levels of MetHb, the SpO2 may tend to read in the low to mid 80s. When elevated levels of MetHb are suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed. Note: Inaccurate SpO2 readings may be caused by:
Intravascular dyes such as indocyanine green or methylene blue Externally applied coloring and texture such as nail polish, acrylic nails, glitter, etc. Elevated levels of bilirubin Severe anemia Low arterial perfusion Motion artifact www.masimo.com 13 Masimo Radical-7 Safety Information, Warnings, and Cautions Note: Inaccurate SpHb and SpOC readings may be caused by:
Intravascular dyes such as indocyanine green or methylene blue Externally applied coloring and texture such as nail polish, acrylic nails, glitter, etc. Elevated levels of bilirubin Low arterial perfusion Motion artifact Low arterial oxygen saturation levels Elevated carboxyhemoglobin levels Elevated methemoglobin levels Difference between patient's finger skin and finger core temperature Hemoglobin synthesis disorders Hemoglobinopathies and synthesis disorders such as thalassemias, Hb s, Hb c, sickle cell, etc. Vasospastic disease such as Raynaud's Elevated altitude Peripheral vascular disease Liver disease EMI radiation interference www.masimo.com 14 Masimo Radical-7 Safety Information, Warnings, and Cautions Note: Inaccurate SpHb readings may be caused by:
Intravascular dyes such as indocyanine green or methylene blue Externally applied coloring and texture such as nail polish, acrylic nails, glitter, etc. Elevated levels of bilirubin Low arterial perfusion Motion artifact Low arterial oxygen saturation levels Elevated carboxyhemoglobin levels Elevated methemoglobin levels Difference between patient's finger skin and finger core temperature Hemoglobin synthesis disorders Hemoglobinopathies and synthesis disorders such as thalassemias, Hb s, Hb c, sickle cell, etc. Vasospastic disease such as Raynaud's Elevated altitude Peripheral vascular disease Liver disease EMI radiation interference Note: Inaccurate SpCO and SpMet readings can be caused by:
Levels of methemoglobin approximately 1.5% or above Intravascular dyes such as indocyanine green or methylene blue Abnormal hemoglobin levels Low arterial perfusion Low arterial oxygen saturation levels including altitude induced hypoxemia Elevated total bilirubin levels Motion artifact SpCO readings may not be provided if SpO2 readings are less than 90%
SpCO readings may not be provided if SpMet readings are greater than 2%
Note: Only SpO2 and pulse rate can be displayed on the multi-parameter monitor with Flexport. Device Related Safety Information, Warnings, and Cautions This section contains device related safety information. www.masimo.com 15 Masimo Radical-7 Safety Information, Warnings, and Cautions Warning Explosion hazard: Do not use the Radical-7 in the presence of flammable anesthetics or other flammable substance in combination with air, oxygen-enriched environments, or nitrous oxide. Warning Do not use the Radical-7 or sensor during magnetic resonance imaging (MRI) scanning. Warning Do not use the Radical-7 during electrocautery. Warning Do not use the Radical-7 or sensor during defibrillation. Warning Do not place the Radical-7 or accessories in any position that might cause it to fall on the patient. Caution Do not place the Radical-7 where the controls can be changed by the patient. Caution Disposal of product - Comply with local laws in the disposal of the instrument and/or its accessories. Caution During SatShare operation, do not use the plethysmographic waveform display on the multi-parameter monitor for diagnostic purposes. Instead, use the plethysmographic waveform displayed on the Radical-7 screen. Caution Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore may not detect certain arrhythmias. The Radical-7 should not be used as a replacement or substitute for ECG-based arrhythmia analysis. Note: Do not place containers with liquids on or near the Radical-7. Liquids spilled on the instrument may cause it to perform inaccurately or fail. Note: EMI radiation interference such as computer displays and/or LCD/plasma TVs can cause error or incorrect measurements on the Radical-7. www.masimo.com 16 Masimo Radical-7 Safety Information, Warnings, and Cautions Note: If the Radical-7 fails any part of the setup procedures or leakage spot check, remove the instrument from operation until qualified service personnel have corrected the situation. Note: A functional tester cannot be used to assess the accuracy of the Radical-7. Note: Do not autoclave, pressure sterilize, or gas sterilize the Radical-7. Note: Do not touch, press, or rub the display panels with abrasive cleaning compounds, instruments, brushes, rough-surface materials, or bring them into contact with anything that could scratch the panel. Note: Do not use petroleum-based or acetone solutions, or other harsh solvents, to clean the Radical-7. These substances affect the devices materials and instrument failure can result. Note: An operator may only perform maintenance procedures specifically described in the manual. Refer servicing to qualified service personnel trained in the repair of this equipment. Note: SatShare signals are ideal simulated waveforms corresponding to the calculated saturation and pulse rate values and do contain all of the information contained in physiological waveforms. The multi-parameter patient monitor decodes these signals into saturation and pulse rate values. Note: Simultaneous use of SatShare and serial port is not supported. Note: If the Radical Docking Station is compatible with SafetyNet, Vuelink is not supported. Use the Radical-7 in accordance with Environmental Specifications section in of this manual. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Cleared Use Only: The device and related accessories are cleared by the Food and Drug Administration (FDA) for noninvasive patient monitoring and may not be used for any processes, procedures, experiments or any other use for which the device is not intended or cleared by the FDA, or in any manner inconsistent with the instructions for use or labeling. Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. In accordance with international telecommunication requirements, the frequency band of 2.4 GHz and 5.15 to 5.25 Ghz is only for indoor usage to reduce potential for harmful interference to co-channel mobile satellite systems. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. www.masimo.com 17 Masimo Radical-7 Safety Information, Warnings, and Cautions However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. This equipment has been tested and found to comply with the limits for medical devices to the EN 60601-1-2: 2002, Medical Device Directive 93/42/EEC and Class B. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. This Class B digital apparatus complies with Canadian ICES-003. Electrical Safety Information, Warnings, and Cautions This section contains electrical related safety information. Warning Fire Hazard: To protect against fire hazard, replace only with fuses of same type, current rating, and voltage rating. Caution Do not place the Radical-7 on electrical equipment that may affect the instrument, preventing it from working properly. Caution Dispose of used batteries according to required country or regional instructions. Caution Risk of explosion if battery is replaced with an incorrect type. Replace with Masimo supplied parts only. Caution At Low Battery, connect the Radical-7 to AC power to prevent loss of power. Caution Do not incinerate battery. www.masimo.com 18 Masimo Radical-7 Safety Information, Warnings, and Cautions Caution Electric shock hazard: Do not open the Radical-7 cover except to replace the battery or batteries. Caution To protect against injury from electric shock, follow the directions below:
Avoid placing the device on surfaces with visible liquid spills. Do not soak or immerse the device in liquids. Use cleaning solutions sparingly. Caution Electrical shock and flammability hazard: Before cleaning the Radical-7, always turn it off and disconnect the power cord from the AC power supply. Caution Do not under any circumstances remove the grounding conductor from the power plug. Caution Do not use extension cords or adapters of any type. The power cord and plug must be intact and undamaged. Caution To ensure patient electrical isolation, connect only to other equipment with electrically isolated circuits. Caution Do not connect to an electrical outlet controlled by a wall switch or dimmer. Note: All external instrument connections to the Analog Output/Nurse Call connector must be IEC-60950 compliant. Note: It is recommended that the Radical-7 Handheld is docked to the Docking Station that is attached to an AC power source when it is not in use to ensure that the battery remains fully charged. Note: External instrument connections to the SatShare port must be IEC-60601-1 compliant. Note: Only use a SatShare cable that has a ferrite bead installed. www.masimo.com 19 Masimo Radical-7 Safety Information, Warnings, and Cautions Note: Use the power cord as the means to disconnect the instrument from the main power supply. Note: If the Radical-7 Handheld has not been used or charged within seven (7) days or more, then recharge the battery prior to use. Note: The instrument must be configured to match your local power line frequency to allow for the cancelation of noise introduced by fluorescent lights and other sources. Note: If there is any doubt about the integrity of the protective earth conductor arrangement, operate the Radical-7 on internal battery power until the AC power supply protective conductor is fully functional. Note: To prevent damage, do not soak or immerse the sensor in any liquid solution. Do not attempt to sterilize by irradiation, steam, autoclave or any method other than ethylene oxide as indicated. Note: Only SpO2 and pulse rate can be displayed on the multi-parameter monitor with SatShare. Note: The battery should be installed and/or removed from the Radical-7 by qualified personnel only. Note: All batteries lose capacity with age, thus the amount of run time at Low Battery will vary depending upon the age of the battery. Note: All external device connections to the RS-232 serial port must be IEC-60950 compliant. Note: The Docking Station battery should be installed and/or removed from the Docking Station only by qualified personnel. To conserve battery power, keep the frequency of the audible alarms to a minimum and the volume to a minimum. To conserve battery power, keep the back-lit LCD screen at minimum illumination. When using the SatShare feature, to conserve battery power, always keep the Radical-7 on AC line power. Alarm Related Safety Information, Warnings, and Cautions This section contains alarm related safety information. Caution For home use, ensure that the Radical-7 alarm can be heard from other rooms in the house, especially when noisy appliances such as vacuum cleaners, dishwashers, clothes dryers, televisions, or radios are operating. www.masimo.com 20 Masimo Radical-7 Safety Information, Warnings, and Cautions Caution Do not place the Radical-7 against a surface that may cause the alarm to be muffled. Caution To ensure that alarm limits are appropriate for the patient being monitored, check the limits each time the Radical-7 is used. Caution The Nurse Call feature is disabled when the Audible Alarms are silenced and Nurse Call setting is set to Alarms. Caution Alarms are disabled in Sleep Mode. Caution When the Radical-7 is placed in All Mute, the patient alarms will not audibly sound on the Radical-7 or the SafetyNet. The SafetyNet View will display a visual alarm. Caution During SatShare operation, the audible alarms may be muted on the Radical-7. When the audible alarm is muted (indicated by a bell with a slash through it) on the Radical-7, use the multi-parameter monitor for audible alarm indication. Caution If an alarm condition occurs while the Alarm Silence period is set to All Mute, the only alarm indications will be visual display and symbols related to alarm condition. No alarm will sound. Note: The Desat Index alarm is intended as an adjunct rather than in place of the Low Saturation alarm. Sensor Related Safety Information, Warnings, and Cautions This section contains sensor related safety information. Warning As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation. www.masimo.com 21 Masimo Radical-7 Safety Information, Warnings, and Cautions Warning Always remove the sensor from the patient and completely disconnect the patient from the Radical-7 before bathing the patient. Caution If using the Radical-7 during full body irradiation, keep the sensor out of the radiation field. If the sensor is exposed to the radiation, the reading might be inaccurate or the instrument might read zero for the duration of the active irradiation period. Note: Do not loop the sensor cable into a tight coil or wrap around the device, as this can damage the sensor cable. Note: Patient Safety - If a sensor is damaged in any way, discontinue use immediately. Note: Failure to apply the sensor properly may lead to incorrect measurements. Note: Additional information specific to the Masimo sensors compatible with Radical-7, including information about parameter/measurement performance during motion and low perfusion, may be found in the sensor's directions for use (DFU). Note: Do not expose the Masimo sensors used with Radical-7 to moisture, liquids or a humid environment, as this may make the sensor perform inaccurately or fail. Note: High-intensity extreme lights (including pulsating strobe lights and direct sunlight) directed on the sensor, may not allow the Radical-7 to obtain readings. Note: When using the Maximum Sensitivity setting, performance of the Sensor Off detection may be compromised. If the Radical-7 is in this setting and the sensor becomes dislodged from the patient, the potential for false readings may occur due to environmental "noise"
such as light, vibration, and excessive air movement. www.masimo.com 22 Masimo Chapter 1: Technology Overview The following chapter contains general descriptions about parameters, measurements, and the technology used by Masimo products. Signal Extraction Technology (SET) Masimo Signal Extraction Technology's signal processing differs from that of conventional pulse oximeters. Conventional pulse oximeters assume that arterial blood is the only blood moving (pulsating) in the measurement site. During patient motion, however, the venous blood also moves, causing conventional pulse oximeters to read low values, because they cannot distinguish between the arterial and venous blood movement (sometimes referred to as noise). Masimo SET pulse oximetry utilizes parallel engines and adaptive digital filtering. Adaptive filters are powerful because they are able to adapt to the varying physiologic signals and/or noise and separate them by looking at the whole signal and breaking it down to its fundamental components. The Masimo SET signal processing algorithm, Discrete Saturation Transform (DST), in parallel with Fast Saturation Transform (FST), reliably identifies the noise, isolates it and, using adaptive filters, cancels it. It then reports the true arterial oxygen saturation for display on the monitor. Masimo rainbow SET Parallel Engines This figure is for conceptual purposes only. www.masimo.com 23 Masimo Radical-7 Masimo SET DST This figure is for conceptual purposes only. Chapter 1: Technology Overview General Description for Oxygen Saturation (SpO2) Pulse CO-Oximetry is a continuous and noninvasive method of measuring the level of arterial oxygen saturation in blood. The measurement is taken by placing a sensor on a patient, usually on the fingertip for adults and the hand or foot for neonates. The sensor is connected to the Pulse CO-Oximetry instrument with a patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data in three ways:
As a percent value for arterial oxygen saturation (SpO2) As a pulse rate (PR) 1. 2. 3. As a plethysmographic waveform Successful Monitoring for SpO2, PR, and PI Stability of the SpO2 readings may be a good indicator of signal validity. Although stability is a relative term, experience will provide a good feeling for changes that are artifactual or physiological and the speed, timing, and behavior of each. The stability of the readings over time is affected by the averaging mode being used. The longer the averaging time, the more stable the readings tend to become. This is due to a dampened response as the signal is averaged over a longer period of time than during shorter averaging times. However, longer averaging times delay the response of the oximeter and reduce the measured variations of SpO2 and pulse rate. www.masimo.com 24 Masimo Radical-7 Chapter 1: Technology Overview Functional Oxygen Saturation The Radical-7 is calibrated to measure and display functional oxygen saturation (SpO2): the amount of oxyhemoglobin expressed as a percentage of the hemoglobin that is available to transport oxygen. Note that carboxyhemoglobin is not capable of transporting oxygen, but is recognized as oxygenated hemoglobin by conventional pulse oximetry. General Description for Pulse Rate (BPM) Pulse rate, measured in beats per minute (BPM) is based on the optical detection of peripheral flow pulse. General Description for Perfusion Index (PI) The Perfusion Index (PI) is the ratio of the pulsatile blood flow to the non-pulsatile or static blood in peripheral tissue. PI thus represents a noninvasive measure of peripheral perfusion that can be continuously and noninvasively obtained from a pulse oximeter. General Description for Pleth Variability Index (PVI) The pleth variability index (PVI) is a measure of the dynamic changes in the perfusion index
(PI) that occur during the respiratory cycle. The calculation is accomplished by measuring changes in PI over a time interval where one or more complete respiratory cycles have occurred. PVI is displayed as a percentage (0-100%). The utility of PVI is unknown at this time and requires further clinical studies. Technical factors that may affect PVI include probe malposition and patient motion. www.masimo.com 25 Masimo Radical-7 rainbow Pulse CO-Oximetry Technology Chapter 1: Technology Overview rainbow Pulse CO-Oximetry technology is governed by the following principles:
1. Oxyhemoglobin (oxygenated blood), deoxyhemoglobin (non-oxygenated blood), carboxyhemoglobin (blood with carbon monoxide content), methemoglobin
(blood with oxidized hemoglobin) and blood plasma constituents differ in their absorption of visible and infrared light (using spectrophotometry). The amount of arterial blood in tissue changes with your pulse
(photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well. 2. www.masimo.com 26 Masimo Radical-7 Chapter 1: Technology Overview The Radical-7 uses a multi-wavelength sensor to distinguish between oxygenated blood, deoxygenated blood, blood with carbon monoxide, oxidized blood and blood plasma. The Radical-7 utilizes a sensor with various light-emitting diodes (LEDs) that pass light through the site to a diode (detector). Signal data is obtained by passing various visible and infrared lights (LEDs, 500 to 1400nm) through a capillary bed (for example, a fingertip, a hand, a foot) and measuring changes in light absorption during the blood pulsatile cycle. This information may be useful to clinicians. The maximum radiant power of the strongest light is rated at 25 mW. The detector receives the light, converts it into an electronic signal and sends it to the Radical-7 for calculation. 1. Light Emitting Diodes (LEDs)
(7 + wavelengths) 2. Detector Once the Radical-7 receives the signal from the sensor, it utilizes proprietary algorithms to calculate the patients functional oxygen saturation (SpO2 [%]), blood levels of carboxyhemoglobin (SpCO [%]), methemoglobin (SpMet [%]), total hemoglobin concentration (SpHb [g/dL]) and pulse rate (BPM). The SpCO, SpMet and SpHb measurements rely on a multi-wavelength calibration equation to quantify the percentage of carbon monoxide and methemoglobin and the concentration of total hemoglobin in arterial blood. In an ambient temperature of 35 C the maximum skin surface temperature has been measured at less than 106 F (41 C), verified by Masimo sensor skin temperature test procedure. Pulse CO-Oximetry vs. Drawn Whole Blood Measurements When SpO2, SpCO, SpMet, and SpHb measurements obtained from the Radical-7
