all | frequencies |
|
|
|
|
|
exhibits | applications |
---|---|---|---|---|---|---|---|---|
manual | photos | labels |
app s | submitted / available | |||||||
---|---|---|---|---|---|---|---|---|
1 2 |
|
User Manual | Users Manual | 4.40 MiB | August 27 2021 | |||
1 2 |
|
int. Photos | Internal Photos | 1.34 MiB | August 27 2021 | |||
1 2 |
|
ext. Photos | External Photos | 1.16 MiB | August 27 2021 | |||
1 2 |
|
Label | ID Label/Location Info | 102.04 KiB | August 27 2021 | |||
1 2 |
|
LabelLoc | ID Label/Location Info | JPEG Image | 2.02 MiB | August 27 2021 | ||
1 2 |
|
ConReq | Cover Letter(s) | 39.76 KiB | August 27 2021 | |||
1 2 |
|
New ID Statement module | Cover Letter(s) | 51.14 KiB | August 27 2021 | |||
1 2 |
|
Power of Attorney Letter | Cover Letter(s) | 33.35 KiB | August 27 2021 | |||
1 2 |
|
RFExp G0M-2103-9714-TFC093PE-V02 | RF Exposure Info | 981.93 KiB | August 27 2021 | |||
1 2 |
|
Test Setup Photos | Test Setup Photos | 1.23 MiB | August 27 2021 | |||
1 2 |
|
TestRpt 20CH-00664.R02 Medela RFID FCC 2 | Test Report | 2.90 MiB | August 27 2021 | |||
1 2 |
|
TestRpt FCC-Testreport-NINA-B3(F181014E7) | Test Report | 1.56 MiB | August 27 2021 |
1 2 | User Manual | Users Manual | 4.40 MiB | August 27 2021 |
Invia® Ease
Negative Pressure Wound Therapy system
EN Clinician Instructions for Use
For Clinician Use Only
Please read all instructions before using this product.
Keep these instructions for future reference.
– Misuse or failure to follow the instructions in this manual may result in serious or fatal injury
to the patient.
– Clinician Instructions for Use are intended for healthcare professionals only.
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a physician.
Medela Customer Service phone: 1-800-435-8316
Table of contents
N
E
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2
Indications for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3
4 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
5 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Invia Ease NPWT system and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
6
7
Initial pump setup: first time the pump is turned on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
8 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
9 Access the Clinician menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
10 Change therapy settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
11 Change pump settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
12 View therapy history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
13 End therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
14 Alarms and troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
15 Accessories setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
16 Safety-related checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
17 Maintenance and service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
18 Sterility requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
19 Cleaning, disinfection and storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
20 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 51
21 International Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
22 Technical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
23 Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 58
24 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
25 Service life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
26 Meaning of symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
27 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
1 Introduction
The Invia Ease NPWT system is designed to provide Negative Pressure Wound Therapy
(NPWT), to help promote wound healing. The Invia Ease system includes the Invia Ease
pump, an Invia Ease canister and tubing in a range of sizes (300 ml, 500 ml and 1000 ml) and
an Invia Ease charger.
Invia Ease NPWT system is intended to be used in conjunction with the Invia dressings only.
2 Intended use
2.1 Intended user
– The Invia Ease NPWT system is intended to be used by clinicians or by adequately trained
– Medical situations must be addressed by a physician. Therapy changes should only be
lay users (patients or caregivers).
done as prescribed by a physician.
– To ensure safe use of the Invia Ease NPWT system, clinicians are responsible to train lay
users (patients and caregivers) according to the Patient Instructions for Use and explain
all related safety information.
2.2 Intended patient population
– The Invia Ease NPWT system is intended to be used only on patients exhibiting conditions
as described in the Indications for Use.
– Appropriate patient selection according to Indications for Use is important to prevent
complications and help ensure the efficacy and successful treatment with NPWT.
– The Invia Ease NPWT system has not been studied on pediatric patients.
– The lay user (patient or caregiver) must have the visual and hearing acuity to appropriately:
– respond to notifications displayed by the pump
– understand directions from a clinician
– follow the instructions provided in the Patient Instructions for Use
2.3 Intended use environment
The Invia Ease NPWT system is intended for use in acute, extended and home care settings.
4 | Introduction/Intended use
3 Indications for use
N
E
The Invia Ease Negative Pressure Wound Therapy (NPWT) system is indicated for patients
who would benefit from a suction device (Negative Pressure Wound Therapy) as when used
on open wounds it creates an environment that promotes wound healing by secondary or
tertiary (delayed primary) intention by preparing the wound bed for closure, reducing
edema, promoting granulation tissue formation and perfusion, and by removing exudates
and infectious material.
When used on closed surgical incisions, the Invia Ease NPWT system is also intended to
manage the environment of surgical incisions that continue to drain following sutured or
stapled closure by maintaining a closed environment and removing exudate via the
application of Negative Pressure Wound Therapy.
The Invia Ease NPWT system is appropriate for use for the following indications:
– Acute or subacute wounds
– Chronic wounds
– Dehisced wounds
– Pressure ulcers
– Diabetic/Neuropathic ulcers
– Venous insufficiency ulcers
– Traumatic wounds
– Partial thickness burns
– Flaps and grafts
– Closed surgical incisions
4 Contraindications
The Invia Ease NPWT system is contraindicated in the presence of:
– Necrotic tissue with eschar present
– Untreated osteomyelitis
– Non-enteric and unexplored fistulas
– Malignancy in the wound (with exception of palliative care to enhance quality of life)
– Exposed vasculature
– Exposed nerves
– Exposed anastomotic site of blood vessels or bypasses
– Exposed organs
Indications for use/Contraindications | 5
5 Safety information
Indicates a potentially hazardous situation that, if not avoided, could result in death
or serious injury.
Indicates a potentially hazardous situation that, if not avoided, could result in minor
or moderate injury.
WARNING
CAUTION
NOTICE
Can lead to material damage (not related to personal injury).
i
INFORMATION
Useful or important information that is not related to safety.
When using electrical products, especially when children are present, basic safety
precautions must always be followed.
WARNING
5.1.
1. Therapy interruption: The recommended application of the therapy with Invia Ease
NPWT system is 24 hours without interruption. If therapy is interrupted for more than
2 consecutive hours (120 minutes), the dressing should be changed and therapy
restarted by a clinician in order to avoid the risk of infection or sepsis.
2. Bleeding: Patient with an increased risk of bleeding complications must be closely
monitored and should be treated in the appropriate care settings (e.g. hospital).
If sudden or increased bleeding is observed, immediately stop the use of the pump,
apply pressure on wound dressing and seek immediate Emergency Medical Attention.
Patients with an increased risk for bleeding include:
– Patients who have been administered anticoagulants or platelet aggregation
inhibitors
– Patients with inadequate wound hemostasis
– Patients who are more likely to experience vascular anastomosis
– Patients with friable vessels, infected blood vessels or organs in or around the wound
as a result of but not limited to suturing of the blood vessels or organs, infection,
trauma, radiation.
3. Canister size: Do NOT use the 1000 ml canister on patients with an increased risk of
bleeding complications or patients, who should not be permitted to lose a large
amount of fluid. Assess patient’s physical condition and wound type prior to choosing
the canister. The 1000 ml canister is recommended for acute care (hospital) setting only.
4. Patient monitoring:
– A patient undergoing NPWT therapy requires frequent supervision. Signs of possible
infection or any complication must be addressed immediately.
6 | Safety information
N
E
– Small adults and elderly patients as well as patients with highly exudating wounds
should be closely monitored for fluid loss and dehydration. When needed, fluid
output in the canister and tubing should be compensated accordingly to avoid excess
fluid loss and dehydration.
– Monitor the patient status and patient comfort level, the exudate collected in the
canister, the wound, and the surrounding skin frequently to ensure efficient and safe
treatment as well as patient comfort. Non-observance can cause pain, injury or lead
to considerable danger to the patient
– Y-Connector: The use of NPWT with single-lumen drain or with the curved arm
of the Invia Y-Connector may increase the risk of exudate accumulation, infection,
maceration, or loss of NPWT while blockage occurs in the vacuum system.
These conditions may only be detected by frequent monitoring and may require
more frequent dressing changes.
5. Wound assessment: Observe wound/periwound tissue and exudate for signs of
infection or other complications. Most common signs of infection include redness,
tenderness, swelling, pain, itching, increased warmth in the wound area, strong odor
or purulent discharge. Additional symptoms to be noted in patient include nausea,
vomiting, diarrhea, headache, dizziness, fainting, sore throat with swelling of the
mucous membranes, disorientation, high fever (>102 °F, >38.8 °C), refractory hypo-
tension, orthostatic hypotension, or erythroedema (a sunburn like rash).
More serious complications include infection gangrene, toxic or septic shock.
If more serious complications of infection occur, discontinue therapy and consult a
physician immediately.
