Allocation of this Manual Product:
Automatic wheel loader scale WK60 / WK60-S / WK60-XS English (US) Language:
Translation of the original German manual WK60-RAD-KBA-A-(2016.10)-EN(US) First edition:
Associated appli-
cable document.: WK60-RAD-BA-*-(20**.**)-EN(US) Instruction manual as updated. Copyright We reserve all copyrights to this documentation. Improper use, especially copying and disclosure to third parties is not permitted. 1. Use of this Manual These brief instructions will describe the first steps to take when operating the WK60 wheel loader scale. Please refer to the Instruction Manual on the included CD for a comprehensive description of all the functions. The Manual is part of the product. Please store it for easy access. Danger symbols used in this Manual Danger point / hazardous situation 2. Safety notes 2.1 Designated use This scale is designed exclusively for weighing of bulk goods using:
Wheel loaders and similar utility vehicles (e. g. Forklift trucks or telehandlers with a lifting fork or loading bucket). The bulk goods should not have a tendency to stick. Any adhering material must be removed from the loading bucket after use. Structural changes to the scale and its components may result in a threat to life or damage to equipment. Structural changes to the scale and its components are prohibited. Conversion of and repairs to the scale may only be carried out by persons trained and authorized by the manufacturer. Only original spare parts and accessories approved by the manufacturer may be used. Please use only original PFREUNDT weighing tickets For optimal printing quality and maximum printer service life, please use only suitable Original PFREUNDT weighing tickets. You are welcome to phone us for advice. 1 Damages resulting from the use of weighing tickets other than the original PFREUNDT tickets intended for the specific printer shall be excluded from our warranty. Any use other than as designated shall be deemed incorrect. 2.2 FCC and IC Certifikation (only for WK60-XS) NOTICE This device complies with Part 15 of the FCC Rules and with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions:
this device may not cause harmful interference, and this device must accept any interference received, inclu-
ding interference that may cause undesired operation. Le prsent appareil est conforme aux CNR dIndustrie Canada applicables aux appareils radio exempts de licence. Lexploitation est autorise aux deux conditions suivantes:
(1) lappareil ne doit pas produire de brouillage, et
(2) lutilisateur de lappareil doit accepter tout brouil-
lage radiolectrique subi, mme si le brouillage est susceptible den compromettre le fonctionnement. NOTICE Changes or modifications made to this equipment not expressly approved by Pfreundt GmbH may void the FCC authorization to operate this equipment. Contains transmitter module FCC ID: PV7-WIBEAR11N-SF2 IC: 7738A-WB11NSF2 NOTE This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential instal-
lation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or te-
levision reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit dif-
ferent from that to which the receiver is connected. Consult the dealer or an experienced radio/TV techni-
cian for help. Radiofrequency radiation exposure Information This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance of 7,87/20 cm between the radiator and your body. This trans-
mitter must not be co-located or operating in conjunction with any other antenna or transmitter. This device IC: 21743-WK60XS has been approved by Industry Canada to operate with the antenna types listed below with the maximum permissible gain indicated. Antenna types not included in this list, having a gain greater than the maximum gain indicated for that type, are strictly prohibited for use with this device. Cet quipement est conforme aux limites dexposition aux rayonnements IC tablies pour un environnement non contrl. Cet quipement doit tre install et utilis avec un minimum de 7,87/20 cm de distance entre la source de rayonnement et votre corps. Ce transmetteur ne doit pas etre place au meme endroit ou utilise simultanement avec un autre transmetteur ou antenne. Cet metteur appareil IC: 21743-WK60XS t approuv par Industry Canada pour fonctionner avec les types dantenne numrs ci-dessous avec le gain maximum autoris et limpdance ncessaire pour chaque type dantenne indiqu. Les types dantenne ne figurant pas dans cette liste et ayant un gain suprieur au gain maximum indiqu pour ce type-
l sont strictement interdits dutilisation avec cet appareil. Antenna Delock 88395 Delock 88985 Gain 2dBi 0.7 dBi 2.3 Risks posed by incorrect use Incorrect use of the scale may lead to life-threatening situations. The scale has not been approved for operation in:
potentially explosive environments aggressive environments (e. g. in rooms with acid /
etching or galvanizing baths) in environments prone to stray currents or heat
(e. g. due to welding) Overloading transport vehicles is life-threatening. Obtain information on the permissible loading of such vehicles and strictly comply with such limits. Avoid incorrect weighing, e.g. by ensuring correct positioning of wheel loaders during weighing. 2.4 Training and qualification of users The WK60 wheel loader scale may only be used by trained persons. The company operating the scale must instruct the operating staff on safe handling of the scale. 3. Operating conditions suitable for correct weighing Carry out regular calibrations Scales of this type subject to mandatory calibration may only be used compliant with national statutory regulations of the country where they are used. The company operating the scale is responsible for regular calibration in accordance with national regulations. Official calibration checks are required after changes Official calibration checks are required immediately after changes to the lifting arms, lifting cylinders, switching points of proximity switches or the load carrier. Scales with loading buckets are not suited for very sticky goods ... ..which may shift the zero point by more than 0.5 e (calib-
ration value) between weighing buckets. Using the scale to weigh aggregate for building material preparation: (e. g. concrete mixing) Observe the applicable regulations on the accuracy of weig-
hing individual components. The lower limit of the minimum load is 10 e (e = calibration value / = the least significant digit of the scale). Only carry out weighing during smooth travel Carry out weighing whilst travelling smoothly, if possible, or in horizontal stationary position. A warning will be displayed and weighing will be blocked if the vehicle is not level enough. Weighing should only be carried out within the permitted temperature range:
Do not operate a scale outside its permissible tempe-
rature range, see the Technical data chapter in the Instruction Manual. When using external components (printer, etc.), obser-
ve the temperatures given in their specific data sheets. 4. Using the touch-screen Tap short tap with a finger:
to open applications to operate controls
(e.g. confirm entry) for entries via the on-screen keyboard Pressing and holding the desired button to call up available options
(e.g. to show context menus) Swiping or stroking (scrolling) to scroll through a long list to call up the previous / next page PFREUNDT GmbH | Brief operating instructions - Wheel loader scale WK60 / WK60-S / WK60-XS | WK60-RAD-KBA-B-(2007.04)-EN(US) 5. Software operation and navigation 7. Zeroing the scale The main pages a) b) c) d) Page 1 (example) The first main page appears when the scale is switched on. Structure of the main pages a) Metrological information field b) Field of application c) Page display d) Menu bar Depending on configuration, scale zeroing will be required in the following situations:
After switching on When completing the loading procedure Manually calling up zeroing You may at any time call up zeroing in the weighing program menu. 7.1 Calling up zeroing in the menu 1. Weighing with a dynamic scale 3. Move the lifting arms through the weighing positi-
on from the bottom up. 4. The measured weight is then prominently displayed and added to the total. Weighing position The General Menu bar The General Menu bar has four basic functions which are always shown in the same position in the first row. Left / Right continue to previous / next page 7.2 2. The WHEEL LOADER SCALE - Information page will open. 3. Open the menu and activate the function Reset to zero.
+
4. The scale zeroing prompt will then appear. The current total will not be deleted. Zeroing the scale Back to Main page (of an Apps first page) one step back (within an App) Menu Open/Close menu 6. Switching on the scale 1. Hold down the ON/OFF key for about 2 seconds. WK60 / WK60-S WK60-XS 2. The Starting screen will then appear. 3. The following actions will be required after switching on, depending on scale configuration:
Logging on a user *
Performing warm-up strokes *
Zeroing the scale
* see the Instruction Manual 1. Lower the lifting arms and tilt the empty loading bucket fully upward. 2. Zeroing a dynamic scale:
2.a Move the lifting arms through the weighing position from the bottom up. 2.b Zeroing will be automatic. 3. The first main page of the weighing program will then show. Weighing position If the buckets tare weight is outside the zeroing range of, a warning will be displayed. Proceed as follows in this case:
Check the bucket and remove any load residues. Repeat zeroing. 8. Carry out weighing 1. Lower the lifting arms and fill the bucket. 2. Tilt the loaded bucket fully upward. a) b) Description of the display:
a) Measured weight b) Total c) Number of weighings c) Brief operating instructions Wheel loader scale WK60 / WK60-S / WK60-XS Valid from program version: Version 1.1.60 Document no.: WK60-RAD-KBA-B-(2007.04)-EN(US) 5. Empty the loading bucket when you are prompted in the display. 6. Continue loading until the transport vehicle is full or the specified target weight has been reached. 7. Finish the loading procedure - see next chapter. 8.1 Finish loading 2. 1. Push the weighing slip into the printer * as soon as the vehicle has been loaded.
* if a printer is connected. If the master data are deacti-
vated, the weighing data will be automatically printed and loading will be complete. The page LOADING - Master data will automatically open with the most recently selec-
ted master data if the master data are activated. The LOADING - Master data page will open. Terminate loading without printer and master data 1. After the last weighing, tap on the Loading mode. 2. 3. 4. The weighing data will be automatically saved. 5. The main page of the scale will then be shown again. Tap on Finish loading. 2 PFREUNDT GmbH | Brief operating instructions - Wheel loader scale WK60 / WK60-S / WK60-XS | WK60-RAD-KBA-B-(2007.04)-EN(US) PFREUNDT GmbH | Robert-Bosch-Str. 5 | 46354 Sdlohn www.pfreundt.de | info@pfreundt.deSales +49(0)2862/9807-100 | Service +49(0)2862/9807-200 | Calibration service +49(0)2862/9807-400 | Weighing tickets +49(0)2862/9807-600 | Head office +49(0)2862/9807-0