TEST REPORT #010419 STANDARD: FCC PART 15 SUBPART C--INTENTIONAL RADIATORS SECTION 15. 247 OPERATION WITHIN THE BANDS 902-928 MHZ, 2400-2483.5 MHZ, AND 5725-5850 MHZ EQUIPMENT TESTED:
MX3 DIAGNOSTICS, INC. MODEL: HYD-PRO-18 FCC ID: 2ASD4-HYD-PRO-18 TEST DATE: 01 APRIL 2019 INTERNATIONAL 1100 Falcon Avenue Glencoe, MN 55336 Tele: 320-864-4444 Fax: 320-864-6611 CERTIFICATION SERVICES, INC. MX3 Diagnostics, Inc. 2701 Stratford Drive Austin, TX 78746 International Certification Services, Inc. 1100 Falcon Avenue Glencoe, MN 55336 Tele: 320-864-4444 Fax: 320-864-6611 International Certification Services, Inc. 1100 Falcon Avenue Glencoe, MN 55336 Tele: 320-864-4444 Fax: 320-864-6611 International Certification Services, Inc. 1100 Falcon Avenue Glencoe, MN 55336 Prepared for:
Test agent:
Test location:
Prepared by:
International Certification Services represents to the client that testing is done in accordance with standard procedures applicable and that reported test results are accurate within generally accepted commercial ranges of accuracy. This report only applies to the specific samples tested under stated test conditions. It is the manufacturers responsibility to assure that additional production units of this model are manufactured with identical electrical and mechanical components. International Certification Services shall have no liability for any deductions, inferences or generalizations drawn by the client or others from this report. This report is the confidential property of the client. As a mutual protection to our clients, the public and ourselves, extracts from the test report shall not be reproduced except in full without our written approval.
. 2 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. Table of Contents
. ........................................................................................................................................................ 2 1.0 TEST SUMMARY ................................................................................................................... 4 2.0 Applicable Standards ............................................................................................................. 5 3.0 Referenced Standards ........................................................................................................... 5 4.0 Equipment Units Tested ......................................................................................................... 5 5.0 Equipment and Cable Configuration ...................................................................................... 5 6.0 List of Test Equipment ............................................................................................................ 6 7.0 Units of Measurement. ........................................................................................................... 6 8.0 Location of Test Site............................................................................................................... 7 9.0 Measurement Procedures ...................................................................................................... 7 10.0 Reporting Measurement Data ............................................................................................. 7 Radiated Emissions Data .................................................................................................... 7 11.0 Summary of Results ............................................................................................................ 7 12.0 13.0 Attachment A Test Set Up ................................................................................................. 8 Attachment B Test Data .................................................................................................... 11 14.0 15.247 (a) (1) Not Applicable ............................................................................................ 11 15.0 15.247 (a) (2) DTS Bandwidth PASS ................................................................................ 11 16.0 17.0 15.247 (b) (1) Not Applicable ............................................................................................ 13 15.247 (b) (2) Not Applicable ............................................................................................ 13 18.0 15.247 (b) (3) Output Power PASS .................................................................................. 13 19.0 15.247 (b) (4) Antenna PASS ........................................................................................... 