1 | User manual | Users Manual | 3.86 MiB | / November 11 2018 |
Trilogy Evo Universal Instructions for Use Instructions for Use | Introduction For Technical Support and Customer Service, contact Philips Customer Service:
USA Respironics Inc. 1001 Murry Ridge Lane Murrysville, PA 15668 Email: service@philips.com, clinical@philips.com Web: www.philips.com\healthcare Authorized European Representative:
Respironics Deutschland GmbH & Co. KG Gewerbestrasse 17 D-82211 Herrsching Germany
+49 8152 93060 Copyright 2018 Koninklijke Philips N.V. All rights reserved. This work is protected under Title 17 of the United States copyright code and is the sole property of Philips Respironics. No part of this document may be copied or otherwise reproduced, or stored in any electronic information retrieval system, except as specifically permitted under United States copyright law, without the prior written consent of Philips Respironics. Trilogy Evo Universal is the registered trademark of Respironics Inc. The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Philips Respironics is under license. Dawn Ultra is a registered trademark of Procter & Gamble. Other trademarks and trade names are those of their respective owners. Illustrations and screen images in this document are representative depictions. Contents 1. Introduction ................................................................ 4 Intended Use................................................................... 4 Environments of Use ....................................................... 4 Contraindications ............................................................ 4 Package Contents............................................................ 5 Warnings ......................................................................... 5 Symbols Glossary ............................................................ 9 How to Contact Philips Respironics .............................. 11 2. About Trilogy Evo Universal ...................................... 12 Parts of Trilogy Evo Universal ........................................... 12 Parts of the User Interface ............................................ 14 Monitoring Window ...................................................... 15 3. Therapy Modes and Controls .................................... 17 Overview ....................................................................... 17 Control Modes .............................................................. 19 Spontaneous Modes ..................................................... 21 Mixed Modes ................................................................ 23 Therapy Features .......................................................... 25 Therapy Actions ............................................................ 27 Therapy Control Settings .............................................. 27 Dynamic Therapy Parameters ....................................... 29 4. Device Setup ............................................................. 31 Setup Overview ............................................................. 31 Placement ..................................................................... 31 Connecting AC Power ................................................... 31 Installing Filters ............................................................. 31 Connecting a Circuit ...................................................... 33 Adding Oxygen .............................................................. 35 Starting Trilogy Evo Universal ....................................... 36 5. Device Operation ...................................................... 37 Clinical Assessment ....................................................... 37 Entering New Patient Information ................................ 37 About Prescriptions ...................................................... 37 Starting and Stopping Ventilation ................................. 39 Actions during Ventilation ............................................ 39 Instructions for Use | Introduction 6. Alarms and System Messages .................................. 41 Overview ...................................................................... 41 About Alarms ................................................................ 41 The Alarm List ............................................................... 42 Setting and Changing Alarms........................................ 42 Setting the Alarm Volume ............................................ 43 Responding to an Alarm ............................................... 43 Resetting Alarms .......................................................... 43 Alarms and System Messages ...................................... 44 Prescription Alarm Availability by Therapy Mode ........ 60 Testing Alarms .............................................................. 60 Alarm and Event Log .................................................... 63 7. Device Options ......................................................... 65 Device Settings ............................................................. 65 Calibration .................................................................... 66 Data Transfer ................................................................ 68 Information .................................................................. 70 8. Cleaning and Disinfection ......................................... 71 Exterior Cleaning and Disinfection ............................... 71 Cleaning the Detachable Battery .................................. 72 Rinsing the Air-Inlet Foam Filter ................................... 72 External Active Exhalation Valve Cleaning and Disinfection .................................................................. 73 Service and Maintenance ......................................... 75 Service .......................................................................... 75 Disposal ........................................................................ 75 Routine Maintenance ................................................... 75 Replacing the Air-Inlet Foam Filter ............................... 75 Replacing the Particulate Filter .................................... 76 Preparing the Device for a Use by a Different Patient.. 76 Trilogy Evo Universal Accessories ......................... 77 Power Accessories ........................................................ 77 Patient Monitors .......................................................... 77 Filters ............................................................................ 80 Patient Circuits ............................................................. 81 Circuit Accessories ........................................................ 81 Dual Limb Active Exhalation Valve ............................... 82 Oxygen ......................................................................... 82 Portability and Travel Accessories ................................ 83 10. 9. 11. Power Management ............................................. 84 Power Sources .............................................................. 84 Battery Power Indicator ................................................ 84 Battery Status Indicator ................................................ 84 AC Power ........................................................................ 84 External Battery ............................................................ 85 Detachable Battery ....................................................... 85 Internal Battery ............................................................. 86 Power Loss .................................................................... 86 Power Indicator Icons ................................................... 86 Technical Specifications ........................................ 88 12. Instructions for Use | Introduction Specifications ............................................................... 88 Standards Compliance .................................................. 91 Pneumatic Diagram ...................................................... 92 EMC Information .......................................................... 92 13. Wireless Connectivity ........................................... 96 Bluetooth Actions ......................................................... 96 NFC Actions .................................................................. 97 Troubleshooting ........................................................... 97 14. Glossary ................................................................ 98 15. Warranty ............................................................ 100 Intended Use 3 1. Introduction Instructions for Use | Introduction The Trilogy Evo Universal ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician according to its technical specifications. Intended Use The Trilogy Evo Universal ventilator provides invasive and non-invasive positive pressure ventilation for the care of patients 2.5 kg through adults. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, respiratory rate, and pulse rate data when integrated with the appropriate accessories. The ventilator is suitable for use in institutional and hospital settings and non-emergency transport settings; for example, wheelchair, personal vehicle, or ambulance. Environments of Use The Trilogy Evo Universal ventilator is intended to be used:
In institutional environments. While attached to a wheelchair, bedrail, gurney, roll stand, or sitting on a flat surface such as a table or nightstand. While transporting patients within and between facilities such as automobile or commercial aircraft. Contraindications If the patient has any of the following conditions, consult the patients health care professional before using noninvasive ventilation:
An inability to maintain a patent airway or adequately clear secretions At risk to aspirate gastric contents Acute sinusitis or otitis media Epistaxis, causing pulmonary aspiration of blood Hypotension Instructions for Use | Introduction Package Contents Warnings Environmental You should not operate Trilogy Evo Universal in the presence of flammable gasses. Do not cover the ventilator or place in a position that affects proper operation. Do not block the cooling and intake air vents. Do not operate Trilogy Evo Universal in an environment that is outside the specified ranges. Using the ventilator outside of this temperature range or above this altitude can affect the ventilator performance. The blower warms the temperature of the airflow. Use of the device at room temperatures warmer than 40C may cause thermal irritation. Do not expose the device or detachable battery to temperatures above 60 C (140 F) during use or above 70 C (158 F) during storage. This will reduce battery life and may increase the risk of fire or damage the battery. The Trilogy Evo Universal is not intended for MRI or anesthesia applications, and is not intended to be permanently mounted in EMS vehicles. When disposing of this device or any accessories, ensure you comply with your local regulations. Dispose of any potentially biohazardous waste according to your local regulations. Warnings 5 Clinical Instructions for Use | Introduction This device is intended for use in the electromagnetic environment specified in the EMC Information chapter. Ensure the environment is compatible. Portable and mobile RF communications equipment, including cables, should be no closer to any part of the device than the recommended separation distance indicated in the EMC Information chapter. This device is not made with natural rubber latex. Do not use the ventilator in a hyperbaric chamber. Do not use the ventilator in the presence of nitric or nitrous oxide. Do not use the ventilator with helium or in the presence of mixtures in combination with helium. Route all cables in a manner to prevent injury, such as tripping or strangulation, to the patient and caregiver. Before placing a patient on the ventilator, perform a clinical assessment. Considerations should include:
Choosing alarm settings
-
- Whether alternative ventilation equipment is required
- Whether alternative monitors are required, such as Vte monitoring for Active PAP circuit, pulse oximeter or respiratory monitor with alarm Trilogy Evo Universal is a restricted medical device. It is designed for use by respiratory therapists or other trained and qualified caregivers under the supervision of a physician. Only the supervising physicians orders authorize changes to the prescription and other device settings. Before using Trilogy Evo Universal, you must read and understand this manual. When using the ActivePAP circuit, CO2 monitoring is required to measure exhaled carbon dioxide, in accordance with ISO 80601-2-55. Unintentional leaks cause exhaled volume and expired CO2 values to differ from actual patient values. The caregiver or health care professional is responsible for verifying any changes to the device, prescription, or other settings before applying changes. The caregiver or health care professional is responsible for ensuring settings are correct and compatible with the patient. Using the wrong prescription for a patient may result in improper therapy, lack of appropriate safety monitoring, or risk of death or injury to the patient. Alternate Ventilation To avoid patient death or serious injury, ventilator-dependent patients require immediate access to alternate ventilation equipment, such as a back-up ventilator or manual resuscitator. Qualified personnel should monitor ventilator-dependent patients continuously. Personnel should be prepared to provide alternate therapy in the event of ventilator failure or inoperative equipment. Alarms Do not rely on any single alarm to detect a disconnected circuit. Respond immediately to any high priority alarm. It may indicate a potentially life-threatening condition. Visually monitor the patient and ventilator at all times during an Alarm Silence period. Allowing alarms to continue without intervention may result in harm to the patient. If the high-priority Low Battery alarm occurs, immediately connect the ventilator to an alternate power source. If no alternate power source is available, immediately place the patient on an alternate source of ventilation. When using a remote alarm, make sure you fully test the remote alarm connector and cable by verifying that:
-
-
You can hear the ventilators audible alarms on the remote alarm. The remote alarm signals when you disconnect the remote alarm cable from the ventilator or from the remote alarm Test the operation of the circuit disconnect function daily and whenever the patient circuit is changed. An increase in circuit resistance can prevent proper operation of some alarms. Warnings 6 Instructions for Use | Introduction Speaking valves, heat moisture exchangers (HMEs), humidifiers, and filters create additional circuit resistance and may affect the performance of alarms chosen for circuit disconnect protection. Do not set the Low Peak Inspiratory Pressure alarm too low, or the system may not detect large circuit leaks or a patient disconnect. Accessories Use Trilogy Evo Universal only with accessories intended for use with this device. Otherwise, adverse performance including increased electromagnetic emissions or decreased electromagnetic immunity of this equipment can occur. For a list of accessories, such as patient interfaces, circuits, exhalation ports, and cables, see the Trilogy Evo Universal accessories guide. Ensure accessories and parts are compatible before you connect a patient to the device. Consult the accessorys instructions before use. The air-inlet foam filter is required to protect the ventilator from dirt and dust. See the Service and Maintenance chapter for maintenance instructions. Be certain that any breathing system filter used with this device complies with ISO 23328-1 and ISO 23328-2. To prevent patient or ventilator contamination, we recommend you use a Respironics-approved main flow bacteria filter (Part Number 342077) on the patient gas outlet port. Filters not approved by Respironics may degrade system performance. When adding any components (such as humidifiers, speaking valves, heat moisture exchangers, and filters) to the breathing system, consider the flow resistance and dead space in relation to the potential for adverse effects on the patients ventilator management and device alarms. Nebulization or humidification can increase the resistance of breathing system filters. Monitor the breathing system frequently for increased resistance and blockage. Gas added by the use of a pneumatic nebulizer can adversely affect ventilator accuracy. When using a passive circuit an exhalation port is required. At low expiratory pressures, the flow through the exhalation port may be inadequate to clear all exhaled gas from the tubing some rebreathing may occur. Do not use antistatic or conductive hoses or conductive patient tubing with the device. The ventilator system (used with patient circuit accessories, such as patient interface devices, humidifiers, water traps, and circuit tubing) may contain small parts that could result in a choking hazard. Be certain that any humidifier in use, including any heated breathing tube, complies with ISO 8185. This device is equipped with an oxygen blender that can deliver oxygen to the patient within a range of 21-100%