(noninvasive) are compared to drawn whole blood (invasive) measurements by blood gas and/or laboratory CO-Oximetry methods, caution should be taken when evaluating and interpreting the results. The blood gas and/or laboratory CO-Oximetry measurements may differ from the SpO2, SpCO, SpMet, SpHb, and SpOC measurements of the Radical-7. Any comparisons should be simultaneous, meaning the measurement on the device should be noted at the exact time that blood is drawn. In the case of SpO2, different results are usually obtained from the arterial blood gas sample if the calculated measurement is not appropriately corrected for the effects of variables that shift the relationship between the partial pressure of oxygen (PO2) and saturation, such as:
pH,temperature, the partial pressure of carbon dioxide (PCO2), 2,3-DPG, and fetal hemoglobin. In the case of SpCO, different results are also expected if concentration of methemoglobin in the blood gas sample is abnormal (greater than 2% for methemoglobin concentration). High levels of bilirubin may cause erroneous SpO2, SpMet, SpCO, and SpHb readings. As blood samples are usually taken over a period of 20 seconds (the time it takes to draw the blood) a meaningful comparison can only be achieved if the oxygen saturation, carboxyhemoglobin, and methemoglobin concentration of the patient are stable and not changing over the period of time that the blood gas sample is taken. Subsequently, blood gas and laboratory CO-Oximetry measurements of SpO2, SpCO, SpMet, SpHb, and SpOC may www.masimo.com 27 Masimo Radical-7 Chapter 1: Technology Overview vary with the rapid administration of fluids and in procedures such as dialysis. Additionally, drawn whole blood testing can be affected by sample handling methods and time elapsed between blood draw and sample testing. Measurements with Low Signal IQ should not be compared to laboratory measurements. General Description for Total Hemoglobin (SpHb) Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of total hemoglobin (SpHb) in arterial blood. It relies on the same principles of pulse oximetry to make its SpHb measurement. The measurement is taken by a sensor capable of measuring SpHb, usually on the fingertip for adult and pediatric patients. The sensor connects directly to the Pulse CO-Oximeter or with a patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as measurement of total hemoglobin concentration. Successful Monitoring for SpHb A stable SpHb reading is associated with correct sensor placement, small physiological changes during the measurement and acceptable levels of arterial perfusion at the measurement site. Physiological changes at the measurement site are mainly caused by fluctuations in the oxygen saturation, blood concentration and perfusion. See Safety Information, Warnings, and Cautions on page 11 and Troubleshooting Measurements on page 107. General Description for Total Arterial Oxygen Content (CaO2) Oxygen (O2) is carried in the blood in two forms, either dissolved in plasma or combined with hemoglobin. The amount of oxygen in the arterial blood is termed the oxygen content
(CaO2) and is measured in units of ml O2/dL blood. One gram of hemoglobin (Hb) can carry 1.34 ml of oxygen, whereas 100 ml of blood plasma may carry approximately 0.3 ml of oxygen*. The oxygen content is determined mathematically as:
CaO2 = 1.34 (ml O2/g Hb) x Hb (g/dL) x HbO2 + PaO2 (mm Hg) x (0.3 ml O2/100 mm Hg/dL) Where HbO2 is the fractional arterial oxygen saturation and PaO2 is the partial pressure of arterial oxygen. For typical PaO2 values, the second part of the above equation (PaO2 [mm Hg] x [0.3 ml O2/
100 mm Hg/dL]) is approximately 0.3 ml/dL. Furthermore, for typical carboxyhemoglobin and methemoglobin levels, the functional saturation (SpO2) as measured by a pulse oximeter is given by:
SpO2 = 1.02 x HbO2
*Martin, Laurence. All You Really Need to Know to Interpret Arterial Blood Gases, Secon d Edition. New York: Lippincott Williams & Wilkins, 1999. www.masimo.com 28 Masimo Radical-7 Chapter 1: Technology Overview General Description for SpOC The above approximations result in the following reduced equation for oxygen content via the Pulse CO-Oximeter:
SpOC (ml/dL*) = 1.31 (ml O2/g Hb) x SpHb (g/dL) x SpO2 + 0.3 ml/dL
*When ml O2/g Hb is multiplied by g/dL of SpHb, the gram unit in the denominator of ml/g cancels the gram unit in the numerator of g/dL resulting in ml/dL (ml of oxygen in one dL of blood) as the unit of measure for SpOC. See Safety Information, Warnings, and Cautions on page 11. General Description for Carboxyhemoglobin (SpCO) Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of carboxyhemoglobin concentration (SpCO) in arterial blood. It relies on the same basic principles of pulse oximetry (spectrophotometry) to make its SpCO measurement. The measurement is obtained by placing a sensor on a patient, usually on the fingertip for adults and the hand or foot for infants. The sensor connects either directly to the Pulse CO-Oximetry instrument or through an instrument patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as percentage value for the SpCO, which reflect blood levels of carbon monoxide bound to hemoglobin. Successful Monitoring for SpCO A stable SpCO reading is associated with correct sensor placement, small physiological changes during the measurement and acceptable levels of arterial perfusion in the patients fingertip (measurement site). Physiological changes at the measurement site are mainly caused by fluctuations in the oxygen saturation, blood concentration and perfusion. General Description for Methemoglobin (SpMet) Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of methemoglobin concentration (SpMet) in arterial blood. It relies on the same basic principles of pulse oximetry (spectrophotometry) to make its SpMet measurement. The measurement is obtained by placing a sensor on a patient, usually on the fingertip for adults and the hand or foot for infants. The sensor connects either directly to the Pulse CO-Oximetry instrument or through a patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as percentage value for the SpMet. www.masimo.com 29 Masimo Radical-7 Chapter 1: Technology Overview Successful Monitoring for SpMet A stable SpMet reading is associated with correct sensor placement, small physiological changes during the measurement and acceptable levels of arterial perfusion in the patients fingertip (measurement site). Physiological changes at the measurement site are mainly caused by fluctuations in the oxygen saturation, blood concentration and perfusion. See Safety Information, Warnings, and Cautions on page 11. SpCO, SpMet, and SpHb Measurements During Patient Motion The Radical-7 displays measurements of SpCO, SpMet, and SpHb during patient motion. However, because of the changes in the physiological parameters such as blood volume, arterial-venous coupling, etc. that occur during patient motion, the accuracy of such measurements may not be reliable during excessive motion. In this case, the measurement value for SpCO, SpMet, or SpHb displays as dashes (---) and a message (Low SpCO SIQ, Low SpMet SIQ, or Low SpHb SIQ) displays to alert the clinician that the instrument does not have confidence in the value due to poor signal quality caused by excessive motion or other signal interference. rainbow Acoustic Monitoring (RAM) Technology rainbow Acoustic Monitoring (RAM) continuously measures a patients respiration rate based on airflow sounds generated in the upper airway. The Acoustic Sensor translates airflow sounds generated in the upper airway to an electrical signal that can be processed to produce a respiration rate, measured as breaths per minute. Respiratory sounds include sounds related to respiration such as breath sounds (during inspiration and expiration), adventitious sounds, cough sounds, snoring sounds, sneezing sounds, and sounds from the respiratory muscles [1]. These respiratory sounds often have different characteristics depending on the location of recording [2] and they originate in the large airways where air velocity and air turbulence induce vibration in the airway wall. These vibrations are transmitted, for example, through the lung tissue, thoracic wall and trachea to the surface where they may be heard with the aid of a stethoscope, a microphone or more sophisticated devices. rainbow Acoustic Monitoring Architecture The following figure illustrates how a respiratory sound produced by a patient can be turned into a numerical measurement that corresponds to a respiratory parameter. www.masimo.com 30 Masimo Radical-7 Patient Chapter 1: Technology Overview The generation of respiratory sounds is primarily related to turbulent respiratory airflow in upper airways. Sound pressure waves within the airway gas and airway wall motion contribute to the vibrations that reach the body surface and are recorded as respiratory sounds. Although the spectral shape of respiratory sounds varies widely from person to person, it is often reproducible within the same person, likely reflecting the strong influence of individual airway anatomy [2-6]. Sensor The sensor captures respiratory sounds (and other biological sounds) much like a microphone does. When subjected to a mechanical strain, (e.g., surface vibrations generated during breathing), the sensor becomes electrically polarized. The degree of polarization is proportional to the applied strain. The output of the sensor is an electric signal that includes a sound signal that is modulated by inspiratory and expiratory phases of the respiratory cycle. Acquisition System The acquisition system converts the electric signal provided by the sensor into a digital signal. This format allows the signal to be processed by a computing device. www.masimo.com 31 Masimo Radical-7 Signal Processing Chapter 1: Technology Overview The digital signal produced by the acquisition system is converted into a measurement that corresponds to the respiratory parameter of interest. As shown in the previous figure, this can be performed by, for example, determining the digital signal envelope or outline which in turn may be utilized to determine the respiratory rate. In this way, a real-time, continuous breath rate parameter can be obtained and displayed on a monitor which, in many cases, may be real-time and continuous. The respiratory cycle envelope signal processing principle is similar to methods that sample airway gasses and subsequently determine a respiratory rate.
[1] A.R.A. Sovijrvi, F. Dalmasso, J. Vanderschool, L.P. Malmberg, G. Righini, S.A.T. Stoneman. Definition of terms for applications of respiratory sounds. Eur Respir Rev 2000; 10:77, 597-610.
[2] Z. Moussavi. Fundamentals of respiratory sounds analysis. Synthesis lectures on biomedical engineering #8. Morgan & Claypool Publishers, 2006.
[3] Olsen, et al. Mechanisms of lung sound generation. Semin Respir Med 1985; 6: 171-179.
[4] Pastercamp H, Kraman SS, Wodicka GR. Respiratory sounds Advances beyond the stethoscope. Am J Respir Crit Care Med 1977; 156: 974-987.
[5] Gavriely N, Cugell DW. Airflow effects on amplitude and spectral content of normal breath sounds. J Appl Physiol 1996; 80: 5-13.
[6] Gavrieli N, Palti Y, Alroy G. Spectral characteristics of normal breath sounds. J Appl Physiol 1981; 50: 307-314. www.masimo.com 32 Masimo Chapter 2: Radical-7 Descriptions The following chapter contains the Radical-7 descriptions, including descriptions of the Handheld monitor , the Standalone monitor, and the optional SatShare monitor interface. General System Description The Radical-7 system includes the following:
1. 2. 3. Instrument Patient Cable Sensor www.masimo.com 33 Masimo Radical-7 Chapter 2: Radical-7 Descriptions Functionality of the Radical-7 The Radical-7 provides the functionality of three instruments in one:
Handheld Pulse CO-Oximeter The Radical-7 is a fully featured Handheld. The Handheld contains the majority of the device features. All measurements and instrument status datum are displayed on the touchscreen. All user input is performed through the touchscreen and control buttons. The sensor cable connector is located on the Handheld. Standalone Pulse Oximeter The Radical-7 is a fully featured Standalone Pulse-Oximeter, and Acoustic Monitor. The Handheld snaps into the Docking Station to provide a fully featured standalone monitor. The Docking Station connects to AC power for standalone operation or charging of the Handheld. An optional Docking Station battery is available. The Standalone features Nurse Call, analog output, and serial output. Monitor Interface The Radical-7 interfaces to the SpO2 input module of multi-parameter patient monitors to upgrade conventional pulse oximetry technology on the multi-parameter monitor to Masimo SET technology. Utilizing a SatShare cable, the standalone Radical-7 also interfaces with the SpO2 input of a validated multi-parameter patient monitor, instantly upgrading the conventional pulse oximetry to Masimo SET pulse oximetry. The SatShare cable attaches to the back of the Radical Docking Station, and SatShare cables are available to interface with most multi-parameter patient monitors. www.masimo.com 34 Masimo Radical-7 Handheld Chapter 2: Radical-7 Descriptions All user input and displays are controlled by this component. The patient cable connects into the connector on the Handheld instrument. The Handheld is battery powered and can be used either as a transport monitor or as a Handheld Pulse CO-Oximeter for spot checks. Handheld Front Panel The following figure numbers and corresponding table describes the hardware features of the Radical-7. 1 Handheld Release button Press down the Handheld Release Button to pull the Handheld off the Docking Station. 2 Touchscreen Display The Touchscreen Display refers to the interactive area on the Handheld. There are different Display Views that can appear in this area. For more about using the Touchscreen and Display Views, see Changing the Size of Parameter Values on page 51. 3 Profile button The Profile button provides instant access to the Profile Screen. See Chapter 5: Configurations on page 89. 4 Power button To turn on the Radical-7, press the Power button. To turn off, press and hold the button for more than 2 seconds. 5 Home button The Home button provides instant access to the Display View screen. 6 Alarm Silence Button The Alarm Silence button temporarily silences alarms. See Silencing the Alarms on page 95. 7 Speaker The speaker indicates audio alarms. Care should be taken not to cover the speaker and muffle the audible alarm volume. 8 Patient Cable Connector Connect a patient cable or a direct cable sensor into the Radical-7 35 Masimo www.masimo.com Radical-7 Handheld Back Panel Chapter 2: Radical-7 Descriptions The Handheld back panel features the interconnection to the Docking Station, an accessory mount for the pole clamp accessory and access to the Handheld battery pack. Item Description The Radical-7 Handheld interfaces with the Docking Station through this connector. The optional Pole Clamp accessory attaches to this holder. See the Directions for Use of the Pole Clamp accessory for attachment instructions. The Radical-7 Handheld is powered by a Lithium Ion battery located in this compartment. For battery care and replacement, see Battery Operation and Maintenance on page 127. 1 2 3 www.masimo.com 36 Masimo Radical-7 Standalone Chapter 2: Radical-7 Descriptions When the Radical-7 Handheld is placed into the Docking Station, they become a full-featured standalone system. In this manual, when the Handheld and the Docking Station are connected, they are referred to Standalone. The Standalone acts as a battery charger for the Handheld and has AC power connection capabilities. If the AC power from the wall outlet is temporarily interrupted, then the battery in the Handheld will allow continuous operation. The Standalone can also interface with serial instruments, Nurse Call, or analog output instruments, and multi-parameter patient monitors through a SatShare cable. There are several models of Docking Stations available. The following table outlines which features are available for each model of Docking Station. Docking Station Features RDS-1 RDS-2 RDS-3 SatShare Interface Serial RS-232 interface Nurse Call/Analog Output interface 10-Hour Extended Battery Automatic Display Rotation Support (Gravity Detector) Docking Station Battery Charging indicator Handheld Battery Charging indicator Visual Alarm indicator AC Power indicator Docking indicator The RDS-1 and RDS-3 are optionally available with SafetyNet capability. www.masimo.com 37 Masimo Radical-7 Standalone Front Panel Chapter 2: Radical-7 Descriptions The following figure and corresponding text review the features of the Radical-7 Standalone. Note that when the Standalone is turned on, all indicator LEDs initially turn on and off at start up. Item Description 1 2 3 4 5 Docking Station Battery Charging Indicator Handheld Battery Charging indicator The Docking Station Battery Charging indicator is illuminated when the Docking Station battery is charging. The indicator blinks just prior to charging. The Charging Indicator does not illuminate when the battery is fully charged or when the battery is not present. The Handheld Battery Charging indicator is illuminated when the Handheld battery is charging. The indicator blinks just prior to charging. The Charging Indicator does not illuminate when the battery is fully charged or when the battery is not present. Visual Alarm Indicator The Visual Alarm indicator is illuminated when an alarm condition is active and the Alarm Status Indicator is shown. AC Power Indicator Docking Indicator The AC Power indicator is illuminated when the Radical-7 Docking Station is plugged into AC line power. The Docking indicator is illuminated when the Handheld instrument is turned on and is properly interfaced to a Docking Station. www.masimo.com 38 Masimo Radical-7 Standalone Back Panel Chapter 2: Radical-7 Descriptions The following figure and corresponding text review the features of the Radical-7 Standalone. Item Connector Description 1 2 3 Serial Output connector Analog Output/Nurse Call connector SatShare Cable connector Use the Serial Output connector with a ferrite bead installed to connect a serial instrument, including a serial printer, a monitoring system or PC to the Radical-7. The data is provided in standard RS-232C format. All external instrument connections to the Serial Output connector must be IEC-60950 compliant. Use the Analog Output connector with a ferrite bead installed to interface with an analog output instrument, such as a chart recorder or Nurse Call system. All external instrument connections to the Analog Output/Nurse Call connector must be IEC-60950 compliant. See Serial Interface Specifications on page 118. Use the SatShare Cable connector to connect a SatShare cable to the SpO2 input connector of a multi-parameter patient monitor. All external instrument connections to the SatShare Cable Connector must be IEC-60601-1-1 compliant. SatShare cables are available to interface with most major multi-parameter patient monitors. Check the label on the SatShare cable and the SatShare Directions for Use to ensure that the correct cable is used for each type of patient monitor. Visit www.masimo.com for the latest SatShare cables and validated instruments. www.masimo.com 39 Masimo Radical-7 Chapter 2: Radical-7 Descriptions Item Connector Description 4 5 Power Entry module The Power Entry module contains the input connector for AC power and two fuses. The AC input provides power to the system from the AC line. Always connect the Radical-7 to the mains power for continuous operation and/ or battery recharging. Note: Use the power cord as the means to disconnect the instrument from the mains power supply. Equipotential Ground connector Use the Equipotential Ground connector for grounding. Monitor Interface The Radical-7 has a unique SatShare interface that links it to most existing multi-parameter patient monitors through the pulse oximetry patient cable or SpO2 input connector. Upgrades any approved and validated monitor to Masimo SET performance by using the calculated SpO2 and pulse rate determined by Radical-7 to simulate an ideal plethysmograph waveform, which is sent to the validated multi-parameter patient monitor. Connects into the SpO2 patient cable or SpO2 input connector of the multi-parameter patient monitor. See Setting Up and Using SatShare on page 44. www.masimo.com 40 Masimo Chapter 3: Setup The following chapter contains information about setting up the Radical-7 before use. Unpacking and Inspection To unpack and inspect the device 1. 2. 3. Remove the instrument from the shipping carton and examine it for signs of shipping damage. Check all materials against the packing list. Save all packing materials, invoice and bill of lading. These may be required to process a claim with the carrier. If anything is missing or damaged, contact the Technical Service Department. See Return Procedure on page 132. Docking Station Power Requirements Always use a hospital-grade, AC power cable to connect the Docking Station to an AC power source. Do not connect the Docking Station to an AC outlet that is controlled by a switch because the power to the Docking Station may be inadvertently switched off. Verify the AC power voltage and line frequency before use. Verify that the power source can provide an adequate power rating as indicated on the rear panel of the Docking Station. The Radical-7 is designed to operate on 100 to 240VAC, 47-63 Hz. The Radical-7 is rated at 55 VA max. Connect a hospital-grade power cable (IEC-320 connector type at the instrument) to the Power Entry module on the Docking Station. Connect the power cable to an AC power source. Ensure that the instrument is adequately powered by verifying that the AC power indicator on the Docking Station is illuminated. See Safety Information, Warnings, and Cautions on page 11. www.masimo.com 41 Masimo Radical-7 Chapter 3: Setup Setting Up the Docking Station Place the Docking Station on a stable hard flat surface near the patient. Always place the Radical-7 on a dry surface. Maintain a minimum of 3 cm (1 inch) free space around the Radical-7. Make sure that the Radical-7 speaker is not covered to avoid a muffled alarm sound. The Radical-7 Handheld, Docking Station or Standalone should not be operated outside the following environmental conditions:
Operating Environmental Conditions Temperature humidity Operating Altitude See Device Settings on page 74. Initial Battery Charging
+5C to +40C, +41F to +104F 5% to 95%, non-condensing 1060 mbar to 500 mbar pressure -1000 ft to 18,000 ft (-304 m to 5,486 m) Before use, the Radical-7 Handheld battery and the Docking Station battery must be charged completely. See Electrical Safety Information, Warnings, and Cautions on page 18. To charge the Handheld and Docking Station for the first time 1. 2. 3. 4. Attach the Handheld to the Docking Station. Plug in the AC power cord to power entry module. Make sure it is securely plugged in. Plug the AC power cord into an AC power source. Verify that the batteries are charging. The Battery Charging indicators on the Docking Station flash prior to charging and remain illuminated while the batteries are charging. See Standalone Front Panel on page 38 and Battery Operation and Maintenance on page 127. Setting Up for Philips, Agilent, or HP VueLink To set up for use with VueLink compatible monitors (Philips, Agilent, or HP) 1. On the Radical-7, on the device output screen, for the serial option, select Hp 2. 3. VueLink. Connect one end of the VueLink cable to the Serial Output connector on the Docking Station. Connect the other end of the VueLink cable to the VueLink module and insert the module into the VueLink compatible monitor rack. www.masimo.com 42 Masimo Radical-7 Chapter 3: Setup 4. 5. The SpO2 and pulse rate values appear on the VueLink compatible monitor. In order for the plethysmographic waveform to be displayed on the VueLink compatible monitor, and for the VueLink monitor to convey alarm conditions measured by the Radical-7, the VueLink compatible monitor must be properly configured. See instructions for use provided with the VueLink compatible monitor and the VueLink module. See Device Related Safety Information, Warnings, and Cautions on page 15 and Serial Interface Specifications on page 118. Setting Up for SpaceLabs Flexport To set up for use with SpaceLabs Flexport 1. On the Radical-7, on the device output screen, for the serial option, select SpaceLabs Flexport. Connect one end of the Spacelabs Flexport cable to the Serial Output connector on the Docking Station. Connect the other end of the Spacelabs Flexport cable to the Spacelabs Universal Flexport connector. The SpO2 and pulse rate values appear on the Spacelabs screen. In order for the plethysmographic waveform to be displayed on the Spacelabs screen, and for the Spacelabs monitor to convey alarm conditions measured by the Radical-7, the Spacelabs monitor must be properly configured. See instructions for use provided with the Spacelabs monitor. See Device Related Safety Information, Warnings, and Cautions on page 15 and Serial Interface Specifications on page 118. 2. 3. 4. 5. www.masimo.com 43 Masimo Radical-7 Chapter 3: Setup Setting Up and Using SatShare Parameter values from the Radical-7 can be displayed on a multi-parameter monitor through the SatShare feature. The SatShare feature provides an ideal, simulated plethysmographic waveform that corresponds to the parameter values determined the by Radical-7. This waveform may be used to display these values on multi-parameter monitors through the multi-parameter oximetry sensor or input connector. It is recommended that the Radical-7 be positioned near the multi-parameter monitor, with the Radical-7 screen displaying the plethysmographic waveform and the parameter values. Refer to the instructions for use provided with the multi-parameter monitor. See Device Related Safety Information, Warnings, and Cautions on page 15. To set up for use with SatShare interface 1. 2. 3. 4. Select the SatShare cable that is appropriate for the multi-parameter monitor. For the latest list of available SatShare cables and validated instruments, see www.masimo.com. Connect the labeled end of the SatShare cable to the SatShare Cable connector on the Docking Station. See Standalone Back Panel on page 39. For a secure connection, tighten the cable connector screws. Connect the other end of the SatShare cable to one of the following:
Sensor connector of the multi-parameter monitor cable Directly to the multi-parameter monitor Verify that the Radical-7 recognizes the SatShare cable. If functional, the name of the SatShare cable displays on the Radical-7 screen. 5. As appropriate, configure alarm limits on the multi-parameter monitor. 6. Set the averaging time for the multi-parameter monitor to its lowest setting (or fastest response). The ideal waveform for the Radical-7 requires additional averaging by the monitor. If the averaging time of the multi-parameter monitor is not changed, the time to display physiological changes in saturation on the monitor is increased with SatShare. However, the delay can be minimized by reducing the averaging time on the multi-parameter monitor. 7. While in the SatShare mode, if there are any significant discrepancies between the readings from Radical-7 and those on the monitor displaying the values obtained from SatShare, the values reported by the Radical-7 are to be considered the correct values. It is possible to use the Radical-7 with SatShare while the Radical-7 is not connected to AC power. However, in this configuration, battery run time is reduced. See Battery Operation and Maintenance on page 127. 8. 9. On the Radical-7, turn on the Satshare Numbers option. See Device Output on page 79. 10. If displaying the simulated waveform is not desirable, it is recommended to turn off the plethysmographic waveform display of the multi-parameter patient monitor. See Serial Interface Specifications on page 118. www.masimo.com 44 Masimo Chapter 4: Operation The following chapter contains information about using the Radical-7. Using the Touchscreen and Buttons 1. 2. Display View To access other screens, tap a value on the Display View. See About the Display View on page 48. Profiles button To the access the Profiles screen, press Profiles. See Chapter 5: Configurations on page 89. 3. Alarm Silence button To temporarily silence audible alarms, press Alarm Silence. See Silencing the Alarms on page 95. 4. Home button 5. To return to the Display View, press Home. Power button To turn on the Radical-7, press the Power button. To turn off, press and hold the button for more than 2 seconds www.masimo.com 45 Masimo Radical-7 Using Screenlock Chapter 4: Operation When turned on, the Screenlock feature may prevent accidental interaction with Display View. Using the Screenlock feature 1. When turned on, any interaction with the Display View triggers the Screenlock 2. feature. To bypass Screenlock when it appears, press and hold the Lock icon until it unlocks. 3. To turn on or turn off Screenlock, see Access Control on page 77. Using the Home Button One option to return to Display View is by using the Home button. To return to Display View using the Home button From any screen, press Home. www.masimo.com 46 Masimo Radical-7 Navigating the Radical-7 Chapter 4: Operation Navigate the Radical-7 screens via the Display View or the Main Menu. Display View The following is the primary interactive screen that the user views. To access the Main Menu screen Tap the gear icon at the lower right corner of the display. Main Menu The following is the Main Menu screen where users can access additional screens and information. Users can swipe the screen or tap the arrow icon to return to the Display View. See Accessing the Main Menu on page 55. www.masimo.com 47 Masimo Radical-7 About the Display View The Display View consists of different areas:
Status Bar See About the Status Bar on page 49. Chapter 4: Operation Parameter Display See Changing the Size of Parameter Values on page 51. Trend Field See Waveform and Trend Views on page 52. www.masimo.com 48 Masimo Radical-7 Chapter 4: Operation Small Parameter View. The gear icon on the right of the Small Parameter View is one way to access the Main Menu. See Accessing the Main Menu on page 55. About the Status Bar The Status Bar is visible on the top portion of the Display View. Status Bar Access additional screens, more information, or toggle features by tapping directly on any of the following indicators in the Status Bar. Sensitivity Modes. See Overview of Sensitivity Modes on page 50. Messages on page 100. (read only) WiFi on page 76. Battery on page 77. Sounds on page 73. Time settings. See Localization on page 75. www.masimo.com 49 Masimo Radical-7 Overview of Sensitivity Modes Chapter 4: Operation Three sensitivity levels enable a clinician to tailor the response of the Radical-7 to the needs of the particular patient situation. Access the menu by tapping on the indicator in the upper left corner of the Display View. The sensitivity levels are as follows:
Normal Sensitivity This is the recommended mode for patients that are experiencing some compromise in blood flow or perfusion. It is advisable for care areas where patients are observed frequently, such as an intensive care unit (ICU). Adaptive Probe Off Detection (APOD) Sensitivity This is the recommended monitoring mode where there is a high probability of the sensor becoming detached. It is also the suggested mode for care areas where patients are not visually monitored continuously. This mode delivers enhanced protection against erroneous pulse rate and arterial oxygen saturation readings when a sensor becomes inadvertently detached from a patient due to excessive movement. Maximum Sensitivity (MAX) This mode is recommended for patients with low perfusion or when the low perfusion message is displayed on the screen in APOD or normal sensitivity mode. This mode is not recommended for care areas where patients are not monitored visually, such as general wards. It is designed to interpret and display data at the measuring site when the signal may be weak due to decreased perfusion. When a sensor becomes detached from a patient, it will have compromised protection against erroneous pulse rate and arterial saturation readings. Toggle between the sensitivity modes by pressing the indication on the top left of the Display View. The sensitivity mode also can be changed in the Profile menu which is accessible from the Main Menu. Note that the device will revert to APOD mode after a power cycle. See Changing Profiles Settings on page 90. www.masimo.com 50 Masimo Radical-7 Chapter 4: Operation Changing the Size of Parameter Values To change the size of parameter values on the Display View 1. On the Small Parameter view, press and hold any one of the active parameters
(shown above). 2. When the parameter value dims, shakes, and grows in size, drag and drop that 3. 4. parameter onto the center of the screen. The parameter value appears on the screen in a larger format. Note that the device automatically configures the screen for optimal display of the parameter values. To remove parameter values from the larger format display, press and hold the larger parameter value. Then drag and drop the parameter value back to the Small Parameter view. 5. When you finish making changes, you can lock the customized format. See Parameter Layout. Trend Field The Trend Field allows users to access various customizable views. See Trends on page 81. www.masimo.com 51 Masimo Radical-7 Chapter 4: Operation To access trend, waveform, or customize the views on the Display View screen 1. 2. Tap the Trend Field. The following screen appears. 3. 4. 5. Swipe up or swipe down the available options. Tap on the desired option. The Trend Field displays trend data specific to the selected option. Waveform and Trend Views The following section contain information about trends and waveforms available from the Trend Field on the Display View screen. The following are examples of some of the views that are available. www.masimo.com 52 Masimo Radical-7 Chapter 4: Operation Pleth + Siq IQ + Acoustics View Shows the parameter values on the top of the screen. The waveform is located below the parameter values. The Small Parameter view is located along the bottom of the screen. This view contains the Pleth Waveform with signal quality indications superimposed on top of the acoustic waveform. Signal Quality (Sig IQ) Indicators The signal quality, on each individual pulsation, is conveyed by vertical bars, as shown below. The height of the bar indicates signal quality. The higher the bar, the better the signal quality. www.masimo.com 53 Masimo Radical-7 Pleth + Sig IQ View Chapter 4: Operation Shows the parameter values on the top of the screen. The waveform is located below the parameter values. The Small Parameter view is located along the bottom of the screen. This view contains the Pleth Waveform with signal quality indications only. RRa Waveform View Shows the parameter values on the top of the screen. The RRa waveform is located below the parameter values. The Small Parameter view is located along the bottom of the screen. Acoustic Respiratory Rate (RRa) must be available for this feature to be shown. This view contains acoustic respiratory rate waveform only. www.masimo.com 54 Masimo Radical-7 Parameter Quick Trend View Chapter 4: Operation This view displays the quick trend of the selected parameter over an adjustable period of time. The default is 1 hour. Expand the quick trend to the full trend view by tapping the expand icon of the waveform display. Using two fingers, the user can expand and zoom the information that is inside this Trend Field. Accessing the Main Menu To access Main Menu from the Display View, tap the gear icon on the bottom right corner of the Small Parameter View. www.masimo.com 55 Masimo Radical-7 Navigating the Main Menu Chapter 4: Operation From the Main Menu screen, tap the icons for any of the following screens:
Device Settings See Device Settings on page 74. Parameter Settings See Parameter Settings on page 58. Profiles See Changing Profiles Settings on page 90. 3D Alarms See 3D Alarms on page 98. www.masimo.com 56 Masimo Radical-7 Chapter 4: Operation Trends See Trends on page 81 and Trend Field on page 51. Sound See Sounds on page 73. About See About on page 87. www.masimo.com 57 Masimo Radical-7 Parameter Settings Chapter 4: Operation The following is an example of the Parameter Settings screen. Only parameters that have been loaded onto your system will be visible. To access any of the available parameter setting screens 1. 2. From the Parameter Settings screen, to access your desired parameter, flick the on-screen icons left or right. Tap the icon of your desired parameter. For details, see any of the following sections. SpO2 Settings on page 61. RRa Settings on page 66. RRp Settings on page 67. SpHb Settings on page 62. PVI Settings on page 69. SpCO Settings on page 70. SpMet Settings on page 71. SpOC Alarms on page 72. www.masimo.com 58 Masimo Radical-7 Chapter 4: Operation Accessing Parameter Information Additional information about each parameter is available. To access additional information about parameters 1. From the parameter settings screen, tap the About icon. The following is an example for SpHb. 2. An About screen appears for the selected parameter. www.masimo.com 59 Masimo Radical-7 Chapter 4: Operation Overview of In Vivo Adjustment The In Vivo Adjustment feature lets clinicians manually adjust one or more clinical parameters to match that of a corresponding laboratory reference for continuous trending. To remind clinicians that the feature is active, an offset value displays alongside the adjusted parameter value. When the In Vivo Adjustment is set to Yes, the feature is active (turned on) and a positive or a negative offset value appears, as shown in the following illustration. In Vivo offset is set to zero for any of the following:
Cable or sensor disconnect from instrument. Sensor goes off patient causing a sensor initialization to occur. 8 hours has elapsed since the In Vivo value was activated. The user turns off In Vivo. Offset value Offset value appears and indicates that In Vivo Adjustment is active. A positive value means that the value is increased (according to a laboratory reference value as entered by a clinician) and a negative value means the value is decreased (according to a laboratory reference value as entered by a clinician). In the example below, the SpO2 value is offset (highlighted) by -1.0 and SpHb is offset by
+0.4. The In Vivo Adjustment feature can be set to Yes or No. the factory default setting is No. If set to Yes, the parameter value is adjusted and an offset value appears. The offset value is set by the user. The feature applies to any of the following parameters:
In Vivo for SpO2 on page 62. In Vivo for SpHb on page 64. In Vivo for SpCO on page 71. In Vivo for SpMet on page 72. www.masimo.com 60 Masimo Chapter 4: Operation Radical-7 SpO2 Settings Access any of the following options:
Alarms for SpO2. Additional Settings for SpO2 on page 62. Accessing Parameter Information on page 59. Alarms for SpO2 From the Alarms screen, change any of the following options:
Options Description Factory Default Settings Configurable Options High Limit High Limit is the upper threshold that triggers an alarm. Off 2% to 99% in steps of 1, or Off When set to Off, alarms are disabled Low Limit Low Limit is the lower threshold that triggers an alarm. 88%
1% to 98% in steps of 1 Rapid Desat Sets the Rapid Desat limit threshold to the selected amount below the Low Alarm Limit. When SpO2 value falls below rapid desat limit the audio and visual alarm are immediately triggered without respect to the alarm delay.