14 20.0 21.0 15.247 (c) Antenna Gain Not Applicable ........................................................................... 14 15.247 (d) Out of Band Spurious Emissions PASS .......................................................... 14 22.0 15.247 (e) Power Spectral Density PASS ......................................................................... 18 23.0 24.0 15.247 (f) Hybrid Systems Not Applicable......................................................................... 20 15.247 (g) FHSS systems Not Applicable ......................................................................... 20 25.0 15.247 (h) FHSS Systems Not Applicable ........................................................................ 20 26.0 15.247 (i) PASS ................................................................................................................ 21 27.0 28.0 Lower Band Edge: PASS .................................................................................................. 21 Upper Band Edge: PASS .................................................................................................. 22 29.0 Power Line Conducted Emissions: PASS ......................................................................... 23 30.0 SAR Requirements: PASS ................................................................................................ 23 31.0 32.0 ATTACHMENT C Product Information ............................................................................. 25 3 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 1.0 TEST SUMMARY 320-864-4444 01 April, 2019 TEST REPORT: #010419 COMPANY:
AGENT:
PHONE:
TEST DATE:
EQUIPMENT UNDER TEST:
GENERAL TEST SUMMARY:
VERIFICATION / CERTIFICATION STATUS:
MODIFICATIONS NECESSARY:
MX3 Diagnostics, Inc. International Certification Services, Inc. Hydration Measuring Instrument FCC ID: 2ASD4-HYD-
PRO-18 Model: HYD-PRO-18 The testing was performed at International Certification Services, Inc. at 1100 Falcon Ave, Glencoe, MN 55336 The Hydration Measuring Instrument FCC ID: 2ASD4-HYD-
PRO-18 Model: HYD-PRO-18 was found to be in compliance with the FCC Part 15 Subpart C, Section 15.247 requirements for a DTS device. None TESTED BY Duane R. Bagdons WRITTEN BY Duane R. Bagdons 4 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 2.0 Applicable Standards 47 CFR Ch.1 (10-1-98 Edition) FCC Part 15 Subpart C Intentional Radiators Radio Frequency Devices Section 15.247 Operation within the bands 902-928 Mhz, 2400-2483.5 Mhz and 5725-5850 Mhz. 3.0 Referenced Standards ANSI C63.4-2014 Methods of Measurement of Radio-Noise Emissions from Low-Voltage FCC DA- 00-705 Electrical and Electronic Equipment in the Range of 9 Khz to 40 Ghz. Filing and Measurement Guidelines for Frequency Hopping Spread Spectrum Systems ANSI C63.4-2014 American National Standard for Methods of Measurement of Radio Noise Emissions from Low Voltage Electrical and Electronic Equipment in the Range of 9 kHz to 40 Ghz ANSI C63.10-2013 American National Standard of Procedures for Compliance Testing of FCC KDB 447498 D01 RF Exposure Procedures and Equipment Authorization Policies Unlicensed Wireless Devices, Section 11 FCC KDB 558074 D01 Guidance for Compliance Measurements on Digital Transmission for Mobile and Portable Devices. System, Frequency Hopping Spread Spectrum System, and Hybrid System Devices Operating Under Section 15.247 of the FCC Rules V05R01 FCC KDB 447498 D01 General Exposure Guidance V06 Section 4.3.1 4.0 Equipment Units Tested The equipment tested was a battery and AC powered 2400 to 2483.5 MHz transmitter. AC power comes from a plugin wall wart type USB charger. The product is typically used only with the internal battery power, but the battery can be charged from the AC public Mains during operation, so the unit was tested with AC power applied to it. The antenna is an integrated chip soldered onto the PC board. The Antenna is a Model: 2450AT42E010B surface mount, above metal, low profile mini chip antenna made by Johanson Technology. The gain of the antenna is specified at -2 peak dBi. This device uses a Nordic Semiconductor Model: nRF52832 for the Bluetooth Transceiver circuitry. The Hydration Measuring Instrument Model: Hyd-Pro-18 was configured as normal operation and was programmed to provide the special test sequences that were required to test to the FCC 15.247 requirements. 5.0 Equipment and Cable Configuration See photo of the EUT test configuration setup in Attachment A 5 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 6.0 List of Test Equipment Test Equipment Spectrum Analyzer Preamp Preamp Preamp Loop Antenna Loop Antenna Biconical Antenna Log Periodic Antenna