6. Osteomyelitis: Do NOT use Invia Ease NPWT system on wounds with untreated
osteomyelitis. In the presence of osteomyelitis, thoroughly debride all necrotic or
infected bone tissue and start systemic antibiotic therapy prior to initiating NPWT.
7. Autonomic hyperreflexia/spinal cord injury: Special care is required when NPWT is
used near vagus nerve (bradycardia) or on patients with a history of spinal cord injury
(stimulation of sympathetic nervous system). Should a patient with a spinal cord injury
experience autonomic hyperreflexia in response to stimulation of the sympathetic
nervous system (characterized by a sudden elevation in blood pressure and one of the
following signs/symptoms: bradycardia, headache, flushing, and/or profuse sweating
above the lesion level), discontinue treatment with the Invia Ease NPWT system and
consult a physician immediately.
8. Defibrillation: In the event that defibrillation is required, clamp the tubings and
disconnect the Invia Ease pump from the wound dressing before the patient is
defibrillated.
9. Hyperbaric Oxygen Therapy (HBOT): Do not take the pump into the hyperbaric
oxygen chamber (HBO). When going into the hyperbaric oxygen chamber, clamp the
canister tubing and disconnect the Invia Ease pump prior to patient entering the
hyperbaric oxygen chamber. Note: Do not clamp the dressing tubing and protect the
end of the dressing tubing at the Quick-Connector.
10. Magnetic Resonance Imaging (MRI): The Invia Ease pump is Magnetic Resonance
unsafe. Do not take the Invia Ease pump into MRI environment. When going into MR
environment, clamp the canister tubing and disconnect the Invia Ease pump prior to
patient entering the MR environment. Note: Do not clamp the dressing tubing and
protect the end of the dressing tubing at the Quick-Connector.
11. Risk of fire, electric shock, or serious burns:
– Due to the fire hazard the Invia Ease pump is not for use in potentially explosive
environments including oxygen enriched environments and in areas of flammable
anesthetics.
– Only use the charger that comes with the pump. An inadequate power source may
result in a fire hazard, electric shock or malfunction of the pump.
Safety information | 7
– Prior to charging, always inspect the charger and the cord for damage. If damage
is found, immediately disconnect the charger and call the Medela Customer Service
or the system provider.
– Never operate the pump if it is not working properly, if it has been damaged
or dropped into water.
– Do not take the pump into a bathtub or shower.
– Never place or drop the pump into water or other liquids.
– If the pump has been exposed to water or other liquids, do not touch it. Disconnect
the pump from electrical outlet port, turn the pump off and contact the manufacturer.
– This pump contains lithium-ion batteries, which bears risk of fire, explosion and
burns.
– Do not disassemble, crash, or heat the Invia Ease pump above 212 °F (100 °C ),
do not incinerate or dispose of in fire.
12. System modification: Do not modify the Invia Ease pump or any of the system
components without authorization from the manufacturer.
13. Magnet: The Invia Ease pump contains a magnet, which may interfere with other
medical devices. Keep a distance of 00 xx between the pump and other medical
devices
14. Sterile products: Do not use the sterile products if the sterile packaging is damaged,
was opened prior to use or has expired.
15. Dressing application:
– For detailed dressing application information, indications, contraindications,
warnings and precautions consult the appropriate Invia dressing Instructions for Use.
– Wound dressing to be applied and changed by clinicians only.
– Perform a thorough wound cleansing per physician’s orders or facility protocol prior
to performing a dressing application.
– Always count the number of dressing filler pieces used in the wound and document
that number on the Transparent Film and in the patient’s chart. Make sure all pieces
are removed when a dressing is changed.
– Document the dressing change date on the Transparent Film and in the patient’s
chart.
– Sharp edges/bone fragments must be eliminated from the wound area or covered
to prevent puncturing blood vessels or organs.
– Do not place dressing filler into blind and/or unexplored tunnels.
– Do not force dressing fillers into any area of the wound, as this may damage tissue,
alter the delivery of negative pressure, or hinder exudate and dressing filler removal.
– Do not place the dressing filler directly on exposed blood vessels, organs, nerves,
tendons, bones or ligaments. When using the Invia Ease NPWT system in close
proximity to these structures, a protective barrier such as a non-adherent wound
contact layer must be used.
– Avoid circumferential dressing application.
– Consider the use of a protective barrier on skin that may come in contact with the
tubing, especially in patients with fragile skin.
– When treating infected wounds or wounds more susceptible to tissue in-growth into
wound filler material, more frequent dressing changes or a non-adherent wound
contact layer may be needed. The frequency of dressing changes should be based
on an evaluation of the wound characteristics rather than standard recommendations.
– For closed surgical incision the recommended pressure level is –125 mmHg at
constant mode.
8 | Safety information
CAUTION
5.2
1. Incorrect use of the Invia Ease NPWT system may cause pain or injury to the patient.
2. Routinely check that therapy is running and that the negative pressure level and therapy
N
E
mode are at the prescribed setting.
3. The patient should be monitored regularly according to the physician instructions and
facility guidelines. Monitor patient comfort, signs of wound infection, integrity of the
dressing, volume of fluid in the canister and therapy compliance.
4. Wounds that involve a fistula may require special care.
5. Excessive negative pressure, a too tight adhesive film dressing, or an infection of the
wound may cause pain to the patient. In each case, the dressing must be removed and
the wound assessed.
6. Constant versus intermittent pressure: use continuous pressure over unstable structures
(such as unstable chest wall or non-intact fascia) in order to help minimize movement
and stabilize the wound bed. Continuous pressure is also recommended for patients
with increased risk of bleeding, flaps and grafts, highly exudating wounds and when
used on closed surgical incisions.
7. Infection control: To reduce the risk of cross contamination, apply universal precautions
for infection control per institutional protocol. Use gloves and other protective equip-
ment if exposure to body fluids is likely.
8. Protect the canister tubing from coming in direct contact with the wound area.
Safety information | 9
5.3 Safety measures
1. Only use products specified in these Instructions for Use.
2. Do not modify the Invia Ease pump or any of the system components without authoriza-
tion from the manufacturer.
3. The Invia Ease pump is verified within the scope of conformity evaluation and is only to be
used with products included in the Invia Ease NPWT system and distributed by Medela.
Medela can only guarantee the effective performance of the system with these products.
4. Before you plug in the Invia Ease pump, please check that your local power supply is the
same as the voltage given on the specification type plate located at the base of the pump.
5. Wireless communication equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, and walkie-talkies can interfere with
the Invia Ease pump and should be kept at a minimum distance of 1 foot (30 cm) away
from the Invia Ease pump.
6. Supervision is necessary when the Invia Ease pump is used in the vicinity of children.
7. Do not use the Invia Ease pump if:
– The power cord or charger are damaged
– The pump is damaged
– The pump is not functioning properly
– The pump has apparent safety defects
8. Do not use the Invia Ease products if they are damaged.
9. Never pull the plug out of the main socket by pulling on the connecting cord.
10. Keep the Invia Ease pump and the associated products away from hot surfaces.
11. Never place the Invia Ease pump or charger in water or other liquids and keep the
charger connector away from moisture or immersion in water.
12. The Invia Ease pump must not be used for suctioning explosive, easily flammable or
corrosive liquids.
5.4 Physician’s orders
A physician must assess each wound when applying this therapy and determine
an appropriate negative pressure setting according to the wound characteristics.
Changes to the therapy should only be done as prescribed by a physician.
10 | Safety information
6 Invia Ease NPWT system and accessories
N
E
For ordering information refer to chapter 26.
Canister and tubing 300 ml, 500 ml and 1000 ml
with Quick-Connector
Handle
IV pole/bed holder
Pump
Charger
Y-Connector with
Quick-Connector
Carrying case
Invia Ease NPWT system and accessories | 11
6.1 Invia Ease pump
The Invia Ease pump is a reusable suction pump that provides adjustable negative pressure,
constant and intermittent therapy modes, and an electronic measuring and monitoring
system. The Invia Ease pump is compact, portable and includes a long-lasting rechargeable
battery. The light indicator conveniently shows the therapy status and is easily visible from
a distance. The user-friendly touch screen guides the user through troubleshooting steps to
help resolve audio and visual alarms as well as provides instructions on how to change a
canister and to detach the patient from the pump. The complete history file can be accessed
via the touch screen throughout the therapy. The pump dynamically regulates the air flushing
cycles to help prevent clogging of the tubing.
Therapy mode
Default setting
Pressure range
Constant
–125 mmHg
–40 to –200 mmHg
Intermittent
–125 mmHg for 5 min /
–40 mmHg for 2 min
High values: –40 to –200 mmHg
Low values: 0 to –175 mmHg
Touch screen
Mute button
On/Off button
Light indicator
Charging port
6.2 Light indicator status
Therapy is running
Green
Yellow
– Therapy is paused
– Alarm state
i For instructions, refer to chapter “Alarms and troubleshooting”
on page 41.