-10%
Off, -5%, or -10%
Alarm Delay When an alarm condition is met, this feature delays the audible part of an alarm. 5 seconds 0, 5, 10, or 15 seconds ATA establishes patient-specific limit thresholds based upon the baseline value of the parameter. See Adaptive Threshold Alarm (ATA) Feature on page 97. Adaptive Threshold Alarm (ATA) Off Off or On www.masimo.com 61 Masimo Radical-7 Additional Settings for SpO2 Chapter 4: Operation From the Additional Settings screen, you change any of the following options:
Options Description Factory Default Settings User Configurable Settings Averaging Time The length of time over which 8 seconds the system calculates the average of all data points. 2-4, 4-6, 8, 10, 12, 14, or 16 seconds FastSat See Overview of FastSat on page 62. Off On or Off Overview of FastSat FastSat enables rapid tracking of arterial oxygen saturation changes. Arterial oxygen saturation data is averaged using pulse oximeter averaging algorithms to smooth the trend. When the Radical-7 is set to FastSat On, the averaging algorithm evaluates all the saturation values providing an averaged saturation value that is a better representation of the patients current oxygenation status. With FastSat, the averaging time is dependent on the input signal. In Vivo for SpO2 From the In Vivo screen, you change any of the following options:
Options Description Factory Default Settings User Configurable Settings Enabled See Overview of In Vivo Adjustment on page 60. No Yes or No Offset Amount See Overview of In Vivo Adjustment on page 60. 0 when turned on
%SpO2 can be adjusted to a difference of 6%
SpHb Settings From the SpHb Settings screen, you access any of the following screens:
SpHb Alarms on page 63. Additional Settings for SpHb on page 63. Accessing Parameter Information on page 59. www.masimo.com 62 Masimo Radical-7 SpHb Alarms Chapter 4: Operation From the Alarms screen, change any of the following options:
Options Description High Limit The High Limit is upper threshold that triggers an alarm. Factory Default Settings 17.0 g/dL
(11.0 mmol/L) Low Limit The Low Limit is lower threshold that triggers an alarm. 7.0 g/dL
(4.0 mmol/L) User Configurable Settings 2.0 g/dL to 24.5 g/dL in steps of 0.1 g/dL, or Off
(2.0 mmol/L to 15.0 mmol/L in steps of 0.1 mmol/L, or Off) When SpHb Precision is set to 1.0, the values are rounded down. When set to Off, alarms are disabled. Off, or 1.0 g/dL to 23.5 g/dL in steps of 0.1 g/dL
(Off, or 1.0 mmol/L to 14.5 mmol/L, in steps of 0.1 mmol/L) When SpHb Precision is set to 1.0, values are rounded down. When set to Off, alarms are disabled. Additional Settings for SpHb From the Additional Settings screen, you change any of the following options:
Options Description Averaging Time The length of time over which the system calculates the average of all data points. Calibration Provides an Arterial (SpHb) or Venous (SpHbv) that displays on the main screen. Factory Default Settings User Configurable Settings Medium Short, Medium, or Long Venous Arterial or Venous www.masimo.com 63 Masimo Radical-7 Precision Unit of Measure Chapter 4: Operation Allows the user to set the decimal for SpHb. 0.1 0.1, 0.5, or 1.0 (whole numbers) Displays total hemoglobin
(SpHb) as g/dL (grams per deciliter) or mmol/L
(milimoles per liter). g/dL mmol/L or g/dL In Vivo for SpHb From the In Vivo screen, change any of the following options:
Options Description Factory Default Settings User Configurable Settings Enable/Disable No Yes or No See Overview of In Vivo Adjustment on page 60. See Overview of In Vivo Adjustment on page 60. 0.0 SpHb ( 3 g/dL with steps of 0.1 g/dL) In Vivo Calibration In Vivo Calibration Offset PR Settings From the Additional Settings screen, you change any of the following options:
PR Alarms on page 64. Accessing Parameter Information on page 59 PR Alarms From the PR Alarms screen, change any of the following options:
Options Description Factory Default Settings Options High Limit The High Limit is upper threshold that triggers an alarm. 140 BPM 35 to 235 BPM, in steps of 5 BPM Low Limit The Low Limit is lower threshold that triggers an alarm. 50 BPM 30 to 230 BPM, in steps of 5 BPM www.masimo.com 64 Masimo Radical-7 Perfusion Index (PI) Settings Chapter 4: Operation From the PI Settings screen, access any of the following screens:
PI Alarms on page 65. Additional Settings for PI on page 65. Accessing Parameter Information on page 59. PI Alarms From the Alarms screen, change any of the following options:
Options Description High Limit The High Limit is upper threshold that triggers an alarm. Low Limit The Low Limit is lower threshold that triggers an alarm. Factory Default Settings User Configurable Settings Off Step Size:
0.04 to 0.09 in steps of 0.01 0.10 to 0.90 in steps of 0.10 1 to 19 in steps of 1, or Off Off Step Size:
Off, or 0.03 to 0.09 in steps of 0.01 0.10 to 0.90 in steps of 0.10 1 to 18 in steps of 1 Additional Settings for PI From the Additional Settings screen, change any of the following options:
Options Description Averaging Time The length of time over which the system calculates the average of all data points. Factory Default Settings User Configurable Settings Long Short or Long www.masimo.com 65 Masimo Radical-7 RRa Settings Chapter 4: Operation From the RRa Settings screen, access any of the following screens:
RRa Alarms on page 66. Additional Settings for RRa on page 67. Accessing Parameter Information on page 59. RRa Alarms From the Alarms screen, change any of the following options:
Options Description High Limit Low Limit Respiratory Pause Alarm Delay The High Limit is upper threshold that triggers an alarm. The Low Limit is lower threshold that triggers an alarm. The duration of time that triggers an alarm if no breaths are detected. When a High or Low alarm condition occurs, this feature delays the audible part of an alarm. Configurable Options Factory Default Settings 30 breaths per minute 5 to 69 breaths per minute in steps of 1, or Off 4 breaths per minute 4 to 68 breaths per minute in steps of 1 15 seconds 20, 25, 30, 35, 40, or 15 seconds 30 seconds 60, 0, 10, 15, or 30 seconds www.masimo.com 66 Masimo Radical-7 Additional Settings for RRa Chapter 4: Operation From the Additional Settings screen, change any of the following options:
Factory Default Settings Slow User Configurable Settings Trending, No Averaging, Fast, Medium, or Slow 5 minutes 10, 15, 0, 1, or 5 minutes Options Description Averaging Time Freshness The length of time over which the system calculates the average of all data points. The duration of time that, during interference, the system displays the last valid reading. RRp Settings If RRa is loaded on the Radical-7, RRp alarms and RRp settings are used instead of RRa. There are no separate screens for RRp. From the RRp Settings screen, access any of the following screens:
RRp Alarms on page 68. Additional Settings for RRp on page 69. www.masimo.com 67 Masimo Radical-7 Overview of RRp Chapter 4: Operation Respiration rate can be determined by the plethysmographic waveform (RRp) using a SpO2 sensor. Respiration rate monitoring, derived from the plethysmographic waveform, measures a patient's respiratory rate based on amplitude changes that correspond to the respiratory cycle. Inaccurate measurements may be caused by:
Low arterial perfusion Motion artifact Low arterial oxygen saturation Excessive ambient or environmental noise Improper sensor placement The device can display respiration rate derived from the plethysmographic waveform under the following conditions:
Respiration rate parameter installed on the instrument Active optical channel Dual Rainbow cable is disconnected RRp Alarms From the Alarms screen, change any of the following options:
Options Description Factory Default Configurable Options High Limit The High Limit is upper threshold that triggers an alarm. 30 Low Limit The Low Limit is lower threshold that triggers an alarm. 6 5 to 69 breaths per minute, or Off 4 to 68 breaths per minute, or Off Alarm Delay When an alarm condition is met, this feature delays the audible part of an alarm. 30 0, 10, 15, 30, 45 www.masimo.com 68 Masimo Radical-7 Additional Settings for RRp Chapter 4: Operation From the Additional Settings screen, change any of the following options:
Options Description Factory Default Settings User Configurable Settings Averaging Freshness Slow The length of time over which the system calculates the average of all data points. No, Fast, Medium, Slow, Trending 5 minutes 0, 1, 5, 10, 15 minutes The duration of time that, during interference, the system displays the last valid reading. PVI Settings From the PVI Settings screen, access any of the following options:
PVI Alarms on page 69. Additional Settings for PVI on page 70. Accessing Parameter Information on page 59 PVI Alarms From the Additional Settings screen, change any of the following options:
Options Description User Configurable Settings Factory Default Settings High Limit The High Limit is upper threshold that triggers an alarm. Off 2 to 99, in steps of 1, or Off When set to Off, alarms are disabled. Low Limit The Low Limit is lower threshold that triggers an alarm. Off Off, 1 to 98 in steps of 1 When set to Off, alarms are disabled. www.masimo.com 69 Masimo Radical-7 Additional Settings for PVI Chapter 4: Operation From the Additional Settings screen, change any of the following options:
Factory Default Settings User Configurable Settings Long Short or Long Options Description Averaging Time The length of time over which the system calculates the average of all data points. SpCO Settings From the SpCO Settings screen, you can access the following screens:
SpCO Alarms on page 70. Accessing Parameter Information on page 59. SpCO Alarms From the SpCO Settings screen, access the following screens:
Options Description User Configurable Settings Factory Default Settings High Limit The High Limit is the upper threshold that triggers an alarm. 10 2% to 98%, in steps of 1%, or Off When set to Off, alarms are disabled. Low Limit The Low Limit is the lower threshold that triggers an alarm. Off Off, 1% to 97%, in steps of 1%
When set to Off, alarms are disabled. www.masimo.com 70 Masimo Radical-7 In Vivo for SpCO From the In Vivo screen, access the following screens:
Chapter 4: Operation Options Description Enabled See Overview of In Vivo Adjustment on page 60. Offset Amount See Overview of In Vivo Adjustment on page 60. SpMet Settings Factory Default Settings No 0 User Configurable Settings Yes, No Yes, No 9% in steps of 1%
From the SpMet Settings screen, access the following screens:
SpMet Alarms on page 71. Accessing Parameter Information on page 59. SpMet Alarms From the Alarms screen, change any of the following options:
Options Description High Limit Low Limit The High Alarm Limit is upper threshold that triggers an alarm. The Low Alarm Limit is lower threshold that triggers an alarm. Factory Default Settings User Configurable Settings 3.0 1.0% to 2.0% in steps of 0.1 2.5% to 99.5% in steps of 0.5%, or Off Off Off, 0.1% to 2.0% in steps of 0.1%
2.5% to 99%, in steps of 0.5%
www.masimo.com 71 Masimo Radical-7 In Vivo for SpMet From the In Vivo screen, access the following screens:
Chapter 4: Operation Options Description User Configurable Settings Factory Default Settings Enabled Match the corresponding laboratory reference for continuous trending. No Yes or No
3.0% in steps of 0.1%
Offset Amount Helps offset individual patient bias that is expected when comparing a noninvasive measurement to a laboratory reference. SpOC Settings From the SpOC Settings screen, access the following screens:
SpOC Alarms on page 72. Accessing Parameter Information on page 59. SpOC Alarms From the SpOC Alarms screen, access the following screens:
Options Description High Limit The High Limit is the upper threshold that triggers an alarm. Low Limit The Low Limit is the lower threshold that triggers an alarm. User Configurable Settings Factory Default Settings Off 2% to 34% in steps of 1, or Off Off Off, or 1% to 33% in steps of 1 www.masimo.com 72 Masimo Radical-7 Sounds Chapter 4: Operation From the Sounds screen, change any of the following options:
Options Description User Configurable Settings Factory Default Settings Alarm Volume Sets the alarm volume and provides a sample of the alarm volume. Level 4 Level 1 to 4 Pulse Tone Volume Sets the volume of the tone that conveys the pulse rate. Level 3 Off, Level 1 to 4 Silence Duration Length of time that the audible alarm remains muted. 120 second 30, 60, 90, or 120 seconds www.masimo.com 73 Masimo Radical-7 Device Settings Chapter 4: Operation The following is an example of the Device Settings screen From the Device Settings screen, you access any of the following options:
Screen Orientation on page 75. Localization on page 75. WiFi on page 76. Battery on page 77. Brightness on page 77. Parameter Layout. Access Control on page 77. Device Output on page 79. www.masimo.com 74 Masimo Radical-7 Screen Orientation Chapter 4: Operation From the Screen Orientation screen, change any of the following options:
Options Description Factory Default Settings User Configurable Settings Auto Orientation Allows the device to automatically adjust the Display screens depending on orientation. On Off or On Orientation Rotates the viewing screens depending on device orientation. Landscape Landscape: rotates the screen to horizontal viewing position Inverted Landscape: rotates the screen to (180 degree) viewing position Portrait: rotates the screen to vertical viewing position Inverted Portrait: rotates the screen to vertical (180 degree) viewing position Localization From the Localization screen, change any of the following options:
Options Description Current Date Default date Current Time Default time Factory Default Settings User Configurable Settings N/A N/A N/A N/A Language Language in which the screens display. English Choose from available languages. www.masimo.com 75 Masimo Radical-7 Time Zone Chapter 4: Operation Setting based on Coordinated Universal Time
(UTC). A (UTC+1hr) Choose your local time zone settings. Date Format Set the format of the date display on the Display View. MM/DD/YYYY MM/DD/YYYY DD/MM/YYYY Time Format Set the format of the time display as it will be shown on the Display View. Line Frequency Set to match regional power line frequency to allow for cancelation of noise introduced by fluorescent lights and other sources. 12 hour 24 hour or 12 hour 60 Hz 50 Hz or 60 Hz Auto Set Date/Time When connector to a WiFi network, automatically sets date and time. Off On or Off Manually set the numerical date if Auto Set Date/Time is Off. MM/DD/YYYY Choose month, date, and year. Manually set the hour and minute, AM or PM, if Auto Set Date/Time is Off. 12-hour format Choose hour and minute. Date Time WiFi When the Radical-7 is connected to a WiFi network, the Wifi icon located on the Status Bar conveys the strength of the Wifi connection. See About the Status Bar on page 49. From the Wifi screen, change any of the following options:
Options Description Factory Default Settings User Configurable Settings WiFi Enables or disables the wireless connection Off On or Off Additional fields in the Wifi screen provide information about WiFi connection. These additional fields are read only and not configurable. www.masimo.com 76 Masimo Radical-7 Battery Chapter 4: Operation From the Battery screen, view the following information:
Battery icon that conveys remaining battery charge as a green color. Battery icon that conveys that battery charging status. See About the Status Bar on page 49. Battery time remaining conveyed as a percent value. See Battery Operation and Maintenance on page 127. Brightness From the Brightness screen, change any of the following options:
Factory Default Settings User Configurable Settings 4 Level 1 to 4 Options Description Brightness The slider option adjusts the brightness level of the display and provides a sample of the brightness level. Access Control The Access Control screen is protected by a Password screen. Password Screen www.masimo.com 77 Masimo Radical-7 Using the Password screen Chapter 4: Operation 1. On the Password screen, enter the following numbers:
6 2 7 4 Tap Enter. 2. 3. To undo numbers, tap Backspace. Access Control Screen From the Access Control screen, change any of the following options:
Options Description All Mute All patient alarm conditions are silenced. Only system alarms will be indicated by an audible alarm. Lock Alarm Volume When set to 3 or 4, 3 or 4 shows dimly lit in the Alarm Volume section of the Alarms Menu screen and cannot be changed. SpO2 Low %
Limit Threshold at which SpO2 Low Alarm Limit cannot be reduced. Lock Layout Prevents the user from making changes to the parameter layout. Factory Default Settings User Configurable Settings Off On or Off Off 3, 4, or Off Off Off 1% to 98% in steps of 1, or Off On or Off Screen Lock Prevents the user from interacting with the device. Off On or Off Legacy Mode Changes the Display View from color to simplified monochrome. Color Mono or Color www.masimo.com 78 Masimo Radical-7 Save as Adult Saves pre-configured profiles for adult patients. N/A Save as Neo Saves pre-configured profiles for neonatal patients N/A Factory Defaults Options are restored to factory values. N/A Chapter 4: Operation Press Save to load all device configuration settings to adult profile. Press Save to load all device configuration settings to neonatal profile. Press Restore to return to factory default values. Device Output From the Device Output screen, change any of the following options:
Options Description Serial Analog 1 Analog 2 All serial output is RS-232 based. An interface with various analog recording instruments and/or strip chart recorders through connector located on Docking Station. Depending on the configuration, the following parameters are output continuously on the Analog 1 and Analog 2. Factory Default Settings ASCII 1 SpO2 0%
to100%
User Configurable Settings ASCII 1, IAP, HP Vuelink, and SpaceLabs Flexport SpO2 50% to 100%, Pulse rate, Pleth, SIQ, 0V Output, 1V Output, SpO2 0% to 100%
Pulse Rate Pleth, SIQ, 0V Output, 1V Output, SpO2 0% to 100%, SpO2 50% to 100%, or Pulse rate Nurse Call Trigger The nurse call output will be activated based on the alarm events. The nurse call with be activated based on Low Signal or Alarm and Low Signal IQ events. Alarms Alarms + SIQ, Low SIQ, Alarms www.masimo.com 79 Masimo Radical-7 Nurse Call Polarity Chapter 4: Operation Can be inverted to accommodate various nurse call station requirements. Normal Normal or Inverted www.masimo.com 80 Masimo Radical-7 Trends Chapter 4: Operation The following sections describe Trend Views and how to adjust trend settings. About Trend Views There are different ways to view trend information. The following is an example of trend information for SpO2 as it appears within the Display View screen. The following is an example of trend information for SpO2 as it appears in the Full Trend screen. www.masimo.com 81 Masimo Radical-7 Chapter 4: Operation Changing Between Trend Views To toggle between Display View and Full Trend From the Display View, in the Trend Field, tap the icon as shown below. From the Full Trend screen, tap the icon as shown below. Manipulating Trend Data To manipulate the view of trend data On the Full Trend screen, using two fingers, the user can zoom in or zoom out of the trend time scale. www.masimo.com 82 Masimo Radical-7 Chapter 4: Operation The user can add parameters to the trend view by dragging and dropping parameters from the Small Parameter view. To view past patient trend data, swipe the trend display to the left or to the right. www.masimo.com 83 Masimo Radical-7 Chapter 4: Operation To exit a trend view, press the Home button. www.masimo.com 84 Masimo Radical-7 Changing Trend Settings Chapter 4: Operation Change the maximum value and the minimum value of the Y axis for any of the available parameters. To adjust the trend settings for any of the available parameters 1. From the Main Menu screen, tap the Trends icon. 2. From the Trends screen, tap any of the available parameters. 3. From the Parameter Trends screen, tap the slider for the Y-axis maximum or the Y-axis minimum.The following is an example of the SpO2 Trend screen. www.masimo.com 85 Masimo Radical-7 Chapter 4: Operation 4. Select the desired setting by scrolling up or down. 5. When finished, select OK. The following example shows the Y axis range for SpO2. www.masimo.com 86 Masimo Radical-7 About Chapter 4: Operation From the About screen, access any of the following options:
Options Description Serial Number Displays the serial number of the Handheld. MCU Displays the version number of the instrument board software. MX Board Displays the version number of the technology level software. Processor Displays the version number of the system level software. If docked, displays the current software version of the Docking Station. Docking Station www.masimo.com 87 Masimo Chapter 5: Configurations The Radical-7 can be configured for various patient populations. About Profiles and Default Settings The Radical-7 contains a Profiles feature. Profiles are custom settings for different patient populations. These settings can be changed and the Radical-7 remembers the preferred settings after a power cycle. The default settings are settings that are pre-loaded as the device is initially powered on. There are different ways to access the Profiles settings. The first way is by the tapping the Profiles shortcut that is located in the Status Bar, as show below. Another way to access Profiles feature is by pressing the Profile button, as shown below. www.masimo.com 89 Masimo Radical-7 Chapter 5: Configurations A third way, from the Main Menu screen, is by tapping the Profiles icon. The Radical-7 stores three types of Profiles:
Custom (Default) Any changes to Alarm settings are retained after a power cycle. Adult Any changes to Alarm settings are reset to pre-defined settings. The Adult profile is the default profile after a power cycle. Neonatal Any changes to Alarm settings are reset to pre-defined Neonatal defaults after a power cycle. After Custom values are changed by the user, the Radical-7 retains these values after a power cycle. The Adult profile and the Neonatal profile can be changed to meet specific hospital or instrument requirements. If the Alarm settings are changed by the user, then these settings return to the instrument default values after it is powered down. A user can also load configurations to the device, which can be enabled by using the Profile button. Changing Profiles Settings From the Profiles screen, change any of the following options:
Options Description Patient Type Sensitivity Modes Defines the patient population for which the device will operate. See About Patient Types on page 92. Defines the sensitivity level for which the device will operate. See Overview of Sensitivity Modes on page 50. Factory Default Settings User Configurable Settings Custom Neonatal, Adult, Custom, or Custom 1 to 5 APOD MAX, APOD, or NORM www.masimo.com 90 Masimo Radical-7 Silence Duration Smart Tone Chapter 5: Configurations The amount of time for which the audible part of an alarm will be silenced. 120 30, 60, 90, 120 seconds Allows the audible pulse to continue to beep when the pleth graph shows signs of motion. Off On or Off www.masimo.com 91 Masimo Radical-7 About Patient Types Chapter 5: Configurations The Patient Type option defines the patient population for which the device will operate. As a safety measure, one of the ways the Radical-7 conveys the active patient type is by changing the color of the Profiles button. The following lists three examples of patient types and their respective colors. Patient Type Adult (blue) Example of Profile button Neonatal (purple) Custom (colors vary depending on the selected custom setting) www.masimo.com 92 Masimo Radical-7 Chapter 5: Configurations Powering Off the Radical-7 When turning off the Radical-7, the device remembers the preferred settings. To turn off the Radical-7 1. Press and hold the button for more than 2 seconds. 2. To confirm the shutdown process, the following screen appears. www.masimo.com 93 Masimo Chapter 6: Alarms and Messages The following chapter contains information about alarms and messages. For more information, see Chapter 7: Troubleshooting on page 107. About Alarms The Radical-7 visually and audibly indicates alarm conditions that the system detects. Audible alarms may be silenced, without affecting the operation of visual alarms. See Safety Information, Warnings, and Cautions on page 11 and Alarm Related Safety Information, Warnings, and Cautions on page 20. There are three priorities for alarms:
High Medium Low Alarm Delay When an alarm condition is met, this feature delays the audible part of an alarm. Silencing the Alarms Alarms are conveyed in ways: audible, visual, or both. The following is an example of a visual alarm for an exception message:
www.masimo.com 95 Masimo Radical-7 Chapter 6: Alarms and Messages The following is an example of a typical alarm. To silence or dismiss alarms:
Tap Silence (the highlighted area of the Status Bar). Audible alarms can be temporarily suspended. With the exception of Sleep mode, there are two audible alarm suspension settings, all controlled by the Alarm Silence button. Repeated pressing of the Alarm Suspend button will cycle through two alarm suspend options. To silence audible alarms When an audible is active, push Alarm Silence one time. 1. 2. The audible alarm will be silenced for 120 seconds. Pushing the button a second time restores the previous alarm settings. www.masimo.com 96 Masimo Radical-7 Chapter 6: Alarms and Messages Adaptive Threshold Alarm (ATA) Feature The Adaptive Threshold Alarm (ATA) feature is an optional feature that provides continuous SpO2 surveillance while allowing the clinician a useful tool to help reduce the frequency of audible alarms. ATA establishes the alarm limit threshold based upon the patient-specific baseline value of the SpO2 parameter which is determined from the recent history of SpO2 values. An Adaptive Threshold Limit is continuously determined for the patient and SpO2 values outside the Adaptive Threshold Limit trigger an audible alarm. The Adaptive Threshold Limit is bound by the standard SpO2 low alarm limit and the Rapid Desat low alarm limit. SpO2 values that exceed the Rapid Desat limit, whether it occurs rapidly or not, will activate an audible alarm. Prior to activating ATA, please review and select the appropriate standard low alarm limit and other alarm settings. Once ATA is selected, the Rapid Desat Alarm protection is always active. If the ATA low alarm limit is violated, ATA generates an audible alarm. It is important to note that once activated, ATA has the following automatic safety features:
Reminder Tones If an SpO2 value from a patient drops below the standard low alarm limit set by the user, a visual alert will display and a reminder tone will repeat every 15 minutes as long as the condition persists. If the SpO2 value drops below the ATA low alarm limit, an audible alarm will be activated. Rapid Desat Alarm Protection The Rapid Desat feature is always active when ATA is turned on. This means that deep desaturations (5% or 10%) from the standard SpO2 low alarm limit immediately generate an audible alarm. When used with ATA, it also serves as absolute low alarm limit protection. SpO2 values exceeding the Rapid Desat low alarm limit, whether rapid or not, will activate an audible alarm. You can change the Rapid Desat default setting from 5% to 10%. ATA does not allow a Rapid Desat default setting of 0%. When ATA is turned Off, the instrument uses the standard alarm limits and standard alarm delays. About Alarms on page 95. Alarms for SpO2 on page 61. www.masimo.com 97 Masimo Radical-7 3D Alarms Chapter 6: Alarms and Messages 3D Alarms include the Desat Index Alarm and the PI Delta Alarm. Overview of Desat Index Alarm on page 98. Overview of Perfusion Index (PI) Delta Alarm on page 99. Overview of Desat Index Alarm The Desat Index Alarm is a user-selectable feature which allows a clinician to request an audible and visual alarm if a patient experiences a specified number of desaturations over a specific period of time. Traditional high and low SpO2 alarm limits alert clinicians to saturation levels that exceed user selected thresholds, and these thresholds are typically established at a considerable change from the patients baseline saturation level. However, in select patient populations, substantial desaturation events that exceed a typical low alarm limit threshold may be preceded by a cycle of transient desaturations over a limited timeframe, and the ability to alert clinicians to a cycle of these smaller desaturations may provide an earlier indication of a potential significant decline in the patients status and the need for more focused monitoring and/or a change in treatment. To address patient populations at risk for cyclic, moderate desaturations, the option includes a user-selectable Desat Index Alarm which allows the clinician to request an audible and visual alarm in the event the patient experiences a specified number of desaturations beyond a defined level from the patients baseline saturation over a specific window of time, with each of these variables selectable by the user within established ranges as noted in Desat Index Settings on page 98. Desat Index Settings From the Desat Index Settings screen, access the following screens:
Desat Index Alarms on page 99. Accessing Parameter Information on page 59. www.masimo.com 98 Masimo Radical-7 Desat Index Alarms Chapter 6: Alarms and Messages From the Alarms screen, change any of the following options:
Options Description Factory Default Settings User Configurable Settings Delta Time See Overview of Desat Index Alarm on page 98. 4%
2% to 10% in steps of 1%. See Overview of Desat Index Alarm on page 98. 1 hour 1 to 4 hours, in steps of 1.
