Version 0 Date: 2014-01-02 1417WGB User / Installation Manual R-USM-503 1 Copyright 2014, Rayence Co. Ltd. All pages of this document contain proprietary and confidential information of Rayence Corporation and are intended for exclusive use by Rayence Corporation personnel or customers. Copying, disclosure to others or other use is prohibited without the express written authorization from Rayence Corporation. Please report any violations of requirement to Rayence Corporation. R-USM-503 2 Attention For improvement of product performance, supplementation, or follow-up of information; the contents of this manual are subject to change without separate prior notice. Please note that our company has neither responsibility for any accidents nor obligation to do free repair service for any damage of the equipment due to user's mistake, which resulted from failure to follow the contents in this manual. Make sure to be familiar with the safety precautions and usage procedures. Also note that the product may slightly differ from the contents of this manual depending on specification. The following marks are used for the effective use of the product in this manual. Attention, consult accompanying documents. This is used to emphasize essential information. Be sure to read this information to avoid incorrect operation. This indicates hazardous situation which, if not heeded, may result in minor or moderate injury to you or others, or may result in machine damage. This indicates a potentially hazardous situation which, if not heeded, could result in death or serious injury to you or others. Federal Law restricts this device to sale by or the order of a radiologist or any other practitioners licensed by the law of the state in which that person practices to use or order the use of the device. R-USM-503 3 Contents 1. Introduction...................................................................................... 5
(1) Overview .......................................................................................... 5
(2) Product features .............................................................................. 5
(3) Product component ........................................................................ 6
(4) Components Description ............................................................... 8
(5) Warning .......................................................................................... 11
(6) Caution ........................................................................................... 13 2. Notes for Using the Detector ........................................................ 15
(1) Handling ......................................................................................... 15
(2) Handle Assembly .......................................................................... 15
(3) Before Exposure............................................................................ 16
(4) During Exposure ........................................................................... 16
(5) Disinfection and Cleaning ............................................................ 16
(6) Others ............................................................................................. 16 3. Technical Features ........................................................................ 17
(1) Mechanical Features ..................................................................... 17
(2) Electrical Features ........................................................................ 17
(3) Environmental requirement ......................................................... 19
(4) PC Requirements .......................................................................... 20 4. Installation ..................................................................................... 21
(1) Program setup ............................................................................... 21
(2) Connection (Manual Trigger) ....................................................... 22
(3) Connection (Auto Trigger) ............................................................ 30
(4) IP set up ......................................................................................... 34
(5) Checking Connection ................................................................... 34
(6) Using Web Manager (IP, SSID Change / Upgrade FW) .............. 35 5. Calibration ...................................................................................... 39
(1) General Principle ........................................................................... 39
(2) Calibration ...................................................................................... 41 6. Image Acquisition Test.................................................................. 46
(1) Get Image ....................................................................................... 46
(2) View Images ................................................................................... 47 7. Operation ........................................................................................ 52
(1) Recommend ................................................................................... 52
(2) Switching power on / off ............................................................... 52
(3) Storage ........................................................................................... 52 8. Radio Frequency compliance ........................................................ 53
(1) FCC Notice (U.S.A) ........................................................................ 53
(2) IC Notice (CANADA) ...................................................................... 55
(3) R&TTE Notice (European Union) ................................................. 56 9. Maintenance .................................................................................. 57
(1) Maintenance ................................................................................... 57
(2) Cleaning ......................................................................................... 57
(3) Inspection ...................................................................................... 58
(4) Marking and labeling symbols ..................................................... 59 Appendix .............................................................................................. 61 R-USM-503 4 1. Introduction
(1) Overview The 1417WGB is a digital X-ray flat panel detector that can generate images of NDT. This X-ray imaging system consists of a scintillator directly coupled to an a-Si TFT sensor. It makes high-resolution, high-sensitive digital images.
(2) Product features Based on a-Si TFT active matrix Compact (18mm thickness) and light weight High resolution : Min. 2.5 lp/mm, Max 3.9 lp/mm 14-bit digital output Easy integration R-USM-503 5
(3) Product component Industrial Image Processing Unit Photo Item Part Name Quantity Detector RD1417WGB 1 Handle
-
1 Battery pack RB37WH 2 Battery charger RC120W 1 Charger adapter PMP120-13-3 1 AGI RA001A 1 Power box supply RS1417 1 R-USM-503 6 Cable Item Part Name Length Quantity Link cable VRH076A 6m 1 P-
Interface cable USB cable(A to B) VRH017A 8m 1 VRH078A 1.8m 1 LAN cable VRH027A 10m 1 AC Power cord VRH018A/019A 1.8m 1 Installation CD o Manual o Detector Library Option o Trigger cable R-USM-503 7
(4) Components Description The detector is designed to capture radiographic images. Captured images are transmitted to PC using the wired connection.