concentration. To ensure accuracy of oxygen administration when using the oxygen blender, use the internal FiO2 accessory or an external oxygen monitor that complies with ISO 80601-2-55 to verify the oxygen concentration in the delivered gas. Substantial leaks may reduce the inspired oxygen concentration to less than the expected value. Use appropriate patient monitoring, as medically indicated, such as an alarming pulse oximeter. Do not connect the device to an unregulated oxygen source. Do not use oxygen while smoking or in the presence of an open flame. Turn off the oxygen flow when the device is not in use. Cleaning and Maintenance To avoid electric shock, do not remove the enclosure cover. Only service personnel should remove the enclosure. Do not immerse the device or allow liquids into any of the controls or the interior of the enclosure as the device may be damaged. If this occurs, contact your equipment provider for assistance. Use only the agents and methods described in Warnings 7 Oxygen Instructions for Use | Introduction this manual to clean and disinfect the device. After cleaning and disinfecting, ensure the device is completely dry before reattaching accessories and connectors and before reconnecting it to a power source. Do not use solvents, polishes, or any oily substances on the device, as they are flammable. Repairs and adjustments must be performed by service personnel only. Unauthorized service could cause death or injury, If the device has been exposed to rain or dampness, dry the device including the area around the power cord connection with the power cord disconnected from the device before applying AC power. invalidate the warranty, or result in costly device damage. detachable battery is dropped, if water is spilled into the enclosure, or if the enclosure is cracked or broken, discontinue use and contact Philips Respironics. If you notice any unexplained changes in the performance of the device, if it is making unusual sounds, if the device or To avoid electrical shock, always unplug the power cord from the wall outlet before cleaning the ventilator. Periodically inspect electrical cords, cables, and the detachable battery pack for damage or signs of wear. Discontinue use and replace if damaged. approved by Respironics may void the users authority to operate the equipment. No modification of this equipment is allowed. Any changes or modifications made to the device that are not expressly Do not connect the ventilator to the battery of a battery-powered wheelchair as this can affect the ventilator performance, Power which can result in patient death. Cable. This cable is fused, pre-wired, and properly terminated to ensure safe connection An external battery should only be connected to the ventilator using the Philips Respironics approved External Battery Warnings 8 Instructions for Use | Introduction Symbols Glossary This glossary contains the symbols on the model label and on the device exterior. Software symbol explanations appear throughout this book. For a complete explanation of symbols appearing on the device and associated labels, go to the following web address:
http://www.symbols.philips.com Symbol Definition Symbols on the Model and Warning Label Refer to instruction manual Symbol Definition Oxygen inlet Prescription device Not made with natural rubber latex For airline use. Complies with RTCA D0160 section 21, category M Bluetooth symbol This equipment includes RF transmitters. IP22: protection against finger-sized objects and protected against dripping water when tilted up to 15 degrees. Catalog number Batch code Serial number Manufacturer Date of manufacture Class II equipment BF applied part LI-ion recycling Waste electrical & electronic equipment Humidity limit Temperature limit Symbols on the Device On/Off (Standby) button Alarm Silence button USB connection Nurse call connection DC power (direct current) AC power (alternating current) Low flow oxygen inlet Flow sensor cable connection Proximal pressure out AEV control line Patient in Patient out Symbols on the Screen - General Prescription settings Home screen Options Help Manual breath Deliver 100% oxygen Delete prescription Patient type indicators Touch screen lock Edit Wi-Fi Symbols on the Screen - Alarms Alarms tab Alarm Silence High priority alarm Symbols Glossary 9 Instructions for Use | Introduction Medium or low priority alarm System message Alarm reset Symbols on the Screen Monitoring Views See chapter 2, About Trilogy Evo. Symbols on the Screen - Power See the Power Management chapter. Symbols on the Screen Connectivity Bluetooth enabled Bluetooth connected USB is exporting data Symbols Glossary 10 Instructions for Use | Introduction How to Contact Philips Respironics If you need help setting up, using or maintaining Trilogy Evo Universal, or if this device does not perform as expected, contact Philips Respironics. Call Philips Respironics Customer Service at:
1-724-387-4000 1-800-345-6443 (toll-free)
+49 8152 93060 (international) How to Contact Philips Respironics 11 Instructions for Use | About Trilogy Evo Universal 2. About Trilogy Evo Universal Parts of Trilogy Evo Universal Front Panel 1. On/off (standby) button 2. AC power indicator 3. Alarm indicator/alarm silence 4. Alarm bar 5. Touch screen 6. Ambient light sensor 1. Carrying handle 2. FiO2 sensor access panel 3. Power cord retention clip 4. Air inlet 5. Oxygen blender 6. High pressure oxygen inlet 7. Air vents Parts of Trilogy Evo Universal 12 Back Panel Patient Panel Utility Panel Instructions for Use | About Trilogy Evo Universal 1. Accessory USB port
(pulse oximeter, CO2 monitor) Inspiratory port (to patient) 2. 3. Proximal pressure port 4. Active exhalation valve line connection for ActivePAP and Active Flow circuits 5. Dual limb active exhalation valve connection (from patient) Flow sensor cable connector 6. Low flow oxygen inlet 1. AC power connector 2. Air vents 3. 4. Detachable battery access door 5. Micro USB Port for device service 6. Accessory USB Port (USB external storage device, communication cables) 7. Remote alarm or nurse call connector (RJ9) 8. DC power connector Parts of Trilogy Evo Universal 13 Instructions for Use | About Trilogy Evo Universal Parts of the User Interface Standard Screen Elements 1. Menu bar 2. Main window 3. Monitored parameters pane 4. Menu Bar Status bar Use the menu bar to navigate, manage alarms, and see the active prescription at a glance. 1. Home: view the main window 2. Prescriptions: manage patient prescriptions 3. Options:
-
-
-
-
-
Device Settings Calibration Data Transfer Information Alarm & Event Log 4. Patient type indicator Parts of the User Interface 14 Instructions for Use | About Trilogy Evo Universal Main Window The main window contents vary depending on the action you are performing. The main window can show the standby window, prescription window, monitoring window, and others. Monitored Parameters Pane Use the monitored parameters pane to see measured and calculated values while delivering therapy. These values vary based on the circuit, therapy mode, and accessory type. Depending on the accessories you use, values such as SpO2 appear during active ventilation and during standby. Parameters that may appear are:
PIP: peak inspiratory pressure Vte: exhaled tidal volume RR: respiratory rate
-
-
-
- MinVent: minute ventilation
-
-
-
SpO2: saturation of peripheral oxygen Pulse Rate etCO2: end tidal carbon dioxide Status Bar Use the status bar to monitor device status and the availability of manual therapeutic actions. 1 Manual breath 2 3 4 5 6 7 Deliver 100% Oxygen 100% Oxygen timer CMD USB data transfer Bluetooth Bluetooth data transfer 8 9 10 11 12 13 14 Wi-Fi Alarm silence Power sources and their status Device Actions Menu System time Monitoring Window During ventilation, a monitoring window, or home screen, contains information such as measured parameters and battery status. You can select the type of information you want to see. Selecting a Monitoring View To select a monitoring view during ventilation:
In the Menu Bar, tap the Home button. In the monitoring window, tap the Views button In the Views menu, tap the type of view you want to use. Monitoring Window 15 1. 2. 3. Instructions for Use | About Trilogy Evo Universal Types of monitoring windows Monitoring windows may vary based on your model. Views menu icon Small manometer Measured and calculated parameters Large manometer Waveform graphs Monitoring window contents
- Manometer pressure indicator
-
Set parameters Set parameters
-
- Measured and calculated parameters
-
Additional parameters based on the prescription (including accessories) This is the default view.
-
An explanation of dynamic parameters is in chapter 3, Therapy Modes.
-
-
Large manometer pressure indicator Six measured and calculated parameters Customizable scalar waveform graphs
-
To customize the graphs, use the buttons in the window as follows:
Button Description Select the waveforms to graph. On the Select Waveforms dialog box, select data for the top and bottom graphs. Pause graphing. Automatically size the vertical scale to fit the data. Tap to change the time scale, and then select a new time scale from the list. Monitoring Window 16 Instructions for Use | Therapy Modes and Controls 3. Therapy Modes and Controls Overview Trilogy Evo Universal therapy modes can be used for invasive and noninvasive ventilation with all circuit types, including mouthpiece ventilation. Breath Types Trilogy Evo Universal can deliver the following breath types:
o Mandatory: Ventilator-initiated, time-cycled o Assist-Control: Patient-initiated, time-cycled o Spontaneous: Patient-initiated, patient-cycled Triggering and Cycling Patient triggers Auto-Trak is a flow trigger with rules that make patient-triggering and cycling more comfortable for the patient. The system uses multiple algorithms to detect the start and end of the breath. Also, it automatically adjusts the trigger and cycle sensitivity to optimize synchronization between the patient and the ventilator. Sensitive Auto-Trak is a more sensitive version of Auto-Trak Flow trigger initiates a breath when the patients inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. A lower number is more sensitive. As inspiratory flow begins to decrease, the device cycles to expiration when the patient flow is less than the percentage of peak flow, based on the flow-cycle sensitivity setting. Ventilator trigger Time Trigger is time-based, defined by the Breath Rate setting. Flow patterns Ramp wave pattern: airflow starts high and decreases throughout inspiration of the breath. Square wave pattern: airflow is generally constant throughout inspiration of the breath. Therapy Modes Overview Control Modes: breaths are assist-control or mandatory. A/C-PC: Assist control assist-control and mandatory breaths with pressure control A/C-VC: Assist control assist control and mandatory breaths with volume control Spontaneous modes: the patient initiates all breaths. CPAP: Continuous positive airway pressure PSV: Pressure support ventilation Overview 17 Instructions for Use | Therapy Modes and Controls Mixed modes: breaths are spontaneous, assist-control, or mandatory. S/T: Spontaneous/timed ventilation spontaneous breaths with pressure support and mandatory breaths with pressure control SIMV-PC: Synchronized intermittent mandatory ventilation (pressure control) spontaneous breaths with pressure support, assist control breaths and mandatory breaths with pressure control SIMV-VC: Synchronized intermittent mandatory ventilation (volume control) spontaneous breaths with pressure support, assist control breaths and mandatory breaths with volume control Low Tidal Volume Therapy For low tidal volume therapy, use the infant/pediatric external flow sensor. See the instructions included with the sensor. When setting volumes greater than or equal to 50ml, use any circuit type. When setting volumes greater than or equal to 35ml, use either the active flow or dual limb circuits. The following pressure modes are available for patients who require a tidal volume less than 35ml:
A/C-PC PSV S/T SIMV-PC Therapy Mode Comparison Table For all modes, the breath type varies based on time of patient inspiration. The breath type is always ventilator-initiated and mandatory when the Trigger Type is set to Off. Breath Types Trigger Source Inspiration Cycle Exhalation Spontaneous Patient CPAP Patient PSV Spontaneous Patient Pressure Support + PEEP Patient Mode Control Modes A/C-PC A/C-VC Spontaneous Modes CPAP Assist-Control Patient Mandatory Ventilator
(Breath Rate) Assist-Control Patient Mandatory Ventilator
(Breath Rate) Spontaneous Patient Mandatory Ventilator
(Breath Rate) Spontaneous Patient Assist-Control Mandatory Patient Ventilator
(Breath Rate) Spontaneous Patient Assist-Control Patient Mandatory Ventilator
(Breath Rate) Mixed Modes S/T SIMV-PC SIMV-VC Pressure Control +PEEP Inspiratory Time PEEP Tidal Volume Inspiratory Time PEEP CPAP PEEP EPAP PEEP PEEP IPAP Patient Inspiratory Time Pressure Support + PEEP Patient Pressure Control + PEEP Inspiratory Time Pressure Support + PEEP Patient PEEP Tidal Volume Inspiratory Time PEEP Overview 18 Instructions for Use | Therapy Modes and Controls Control Modes A/C-PC: Assisted/Control-Pressure Control DESCRIPTION:
The A/C-PC mode provides pressure-controlled mandatory or assist-control breaths. When the Trigger Type is set to Off, the ventilator triggers and cycles all breaths. When the Trigger Type is not set to Off, then the ventilator or the patient can trigger a breath, and the ventilator cycles all breaths. SETTINGS:
Setting Name Pressure Control PEEP Rise Time Breath Rate Inspiratory Time Trigger Type Flow Trigger Sensitivity Flow Cycle Sensitivity FiO2
(optional) SETTABLE ALARMS:
Description Inspiratory pressure above PEEP Positive end expiratory pressure Time required for the ventilator to change from the expiratory pressure setting to the inspiratory pressure setting when the breath is triggered. Minimum rate of breaths per minute Length of the inspiratory phase Off This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the patients inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. This control is available when the trigger type is Flow Trigger. As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the device cycles to expiration. Requires model with oxygen blender Auto-Trak (passive circuits only) Sensitive Auto-Trak (passive circuits only) Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) Circuit disconnect High tidal volume Low tidal volume High minute ventilation Low minute ventilation High respiratory rate Low respiratory rate ILLUSTRATION 1. EPAP 2. Pressure control Control Modes 19 Instructions for Use | Therapy Modes and Controls A/C-VC: Assisted/Control-Volume Control DESCRIPTION:
The A/C-VC mode provides volume-controlled mandatory and assist breaths. When the Trigger Type is set to Off, both triggering and cycling are performed by the ventilator. When the Trigger Type is not Off, then the trigger can be performed by the vent or patient and the cycle is always performed by the ventilator. To deliver the set volume in the set time, the ventilator alters the flow rate. The flow pattern setting defines the shape of the flow delivery pattern. SETTINGS:
Setting Name Tidal Volume PEEP Inspiratory Time Breath Rate Flow Pattern Trigger Type Flow Trigger Sensitivity Flow Cycle Sensitivity SETTABLE ALARMS:
Description Set inspiratory volume Positive end expiratory pressure Length of the inspiratory phase Minimum rate of mandatory breaths per minute Sets the shape of the waveform as a ramp or square Off This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the patients inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. This control is available when the trigger type is Flow Trigger. As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the device cycles to expiration. Auto-Trak Sensitive Auto-Trak (passive circuits only) Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) Circuit disconnected High tidal volume Low tidal volume High minute ventilation Low minute ventilation High respiratory rate Low respiratory rate High inspiratory pressure Low inspiratory pressure ILLUSTRATION Control Modes 20 Instructions for Use | Therapy Modes and Controls Spontaneous Modes CPAP: Continuous Positive Airway Pressure DESCRIPTION:
In CPAP mode, the pressure delivered to the patient during both inhalation and exhalation is the CPAP pressure setting. All breaths in this mode are spontaneous breaths. The ventilator monitors inspiratory and expiratory tidal volume. SETTINGS:
Setting Name CPAP Trigger Type Flow Trigger Sensitivity Flow Cycle Sensitivity SETTABLE ALARMS:
Auto-Trak Sensitive Auto-Trak (passive circuits only) Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) Description Continuous positive airway pressure range This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the patients inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. This control is available when the trigger type is Flow Trigger. As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the device cycles to expiration. Circuit disconnected High tidal volume Low tidal volume High minute ventilation Low minute ventilation High respiratory rate Low respiratory rate ILLUSTRATION Spontaneous Modes 21 Instructions for Use | Therapy Modes and Controls PSV: Pressure Support Ventilation DESCRIPTION:
PSV mode is patient-triggered, pressure-limited, and flow-cycled. With this strategy, breaths are assisted by a set inspiratory pressure that is delivered until inspiratory flow drops below a set threshold. In the PSV mode, the ventilator delivers spontaneous, pressure-supported, breaths and user-initiated mandatory breaths. The ventilator functions as a demand flow system, with the patient triggering breaths and determining their timing and volume. The ventilator can support the breaths with the set pressure support. The Pressure Control setting defines the applied pressure above PEEP. The patient determines the breath timing. As in other dual limb modes, you also set PEEP, inspiratory trigger, and O2. It is recommended that you set backup ventilation in PSV mode. SETTINGS:
Setting Name Pressure Support PEEP Rise Time Trigger Type Flow Trigger Sensitivity Flow Cycle Sensitivity SETTABLE ALARMS:
Description Target pressure that the device delivers during the inspiratory phase of a spontaneous breath Positive end expiratory pressure Time required for the ventilator to change from the expiratory pressure setting to the inspiratory pressure setting when the breath is triggered. This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the patients inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. This control is available when the trigger type is Flow Trigger As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the device cycles to expiration. Auto-Trak Sensitive Auto-Trak (passive circuits only) Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) Circuit disconnected High tidal volume Low tidal volume High minute ventilation Low minute ventilation High respiratory rate Low respiratory rate ILLUSTRATION:
Spontaneous Modes 22 Instructions for Use | Therapy Modes and Controls Mixed Modes S/T: Spontaneous/Timed DESCRIPTION:
A bi-level therapy mode where each breath is patient-triggered and patient-cycled or ventilator-triggered and ventilator-cycled. In this mode, an IPAP is delivered during inhalation and a lower EPAP is delivered during exhalation. The duration of a spontaneous breath is determined by the patient effort. The duration of a mandatory breath is determined by the inspiratory time setting. Remember that the IPAP setting is the maximum pressure the ventilator will deliver; it is not in addition to the EPAP setting. SETTINGS:
Setting Name IPAP EPAP Rise Time Breath Rate Inspiratory Time Trigger Type Flow Trigger Sensitivity Flow Cycle Sensitivity Description Inspiratory positive airway pressure Must be greater than or equal to EPAP Expiratory positive airway pressure Time required for the ventilator to change from the expiratory pressure setting to the inspiratory pressure setting when the breath is triggered. Minimum rate of breaths per minute. If the patient doesnt trigger a breath within this time, the ventilator triggers the breath. For a mandatory breath, length of the inspiratory phase This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the patients inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. This control is available when the trigger type is Flow Trigger As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the device cycles to expiration. Auto-Trak Sensitive Auto-Trak (passive circuits only) Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) SETTABLE ALARMS:
Circuit disconnect High tidal volume Low tidal volume High minute ventilation Low minute ventilation High respiratory rate Low respiratory rate ILLUSTRATION:
Mixed Modes 23 Instructions for Use | Therapy Modes and Controls SIMV-PC: Synchronous Intermittent Mandatory Ventilation- Pressure Control DESCRIPTION:
SIMV-PC mode is a pressure control mode that provides a mixture of mandatory and spontaneous breaths. SIMV-PC mode guarantees one mandatory breath in each cycle. Spontaneous breaths can be delivered with pressure support. The breath rate determines the length of the cycle. The first phase of the cycle is reserved for synchronizing a mandatory breath with patient effort. If the patient triggers a breath during this phase of the cycle, the ventilator delivers a synchronized mandatory breath. If a patient does not trigger a breath during the mandatory phase of the cycle, then the ventilator delivers a mandatory breath. Breaths triggered by the patient after the mandatory breath in the cycle are spontaneous breaths. This process is repeated at the start of every cycle. SETTINGS:
Setting Name Pressure Control Pressure Support PEEP Inspiratory Time Rise Time Breath Rate Trigger Type Flow Trigger Sensitivity Flow Cycle Sensitivity Description Defines the applied pressure for all breaths Target pressure that the device delivers during the inspiratory phase of a spontaneous breath Positive end expiratory pressure Positive pressure maintained in the patient circuit during exhalation: must be less than or equal to the pressure setting. For a mandatory breath, length of the inspiratory phase Time required for the ventilator to change from the expiratory pressure setting to the inspiratory pressure setting when the breath is triggered. Minimum rate of mandatory breaths per minute This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the patients inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. This control is available when the trigger type is Flow Trigger As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the device cycles to expiration. Auto-Trak Sensitive Auto-Trak (passive circuits only) Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) SETTABLE ALARMS:
Circuit disconnect High tidal volume Low tidal volume High minute ventilation Low minute ventilation High respiratory rate Low respiratory rate ILLUSTRATION Mixed Modes 24 Instructions for Use | Therapy Modes and Controls SIMV-VC: Synchronous Intermittent Mandatory Ventilation- Volume Control DESCRIPTION:
Similar to SIMV-PC, but with volume control. SETTINGS:
Setting Name Tidal Volume Pressure Support PEEP Inspiratory Time Rise Time Breath Rate Flow Pattern Trigger Type Flow Trigger Sensitivity Flow Cycle Sensitivity SETTABLE ALARMS:
Description Target gas volume that the device delivers during a spontaneous breath Target pressure that the device delivers during the inspiratory phase of a spontaneous breath Positive end expiratory pressure Positive pressure maintained in the patient circuit during exhalation: must be less than or equal to the pressure setting. For a mandatory breath, length of the inspiratory phase Time required for the ventilator to change from the expiratory pressure setting to the inspiratory pressure setting when the breath is triggered. Minimum rate of mandatory breaths per minute Sets the flow-pressure waveform This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the patients inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. This control is available when the trigger type is Flow Trigger As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the device cycles to expiration. Auto-Trak Sensitive Auto-Trak (passive circuits only) Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) Circuit disconnect High tidal volume Low tidal volume High minute ventilation Low minute ventilation High respiratory rate Low respiratory rate High inspiratory pressure Low inspiratory pressure ILLUSTRATION See SIMV-PC. Therapy Features The following features are available in addition to the therapy modes. Backup Ventilation Enable DESCRIPTION Set the device to deliver ventilator-initiated breaths when patient-initiated breaths are not detected, based on the Apnea alarm interval. When you turn Backup Ventilation on, set an Apnea interval in the alarm settings tab. Within the apnea interval; if no breaths are triggered by the patient, the ventilator delivers breaths at the set pressure or volume based on the Backup Rate. When an Apnea alarm occurs, the ventilator automatically starts backup ventilation. When two consecutive patient-initiated breaths are detected, the ventilator automatically reverts to patient-initiated breaths. Backup ventilation settings take precedence over standard therapy mode settings. Therapy Features 25 SETTINGS:
Instructions for Use | Therapy Modes and Controls Backup Ventilation (On/Off): when you turn this setting On, set an Apnea interval in the alarm settings tab. Backup Rate: when in backup ventilation, the backup breath rate takes precedence over any breath rate set in the therapy mode. The rate cannot be less than the Breath Rate set in the current therapy mode. Backup Inspiration Time (CPAP and PSV modes only) when in backup ventilation the Back Up Tinsp controls the duration of inspiration. Backup PS (CPAP mode only) Backup Rise Time (CPAP mode only) To access the Backup Ventilation feature, in the Prescription window, tap Advanced. When you turn the feature on, the additional settings appear in the prescription window. If Trigger Type is Off, this feature is unavailable. APPLICABLE THERAPY MODES A/C-PC A/C-VC CPAP PSV S/T SIMV-PC SIMV-VC Insp Time Min/Max Enable DESCRIPTION Set the minimum and maximum inspiratory time for pressure support breath types. This feature changes inspiration time from a constant to a variable value so you can select a range for the inspiration time. The range allows the patient to have a chance to cycle. When the maximum time has passed with no patient-initiated breath, then the ventilator automatically cycles the breath. To access this feature, in the Prescription window, tap Advanced. When you turn the feature on, the additional settings appear in the prescription window. APPLICABLE THERAPY MODES PSV S/T SIMV-PC SIMV-VC Sigh Enable DESCRIPTION Delivers a periodic, larger volume breath. SETTINGS:
Sigh (On/Off) Sigh Volume Sigh Frequency (deliver a sigh after X number of patient- or ventilator- triggered breaths) To access the Sigh feature, in the Prescription window, tap Advanced. When you turn the feature on, the additional settings appear in the prescription window. APPLICABLE THERAPY MODE A/C-VC Therapy Features 26 Instructions for Use | Therapy Modes and Controls Therapy Actions Oxygen Flush DESCRIPTION This feature requires the oxygen blender. When active, the device delivers 100% oxygen for two minutes. This feature functions independent of any oxygen blending setting. During an oxygen flush, the High FiO2 alarm is disabled. WORKING WITH OXYGEN FLUSH To start an oxygen flush:
Tap 100% O2 in the status bar and then tap Start. A timer appears that counts down the two minutes. To stop an oxygen flush:
Tap 100% O2 in the status bar and then tap Stop. APPLICABLE THERAPY MODES All Manual Breath DESCRIPTION Delivers a breath based on the current therapy mode settings. To deliver a manual breath:
Tap the Manual Breath button in the status bar and then tap Start. APPLICABLE THERAPY MODES A/C-PC A/C-VC PSV S/T SIMV-PC SIMV-VC Therapy Control Settings Therapy control settings can be interdependent. For guidance, see the previous therapy mode descriptions. Setting Name Backup Pressure Support All circuits but passive: 0-60 cmH2O Passive circuit: 0-57 cmH2O Increments: 1 cmH2O Setting Range/Increment Adult patient type:
Pediatric patient type:
Infant patient type:
All circuits: 0-30 cmH2O Increments: 1 cmH2O Therapy Control Settings 27 Setting Name Breath Rate CPAP EPAP Min/Max FiO2 Flow Cycle Sensitivity Flow Pattern Instructions for Use | Therapy Modes and Controls All circuits: 0-20 cmH2O Increments: 1 cmH2O Setting Range/Increment Adult patient type: 0-80 BPM, 1BPM increments Pediatric patient type: 0-60 BPM, 1BPM increments Infant patient type: 0-40 BPM, 1 BPM increments Adult and pediatric patient types: 3-25 cmH2O, 1 cm H2O increments Infant patient type: 3-15 cmH2O, 1 cm H2O increments Adult and pediatric patient types: 3-25 cmH2O, 1 cm H2O increments Infant patient type: 3-15 cmH2O, 1 cm H2O increments 21-100%, 1% increments
(21% = ambient condition, no control) 10-90%, 1% increments Square: airflow is constant Ramp: inspiration airflow starts high and decreases Flow Trigger Sensitivity 0.5 (high sensitivity) to 9 L/min (low sensitivity) Inspiratory Time IPAP Max Pressure PEEP Pressure Control Pressure Support PS Min/Max Rise Time Tidal Volume 6-60 cmH2O, 1 cmH2O increments Active circuit: 0-35 cmH2O Passive circuit: 3-25 cmH2O Increments: 1 cmH2O Active circuit: 0-15 cmH2O Passive circuit: 3-15 cmH2O Increments: 1 cmH2O Active circuit: 0-25 cmH2O Passive circuit: 3-25 cmH2O Increments: 1 cmH2O All circuits but passive: 0-60 cmH2O Passive circuit: 0-57 cmH2O Increments: 1 cmH2O Adult patient type: 0.5-5.0 seconds, 0.1 second increments Pediatric patient type: 0.3-2.0 seconds, 0.1 second increments Infant patient type: 0.3-1.0 seconds, 0.1 second increments Adult patient type: 3-60 cmH2O, 1 cmH2O increments Pediatric patient type: 3-45 cmH2O, 1 cmH2O increments Infant patient type: 3-35 cmH2O, 1 cmH2O increments Adult patient type:
Pediatric patient type:
Infant patient type:
Adult patient type:
Pediatric patient type:
Infant patient type:
Adult patient type:
Pediatric patient type:
Infant patient type:
0 (faster) to 6 (slower) Adult patient type:
70-1200 ml Increment: 5 ml All circuits: 0-20 cmH2O Increments: 1 cmH2O All circuits but passive: 0-60 cmH2O Passive circuit: 0-57 cmH2O Increments: 1 cmH2O All circuits but passive: 0-60 cmH2O Passive circuit: 0-57 cmH2O Increments: 1 cmH2O All circuits: 0-30 cmH2O Increments: 1 cmH2O All circuits: 0-20 cmH2O Increments: 1 cmH2O All circuits: 0-30 cmH2O Increments: 1 cmH2O Therapy Control Settings 28 Instructions for Use | Therapy Modes and Controls Setting Name Setting Range/Increment Pediatric patient type:
Dual limb or active flow: 35-400 ml Passive or Active PAP: 50-400 ml Increment: 5 ml Auto-Trak (passive circuits only) Sensitive Auto-Trak (passive circuits only) Flow Trigger (all circuits) Off Trigger Type Dynamic Therapy Parameters It is unnecessary to perform an inspiratory hold to assess the plateau pressure and other lung parameters. The advanced measurement system of Trilogy Evo estimates lung compliance, airway resistance, AutoPEEP and plateau pressure during normal mechanical ventilation without requiring a static maneuver. Dyn R Airway resistance is the opposition to the motion of gas within the airways. In the Measured and Calculated Parameters window, this value is Dyn R
(dynamic resistance), so named because it is estimated without requiring a static maneuver. At the end of inhalation, Trilogy Evo estimates the airway resistance by computing the ratio between the driving pressure from within the lung to the air flow. The flow term is corrected to take into account the contributions of the following:
Intrinsic PEEP, by subtracting the expiratory flow at the end of exhalation The elastic recoil of the lungs, by adding the tidal volume divided by the respiratory time constant, .