# of Events See Overview of Desat Index Alarm on page 98. Off Off, 1 to 24 desaturations in steps of 1. Overview of Perfusion Index (PI) Delta Alarm The PI Delta Alarm is a user-selectable feature which allows a clinician to request an audible and visual alarm if perfusion at the monitored site decreases by a specified level (delta) over a specific period of time. Perfusion Index gives an indication of the level of perfusion at the monitored site. The Radical-7 measures perfusion at the SpO2 site by comparing the pulsatile signal to the non-pulsatile signal, and expressing that ratio as a percentage. PI has been clinically proven to be useful as a predictor of the level of illness in neonates and adults and that PI may change dramatically in response to sympathetic changes caused by inhalational agents and pain stimulation*. If PI decreases over time, there may be underlying physiological reasons that may need to be addressed. The PI Delta provides an audible and visual alert to important changes in perfusion compared to the patients baseline PI rate. The baseline is set by the Radical-7 once the user has enabled the alarm. The baseline is 30 seconds of currently averaged PI. The feature includes a user-selectable PI Delta Alarm. This allows the clinician to request an audible and visual alarm if perfusion at the monitored site decreases by a specified level (delta) over a specified window of time. Three of the variables are selectable by the user within established ranges as noted in PI Delta Settings on page 99.
*De Felice C, Latini G, Vacca P, Kopotic RJ. The pulse oximeter perfusion index as a predictor for high illness severity in neonates. Eur J Pediatr. 2002;161:561-562. PI Delta Settings From the PI Delta Settings screen, you can access the following screens:
Overview of Perfusion Index (PI) Delta Alarm on page 99. Accessing Parameter Information on page 59. www.masimo.com 99 Masimo Radical-7 PI Delta Alarms Chapter 6: Alarms and Messages From the Alarms screen, change any of the following options:
Options Description User Configurable Settings Factory Default Settings Set Baseline See Overview of Perfusion Index (PI) Delta Alarm on page 99. Off On or Off See Overview of Perfusion Index (PI) Delta Alarm on page 99. Off, or PI baseline N/A See Overview of Perfusion Index (PI) Delta Alarm on page 99. 50%
10% to 99% in steps of 1 See Overview of Perfusion Index (PI) Delta Alarm on page 99. None 1 min, 5 min, 30 min, 1 hr, 4 hr, 8 hr, 12 hr, 24 hr, 36 hr, 48 hr, or None. Baseline Percent Change Timeout Messages The following section lists common messages, their potential causes, and next steps. Replace Sensor Message Message:
(Pulse Ox) Replace Sensor
(RAM) Replace Sensor SpHb reusable sensor has used all its available monitoring time. Sensor is non-functional. Defective sensor. Next steps: Replace sensor. www.masimo.com 100 Masimo Chapter 6: Alarms and Messages Radical-7 Replace Cable Message Message:
(Pulse Ox) Replace Cable
(RAM) Replace Cable The patient cable is non-functional or the life of the cable has expired. Next steps: Replace the patient cable. Replace Adhesive Sensor Message Message:
(Pulse Ox) Replace Adhesive Sensor
(RAM) Replace Adhesive Sensor When a single-patient-use sensor is used, the adhesive portion of the sensor is non-functional, or the life of the adhesive portion of the sensor has expired. (Applies to Rainbow ReSposable CO-Oximeter Sensor Systems or ReSposable Pulse Oximeter Sensor Systems only.) Next steps: Replace the adhesive portion of the sensor. Incompatible Sensor Message Message:
(Pulse Ox) Incompatible Sensor
(RAM) Incompatible Sensor Not a proper Masimo sensor. Next steps: Replace with a proper Masimo sensor. SpHb sensor is attached to a instrument without SpHb installed. Next steps: Use a non-SpHb sensor. Contact your local Masimo Representative to learn more about the optional SpHb upgrade. Incompatible Adhesive Sensor Message Message:
(Pulse Ox) Incompatible Adhesive Sensor
(RAM) Incompatible Adhesive Sensor Not a proper Masimo sensor. Next steps: Replace with a proper Masimo sensor. SpHb sensor is attached to a instrument without SpHb installed. Next steps: Use a non-SpHb sensor. Contact your local Masimo Representative to learn more about the optional SpHb upgrade. www.masimo.com 101 Masimo Radical-7 Chapter 6: Alarms and Messages No Adhesive Sensor Connected Message Message:
(Pulse Ox) No Adhesive Sensor Connected
(RAM) No Adhesive Sensor Connected When a single-patient-use sensor is used, the adhesive portion of the sensor is not connected. (Applies to Rainbow ReSposable CO-Oximeter Sensor Systems or ReSposable Pulse CO-Oximeter Sensor Systems only.) Next steps: Ensure the adhesive portion is firmly connected to the sensor. Interference Detected Message Message:
(Pulse Ox) Interference Detected
(RAM) Interference Detected High intensity light such as pulsating strobe lights, excessive ambient light sources such as surgical lights or direct sunlight, or other monitor displays. Next steps: Place a Masimo Optical Light Shield over the sensor. Incorrect monitor line frequency setting (Hz). Next steps: . Adjust the Line Frequency to the correct Hz setting. See Device Settings on page 74. SpO2 Only Mode Message Message: (Pulse Ox) SpO2 Only Mode Occurs during an unsuccessful sensor initialization/pulse search routine or during monitoring. Next steps: See the directions for use provided with your sensor. Use a Masimo light shield to cover the sensor and adjust the sensor. Low Battery Message Battery charge is low. Next steps: Charge battery by placing the Handheld into the Docking Station and powering the instrument with AC line power. Replace battery if necessary. Low Perfusion Index Message Message: (Pulse Ox) Low Perfusion Index Signal too small. Next steps: Move sensor to better perfused site. See Low Perfusion on page 108. www.masimo.com 102 Masimo Radical-7 Low Signal IQ Message Message: (Pulse Ox) Low Signal IQ Chapter 6: Alarms and Messages Low signal quality. Next steps: Ensure proper sensor application. Move sensor to a better perfused site. See Signal Indication and Quality Indicator (SIQ) on page 107. Low SpCO SIQ Message SpCO measurement reading is obscured. Next steps: Ensure proper sensor application. Check sensor to see if it is working properly. If not, replace the sensor. See Successful Monitoring for SpCO on page 29. Low SpMet SIQ Message SpMet measurement reading is obscured. Next steps: Ensure proper sensor application. Check sensor to see if it is working properly. If not, replace the sensor. See Successful Monitoring for SpCO on page 29. Low SpHb SIQ Message SpHb measurement reading is obscured. Next steps: Ensure proper sensor application. Check sensor to see if it is working properly. If not, replace the sensor. See Successful Monitoring for SpHb on page 28. Speaker Failure Message Instrument requires service. Next steps: Contact Masimo Tech Support. Chapter 9: Service and Maintenance on page 127. No Cable Connected Message Message:
(Pulse Ox) No Cable Connected
(RAM) No Cable Connected Cable not attached or not fully inserted into the connector. Next steps: Disconnect and reconnect cable into connector. www.masimo.com 103 Masimo Radical-7 Chapter 6: Alarms and Messages No Sensor Connected Message Message:
(Pulse Ox) No Sensor Connected
(RAM) No Sensor Connected Sensor not fully inserted into the connector. May be an incorrect sensor or a defective sensor or cable. Next steps: Disconnect and reconnect sensor. See the instructions for use provided with your sensor. Instrument is searching for patients pulse. Next steps: Disconnect and reconnect the sensor into the Patient Cable connector. Sensor is disconnected from patient cable. Sensor connected upside down into patient cable. Next steps: Check to see if the sensor LED is flashing. Disconnect and reconnect the sensor. If the LED fails to operate, replace the sensor. Pulse Search Message Message: (Pulse Ox) Pulse Search Instrument is searching for pulse. Next steps: If instrument fails to display within 30 seconds, disconnect and reconnect. If pulse search continues, move sensor to better perfused site. Sensor Initializing Message Message: (Pulse Ox) Sensor Initializing Instrument is checking the sensor for proper functioning and performance. Next steps: If values are not displayed within 30 seconds, disconnect and reconnect sensor. If values are still not displayed, replace with a new sensor. Sensor Off Patient Message Message:
(Pulse Ox) Sensor Off Patient
(RAM) Sensor Off Patient Sensor off patient. Next steps: Disconnect and reconnect sensor. Reattach sensor. Sensor not connected to patient properly. Sensor is damaged. Next steps: Properly reapply the sensor on the patient and reconnect the sensor to the instrument or patient cable. If the sensor is damaged, replace the sensor. www.masimo.com 104 Masimo Radical-7 Chapter 6: Alarms and Messages Service Required Message Internal Failure. Instrument requires service. The message fills the entire display. This is a numeric error code. Next steps: Contact Masimo for service, see Chapter 9: Service and Maintenance on page 127. Incompatible Cable Message Message: (Pulse Ox) Incompatible Cable Not a proper cable. Next steps: Replace with a proper cable. www.masimo.com 105 Masimo Chapter 7: Troubleshooting The following chapter contains information about troubleshooting the Radical-7 system. Troubleshooting Measurements See Parameter Related Safety Information, Warnings, and Cautions on page 11. Signal Indication and Quality Indicator (SIQ) The Radical-7 Pulse CO-Oximeter display provides a visual indicator of the plethysmogram signal quality and an alert when the displayed SpO2 values are not based on adequate signal quality. The signal quality indicator displayed on the Radical-7 Pulse CO-Oximeter is called the SpO2 SIQ. The SpO2 SIQ can be used to identify the occurrence of a patients pulse and the associated signal quality of the measurement. With motion, the plethysmographic waveform is often distorted and may be obscured by artifact. The SpO2 SIQ, shown as a vertical line, coincides with the peak of an arterial pulsation. Even with a plethysmographic waveform obscured by artifact, the Radical-7 Pulse CO-Oximeter locates the arterial pulsation. The pulse tone (when enabled) coincides with the vertical line of the SpO2 SIQ. The height of the vertical line of the SpO2 SIQ indicates the quality of the measured signal. A high vertical bar indicates that the SpO2 measurement is based on a good quality signal. A small vertical bar indicates that the SpO2 measurement is based on data with low signal quality. When the signal quality is very low the accuracy of the SpO2 measurement may be compromised. See About the Status Bar on page 49. When parameters are dimly lit, proceed with caution and do the following:
Assess the patient. Check the sensor and ensure proper sensor application. The sensor must be well secured to the site for the Radical-7 Pulse CO-Oximeter to maintain accurate readings. Misalignment of the sensors emitter and detector can result in smaller signals and cause erroneous readings. Determine if an extreme change in the patients physiology and blood flow at the monitoring site occurred, (e.g. an inflated blood pressure cuff, a squeezing motion, sampling of an arterial blood specimen from the hand containing the pulse oximetry sensor, severe hypotension, peripheral vasoconstriction in response to hypothermia, medications, or an episode of Raynauds syndrome.) With neonates or infants, check that the peripheral blood flow to the sensor site is not interrupted. Interruption, for example, may occur while lifting or crossing their legs during a diaper change. After performing the above, if the parameter remains dimly lit frequently or continuously, obtaining an arterial blood specimen for CO-Oximetry analysis may be considered to verify the oxygen saturation value. See Parameter Related Safety Information, Warnings, and Cautions on page 11. www.masimo.com 107 Masimo Radical-7 Dimly Lit Parameters Chapter 7: Troubleshooting When the signal quality is very low, the accuracy of measurements may be compromised, the parameter may be dimly lit, and the parameter may display dashes instead of a numeric value. Low Perfusion It has been suggested that at extremely low perfusion levels, pulse oximeters can measure peripheral saturation, which may differ from central arterial saturation This "localized hypoxemia" may result from the metabolic demands of other tissues extracting oxygen proximal to the monitoring site under conditions of sustained peripheral hypoperfusion. This may occur even with a pulse rate that correlates with the ECG heart rate. Low Signal Quality Improper sensor type or application. Next steps: Excessive motion relative to perfusion. Sensor is damaged or not functioning. Check and see if blood flow to the site is restricted. Check the placement of the sensor. Reapply sensor or move to a different site. See Appendix: Best Practices for Comparisons to Reference Measurements on page 137. SpO2 Values Do Not Correlate With Clinical Assessment or Arterial Blood Gas Measurements Low perfusion or sensor displacement. Next steps: Check for error messages. See Chapter 6: Alarms and Messages on page 95. Check placement of sensor or if it is too tight. Reapply sensor or select a new site. Set to MAX sensitivity and confirm that the sensor is securely on the patient. See the directions for use provided with your sensor. Unexpected SpO2, SpCO, SpMet, or SpHb Reading Low SIQ or PI values. Next steps: Reposition sensor to site with strong SIQ and PI. Average readings taken from three different sites to improve accuracy. Submit blood sample for laboratory CO-Oximetry test for comparison. Inappropriate sensor size or sensor measurement location. Next steps:Verify proper sensor for patient size. Verify proper sensor site. See Appendix: Best Practices for Comparisons to Reference Measurements on page 137. www.masimo.com 108 Masimo Radical-7 Chapter 7: Troubleshooting Unexpectedly High SpCO Reading Possible elevated methemoglobin level. Next steps: Submit blood sample for laboratory CO-Oximetry test. See Appendix: Best Practices for Comparisons to Reference Measurements on page 137. Difficulty Obtaining a Reading Low battery/ not plugged into AC power supply. Next steps: Insert Handheld into Docking Station, verify Docking Station power cord plugged in and Docking Station power indicator light is illuminated. Interference from line frequency induced noise. Next steps: Verify/set 50/60hz menu setting. See Localization on page 75. Inappropriate sensor or sensor size. Next steps: Verify proper sensor and sensor size for the patient. Excessive ambient or strobing light. Next steps: Shield the sensor from excessive or strobing light. Minimize or eliminate motion at the monitoring site. See Appendix: Best Practices for Comparisons to Reference Measurements on page 137. SpCO Reading Displays as Dashes SpO2 value below 90% Next steps: Assess/address patient condition. SpMet value greater than 2% Next steps: Laboratory analysis of a blood sample should be performed. SpCO parameter has not yet stabilized during initial startup Next steps: Verify proper sensor and sensor size for the patient. Allow time for parameter reading to stabilize. See Appendix: Best Practices for Comparisons to Reference Measurements on page 137. Troubleshooting the Radical-7 For more information, see Chapter 6: Alarms and Messages on page 95. Instrument Does Not Turn On One or both of the fuses are not operating properly. Next steps: Replace the fuses. For details, see Replacing the Fuses on page 129. Instrument Turns On But Screen is Blank The viewing contrast is not correct. Next steps: Adjust the brightness setting. See Brightness on page 77. If the condition persists, requires service. See. Contacting Masimo on page 133. www.masimo.com 109 Masimo Radical-7 Continuous Speaker Tone Chapter 7: Troubleshooting Internal failure. Next steps: To silence an alarm, press the Alarm Silence button. If alarm continues to sound, turn off the Radical-7. If necessary, remove Handheld battery. Requires service. See Contacting Masimo on page 133. Buttons Do Not Work When Pressed Internal failure. Next steps: Requires service. See Contacting Masimo on page 133. Handheld Battery Does Not Charge AC power cable may be disconnected. Next steps: Restore power to the instrument. Battery Run Time Significantly Reduced Battery memory effects. Next steps: See Battery Operation and Maintenance on page 127. Indicators on Docking Station Continuously Flash Incompatible version of software on Handheld and Docking Station. Next steps: Upgrade to current software versions. Match Handheld to Docking Station with compatible software versions. www.masimo.com 110 Masimo Chapter 8: Specifications The following chapter contains specifications for the Radical-7 Handheld, compatible Docking Stations, and the Standalone system. Performance Measurement SpO2 SpMet SpCO SpHb SpOC Pulse Rate Perfusion Index Pleth Variability Index Range 0 to 100%
