WIRELESS TEMPERATURE MONITOR Model No.: ST323C Operation Manual Reading Before Using 0434 Thank you for purchasing MSR Wireless Temperature Monitor with Remote Monitor. Before you start using this unit, please read this operation manual thoroughly. Contents Introduction . Specification. Safety Note Components . Product Descriptions Preparation Before Using . Operation .. Other applications Ambient Temperature Measurement . Clock . Alarm -----------------------------------------------------------------------------------
Responding to Temperature Alarm -------------------------------------------------
Display Screen Messages .. Light Indicator Battery Replacement .. Care and Storage . Warranty.. iINTROUDCTIONj The Wireless Temperature Monitor (called device) is a newly innovative medical device intended for remote monitor of the armpit temperature up to 100 hours continuously for patients of 3 years older to adult when they are developing fever. The device consists of two units: the transmitter and the receiver. The transmitter functions as the unit not only to monitor the armpit temperature but also to transmit the temperature measured to the receiver that is remotely located by about 160 feet away. The receiver functions as the unit to receive the transmitted temperature signal from the transmitter and display on the LCD screen the temperature monitored. The frequency of the transmission is at 433.92 MHz that is approved by FCC for medical use. The receiver can automatically register the temperature in the memory at an interval of every 30 minutes since the beginning of monitoring. The care taker can set the alarm for upper and lower temperature limits of the monitored armpit temperature. If the monitored temperature exceeds the set limit, either upper or lower, the receiver will begin to alarm and flash to alert the care taker for taking proper action in response to the alarm. The device is also designed to function as a regular clock and ambient thermometer when armpit temperature monitor is dysfunctional. Also the receiver is equipped with the backlight function for easy night reading. The device is already approved by FDA and the CE mark is obtained as well. The technology is proprietary and the patent application to the US Patent and Trademark office is already submitted. iSpecificationsj 200hrs 500hrs Receiver: ABS Receiver: Approx. Size: Transmitter: 56.95 x 40.64 x 18.41mm Receiver : 72.4 x 23 mm Weight: Transmitter: approx. 11.20g (without battery) Receiver: approx. 41g (without battery) Product Material: Transmitter: ABS Type of Batteries: Transmitter: 3V x 1 (CR2032) Receiver: AAA / 1.5V x 2 Life of Batteries: Transmitter: Approx. Backlight: Backlight will be ON for 5 sec. whenever you press any button Effective Distance for Transmission and Reception:
160 feet distance (approx.50M) in an open space. If there are objects in that space, the transmission distance will be shortened. Range of Temperature Settings: 10.0J ~ 43.0J (50.0K~ 109.4K) Temperature Accuracy: 0.1J / 0.2K between 32J ~ 43.0J (89.6K ~ 109.4K) 0.5J / 1.0K beyond above range Storage environment: Transmitter / Receiver Ambient temperature: -10.0J ~ 60.0J (14.0K ~ 140.0K) Humidity: 30% -80% RH Temperature Measurement Interval: Transmitter measures body temperature once every 15 seconds. Transmitting Time: When transmitter detects body temperatures exceeding the Hi-Lo range three (3) times in a row, it transmits a signal causing the receiver to sound an alarm. Sound of Receiver Alarm: Lo: Long Bi. BiBi (repeats 6 times) Hi: Short Bi . Bi (repeats 6 times) Frequency of Transmission: 433.92 MHZ Memory Function: 200 sets During measurement mode, press or to review lasting readings. J/K : Can be selected on the receiver ON: Transmitter: Press ON/OFF button 3 seconds Receiver: Press On/OFF button for 3 seconds OFF: Transmitter: Press ON/OFF button over 4 second Receiver: Press ON/OFF button for 3 seconds iSafey Notej Warning 1. Before you start using this unit, please read this operation manual thoroughly for the correct methods of usage. 2. This device is not waterproof and must be never submersed in any liquid. 3. Do not drop or disassemble the units. 4. Avoid exposing the thermometer to direct sunlight, high temperature, moisture or dirt. 5. If the unit will not be used for a long time, it is advisable that the battery be removed and stored from the reach of children. 6. If the unit becomes contaminated, clean it with a damp cloth. Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user authority to operate the equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. The users manual or instruction manual for an intentional or unintentional radiator shall caution the user that changes or modifications not expressly approved by the party responsible for compliance could void the users authority to operate the equipment. iComponentsj Components in packing Fasten belt Sensor patch Battery for receiver Battery for Transmitter Receiver Transmitter Components assembly iProduct Descriptionsj Light Indicator Display Receiver group:
Selection button On/Off button Selection button Battery for Recevier Set button Set button For Year/Month/Date/Time On/Off button Transmitter group:
Sensor Patch Light indicator Battery for Transmitter Fasten belt Plug to sensor wire Plug to transmitter iPreparation Before Usingj Step one: Install battery
One CR2032 (3 volt) battery
Remove the battery compartment cover Insert one battery in the battery compartment, as indicated by the positive (+) faced up.
Replace the battery cover Two AAA (1.5V) batteries
Remove the battery compartment cover Insert two batteries in the battery compartment, as indicated.
Replace the battery cover Step one: Connect the patch and fasten belt and transmitter when you use it for temperature monitor Take the fasten belt out Plug the patch and transmitter to both ends of the sensor wire on the fasten belt Fix the fasten belt on the transmitter Wear it on the upper arm Note: The receiver should be kept with the caretaker remotely. Peel the cover on the patch and affix the patch on the armpit of the patient. iOperationj To activate the I/O button Transmitter
(including sensor ) Receiver Selection Button On/Off button Set button Press the I/O button approx. 3 second to activate this device. The LCD show all segments To Change K/J mode (the default is J) When the device is activated, press set button till you see C on the display. Press or to select K/J mode. Note:
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.