12 | Invia Ease NPWT system and accessories
6.3 Touch screen navigation
Main screen
N
E
Motor stops running / Therapy is paused
Pause screen
u
n
e
m
n
a
c
n
i
i
i
l
C
s
s
e
c
c
A
u
n
e
m
n
a
c
n
i
i
i
l
C
s
s
e
c
c
A
Clinician menu
Therapy setting
Battery status
Air leak indicator:
Low/Medium/High
Therapy running status:
moving waves
Pause therapy timer
Open instructions
Log dressing change
T
h
e
r
a
p
y
r
e
s
t
a
r
t
s
r
u
n
n
n
g
/
R
e
t
u
r
n
t
i
i
o
M
a
n
s
c
r
e
e
n
Exit Clinician menu
View therapy history
Change therapy and pump
settings
End current therapy
Invia Ease NPWT system and accessories | 13
6.4 Invia Ease canister and tubing
NOTICE
Select the appropriate canister size depending on the wound type.
Canister sizes:
Tubing length:
Tubing inner diameter:
300 ml, 500 ml and 1000 ml
1.5 m
2.0 mm/3.0 mm
Canister release button
Control lumen
Removal lumen
Graduation scale
Clamp
Quick-Connector
to dressing tubing
Canister release button
Overflow protection/Carbon bacteria filter
– Canister release button disconnects the canister from the pump with one touch
– Control lumen (smaller tubing) regulates the pressure between the pump and the wound site
– Removal lumen (larger tubing) removes the fluid from the wound into the canister
– Graduation scale shows the amount of fluid in the canister
– Clamp closes the tubing to prevent fluid from spilling out
– Quick-Connector securely attaches canister tubing to the dressing tubing
– Overflow protection/carbon bacteria filter automatically clogs when it comes in contact
with fluids to prevent any fluids from entering the pump
– Solidifier is activated when exposed to fluids
14 | Invia Ease NPWT system and accessories
N
E
6.5 Invia Ease charger
WARNING
• Only use the charger that comes with the pump. An inadequate power source may result
in a fire hazard, electric shock or malfunction of the pump.
• Prior to connecting the charger, always inspect the charger and the cord for damage.
If damage is found, immediately disconnect the charger and call the Medela Customer
Service or the system provider.
The pump operates when the pump battery is sufficiently
charged or when the pump is connected to an electrical
outlet port via the charger provided with the pump.
Invia Ease NPWT system and accessories | 15
6.6 Accessories
Invia Y-Connector with Quick-Connector
Invia Ease carrying case
Invia Ease handle
NOTICE
For warnings and cautions of the Invia Y-Connector
with Quick-Connector consult the Invia Y-Connector
with Quick-Connector Instructions for Use.
Invia Y-Connector with Quick-Connector is intended to
connect two dressing tubings to a single Invia Ease pump.
NOTICE
The carrying case is for a single patient use only.
The carrying case is designed to more conveniently carry
around a pump, which does not include a handle. The
carrying case is compatible with the 300 ml and 500 ml
canisters and includes a removable shoulder strap for
easier transport. Canister exchange and battery charging
can be done while the pump is in the carrying case. The
outer pocket of the carrying case should be used to store
the quick reference card.
The Invia Ease pump is designed with an option to attach
a handle to the pump to more conveniently carry around
the pump. The handle should be affixed to the pump via
two screws, which are supplied with the handle.
Invia Ease IV pole/bed holder
The IV pole/bed holder is designed to be used with the
Invia Ease pumps, which include a handle. The IV pole/bed
holder allows for easy attachment and removal of the
pump to the IV pole or the bed board.
16 | Invia Ease NPWT system and accessories
6.7 Spare parts
For ordering information refer to chapter 26.
N
E
Filter gasket with orange O-ring
Handle attachment covers
Pump feet
Invia Ease NPWT system and accessories | 17
7 Initial pump setup:
first time the pump is turned on
Never operate the pump if it is not working properly, if it has been damaged or dropped
into water. If damage is found, immediately disconnect the charger and call the Medela
Customer Service or the system provider.
WARNING
NOTICE
Medela recommends to fully charge the pump battery before its first use.
i
INFORMATION
For instructions, refer to chapter “Battery status and charging” on page 20.
To carry out the initial pump setup:
Turn on the pump by pressing the “On/Off” button
The light indicator will illuminate yellow and the start-up
screen will appear.
Click to begin the initial setup.
18 | Initial pump setup: first time the pump is turned on
N
E
Select the preferred language from the list.
Click to continue.
Program the date and time:
– Choose between / or format
– To set date (day, month, year) and time (hours, minutes)
press or
Click to continue.
Select the method to access the Clinician menu. This prevents patients from changing any
settings. Choose between a Disclaimer or a Code.
Click to continue.
Disclaimer
Access the clinician menu with a
swipe.
Code
Access the Clinician Menu by entering a code:
– The default code is programmed as 1111.
– To set an additional code refer to chapter “Pump
settings/Clinician menu access” on page 36.
OR
The Invia Ease pump setup is now complete.
To proceed either:
Turn pump off
Turn the pump off by clicking
Start therapy
Start a new therapy by clicking
Initial pump setup: first time the pump is turned on | 19
8 Operating instructions
WARNING
Never operate the pump if it is not working properly, if it has been damaged or dropped
into water. If damage is found, contact the Medela Customer Service or the system
provider.
The Invia Ease NPWT system is intended to be used by clinicians or by adequately trained
lay users (patients or caregivers).
8.1 Pump positioning
its back during use.
– The Invia Ease pump should remain in an upright position or can be placed on its front or
– If the Invia Ease pump is placed in an inappropriate position (upside-down or on its side),
it will lead to the “Pump in the wrong position” alarm.
i
For troubleshooting instructions refer to chapter “Alarms and troubleshooting/Pump in
the wrong position” on page 42.
– If the canister filter becomes clogged because the pump is in an inappropriate position
for too long, it will lead to the “Canister full” alarm. The canister must then be replaced.
i
For troubleshooting instructions refer to chapter “Alarms and troubleshooting/Canister
full” on page 44.
✓
8.2 Battery status and charging
WARNING
• This product contains lithium-ion batteries, which bear risk of fire, explosion and burns.
• Only use the charger that comes with the pump. An inadequate power source may result
in a fire hazard, electric shock or malfunction of the pump.
• Prior to charging, always inspect the charger and the cord for damage. If damage is
found, immediately disconnect the charger and call the Medela Customer Service or
the system provider.
NOTICE
full charge.
• Medela recommends continuous charging while the pump is running overnight to ensure a
• If the pump is not being used, the battery must be charged approximately once every
6 months to ensure optimum function.
20 | Operating instructions
The battery level (as a percentage of fully charged) and the battery symbol are displayed
at the top of the touch screen:
N
E
Battery status/battery level and
symbol
Additional battery symbols/status
Battery is charging
Battery fault
Charging is not possible at this time
To charge the pump battery:
1. Connect the charger to the charging port.
The LED light on the charger will illuminate white when
the charger is correctly connected to the pump’s charging
port.
2. Plug the charger into an electrical outlet port.
Operating instructions | 21
When the pump begins to charge:
Therapy is running
– The touch screen lights up and the battery symbol
indicates that the pump is charging.
– If there is no user interaction after 1 minute, the touch
screen will enter sleep mode.
Pump is turned off
– The touch screen will display that the pump is charging.
– If there is no user interaction after 1 minute, the touch
screen will enter sleep mode.
– The pump is fully charged after approximately 4 hours.
– A fully charged pump should work for at least 20 hours. If there is an air-leak in the system,
the pump motor may need to work harder to maintain the target level of negative pressure.
Under these circumstances, the pump may need to be recharged more frequently.
– When the battery charge drops to 20%, an acoustic alarm will sound and the “Battery low”
alarm will appear on the touch screen. This alerts the user to recharge the battery.
i
For troubleshooting instructions refer to chapter “Alarms and troubleshooting/Battery
low” on page 42.
– When the battery charge drops below 5%, an acoustic alarm will sound, the “Battery
empty” alarm will appear on the touch screen and therapy will be automatically paused.
i For troubleshooting instructions refer to chapter “Alarms and troubleshooting/Battery
empty” on page 43.
22 | Operating instructions
N
E
8.3 Attach canister to the pump
WARNING
NOTICE
Sterile products: Do not use the sterile products if the sterile packaging is damaged,
was opened prior to use or has expired.
• Wear gloves for all operations and apply universal precautions.
• Canisters are designed for single patient use only and cannot be reused.
1. Select the appropriate canister size depending
on the wound type (300 ml, 500 ml, 1000 ml).
2. Unpack a new sterile canister.
3. Slip the pegs at the bottom of the canister to secure
into the slots at the base of the pump.
click
place.