(1) Detector 1. LED indicator: The LED indicates the state of detector. i. Green: Power ii. Blue: Link iii. Orange: Low battery 2. Handle : Hold this handle when carrying the sensor unit. It is removable. (Horizon) 3. Handle unlock-lever : This is an unlock-lever to remove handle. 4. Link cable connector : This is a connector for Wire communication. Connect to provided link cable and another end of the cable connects to the PC. 5. CFRP(Carbon Fiber Reinforced Plastic) : The object to which an image is to be taken should be placed against this plate.. R-USM-503 8
(2) Power supply 1. Power plug connector: Connect with AC power supply cord 2. LAN connector : Ethernet port for transmitting the image/command between detector and PC 3. Link cable connector: Power connector for detector operating 4. LED indicator: Four LEDs to display the status Changeable part: Fuse T3.15 AL 250V
(3) AGI
<AGI is needed on only Manual Trigger Mode>
1. Trigger connector: This is a connector for synchronization between detector and generator. Connect the AGI to the generator using P-interface cable.. 2. USB connector: This is a connector for communication between AGI and PC. Connect the AGI to the PC using USB cable R-USM-503 9
(4) Battery & Charger 1. Battery : Lithium ion battery. The number of times being acquired image is 960 images(@ cycle time : 15s). The batteries last 2.5 hours and are rechargeable. 2. Charger : Two port cradle type. 3. LED indicator :
i. Orange on : charging ii. Green on : Charging is completed. R-USM-503 10
(5) Warning Environment of Use and Storage Follow the specified process of operational instructions written in this manual for the safety of the users Does not use or store the instrument near any flammable chemicals such as thinner, benzene, etc. Also, this instrument is not a category AP or APG equipment. If chemicals are spilled or evaporate, it may result in fire or electric shock through contact with electric parts inside the instruments. Also, some disinfectants are flammable. Be sure to take care when using them. Connection Do not connect the instrument with anything other than specified. Otherwise, it may result in fire or electric shock. To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth. Handling Always be sure to keep checking the condition of the system to ensure they are normal during the use of the instrument. If any problem is found, take appropriate measures, such as stopping the operation of the instrument, as required. Never disassemble or modify the product as it may result in fire or electric shock. Also, since the instrument incorporates parts that may cause electric shocks and other hazardous parts, touching them may cause death or serious injury. Do not hit or drop the instrument. The instrument may be damaged if it receives a strong jolt, which may result in fire or electric shock if the instrument is used without it being repaired. When Problem Occurs R-USM-503 11 Make sure to observe the following right Should any of the following occur, immediately turn OFF the power of each instruments, unplug the power supply cord from the AC outlet, and contact Rayence representative or distributor. When there is smoke, odd smell or abnormal sound. When liquid has been spilled into the instrument or a metal object has entered through an opening. When the instrument has been dropped and it is damaged. Maintenance and Inspection For safety reasons, be sure to turn OFF the power of each instrument when the inspections indicated in this manual are going to be performed. Otherwise, it may result in electric shock. When the instrument is going to be cleaned, be sure to turn OFF the power of each instrument, and unplug the power supply cord from the AC outlet. Never use benzene, thinner or any other flammable cleaning agents. Otherwise, fire or electric shock may result. The instrument must be repaired by a qualified engineer only. If it is not repaired properly, it may cause fire, electric shock, or accident. R-USM-503 12 Make sure to observe the following right.
