Instruction Manual of WatchBP Pro 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro operational sequence Actions LCD display Description Remark
Locate the black power switch on the back side.
Turn it on
The green indicator light on the top right front of the WatchBP Pro will illuminate and remain lit.
Locate the silver BT switch on the back side.
Turn it on
Antenna icon is flashing and waiting for connection
Please refer to Users guideline for WatchBP Pro Middleware
* Power Up the WatchBP Pro
*Power Up the Bluetooth
*First time switch on
Switch on the O/I button at the back side of the unit, LCD full segments display for 1 second with backlight on Auscultative Mode (one time measurement, on left arm).
*Measuring mode selection
Mode switch to AUSCULT., device ready for measuring.
Plug icon shows on LCD if 7.5V, 2.0A adaptor is used. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Cuff connection
Binding left-arm cuff on your left arm and place chest-piece of stethoscope on your brachial artery correctly.
*Press Start/Stop button
LCD full segment displays for 2 seconds with backlight on
*Auto zero reset
Display shows 0 when device
5 short Bs in auto zero reset.
* Fuzzy inflation
* Deflation
Pumping and starts, cuff pressure increases to working pressure (for fuzzy inflation).
Pumping stops, pressure deflation by 2~4 mmHg/sec. in 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Catch the SYS pressure.
*Catch the DIA pressure
* Measurement finishes
*Display measured values
Press M button when a faint, rhythmic tapping or thumping sound is heard, which stands for your SYS reading.
Press M button when your diastolic reading reaches and thumping sound stops, this is DIA reading.
A long beep
When measuring finishes, cuff deflates completely and a long beep sounds.
AUT (auto-measured result), AUS result) display for about 2 sec with backlight on simultaneously.
(Aus-measured
Device first displays of SP/DP, followed by MAP & PP values.
SP/DP and MAP/PP values
PP=SP-DP
MAP=
DP+1/3 PP display one after the other.
Backlight remains on 40 sec.
*Under auscult. mode, device auto measures patients BP & pulse and shows the results on the upper side of LCD (AUT). 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Memory reading
Device auto memorizes one set of measured values. Press M button, LCD shows AUE/AUS, then memory readings, with backlight on simultaneously
Memory SP/DP and MAP/PP displays one after the other.
Backlight remains on 40 sec.
*Withdraw from readings
Press Start/Stop button or put the device untouched for 5 to terminate measurement display.
First Visit mode (3 times measurement in succession, on both arms).
*Measuring mode selection
Mode switch to 1st and device ready for measuring. and blue
*Cuff connection tooth
Binding left & right arm cuffs on your both arms correctly.
Push blue-tooth switch on the back side of device (ON/OFF sculpted) to ON for computer connection if remote control. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Press Start/Stop button or remote control.
LCD full segment displays for 1 second with backlight on
*60 sec count down before measurement.
Beeper sounds 5 seconds before
5 short Bs the 1st measurement.
Press Start/Stop button or remote control to skip over count down and begin 1st turn measuring immediately.
* Auto zero reset
Display shows 0 for 2 sec when device in auto zero reset.
5 short Bs
* First turn fuzzy inflation.
Pumping starts & cuff pressure increases to working pressure
(for fuzzy inflation).
*First turn deflation.
Pumping stops, pressure deflation by 4~6 mmHg/sec. in 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Pulse detected turn measurement
*First finishes
*60 sec count down
Detect the pulse, icon start to flash and a beep is heard for every heartbeat.
Beeper each pulse
After first turn measuring, cuff deflates completely, and a long beeper sounds
A long B, and 2 short
After another two close beeps, device comes into 60 sec count down.
Beeper sounds 5 seconds before
5 short Bs the 2nd measurements.
Press Start/Stop button or remote control to skip over count down and begin 2nd turn measuring immediately.
* Auto zero reset
Display shows 0 for 2 sec when device in auto zero reset.
5 short Bs
* 2nd turn fuzzy inflation.
Pumping starts & cuff pressure increases to working pressure
(for fuzzy inflation).