4. Push the canister towards the pump until it clicks into
Operating instructions | 23
8.4 Connect canister tubing to dressing tubing
1. Align both parts of the Quick-Connector and push
the Quick-Connector together until you hear a click.
click
8.5 Turn pump ON
1. Turn on the pump by pressing the “On/Off” button
2. The light indicator will illuminate yellow and the start-up
screen will appear.
24 | Operating instructions
8.6 Start new therapy
NOTICE
• New therapy should only be started by a clinician.
• The default therapy setting of the Invia Ease pump is –125 mmHg in constant mode.
N
E
i For instructions, refer to chapter “Turn pump ON” on page 24.
1. Turn the pump ON.
2. To start new therapy:
In default settings
Click
Change settings first
Click
i
For instructions on how to change therapy settings
refer to chapter “Change therapy settings” on page 34.
For instructions on how to change pump settings refer
to chapter “Change pump settings” on page 35.
Click
3. The light indicator will illuminate green when therapy is running and the target level
of negative pressure is reached.
8.7 Activate the display during running therapy
Touch anywhere on the touch screen to activate the display.
The touch screen will automatically activate in the event of an alarm.
8.8 Dim the display/sleep mode during therapy
The touch screen will dim and go into sleep mode 1 minute after the last user interaction.
Operating instructions | 25
8.9 Pause therapy
WARNING
Therapy interruption: The recommended application of the therapy with Invia Ease
NPWT system is 24 hours without interruption. If therapy is interrupted for more than
2 consecutive hours (120 minutes), the dressing should be changed and therapy restarted
by a clinician in order to avoid the risk of infection or sepsis.
The pump may be paused for a variety of reasons including:
– To change the wound dressing. For detailed information regarding dressing applications,
consult the appropriate Invia dressing Instructions for Use.
– To change the cannister.
– To temporarily detach the patient from the pump.
The pump allows the therapy to be paused for up to 30 minutes. After therapy has been
paused for 30 minutes, the “Pause reminder” alarm will appear on the touch screen.
i
For troubleshooting instructions refer to chapter “Alarms and troubleshooting/Pause
reminder” on page 43.
To pause therapy:
Click and hold for 3 seconds.
Pause screen will appear, the pump motor will stop working and the light indicator will switch
from green to yellow.
Pause screen
Access Clinician menu
Pause therapy timer
Open instructions
Log dressing change
Restart therapy
26 | Operating instructions
N
E
8.10 Change a canister
WARNING
Sterile products: Do not use the sterile products if the sterile packaging is damaged,
was opened prior to use or has expired.
NOTICE
• To change a canister, therapy must be paused.
• Canisters are designed for single patient use only and cannot be reused.
• Change the canister a minimum of once a week or when:
– The canister appears full
– “Canister full” alarm appears on the touch screen
1. Click and hold for 3 seconds on the Main screen.
i For instructions to change the canister click on the Pause screen
i To exit the instructions at any time, click
2. Position the clamps next to the Quick-Connector and
close clamps on the canister tubing and the dressing
tubing.
3. Disconnect the canister tubing from the dressing tubing
by pressing the sides of the Quick-Connector.
Prevent the end of the dressing tubing from becoming
contaminated.
Operating instructions | 27
4. Press the canister release button on top of the canister
to release and remove the used canister.
5. Unpack a new sterile canister and attach it to the pump.
i
For instructions, refer to chapter “Attach canister to the pump” on page 23.
6. Connect canister tubing to the dressing tubing.
i
For instructions, refer to chapter “Connect canister tubing to dressing tubing” on page 24.
7. Open the clamp on the dressing tubing and make sure
the canister tubing is unclamped.
8. To resume the therapy click on the Pause screen.
The pump motor will start to work, the Main screen will appear and the light indicator
should switch from yellow to green when the target level of negative pressure is reached.
9. Discard the used canister in accordance with local environmental guidelines.
For instructions, refer to chapter “Disposal” on page 50.
28 | Operating instructions
N
E
8.11 Detach the pump from the patient
NOTICE
To detach the pump from the patient, therapy must be paused.
1. Click and hold for 3 seconds on the Main screen.
For instructions to detach from the pump click on the Pause screen
i
i
To exit the instructions at any time, click
2. Position the clamps next to the Quick-Connector and
close clamps on the canister tubing and the dressing
tubing.
3. Disconnect the canister tubing from the dressing tubing
by pressing the sides of the Quick-Connector. Prevent
the end of the dressing tubing from becoming contami-
nated.
4. The patient can perform the activity while being disconnected from the pump, such as
taking a shower.
5. When the activity is completed, re-connect the canister tubing to the dressing tubing.
i
For instructions, refer to chapter “Connect canister tubing to dressing tubing” on page 24.
Operating instructions | 29
6. Open the clamps on the dressing tubing and
the canister tubing.
7. To resume the therapy click on the Pause screen.
The pump motor will start to work, the Main screen will appear and the light indicator
should switch from yellow to green when the target level of negative pressure is reached.
8.12 Change a dressing
NOTICE
To change a dressing, therapy must be paused.
i
For detailed information regarding dressing applications, consult the appropriate Invia
dressing Instructions for Use.
1. Click and hold for 3 seconds.
2. Log the dressing change by moving the toggle bar from
left to right.
When the DRESSING CHANGE toggle is activated:
– The “Pause reminder” alarm is suspended
– The pump will provide an acoustic sound every 15 min
– A dressing change is logged in the pump’s History
3. Perform the dressing change according to the facility guidelines.
4. To resume the therapy click on the Pause screen.
The pump motor will start to work, the Main screen will appear and the light indicator
should switch from yellow to green when the target level of negative pressure is reached.
30 | Operating instructions
N
E
8.13 Turn the pump off to temporarily interrupt therapy
WARNING
Therapy interruption: The recommended application of the therapy with Invia Ease
NPWT system is 24 hours without interruption. If therapy is interrupted for more than
2 consecutive hours (120 minutes), the dressing should be changed and therapy restarted
by a clinician in order to avoid the risk of infection or sepsis.
For instructions on how to end therapy refer to chapter “End therapy” on page 40.
i
To turn the pump off either:
OR
1. Press the “On/Off” button
1. Press and hold the “On/Off” button
for 3 seconds
2. Click
Operating instructions | 31
8.14 Resume existing therapy (after the pump was turned off)
1. Press the “On/Off” button to turn the pump on.
i
i
2. Click to resume existing therapy.
To start new therapy click
For instructions, refer to chapter “Start new therapy”
on page 25.
NOTICE
New therapy should only be started by a clinician.
32 | Operating instructions
9 Access the Clinician menu
N
E
NOTICE
the Clinician menu.
• The Clinician menu should be used by clinicians only.
• The pump automatically switches to the Main screen 1 minute after the last interaction in
The Clinician menu can be accessed during running therapy or while therapy is paused.
From the Clinician menu it is possible to:
– Change therapy settings
– Change pump settings
– Check the history and log file of the existing therapy
– End therapy
To access the Clinician menu:
1. Select at the top of the Main
screen.
2. Acknowledge that you are a clinician:
Disclaimer
Read the disclaimer and swipe right to unlock.
Code
Enter the code (default code 1111 or a pre-set personal
code).
3. To exit the Clinician menu click
Access the Clinician menu | 33
10 Change therapy settings
NOTICE
• Changes to therapy settings should only be made as prescribed by a physician.
• The default therapy setting of the Invia Ease pump is –125 mmHg in constant mode.
1. Access the Clinician menu.
For instructions, refer to chapter “Access the Clinician menu” on page 33.
i
2. Click
3. Select between or
therapy modes.
4. Select the field that needs to be changed and then
use or to adjust:
– Pressure between –40 and –200 mmHg in increments
– Time between 2 and 60 minutes (in the intermittent
of 5 mmHg
mode only)
5. Click to save the new therapy settings.
If the new therapy settings are not saved, the pump will
revert to the previous therapy settings.
6. Click to return to the Clinician menu.
34 | Change therapy settings
N
E
11 Change pump settings
The following settings can be changed from the pump settings menu:
– Date and time
– Clinician access
– Alarm volume
– Vacuum intensity
– Language
– Bluetooth on/off
– Reset factory settings
1. Access the Clinician menu.
For instructions, refer to chapter “Access the Clinician menu” on page 33.
i
2. Click
3. Click
4. Select the field that needs to be changed
per instructions on the following pages.
5. Click to return to the Clinician menu.
Change pump settings | 35
11.1 Date and time
1. Choose between / or format.
2. Click or to set date (day, month, year) and time (hours, minutes).
3. Click to return to the pump settings menu.
11.2 Clinician menu access
1. Choose between a disclaimer or a code.
Disclaimer
Access the clinician menu with a swipe.
Code
Access the Clinician menu with a default code 1111
or set an additional code.