(6) Caution Environment of Use and Storage Do not install the instrument in a location with the conditions listed below. Otherwise, it may result in failure or malfunction, fall or cause fire or injury. Close to facilities where water is used. Close to air-conditioner or ventilation equipment. Close to heat source such as a heater. Prone to vibration. Insecure place. Dusty environment. Saline or sulfurous environment. High temperature or humidity. Freezing or condensation. Do not place the storage case in a location with the conditions listed below. Where the cable of the sensor unit will be strongly pulled when the sensor unit is put into the case, otherwise, the cable may be damaged, resulting in fire or electric shock. Where someone might get their foot caught in the cable of the sensor unit is put in the case. Otherwise they could trip over, resulting in injury R-USM-503 13 Handling Wipe the CFRP plate of the sensor unit with ethanol or glutaraldehyde solution Turn off the power of each instrument for safety when they are not going to be used. Maintenance and Inspection For safety reasons, be sure to inspect the instrument before using it. In addition, carry out a regular inspection at least once a year. If the device does not work, do not disassemble yourself. Please contact to the Rayence C/S team. R-USM-503 14 2. Notes for Using the Detector
(1) Handling Handle the instrument carefully, as it may be damaged if something is hit against it, dropped, or receives a strong jolt.
(2) Handle Assembly Please confirm lever position. When the handle is locked, the lever should be same position with the right picture. If the handle is not locked, the detector can be dropped. Insert the handle in the same direction Insert the handle always in the same R-USM-503 15
(3) Before Exposure Be sure to check the equipment daily and confirm that it works properly. Sudden heating of the room in cold areas will cause condensation to form on the instrument. In this case, wait until condensation disappears before performing exposure. If the instrument is used with condensation formed on it, problems may occur in the quality of the instrument. When an air-
conditioner is going to be used, be sure to raise/lower the temperature gradually so that a difference in temperature in the room and in the instrument does not occur, to prevent forming of condensation.
(4) During Exposure Do not use the detector near devices generating a strong magnetic field. Doing so may produce image noise or artifacts. Do not apply excessive weight to the sensor unit. Otherwise, the sensor may be damaged.
(5) Disinfection and Cleaning Do not spray the detector directly with disinfectants or detergents. Do not use anything other than neutral detergent for cleaning the cover of the instrument. Otherwise, the coating will be corroded.
(6) Others Be sure to reconnect the cables to the proper connectors. Otherwise, the instrument may malfunction or may be damaged. R-USM-503 16 3. Technical Features
(1) Mechanical Features Size Weight 475 x 399 x 18 mm 5.3 kg Encapsulation Material Mg Window Material Carbon fiber plate
(2) Electrical Features Detector Sensor Type Amorphous Silicon with TFT (Single Panel) X-ray Converter Gd2O2S:Tb Total Pixel Number 3328 2816 pixels Total Pixel Area 422.7 x 357.6 mm Effective Pixel Number 3268 2756 pixels Effective Pixel Area 415.0 x 350.0 mm Pixel Pitch 127 m Limiting Resolution Max. 3.9 lp/mm Energy Range 40 - 150 kV A/D Conversion 14 bits Preview Time Data Interface Table 1: Product specification
< 2.0 sec (wired)
< 5.0 sec (wireless) Gigabit Ethernet Wi-Fi (802.11a/b/g/n) Under RQA5 condition (70kVp, 21mmAl) Preview time is not included post-processing time R-USM-503 17 Battery Size Weight Input Output 232.5 x 132.8 x 7 mm Typ. 0.3 kg 12.5 VDC 11.1 VDC Charging time Typ. 2.5 hours Capacity Typ. 3400 mAh The number of times being acquired image 960 images (@cycle time = 15s) The Battery level can be displayed on the LED status of battery. If the battery level goes down under 25%, please charge the battery Display Battery level 75~100 %
50~75 %
25~50 %
0~25 %
267.5 x 100 x 77 mm Typ. 0.86 kg 20 VDC 11.1 VDC 160 x 76 x 43.7mm (cable length: 1.28m) Typ. 0.92 kg 100~240 VAC, 47~63 Hz, 1.4~0.6A 20 VDC, Max 6.0A R-USM-503 18 Charger Size Weight Input Output Adapter Size Weight Input Output Wireless Spec Standard Without DFS (5.25GH to 5.35GHz and 5.47 802.11a/b/g/n compliance Peak Rate Frequency Bandwidth MIMO to 5.725) Band 300Mbps 2.4 GHz / 5 GHz 20MHz/40MHz 2x2 Recommended Maximum operable distance : 7m
(From the Access Point) 5150~5250 MHz band is restricted to indoor operations only. (for FCC) 5150~5250 MHz band is restricted to indoor operations only. (for CE) 5150~5250 MHz is indoor use only. (for Japan)
(3) Environmental requirement Item Min Typ. Max. Unit Note Temperature 5 Humidity 30 Pressure 70 40 75
%
H.R. 106 kPa Vibrations
(Wrapping condition) 2G
(8G) 10-150Hz, 10Sweeps, 1min/Octa ve, XYZ axis R-USM-503 19
(4) PC Requirements Processor: At least Intel Pentium IV HT with 2.8GHz, Intel Core Duo / Core 2 or comparable AMD Dual Core processor At least 3GB of RAM requirement (4GB for 32 BITS OS and 8GB for 64 BITS OS recommended) At least 500GB for application and archiving. Recommended 500GB for applications and secondary drive of 1TB for image archiving.. Dual 10/100/1000 network card system required. One for network (Internet) and one for the DR Panel communication Graphics card / monitor: Resolution of at least 1,600 x 900 for desktop and 1366 x 768 for laptop. For diagnostics purpose we recommend 1920 x 1080 resolution (2 mega pixels) monitor Microsoft Windows 7 32BIT/64BIT No antivirus except for Microsoft Security Essentials. R-USM-503 20 4. Installation Portable Imaging processing unit must be installed in a way that enables the user to achieve optimal use The Detector is composed of sensitive electronic parts and components. It is recommended to use the product in a clean place and to exercise caution to ensure that it is not affected by dust or liquids. It is recommended to Use a dry and soft cloth to clean the detector housing.