0 to 99.9%
0 to 99%
0 to 25 g/dL 0 to 35 ml of O2/dL of blood 25 to 240 (bpm) 0.02% to 20%
0 to 100%
RRa (Respiration Rate) 0 to 70 breaths per minute RRp (Respiration Rate) Accuracy 0 to 70 breaths per minute Oxygen Saturation Accuracy [1]
No Motion 60% to 80%
Adults, Infants, Pediatrics 3%
www.masimo.com 111 Masimo Chapter 8: Specifications Radical-7 No Motion [2]
Adults, Infants, Pediatrics Neonates Motion [3]
70% to 100%
2%
3%
70% to 100%
Adults, Infants, Pediatrics, Neonates 3%
Low Perfusion [4]
Adults, Infants, Pediatrics, Neonates 2%
Pulse Rate Accuracy Pulse rate range No Motion 25 to 240 BPM Adults, infants, pediatrics, neonates 3 BPM Motion [4]
Adults, infants, pediatrics, neonates 5 BPM Low Perfusion Adults, infants, pediatrics, neonates 3 BPM Carboxyhemoglobin Saturation Accuracy [1]
Adults, infants, pediatrics 1% to 40% 3%
Methemoglobin Saturation Accuracy [1]
Adults, infants, pediatrics, neonates 1% to 15% 1%
Total Hemoglobin Accuracy [6]
Adults, pediatrics 8 to 17 g/dL 1 g/dL www.masimo.com 112 Masimo Radical-7 Respiratory Rate Accuracy (RRa) [11]
Chapter 8: Specifications Adults 40 to 70 1 breath per minute Respiratory Rate Accuracy (RRp) [11]
40 to 70 1 breath per minute Step Size 1%
1%
0.1%
0.1 g/dL 1 BPM 1 breath per minute Adults Resolution Parameter
%SpO2
%SpCO
%SpMet SpHb g/dL Pulse Rate (BPM) Respiration Rate Electrical Standalone AC Power requirements 100 to 240 VAC, 47 to 63 Hz Power consumption 55 VA Fuses Handheld Battery Type Capacity www.masimo.com 1 Amp, Fast Acting, Metric, (5x20mm), 250V Lithium ion 4 hours [7]
113 Masimo Radical-7 Time Environmental Operating Temperature Transport/Storage Temperature Chapter 8: Specifications 3 hours 41F to 104F
(5C to 40C)
-40F to 158F
(-40C to +70C) [8]
Operating Humidity 5% to 95%, non-condensing Operating Altitude 500 mbar to 1060 mbar
-1000 ft to 18,000 ft (-304 m to 5,486 m) Physical Characteristics Dimensions Handheld Standalone Weight Handheld 8.9 x 3.5 x 2.1
(22.6 cm x 8.9cm x 5.3 cm) 3.5 x 10.5 x 7.7
(8.9 cm x 26.7 cm x 19.6 cm) 1.2 lbs. (0.54 kg) Docking Station (RDS-1, RDS-2, RDS-3) 2.5 lbs. (1.14 kg) Standalone (RDS-1, RDS-2, RDS-3) 3.8 lbs. (1.73 kg) Trending 72 hours of trending at 2 second resolution
> 10 days of trending at 10 second resolution Output to serial printer or other serial instruments SpO2 Averaging www.masimo.com 2, 4, 8,10, 12, 14, or 16 seconds 114 Masimo Radical-7 Sensitivity RRa Averaging Alarms Parameter SpO2 SpCO SpMet SpHb RRa PI PVI Pulse Rate SpOC Chapter 8: Specifications Normal, Maximum, and APOD [10]
0, 10, 20, 30, 60 seconds Alarm Range 1% to 99%
1% to 98%
0.1% to 99.5%
1.0 g/dL to 24.5 g/dL 4 to 69 breaths per minute 0.03% to 19%
1% to 99%
30 to 235 BPM 1% to 34%
Sensor condition, system failure, and low battery alarms High Priority Medium Priority 571 Hz tone, 5 pulse burst, pulse spacing:
0.250s,0.250s, 0.500s, 0.250s, repeat time:10s 550 Hz tone, 3 pulse burst, pulse spacing:
0.375s, 0.375s, repeat time: 7s Low Priority 500 Hz tone, 1 pulse burst, repeat time: 5s Alarm Muted Reminder 500Hz tone, 2 pulse burst, pulse spacing 0.375s, repeat time: 3min. www.masimo.com 115 Masimo Chapter 8: Specifications High Priority: 70 dB (min) Medium Priority: 70 dB (min) Low Priority: 45 dB (min) Radical-7 Alarm Volume Display Indicators Display Update Rate 1 second Response Time
<20 second delay Type Pixels Dot Pitch Backlit Active Matrix TFT LCD 480 x 272 dots 0.25 mm www.masimo.com 116 Masimo Radical-7 Compliance EMC Compliance Chapter 8: Specifications EN55011: Radiated Emissions (CISPRR 11, 2009 Amendment A1:2010, Class B) EN55011: Conducted Emissions (CISPRR 11, 2009 Amendment A1:2010, Class B) EN301 489-17: Radiated Emissions (EN 301 489-01 V1.8.1:2008, Class B) EN301 489-17: Conducted Emissions (EN 301 489-01 V1.8.1:2008, Class B) EN61000-3-2: Limitations of quasi-stationary current Harmonic Emissions
(Requirements from EN 60601-1-2:2007) EN301 489-17: Harmonic Emissions (EN 301 489-01 V1.8.1:2008) EN61000-3-3: Limitations of voltage fluctuations and Flicker EN301 489-17: Flicker (EN 301 489-01 V1.8.1:2008) EN61000-4-2: ESD: Direct, Indirect, Vertical and Horizontal coupling plane EN301 489-17: ESD: Direct, Indirect, Vertical and Horizontal coupling plane (EN 301 489-01 V1.8.1:2008) EN61000-4-3: Radio Frequency Electromagnetic Field (80MHz to 2500MHz) at 3V/m, 2 Hz sine wave, 3 sec dwell EN301 489-17: Radio Frequency Electromagnetic Field (80MHz to 2500MHz) at 3V/m (EN 301 489-01 V1.8.1:2008) EN61000-4-4: Electric Fast Transient / Burst Immunity EN301 489-17: Electrical Fast Transient / Burst Immunity EN61000-4-5: Surge Immunity EN301 489-17: Surge Immunity EN61000-4-6: Conducted Immunity, Disturbances induced by RF Fields (150 KHz to 80 MHz) EN301 489-17: Conducted Immunity , Disturbances induced by RF Fields EN61000-4-8: Power Frequency Magnetic Field Immunity EN301 489-17: Voltage dips and interruptions EN61000-4-11: AC voltage dips EN61000-4-11: AC voltage interruptions Equipment Classification Type of Protection www.masimo.com IEC 60601-1 / UL 60601-1 EN60601-1: 1990 + A!: 1993 +
A2: 1995 Class 1 (on AC power), Internally powered
(on battery power) 117 Masimo Radical-7 Chapter 8: Specifications Degree of Protection-Patient Cable Type BF-Applied Part Degree of Protection-SatShare Cable Type CF-Applied Part Mode of Operation Output Interface Continuous SatShare (RDS-1). See Serial Interface Specifications on page 118. Serial RS-232 (RDS-1, RDS-3) Nurse Call/Analog Output (RDS-1, RDS-3) VueLink, (Philips, Agilent, HP, Spacelabs Flexport, RadNet, SafetyNet (RDS-1, RDS-3) Wireless Radio (If Installed) Radio Modes Encryption Authentication Compliance USA Canada Europe 802.11 a/b/g 64/128-bit WEP, Dynamic WEP, WPA-TKIP, WPA2-AES Open System, Shared Key, Pre-Shared Key (PSK), 802.1X: LEAP, PEAP< TTLS, TLS, EAP-FAST FCC ID: VKF-RAD7CA, FCC parts 15.207, 15.209, 15.247, and 15.407 IC: 7362A-RAD7CA RSS-210 EN 3000 328, EN 301 893, EN 301 489-17 Serial Interface Specifications The digital interface for serial communication is based on the standard RS-232 protocol. The Radical-7 by default always outputs ASCII 1 text data through the serial port, unless the www.masimo.com 118 Masimo Radical-7 Chapter 8: Specifications user selects a different output mode. To interface with the Radical-7 and receive serial text data, connect a serial interface cable with a ferrite bead installed to the serial output connector located on the back of the Radical-7 Docking Station. The Radical-7 serial interface is only available when the Radical-7 Handheld is properly attached to the Docking Station. Once serial communication is established, packets of data are communicated at 1 second intervals. See Device Settings on page 74. Serial Interface Setup To interface with the Radical-7 serial port, set the following communication parameters on the interfacing serial instrument:
Parameter Baud rate Setting 9600 baud bi-directional Number of bits per character Parity Bits Handshaking Connector type 8 None 1 start, 1 stop None Female DB-9 The pin-outs for the RS-232 connector are shown in the following table:
Pin 1 2 3 4 5 6 7 Signal name No Connection Receive data RS-232 9 V (5 Vmin) Transmit data RS-232 9 V (5 Vmin) No Connection Signal Ground Reference for COM signals No Connection No Connection www.masimo.com 119 Masimo Radical-7 Chapter 8: Specifications Pin 8 9 Signal name No Connection No Connection Analog Output and Nurse Call Specifications Analog Out and Nurse Call are accessible on the same female high-density DB-15 connector. Analog Output and Nurse Call interface are only available when the Handheld is attached to the Docking Station. Only use an Analog and Nurse Call cable that has a ferrite bead installed. Analog Output and Nurse Call interface is not available in all versions of the Docking Station. See Nurse Call Test on page 131 and Handheld Front Panel on page 35. The following table shows the pin out of the Analog Output and Nurse Call. Pin Signal Name 1 2 3 4 5 6 7 8 9 10 11 12 13 14
+5V (60mA max.) Ground Ground Ground Ground Nurse Call (Normally Open) Nurse Call (Normally Closed) Ground Analog 1 Ground Ground Nurse Call -Common Ground Ground www.masimo.com 120 Masimo Radical-7 Pin 15 Analog Output Signal Name Analog 2 Chapter 8: Specifications The Radical-7 Pulse CO-Oximeter can interface with various analog recording instruments or strip chart recorders through its Analog Output connector located on the back of the Docking Station. The output signals vary from approximately 0 to 1 volt in a linear fashion. The actual analog output voltage generated may not exactly range between 0.0V to 1.0V. A variance of 40 mV is acceptable. Calibration For instrument calibration purposes, the analog output signals can be set to either 0 Volts or 1 Volt. Calibrate your analog recording system to those levels before use. Nurse Call The Nurse Call feature is available when the Radical-7 is operating as a standalone. Nurse Call is based on the relay closing or opening depending on alarm, Low Signal IQ events, or both. For maximum flexibility, either normally open (pin 6) or normally closed (pin 7) signals are available. Only qualified personnel should connect one of these two signals and common (pin 12) to a hospitals Nurse Call system. During an alarm condition or a Low Signal IQ event, depending on the configuration, the normally open pin will be connected to the common pin and the normally closed will be disconnected. The Nurse Call polarity can be inverted to accommodate various Nurse Call station requirements. Parameter Max voltage Max Current Specification 100V DC or AC peak 100mA www.masimo.com 121 Masimo Radical-7 Symbols Chapter 8: Specifications The following symbols are found on the Radical-7, Docking Station, or packaging and are defined below. Some of the interfaces and symbols are not available on all versions of the Docking Station. Symbols Definition RS-232 interface SatShare interface Equipotential ground terminal See Instructions for Use Fuse replacement Analog Out interface Nurse Call interface WEEE Compliant Symbols Definition Relative humidity storage range: 5% to 95%
Storage temperature range: +70C to -40C;
Storage altitude range:
+1600hPa to +500hPa Keep dry Fragile/breakable, handle with care Year of Manufacture IPX1 Protection against liquid drops falling vertically Defibrillation proof type BF EU authorized representative www.masimo.com 122 Masimo Radical-7 Chapter 8: Specifications Mark of conformity to European Medical Instrument Directive 93/42/EEC Federal law restricts this device to sale by or on the order of a physician (USA FDA) Underwriter's Laboratories Inc. certification Non-ionizing electromagnetic radiation Caution Name of Manufacturer Wireless features can be used in member states with the restriction of indoor use in France Federal Communications Commission (FCC) licensing IC Model RAD7CA ISO Country Codes (FCC and EU) This equipment may be operated in the following countries:
Country Albania Austria Bangladesh Cambodia Costa Rica Code Country Code Country AL AT BD KH CR Andorra AD Aruba Bahamas Belgium Chile BS BE CL Bahrain Brunei Colombia Croatia HR Cyprus Czech Republic CZ Denmark DK Ecuador El Salvador SV Estonia EE Finland Code AW BH BN CO CY EC FI www.masimo.com 123 Masimo Chapter 8: Specifications France Greece FR Germany GR Guatemala Hong Kong HK Hungary Iraq Jordan Kuwait IQ Israel JO Kazakhstan KW Laos Libya LY Malta Monaco MC Morocco DE GT HU IL KZ LA MT MA Netherlands NL New Caledonia NC Norway NO Oman OM Panama PA Papua New Guinea PG Radical-7 San Marino Ghana Honduras India Italy Kenya Lebanon Mauritius Mozambique Nigeria Palestinian Territory Peru Portugal SM GH HN IN IT KE LB MU MZ NG PS PE PT Philippines PH Poland Puerto Rico PR RE Qatar Russia Republic of Serbia RS Reunion Saudi Arabia Slovakia Spain Trinidad and Tobago SA SK ES TT Senegal SN Singapore Slovenia Sweden Turkey SI SE TR South Africa Thailand Uganda United Kingdom GB Uruguay UY Vietnam PL QA RU SG ZA TH UG VN www.masimo.com 124 Masimo Radical-7 Citations Chapter 8: Specifications
[1] SpO2, SpCO, and SpMet accuracy was determined by testing on healthy adult volunteers in the range 60% to 100% SpO2, 0% to 40% SpCO, and 0% to 15% SpMet against a laboratory CO-Oximeter. SpO2 and SpMet accuracy was determined on 16 neonatal NICU patients ranging in age from 7 days to 135 days old and weighting between 0.5 kgs and 4.25 kgs. Seventy-nine (79) data samples were collected over a range of 70% to 100% SaO2 and 0.5%
to 2.5% HbMet with a resultant accuracy of 2.9% SpO2 and 0.9% SpMet. Contact Masimo for testing specifications.
[2] The Masimo Rainbow SET technology with Masimo sensors has been validated for no motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-Oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population weight.
[3] The Masimo Rainbow SET technology with Masimo sensors has been validated for motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies while performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-Oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
[4] The Radical-7 has been validated for low perfusion accuracy in bench-top testing against a Biotek Index 2TM* simulator and Masimo's simulator with signal strengths of greater than 0.02% and transmission of greater than 5% for saturations ranging from 70-100%. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
[5] Masimo Rainbow SET technology with Masimo sensors has been validated for pulse rate accuracy for the range of 25-240 bpm in bench top testing against a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
[6] SpHb accuracy has been validated on healthy adult male and female volunteers and on surgical patients with light to dark skin pigmentation in the range of 8 g/dL to 17 g/dL SpHb against a laboratory CO-Oximeter. The variation equals plus or minus one standard deviation which encompasses 68% of the population. The SpHb accuracy has not been validated with motion or low perfusion.
[7] This represents approximate run time at the lowest indicator brightness and pulse tone turned off using a fully charged battery.
[8] If the batteries are to be stored for extended periods of time, it is recommended that they be stored between -20C to +30C, and at a relative humidity less than 85%. If stored for a prolonged period at environmental conditions beyond these limits, overall battery capacity may be diminished, and lifetime of the batteries may be shortened.
[9] With FastSat the averaging time is dependent on the input signal. For the 2 and 4 second settings the averaging time may range from 2-4 and 4-6 seconds, respectively.
[10] Maximum sensitivity mode fixes perfusion limit to 0.02%.
[11] Respiration rate accuracy for the Masimo Acoustic Respiration Sensor and Instrument has been validated for the range of 4 to 70 breaths per minute in bench top testing. Clinical validation for up to 30 breaths per minute was also performed with the Masimo Acoustic Respiration Sensor and Instrument.
*Registered trademark of Fluke Biomedical Corporation, Everett, Washington. www.masimo.com 125 Masimo Radical-7 Chapter 8: Specifications www.masimo.com 126 Masimo Chapter 9: Service and Maintenance The following chapter contains information about cleaning, battery operation, performance verification, service, repair, and warranty. Cleaning Under normal operation, no internal adjustment or recalibration is required. Safety tests and internal adjustments should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance with local and governmental regulations. Safety Information, Warnings, and Cautions on page 11. The Radical-7 Pulse CO-Oximeter is a reusable instrument. The instrument is supplied and used non-sterile. To surface clean the Radical-7 The outer surface of the instrument can be cleaned with a soft cloth dampened with a mild detergent and warm water solution. Do not allow liquids to enter the interior of the instrument. The outer surface of the instrument can also be wiped down using any of the following solvents:
Cidex Plus (3.4% glutaraldehyde) 10% bleach solution 70% isopropyl alcohol solution Using the recommended cleaning solutions on the touchscreen panel will not affect the performance of the Handheld. Battery Operation and Maintenance The Radical-7 Handheld includes a lithium ion rechargeable battery. The Radical-7 Docking Station may include the optional 6.5 amp-hour nickel metal hydride rechargeable battery. Before using the Radical-7 as a Handheld or as a transport monitor, the Handheld rechargeable battery and the optional Docking Station rechargeable battery must be fully charged. To charge the Handheld rechargeable battery and the Docking Station rechargeable battery Attach the Handheld to the Docking Station. Connect the Docking Station to AC power. 1. 2. 3. Verify that the batteries are charging. The Docking Station Battery Charging indicator momentarily flashes and then remains illuminated while the batteries are actively charging. If the internal battery temperature exceeds recommended operating conditions for proper battery charging, the Handheld Battery Charging www.masimo.com 127 Masimo Radical-7 Chapter 9: Service and Maintenance indicator continuously flashes. When the temperature returns to recommended operating conditions, proper battery charging proceeds. The Handheld battery requires approximately 4 hours for charging. The Docking Station battery requires approximately 6 hours for charging. When both the Handheld Battery Charging indicator and the Docking Station Battery Charging indicator turn off, additional trickle charging may occur to complete charging. Memory effects of the battery may shorten run-time. When battery run time is significantly reduced, it is advisable to completely discharge and fully recharge the battery. Charging can occur while the Handheld is docked and turned on, the most efficient charge times are achieved when the Handheld is turned off. During battery operation of the Radical-7, note that the following operating conditions affect the estimated run time of the included rechargeable batteries:
Estimated Run Times of Battery Power The following tables outline the estimated run times of the battery-powered Radical-7. The time estimates are based on a Radical-7 with fully charged batteries. The time estimates are also based on a Radical-7 with and without the back-light illuminated. The Radical-7 is always configured to include the Handheld battery. It may optionally be configured to include the Docking Station battery. Determine the configuration of your system before referencing the following tables. Run Time for Handheld Only In this configuration, the Radical-7 is configured to only include the Handheld battery
(standard configuration). When running on battery power, it is advisable to operate only the Handheld. On battery power, it is possible to operate the Standalone (Handheld attached to the Docking Station with the Handheld battery providing power to the Docking Station). However, the capacity of the Handheld battery pack is not sufficient to support this mode for long periods of time. For optimal battery run time, configure the device to automatically adjust the brightness. See Brightness on page 77. Configuration Operation Mode Minimum run time Handheld only Handheld, undocked, not connected to AC power 4 hours Handheld only Handheld docked, not connected to AC power 1 hour www.masimo.com 128 Masimo Radical-7 Replacing the Batteries Chapter 9: Service and Maintenance Before installing or removing the battery, make sure the AC power cord is removed and power to the Radical-7 is turned off. To replace the rechargeable Handheld battery 1. 2. 3. 4. 5. Turn off the Radical-7 Handheld off and remove the patient cable connection. If docked, detach the Handheld from the Docking Station. Loosen the closure screw on the battery compartment door and lift out the battery. Take a new battery and place it in the compartment. Tighten the closure screw. Place the Handheld into Docking Station, turn on line power and charge battery. See Battery Operation and Maintenance on page 127. Replacing the Fuses Should a power problem blow one or both of the fuses in the power entry module on the rear panel, the fuse(s) will need to be replaced. Before you begin, you will need a 5-mm or 3/16-in screwdriver. To replace the fuse(s) 1. 2. Disconnect instrument from AC power. Remove AC power cord from the power entry module at the rear of the Docking Station. 3. Using the screwdriver, gently pry loose the fuse cover in the left portion of the power entry module, exposing the fuse holder. 4. Using the screwdriver, gently remove the fuse holder. 5. Note how the fuse(s) are placed in the fuse holder for installation of the new 6. 7. 8. 9. fuse(s). To remove the fuses from the fuse holder, use the edge of the screwdriver blade to pry against the bottom of the metal portion of the fuse where it is secured to the glass portion of the fuse. Place the fuse(s) (1 amp, metric, fast-acting, 5x20mm, 250V) in the fuse holder, properly orienting the fuse(s). Slide the fuse holder back into the power entry module and press firmly to make sure it is completely seated. Close the fuse cover and press gently until it seats completely, flush with the back of the Docking Station. The instrument is ready to be reconnected to AC power. If the fuses blow shortly after replacement, the instrument requires service. www.masimo.com 129 Masimo Radical-7 Chapter 9: Service and Maintenance Performance Verification Under normal operation, no internal adjustment or recalibration is required. Safety tests and internal adjustments should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance with local and governmental regulations. To test the performance of the Radical-7 following repairs or during routine maintenance, follow the procedure outlined in this chapter. If the Radical-7 fails any of the described tests, discontinue its use and correct the problem before returning the instrument back to the user. Before performing the following tests, do the following:
Place the Handheld into the Docking Station. Connect the Docking Station to AC power and fully charge the Handheld battery. Disconnect any patient cables or pulse oximetry probes. Disconnect any SatShare, serial or analog output cables from the instrument. Set the Radical-7 to Normal operating mode by going to the Main menu and the Home Use feature to No. Power-On Self Test To conduct a Power-On Self Test 1. 2. 3. Connect the monitor to AC power and verify that the AC Power indicator is illuminated. Turn on the monitor. Within 5 seconds, all available indicators illuminate, you hear a tone, and the Masimo logo displays. The Docking Station indicator continuously illuminates and the Radical-7 begins normal operation. Alarm Limit Test To conduct an Alarm Limit Test 1. Change the High SpO2 Alarm parameter to a value two points below the currently selected value. See Alarms for SpO2 on page 61. Verify that the newly set parameter is shown on the Display screen. 2. 3. Return the parameter to its original setting. 4. Repeat steps 1 to 3 for all active parameters. 5. Reset the alarm limits again to the original settings. www.masimo.com 130 Masimo Radical-7 Chapter 9: Service and Maintenance Testing with the optional Masimo SET Tester To conduct a test with the optional Masimo SET Tester Turn off and then turn on the Radical-7. 1. 2. Use the Patient Cable connector on the Radical-7 to connect the Masimo SET Tester to the Radical-7. See the directions for use that were provided with the Masimo SET Tester. 3. Nurse Call Test To conduct a Nurse Call test 1. 2. Disconnect any patient cables, sensors, or accessories from the Radical-7. Turn off the Radical-7 and then turn on again. Ensure that there are no audible alarms and that the Audible Alarm feature is not set to silenced. 3. Verify the Nurse Call polarity is set to normal (default). 4. Connect the common lead of a digital multi-meter to the pin 12 (Nurse Call Common) of the Analog Output connector on the Radical-7. Connect the positive lead of the multi-meter to pin 6 (Nurse Call - Normally Open) of the Analog Output connector and measure that the resistance is greater than 1 MW (open circuit). Trigger an alarm on the monitor (for example, by disconnecting a sensor after it was measuring data). Verify that the resistance is less than 35 ohms. 5. Analog Output Test To conduct an Analog Output test 1. 2. Disconnect any patient cables, sensors, or accessories from the Radical-7. Turn off the Radical-7 and then turn on again. Connect the common lead of a digital voltmeter to the pin 2 (Ground) of the analog output connector on the Radical-7. Connect the positive lead of the voltmeter to pin 9 (Analog 1) of the analog output connector. 3. On the device output screen, on the analog 1 option, select 0V Output. See Device Output on page 79. Verify that the voltmeter measures a voltage of approximately 0V. Change the analog 1 option to 1V Output. 4. 5. 6. Verify that the voltmeter measures a voltage of approximately 1.0V. 7. Repeat steps 5 and 6, with the positive lead of the voltmeter connected to pin 15
(analog 2). See Serial Interface Specifications on page 118. Connect a patient cable and sensor and verify that the voltage on pins 9 and 15 are between 0V and 1.0V while measuring a saturation and pulse rate. 8. www.masimo.com 131 Masimo Radical-7 Battery Test To conduct a Battery test Chapter 9: Service and Maintenance 1. Fully charge the Radical-7 by placing the Handheld into the Docking Station and then connect the AC power. Verify that the Handheld Battery Charging indicator is illuminated. 2. 3. When the Radical-7 is fully charged, the Handheld Battery Charging indicator turns off. Turn on the Radical-7 on and verify that the Battery indicator shows a full charge. 4. Repair Policy Masimo or an authorized Service Department must perform warranty repair and service. Do not use malfunctioning equipment. Have the instrument repaired. Clean contaminated and/or dirty equipment before returning, following the cleaning procedure described in Cleaning. Make sure the equipment is fully dry before packing. To return the instrument for service, see Return Procedure on page 132. Return Procedure Clean contaminated/dirty equipment before returning, following instructions in Cleaning on page 127. Make sure the equipment is fully dry before packing. Call Masimo at 800-326-4890 and ask for Technical Support. Ask for an RMA number. Package the equipment securely, in the original shipping container if possible, and enclose or include the following information and items:
A letter describing in detail any difficulties experienced with the Radical-7. Include the RMA number in the letter. Warranty information, a copy of the invoice or other applicable documentation must be included. Purchase order number to cover repair if the Radical-7 is not under warranty, or for tracking purposes if it is. Ship-to and bill-to information. Person (name, telephone/Telex/fax number, and country) to contact for any questions about the repairs. A certificate stating the Radical-7 has been decontaminated for bloodborne pathogens. Return the Radical-7 to the following shipping address listed in Contacting Masimo on page 133. www.masimo.com 132 Masimo Radical-7 Contacting Masimo To contact Masimo, refer to the following:
Chapter 9: Service and Maintenance USA, Canada, and Asia Pacific Masimo Corporation 40 Parker Irvine, California 92618
(949) 297-7000 Fax: (949) 297-7001 Warranty Europe:
All Other Locations:
Contact your local Masimo Representative. Masimo International Srl Puits-Godet 10 2000 Neuchatel-
Switzerland Tel:+41 32 720 1111 Fax: +41 32 724 1448 Masimo warrants to the initial Purchaser for a period of one (1) year from the date of purchase that: (i) each new Product and the Software media as delivered are free from defects in workmanship or materials, and (ii) the Product and Software will perform substantially as labeled in the directions for use. Masimos sole obligation under this warranty is to repair or replace any Product or Software that is covered under warranty. Batteries are warranted for six (6) months. To request a replacement under warranty, Purchaser must contact Masimo for a returned goods authorization. If Masimo determines that a Product must be replaced under warranty, it will be replaced and the cost of shipment covered. All other shipping costs shall be the responsibility of Purchaser. Exclusions The warranty does not extend to, and Masimo is not responsible for, repair, replacement, or maintenance needed because of: a) modification of the Product or Software without Masimos written authorization; b) supplies, instruments or electrical work external to the Product or not manufactured by Masimo; c) disassembly or reassembly of the Product by anyone other than an authorized Masimo agent; d) use of the Product with sensors or other accessories other than those manufactured and distributed by Masimo; e) use of the Product and Software in ways or in environments for which they are not labeled; and f) neglect, misuse, improper operation, accident, fire, water, vandalism, weather, war, or any act of God. This warranty does not extend to any Product that has been reprocessed, reconditioned or recycled. This warranty also does not apply to any Products provided to Purchaser for testing or demonstration purposes, any temporary Products modules or any Products for which Seller does not otherwise receive a usage or purchase fee; all such Products are provided AS-IS without warranty. This warranty, together with any other express written warranty that may be issued by masimo is the sole and exclusive warranty as to the product and software. this warranty is expressly in lieu of any oral or implied warranties, including without limitation any implied warranty of merchantability or fitness for a particular purpose. masimo shall not be liable for any incidental, special or consequential loss, damage or expense directly or indirectly arising from the use or loss of use of any products or software. in no event shall masimos www.masimo.com 133 Masimo Radical-7 Chapter 9: Service and Maintenance liability arising from any product and software (under contract, warranty, tort, strict liability or other claim) exceed the amount paid by purchaser for the products giving rise to such claim. the limitations in this section shall not be deemed to preclude any liability that cannot legally be disclaimed by contract. Sales & End-User License Agreement This document is a legal agreement between you (purchaser) and Masimo corporation
(Masimo) for the purchase of this product (product) and a license in the included or embedded software (software) except as otherwise expressly agreed in a separate contract for the acquisition of this product, the following terms are the entire agreement between the parties regarding your purchase of this product. if you do not agree to the terms of this agreement, promptly return the entire product, including all accessories, in their original packages, with your sales receipt to masimo for a full refund. 1. Grant of License: In consideration of payment of the Software license fee, which is part of the price paid for the Product, Masimo grants to Purchaser a nonexclusive, nontransferable (except as set forth below) license (License), without right to sublicense, to use the copy of the Software in connection with Purchasers use of the Product for its labeled purpose as set forth in these directions for use. Masimo reserves all rights not expressly granted to Purchaser. 2. Ownership of Software: The Software is licensed not sold; all rights and interests in the Software and all copies thereof remain at all times vested in Masimo, and do not pass to Purchaser. Any references in this Agreement to the purchase or sale of the Software shall be deemed the purchase or sale of a Software License as set forth herein. Restrictions 1. Copyright Restrictions: The Software and the accompanying written materials are copyrighted. Unauthorized copying of the Software, including Software that has been modified, merged, or included with other software, or the written materials is expressly forbidden. Purchaser may be held legally responsible for any copyright infringement that is caused or incurred by Purchasers failure to abide by the terms of this Agreement. Nothing in this License provides any rights beyond those provided by 17 U.S.C. 117. 2. Use Restrictions: Purchaser may physically transfer the Product from one location 3. to another provided that the Software is not copied. Purchaser may not electronically transfer the Software from the Product to any other instrument. Purchaser may not disclose, publish, translate, release, distribute copies of, modify, adapt, translate, reverse engineer, decompile, disassemble, or create derivative works based on the Software or the written materials. Transfer Restrictions: In no event may Purchaser transfer, assign, rent, lease, sell, or otherwise dispose of the Product or the Software on a temporary basis. Purchaser shall not assign or transfer this License, in whole or in part, by operation of law or otherwise without Masimo's prior written consent; except that the Software and all of Purchasers rights hereunder shall transfer automatically to any party that legally acquires title to the Product with which this Software is included. Any attempt to assign any rights, duties or obligations arising hereunder other than as set forth in this paragraph shall be void. 4. U.S. Government Rights: If Purchaser is acquiring Software (including the related documentation) on behalf of any part of the United State Government, the following provisions apply: the Software and documentation are deemed to be www.masimo.com 134 Masimo Radical-7 Chapter 9: Service and Maintenance commercial software and commercial computer software documentation, respectively pursuant to DFAR Section 227.7202 FAR 12.212, as applicable. Any use, modification, reproduction, release, performance, display or disclosure of the Software (including the related documentation) by the U.S. Government or any of its agencies shall be governed solely by the terms of this Agreement and shall be prohibited except to the extent expressly permitted by the terms of this Agreement. www.masimo.com 135 Masimo Appendix: Best Practices for Comparisons to Reference Measurements Best Practices Checklist for Continuous SpHb Comparisons 1. 2. Ensure SpHb device is turned on. If applicable, connect to computer with automatic data capture. Sensor site selection:
Remove anything from patient's arm that can impede blood flow to the sensor site, such as restrictive garments, accessories, purses, backpacks, watches, jewelry, and blood pressure cuff. Do not use sites with any of the following conditions:
An anatomically abnormal finger (e.g. damaged, clubbed, deviated, etc.). A finger or arm that has experienced previous surgical procedures. A finger or arm that is currently receiving an IV infusion. A finger or arm that is currently used for blood pressure cuff. Site should be cleaned of debris and dry prior to testing. Nail polish should be removed prior to testing. Select the patients testing finger in the following priority. Non-dominant ring or middle finger. Sensor selection:
Dominant ring or middle finger. 3. For reusable sensors, measure the patient's finger size (diameter) at the cuticle, using the sensor size gauge, to determine the correct sensor size. When using ReSposable sensors, connect the reusable optical sensor