To set an additional code:
– Check the box “Set additional code”
– Enter a four digit code
– Click to return to the pump settings menu.
To change an additional code:
– Click and enter a new four digit code.
– Click to return to the pump settings menu.
36 | Change pump settings
N
E
Select the sound of the acoustic alarm. The sound always starts low and continues to get
louder until it reaches its maximum volume. The default alarm volume setting is Medium.
Select the speed at which the negative pressure is applied and the dressing compresses. The
Low vacuum intensity requires the longest time to apply the target negative pressure while
the High vacuum intensity is the quickest to apply the target negative pressure at the
dressing. The default vacuum intensity setting is Medium.
11.3 Alarm volume
11.3 Vacuum intensity
11.5 Language
1. Select the language from the list.
2. Click to return to the pump settings menu.
11.6 Bluetooth
– Bluetooth functionality is presently inactive regardless of Bluetooth setting (On or Off).
– Default setting of the Invia Ease pump is Bluetooth Off.
– Bluetooth interface is used by the Medela service centers only.
11.7 Factory settings reset
The Invia Ease pump can be reset to the default factory settings:
– Therapy setting is constant mode in –125 mmHg
– Language is English
– Clinician access is Disclaimer
– Alarm volume is medium
– Vacuum intensity is medium
– Unit pressure is mmHg
– Bluetooth is Off
Change pump settings | 37
12 View therapy history
1. Access the Clinician menu.
For instructions, refer to chapter “Access the Clinician menu” on page 33.
i
2. Click
3. To check the detailed therapy events , click
History screen
Exit history screen
Therapy start date
Therapy ID number
Number of dressing changes
performed during therapy
Indicates an alarm on the therapy
graph
Number of canisters used during
therapy
Check detailed therapy events
Therapy run rate
Therapy duration
Therapy graph showing dates,
pressure setting(s), alarm(s) and
pause(s)
4. Scroll down to view the log file. To filter the events
recorded in the log file, click to select the respective
fields
and scroll down to view the full list. Filterable events
include alarms, dressing changes, canister changes and
therapy pauses.
5. Click to return to the Clinician menu.
38 | View therapy history
13 End therapy
N
E
NOTICE
• The therapy history of the current therapy will be erased once therapy is ended
and the pump is turned off.
• Therapy should only be ended by a clinician.
i
If you wish to temporarily interrupt therapy, please refer to chapter “Turn pump off
to temporarily interrupt therapy” on page 31.
To end existing therapy either:
1. Access the Clinician menu.
i
For instructions, refer to chapter
“Access the Clinician menu” on page 33.
OR
1. Press the “On/Off” button
2. Click and hold
for 3 seconds.
3. Acknowledge you are a clinician
via a disclaimer or a code to confirm
the end of therapy.
2. Click and hold
for 3 seconds.
3. Therapy history of the current therapy will
be shown on the touch screen.
4. To see the detailed log file, click
5. To turn the pump off, click
End therapy | 39
14 Alarms and troubleshooting
WARNING
Therapy interruption: The recommended application of the therapy with Invia Ease
NPWT system is 24 hours without interruption. If therapy is interrupted for more than
2 consecutive hours (120 minutes), the dressing should be changed and therapy restarted
by a clinician in order to avoid the risk of infection or sepsis.
When the pump detects a fault, an acoustic alarm sounds, troubleshooting instructions
appear on the touch screen and the light indicator illuminates yellow. The pump differentiates
between “Low Priority” and “Medium Priority” alarms (IEC 60601-1:2005, 3rd edition and
IEC 60601-1-8:2006):
“Low priority” alarms:
– Battery low
– Temperature high
– No canister detected
– Pump in wrong position
“Medium priority” alarms:
– Pause reminder
– Battery empty
– Internal temperature exceeded
– High leakage
– Blockage
– Canister full
– Defective charger
– Pump error
To resolve an alarm:
1. Press the Mute button to silence the alarm.
The mute symbol will appear on the screen to
indicate that the pump has been silenced. The acoustic
alarm will resume in 5 minutes if the problem has not
been resolved.
2. Follow the instruction shown on the touch screen or refer to the chapter “Alarm table”
on page 42.
3. If the problem cannot be resolved, turn the pump off and contact the Medela Customer
Service or the system provider for further instructions. For instructions, to turn the pump off
refer to chapter “Turn pump Off to temporarily interrupt therapy” on page 31.
40 | Alarms and troubleshooting
Alarm table
“Low priority” alarms:
– Therapy continues to run
– Status indicator illuminates yellow
N
E
NOTICE
”Low Priority” alarms that are not resolved will lead to “Medium Priority” alarms.
To close the instructions after acknowledging the alarm and to return to the Main screen
click
Alarm type
Potential cause
Notes
Battery low
• Battery charge has
fallen to 20 %
• Alarm will repeat if
the battery charge
drops to 15 % and
subsequently to 10 %
Troubleshooting
instructions
Charge the pump
battery
i
For instructions,
refer to chapter
“Battery status
and charging” on
page 20.
Temperature high
The pump is next to a
heat source or is under
a cover
Remove the pump
from the heat source
or uncover the pump
If battery charge
drops to 5 %, it will
lead to the “Battery
empty” alarm and
therapy will be auto-
matically paused.
The pump is at risk
of overheating. If the
pump is not cooled, it
will lead to “Tempera-
ture exceeded” alarm
and therapy will be
automatically paused.
If a canister is missing
or is not correctly
attached, it will lead to
“High leakage” alarm.
No canister detected
The pump does not
detect a canister
• Attach a canister to
the pump
• Release and
reposition
the canister if
needed
Pump in wrong
position
The pump is positio-
ned on its side or is
upside down
Place the pump
in the upright position
If the pump is in a
wrong position for too
long, it could lead to
“Canister full” alarm.
Alarms and troubleshooting | 41
“Medium priority” alarms:
– Therapy may automatically pause
– Status indicator illuminates yellow and flashes
Alarm type
Potential cause
Notes
Pause reminder
The therapy has been
paused for 30 minutes
or longer
Troubleshooting
instructions
• To resume
the therapy click
• To remain in the
pause therapy
mode, click
After 90 minutes in the
pause therapy mode,
the “Pause reminder”
alarm will display an
additional warning
regarding dressing
change
Battery empty
Battery charge is 5 %
or below
Internal temperature
exceeded
The pump
has overheated
• Remove the pump
• Therapy is automati-
from the heat source
or uncover the pump
cally paused.
• To acknowledge
Charge the pump
battery
i
For instructions,
refer to chapter
“Battery status
and charging”
on page 20.
Therapy is automati-
cally paused. Resume
the therapy once the
pump is charging by
clicking
• Wait a minimum of
30 minutes for the
pump to cool down
the instructions and
switch to the Pause
screen click
• Resume the therapy
after the pump is
cooled off by
clicking
Blockage
• Tubing may be
• Clamp may be
twisted
closed
• Check the tubing
to make sure it is not
twisted
• Make sure that both
clamps are open
If the tubing is blocked
or the blockage
cannot be identified,
the canister and/or
the dressing should be
changed.
42 | Alarms and troubleshooting
Alarm type
Potential cause
Troubleshooting
instructions
Notes
High Leakage
The pump detects an
air leak in the system
and cannot maintain
the target level of
negative pressure
N
E
• If the leakage is not
resolved in the first
five minutes, the
pump will switch
into pause therapy
mode.
• Inspect the dressing
for an air leak
• Press firmly around
the edges of the
dressing and around
the Invia FitPad
• Apply additional film
around the dressing
if needed to seal the
leak
• Check that the
Quick-Connector is
securely attached
• Check that canister
is fully connected to
the pump
• Release and
reposition
the canister if
needed
• Ensure that the two
filter gasket orange
O-rings on the canis-
ter side of the pump
are not missing
Filter gasket with orange O-ring
Change the canister
For instructions,
i
refer to chapter
“Change a
canister” on
page 27.
• If the therapy has
been automatically
paused, click
to check if the
leakage has been
resolved or continue
to look for the leak
by following the
troubleshooting
instructions.
• If an air leak cannot
be identified or re-
solved, the dressing
should be changed.
If the pump is not
positioned upright,
it may cause the
canister filter to clog.
• Unplug the charger
immediately
• Inspect the charger
for damage
• Replace the charger
If the charger is not
unplugged imme-
diately, it may cause
damage to the pump.
• Click
to turn the pump off
• Press the “On/Off”
button on the
side of the pump to
restart the pump
If the problem cannot
be resolved, turn the
pump off and contact
the Medela Customer
Service or the system
provider for further
instructions.
Alarms and troubleshooting | 43
Canister full
• Canister is full
• Canister filter is
clogged
Defective charger
The charger may be
damaged
Pump error
The pump has an
internal error
15 Accessories setup
15.1 Invia Ease carrying case
To place the pump inside the carrying case:
1. Open the Velcro on the top of the carrying case.
2. Slide the pump with canister attached into
the designated slot.