(1) Program setup To acquire images, run _vadav.lnk program. Please set the following figures Detectors IP : Wireless-192.168.1.80, Wired-192.168.1.81 Detectors number of ADC : 0 Detectors size of image : 3328 x 2816 21 R-USM-503
(2) Connection (Manual Trigger) Power Connection A. Connect the battery pack or power cable to the equipment.
* In wired mode, the frame ground is necessary. B. Connect the USB cable from your PC to AGI. Be sure to sure only the dedicated battery pack, RB37WH for 1417WGB. R-USM-503 22 Wireless Communication A. AP Router(Line sharer) setting
- SSID : Griffon
- Internal network
- IP address : 192.168.1.80
- Subnet mask : 255.255.255.0
- Dynamic IP allocation range
: 192.168.1.2 ~
192.168.1.254
- Pre-Shared Key(Password) : project302
- Authentication methods : WPAPSK or WPA2PSK
- Password methods : TKIP/AES
- AP IP : 192.168.1.1
- Channel (Frequency)
- Avoid crowded channel
(Using wireless detector under crowded channel result in low performance)
- If available, Use Auto-Channel Selection function of router to find optimal channel B. Reception Indicator Link LED flickering Blink Speed : Slow Low link quality Fast High link quality R-USM-503 23 C. Checking Link Quality & Battery Remain
- After wireless connection is established, perform Get Bright in Calibration tap.
- Check the value named Wireless Signal in black log screen. Wireless Signal = Link Quality (Max. 100)
- Check the value named Battery Remain in black log screen. Battery Remain = Battery Remain(Max. 100) The value Wireless Signal will be displayed here R-USM-503 24 Software Setting R-USM-503 25 Trigger Connection A. Connect the P-interface cable or trigger cable to the generator X-ray Generator Connection A. Mode 1 : P-interface cable mode Connect the P-interface cable between the AGI box and X-ray generator. Make assurance doubly sure SIGNAL RATING before connection. Operating description R-USM-503 26 The window time can be changed. Refer to the following pictures 27 R-USM-503 Make assurance doubly sure SIGNAL RATING before connection. B. Mode 2 : Trigger cable mode Connect the Trigger cable between the X-ray enable connector of AGI with X-ray generator. AGI Box Figure 8: <2nd example> wiring for the interface between portable generator & detector Connection description Signal New Label Old Label Color Input / Output 12V~24VDC R.E.C. +12-24V Red READY IN Ready Input Signal Black Input Input READY DONE Exposure Output SW1 Green Output READY DONE COM Exposure Output SW2 Blue Output READY IN READY DONE R-USM-503 28 Operating description t1: It will be occurred when exposure switch is pushed completely after. t2: Window time of detector is varying 0 sec ~ 15 sec. which can be control by S/W. Expose relay switch ON time (Window Time) and X-ray exposure time must be same. t2 time can be control by S/W The window time can be changed. Refer to the following pictures R-USM-503 29
(3) Connection (Auto Trigger) Power Connection A. Connect the battery pack or power cable to the equipment.