(200-1000 MHz) Horn Antenna (1-18 Ghz) Horn Antenna (3.95-5.85 Ghz) Horn Antenna (5.4-8.2 Ghz) Horn Antenna (8.2-12.4 Ghz) Horn Antenna (12.4-18 Ghz) Horn Antenna (18-26 Ghz Model Hewlett-Packard 8566B MiniCircuits ZKL-2R7 Comlilnear Corporation CLC-102 Nextec NB00391 Electrometrics ALP-11 /
EM-6870 EMCO 6512 AH Systems Model SAS-
200/540 EMCO 3146 EMCO 3115 Systron Donner DBK-520-
15 Narda 642 Scientific Atlanta 8.2-12.4 Alpha Industries 61932400 Alpha Industries 61932500 S/N Next Calibration 2421A00458 N/A 22773 378 286 8912-1074 328 9111-3280 5697 207 1172 30 55 Date 12/01/19 12/26/18 03/07/18 03/22/19 06/01/18 06/01/18 04/17/18 04/17/18 08/17/18 05/26/18 05/26/18 05/26/18 05/26/18 05/26/18 All equipment is on a 2 year calibration cycle Measurement cable losses, and antenna correction factors are included in the data sheets. 7.0 Units of Measurement. Field strength measurements were recorded in dBm/m and dBuV/m with the antenna located at 3 meters distance from the EUT. Frequency measurements are recorded in Mhz 6 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 8.0 Location of Test Site The open area test site (OATS) measurement facility used to collect the data was International Certification Services, Inc. at 1100 Falcon Ave in Glencoe, MN 55336. This site has been certified to be in spec of the normalized site attenuation per ANSI C63.4-
2014. 9.0 Measurement Procedures The antenna was placed at a distance of 3 meters from the EUT. The EUT was set on an insulating table in the OATS site and rotated through 360 degrees to determine the worst case EUT orientation. The antenna was then positioned vertical and horizontal to determine which antenna polarity orientation was worst case. Then certification data was recorded at all the transmitter frequencies from the fundamental to the 10th harmonic at an antenna height variation of from 1-4 meters. 10.0 Reporting Measurement Data See data sheets and plots in Attachment B. 11.0 Radiated Emissions Data The frequency and amplitude of the tuned frequency of the EUT along with the frequencies and amplitudes of the harmonics up to the 10th harmonic are reported in the data sheets in Attachment B. This information is plotted against the limit of section 15.247 of FCC Part 15 subpart C. Both Horizontal and Vertical antenna polarities as well as antenna heights of 1 to 4 meters were observed but all maximum signal strengths occurred in the Horizontal antenna polarity and at 1.5 meter antenna height. The Final Level, expressed in dBuV/m, is arrived at by taking the reading from the spectrum analyzer (Level dBuV) and adding the antenna correction factor and cable loss factor
(Factor dB) and subtracting the preamp gain. This result then has the FCC limit subtracted from it to provide the margin which gives the tabular data as shown in the data sheets in Attachment B. Example:
Frequency
(MHz) 100.0 12.0 Summary of Results Level + Factor = Corr Data - FCC Limit =
=
(dBuV) + (dB) = (dBuV/m) - (dBuV/m) 20.6
=
+ 11.0 = 31.6
- 43.5 Margin
(dB)
-11.9 The EUT passed the requirements of FCC Part 15 Subpart C, Section 15.247 as a DTS device. No modifications were necessary to accomplish this compliance. 7 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 13.0 Attachment A Test Set Up RADIATED AND CONDUCTED MEASUREMENT TEST SET UP 8 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. MX3 Diagnostics, Inc. Model: Hyd-Pro-18 FCC ID: 2ASD4-HYD-PRO-18 Radiated Emissions Test Configuration 9 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. MX3 Diagnostics, Inc. Model: Hyd-Pro-18 FCC ID: 2ASD4-HYD-PRO-18 Conducted Emissions Test Configuration 10 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 14.0 Attachment B Test Data MX3 Diagnostics, Inc. Model: HYD-PRO-18 FCC ID: 2ASD4-HYD-PRO-18 Temperature: 10.5 Deg C. Humidity: 68 % R.H. Test Technician: Duane R. Bagdons Center Frequency: 2402.159 Mhz (low channel) 2440.11 Mhz (mid channel) 2480.105 Mhz (high channel) Preliminary testing was done to determine what antenna polarity and antenna height generated the highest signal levels. Tests were performed at this test configuration and then each frequency was maximized to 0-360 degrees orientation and antenna height of 1-4 meters. 15.0 16.0 15.247 (a) (2) DTS Bandwidth PASS 15.247 (a) (1) Not Applicable Systems using digital modulation techniques may operate in the 902-928 Mhz, 2400-2483.5 Mhz and 5725-5850 Mhz bands. The minimum 6 dB bandwidth shall be at least 500 Khz. Channel (Frequency) DTS Bandwidth Low: 2402.159 Mhz Middle: 2440.11 Mhz High: 2480.105
(Mhz) 0.862 0.542 0.823 11 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 12 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.Freq (Mhz)2398239924002401240224032404dBm/m-70-65-60-55-50-45-40-35-30MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247 (a) (2)DTS Bandwidth Modulation ON Low Channel (2402.159 MhzKDB558074 Option 1April 1, 2019International Certification Services, Inc.Maximum -32.96 dBm2401.699 Mhz2402.562 MhzLow Channel Bandwidth = 0.862 MhzFreq (Mhz)24382439244024412442dBm/m-80-75-70-65-60-55-50-45-40MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247 (a) (2)DTS Bandwidth Modulation ON Mid Channel (2440.11 Mhz)KDB558074 Option 1April 1, 2019International Certification Services, Inc.Maximum -43.2518 dBm2439.844 Mhz2440.387 MhzMiddle Channel Bandwidth = 0.542 Mhz 17.0 18.0 19.0 15.247 (b) (1) Not Applicable 15.247 (b) (2) Not Applicable 15.247 (b) (3) Output Power PASS For systems using digital modulation in the 902-928 MHz, 2400-2483.5 MHz, and 5725-5850 MHz bands: 1 Watt. As an alternative to a peak power measurement, compliance with the one watt limit can be based on a measurement of the maximum conducted output power. Maximum Conducted Output Power is defined as the total transmit power delivered to all antennas and antenna elements averaged across all symbols in the signaling alphabet when the transmitter is operating at its maximum power control level. Power must be summed across all antennas and antenna elements. The average must not include any time intervals during which the transmitter is off or is transmitting at a reduced power level. If multiple modes of operation are possible (e.g., alternative modulation methods), the maximum conducted output power is the highest total transmit power occurring in any mode. This data was measured in units of dBm/m. Equations from ANSI C63-10: 2013 Section G5.2 were used to convert from dBm to watts 13 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.Freq (Mhz)24782479248024812482dBm/m-70-65-60-55-50-45-40-35-30MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247 (a) (2)DTS Bandwidth Modulation ON High Channel (2480.105 Mhz)KDB558074 Option 1April 1, 2019International Certification Services, Inc.Maximum -33.531 dBm2479.716 Mhz2480.539 MhzHigh Channel Bandwidth = 0.82331 Mhz
(31) EIRP = P + L PSD = Measured Power Spectral Density (including all correction factors) (dBm/m) L = Free Space Propagation Loss (dB) mW = 10 LOG dBm W = mW / 1000 L = 20 LOG (F) + 20 LOG (d) - 27.5
(32) EIRP = PSDmeasured + L EIRP(W)= 10(PSD dBm
/ 10)/ 1000 = 10((PSD dBm
- 30) / 10) Freq (Mhz) 2402.159 2440.11 2480.105 PSDMeasured
(dBm/m)
-52.71
-54
-53.68 L (dB) 49.65 49.79 49.93 EIRP
(dBm/m)
-3.06
-4.21
-4.05 EIRP
(mW) 0.494 0.379 0.393 EIRP (W) 0.000494 0.000379 0.000393 See graphs in previous section 20.0 15.247 (b) (4) Antenna PASS The conducted output power limit specified in paragraph (b) of this section is based on the use of antennas with directional gains that do not exceed 6 dBi. Except as shown in paragraph (c) of this section, if transmitting antennas of directional gain greater than 6 dBi are used, the conducted output power from the intentional radiator shall be reduced below the stated values in paragraphs
(b)(1), (b)(2), and (b)(3) of this section, as appropriate, by the amount in dB that the directional gain of the antenna exceeds 6 dBi. The antenna used is a mini chip antenna made by Johanson Technology Model: 2450AT42E010B with a gain of -2.0 dBi. 21.0 Antenna Gain Not Applicable 15.247 (c) Operation with directional antenna gains greater than 6 dBi. 22.0 15.247 (d) Out of Band Spurious Emissions PASS In any 100 kHz bandwidth outside the frequency band in which the spread spectrum or digitally modulated intentional radiator is operating, the radio frequency power that is produced by the intentional radiator shall be at least 20 dB below that in the 100 kHz bandwidth within the band that contains the highest level of the desired power, based on either an RF conducted or a radiated measurement, provided the transmitter demonstrates compliance with the peak conducted power limits. If the transmitter complies with the conducted power limits based on the use of RMS 14 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. averaging over a time interval, as permitted under paragraph (b)(3) of this section, the attenuation required under this paragraph shall be 30 dB instead of 20 dB. Attenuation below the general limits specified in 15.209(a) is not required. In addition, radiated emissions which fall in the restricted bands, as defined in 15.205(a), must also comply with the radiated emission limits specified in 15.209(a) (see 15.205(c)). No signals from the Intentional Radiator were observed in the frequency spectrum of 30-1000 Mhz Above 1000 Mhz the following signals were detected coming from the Intentional Radiator:
Channel Frequency
(Mhz) Maximum Fundamental Power Output Harmonic
(dBm/m) Power Difference
(dBm) 2402.159
(dBm/m)
-38.32 4804.32 7206.48 2440.11
-43.25 4880.22 7320.33 2480.105
-33.53 4960.21 7440.32
-62.13
-69.52
-63.88
-69.07 23.81 31.20 20.63 25.82
-65.37
-65.8 31.84 32.27 15 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 16 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.Freq (Mhz)100010000dBm/m-100-90-80-70-60-50-40-30MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247 (d) Spurious Harmonics Low Channel 2402.159 MhzJune 27, 2019International Certification Services, Inc.2402.159 Mhz -38.32 dBm/m4804.32 Mhz -62.13 dBm/m7206.48 Mhz, -69.52 dBm/mFreq (Mhz)100010000dBm/m-100-90-80-70-60-50-40-30MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247 (d) Spurious Harmonics Middle Channel 2440.11 MhzJune 27, 2019International Certification Services, Inc.4880.22 Mhz -63.88 dBm/m2440.11 Mhz -39.79 dBm/m7320.33 Mhz, -69.07 dBm/m 17 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.Freq (Mhz)100010000dBm/m-100-90-80-70-60-50-40-30MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247 (d) Spurious Harmonics High Channel 2480.105 MhzJune 28, 2019International Certification Services, Inc.2480.105 Mhz -38.44 dBm/m4960.21 Mhz -65.37 dBm/m7440.32 Mhz, -65.8 dBm/m 23.0 15.247 (e) Power Spectral Density PASS For digitally modulated systems, the power spectral density conducted from the intentional radiator to the antenna shall not be greater than 8 dBm in any 3 kHz band during any time interval of continuous transmission. This power spectral density shall be determined in accordance with the provisions of paragraph (b) of this section. The same method of determining the conducted output power shall be used to determine the power spectral density. Channel Frequency
(Mhz) 2402.159 2440.11 2480.105 Power Spectral Density (dBm) FCC 15.247 (e)
(dBm)
-52.71
-54
-53.68 8 8 8 18 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.Freq (Mhz)2400.02400.52401.02401.52402.02402.52403.02403.52404.0dBm/m-100-90-80-70-60-50MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247 ePower Spectral DensityLow Channel (2402.159 Mhz)April 1, 2019International Certification Services, Inc.2402.159 MhzPower Spectral Density -52.7 dBm/m 19 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.Freq (Mhz)2438.02438.52439.02439.52440.02440.52441.02441.52442.0dBm/m-100-90-80-70-60-50MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247 ePower Spectral DensityMiddle Channel (2440.11 Mhz)April 1, 2019International Certification Services, Inc.2440.11 MhzPower Spectral Density -54 dBm/mFreq (Mhz)2478.02478.52479.02479.52480.02480.52481.02481.52482.0dBm/m-100-90-80-70-60-50MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247 ePower Spectral DensityHigh Channel (2480.105 Mhz)April 1, 2019International Certification Services, Inc.2480.105 MhzPower Spectral Density -53.68 dBm/m 15.247 (f) Hybrid Systems Not Applicable For the purposes of this section, hybrid systems are those that employ a combination of both frequency hopping and digital modulation techniques. The frequency hopping operation of the hybrid system, with the direct sequence or digital modulation operation turned-off, shall have an average time of occupancy on any frequency not to exceed 0.4 seconds within a time period in seconds equal to the number of hopping frequencies employed multiplied by 0.4. The power spectral density conducted from the intentional radiator to the antenna due to the digital modulation operation of the hybrid system, with the frequency hopping operation turned off, shall not be greater than 8 dBm in any 3 kHz band during any time interval of continuous transmission. 24.0 15.247 (g) FHSS systems Not Applicable Frequency hopping spread spectrum systems are not required to employ all available hopping channels during each transmission. However, the system, consisting of both the transmitter and the receiver, must be designed to comply with all of the regulations in this section should the transmitter be presented with a continuous data (or information) stream. In addition, a system employing short transmission bursts must comply with the definition of a frequency hopping system and must distribute its transmissions over the minimum number of hopping channels specified in this section. 25.0 15.247 (h) FHSS Systems Not Applicable
(h) The incorporation of intelligence within a frequency hopping spread spectrum system that permits the system to recognize other users within the spectrum band so that it individually and independently chooses and adapts its hopsets to avoid hopping on occupied channels is permitted. The coordination of frequency hopping systems in any other manner for the express purpose of avoiding the simultaneous occupancy of individual hopping frequencies by multiple transmitters is not permitted. 20 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 26.0 15.247 (i) PASS Systems operating under the provisions of this section shall be operated in a manner that ensures that the public is not exposed to radio frequency energy levels in excess of the Commission's guidelines. See 1.1307(b)(1) of this chapter. 27.0 Lower Band Edge:
PASS 21 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.Freq (Mhz)24002401240224032404dBm/m-70-65-60-55-50-45-40-35-30MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247Lower Band Edge Low Channel (2402.159 Mhz)KDB558074 International Certification Services, Inc.April 1, 20192402.159 Mhz -32.96 dBm/m2400 Mhz -64.2 dBm/m 28.0 Upper Band Edge:
PASS 22 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.Freq (Mhz)2478247924802481248224832484dBm/m-80-75-70-65-60-55-50-45-40MX3 Diagnostics, Inc.Model: HYD-PRO-18FCC ID: 2ASD4-HYD-PRO-18FCC 15.247Upper Band Edge High Channel (2480.105 Mhz)KDB558074 International Certification Services, Inc.April 1, 20192480.105 Mhz -44.23 dBm/m2483.5 Mhz -75.6 dBm/m 29.0 Power Line Conducted Emissions:
PASS 30.0 SAR Requirements: PASS Using the equation from FCC KDB 447498 D01 General Exposure Guidance V06 Section 4.3.1 4.3.1 Standalone SAR test exclusion considerations a) For 100 Mhz to 6 Ghz and test separation distances <= 50 mm
(max power of channel, including tune-up tolerance, mw) X fGhz
(min test separation distance, mm) Low Frequency Channel: 2402.159 SAR =0.494 X 25.4 SAR = 0.03 2.402159 23 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.Freq (Mhz)0.1110dBuV01020304050607080MX3 DiagnosticsModel: HYD-PRO-18 (with Charger connected)FCC ID: 2ASD4-HYD-PRO-18FCC Part 15.207 Conducted Emissions (120 VAC, 60 Hz)Power Supply Qualcomm Model: PT-AC1UQC3 International Certification Services, Inc.April 1, 2019Quasi Peak LimitAverage LimitAverage Detected Data Middle Frequency Channel: 2440.11 SAR =0.379 X 25.4 SAR = 0.023 2.44011 High Frequency Channel: 2480.105 SAR =0.393 X 25.4 SAR = 0.024 2.480105 Channel (Frequency) Maximum SAR SAR Limit Power Spectral Density (mW) Low: 2402.159 Mhz Middle: 2440.11 Mhz High: 2480.105 0.494 0.379 0.393 0.03 0.023 0.024
<=3
<=3
<=3 24 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. 31.0 ATTACHMENT C Product Information Product Information Sheet as supplied by the Customer 25 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. COMPANY NAME: MX3 Diagnostics, Inc. CUSTOMER REPRESENTATIVE: International Certification Services, Inc. EQUIPMENT DESCRIPTION: Hydration Analyzer MODEL NUMBER: HYD-PRO-18 FCC ID: 2ASD4-HYD-PRO-18 SERIAL NUMBER: N/A TYPE OF TEST:
OSCILLATOR FREQUENCIES:
Bluetooth Radio: 2.4 to .48 Ghz Crystals: 32 Mhz, 32.768 Khz POWER INTERFACE:
Frequency: 50/60 Hz Voltage: 100-240 VAC POWER CABLE:
POWER SUPPLY:
Type: Switching Manufacturer: Qualcomm Model: PT-AC1UQC3 Development Initial Design Verification Design Change (Please describe exact changes below) X Production Sample (Audit Test) Hardwired Flexible Shielded Unshielded POWER LINE FILTER: None Current Removable CABINET SHIELDING PROVISION:
Plastic enclosure SOFTWARE AND / OR OPERATING MODES:
I-v2.0.1 (Rev 2) . INTERFACING EQUIPMENT OR SIMULATORS: None I/O CABLES: None 26 of 27 TEST REPORT 010419 InternationalCertification Services, Inc. MX3 Diagnostics, Inc. Model: HYD-PRO-18 FCC ID: 2ASD4-HYD-PRO-18 Power Supply: Qualcomm Model: PT-AC1UQC3 Test Configuration EUT AC Charger 120 VAC 60 Hz 27 of 27 TEST REPORT 010419 InternationalCertification Services, Inc.