(Respiratory time constant is the airway resistance times the summed compliance of the lung and chest wall) Trilogy Evo calculates Dyn R using the following formula:
=
(=)(=)+
Where:
PIP is the peak inspiratory pressure (pressure at the end of inhalation) is the extrinsic pressure (pressure applied by the ventilator) at the end of the breath is the tidal volume
(=) is the patient flow at the end of the exhalation (EOE)
(=) is the patient flow at the end of inhalation(EOI) To understand the calculations adopted to compute Dyn R, note that the above equation can be rewritten as the classic equation for airway resistance:
(++/ )= ()
=() this value is the intrinsic PEEP or AutoPEEP (see the section, AutoPEEP below)
+ is the total pressure (extrinsic plus intrinsic) at the end of the breath That is to say, pressure across the resistance equal to resistance times flow, where:
Dyn C Lung Compliance is the ratio between the tidal volume and the changes in pressure. In the Measured and Calculated Parameters window, this value is Dyn C (dynamic compliance), so named because it is estimated without requiring a static maneuver. Trilogy Evo estimates the integrated compliance of the pulmonary system, (the summed compliance of the lung and chest wall). The compliance of the respiratory system can be derived from the measurement of plateau pressure, , using the relationship between the tidal volume, , and the difference between the and PEEP. Trilogy Evo calculates Dyn C using the following formula:
Dynamic Therapy Parameters 29 Note that the compliance is related to the airway resistance by the respiratory time constant, , by the relationship described above. Where:
Instructions for Use | Therapy Modes and Controls PEEP is the total pressure (intrinsic plus extrinsic) at the start of the breath (= + ) is the tidal volume
=
=
= ++
is the tidal volume is dynamic compliance PEEP is the total pressure (intrinsic plus extrinsic) at the start of the breath (= +
=() this value is the intrinsic PEEP or AutoPEEP Dyn Pplat Plateau pressure is the pressure applied to small airways and alveoli during positive-pressure mechanical ventilation. In the Measured and Calculated Parameters window, this value is Dyn Pplat. Having already estimated the compliance (Dyn R and Dyn C above), Trilogy Evo calculates Dyn Pplat as follows:
Intrinsic PEEP, , is the resistive pressure at the end of exhalation (EOE), that occurs when a new breath is initiated before the previous breath is completed. In the Measured and Calculated Parameters window, this value is AutoPEEP. AutoPEEP can be used as a guide to detect the presence of dynamic hyperinflation. In most cases, this number represents the pressure in the small airways at the start of the breath in excess of the PEEP applied by the ventilator. In some cases, such as a ventilated, complex, COPD patient, the displayed pressure may not be accurate. However, in these complex cases, any non-zero pressure displayed (accurate or not) indicates the presence of AutoPEEP. When Auto PEEP is zero, this indicates that there is no intrinsic PEEP. Where:
AutoPEEP Trilogy Evo calculates AutoPEEP using the following formula: = (=)
(=) is the patient flow at the end of the exhalation (EOE) Dyn R is the dynamic resistance (explained in a previous section, Dyn R above) Where:
Dynamic Therapy Parameters 30 Instructions for Use | Device Setup 4. Device Setup Setup Overview To set up Trilogy Evo, follow the steps shown below. See the accompanying section for instructions. Step 1. Place the device. 2. Connect AC power. 3. 4. Connect a circuit. 5. Add oxygen (optional). 6. Start the device. Install filters. Section Placement Connecting AC Power Installing Filters Connecting a Circuit Adding Oxygen Starting Trilogy Evo Universal Page 31 31 32 32 35 36 Placement Place Trilogy Evo on a stable, flat, hard surface. Air must flow freely. Do not block the air vents with items such as bedding or curtains. Do not place Trilogy Evo near any heating or cooling equipment or air supplies such as forced air vents, radiators, or air conditioners. Ensure the USB and detachable battery panel doors remain closed when not in use. If the device has been stored outside the normal operating temperature stated in Technical Specifications, ensure the device reaches operating temperature before connecting power. See the EMC Information chapter for guidance on possible electromagnetic interference. Connecting AC Power Use the AC cord provided to connect AC power before pressing the power button. Verify Trilogy Evo Universal is using AC power, indicated by the green LED light next to the power button. To use other power sources, such as the detachable battery or an external battery, see the Power Management chapter. Installing Filters CBRN Filter To install the CBRN filter:
1. Ensure the CBRN filter adapter gasket is securely seated. 2. Place the filter adapter onto the bayonet mount and twist clockwise to lock in place. Screw the CBRN filter onto the filter adapter. 3. Installing Filters 31 Instructions for Use | Device Setup Air-Inlet Foam Filter Ensure the air-inlet foam filter is installed correctly. To install the air-inlet foam filter, pinch the filter as you press it into the filter cover as shown. Position it securely behind the top and bottom restraints. Particulate Filter To install a particulate filter:
1. 2. 3. 4. 1. Twist the filter cover counterclockwise and pull out to remove it. 2. Place the filter over the bayonet mount. 3. Twist clockwise a quarter of a turn. 4. Replace the filter cover and turn clockwise to secure. Installing Filters 32 Instructions for Use | Device Setup Connecting a Circuit Be certain that any breathing system filter used with this device complies with ISO 23328-1 and ISO 23328-2. To prevent patient or ventilator contamination, we recommend you use a Respironics-approved main flow bacteria filter (Part Number 342077) on the patient gas outlet port. Filters not approved by Respironics may degrade system performance. Note: A leak device is mandatory during invasive ventilation or when using a circuit with a non-vented mask. After you connect the circuit, you may calibrate the circuit. See chapter 7, Device Options, Calibration. For low tidal volumes, see the Therapy Modes chapter, section titled Low Tidal Volume Therapy. Passive Single Limb Circuits 1 2 3 Bacteria filter Tubing Exhalation port 1. Connect the bacteria filter (1) on the circuit to the Inspiratory Port. Connecting a Circuit 33 Active PAP Circuits Instructions for Use | Device Setup 1 2 3 4 Bacteria filter Proximal pressure port Active exhalation valve line connection Tubing 1. Connect the bacteria filter (1) on the circuit to the Inspiratory Port. 2. Connect the proximal pressure line to the Proximal Pressure Port (2). 3. Connect the active exhalation valve pressure line to the active exhalation valve line connection (3). Active Flow Circuits 1 2 3 4 5 Bacteria filter Proximal pressure port Active exhalation valve line connection Flow sensor cable connector Tubing 1. Connect the bacteria filter (1) on the circuit to the Inspiratory Port. 2. Connect the proximal pressure line to the Proximal Pressure Port (2). 3. Connect the active exhalation valve pressure line to the active exhalation valve line connection (3). Connecting a Circuit 34 Instructions for Use | Device Setup 4. Connect the flow sensor to the Flow Sensor Cable Connector (4). Dual Limb Circuits 1 2 3 4 5 6 Bacteria filters Proximal pressure port Active exhalation valve connection Flow sensor cable connector Tubing Flow sensor connected to circuit Install the external active exhalation valve (AEV) according to the instructions provided with it (3). 1. Attach the bacteria filter (1) end of the colored inspiratory tube to the Inspiratory Port. 2. Attach the proximal pressure line (2) to the Proximal Pressure Port. 3. 4. Attach the bacteria filter end of the clear expiratory tube to the AEV (3). 5. Attach the flow sensor cable to the flow Sensor Cable Connector (4). 6. Attach the flow sensor to the Y-shaped connector on the circuit (6). Low Tidal Volume Ventilation When using volumes between 35 and 50 ml, use the infant/pediatric external flow sensor with either the active flow or dual limb circuits. See the operating instruction sheets included with those accessories. Adding Oxygen High Pressure Oxygen To ensure accurate oxygen administration and to monitor for the presence of contamination, use an FiO2 sensor or external oxygen monitor to verify the oxygen concentration in the delivered gas. To connect high-pressure oxygen: Connect an oxygen hose to the high-pressure oxygen connector on the back panel, and then connect the other end of the hose to the source. After you have completed the connection, calibrate the O2 sensor. See chapter 7, Device Options, O2 Sensor Calibration. To maintain accuracy, calibrate the oxygen sensor daily. Adding Oxygen 35 Low Flow Oxygen To connect low flow oxygen:
Instructions for Use | Device Setup 1. Connect the oxygen tubing to the O2 adapter supplied with the device. 2. Connect the O2 adapter to the low flow oxygen inlet on the Utility Panel by pressing down on the valve. Oxygen-Related Alarms After the oxygen is connected, set the oxygen-related alarms. See chapter 6, Alarms and System Messages for instructions. Oxygen alarms are as follows:
Oxygen Regulation Low SpO2 High SpO2 Low FiO2 High FiO2 Low Oxygen Input Pressure High Oxygen Input Pressure Starting Trilogy Evo Universal To start Trilogy Evo Universal:
1. Visually inspect Trilogy Evo Universal and all accessories, cords, and tubes attached to the device. 2. Verify that circuit connections are secure. 3. Press the On/Off (Standby) button. 4. Listen for a minimum of three beeps as Trilogy Evo Universal performs system start up checks. The beeps test all alarm signals to ensure proper functioning. Ensure no system messages appear. 5. Watch as the light bar and Alarm Silence button blink once red and once yellow. 6. Confirm that the power sources you have connected are functioning properly and that power is sufficient. For help, see the Power Management chapter. Starting Trilogy Evo Universal 36 Instructions for Use | Device Operation 5. Device Operation Clinical Assessment Warning Before placing a patient on the ventilator, perform a clinical assessment. Considerations should include:
Choosing alarm settings Assessing whether alternative ventilation equipment is required Selecting additional accessories, including the patient monitoring accessories you will use For ventilator-dependent patients, always have alternate ventilation equipment, such as a back-up ventilator or manual resuscitator available. Ventilator-dependent patients should be continuously monitored by qualified personnel. These personnel should be prepared to provide alternate therapy in the event of ventilator failure or inoperative equipment. Entering New Patient Information To enter new patient information 1. 2. 3. 4. Infant Pediatric Adult In the Home window, tap the New Patient button. This button clears all existing patient data. In the New Patient window, select a Patient Type:
Select the Patient Sex. For infant patients, in the Weight section, use the slider or the plus and minus buttons to select the patients weight. For pediatric or adult patients, in the Height section, select the patients height. Note: this information is used to establish default therapy and alarm settings, including tidal volume and alarms based on tidal volume. In the title bar, tap Accept to save your choices. 5. 6. Acknowledge the reminder to ensure a viral/bacterial filter is installed on the outlet of the device. 7. Edit the prescription settings according to the procedure in the next section, About Prescriptions. Therapy and alarm settings differ based on Patient Type. See the Therapy Modes and Controls and Alarms chapters. About Prescriptions System Timeout When working in the Prescription Settings window, ensure you save your changes. After a period of inactivity, the system reverts to the previous setting and your changes are not saved. If you are making changes during therapy delivery, the period of inactivity is 30 seconds. If you are making changes when the system is in standby, the period of inactivity is 5 minutes. About Prescriptions 37 Instructions for Use | Device Operation Parts of the Prescription Settings Window 7-1: Parts of the Prescription Settings Window Editing Prescription and Alarm Settings To edit prescription settings:
In the menu bar, tap the prescription settings icon. In the prescription grid, tap Circuit. In the workspace below the grid, select the circuit Type and Size. If you are using a humidifier, in the Active Humidification section, tap Yes. Otherwise, tap No. 1. 2. 3. 4. To save your changes, tap Accept. 5. 6. Tap a prescription parameter. In the lower pane, move the slider or use the plus and minus buttons to change the parameter value. 7. Tap the Alarm tab to view and edit the associated alarm settings. In the prescription grid, tap Mode. In the workspace below the grid, select the mode. For more information about alarms, including details about each alarm, see chapter 6: Alarms. 8. Change the alarm parameters in the lower pane and then click Accept at the top of the window. 9. Continue editing prescription and alarm parameters. To work with advanced options, tap Advanced. 10. Test the alarms. See Testing Alarms in chapter 6. 11. When you are finished, tap Accept to save your changes. Adding a Prescription If you want to add another prescription:
In the menu bar, tap the prescription settings icon. 1. 2. Tap the Prescription list to expand it and then tap Add New. 3. On the Select Prescription Name dialog box, tap the prescription name that you want to use. 4. Edit the prescription settings as you would for a new prescription. About Prescriptions 38 Instructions for Use | Device Operation Deleting a Prescription To delete a prescription: In the Home window, tap the prescription you want to delete and then tap the trashcan icon. Starting and Stopping Ventilation To start ventilation:
In the Prescription Settings window, select the prescription you want to use and then tap Start Therapy. To stop ventilation and put the ventilator on standby:
Press the On/Off (Standby) button on the front panel. On the confirmation window, tap Standby. To turn the device off:
Press the On/Off (Standby) button on the front panel. On the confirmation window, tap Power Off. Actions during Ventilation Using Different Prescriptions If a clinician has set up more than one prescription; for example, a daytime and nighttime prescription, you can use the prescription list to change the active prescription. To select a prescription that uses a circuit type that is different from the current prescription:
1. Press the On/Off (Standby) button on the front panel. On the confirmation window, tap Standby. 2. Attach the circuit type that corresponds to the prescription. 3. On the menu bar, tap the active prescription to expand the prescription list. 4. Tap the prescription you want to use and then confirm your choice. 5. Tap Start Therapy. To select a prescription that uses a circuit type that is the same as the current prescription:
1. On the menu bar tap the active prescription to expand the prescription list. 2. Tap the prescription you want to use and then confirm your choice. 3. Tap Start Therapy. Locking and Unlocking the Screen To lock the screen, expand the Device Actions Menu then tap the Lock Screen button. Actions during Ventilation 39 Instructions for Use | Device Operation To unlock the screen, tap the screen. On the screen unlock dialog box, tap and hold Yes for three seconds. When an alarm or system message becomes active, the screen saver is stopped and the automatic screen lock is disabled. For more information, see the Alarms chapter. Delivering a Manual Breath To deliver a manual breath, tap the Manual Breath icon in the Status Bar. Delivering 100% Oxygen To deliver 100% oxygen for two minutes, tap the 100% O2 icon in the Status Bar. To stop delivery, tap the icon again and then tap Stop. Actions during Ventilation 40 Instructions for Use | Alarms and System Messages 6. Alarms and System Messages Overview Trilogy Evo Universal generates audible and visual alarms to alert you when conditions require attention. Warning: To prevent death or serious injury, monitor the patient and the ventilator regularly in order to determine the need to provide emergency ventilation when an alarm sounds or the ventilator malfunctions. Always test the alarms after changing the circuit or prescription. An increase in circuit resistance can prevent proper operation of some alarms. Speaking valves, heat moisture exchangers (HMEs), and filters create additional circuit resistance and may affect the performance of alarms chosen for circuit disconnect protection. Alarm settings are retained when power is lost. Do not rely on any single alarm to detect a disconnected circuit. Certain components may affect the performance of the alarms chosen to signal that a circuit is disconnected. Use the apnea, low tidal volume, low minute ventilation, and low respiratory rate alarms in conjunction with the circuit disconnect alarm. Test these alarms daily and after you change ventilator settings. About Alarms When an alarm is active, the following indicators occur:
An Alarm List appears in the menu bar. An Alarm light bar flashes red or yellow, or glows steady yellow, depending on the alarm level (To configure this indicator, see chapter 7, Device Options. The Alarm Silence button on the device flashes red or yellow, or glows steady yellow, depending on the alarm level An audible alarm sounds. Trilogy Evo Universal uses three alarm levels:
High Priority Requires an immediate response Medium Priority Requires a prompt response Trilogy Evo Universal also shows system messages to inform you about changing conditions. These messages are described in Alarms and System Messages. Low Priority Requires awareness. When an alarm or system message becomes active, Trilogy Evo Universal stops the screen saver and disables the automatic screen lock. When external monitors are used, such as an FiO2 sensor or SpO2 sensor, associated alarm settings only appear when the ventilator detects sensor connection. Alarm settings are stored in the system so if a sensor becomes disconnected, the alarm settings are restored upon reconnection. To delete external monitor alarm settings, clear all patient data according to the procedure described in the Device Operation chapter. Some patient-related alarms have configurable settings, while others have fixed settings. All system and power alarm settings are fixed. Trilogy Evo Universal also shows system messages to inform you about changing conditions. Alarm and system message details appear in the Alarms and System Messages section in this chapter. Alarm and Message Indictors Icon Description Light and sound indicators High alarm Medium alarm Light bar flashes red Audible alarm repeats rapidly Light bar flashes yellow Audible alarm repeats moderately About Alarms 41 Instructions for Use | Alarms and System Messages Alarm and Message Indictors Icon Description Light and sound indicators Low alarm Message Light bar glows steady yellow Audible alarm repeats slowly Single beep Resolved alarm None or message The Alarm List The Alarm List appears in the menu bar. Tap the list to expand it and view the alarms. Tap the up and down arrows to navigate through the list. The list is ordered by priority and then by time. The most urgent, most recent alarm appears at the top of the list. An alarm counter shows the number of active alarms. Setting and Changing Alarms Warning: A hazard can exist if different alarms are used for the same or similar equipment in any single area. Ensure all alarms are appropriate for the patient before use. You can change alarm settings when creating a new prescription or when editing an existing prescription. To change an alarm setting:
1. On the menu bar, tap the Prescription icon. 2. In the Prescriptions window, tap the Alarm tab. Select the alarm you want to change. 3. 4. Adjust the parameters in the lower pane. 5. To undo changes, tap the Undo button in the lower pane. 6. When you are ready to save your changes, tap Accept. Setting and Changing Alarms 42 Instructions for Use | Alarms and System Messages Setting the Alarm Volume Warning:
Ensure the alarm volume is set loud enough for the caregiver to hear it. Consider the use of a remote alarm. If you do use a remote alarm, test it before starting ventilation. To set the alarm volume:
In the menu bar, tap the Options icon. In the Options window, tap Device Settings. In the Device Settings window, tap Alarm Volume. In the setting dialog box, make your selection. 1. 2. 3. 4. 5. On the Alarm Volume dialog box, select the volume you want and then tap the Accept checkmark. Responding to an Alarm Warning Visually monitor the patient and ventilator at all times during an Alarm Silence period. Allowing alarms to continue without intervention may result in harm to the patient. When an alarm occurs:
1. Ensure the patient has adequate ventilation and oxygen. If required, provide an alternate method of ventilation. 2. Tap the Alarm List to view all alarms and messages. If you see the help icon
, you can tap it for more information. 3. Press the Alarm Silence button on the device to pause all audible alarms for 2 minutes. 4. Take action to resolve the alarm. For help on specific alarms, see the Alarms and System Messages section. Resetting Alarms The Alarm Reset icon resets all active and resolved alarms. When an alarm-triggering condition is no longer present, the alarm status changes to Resolved and the alarm or message remains in the list. Tap the Alarm Reset icon to reset the alarm list. Resetting Alarms 43 Instructions for Use | Alarms and System Messages Alarms and System Messages This section contains the details of each alarm and system message:
Alarm High-priority System Alarms Ventilator Inoperative Ventilator Service Required Obstruction High Expiratory Pressure High Inspiratory Pressure External Flow Sensor Failure External Flow Sensor Cable Disconnected External Flow Sensor Not Connected External Flow Sensor Reversed Active Exhalation Valve Failed Check Active Exhalation Pressure Control Line Proximal Pressure Line Disconnected Oxygen Regulation High-priority Patient Alarms with Variable Settings Apnea Circuit Disconnected Low Minute Ventilation Low Respiratory Rate High Inspiratory Pressure Medium-priority System Alarms Circuit Leakage Rebreathing Detected Volume Under Delivery Loss of CO2 Signal Loss of SpO2 Signal CO2 Sensor Adapter Zero Required Check/Change CO2 Airway Adapter CO2 Sensor Failure Low Expiratory Pressure Low Inspiratory Pressure See page Alarm See page 45 45 45 45 45 46 46 46 46 47 47 47 47 48 48 48 48 48 49 49 49 49 50 50 51 51 51 51 51 52 52 Low Oxygen Input Pressure High Oxygen Input Pressure Medium-priority Patient Alarms with Variable Settings High Tidal Volume Low Tidal Volume High Minute Ventilation High Respiratory Rate Low Inspiratory Pressure Alarm Low SpO2 High SpO2 Low EtCO2 High EtCO2 Low FiO2 High FiO2 Low-priority System Alarms Stuck Key Inlet Filter Blocked Check Proximal Pressure Line Low-priority Patient Alarms with Variable Settings Low Pulse Rate High Pulse Rate Power Alarms Loss of All Power Low Battery AC Power Disconnected Internal Battery in Use Replace Detachable Battery Internal Battery Depleted System Messages 52 53 53 53 53 54 54 55 55 55 55 55 56 56 56 56 56 57 57 57 57 57 57 58 58 58 58 59 59 Alarms and System Messages 44 Instructions for Use | Alarms and System Messages High-priority System Alarms VENTILATOR INOPERATIVE High Priority Why it occurs The system self-test indicates a failure or malfunction of a component that causes therapy to stop or not meet essential performance criteria. Contact Philips Respironics Customer Service Therapy is stopped and both audible and visual alarms are continuous. Depending on the systems impacted, you may or may not see a message on the screen. What to do Device performance during active alarm VENTILATOR SERVICE REQUIRED Priority Why it occurs High This alarm occurs when the system self-test indicates an error that does not affect the ability of the ventilator to meet essential performance criteria. Contact Philips Respironics Customer Service The device continues to operate (possibly in a reduced capacity mode). If the problem is not corrected, the device will generate a reminder message until the issue is corrected. Additionally, if therapy is stopped, a reminder message will immediately appear when therapy is turned on again. What to do Device performance during active alarm OBSTRUCTION Priority Why it occurs What to do Requirements Device performance during active alarm Algorithm summary What to do High The ventilator detects an obstruction in the patients inhalation path, exhalation path, or external flow sensor. The ventilator detects that the exhalation port is missing. Passive, dual limb, active PAP, or active flow circuit The device will automatically open the active exhalation valve and continue to operate. Check the circuit for kinked or pinched tubing. Check the bacteria filter or HME for a blockage or occlusion. Ensure that the exhalation port is not occluded or missing. Check the external flow sensor for a blockage or occlusion. Any circuit, inhalation limb: an obstruction is detected when either of the following conditions exist:
The flow exiting the machine is less than 0.5 lpm for 5 seconds continuously The flow exiting the machine during inspiration is less than 1 lpm for 5 seconds, 2 breaths, or 65 seconds for very low breath rates. Any circuit: The expiratory port is missing and causes an obstruction alarm if the average flow is less than 1 lpm for 2 breaths or 60 seconds. Active flow or dual limb circuit: an obstruction is detected when the external flow sensor measures less than 0.5 lpm for 65 continuous seconds. HIGH EXPIRATORY PRESSURE Priority Why it occurs High During the expiratory phase, the delivered pressure exceeds the target patient pressure by 5 cmH2O or more. Note: This alarm condition may be due to pinched tubing or the patient having a fast breath rate. Check the circuit for kinked or pinched tubing. Ensure that the leak device is not blocked or occluded. Alarms and System Messages 45 Instructions for Use | Alarms and System Messages Requirements Device performance during active alarm The alarm is automatically resolved when the delivered pressure comes within 5 cmH2O of the target patient pressure during the expiratory phase. Device will continue to operate. HIGH INSPIRATORY PRESSURE Priority Why it occurs High During the inspiratory phase, the delivered pressure exceeds the target patient pressure by 5 cmH2O or more Check the patient for:
Coughing or excessive secretions Bronchospasms Tracheotomy tube stability Check the ventilator for:
Kinked, pinched, or blocked tubing Blocked leak device or blocked exhalation device Secretions in the HME o o o o o o Therapy mode must be SIMV-VC or A/C-VC The alarm is automatically resolved when the delivered pressure falls within 5 cmH2O of the target patient pressure during the inspiratory phase. The device will automatically cycle to the expiratory phase and continue to operate. EXTERNAL FLOW SENSOR FAILED Priority Why it occurs What to do High A flow sensor offset or sensor failure is detected when the device is in standby or delivering therapy. Ensure the sensor and cable are connected properly and are not damaged. Clean if needed. Replace if necessary. Active flow or dual limb circuit The alarm remains active until the flow sensor is replaced or reconnected. If the alarm occurs when delivering therapy, the device will continue to operate. If the alarm occurs when the device is in standby, you cannot start therapy. EXTERNAL FLOW SENSOR CABLE DISCONNECTED Priority Why it occurs High The external flow sensor cable becomes disconnected from the ventilator during active therapy. Active flow or dual limb circuit is selected and an external flow sensor cable is not connected. Ensure the sensor and cable are connected properly and are not damaged. Replace if necessary. Active flow or dual limb circuit Prohibits starting therapy if the external flow sensor cable is not connected to the ventilator when the active flow or dual limb circuit is selected EXTERNAL FLOW SENSOR NOT CONNECTED Priority Why it occurs High The external flow sensor becomes disconnected from the external flow sensor cable during active therapy. Ensure the sensor and cable are connected properly and are not damaged. Replace if necessary. Active flow or dual limb circuit Device continues to provide therapy at the set breath rate. Device will not trigger on a patient initiated breath. Alarms and System Messages 46 What to do Requirements Device performance during active alarm Requirements Device performance during active alarm What to do Requirements Device performance during active alarm What to do Requirements Device performance during active alarm Monitored parameters and alarms that use the flow measurement such as tidal volume will not function. Monitored parameters and alarms that use the pressure measurement will continue to function. Instructions for Use | Alarms and System Messages EXTERNAL FLOW SENSOR REVERSED Priority Why it occurs What to do Requirements Device performance during active alarm High The external flow sensor is connected backwards during active therapy. Check the sensor position. The arrow direction should align with the air delivered to the patient. Active flow or dual limb circuit In volume modes, device will not accurately control volume. The peak inspiratory pressure will continue to increase. In pressure modes, therapy will be unaffected. All monitored parameters and alarms will continue to function. ACTIVE EXHALATION VALVE FAILED Priority Why it occurs What to do For a single-limb active circuit, check all connections to the valve and check that the valve is clear. For a dual limb circuit, check that the valve is clear.. High The active exhalation valve is stuck closed. Active flow or dual limb circuit The alarm is automatically resolved and the device continues to operate. Requirements Device performance during active alarm What to do Requirements Device performance during active alarm Requirements Device performance during active alarm CHECK ACTIVE EXHALATION PRESSURE CONTROL LINE Priority Why it occurs High The active-exhalation pressure control line is not connected, becomes disconnected, or contains water droplets that affect the active exhalation valve line pressure reading. Inspect the line. Empty or replace it if necessary. Active flow or dual limb circuit Therapy delivery will be compromised. All monitored parameters and alarms will continue to function. PROXIMAL PRESSURE LINE DISCONNECTED Priority Why it occurs What to do High The proximal pressure line is not connected. Verify the proximal pressure line is connected properly at both ends and the line is clean and untangled. Ensure the main circuit is connected properly and does not have large leaks. Ensure the exhalation valve is intact. Dual limb, active PAP, or active flow circuit The alarm is automatically resolved when the proximal pressure line is connected properly. The device continues to operate. OXYGEN REGULATION Priority Why it occurs What to do High The flow measured FiO2 is not within 10% 3% of the FiO2 setting. Verify the oxygen blender is connected properly. Alarms and System Messages 47 Requirements Device performance during active alarm Algorithm summary Instructions for Use | Alarms and System Messages Ensure that the oxygen source is appropriate. Contact Philips Respironics Customer Service Oxygen source Device will continue to operate. The controller cannot regulate flow to guarantee FiO2 accuracy within 10% 3% of the FiO2 setting. High-priority Patient Alarms with Variable Settings APNEA Priority Why it occurs What to do Ensure the circuit is properly connected to the patient. Check the circuit for a leak or disconnect. Check the circuit for kinked or pinched tubing.. High Time between patient-initiated breaths is greater than the Apnea Interval setting. The alarm is automatically resolved when two patient breaths are detected that occur within set interval. Device will continue to operate. 5 to 60 seconds in increments of 5 seconds Available when Backup Ventilation is enabled Device performance during active alarm Alarm Settings CIRCUIT DISCONNECTED Warning: You should not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnected alarm. Priority Why it occurs What to do Device performance during active alarm Alarm Settings Algorithm summary High The patient is not properly connected to the ventilator breathing circuit or there is a large leak. Ensure the circuit is properly connected to both the patient and the ventilator and that a large unintentional leak does not exist. The alarm is automatically resolved when the circuit is reconnected or the excessive leak is fixed. Device will continue to operate. Off, 5 to 60 seconds in increments of 5 seconds. The alarm occurs when either of the following conditions are met:
The flow out of the device is excessive for a time greater than the alarm setting. The ventilator examines all expiratory flow for a minimum of three breaths or amount of time greater than the alarm setting and determines that the circuit is disconnected or obstructed. LOW MINUTE VENTILATION Priority Why it occurs What to do High The patients minute ventilation is less than or equal to the Low Minute Ventilation alarm setting. Or, no breath has occurred for 15 seconds. Note: A patient that is sleeping or medicated may breathe at a lower rate or with lower tidal volumes. These conditions may contribute to lower minute ventilation. Check the circuit for kinked or pinched tubing. Check the circuit for a leak or disconnect. Remove excessive water from the tubing. Ensure the bacteria filter is not blocked, occluded, or disconnected. Ensure the leak device is not blocked, occluded, or disconnected. Alarms and System Messages 48 Instructions for Use | Alarms and System Messages Device performance during active alarm Alarm Settings The alarm is automatically resolved when the calculated minute ventilation is greater than the low minute ventilation alarm setting. volume under delivery Device will continue to operate. Adult and Pediatric patient types: 0.2 to 30 LPM in increments of 0.1 LPM. Infant patient type: 0.2 to 10 LPM in increments of 0.1 LPM. What to do LOW RESPIRATORY RATE High Priority Why it occurs The patients respiratory rate is less than or equal to the low respiratory rate alarm setting. Or, no breath has occurred for 15 seconds. Note: A patient that is sleeping or medicated may breathe at a lower rate. The alarm is automatically resolved when the measured respiratory rate is greater than the low respiratory rate alarm setting. Device will continue to operate. Off, 1 to 80 BPM in increments of 1 BPM. Check the circuit for kinked or pinched tubing. Check the circuit for a leak or disconnect. Device performance during active alarm Alarm Settings HIGH INSPIRATORY PRESSURE Priority Variable The alarm will sound if the measured patient pressure exceeds the high inspiratory pressure alarm Why it occurs setting. Occurs in stages, escalating from an audible single beep for the first two occurrences to an audible medium priority on the third occurrence, and then to a high priority alarm if the condition continues. Occurrences must be consecutive. Check the patient status Check the ventilator for the following potential causes for this alarm:
What to do o o o Kinked, pinched, or blocked tubing Blocked leak device or blocked exhalation device Secretions in the HME Inspiration is cycled to PEEP with this alarm Device will continue to operate. The alarm is automatically resolved when the peak inspiratory pressure is less than or equal to the high inspiratory pressure alarm setting. The device will cycle out of the breath. 10 to 90 cmH2O in increments of 1 cmH2O. The High Inspiratory Pressure alarm shall generate a High Priority alarm on the 10th consecutive breath or if 30 seconds has elapsed since a previous breath and the Peak Inspiratory pressure from the current breath is greater than or equal to the alarm setting Medium-priority System Alarms CIRCUIT LEAKAGE Priority Why it occurs What to do Check the valve and both lines for kinked or pinched tubing. Check the circuit and both lines for a leak or disconnect at both ends. Check that the valve is not damaged and can properly seal and open. Medium In the Active PAP circuit, a leak in the active exhalation valve is detected Active PAP circuit is selected The alarm is automatically resolved when the circuit is reconnected or the valve is fixed. Device will continue to operate. Device performance during active alarm Alarm Settings Algorithm summary Requirements Device performance during active alarm Alarms and System Messages 49 Algorithm summary The flow from the device at the end of the expiratory phase is greater than a threshold based on a typical flow that would result if a 0.25 diameter orifice leak were present in the circuit. Instructions for Use | Alarms and System Messages REBREATHING DETECTED Priority Why it occurs What to do Device performance during active alarm Algorithm summary Medium The ventilator detects the potential for the inhalation of exhaled gases. Rebreathing is detected at 3 consecutive breaths for passive circuits and 6 consecutive breaths for Active PAP, Active Flow, or Dual Limb circuits. Check for a partially occluded exhalation port or increase expiratory leak flow. Ensure the exhalation valve is attached. The alarm is automatically resolved when exhaled air returns to a non-hazardous level for 6 consecutive breaths. Device will continue to operate. This alarm is based on an estimation of the inhaled fraction of carbon dioxide (FiCO2). For each breath, the ventilator assigns a percentage of FiCO2. When the sum of the percentages for the most recent breaths exceeds 30%, the alarm is announced. This method of using a running sum of concentration for the alarm reporting results in a shorter alarm delay when there are higher concentrations of FiCO2. The concentration estimate is based on nominal volumetric capnography curves, measured flow in the inhalation limb, and patient tidal volume. Any three breaths without rebreathing resets the sum of percentages to zero. This figure shows the assumed shape of the capnograph for a given volume of exhaled gases. VOLUME UNDER DELIVERY Priority Why it occurs What to do Requirements Device performance during active alarm Algorithm summary Medium A system limit is reached and the set volume cannot be reached for 3 consecutive breaths. Check the patient. Verify that the circuit and airway are clear of obstruction. Verify that the high pressure alarm limit is sufficient. Review all settings including Inspiratory Time, Tidal Volume, and Flow Pattern. Verify that patient and ventilator are synchronous by examining the waveforms and adjusting the trigger settings. The mode is A/C-VC or SIMV-VC. Device will continue to operate, attempting to deliver the set therapy All monitored parameters and alarms will continue to function. In modes designed to regulate the tidal volume on each breath such as AC-VC or SIMV-VC, the alarm is generated when the inhaled tidal volume is less than or equal to 85% of the tidal volume setting for 3 consecutive breaths. Alarms and System Messages 50 Instructions for Use | Alarms and System Messages LOSS OF CO2 SIGNAL Priority Why it occurs What to do Requirements Device performance during active alarm Medium The EtCO2 sensor is reporting invalid data, the sensor signal is lost, or no breaths are detected for more than 10 seconds while the device is delivering therapy. Ensure that the EtCO2 sensor is properly attached to the ventilator and the patient circuit. The High EtCO2 alarm or Low EtCO2 alarm is enabled and the sensor was previously reporting valid data for 3 continuous seconds.. The alarm is automatically resolved when the EtCO2 sensor is properly attached to the ventilator and the patient and the sensor reports data. Device will continue to operate. LOSS OF SPO2 SIGNAL Priority Why it occurs What to do Requirements Device performance during active alarm Requirements Device performance during active alarm Ensure that the SpO2 probe is properly attached to the ventilator and the patient. Reposition the probe on the patient, if necessary. Medium The oximeter is reporting invalid data or the oximeter is disconnected for more than 10 seconds while the device is delivering therapy or is in standby. Low SpO2, high SpO2, low pulse rate, or high pulse rate alarm is enabled and the oximeter was previously reporting valid data for 3 continuous seconds. The alarm is automatically resolved when the SpO2 probe is properly attached to the ventilator and to the patient and the oximeter reports data for more than 10 seconds. Device will continue to operate. CO2 SENSOR ADAPTER ZERO REQUIRED Priority Why it occurs What to do Medium The CO2 sensor requests a zero when the device is delivering therapy. Reestablish the baseline CO2 level. Follow the CO2 Sensor Adapter Zero procedure in the Device Options chapter. CO2 sensor Device will continue to operate. CO2 monitored parameters and alarms will not function CHECK/CHANGE CO2 AIRWAY ADAPTER Priority Why it occurs What to do Requirements Device performance during active alarm Medium The CO2 sensor reports that a check is required when the device is delivering therapy. Check the CO2 sensor CO2 airway adapter Device will continue to operate. CO2 monitored parameters and alarms will not function CO2 SENSOR FAILURE Priority Why it occurs What to do Requirements Device performance Disconnect and then reconnect the sensor. Replace the sensor. Medium The CO2 sensor reports a fault while the device is delivering therapy. CO2 sensor Device will continue to operate. CO2 monitored parameters and alarms will not function Alarms and System Messages 51 during active alarm Instructions for Use | Alarms and System Messages LOW EXPIRATORY PRESSURE Medium Priority During the expiratory phase, the delivered pressure is 5 cmH2O or more below the target patient Why it occurs pressure. The alarm is automatically resolved when the delivered pressure comes within 5 cmH2O of the target patient pressure during the expiratory phase. Device will continue to operate. Check the circuit for a leak or disconnect. Check the circuit for kinked, pinched, or blocked tubing. What to do Device performance during active alarm LOW INSPIRATORY PRESSURE Priority Medium During the inspiratory phase, the delivered pressure is 5 cmH2O or more below the target patient Why it occurs pressure. Check the patient for excessive inspiratory effort Check the ventilator for:
What to do o o Kinked, pinched, or blocked tubing Circuit leak or disconnect The alarm is automatically resolved when the delivered pressure comes within 5 cmH2O of the target patient pressure during the inspiratory phase. Device will continue to operate. Device performance during active alarm LOW OXYGEN INPUT PRESSURE Priority Why it occurs What to do Requirements Device performance during active alarm Algorithm summary Medium The pressure at the O2 inlet is too low to support the FiO2 setting. Check the O2 source. Additional oxygen Device will continue to operate. FiO2 delivery may be inaccurate. The machine assigns a minimum pressure threshold for oxygen based on the flow setting to support the FiO2 set point. The following table illustrates the threshold based on flow. The alarm occurs when the inlet pressure is less than or equal to 5psig or the Inlet Pressure value from the table for six consecutive breaths. Inlet Pressure [psig]
10 20 30 40 50 60 70 80 Maximum Flow [slpmd]
66.4 96.1 126.1 156.7 188.1 219.4 251.7 279.9 Alarms and System Messages 52 Instructions for Use | Alarms and System Messages HIGH OXYGEN INPUT PRESSURE Priority Why it occurs What to do Requirements Device performance during active alarm Medium The measured oxygen inlet pressure is greater than or equal to 87psig. Check the O2 source. Additional oxygen Device will continue to operate. Medium-priority Patient Alarms with Variable Settings HIGH TIDAL VOLUME Priority Why it occurs Medium Passive, active flow, or dual limb circuit types:
The estimated exhaled tidal volume is greater than or equal to the High Vte alarm setting for a number of consecutive breaths. The number depends on the therapy mode as follows. Three consecutive breaths: A/C-PC, CPAP, PSV, S/T, SIMV-VC, and SIMV-PC. Six consecutive breaths: A/C-VC Active PAP circuit type:
The delivered tidal volume is greater than or equal to the High Vti alarm setting for three consecutive breaths. Note: For active flow and dual limb, low flow O2 or the use of a nebulizer may increase tidal volumes beyond the alarm setting. The alarm is automatically resolved when a breath occurs in which the exhaled tidal volume does not reach the High Vte alarm setting. Device will continue to operate. 10 to 2000 ml in increments of 5 ml. Check the patient for pain or anxiety. Check the circuit for kinked or pinched tubing. Ensure that the active exhalation valve is attached. Medium Passive, active flow, or dual limb circuit types:
The estimated exhaled tidal volume is less than or equal to the low tidal volume alarm setting for a number of consecutive breaths. The number depends on the therapy mode as follows. Three consecutive breaths: A/C-PC, CPAP, PSV, S/T, SIMV-VC, and SIMV-PC. Six consecutive breaths: A/C-VC Active PAP circuit type:
The delivered tidal volume is less than or equal to the low tidal volume setting. Check the circuit for kinked, pinched, or blocked tubing. Ensure the leak device is not blocked, occluded, or disconnected. Ensure the diaphragm in the active exhalation device is inserted correctly. Check the mask fit or change the mask. Passive, active flow, or dual limb circuit types:
The alarm is automatically resolved when a breath occurs in which the exhaled tidal volume exceeds the low tidal volume alarm setting. Device will continue to operate. Active PAP circuit type:
Alarms and System Messages 53 What to do Device performance during active alarm Alarm Settings LOW TIDAL VOLUME Priority Why it occurs What to do Requirements Instructions for Use | Alarms and System Messages The alarm is automatically resolved when a breath occurs in which the exhaled tidal volume exceeds the low tidal volume alarm setting. Device will continue to operate. 10 to 2000 ml in increments of 5 ml. Alarm Settings HIGH MINUTE VENTILATION Medium Priority The patients minute ventilation is greater than or equal to the high minute ventilation alarm setting. Why it occurs What to do Check if the patient is in pain or in an anxious state. Note: A patient in pain or in an anxious state may breathe at a higher rate or at a higher tidal volume. Such conditions may contribute to higher minute ventilation. The alarm is automatically resolved when the calculated minute ventilation is less than the high minute ventilation alarm setting. Device will continue to operate. Adult and Pediatric patient types: 0.2 to 30 LPM in minimum increments of 0.1 LPM. Infant patient type: 0.2 to 10 LPM in minimum increments of 0.1 LPM. Device performance during active alarm Alarm Settings HIGH RESPIRATORY RATE Priority Why it occurs Check the patient for pain or anxiety Check the ventilator for auto-triggering Medium The respiratory rate is greater than the High Respiratory Rate alarm setting. When the trigger type is Off then the spontaneous respiratory rate will not trigger the alarm. Note: A patient in pain or in an anxious state may breathe at a higher rate. The alarm is automatically resolved when the measured respiratory rate is less than the high respiratory rate alarm setting. Device will continue to operate. Off, 1 to 90 BPM in increments of 1 BPM. LOW INSPIRATORY PRESSURE ALARM Priority Why it occurs Medium The measured peak inspiratory pressure is less than or equal to the low inspiratory pressure alarm setting. Check the patient for changes that may cause this alarm, such as excessive patient inspiratory effort Check the ventilator for o o Kinked, pinched, or blocked tubing Circuit leak or disconnect Volume Modes:
- CV
- SIMV-VC
- AC Device will continue to operate. The alarm is automatically resolved when the measured peak inspiratory pressure is greater than the low inspiratory pressure alarm setting. PEEP+1 to 89 cmH2O in increments of 1 cmH2O. Alarms and System Messages 54 What to do Device performance during active alarm Alarm Settings What to do Requirements Device performance during active alarm Alarm Settings LOW SPO2 Priority Why it occurs What to do Requirements Device performance during active alarm Alarm Settings HIGH SPO2 Priority Why it occurs What to do Requirements Device performance during active alarm Alarm Settings LOW ETCO2 Priority Why it occurs What to do Requirements Device performance during active alarm Alarm Settings HIGH ETCO2 Priority Why it occurs What to do Requirements Device performance during active alarm Alarm Settings Instructions for Use | Alarms and System Messages Ensure that oxygen tubing is attached to the ventilator and the oxygen source. Ensure that the oxygen source is providing oxygen. Increase oxygen flow or increase FiO2. Medium The measured SpO2 is less than or equal to the Low SpO2 alarm setting while the device is delivering therapy or is in standby. Requires pulse oximeter The alarm is automatically resolved when the measured SpO2 rises above the Low SpO2 alarm setting. Device will continue to operate. 50 to 95% in increments of 1%. Ensure that the oxygen source is appropriate. Decrease percentage of oxygen delivery. Medium The measured SpO2 is greater than or equal to the High SpO2 alarm setting. Requires pulse oximeter The alarm is automatically resolved when the measured SpO2 falls below the High SpO2 alarm setting. Device will continue to operate. 90 to 100% in increments of 1%. Medium The measured EtCO2 is less than or equal to the Low EtCO2 alarm setting for 10 seconds while the device is delivering therapy. Check the patient. If there is a tracheal tube, verify it is properly inserted. Verify that leak is not excessive. High leak will reduce the measured EtCO2. Review the ventilation parameters including tidal volume, minute ventilation and respiratory rate and verify that they are appropriate. CO2 sensor The alarm is automatically resolved when the measured EtCO2 falls below the High EtCO2 alarm setting. Device will continue to operate. 1 to 100 mmHg in increments of 1 mmHg. Passive circuit: check for an insufficient leak Active and dual limb circuits: ensure the valve is operable Medium The measured EtCO2 is greater than or equal to the high EtCO2 alarm setting for 10 seconds while the device is delivering therapy. CO2 sensor The alarm is automatically resolved when the measured EtCO2 falls below the high EtCO2 alarm setting. Device will continue to operate. 1 to 100 mmHg in increments of 1 mmHg. Alarms and System Messages 55 LOW FIO2 Priority Why it occurs What to do Requirements Device performance during active alarm Alarm Settings HIGH FIO2 Priority Why it occurs What to do Requirements Device performance during active alarm Alarm Settings Instructions for Use | Alarms and System Messages Confirm proper connection of the oxygen source. If the alarm persists, calibrate the oxygen sensor. See chapter 7, Device Options. Medium The measured FiO2 is less than or equal to the low FiO2 alarm setting for 10 seconds in active therapy. When using low-flow oxygen, check for unnecessary leaks. Increase flow rate if possible. When using high pressure oxygen, check for excessive leaks. FiO2 sensor The alarm is automatically resolved when the measured FiO2 exceeds the low FiO2 alarm setting in active therapy. Device will continue to operate. 21 to 95% in increments of 1%. Confirm proper connection of the oxygen source. If the alarm persists, calibrate the oxygen sensor. See chapter 7, Device Options. Medium The measured FiO2 is greater than or equal to the high FiO2 alarm setting for 10 seconds. When using low-flow oxygen, decrease the flow rate if possible. FiO2 sensor 100% O2 must not be active. This alarm is disabled when 100% O2 is active The alarm is automatically resolved when the measured FiO2 falls below the high FiO2 alarm setting in active therapy (for 10 consecutive seconds). Device will continue to operate. 27 to 100% in increments of 1%
Low-priority System Alarms STUCK KEY Priority Why it occurs What to do Device performance during active alarm Medium A stuck On/Off (Standby) button or Alarm Silence button has been detected for at least 120 seconds. Contact Philips Respironics Customer Service Device will continue to operate. INLET FILTER BLOCKED Priority Why it occurs What to do Requirements Device performance during active alarm Algorithm summary Low The inlet filter becomes blocked and delivered therapy is reduced. Remove the filter. If the filter is the air-inlet foam filter, rinse it according to the instructions in the Cleaning and Disinfection chapter. Otherwise, replace the filter. Inlet filter installed Device will continue to operate. Because the inlet filter is blocked, the pressure generated by the ventilator is less than 75% of the estimated outlet pressure based on the nominal performance characteristics of the device. CHECK PROXIMAL PRESSURE LINE Priority Low Alarms and System Messages 56 Why it occurs What to do Requirements Device performance during active alarm Algorithm summary Instructions for Use | Alarms and System Messages The proximal pressure line may be improperly connected or it may contain water droplets that affect the pressure reading. Verify the proximal pressure line is connected properly at both ends and the line is clean and untangled. If necessary, clear the line and reconnect. Proximal pressure line attached Device will continue to operate. Monitored parameters and alarms that use the pressure measurement will not function Monitored parameters and alarms that use the flow measurement, such as tidal volume, will continue to function During periods of the breath when there is low flow in the inhalation limb, the pressure measured at the outlet of the device is compared to the pressure measured by the proximal line. When the difference is greater than 5 cm H20 at low flow for three consecutive breaths, the alarm occurs. Low-priority Patient Alarms with Variable Settings LOW PULSE RATE Priority Why it occurs Low In the standby and therapy states, the measured pulse rate is less than or equal to the low-pulse-rate alarm setting. Check probe placement and reposition Pulse oximeter reporting correctly for the previous three seconds The alarm is automatically resolved when the measured heart rate exceeds the low pulse rate alarm setting. Device will continue to operate. 18 to 300 beats per minute in increments of 1 beat per minute. Check oxygen flow to the ventilator Low In the standby and therapy states, the measured pulse rate is greater than or equal to the high-pulse-