* In wired mode, the frame ground is necessary. Be sure to sure only the dedicated battery pack, RB37WH for 1417WGB. R-USM-503 30 Wireless Communication A. AP Router(Line sharer) setting
- SSID : Griffon
- Internal network
- IP address : 192.168.1.80
- Subnet mask : 255.255.255.0
- Dynamic IP allocation range
: 192.168.1.2 ~
192.168.1.254
- Pre-Shared Key(Password) : project302
- Authentication methods : WPAPSK or WPA2PSK
- Password methods : TKIP/AES
- AP IP : 192.168.1.1
- Channel (Frequency)
- Avoid crowded channel
(Using wireless detector under crowded channel result in low performance)
- If available, Use Auto-Channel Selection function of router to find optimal channel B. Reception Indicator Link LED flickering Blink Speed : Slow Low link quality Fast High link quality R-USM-503 31 C. Checking Link Quality & Battery Remain
- After wireless connection is established, perform Get Bright in Calibration tap.
- Check the value named Wireless Signal in black log screen. Wireless Signal = Link Quality (Max. 100)
- Check the value named Battery Remain in black log screen. Battery Remain = Battery Remain(Max. 100) The value Wireless Signal will be displayed here R-USM-503 32 Software Setting R-USM-503 33
(4) IP set up
[My Network Places] [Properties] [Local Area Connection]
[Properties] [Internet Protocol (TCP/IP)]
[Use the following IP address]
IP address: Obtain an IP address automatically IP address : Obtain an IP address automatically
(5) Checking Connection Check LED on the detector & power supply Ping test: [Start] [Run] ping -t 192.168.1.80 R-USM-503 34
(6) Using Web Manager (IP, SSID Change / Upgrade FW) Change IP Address of Detector A. Turn on Detector and connect to PC
(wired connection is recommended) B. After detector boot up, Launch web-browser
(Optimized for Chrome/Internet Explorer) C. Connect to http://[Detectors IP]
D. Select IP address to change from drop-down menu R-USM-503 35 E. Click SUBMIT F. Restart detector(Turn Off then On) Change SSID and PSK(Pre-Shared Key) A. Turn on Detector and connect to PC
(wired connection is recommended) B. After detector boot up, Launch web-browser
(Optimized for Chrome/Internet Explorer) C. Connect to http://[Detectors IP]
D. Select SSID to set
(This value should match to Routers setting) R-USM-503 36 E. Type Pre-Shared Key to set (Password)
(This value should match to Routers setting Default:project302 ) F. Click SUBMIT G. Restart detector(Turn Off then On) Upgrade Firmware A. Turn on Detector and connect to PC
(wired connection is recommended) R-USM-503 37 B. After detector boot up, Launch web-browser
(Optimized for Chrome/Internet Explorer) C. Connect to http://[Detectors IP]
D. Select firmware file by click Choose File button.
(Released file is named fw_ppc) E. Click Upload File button F. Click Flash Write button G. Restart detector(Turn Off then On) R-USM-503 38 5. Calibration X-ray detector should be used at stable state within driving temperature range. Acquire the X-ray images after power on and 30 minutes warming up to obtain high quality images.
(1) General Principle Notation Calibration can be done by image acquisition S/W. The gain-
offset correction (under calibration) will be done with one dark, at least one bright and object frame. Parameter Description Offset Dark image, acquired image without X-ray exposure Bright Acquired image with X-ray exposure Object Bright image with object, will be calibrated Gain Gain of imaging system, offset subtracted image Offset correction Offset subtract Gain correction Compensate gain variance of pixel Table 4: Parameter description R-USM-503 39 The calibration range of bright is can be select by which exposure level is maximum level that user want to use. If the maximum level of user want to use is 6500 in this case the level is contained in Bright point of 3(refer Table 5 : Median value). The meaning is you dont have to make bright point for 4(In this case, it will be does not working if you get image on higher level than maximum bright point.) Bright Calibration Point To gain correction, bright frame and dark frame should be acquired. The dark frame is needed only one frame. The bright frame is recommended to be acquired more than 2 different levels of median values of bright frames. The X-ray source condition will be recommended the tube energy level of 70kVp with variable tube current and exposure time. To acquire at least 2 frames at same condition will be recommended. The median values of bright frames are below.