3. Close the Velcro over the pump and over the canister.
Make sure that the canister tubing is not kinked and
the status indicator is visible.
4. Secure additional tubing with a Velcro on the side
of the pump.
44 | Accessories setup
To exchange a canister from the carrying case:
1. Detach the handle strap on the side of the canister.
N
E
2. Open the Velcro over the canister.
3. Unzip the carrying case from the canister side.
4. Change the canister.
i
For instructions, refer to chapter “Change a canister”
on page 27.
5. Zip up the carrying case from the canister side.
Accessories setup | 45
6. Close the Velcro over the canister.
7. Re-attach the handle strap on the side of the canister.
46 | Accessories setup
15.2 Invia Ease handle
To affix the carrying handle to the pump:
1. Remove the two hole covers from the back of the pump.
N
E
2. Align the two handle parts and fixate the screws inside
the two holes.
3. Cover the screws with the two hole covers from the
pump.
Accessories setup | 47
15.3 Invia Ease IV pole/bed holder
To attach the IV pole/bed holder to an IV pole or a bed board (up to 50 mm):
1. Turn the knob left to open the IV pole/bed holder.
2. Position the open holder in the center of the IV pole
or over the bed board.
3. Turn the knob to the right to firmly secure the holder
onto the IV pole or the bed board.
To mount the Invia Ease pump onto the IV pole/bed holder:
1. Hold the pump by the handle and position the base of
the handle over the IV pole/bed holder.
2. Slide the pump down along the handle until the IV pole/
bed holder is secured to the pump handle.
To remove the Invia Ease pump from the IV pole or the bed board:
1. Hold the pump by the handle and lift the pump up to release it from the IV pole/bed
holder.
48 | Accessories setup
16 Safety-related checks
N
E
– The Invia Ease pump is a device in protection class II (EN IEC 60601-1), the safety-related
checks are confined to visual inspection of the housing and charger for damage.
This check must be performed prior to each use.
– If any damage or safety defects are observed, contact the Medela Customer Service
or the system provider.
– Devices of protection class II do not have a protective earth conductor therefore there
is no need to check the earth leakage current.
– The Invia Ease pump enclosure is made entirely of insulated material. Tests of the enclosure
leakage current using common measuring instruments will therefore not reveal measurable
values.
– Even when suctioning a conductive fluid until the overflow protection device activates,
measurements of the patient leakage current using common measuring instruments will
not reveal measurable values.
– The Invia Ease pump does not have patient circuits or functional earth connections.
17 Maintenance and service
– The Invia Ease pump and charger should be maintained and repaired throughout
its service life in compliance with the Medela service procedures.
– Service work may only be carried out by Medela’s authorized personnel.
– To request maintenance service contact the Medela Customer Service.
18 Sterility requirements
WARNING
Sterile products: Do not use the sterile products if the sterile packaging is damaged,
was opened prior to use or has expired.
Invia Ease pump, charger, handle, IV pole/bed holder and carrying case cannot
be sterilized.
The following products are sterile and single use only:
Invia Ease canister and tubing
These are single use products that should be disposed of after use.
Do not reuse.
If reused, the performance of the product may deteriorate, cross-contamination may occur.
Safety-related checks/Maintenance and service/Sterility requirements | 49
19 Cleaning, disinfection and storage
Reusable devices and components from Medela are delivered non-sterile and are intended
for reuse. Prior to use, clean and reprocess the product by following the procedure in
the Cleaning and Disinfection instructions below.
Materials and equipments:
– Personal Protective Equipment (PPE: disposable gloves, proper protective gear)
– Lint free nonwoven wipes
– Clean water (<104 °F, <40 °C)
– Common household cleaner
• Never use steel brushes or steel wool for cleaning.
• Invia Ease pump and accessories (charger, handle and IV pole fixation) should be cleaned
and disinfected after every use.
• Never place or drop the pump into water or other liquids.
• Before cleaning/disinfecting the pump or the charger, disconnect the charger from
the electrical outlet port.
NOTICE
WARNING
19.1 Cleaning
– The Invia Ease pump and the accessories (charger, handle, IV pole fixation and carrying
case) can be wiped with a damp cloth using a common household cleaner throughout
the therapy.
– The carrying case can be washed in the washing machine if needed.
19.2 Disinfection
The Invia Ease pump and the accessories (charger, handle and IV pole fixation) should
be disinfected after every therapy with a disinfectant containing alcohol to avoid
cross-contamination.
NOTICE
• Do not spray disinfectants directly into openings as this may harm electronic components.
• Immersion disinfection, thermal disinfection and ultrasound cleaning are not permitted.
50 | Cleaning, disinfection and storage
N
E
To disinfect the Invia Ease pump and the accessories (charger, handle and IV pole
fixation):
1. Wear suitable protection (clothing, gloves, face mask and goggles) according to the facility
guidelines.
2. Apply the disinfectant agent in accordance with instructions from the manufacturer.
Pay particular attention to the edges, narrow corners, and bottom side.
3. Leave the disinfection agent on for the time recommended by the manufacturer.
4. Thoroughly clean the surface, all edges, housing niches, corners, and bottom side.
5. Wipe dry or air dry according to the instructions from the manufacturer.
6. If needed, repeat step 2–5.
7. Dispose contaminated material in accordance with local environmental guidelines.
For detailed cleaning instructions contact the Medela Customer Service.
19.3 Drying
After the manual disinfection, store the product on a clean dry cloth and let it dry.
Alternatively, follow the instructions provided by the disinfectant manufacturer.
19.4 Inspection
After reprocessing according to the steps described above, the device and its components
must be inspected for signs of degradation that may limit the useful life and/or performance
of the device, such as the following: visible corrosion, mechanical wear, abrasion, damage,
or deformation. Furthermore, check if the silicone O-ring is still in place (if applicable).
Discard the device if any signs of degradation are evident.
19.5 Storage
Store the product dry and dust-free.
20 Disposal
NOTICE
environmental guidelines.
• Invia Ease products should be handled and disposed of in accordance with the local
• The Invia Ease pump or the accessories must not be disposed of with the household refuse.
– The Invia Ease pump is made from various metal and plastic parts, which should be
disposed of in accordance with the local environmental guidelines.
– Before disposal of the pump, the rechargeable battery and electronics must be removed.
For instructions, contact the Medela Customer Service.
Cleaning, disinfection and storage/Disposal | 51
21 International regulations
21.1 Electromagnetic compatibility (EMC)
The Invia Ease pump is intended to be used in the professional healthcare facility environment
and home healthcare environment and is EMC tested in conformity with the requirements of
the standard IEC 60601-1-2:2014 4th Edition according to clause 7 and 8.9.
The pump needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in these instructions for use. Portable and
mobile RF communications can affect the pump.
NOTICE
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reason-
able protection against harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the distance between the equipment and receiver.
• Connect the equipment to an outlet on a circuit different from that to which the receiver
is connected.
• Consult the dealer or an experienced radio/TV technician for help.
WARNING
To prevent malfunction resulting from electromagnetic interference:
• The Invia Ease pump should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the Invia Ease pump should be observed to verify
normal operation in the configuration in which it will be used.
• Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies, RFID can affect
the Invia Ease pump and should be kept at a distance of at least 30 cm away from the
device.
• Use of accessories or cables other than those provided by the manufacturer of this device
could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
52 | International regulations
N
E
Guidance and manufacturer’s declaration – electromagnetic emissions
This Invia Ease pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Invia Ease pump should assure that it is used in such an
environment.
Emission tests
Compliance
Electromagnetic environment – guidance
RF Emissions
CISPR 11
RF emissions
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Group 1
Class B
Not
applicable
Not
applicable
The Invia Ease pump uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The Invia Ease pump is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
Compliance level
IEC 60601
test level
Electrostatic
discharge ( ESD )
IEC 61000-4-2
± 8 kV
contact discharge
± 8 kV
contact discharge
± 15 kV
air discharge
± 15 kV
air discharge
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV
for power supply
lines
± 2 kV
for power supply
lines
Surge
IEC 61000-4-5
± 0,5 kV, ± 1 kV
Line-to-line
± 0,5 kV, ± 1 kV
Line-to-line
Electromagnetic
environment
– guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material,
the relative humidity
should be at least 5 %.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
International regulations | 53
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % UT for 0.5 cycle
at 0°, 45°, 90°, 135°,
180°, 225°, 270°,
and 315°
0 % UT for 0.5 cycle
at 0°, 45°, 90°, 135°,
180°, 225°, 270°,
and 315°
0 % UT for 1 cycle
at 0°
0 % UT for 1 cycle
at 0°
70 % UT
for 25 cycles at
50 Hz at 0°
for 30 cycles at
60 Hz at 0°
0 % UT
for 250 cycles at
50 Hz at 0°
for 300 cycles at
60 Hz at 0°
70 % UT
for 25 cycles at
50 Hz at 0°
for 30 cycles at
60 Hz at 0°
0 % UT
for 250 cycles at
50 Hz at 0°
for 300 cycles at
60 Hz at 0°
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the Invia Ease
pump requires continued
operation during power
mains interruptions, it
is recommended that
the Invia Ease pump
be powered from an
uninterruptible power
supply or a battery.