rate alarm setting. Pulse oximeter reporting correctly for the previous three seconds. The alarm is automatically resolved when the measured heart rate falls below the high pulse rate alarm setting. Device will continue to operate. 18 to 300 beats per minute in increments of 1 beat per minute. What to do Requirements Device performance during active alarm Alarm Settings HIGH PULSE RATE Priority Why it occurs What to do Requirements Device performance during active alarm Alarm Settings Power Alarms LOW BATTERY Warning:
If the high-priority Low Battery alarm occurs, immediately connect the ventilator to an alternate power source. If no alternate power source is available, immediately place the patient on an alternate source of ventilation. Priority Why it occurs Medium priority when the last available battery is the detachable or internal battery and can provide at least 20 minutes of therapy and no more than 45 minutes of therapy Medium priority when the last available battery is the external DC source and can provide at least 20 minutes of therapy. High priority when the last available battery can provide at least 10 minutes of therapy. The last battery available is low or nearly depleted. This includes the internal battery. Alarms and System Messages 57 Instructions for Use | Alarms and System Messages The alarm is medium priority and the icons are yellow when 25 minutes of time remains. The alarm is high priority and the icons are red when 15 minutes of time remains. Switch to alternate battery or AC power to recharge the low battery. If the low battery is recharged and the alarm continues, replace the battery. Device will continue to operate. Low AC power is disconnected while the ventilator is providing therapy or in standby using AC power. Device will continue to operate. Switch to alternate power source. Check AC connection What to do Device performance during active alarm AC DISCONNECTED Priority Why it occurs What to do Device performance during active alarm INTERNAL BATTERY IN USE Low Priority The power source has switched to the internal battery. Why it occurs You start therapy on internal battery power. Device will continue to operate. What to do Confirm the remaining battery capacity. This is the last available power source. Prepare an alternative power source such as AC power or an external battery. Device performance during active alarm REPLACE DETACHABLE BATTERY Priority Why it occurs What to do Low The detachable battery has failed or the battery state of health is less than or equal to 60%. Device will continue to operate. If the condition still exists 60 minutes after you manually reset the alarm, the alarm repeats. Replace the detachable battery. If the alarm continues, plug the device into an alternate AC power source. INTERNAL BATTERY DEPLETED Priority Why it occurs What to do Low The internal battery is depleted. Plug the device into an alternate AC power source, or connect to a fully charged detachable or external battery. Device will continue to operate. Device performance during active alarm Device performance during active alarm Alarms and System Messages 58 System Messages Message External DC Source Disconnected Depleted External DC Check External DC Instructions for Use | Alarms and System Messages Cause Power is supplied from an external DC source and that power source is disconnected or depleted. The external DC power source is depleted. An external battery is connected but cannot supply sufficient power. Check for a faulty cable, bad connection, or bad battery. Depleted Detachable Battery The detachable battery is depleted. Internal Battery Not Charging Temperature The system is unable to charge the internal battery due to temperature. Internal Battery Not Discharging Temperature Detachable Battery Not Charging Temperature Detachable Battery Not Discharging Temperature Start On Battery Circuit Mismatch ActivePAP, passive, or dual limb/active flow circuits only The internal battery is unable to power the device due to high or low temperature. Change the environmental temperature or relocate the device. The system is unable to charge the detachable battery due to temperature. The system is unable to charge the detachable battery due to temperature. Trilogy Evo Universal is turned on and AC power is not detected. Reconnect AC power. Power is restored after losing all power and AC power is not detected. Reconnect AC power. The circuit type selected for use does not match the circuit connected to Trilogy Evo Universal for 3 consecutive breaths. Check the circuit type. Connect a circuit that matches the prescription. Unsupported Accessory Connected An unsupported accessory is connected to Trilogy Evo Universal. FiO2 sensor not calibrated Replace FiO2 sensor Pulse Oximeter Connected When FiO2 monitoring is enabled and an FiO2 sensor is connected but was not calibrated The FiO2 sensor is at the end of life or is no longer functioning. Indicates a pulse oximeter is now connected. Mainstream Capnography Sensor Connected Indicates a Mainstream Capnography Sensor is now connected Alarms and System Messages 59 Instructions for Use | Alarms and System Messages Prescription Alarm Availability by Therapy Mode Most alarms operate regardless of the therapy mode being delivered. However, certain alarms operate only when certain therapy modes are being delivered. The chart below lists those alarms and the therapy modes in which the alarms are available. Alarm Apnea
(requires Backup Ventilation) Circuit Disconnect High Tidal Volume Low Tidal Volume High Minute Ventilation Low Minute Ventilation High Respiratory Rate Low Respiratory Rate High Inspiratory Pressure Low Inspiratory Pressure High SpO2 Low SpO2 High Pulse Rate Low Pulse Rate High etCO2 Low etCO2 High FiO2 Low FiO2 Testing Alarms A/C-PC A/C-VC CPAP X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X PSV X S/T X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X SIMV-PC SIMV-VC X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X X Test alarms any time you make a significant change to the system. Testing Circuit Disconnection Alarms For ventilator-dependent patients, do not rely on any single alarm to detect when a circuit is disconnected. One or more of the following alarms may indicate a disconnected circuit.
-
-
-
-
-
-
Circuit Disconnected Low Tidal Volume Low Minute Ventilation Low Respiratory Rate Low Peak Inspiratory Pressure alarms (user settable for volume modes) Leakage alarm (Active PAP circuit only) To test that these alarms detect a circuit disconnection:
1. Ensure that the patient is connected to the ventilator and that ventilation is stabilized. Ensure that none of the alarms above are active. 2. Disconnect the circuit at the patient end of the circuit. Remove the leak device (passive circuits) and any humidifiers or other circuit accessories at the patient end of the circuit. Only the circuit tube should remain. 3. Confirm that one or more of the above-listed alarms activate. 4. Reconnect the circuit and confirm that any active alarm automatically resets. Testing Circuit Obstruction Alarms For ventilator-dependent patients, do not rely on any single alarm to detect when a circuit is obstructed. One or more of the following alarms may indicate an obstructed circuit.
-
-
Obstruction High Inspiratory Pressure Testing Alarms 60 Instructions for Use | Alarms and System Messages
-
-
-
-
-
-
-
Circuit Disconnected Low Tidal Volume Low Minute Ventilation Low Respiratory Rate Low Peak Inspiratory Pressure alarms (user settable for volume modes) Leakage alarm (Active PAP circuit only) Rebreathing Detected To test that these alarms detect a circuit obstruction:
1. Ensure that the patient is connected to the ventilator and that ventilation is stabilized. 2. For a Passive circuit: Disconnect the circuit at patient connection port and remove the leak device and block the end of the circuit. For Active circuit: Disconnect the circuit at patient connection port and block the end of the circuit. 3. Confirm that one or more of the above-listed alarms activate. 4. Reconnect the circuit and confirm that any active alarm automatically resets. Testing the Leakage Alarm For an Active PAP circuit, the Leakage alarm detects a leak in the active exhalation valve (AEV). To test the Leakage alarm:
1. Ensure that the patient is connected to the ventilator and that ventilation is stabilized. 2. Disconnect the AEV control line from the ventilator. 3. Confirm that the Circuit Leakage, and/or Check Active Exhalation Valve Line alarm activates. 4. Reconnect the AEV control line and confirm that the alarm automatically resets. Testing the Low FiO2 Alarm The Low FiO2 alarm requires an FiO2 sensor to be connected and the FiO2 sensor setting to be turned on. To test the Low FiO2 alarm:
1. Ensure that the patient is connected to the ventilator and that ventilation is stabilized. 2. Disconnect the oxygen from the ventilator 3. Confirm that the Low FiO2 alarm activates. 4. Reconnect the oxygen and confirm that the alarm automatically clears, which may take 30 seconds or more. Testing Power Alarms Power indicators in the Status Bar:
To test power alarms:
1. Connect Trilogy Evo Universal to AC power. 2. Ensure Trilogy Evo Universal is using AC power. You should see the green LED light next to the power button. 3. Disconnect AC power (pull the power cord out of the outlet). 4. Confirm that the AC Disconnected alarm activates. 5. 6. Confirm that the device continues to operate, the external battery icon appears in the Status Bar, and that the Battery-in-use Indicator is If connected to an external battery, go to the next step. Otherwise go to step 9. pointing to the external battery icon. Testing Alarms 61 Instructions for Use | Alarms and System Messages 7. Disconnect the external battery. 8. Confirm that the External DC Source Disconnected system message appears. 9. Confirm that the device continues to operate, the detachable battery icon appears in the Status Bar, and that the Battery-in-use Indicator is pointing to the detachable battery. 10. Remove the detachable battery. 11. Confirm that the Internal Battery in Use alarm activates. 12. Confirm that the device continues to operate, that the internal battery icon appears in the Status Bar, and that the Battery-in-use indicator is pointing to the internal battery. To test the low battery alarm:
1. Ensure that AC power is available. 2. Disconnect all power sources and remove the detachable battery. 3. Allow the device to use internal battery power. 4. Confirm that the medium-priority Low Battery alarm activates and that the device continues to operate. 5. Continue to allow the device to use internal battery power. 6. Confirm that the high-priority Low Battery alarm activates and that the device continues to operate. 7. Insert the detachable battery and reconnect AC power to recharge the batteries. Testing Therapy Setting Alarms When testing alarms, remember that you can use the Alarm Silence button. To test alarms related to therapy settings:
Set the therapy mode to A/C-VC. Set the Low Tidal Volume alarm limit above the measured tidal volume value. Set the High Tidal Volume alarm limit below the measured tidal volume value. 1. 2. Connect a circuit to the device and to a test lung. 3. Observe the measured values. 4. 5. Confirm that the High Tidal Volume alarm activates. 6. Restore the alarm limit. 7. 8. Confirm the Low Tidal Volume alarm activates. 9. Restore the alarm limit. 10. Set the High Minute Ventilation alarm limit below the measured minute ventilation value. 11. Confirm that the High Minute Ventilation alarm activates. 12. Restore the alarm limit. 13. Set the Low Minute Ventilation alarm limit above the measured minute ventilation value. 14. Confirm that the Low Minute Ventilation alarm activates. 15. Restore the alarm limit. 16. Set the High Respiratory Rate alarm limit below the measured respiratory rate value. 17. Confirm that the High Respiratory Rate alarm activates. 18. Restore the alarm limit. 19. Set the Low Respiratory Rate alarm limit above the measured respiratory rate value. 20. Confirm that the Low Respiratory Rate alarm activates. 21. Restore the alarm limit. 22. Set the High Inspiratory Pressure alarm limit below the measured PIP value. 23. Confirm that the High Inspiratory Pressure alarm activates. 24. Restore the alarm limit. 25. Set the Low Inspiratory Pressure alarm limit above the measured PIP value. 26. Confirm that the Low Inspiratory Pressure alarm activates. 27. Restore the alarm limit. Testing Patient-Monitoring Alarms Before testing these alarms, ensure that the patient is connected to the ventilator and that ventilation is stabilized. Ensure that the monitor is connected and functional. To test alarms related to patient monitoring where the alarm limits are settable:
Testing Alarms 62 Instructions for Use | Alarms and System Messages Set the High EtCO2 alarm limit below the measured EtCO2 value. ETCO2 1. 2. Confirm that the High EtCO2 alarm activates. 3. Restore the alarm limit. 4. 5. Confirm the Low EtCO2 alarm activates. 6. Restore the alarm limit. Set the Low EtCO2 alarm limit above the measured EtCO2 value. PULSE RATE Set the High Pulse Rate alarm limit below the measured Pulse Rate value. 1. 2. Confirm that the High Pulse Rate alarm activates. 3. Restore the alarm limit. 4. 5. Confirm the Low Pulse Rate alarm activates. 6. Restore the alarm limit. Set the Low Pulse Rate alarm limit above the measured Pulse Rate value. Set the High SpO2 alarm limit below the measured SpO2 value. SPO2 1. 2. Confirm that the High SpO2 alarm activates. 3. Restore the alarm limit. 4. 5. Confirm the Low SpO2 alarm activates. 6. Restore the alarm limit. Set the Low SpO2 alarm limit above the measured SpO2 value. Alarm and Event Log The Alarm and Event Log is a record of all events related to the device and to therapy. The log shows each event, when it occurred, and a brief description. Information is retained even when you shut off the device or when power is lost. The log stores the most recent 6-months of information with the exception of the event log, which stores the most recent 10,000 records. Older records are overwritten. To access the Alarm and Event Log:
1. 2. In the menu bar, tap the Options icon. In the Options window, tap Alarm & Event Log. Alarm and Event Log 63 Instructions for Use | Alarms and System Messages Parts of the Alarm and Event Log Item 1. Date Range 2. Events 3. Alarm and event list Description Tap to filter events by date. On the Select Date Range dialog box, select the date range and then tap Accept. Tap to filter events so only alarms appear in the list. Tap again to view all event types. Event Type High-Priority Alarm Medium-Priority Alarm Low-Priority Alarm Info-Type Alarm Event Power or battery Physical button press Calibration event Change to therapy setting, alarm setting, or device option Data transfers Icon 4. Ascending/Descending button 5. Page up and down 6. Clear Sort the list by date and time. Tap to scroll through the log. Clear all events from the screen. Alarm and Event Log 64 7. Device Options Instructions for Use | Device Options Use the features contained in the Options window to change device settings, run calibrations and tests, and view and work with data. Device options include the following features:
-
-
-
-
-
Device Settings Calibration Data Transfer Alarm & Event Log see the Alarms and System Messages chapter Information Device Settings Use the Device Settings feature to customize Trilogy Evo Universal. When working with settings, ensure you save your changes. After 30 seconds of inactivity, the system reverts to the previous setting and your changes are not saved. To change a setting:
In the menu bar, tap the Options icon. In the Options window, tap Device Settings. In the Device Settings window, tap the setting you want to change. In the setting dialog box, make your selection. 1. 2. 3. 4. 5. When your selection is complete, on the title bar, tap the Accept checkmark. Setting Language Description Set the device language. Alarm Volume Set system alarm loudness. Screen Brightness Set screen brightness. Light Bar Turn the light bar on or off. Automatic Touchscreen Lock Screen Saver Automatically locks the screen after five minutes of inactivity. The screen automatically unlocks during an alarm. Select the type of screen saver that you want to use. Date and Format Time and Format Manual Breath FiO2 Sensor NFC CMD Device Units Patient Units Bluetooth Set the system date and format. Set the system time and select 12- or 24-hour format. Turn the manual breath feature on or off. Turn the FiO2 sensor on or off. Turn nearfield communication on or off. Turn CMD (communication means device) status on or off. Select the pressure and CO2 measurement units. Select metric or imperial measurement units.