(Additional 21mm Al filter is recommended for calibration) Point Median Value [LSB 1 500 ~
1000 Table 5 : Median value 2 2900 ~
3400 3 6200~
6700 4 9500 ~
10000 The Purpose of Bright Calibration The center of the non calibrated image is brighter than the edge due to hill effect of X-ray exposure. Generally, the intensity of X-
ray flux at center region of exposed area is higher than surroundings due to the X-ray expose like cone shape. A calibration process is used to compensate for this effect. Generally, called it Flat Field Correction (Bright calibration). Figure 8: Comparison of flat field correction processing R-USM-503 40
(2) Calibration Describe the calibration step by step. 1st Step Click on the Get Dark button. The acquired dark frame dark.raw will be generated in the \cal\ folder. Figure 9: Get dark R-USM-503 41 2nd Step Push Get Bright button at different four of X-ray condition. The X-ray condition should be set or tested before, same as the level of 1.2. Push Get Bright button at least 2 times at same condition, and then the offset subtracted bright (gain) is generated which of filename is xNNNNNA.raw(Refer to NOTE) Click button [Get Bright]. It will produce frame with name %CAL%
xNNNNNA.raw, where NNNNN is median pixels value within current image borders after offset calibration
(cut frame edges are never used during calibration). Suffix A (it also could be B,C etc) avoids casual coincidence of file names. Figure 10: Get bright. R-USM-503 42 3rd step After 2nd step, the Generate button will be activated. Click the button Generate, and then calibration point will be generated which of file name is A # of point_ median value of generated point like file of bright frame. The acquired bright frames within tolerance value which is variance of median level of acquired bright frames will be averaged and generated to a calibration point. The tolerance value can be edited. After making calibration point the Generate button will be disabling. Figure 11: Generate R-USM-503 43 4th step After 3rd step, Change Bad Pixels Removal Tab, Click the button Generate Auto BPM, and then Defect Map will be generated which of file name is BPM.raw at the \cal\ folder. Figure 12: Bad pixels removal R-USM-503 44 5th step For additional Defect correction, if BPMM.raw is existed at the install CD, copy to the \cal\ folder. 6th step On Acquisition Tab. Check the box Offset Calibration, Gain Calibration, Bad Pix Map for activate to each calibration and Bad Pixels Removal. Otherwise, it will does not working when going to pre-processing . Figure13: Application of calibration R-USM-503 45 6. Image Acquisition Test
(1) Get Image On Acquisition tab, click the Get Image button to get image. After click the button, you can see pop-up window, which is display window time and process of acquiring image. Figure 14: Get image R-USM-503 46
(2) View Images Frame- and image-files have extension raw and contain pixel data in signed 16-bits little-endian format. One could view those files in Photoshop or another image editor. Figure 15: View images R-USM-503 47 Common controls and displayed statistics Figure 16: Description of viewer 1 Pixel_Min minimum pixel value in frame- or image- data Pixel_Max maximum pixel value Pixel_Black if a pixel Pixel_Black then it is displayed as black (RGB 0, 0, 0) Pixel_White if a pixel Pixel_Black then it is displayed as black (RGB 255, 255, 255) R-USM-503 48 Histograms presentation Relative Histogram Scale [H]=1000 means that that the distance depicted as H on the drawing matches 1% of total number of pixels. Respectively [H]=100 means that H matches 0.1% of pixels and [H]=500 means that H matches 0.5% of pixels. Figure 17: Description of histogram R-USM-503 49 Marker type S Displays local surround of selected location Figure184: Description of marker type "S"
Marker type R Display profile chart of a row. Figure 19: Description of marker type "R"
R-USM-503 50 Marker type C Display profile chart of a column. Figure 20: Description of marker type "C"
R-USM-503 51 7. Operation
(1) Recommend X-ray detector should be used at stable state within driving temperature range. Acquire the X-ray images after power on and 30 minutes warming up to obtain high quality images. The calibration should be performed every 6 months
(2) Switching power on / off All connection should be done, before turn on the power. Press the power button by more than 3 sec, when power on/off. The green light of the LED indicator on the detector is on, the detector power is on. The blue light of the LED indicator on the detector is blinking, the detector is getting prepared to work and initialize. After power off, separate the battery.
(3) Storage Store the sensor unit in clean and dry place. Ensure that storage place should be not affected by dust or liquids. R-USM-503 52 8. Radio Frequency compliance
(1) FCC Notice (U.S.A) Test standards
- 47CFR Part 15.107 (b) / 47CFR Part 15.109 (g) Class A.