2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*2nd turn deflation.
*Pulse detected turn measurement
*2nd finishes
*60 sec count down
Pumping stops, pressure deflation by 4~6 mmHg/sec. in
Detect the pulse, icon start to flash and a beep is heard for every heartbeat.
A beep each pulse
After 2nd turn measuring, cuff deflates completely, and a long beeper sounds
A long B, and 2 short
After another two close beeps, device comes into 60 sec count down.
Beeper sounds 5 seconds before
5 short Bs the 2nd measurements.
Press Start/Stop button or remote control to skip over count down and begin 3rd turn measuring immediately.
*Auto zero reset
Display shows 0 for 2 sec when device in auto zero reset.
5 short Bs 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*3rd turn fuzzy inflation.
Pumping starts & cuff pressure increases to working pressure
(for fuzzy inflation).
*3rd turn deflation.
Pumping stops, pressure deflation by 4~6 mmHg/sec. in
*Pulse detected
Detect the pulse, icon start to flash and a beep is heard for every heartbeat.
A beep each pulse
*The 3rd turn measuring finishes
After 3rd turn measuring, cuff deflates completely, and a long beeper sounds
A long B. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
* Display of average values
*Data transmission.
*Memory reading--average values display.
LCD displays values averaged from 3 times measurement with backlight on simultaneously.
Left arm BP values will flash 5 times by default if difference between L and R values are less than 10 mmHg, otherwise, the higher (L or R) flashes 5 times.
Display of SP/DP, following which, MAP & PP shows on.
SP/DP and MAP/PP values display one after the other till device auto off.
Backlight remains on 40 sec.
Through blue tooth, measured to values computer for further analysis. submitted are
Device auto memorizes 3 times their measured values and average readings.
Press M button, LCD shows an readings, on then average backlight A with simultaneously
Averaged SP/DP and MAP/PP displays one after the other.
Backlight remains on 40 sec. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
* Memory reading--3rd turn measured values display.
Press M button again, LCD shows a 3 then displays of the 3rd turn measured values.
SP/DP and MAP/PP displays one after the other. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Memory reading--2nd turn measured values display.
Press M button again, LCD shows a 2 then displays of the 2nd turn measured values.
SP/DP and MAP/PP displays one after the other. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Memory reading--1st turn measured values display.
Press M button again, LCD shows a 1 then display of the 1st turn measured values.
SP/DP and MAP/PP displays one after the other.
*Withdraw from readings
Press Start/Stop button or put the device untouched for 5 to terminate measurement display. Follow up mode (3 times measurement in succession, optional on left or right arm).
Mode switch to FOLLOW UP and device ready for measuring.
*Measuring mode selection 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro and blue
*Cuff connection tooth
*Choose one preferred arm for measurement.
*Press Start/Stop button or remote control.
*60 sec count down before measurement.
Binding left & right arm cuffs on your both arms correctly.
Push blue-tooth switch on the back side of device (ON/OFF sculpted) to ON for computer connection if remote control.
Hold down M button for 3 sec, memory shows with backlight on, till L or R flashes.
Press M button to choose that preferred arm for measurement, and press Start/Stop button for confirmation.
LCD full segment displays for 1 second with backlight on
Beeper sounds 5 seconds
5 short Bs before the 1st measurement.
Press Start/Stop button or remote control to skip over count down and begin 1st turn measuring immediately.
*Auto zero reset
Display shows 0 for 2 sec when device in auto zero reset.
5 short Bs 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
* First turn fuzzy inflation.
Pumping starts & cuff pressure increases to working pressure (for fuzzy inflation).
*First turn deflation.
Pumping stops, pressure deflation by 4~6 mmHg/sec. in
*Pulse detected
Detect the pulse, icon start to flash and a beep is heard for every heartbeat.
Beeper each turn measurement
*First finishes
After first turn measuring, cuff deflates completely and device comes into 60 sec count down.
*60 sec count down
Beeper sounds 5 seconds
before the 2nd measurements.