It may be necessary
to position the Invia
Ease pump further
from sources of power
frequency magnetic fields
or to install magnetic
shielding. The power
frequency magnetic field
should be measured in
the intended installation
location to assure that it is
sufficiently low.
54 | International regulations
Guidance and manufacturer’s declaration – electromagnetic immunity
This Invia Ease pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Invia Ease pump should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
N
E
Conducted RF
IEC 61000-
4-6
3 Vrms
150 kHz to
80 MHz
outside ISM band
3 Vrms
6 Vrms
d = 2.0√P
6 Vrms
150 kHz to
80 MHz
in ISM and
amateur radio
bands
Radiated RF
IEC 61000-
4-3
10 V/m
80 MHz to
2.5 GHz
10 V/m
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Except as indicated in the table on the
next page, portable and mobile RF
communications equipment should be
used no closer to any part of the Invia
Ease pump, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommend separation distance
d = 1.2√P
d = 0.35√P 80 MHz to 800 MHz
d = 0.7√P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey,a should be less than the
compliance level in each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following
symbol.
International regulations | 55
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the Invia Ease pump is used exceeds the applicable RF compliance
level above, the Invia Ease pump should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating
the Invia Ease pump.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Table of frequencies of portable and mobile transmitters for which the recommended
separation distance is 30 cm (12 inches):
Band (MHz)
Service
380–390
TETRA 400
430–470
GMRS 460, FRS 460
704–787
LTE Band 13, 17
800–960
GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5
1 700–1 990
GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS
2 400–2 570
Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7
5 100–5 800
WLAN 802.11 a/n
WARNING
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the Invia
Ease pump including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
21.2 Radio transmission
FCC (USA) statement
Changes or modifications not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment. This device complies with Part 15
of the FCC Rules. Operation is subject to the following two conditions:
(1)
(2) this device must accept any interference received, including interference that may cause
this device may not cause harmful interference, and
undesired operation.
56 | International regulations
22 Technical specifications
N
E
vacuum range
–40 to –200 mmHg
–5 to –27 kPa
low flow
5 l/min.
without canister
1.1 kg
2.4 lbs
H x W x D
125 x 175 x 90 mm
4.92 x 6.89 x 3.54 inch
max. noise level
35 dBA
(250 ml/min; 125 mmHg)
alarm noise level
62/68/74 dBA
Switching adapter AC
Model: MSA-C2500IS12.0-30C-ZZ
IEC:
Input:
Output: 12 VDC, 2.5 A
60601-1
100–240 VAC, max 0.8 A, 50/60 Hz
Transport/Storage Conditions
Operating Conditions
ISO 13485
Technical specifications | 57
23 Material
Item description
Raw material description
Raw material information
Invia Ease pump (base unit)
Pump housing (upper and lower
part) is made of ABS material.
The silicone seal in between is
made of silicone material.
Housing: ABS
Housing seal: MQ Elastosil
Invia Ease charger
Charger housing is made of PC
material, plug is made of ABS.
PC and ABS material, heat
resistant
Invia Ease canister and tubing
300 ml, 500 ml, 1000 ml
Invia Ease carrying case
Invia Ease handle
Invia Ease IV pole/bed holder
The canister is made of
polypropylene material, the
tube is made of silicone and the
Quick-Connector is made of
ABS material.
Body of carrying case is made
of microfiber material.
Handle is made of PA with 30 %
glass fiber, seal between pump
and handle is made of silicone
material.
Plastic parts are made of PA
with 30 % glass fiber, silicone
pads are made of silicone
material, and metal parts are
stainless steel and aluminum.
Canister: PP
Tubing: MVQ, silicone
Quick-Connector: ABS
Polyester microfibre with
thermoplastic elastomer
backing
Plastic parts: PA
Pad: silicone
Metal parts: aluminium and
inox material
24 Warranty
Warranty period is for 2 years after date of delivery. The manufacturer is not liable for any
damage or consequential damage caused by incorrect operation, inappropriate usage as
well as use by unauthorized persons.
The service life of the device is five years; the internal batteries life included.
25 Service life
58 | Material/Warranty/Service life
26 Meaning of symbols
The following tables explain the meaning of the symbols found on the product parts
and its packaging.
N
E
Symbols used in these instructions
General safety alert symbol, points to information related to safety. 1
Symbols on the motor unit
Indicates the part number
of the device. 7
MD
This symbol indicates the
item is a medical device. X
Defines a relative humidity
range (e.g. for operation,
transport or storage). 3
Read and follow the
instructions for use. 12
Indicates the serial number
of the device. 8
Defines a temperature
range (e.g. for operation,
transport or storage). 2
Identifies the manufactur-
er. 5
This symbol indicates a
Prescription Device U.S.
Federal law restricts
this device to sale by or on
the order of a physician
(for US only). X
Contains fragile goods.
Handle with care. 14
Keep away from rain. Keep
in dry conditions. 16
Keep away from sunlight. 15
Indicates the compliance
with the requirements of
the Federal Communica-
tions Commission. 13
XXX
#
XXX
Indicates that the package
is capable of being
recycled. 17
Indicates the compliance
with additional USA and
Canada safety
requirements for medical
electrical equipment. 11
XXX
Indicates that this device
contains Bluetooth wireless
technology. (Trademarks
of Bluetooth Special
Interest Group (SIG))
Indicates compliance with
international requirements
for protection from electric
shock (Type BF applied
parts). 9
Do not dispose of electric/
electronic devices together
with unsorted municipal
waste (dispose of the
device in accordance with
local regulations). 10*
Indicates the location of
the On/Off button. 4
Indicates the location of
the mute button. 4?
Meaning of symbols | 59
Symbols on the power adapter
Indicates the compliance
with additional USA and
Canada safety
requirements for medical
electrical equipment. 11
The CE mark indicates
conformity with the
European low voltage and
electromagnetic
compatibility directive.*
Indicates alternating
current. 17
Do not dispose of electric/
electronic devices together
with unsorted municipal
waste (dispose of the
device in accordance with
local regulations). 10*
Identifies the manufactur-
er. 5
Indicates the part number
of the device. 7
Indicates the date of
manufacturing. 6
This symbol indicates
manufacturer’s batch
code. X
Symbols on the canister/tubing
Indicates the part number
of the device. 7
x
10
Indicates the quantity (x) of
individual devices in
pack. X
300 ml
Quick reference of the
canister size.
MD
This symbol indicates the
item is a medical device. X
This symbol indicates
manufacturer’s batch
code. X
This symbol indicates that
the device should not be
used after the date
shown. X
This symbol indicates a
single use device. Do not
reuse the device. X
This symbol indicates the
device is sterilized using
ethylene oxide. X
C onne
k
c
t
o
r
Quic
S
ecure H o l d
XXX
CC
XXX
This symbol indicates a
Prescription Device U.S.
Federal law restricts
this device to sale by or on
the order of a physician
(for US only). X
Identifies the manufactur-
er. 5
Contains fragile goods.
Handle with care. 14
Keep away from rain. Keep
in dry conditions. 16
Keep away from sunlight. 15 XXX
Indicates that the package
is capable of being
recycled. 17
Single sterile barrier
system with protective
packaging inside. X
Single sterile barrier
system. X
60 | Meaning of symbols
Symbols on the handle and IV pole/bed holder
N
E
Indicates the part number
of the device. 7
Indicates the date of
manufacturing. 6
Identifies the manufactur-
er. 5
Indicates that the package
is capable of being
recycled. 17
Contains fragile goods.
Handle with care. 14
Keep away from rain. Keep
in dry conditions. 16
This symbol indicates a
carton package. X
References
1 IEC 60601-1, Medical electrical equipment – Part 1: General Requirements for basic safety and essential
performance, Table D.2 Symbol 2 General Warning sign
2 ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and information
to be supplied, Part 1: General requirements, Clause 5.3.7 Temperature Limit/ISO 7000-0632,
Graphical symbols for use on equipment, Temperature Limit
3 ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and information
to be supplied, Part 1: General requirements, Clause 5.3.8 Humidity Limit/ISO 7000-2620,
Graphical symbols for use on equipment, Humidity Limit
4 IEC 60601-1, Medical electrical equipment – Part 1: General Requirements for basic safety and essential
performance, Table D.1 Symbol 29 Stand-by
5 ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and information
to be supplied, Part 1: General requirements, Clause 5.1.1 Manufacturer
6 ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and information
to be supplied, Part 1: General requirements, Clause 5.1.3 Manufacturing Date/ISO 7000-2497,
Graphical symbols for use on equipment, Date of manufacture
7 ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and information
to be supplied, Part 1: General requirements, Clause 5.1.6 Article number/ISO 7000- 2493,
Graphical symbols for use on equipment, Catalogue number
8 ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and information
to be supplied, Part 1: General requirements, Clause 5.1.7 Serial number/ISO 7000-2498,
Graphical symbols for use on equipment, Serial number
9 IEC 60601-1, Medical electrical equipment – Part 1: General Requirements for basic safety and essential
performance, Table D.1 Symbol 20 Type BF applied parts
10 EN 50419, Marking of Electrical and Electronic Equipment in accordance with Article 11(2) of Directive
2002/96/EC (WEEE).