(ROS) to the disposable optical sensor (DOS) after DOS properly placed onto finger. 4. Ensure proper sensor positioning:
Rest patients hand and arm with sensor on a horizontal surface securely to limit the movement of the patient. Examine the finger while placed in the sensors to ensure the emitter and detector are directly aligned on top of each other and there is no gap between the sensor and fingertip. www.masimo.com 137 Masimo Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements Align upper and lower red lines. If using a reusable sensor, make sure finger tip is inserted all the way and touching the finger stop inside the sensor (allowing long finger nails to extend beyond the finger stop). Ensure cable runs flat over the top of the hand directly in the middle of the finger with no kinks or twists so the cable does not pull on the sensor. Cover sensor with Masimo-provided shielding to avoid light interference. www.masimo.com 138 Masimo Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements Secure sensor cable to patients arm with tape. Instruct subjects to remain still without any sensor movement. Connect sensor cable to device without pulling on sensor. Wait until SpHb measurement is stable (one minute) before recording SpHb values. Note any Low SIQ message that displays during measurement. 5. Blood sampling and laboratory analysis:
Record site of blood draw. Document exact time blood withdrawn from patient. When taking blood sample from arterial line, remove adequate dead space to ensure blood sample is not diluted with solution to keep arterial line open. When IV line set-up includes stopcocks, ensure that no fluid is infused through the line from which you are taking the blood. Draw blood directly into the vacutainer, if possible. Mix blood carefully after filling the vacutainer by rotating it gently at least 10 times. (A rotation is one turn of the tube upside down and then right side up.) All blood samples must be analyzed on the same calibrated laboratory hematology analyzer (Coulter, Sysmex, or equivalent) to avoid variation induced by the use of multiple laboratory devices. www.masimo.com 139 Masimo Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements Best Practices Checklist for SpCO Comparisons 1. 2. 3. Ensure SpCO device is powered on and connected to computer with automatic data capture. After venous blood draw, SpCO testing must be initiated within 30 seconds
(COHb can change rapidly). Sensor site selection:
Remove anything from patient's arm that can impede blood flow to the sensor site, such as restrictive garments, accessories, purses, backpacks, watches, jewelry, and blood pressure cuff. Do not use sites with any of the following conditions:
An anatomically abnormal finger (e.g. damaged, clubbed, deviated, etc.). A finger or arm that has experienced previous surgical procedures. A finger or arm that is currently receiving an IV infusion. A finger or arm that currently used for blood pressure cuff. Site should be cleaned of debris and dry prior to testing. Nail polish should be removed prior to testing. Select the patients testing finger in the following priority. Non-dominant ring or middle finger. 4. 5. Sensor selection:
Dominant ring or middle finger. Use the appropriate sensor as defined in the sensor directions for use. Ensure proper sensor positioning:
Rest patients hand and arm with sensor on a horizontal surface securely to limit the movement of the patient. Examine the finger while placed in the sensors to ensure the emitter and detector are directly aligned on top of each other and there is no gap between the sensor and fingertip. Align upper and lower red lines. If using a reusable sensor, make sure finger tip is inserted all the way and touching the finger stop inside the sensor (allowing long finger nails to extend beyond the finger stop). 140 Masimo www.masimo.com Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements Ensure cable runs flat over the top of the hand directly in the middle of the finger with no kinks or twists so the cable does not pull on the sensor. Cover sensor with Masimo-provided shielding to avoid light interference. Secure sensor cable to patients arm with tape. Instruct subjects to remain still without any sensor movement. Connect sensor cable to device without pulling on sensor. Wait until SpCO measurement is stable (one minute) before recording SpHb values. Note any Low SIQ messages that display during measurement. www.masimo.com 141 Masimo Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements Best Practices Checklist for Acoustic Respiration Rate Comparisons Acoustic Sensor placement The Acoustic sensor has a small black arrow on the front (see figure below), when placing the sensor the black arrow should point forward to the anterior of subjects body. Ensure placement site is hair-free, clean of debris, and dry prior to sensor placement. Use an alcohol swab to clean the neck area, if needed. The preferred sensor placement site is either in quadrant 3 or 4 (center of sensor), see figure below to either side of the larynx, avoid center of the neck. The most critical item is to place the sensor at the site where it can receive the strongest breathing signal. To locate this, place two fingers on the neck while the subject makes a continuous Ahhh sound and choose the location that has the most vibration. For pediatric subjects that have limited neck space, the sensor may be placed on the right side of chest, underneath clavicle. The sensor should not be touching the clavicle. Place sensor tape on skin. Gently press on sensor tape from center outward so adhesive forms a good contact with patient's skin. Ensure there are no skin folds or air gaps under sensor pad. Remove the release liner from the anchor pad and place the anchor pad on patients side of the neck; route the sensor cable in front of patient. Do not place anchor pad on clothing. www.masimo.com 142 Masimo Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements Pulse-Oximeter Sensor Placement Place finger sensor on middle or ring finger of hand opposite the blood pressure cuff. For reusable sensors, make sure the fingertip just touches the rubber stopper at the end of the sensor without going over it. Ensure sensor is right side up with the cable running in line with finger, not at an angle on dorsal (back side) of hand. Monitoring Ensure RRa device is turned on and connected to computer with automatic data capture. Connect sensor to cable. If RRa values are not displayed after 2 minutes or if the RRa value has dropped out, check the following:
Confirm appropriate sensor placement, orientation and site selections. Confirm that optical pulse-oximeter sensor is placed properly on the patient's finger. Confirm that all cables are plugged in at each of the various connection points and hubs. Auscultate with stethoscope to listen for air sounds on the side opposite sensor. If breath sounds are present, remove sensor and replace with new sensor on opposite side of neck. Change the sensor out if RRa value continues to not display. Verify that there is not excessive hair or a gap between the sensor and the neck and that it is placed. Simultaneously record the RRa and respiratory rates and from other methods. If comparing RRa to capnography respiration rate, a mask is recommended. Sidestream methods with a nasal cannula are not recommended because of dilution effect in the supplemental flow of gasses, inability to measure both nasal and oral airflow, and nasal cannula mispositioning. When recording values, confirm that there are no SIQ messages displayed on the device. Suggested directions to record manual respiration rate are as follows:
Use stethoscope to listen for breath sounds, count each breath cycle as one breathe, count for 60 seconds. Alternate method to stethoscope, count the number of chest rises/inhalations during a 60 second period. Record manual respiration rate to compare with RRa. Adjust respiratory pause settings as necessary, default is 30 seconds (options are: 15, 20, 25, 30, 35, or 40 seconds). Document patient events and time of event (based on computer clock). Include events that may affect performance, these include:
Patient talking, picking at the sensor or nasal cannula, excessive movement, ambient noise present, fans or air blowing at sensor Inaccurate measurements may be caused by:
www.masimo.com 143 Masimo Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements Excessive ambient or environmental noise (patient speaking, room noise) Improper sensor placement Cable disconnection Movement, picking, or air blowing at sensor www.masimo.com 144 Masimo Index 3 3D Alarms 56, 98 A About 57, 87 About Alarms 95, 97 About Patient Types 90, 92 About Profiles and Default Settings 89 About the Display View 45, 48 About the Status Bar 48, 49, 76, 77, 107 About this Manual 7 About Trend Views 81 Access Control 46, 74, 77 Accessing Parameter Information 59, 61, 62, 64, 65, 66, 69, 70, 71, 72, 98, 99 Accessing the Main Menu 47, 49, 55 Accuracy 111 Acquisition System 31 Adaptive Threshold Alarm (ATA) Feature 61, 97 Additional Settings for PI 65 Additional Settings for PVI 69, 70 Additional Settings for RRa 66, 67 Additional Settings for RRp 67, 69 Additional Settings for SpHb 62, 63 Additional Settings for SpO2 61, 62 Alarm Limit Test 130 Alarm Related Safety Information, Warnings, and Cautions 20, 95 Alarms 115 Alarms for SpO2 61, 97, 130 Analog Output and Nurse Call Specifications 120 Analog Output Test 131 Appendix www.masimo.com Best Practices for Comparisons to Reference Measurements 108, 109, 137 B Battery 49, 74, 77 Battery Operation and Maintenance 36, 42, 44, 77, 110, 127, 129 Battery Run Time Significantly Reduced 110 Battery Test 132 Best Practices Checklist for Acoustic Respiration Rate Comparisons 142 Best Practices Checklist for Continuous SpHb Comparisons 137 Best Practices Checklist for SpCO Comparisons 140 Brightness 74, 77, 109, 128 Buttons Do Not Work When Pressed 110 C Changing Between Trend Views 82 Changing Profiles Settings 50, 56, 90 Changing the Size of Parameter Values 35, 48, 51 Changing Trend Settings 85 Chapter 1 Technology Overview 23 Chapter 2 Radical-7 Descriptions 33 Chapter 3 Setup 41 Chapter 4 Operation 45 Chapter 5 Configurations 35, 45, 89 Chapter 6 Alarms and Messages 95, 108, 109 Chapter 7 Troubleshooting 95, 107 Chapter 8 145 Masimo Radical-7 Specifications 111 Chapter 9 Service and Maintenance 103, 105, 127 Citations 125 Cleaning 127, 132 Compliance 117 Contacting Masimo 109, 110, 132, 133 Continuous Speaker Tone 110 Contraindications 10 D Desat Index Alarms 98, 99 Desat Index Settings 98 Device Output 44, 74, 79, 131 Device Related Safety Information, Warnings, and Cautions 15, 43, 44 Device Settings 42, 56, 74, 102, 119 Difficulty Obtaining a Reading 109 Dimly Lit Parameters 108 Display Indicators 116 Docking Station Power Requirements 41 E Electrical 113 Electrical Safety Information, Warnings, and Cautions 18, 42 Environmental 114 Index General Description for Methemoglobin
(SpMet) 29 General Description for Oxygen Saturation (SpO2) 24 General Description for Perfusion Index
(PI) 25 General Description for Pleth Variability Index (PVI) 25 General Description for Pulse Rate
(BPM) 25 General Description for SpOC 29 General Description for Total Arterial Oxygen Content (CaO2) 28 General Description for Total Hemoglobin (SpHb) 28 General System Description 33 H Handheld 35 Handheld Back Panel 36 Handheld Battery Does Not Charge 110 Handheld Front Panel 35, 120 I In Vivo for SpCO 60, 71 In Vivo for SpHb 60, 64 In Vivo for SpMet 60, 72 In Vivo for SpO2 60, 62 Incompatible Adhesive Sensor Message Estimated Run Times of Battery Power 101 128 Exclusions 133 F Functional Oxygen Saturation 25 Functionality of the Radical-7 34 G General Description for Carboxyhemoglobin (SpCO) 29 www.masimo.com Incompatible Cable Message 105 Incompatible Sensor Message 101 Indications for Use 10 Indicators on Docking Station Continuously Flash 110 Initial Battery Charging 42 Instrument Does Not Turn On 109 Instrument Turns On But Screen is Blank 109 Interference Detected Message 102 ISO Country Codes (FCC and EU) 123 146 Masimo Radical-7 K Key Features 9 L Localization 49, 74, 75, 109 Low Battery Message 102 Low Perfusion 102, 108 Low Perfusion Index Message 102 Low Signal IQ Message 103 Low Signal Quality 108 Low SpCO SIQ Message 103 Low SpHb SIQ Message 103 Low SpMet SIQ Message 103 M Manipulating Trend Data 82 Masimo rainbow SET Parallel Engines 23 Masimo SET DST 24 Messages 49, 100 Monitor Interface 40 N Navigating the Main Menu 56 Navigating the Radical-7 47 No Adhesive Sensor Connected Message 102 No Cable Connected Message 103 No Sensor Connected Message 104 Nurse Call Test 120, 131 O Output Interface 118 Overview of Desat Index Alarm 98, 99 Overview of FastSat 62 Overview of In Vivo Adjustment 60, 62, 64, 71 Overview of Perfusion Index (PI) Delta Alarm 98, 99, 100 www.masimo.com Index Overview of RRp 68 Overview of Sensitivity Modes 49, 50, 90 P Parameter Quick Trend View 55 Parameter Related Safety Information, Warnings, and Cautions 11, 107 Parameter Settings 56, 58 Patient 31 Performance 111 Performance Verification 130 Perfusion Index (PI) Settings 65 Physical Characteristics 114 PI Alarms 65 PI Delta Alarms 100 PI Delta Settings 99 Pleth + Sig IQ View 54 Pleth + Siq IQ + Acoustics View 53 Powering Off the Radical-7 93 Power-On Self Test 130 PR Alarms 64 PR Settings 64 Product Description 9 Product Description, Indications for Use, Contraindications, and Features 9 Pulse CO-Oximetry vs. Drawn Whole Blood Measurements 27 Pulse Search Message 104 PVI Alarms 69 PVI Settings 58, 69 R rainbow Acoustic Monitoring (RAM) Technology 30 rainbow Acoustic Monitoring Architecture 30 rainbow Pulse CO-Oximetry Technology 26 Repair Policy 132 147 Masimo Radical-7 Index Replace Adhesive Sensor Message 101 Silencing the Alarms 35, 45, 95 Replace Cable Message 101 Replace Sensor Message 100 Replacing the Batteries 129 Replacing the Fuses 109, 129 Resolution 113 Restrictions 134 Return Procedure 41, 132 RRa Alarms 66 RRa Settings 58, 66 RRa Waveform View 54 RRp Alarms 67, 68 RRp Settings 58, 67 Run Time for Handheld Only 128 S Safety Information, Warnings, and Cautions 11, 28, 29, 30, 41, 95, 127 Sales & End-User License Agreement 134 Screen Orientation 74, 75 Sensor 31 Sensor Initializing Message 104 Sensor Off Patient Message 104 Sensor Related Safety Information, Warnings, and Cautions 21 Serial Interface Setup 119 Serial Interface Specifications 39, 43, 44, 118, 131 Service Required Message 105 Setting Up and Using SatShare 40, 44 Setting Up for Philips, Agilent, or HP VueLink 42 Setting Up for SpaceLabs Flexport 43 Setting Up the Docking Station 42 Signal Extraction Technology (SET) 23 Signal Indication and Quality Indicator
(SIQ) 103, 107 Signal Processing 32 Signal Quality (Sig IQ) Indicators 53 www.masimo.com Sounds 49, 57, 73 SpCO Alarms 70 SpCO Reading Displays as Dashes 109 SpCO Settings 58, 70 SpCO, SpMet, and SpHb Measurements During Patient Motion 30 Speaker Failure Message 103 SpHb Alarms 62, 63 SpHb Settings 58, 62 SpMet Alarms 71 SpMet Settings 58, 71 SpO2 Only Mode Message 102 SpO2 Settings 58, 61 SpO2 Values Do Not Correlate With Clinical Assessment or Arterial Blood Gas Measurements 108 SpOC Alarms 58, 72 SpOC Settings 72 Standalone 37 Standalone Back Panel 39, 44 Standalone Front Panel 38, 42 Successful Monitoring for SpCO 29, 103 Successful Monitoring for SpHb 28, 103 Successful Monitoring for SpMet 30 Successful Monitoring for SpO2, PR, and PI 24 Symbols 122 T Testing with the optional Masimo SET Tester 131 Trend Field 51, 57 Trending 114 Trends 51, 57, 81 Troubleshooting Measurements 28, 107 Troubleshooting the Radical-7 109 148 Masimo Index Radical-7 U Unexpected SpO2, SpCO, SpMet, or SpHb Reading 108 Unexpectedly High SpCO Reading 109 Unpacking and Inspection 41 Using Screenlock 46 Using the Home Button 46 Using the Touchscreen and Buttons 45 W Warranty 133 Waveform and Trend Views 48, 52 WiFi 49, 74, 76 Wireless Radio (If Installed) 118 www.masimo.com 149 Masimo www.masimo.com 35196/LAB-5476B-0512
frequency | equipment class | purpose | ||
---|---|---|---|---|
1 | 2012-05-24 | 5180 ~ 5240 | NII - Unlicensed National Information Infrastructure TX | Original Equipment |
2 | 5745 ~ 5825 | DTS - Digital Transmission System |
app s | Applicant Information | |||||
---|---|---|---|---|---|---|
1 2 | Effective |
2012-05-24
|
||||
1 2 | Applicant's complete, legal business name |
Masimo Corporation
|
||||
1 2 | FCC Registration Number (FRN) |
0016800161
|
||||
1 2 | Physical Address |
52 Discovery
|
||||
1 2 |
Irvine, California 92618
|
|||||
1 2 |
United States
|
|||||
app s | TCB Information | |||||
1 2 | TCB Application Email Address |
d******@nwemc.com
|
||||
1 2 | TCB Scope |
A4: UNII devices & low power transmitters using spread spectrum techniques
|
||||
app s | FCC ID | |||||
1 2 | Grantee Code |
VKF
|
||||
1 2 | Equipment Product Code |
RAD7CA
|
||||
app s | Person at the applicant's address to receive grant or for contact | |||||
1 2 | Name |
A**** J****
|
||||
1 2 | Title |
EMC Manager
|
||||
1 2 | Telephone Number |
949-2********
|
||||
1 2 | Fax Number |
949-2********
|
||||
1 2 |
a******@masimo.com
|
|||||
app s | Technical Contact | |||||
n/a | ||||||
app s | Non Technical Contact | |||||
n/a | ||||||
app s | Confidentiality (long or short term) | |||||
1 2 | Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | Yes | ||||
1 2 | Long-Term Confidentiality Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | No | ||||
if no date is supplied, the release date will be set to 45 calendar days past the date of grant. | ||||||
app s | Cognitive Radio & Software Defined Radio, Class, etc | |||||
1 2 | Is this application for software defined/cognitive radio authorization? | No | ||||
1 2 | Equipment Class | NII - Unlicensed National Information Infrastructure TX | ||||
1 2 | DTS - Digital Transmission System | |||||
1 2 | Description of product as it is marketed: (NOTE: This text will appear below the equipment class on the grant) | RAD7CA | ||||
1 2 | Related OET KnowledgeDataBase Inquiry: Is there a KDB inquiry associated with this application? | No | ||||
1 2 | Modular Equipment Type | Does not apply | ||||
1 2 | Purpose / Application is for | Original Equipment | ||||
1 2 | Composite Equipment: Is the equipment in this application a composite device subject to an additional equipment authorization? | Yes | ||||
1 2 | Related Equipment: Is the equipment in this application part of a system that operates with, or is marketed with, another device that requires an equipment authorization? | No | ||||
1 2 | Grant Comments | Output power is conducted. Operation of this device in the 5150 MHz to 5250 MHz range is restricted to indoor use only. SAR compliance for body-worn operating conditions is limited to the specific configurations tested for this filing. End users must be informed of the body worn requirements for satisfying RF Exposure compliance. The highest reported SAR values are: Body-worn: 0.511 W/kg | ||||
1 2 | Is there an equipment authorization waiver associated with this application? | No | ||||
1 2 | If there is an equipment authorization waiver associated with this application, has the associated waiver been approved and all information uploaded? | No | ||||
app s | Test Firm Name and Contact Information | |||||
1 2 | Firm Name |
Northwest EMC, Inc.
|
||||
1 2 | Name |
G**** K******
|
||||
1 2 | Telephone Number |
503-8******** Extension:
|
||||
1 2 | Fax Number |
503-8********
|
||||
1 2 |
g******@nwemc.com
|
|||||
Equipment Specifications | |||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Line | Rule Parts | Grant Notes | Lower Frequency | Upper Frequency | Power Output | Tolerance | Emission Designator | Microprocessor Number | |||||||||||||||||||||||||||||||||
1 | 1 | 15E | CC | 5180.00000000 | 5240.00000000 | 0.0157000 | |||||||||||||||||||||||||||||||||||
Line | Rule Parts | Grant Notes | Lower Frequency | Upper Frequency | Power Output | Tolerance | Emission Designator | Microprocessor Number | |||||||||||||||||||||||||||||||||
2 | 1 | 15C | CC | 2412 | 2462 | 0.0561 | |||||||||||||||||||||||||||||||||||
2 | 2 | 15C | CC | 5745 | 5825 | 0.0062 |
some individual PII (Personally Identifiable Information) available on the public forms may be redacted, original source may include additional details
This product uses the FCC Data API but is not endorsed or certified by the FCC