- FCC Part 15 C Section 15.247
- FCC Part 15 C Section 15.407 FCC ID: QIIRY1417WGB 5.15- 5.25 GHz band is restricted to indoor operations only. Compliance with FCC requirement 15.407(c) Data transmission is always initiated by software, which is the passed down through the MAC, through the digital and analog baseband, and finally to the RF chip. Several special packets are initiated by the MAC. These are the only ways the digital baseband portion will turn on the RF transmitter, which it then turns off at the end of the packet. Therefore, the transmitter will be on only while one of the aforementioned packets is being this device automatically discontinues transmission in case of either absence of information to transmit or operational failure. In other words, transmitted. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. R-USM-503 53 FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the users authority to operate the equipment. This transmitter must not be co-located or operated in conjunction with any other antenna or transmitter. When installing it in a mobile equipment This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines in Supplement C to OET65. This equipment has very low levels of RF energy that it deemed to comply without maximum permissive exposure evaluation (MPE). But it is desirable that it should be installed and operated keeping the radiator at least 20cm or more away from persons body (excluding extremities: hands, wrists, feet and ankles). When you use the detector with wire mode, the wireless function is automatically off. R-USM-503 54
(2) IC Notice (CANADA) Test standards
- ICES-003 Issue 4
- IC RSS-210 Issue 7 IC: 10742A-1417WGB This Class A digital apparatus complies with Canadian ICES-003 This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Le present appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisee aux deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioelectrique subi, meme si le brouillage est susceptible d'en compromettre le fonctionnement. R-USM-503 55
(3) R&TTE Notice (European Union) Test standards
- EN 62311
- ETSI EN 301 489-1
- ETSI EN 301 489-17
- ETSI EN 300 328
- ETSI EN 301 893 R-USM-503 56 9. Maintenance
(1) Maintenance Maintenance of the detector should be done by an authorized service provider If the Detector Panel is defective, the detector will be returned as is to the manufacturer for repair Clean the equipment with a dry soft cloth, or a soft cloth lightly moistened with mild detergent solution. Do not use any type of solvent, such as benzene This equipment and accessories are to be disposed of safely after the life span of them and national regulation must be observed. For safety reasons, be sure to inspect the instrument before using it. In addition, carry out a regular inspection at least once a year. Arrange the detector and power supply link cable to prevent the damage of the cables rubber tube. For example, do not press the cable under the legs of the table or the people. If you need more information, please contact to the Rayence C/S team, they will provide it. When the instrument is going to be cleaned, be sure to turn OFF the power of each instrument, and unplug the power cable from the AC outlet. Never use benzene, thinner or any other flammable cleaning agents. Otherwise, fire or electric shock may result.
(2) Cleaning Cover Clean the cover by the left warning box, if it is dirty. CFRP of Detector Wipe the CFRP plate of the sensor unit with ethanol or glutaraldehyde solution. If you are using disinfectant other than those specified above, or you are mixing another disinfectant with R-USM-503 57 ethanol, please also consult a specialist, because they may harm the CFRP plate.
(3) Inspection In order to ensure that the instrument is used safely and normally, please be sure to inspect the instrument before use. If any problem is found during the inspection, please take measures indicated in this chapter. If problem still cannot be corrected, please contact Rayence representative or distributor. It is recommended that a record of the inspection be kept by making copies of the check lists in this section, or making a separate check list. Inspection chart Inspection User Vendor Cycle Check that cables are not damaged or cover of cables is O not torn Check that the plugs and locks of connectors are not loose Check that the cover or parts are not damaged and not loose Check the LED indicator Check the bad pixel map Check the performance of the instrument by performing exposures using a phantom or a O O O resolution chart Daily Daily Daily Daily O Half year O Yearly R-USM-503 58 For safety reasons, be sure to turn OFF the power of the detector when the following inspections are going to be performed. Otherwise, it may result in electric shock.
(4) Marking and labeling symbols Symbols Meaning Caution : Attention, see instructions for use Alternate current Protective earth (Ground) Off (power : disconnect from the main switch) On (power : connect to the main switch) Warning Caution Note R-USM-503 59 Manufacturer Date of manufacture Serial number WEEE
: Waste Electrical and Electronic Equipment Figure 21: Labeling symbols R-USM-503 60 Appendix Dimension
[unit : mm]
R-USM-503 61 Rayence Co., Ltd. 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea www.rayence.com Rayence Co., Ltd. 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea www.rayence.com R-USM-503 62