11 TÜV (Technical Inspection Association) mark indicates that the product is manufactured in compliance
with UL safety requirements for USA and Canada (USA: UL60950-1, CAN: CSA C22.2 NO. 60950-1).
12 IEC 60601-1, Medical electrical equipment – Part 1: General requirements for basic safety and essential
performance, Table D.2 Symbol 10 Refer to instruction manual/booklet
13 Code of Federal Regulations, Title 47, Part 15b/15 c
14 ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and information
to be supplied, Part 1: General requirements, Clause 5.3.1, Fragile, handle with care/ISO 7000-0621,
Graphical symbols for use on equipment, Fragile, handle with care
15 ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and information
to be supplied, Part 1: General requirements, Clause 5.3.2 Keep away from sunlight/ISO 7000-0624,
Graphical symbols for use on equipment, Keep away from sunlight
16 ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and information
to be supplied, Part 1: General requirements, Clause 5.3.4, Keep away from rain/ISO 7000-0626,
Graphical symbols for use on equipment, Keep away from rain
17 ISO 7000-1135, Graphical symbols for use on equipment, General symbol for recovery/recyclable
* These symbols are not applicable for the US, Canada, Mexico market.
Meaning of symbols | 61
27 Ordering information
Product
Invia Ease NPWT system (pump and charger)
Invia Ease canister and tubing 300 ml (5pcs)
Invia Ease canister and tubing 500 ml (5 pcs)
Invia Ease canister and tubing 1000 ml (3 pcs)
Invia Ease carrying case
Invia Ease handle
Invia Ease IV pole/bed fixation
Invia Ease charger US
Filter gasket with orange O-ring
Pump feet
Handle attachment covers
REF number
101037357
101037361
101037362
101037363
101037360
101037364
101041744
101041939
62 | Ordering information
N
E
Medela AG
Lättichstrasse 4b
6340 Baar, Switzerland
www.medela.com
USA
Medela LLC
1101 Corporate Drive
McHenry, IL 60050
USA
Phone +1 877 735 1626
Fax
+1 815 307 8942
info-healthcare@medela.com
www.medelahealthcare.us
Medela Customer Service phone: 1-800-435-8316
A
-
6
0
-
1
2
0
2
/
9
1
4
3
4
0
1
0
1
/
G
A
a
e
d
e
M
©
l
1 2 | Power of Attorney Letter | Cover Letter(s) | 33.35 KiB | August 27 2021 |
m: dela 5 bhp Federal Communications Commission PNUlistelera-lilelelelate| Evaluation Division Equipment Authorization Branch 7435 Oakland Mills Road Columbia, MD 21046 Syelarcvivlas Werner Schnenberger Team Leader HW/SW IW Kre(=1 EW ANG) To whom it may concern:
We, the undersigned, hereby authorize Eurofins Product Service GmbH to act on our behalf in all matters cil aTo CoM Tevellter-li(e late iets equipment authorization FCC ID: 2ATCR608283 including the signing of all documents related to these matters. Any and all acts carried out by Eurofi as acts of our own. ns Product Service GmbH on our behalf shall have the same effect We also hereby certify that no party to this application is subj ect to a denial of benefits, including FCC benefits, pursuant to Section 5301 of the Anti-Drug Abuse Act of 1988, 21 U.S.C. 862. Medela AG Lattichstrasse 4b 6341 Baar Switzerland Phone +41 (0)41 5625151 Fax +41 (0)41 562 54 00 info@medela.ch www.medela.com 4/30/2021
frequency | equipment class | purpose | ||
---|---|---|---|---|
1 | 2021-08-27 | 2402 ~ 2480 | DTS - Digital Transmission System | Original Equipment |
2 | 13.56 ~ 13.56 | DXX - Part 15 Low Power Communication Device Transmitter |
app s | Applicant Information | |||||
---|---|---|---|---|---|---|
1 2 | Effective |
2021-08-27
|
||||
1 2 | Applicant's complete, legal business name |
Medela AG
|
||||
1 2 | FCC Registration Number (FRN) |
0028486975
|
||||
1 2 | Physical Address |
Laettichstrasse 4b
|
||||
1 2 |
Baar, N/A
|
|||||
1 2 |
Switzerland
|
|||||
app s | TCB Information | |||||
1 2 | TCB Application Email Address |
J******@eurofins.de
|
||||
1 2 | TCB Scope |
A4: UNII devices & low power transmitters using spread spectrum techniques
|
||||
1 2 |
A1: Low Power Transmitters below 1 GHz (except Spread Spectrum), Unintentional Radiators, EAS (Part 11) & Consumer ISM devices
|
|||||
app s | FCC ID | |||||
1 2 | Grantee Code |
2ATCR
|
||||
1 2 | Equipment Product Code |
608283
|
||||
app s | Person at the applicant's address to receive grant or for contact | |||||
1 2 | Name |
M**** H****
|
||||
1 2 | Telephone Number |
00414********
|
||||
1 2 | Fax Number |
00414********
|
||||
1 2 |
m******@medela.ch
|
|||||
app s | Technical Contact | |||||
1 2 | Firm Name |
Medela AG
|
||||
1 2 | Name |
I**** G******
|
||||
1 2 | Physical Address |
Laettichstrasse 4b
|
||||
1 2 |
Baar, 6340
|
|||||
1 2 |
Switzerland
|
|||||
1 2 |
i******@medela.com
|
|||||
app s | Non Technical Contact | |||||
n/a | ||||||
app s | Confidentiality (long or short term) | |||||
1 2 | Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | Yes | ||||
1 2 | Long-Term Confidentiality Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | No | ||||
if no date is supplied, the release date will be set to 45 calendar days past the date of grant. | ||||||
app s | Cognitive Radio & Software Defined Radio, Class, etc | |||||
1 2 | Is this application for software defined/cognitive radio authorization? | No | ||||
1 2 | Equipment Class | DTS - Digital Transmission System | ||||
1 2 | DXX - Part 15 Low Power Communication Device Transmitter | |||||
1 2 | Description of product as it is marketed: (NOTE: This text will appear below the equipment class on the grant) | The Invia Ease pump is a suction pump designed to help promote wound healing through Negative Pressure Wound Therapy. | ||||
1 2 | Related OET KnowledgeDataBase Inquiry: Is there a KDB inquiry associated with this application? | No | ||||
1 2 | Modular Equipment Type | Does not apply | ||||
1 2 | Purpose / Application is for | Original Equipment | ||||
1 2 | Composite Equipment: Is the equipment in this application a composite device subject to an additional equipment authorization? | Yes | ||||
1 2 | Related Equipment: Is the equipment in this application part of a system that operates with, or is marketed with, another device that requires an equipment authorization? | No | ||||
1 2 | Grant Comments | Output power listed is maximum peak conducted. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 5 mm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter except for the co-location described in this filing. | ||||
1 2 | Is there an equipment authorization waiver associated with this application? | No | ||||
1 2 | If there is an equipment authorization waiver associated with this application, has the associated waiver been approved and all information uploaded? | No | ||||
app s | Test Firm Name and Contact Information | |||||
1 2 | Firm Name |
Eurofins Electric & Electronic Product Testing AG
|
||||
1 2 | Name |
J******** B****
|
||||
1 2 | Telephone Number |
41 58********
|
||||
1 2 |
J******@eurofins.ch
|
|||||
Equipment Specifications | |||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Line | Rule Parts | Grant Notes | Lower Frequency | Upper Frequency | Power Output | Tolerance | Emission Designator | Microprocessor Number | |||||||||||||||||||||||||||||||||
1 | 1 | 15C | CC | 2402.00000000 | 2480.00000000 | 0.0070000 | |||||||||||||||||||||||||||||||||||
Line | Rule Parts | Grant Notes | Lower Frequency | Upper Frequency | Power Output | Tolerance | Emission Designator | Microprocessor Number | |||||||||||||||||||||||||||||||||
2 | 1 | 15C | CC | 13.56000000 | 13.56000000 |
some individual PII (Personally Identifiable Information) available on the public forms may be redacted, original source may include additional details
This product uses the FCC Data API but is not endorsed or certified by the FCC