Press Start/Stop button or remote control to skip over count down and turn measuring immediately. begin 2nd pulse 5 sh ort B s 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
* Auto zero reset
Display shows 0 for 2 sec when device in auto zero reset.
5 short Bs
* 2nd turn fuzzy inflation.
Pumping starts & cuff pressure increases to working pressure
(for fuzzy inflation).
*2nd turn deflation.
Pumping stops, pressure deflation by 4~6 mmHg/sec. in
*Pulse detected
Detect the pulse, icon start to flash and a beep is heard for every heartbeat.
A beep each pulse 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*2nd turn measurement finishes
*60 sec count down
After 2nd turn measuring, cuff
deflates completely. If a difference of 5mmHg exists between 1st&2nd time measured sys or dia pressure, device auto comes into 60 sec count down and takes the 3rd measurement.
*Measuring finishes with a long beep and average values display if difference between 1st & 2nd turn measured sys or dia pressure is less than 5 mmHg.
Beeper sounds 5 sec before the 3rd turn measurement.
Press Start/Stop button or remote control to skip over count down and turn measuring immediately. begin 3rd
5 sh ort B s
*Auto zero reset
Display shows 0 for 2 sec when device in auto zero reset.
5 short Bs
* 3rd turn fuzzy inflation.
Pumping starts & cuff pressure increases to working pressure
(for fuzzy inflation). 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*3rd turn deflation.
*Pulse detected
Pumping stops, pressure deflation by 2~4 mmHg/sec. in
Detect the pulse, icon start to flash and a beep is heard for every heartbeat.
A beep each pulse turn measurement
*3rd finishes
After 3rd turn measuring, cuff deflates completely, and a long beeper sounds
A long B.
* Display of average values
LCD displays values averaged from 3 times measurement with backlight on simultaneously.
Display of SP/DP, following which, MAP & PP shows on.
SP/DP and MAP/PP values display one after the other till device auto off.
Backlight remains on 40 sec. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Data transmission.
*Memory reading--average values display.
Through blue tooth, measured to values computer for further analysis. submitted are
Device auto memorizes 3 times their measured values and average readings.
Press M button, LCD shows an A then average reading, with backlight on simultaneously.
Averaged SP/DP and MAP/PP displays one after the other.
Backlight remains on 40 sec. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
* Memory reading--3rd turn measured values display.
Press M button again, LCD shows a 3 then displays of the 3rd turn measured values.
SP/DP and MAP/PP displays one after the other. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Memory reading--2nd turn measured values display.
Press M button again, LCD shows a 2 then displays of the 2nd turn measured values.
SP/DP and MAP/PP displays one after the other. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
*Memory reading--1st turn measured values display.
Press M button again, LCD shows a 1 then display of the 1st turn measured values.
SP/DP and MAP/PP displays one after the other.
*Withdraw from readings
Press Start/Stop button or put the device untouched for 5 to terminate measurement display.
*Memories deletion.
3 short B
Hold down M button for 7 sec till 3 short beepers are heard and bars appear on LCD with backlight on simultaneously, which means memory readings are deleted successfully. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
* Battery low indication
*Turn off the device.
When voltage inside batteries is lower to 4.8V, LCD shows low battery icon and batteries begin to be charged if adaptor is connected.
Once voltage inside batteries is lower to 4.6V, LCD shows low battery icon, then batteries must be charged again.
Switch off O/I button at the back side of the unit
Memories wont be stored once the device is turned off. Federal Communications Commission (FCC) Statement 15.21 You are cautioned that changes or modifications not expressly approved by the part responsible for compliance could void the users authority to operate the equipment. 15.105(b) This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which the receiver is 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro connected.
-Consult the dealer or an experienced radio/TV technician for help. Operation is subject to the following two conditions:
1) this device may not cause interference and 2) this device must accept any interference, including interference that may cause undesired operation of the device. FCC RF Radiation Exposure Statement:
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. End users must follow the specific operating instructions for satisfying RF exposure compliance. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro