PUBLIC MED-TAB v.1 USER MANUAL PUB INT EN 003TR Copyright 2003-2015 IMAGE Information Systems Ltd. Release date: 2015-09-02 MED-TAB v.1 User Manual PUB INT EN 003TR 1 4. i Q
-
V TABLE OF CONTENTS 1 Introduction ......................................................................................................... 3 1.1 Scope of the document ......................................................................................... 3 1.2 Summary ........................................................................................................... 3 Intended medical indication ............................................................................ 4 1.2.1 Intended patient population ............................................................................ 4 1.2.2 Intended users and user groups ...................................................................... 4 1.2.3 1.2.4 Designated Conditions of Use .......................................................................... 4 1.2.5 Contraindications ........................................................................................... 4 1.2.6 Exclusions, warnings and precautions ............................................................... 4 2 First steps............................................................................................................. 5 2.1 Unpacking .......................................................................................................... 5 2.2 MED-TAB layout .................................................................................................. 6 2.3 Keys and Buttons ................................................................................................ 8 2.4 Charge the battery .............................................................................................. 8 2.5 Care and cleaning instructions ............................................................................... 8 2.6 Inserting of a memory card (optional) .................................................................... 9 2.7 Leather Case ....................................................................................................... 9 2.8 Pen .................................................................................................................. 10 2.9 Grayscale adjustment according to DICOM Part 14 GSDF ........................................ 10 2.10 Climatic and ambient conditions .......................................................................... 11 3 Getting started ................................................................................................... 12 3.1 Working position................................................................................................ 12 3.2 Turning on or off ............................................................................................... 12 3.3 Switching between DICOM and regular display modes ............................................ 12 3.4 Optimizing the display lifetime ............................................................................ 12 3.5 Using the touch screen ....................................................................................... 12 3.6 Lock or unlock the touch screen and the keys ....................................................... 13 4 DICOM GSDF Grayscale Adjustment using CALCON ............................................ 14 4.1 Introduction ...................................................................................................... 14 4.2 Main menu ........................................................................................................ 14 4.3 DICOM GSDF grayscale image adjustment ............................................................ 15 4.4 Adjustment results............................................................................................. 17 4.5 Determnation of Ambient Light ............................................................................ 18 4.5.1 Getting started ............................................................................................ 18 4.5.2 Ambient light classification ............................................................................ 18 5 Troubleshooting ................................................................................................. 19 6 Safety information .............................................................................................. 20 6.1 Precautions ....................................................................................................... 20 6.2 Environmental information .................................................................................. 22 7 Index .................................................................................................................. 23 MED-TAB v.1 User Manual PUB INT EN 003TR 2 1. i Q
-
V 1 Introduction 1.1 Scope of the document This document represents the MED-TAB v.1 user manual provided by IMAGE Information Systems Ltd. Disclosure level of this document is PUB, which means that this document is freely available to anyone interested, such as resellers, current end users as well as potential customers. Primary color is orange. NOTICE:
The provided data was researched with the utmost care. Nevertheless, the author does not accept any liability for the correctness, completeness and up-to-date nature of the information contained in this document. It is the responsibility of the user or the local administrator to verify the details by considering suitable sources. 1.2 Summary MED-TAB is a 13.3 color LCD monitor including built in tablet computer for viewing and analysis of medical images other than those of mammography from a picture archiving and communication system (PACS). The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 1920 x 1080 pixels (2MP). There is one REGULAR display MODE to visualize non-medical images and one DICOM MODE factory calibrated display mode to display medical images according to DICOM GSDF regulations. CALCON is application software which is preinstalled in each tablet offering quality control of the monitor. The software supports quantitative and visual tests. The MED-TAB has been precisely calibrated to comply with DICOM Part 14 at factory. With built-in brightness stabilization circuit, USB adjustment tool and ambient light sensor, stable brightness and persistent calibration can be guaranteed. The MED-TAB can support both landscape and portrait mode. The anti-reflection coated protective screen can prevent display from damage under heavy operating conditions. The tablet-PC is designed in conformity with IEC60601-1, IEC60601-1-2 and CCC standards. CAUTION Danger of wrong image analysis results. Read the entire manual in full detail to ensure safe and proper use, before using the device first time. Incorrect use due to insufficient environmental conditions or incompatible software might lead to incorrect results of image analysis. MED-TAB v.1 User Manual PUB INT EN 003TR 3 1. i Q
-
V 1.2.1 Intended medical indication MED-TAB is a 13.3 color LCD monitor including built in tablet computer for viewing and analysis of medical images other than those of mammography from a picture archiving and communication system (PACS). 1.2.2 Intended patient population The device is not intended for use by a patient. 1.2.3 Intended users and user groups Users of the MED-TAB are trained medical practitioners or veterinarians. Patients shall not touch the device. 1.2.4 Designated Conditions of Use Designated conditions of use are:
Only intended for medical healthcare professionals. It is recommended to place the device on a desk using the rotating case stand either in portrait or landscape mode without direct flare on the screen. The ambient should be less than 100 lux for image analysis. The intention of the device is for repeated but not continuous use. The operating Temperature is +10C to +35C. The operating air pressure is 600 hPa to 1100 hPa. 1.2.5 Contraindications The MED-TAB is not intended for:
Mammography for diagnostic use. Use at ambient light of more than 1000 lux. Sterile use. Use in an infectious environment. 1.2.6 Exclusions, warnings and precautions Perform a medical display acceptance test with the PACS before clinical use according to local laws and regulations using a calibrated luminance meter. No use for emergency procedures in case of unreliable data connections. No handheld use for image analysis. Regularly clean the device to remove dirt. Only use a pen or a mouse to perform measurements using a PACS. Do not use fingers. MED-TAB v.1 User Manual PUB INT EN 003TR 4 1. i Q
-
V 2 First steps 2.1 Unpacking Please check the product box for the following items:
MED-TAB v.1 MED-TAB CALIBRATOR with a Micro USB plug Leather case with rotating stand Touch pen Switching Power Adaptor for MED-TAB v.1 incl. plugs for US, UK, EU and CN Micro USB to USB cable USB OTG cable Warranty card Quick start guide We recommend to operate the MED-TAB using the leather case to protect the device. CAUTION Danger of incorrect image analysis results. Always transport MED-TAB in the original case or in the original packaging box. The antiglare sheet on the surface of the tablet is soft. Scratches can impair the image quality. MED-TAB v.1 User Manual PUB INT EN 003TR 5 1. i Q
-
V 2.2 MED-TAB layout Power/Lock Button Volume key DICOM-Switch Ambient light sensor Front camera Touch screen 6 1. i Q
-
V MED-TAB v.1 User Manual PUB INT EN 003TR Camera DICOM Switch Volume key Power/Lock Button Headset jack Micro SD Slot Mini HDMI connector USB connector Microphone Kensington lock Power connector MED-TAB v.1 User Manual PUB INT EN 003TR 7 iQ-
VIE W 2.3 Keys and Buttons Key Action Power/Lock To turn on/off the device (press and hold); to lock/unlock Volume For adjusting the volume of the device DICOM-Switch To activate/deactivate the DICOM-Mode 2.4 Charge the battery Before using the MED-TAB for the first time, charge the battery for 8 hours. Danger of electrical shock, fire and device malfunction. WARNING Always use the Switching Power Adaptor for MED-TAB v.1 for charging and operation. Make sure the power connector is fully connected to the power socket. The power adaptor must be used at specified electric voltage, only. CAUTION Danger of delayed treatment due to unavailable image analysis device. A warning will appear along with an alert when the battery charge is low. The battery must be charged for wireless operation. Once the battery is completely discharged, the device cannot be turned on until the battery was charged for a few minutes until the battery has been charged. 2.5 Care and cleaning instructions CAUTION Danger of wrong image analysis. Frequent cleaning of the display using a display cleaner is recommended. MED-TAB v.1 User Manual PUB INT EN 003TR 8 4. i Q
-
V Finger prints and dirt might lead to incorrect image analysis results. Use of rough tissues for cleaning might damage the surface which can impair the image quality. Incorrect display cleaners might dull the surface and impair the image quality. Do a new image acceptance test if you are unsure after any damage of the surface. Periodic cleaning is recommended to keep the display clean and to prolong its operation lifetime. Never use thinner, benzene, alcohol (ethanol, methanol, or isopropyl alcohol), abrasive cleaners, acid, steel wool or other strong solvents, as these may cause damage to the cabinet or LCD panel. The surface can be cleaned with a soft display-cloth, such as cotton or lens paper and special display cleaner. If necessary, stubborn stains can be removed by moistening part of a cloth with water to enhance its cleaning power. CAUTION Danger of incorrect image analysis. Do not scratch the display surface. Be careful with rings, keys or jewelry. Do not use other touch pens than the original one. Scratches might damage the display and impair the image quality. Do a new image acceptance test if you are unsure after damaging the surface. 2.6 Inserting of a memory card (optional) The MED-TAB only accepts microSDTM and microSDHC memory cards with a maximum capacity of 32GB. Not all memory cards might be compatible with MED-TAB. MED-TAB only supports the FAT file system. If you insert a card with a different file structure, the device will ask to reformat the memory card before use. The file directory of inserted memory cards appears in the extSdCard folder of the MED-tAB. If you want to remove a memory card, please be ensure that device is not currently accessing the memory card. Open Settings in the application list, choose the category Storage and select Unmount SD card. After confirming the process, you can unmount the memory card. 2.7 Leather Case The MED-TAB should be used sheltered in its leather case. The business card cover is thought to identify the owner or user of a device, e.g. by inserting a business card. The rotating stand of the MED-TAB can be turned over 360 to enable both landscape and portrait views. MED-TAB v.1 User Manual PUB INT EN 003TR 9 1. i Q
-
V 2.8 Pen The supplied pen allows a dirt-free usage of the MED-TAB and hence preserves its image quality. CAUTION Danger of incorrect image analysis. Use the original touch pen to control the MED-TAB for image analysis to avoid image quality impair by finger prints. Wash your hands before using MED-TAB to reduce oil remaining on the surface. 2.9 Grayscale adjustment according to DICOM Part 14 GSDF The MED-TAB Calibrator is a grayscale adjustment tool. The grayscale adjustment tool adjusts in combination with the MED-TAB CALCON app the MED-TAB to comply with the DICOM Part 14 GSDF specification (see chapter 4 CALCON - Grayscale adjustment tool). MED-TAB v.1 User Manual PUB INT EN 003TR 10 1. i Q
-
V 2.10 Climatic and ambient conditions Key Operating Temperature Air pressure Operating Humidity Storage Temperature Action
+10C~+35C 600 hPa ~ 1100 hPa 30% ~ 95% Non-condensing
-10C ~ +55C Storage/Transportation Humidity 10% ~ 95% Non-condensing Magnetic resistance
< 0,25 T MED-TAB v.1 User Manual PUB INT EN 003TR 11 1. i Q
-
V 3 Getting started 3.1 Working position CAUTION Danger of incorrect image analysis. Use a viewing angle of around 90 for image analysis. Do not operate in direct light to avoid glary reflections which can impair the image quality. Verify that the ambient light is less than 100 lux for therapeutic decisions. Use a 3rd party calibrated lux meter to determine ambient light. The CALCON app provides an indicator for the ambient light using the built in ambient light sensor. 3.2 Turning on or off To turn on the MED-TAB, please press and hold the Power button. To turn off the MED-TAB, please press and hold the Power button for 3 seconds and then select Power off. Confirm the shutdown by OK. 3.3 Switching between DICOM and regular display modes Use the hardware DICOM Switch at the long edge of MED-TAB to switch the DICOM mode on or off. CAUTION Danger of incorrect image analysis. Use the DICOM mode for grayscale image analysis and the REGULAR mode for any non-DICOM applications. Use of the incorrect display mode will impair the image quality. 3.4 Optimizing the display lifetime In case of working with the same image or an application with static image elements for more than 4 hours continuously, it is recommended to change the image content regularly to avoid image retention of the static elements. NOTICE:
Risk of display damage if continuously displaying the same image for more than a few hours. We recommend to set the display sleep timer to 5 minutes of nonuse to protect the screen. 3.5 Using the touch screen The Med-TAB touch screen provides an easy opportunity/possibility to select items or perform functions. MED-TAB v.1 User Manual PUB INT EN 003TR 12 1. i Q
-
V CAUTION Danger of incorrect image analysis. Only use the original touch pen. Only use the touch pen or a mouse for any measurements with a connected PACS. Never use fingers or other tools due to inaccurate measurements. 3.6 Lock or unlock the touch screen and the keys To lock/unlock the touch screen and the buttons manually, press the Power button. MED-TAB v.1 User Manual PUB INT EN 003TR 13 1. i Q
-
V 4 DICOM GSDF Grayscale Adjustment using CALCON 4.1 Introduction MED-TAB CALCON software adjusts the MED-TAB to comply with DICOM Part 14 GSDF specification using the MED-TAB CALIBRATOR. CAUTION Danger of incorrect image analysis. Do an acceptance test of the MED-TAB before first clinical use and periodically according to local rules and regulations. The MED-TAB CALIBRATOR is a grayscale adjustment tool according to DICOM part 14 GSDF but not a measurement tool according to ISO 13485, DIN 6868-157 or other applicable rules and regulations. Introduction to grayscale adjustment procedure: The adjustment is based on a patented technology of generating 11-bit (2048) grayscales on a color LCD panel. The MED-TAB CALCON communicates with the calibration ASIC (application specific integrated circuit) to generate 2048(211) highly accurate grayscales first. For each grayscale generated by CALCON, the MED-TAB CALIBRATOR measures the luminance of each grayscale. After collecting all 2048 luminances of all grayscales, CALCON runs a Barten model to select 256 (28) grayscales which match closest to the Barten standard characteristic function curve. Those 256 values are used to set the lookup table for medical image display. After certain time of the operation of the MED-TAB, the characteristic function curve would shift. Thus, device administrators shall run a readjustment at least after 10.000 hours or after failure of an image display acceptance test or according to local rules and regulations, e.g. in Germany according to DIN 6868-157, in Switzerland according to Weisung R 08 06, in US according to AAPM TG 18, in Japan according to JESRA regulations. Use the MED-TAB CALCON and CALIBRATOR to adjust the MED-TAB again, as part of PACS maintenance to keep all MED-TABs DICOM compliant. Run an acceptance test after each adjustment procedure. 4.2 Main menu Key CAL FILE VIEW IMAGE INSPECT CHECK Action Start or stop the grayscale adjustment process Open or save adjustment results, only used by customer service agents View the adjustment result View test pictures and patterns Only used by customer service engineers Only used by customer service engineers MED-TAB v.1 User Manual PUB INT EN 003TR 14 1. i Q
-
V Key SETTING HELP Action Ambient Light Sensor setting User help and end-user software license agreement 4.3 DICOM GSDF grayscale image adjustment Prepare the MED-TAB CALIBRATOR. In order to adjust the MED-TAB, please ensure that the battery of MED-TAB is charged for
>50%, or connected with the Switching Power Adaptor for MED-TAB v.1 to the power supply. Step Picture/Screenshot 1. Connect the MED-TAB CALIBRATOR with the MED-TAB via Micro USB 2. Open the CALCON app and touch the CAL-button MED-TAB v.1 User Manual PUB INT EN 003TR 15 1. i Q
-
V Step Picture/Screenshot 3. Confirm the permission request by selecting OK 4. Place the MED-TAB CALIBRATOR on the center of the MED-TAB display and press START 5. The app now generates 2048 grayscale shades for adjustment. After the adjustment, the app will indicate to save the LUX and LCD table. 6. Click OK to finish the DICOM GSDF grayscale adjustment. MED-TAB v.1 User Manual PUB INT EN 003TR 16 1. i Q
-
V 4.4 Adjustment results Step Picture/Screenshot 1. After the adjustment, users can view the results by selection VIEW. 2. MED-TAB v.1 User Manual PUB INT EN 003TR 17 1. i Q
-
V 4.5 Determnation of Ambient Light Applicable rules and regulations, e.g. DIN 6868-157, define various reading conditions according to maximum ambient light. 4.5.1 Getting started To start the visualization of the ambient light, start the CALCON app. A colored circle appears, showing the estimated ambient light in Lux, e.g.:
This button indicates the ambient light and can be moved anywhere on the screen. Just tap on it and hold to move it freely all over the display. The button stays visible even when CALCON is minimized. 4.5.2 Ambient light classification Ambient light Classification
<50 lx
<100 lx
<500 lx
<1000 lx
>1000 lx Green: diagnostic reading conditions Yellow/Green: Conditions for therapy relevant decisions Yellow: Conditions to perform medical procedures Yellow/Red: viewing conditions Red: no viewing conditions MED-TAB v.1 User Manual PUB INT EN 003TR 18 1. i Q
-
V 5 Troubleshooting Problems Points to check with possible solutions The screen has defective pixels (e.g. slightly light or dark). This is due to the characteristics of the panel itself, not the LCD monitor. MED-TAB v.1 User Manual PUB INT EN 003TR 19 1. i Q
-
V 6 Safety information To reduce the risk of fire or electric shock, do not expose this apparatus to rain or moisture. The MED-TAB is primarily for use in a health care facilities without direct contact to patients (no applied part). Do not connect this apparatus to household or consumer devices. We will try the best to assure the newest information be shown in this manual, but the information in this document is subject to change without prior notice. 6.1 Precautions If the device emits smoke, smells like something is burning, or makes strange noise, disconnect all power connections immediately and contact your dealer for advice. Attempting to use a malfunctioning unit may result in fire, electric shock, or damage of the device. Do not open the case or modify it. Opening the case or modifying the device may result in fire, electric shock, or burn. Refer all service to manufacturer authorized qualified service personnel. Do not attempt to service this product yourself as opening or removing covers may result in fire, electric shock, or equipment damage. If an object or liquid falls/spills into the cabinet, unplug the unit immediately. Have the unit checked by a qualified service engineer before using it again. Set up the unit in an appropriate location. Not doing so may result in fire, electric shock, or damage of the device. Do not place the unit in a dusty or humid environment. Do not place the unit in a location where steam flows directly to the device. Do not place the unit near heat generating devices or a humidifier. Do not place the unit in an inflammable gas environment. To avoid danger of suffocation, keep the plastic packing bags away from babies and children. To disconnect the power cord, grasp the plug firmly and pull. Tugging on the cord may damage and result in fire or electric shock. Use the correct voltage. The unit is designed for use with a specific voltage only. Connection to another voltage than specified in this User Manual may cause fire, electric shock, or equipment damage. Do not overload your power circuit, as this may result in fire or electric shock. MED-TAB v.1 User Manual PUB INT EN 003TR 20 1. i Q
-
V Handle the power cord with care. If the power cord is broken, stop to use it immediately. Using broken power cord may result in fire or electric shock. Do not put the power cord under the unit or heavy objects. Do not pull up or twist the power cord. Never touch the plug and power cord if it begins to thunder. Touching them may result in electric shock. Do not touch a damaged LCD panel directly with bare hands. The liquid crystal that may leak from the panel is poisonous if it enters the eyes or mouth. If any part of the skin or body comes in direct contact with the panel, please wash thoroughly. If some physical symptoms result, please consult your doctor. Follow local regulation or laws for safe disposal. Handle with care when carrying the unit. Disconnect the power cord and data cables and remove the optional unit. Moving the unit with the cord or the option attached is dangerous. It may result in injury. Do not block the ventilation slots of the cabinet. Do not place any objects on the ventilation slots. Do not install the unit in a closed space. Do not use the unit laid down or upside down. Blocking the ventilation slots prevents proper airflow and may result in fire, electric shock, or equipment damage. Do not touch the plug with wet hands. Doing so may result in electric shock. Periodically clean the area around the plug. Dust, water, or oil on the plug may result in fire. Unplug the unit before cleaning it. Cleaning the unit while it is plugged into a power outlet may result in electric shock. If you plan to leave the unit unused for an extended period, disconnect the power cord from the wall socket after turning off the power switch for the safety and the power conservation. When the MED-TAB is cold and brought into a room or the room temperature goes up quickly, dew condensation may occur inside and outside the display. In that case, do not turn the MED-TAB on and wait until dew condensation disappears, otherwise it may cause some damages to it. Do not scratch or press on the panel with any sharp objects, such as a pencil or pen
(except the supplied Pen) as this may result in damage to the panel. Do not attempt to brush with tissues as this may scratch the LCD panel. MED-TAB v.1 User Manual PUB INT EN 003TR 21 1. i Q
-
V NOTICE:
Danger of patient data breach. We recommend you do not store patient data on the MED-TAB or an inserted memory card. The MED-TAB file system must be encrypted if patient data will be stored on the device. Ask the manufacturer for further details. We recommend you use up-to-date anti-virus software on the MED-TAB. The virus definitions must be updated regularly (they should not be older than 2 weeks). Older virus definitions might lead to data breaches. 6.2 Environmental information The MED-TAB and its battery must not be disposed of with other municipal waste. Please dispose the device and its accessories by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment. For more information about recycling of this product, please contact your municipal waste disposal service or local city office. MED-TAB v.1 User Manual PUB INT EN 003TR 22 1. i Q
-
V 7 Index B S Battery ................................................. 8 Buttons ................................................ 8 Scope of the document........................... 3 Storage conditions ...................... 9, 10, 11 C T Care instructions ................................... 8 K Touch screen ................................. 12, 17 Troubleshooting .................................. 19 Keys .................................................... 8 U L Unpacking ............................................ 5 Layout ................................................. 6 Lock screen .........................................13 W Wi-Fi ............................................... 4, 18 M Memory card ......................................... 9 MED-TAB v.1 User Manual PUB INT EN 003TR 23 1. i Q
-
V This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. Caution: Any changes or modications to this device not explicitly approved by manufacturer could void your authority to operate this equipment. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. IMAGE INFORMATION SYSTEMS LTD. 3RD FLOOR | 207 REGENT STREET | LONDON W1B 3HH | UNITED KINGDOM TEL. UK: +44 207 193 06 20 | TEL. GER: +49 381 496 58 20 | TEL. US: +1 704 323 66 63 FAX UK: +44 207 976 48 97 | FAX GER: +49 381 496 5 82 99 WWW.IMAGE-SYSTEMS.BIZ | INFO@IMAGE-SYSTEMS.BIZ MED-TAB v.1 User Manual PUB INT EN 003TR 24 1. i Q
-