-
To enable a Bluetooth connection: tap Bluetooth. On the dialog box, tap On. Note that a Bluetooth symbol appears in the Status Bar to indicate when devices are connected. To disable the Bluetooth connection: tap Bluetooth. On the dialog box, tap Off. To clear all devices from memory: tap Bluetooth. On the dialog box, tap Forget All Devices.
(Disabled when Bluetooth is not enabled.)
-
-
Note: Bluetooth functionality may not be present in all models. Device Settings 65 Instructions for Use | Device Options Calibration Circuit Calibration Concepts The Trilogy Evo Universal is optimized for circuits that are within the specifications shown in the Circuit section of the Accessories chapter. If you want to use a different circuit, you can perform an optional Circuit Calibration, intended to characterize compliance and resistance. The circuit calibration process includes the following procedures, based on the type of prescription:
Depending on the circuit type:
Active circuits: calibrates according to the results of the leak test, compliance, and resistance Passive circuits: calibrates according to the compliance and resistance When you start the circuit calibration, follow the instructions on the screen as the system completes the tests. If the calibration is successful, then you will see a confirmation message. If the circuit fails any part of the test, the reason appears in the window. Adjust the circuit and repeat the calibration. If you repeat the calibration but the circuit still fails, you can either replace the circuit and try again or you can use the default settings. Information about the circuit calibration is recorded in the Event Log. For more information about the Event Log, see Alarm and Event Log in the Alarms chapter. CALIBRATING A CIRCUIT Requirements:
-
-
The patients demographics and prescription are entered into the system The patient interface is not attached to the circuit. To calibrate a circuit:
In the menu bar, tap the Options icon. 1. In the Options window, tap Calibration & Setup. 2. In the Calibration & Setup window, tap Circuit Calibration. 3. 4. In the Calibrate Circuit window, in the Current Prescriptions list, locate the prescription you want to calibrate and then tap Calibrate. Follow the instructions on the screen.
-
-
If any part of the test fails, correct the issue suggested on the screen and then tap Retest to continue the test. To cancel the test, tap the Close icon at the top of the window. USING DEFAULT SETTINGS If you want to stop using calibrated settings and return to the default settings, follow these instructions:
In the menu bar, tap the Options icon. 1. In the Options window, tap Calibration & Setup. 2. 3. In the Calibration & Setup window, tap Circuit Calibration. In the Calibrate Circuit window, in the Current Prescriptions list, locate the prescription you want to calibrate and then tap Set to Defaults. 4. On the confirmation window, tap Yes. Calibration 66 Instructions for Use | Device Options Leak Test A leak test is available for times when you want to perform only a leak test and not complete an entire circuit calibration. Prerequisites:
Ensure the prescription is for an active circuit:
-
-
-
Active Flow Active PAP Dual Limb Remove the patient interface from the circuit. Block the end of the circuit where the patient interface would be. If the circuit includes a heated humidifier, ensure the humidifier chamber is filled. Ensure the external active exhalation valve is properly assembled and connected. To perform a leak test:
In the menu bar, tap the Options icon. 1. In the Options window, tap Calibration and Setup. 2. 3. Review the prerequisites and then tap Start. 4. The system performs the test. The results appear in the test progress pane. If the test:
In the Calibration & Setup window, tap Leak Test. Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap the cancel button at the top of the window. Passes: tap OK. O2 Sensor Calibration Prerequisites:
Ensure the patient circuit is connected. Remove the patient interface from the circuit. Remove any passive exhalation device. Connect high-pressure O2. Ensure that low-flow O2 is not connected. To calibrate the O2 sensor:
In the menu bar, tap the Options icon. In the Options window, tap Calibration and Setup. In the Calibration & Setup window, tap O2 Sensor Calibration. 1. 2. 3. 4. Review the prerequisites and then tap Start. 5. The system performs the Circuit Flush test. The results appear in the progress pane. If the test:
Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap Quit. Passes: go to the next step. 6. Block the end of the circuit and then tap Continue. 7. The system performs the 21% Calibration test. The results appear in the progress pane. If the test:
Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap Quit. Passes: go to the next step. If you are using an oxygen-blending module, confirm that high-pressure O2 is connected and the circuit outlet is blocked and then tap Continue. Otherwise, the test is complete. Calibration 67 8. 9. The system performs the test. The results appear in the test progress pane. If the test:
Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap Quit. To deactivate the FiO2 sensor, turn off the FiO2 sensor in Device Options, Device Settings. Passes: tap OK. Instructions for Use | Device Options CO2 Sensor Adapter Zero Perform this task when installing a CO2 sensor (capnograph) To establish a baseline CO2 level:
1. Place the CO2 sensor onto a clean and dry airway adapter that is exposed to room air, but is away from all CO2 sources including the ventilator, your breath, and the patients breath. In the menu bar, tap the Options icon. In the Options window, tap Calibration and Setup. In the Calibration & Setup window, tap CO2 Sensor Adapter Zero. 2. 3. 4. 5. Review the prerequisites and then tap Start. 6. The system performs the test. The results appear in the test progress pane. If the test:
Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap the Quit. Passes: tap OK. Data Transfer Use data transfer functions to import and export various sets of data. When working with patient data, ensure the data is kept secure. When working with an external storage device that contains patient information, such as a USB flash drive, ensure you delete that data by reformatting the storage device. To access the data transfer functions:
1. Connect a data storage device to the USB port on the Service Panel. 2. 3. In the menu bar, tap the Options icon. In the Options window, tap Data Transfer. Data Transfer 68 In the Data Transfer window, tap the procedure you would like to perform and then follow the prompts on the screen:
Instructions for Use | Device Options 4. Function Export Data - USB Export Data - Bluetooth Description Export details about the patients therapy to a storage device connected to the USB port. Data is identified by the device serial number. Data type Changes to the ventilator, alarms, prescriptions, accessories and oxygen 30-second averages of each therapy parameter related to the therapy provided to the patient Waveform data including pressure, flow, volume, and total leak Patient monitors:
Data stored 6 months 31 days
-
-
-
CO2 monitor data Oximeter data FiO2 monitor data Export details about the patients therapy to a storage device connected through Bluetooth. Data is identified by the device serial number. Data type Changes to the ventilator, alarms, prescriptions, accessories and oxygen 30-second averages of each therapy parameter related to the therapy provided to the patient Waveform data including pressure, flow, volume, and total leak Patient monitors:
Data stored 6 months 31 days
-
-
-
CO2 monitor data Oximeter data FiO2 monitor data Export Alarm & Event Log Install Prescriptions Install Software Update Export the alarm and event logs. If two storage devices are connected, the logs are saved to the device that was connected first. Installs prescriptions that were previously exported. Ensure the version number that you are installing is higher than the current software version. After installing the update, the system automatically restarts. As data is transferred, a data transfer icon appears in the Status Bar. USB Data Transfer Icon Bluetooth Data Transfer icon Data Transfer 69 Information Instructions for Use | Device Options The information window shows general information about the Trilogy Evo Universal device, including the following:
To change pages, tap the page icons at the bottom of the window. To view software licenses and photography credits, tap the item you want to view. Page 1
-
-
-
-
Page 2
-
-
-
-
Page 3
-
-
DME Representative Name DME Contact Phone DME Contact Email DME Company
- Model Number
-
Serial Number
-
Software Version Number
-
Hardware Version Number Internal Battery State of Health Detachable Battery State of Health Internal Battery Serial Number Detachable Battery Serial Number
-
-
Operational Hours Total Blower Hours Operational Hours Total Patient Hours Software licenses Photography credits Information 70 Instructions for Use | Cleaning and Disinfection 8. Cleaning and Disinfection This chapter contains instructions for cleaning and disinfecting Trilogy Evo Universal, and cleaning, disinfecting, and sterilizing Trilogy Evo Universal accessories. Because Trilogy Evo Universal is intended for multi-patient use, ensure you follow the cleaning, disinfection, and sterilization instructions in this chapter. Exterior Cleaning and Disinfection Warning: To avoid electric shock, do not remove the enclosure cover. Only service personnel should remove the enclosure. After cleaning and disinfecting, ensure the device is completely dry before reattaching accessories and connectors and before reconnecting it to a power source. To avoid electrical shock, always unplug the power cord from the wall outlet before cleaning the ventilator. If the device has been exposed to rain or dampness, dry the device including the area around the power cord connection with the power cord disconnected from the device before applying AC power. Caution: Do not immerse the device or allow liquids into any of the controls or the interior of the enclosure as the device may be damaged. If this occurs, contact your equipment provider for assistance. Use only the cleaning agents and methods described in this section to clean and disinfect the device. Cleaning the Exterior Frequency: Clean Trilogy Evo Universals exterior surface weekly and between patients. Requirements: lint-free cloth, soft-bristle brush, and liquid dishwashing detergent solution: 1 teaspoon of liquid dishwashing detergent (such as Dawn Ultra Dishwashing Liquid) per gallon of water 1. Turn Trilogy Evo Universal off and disconnect it from the power source. 2. Detach all accessories and connectors. 3. Use a lint-free cloth dampened (not dripping) with a liquid dishwashing detergent solution to clean the exterior of the enclosure. 4. Use a soft-bristle brush in the areas around the screen, buttons, and any other areas where soil may be difficult to remove. Ensure you remove all visible soil. 5. Use a lint-free cloth dampened (not dripping) with clear water to remove all detergent residue. 6. Use a lint-free cloth to dry the enclosure. 7. 8. Repeat the cleaning steps until the surfaces are visibly clean. 9. Inspect the device for cleanliness. Inspect the device for damage after cleaning. If any parts are damaged, contact Philips Respironics Customer Service. Disinfecting the Exterior Frequency: Disinfect Trilogy Evo Universals exterior surface weekly and between patients. Prerequisite: Before disinfecting Trilogy Evo Universals exterior, ensure you have cleaned Trilogy Evo Universal as instructed in the previous section, Cleaning the Exterior. ISOPROPYL ALCOHOL Use a minimum of 70% isopropyl alcohol but no more than 91% by volume. 1. Use a lint-free cloth dampened with alcohol to clear visible soil from the surfaces. 2. Use a second lint-free cloth to wipe the alcohol onto the exterior, thoroughly wetting the surfaces. 3. Keep wet 10 minutes. 4. Allow to air dry. CHLORINE BLEACH Use a household chlorine bleach containing 8.25% sodium hypochlorite. 1. Combine 10 parts water to 1 part bleach. 2. Use a lint-free cloth dampened with the bleach solution to clear visible soil from the surfaces. 3. Use a second lint-free cloth to wipe the bleach solution onto the exterior, thoroughly wetting the surfaces. 4. Keep wet 10 minutes. 5. Allow to air dry. Exterior Cleaning and Disinfection 71 Instructions for Use | Cleaning and Disinfection Cleaning the Detachable Battery Frequency: Clean the detachable battery monthly and between patients. Requirements: lint-free cloth, soft-bristle brush, and liquid dishwashing detergent solution: 1 teaspoon of liquid dishwashing detergent (such as Dawn Ultra Dishwashing Liquid) per gallon of water 1. Remove the detachable battery. Open the detachable battery access door. Lift the battery handle and pull the battery to remove it from the battery bay. 2. Use a lint-free cloth dampened (not dripping) with a liquid dishwashing detergent solution to wipe the battery. Ensure you remove all visible soil. 3. Use a soft-bristle brush to clean any small areas, such as crevices or small openings that are not accessible with the cloth. 4. Use a lint-free cloth dampened (not dripping) with clear water to remove all detergent residue. 5. Allow the battery to air dry completely. 6. Inspect the battery for damage after cleaning. If any part is damaged, contact Philips Respironics Customer Service. 7. Replace the battery. Open the detachable battery access door. Slide the battery into the bay until you hear a click. Rinsing the Air-Inlet Foam Filter The air-inlet foam filter is the gray foam located on the back panel. It protects Trilogy Evo Universal from dirt and dust. Only use Philips Respironics-supplied filters. Ventilation can continue while you are replacing the filter. Frequency: For invasive ventilation, rinse daily and dispose monthly. For non-invasive ventilation, rinse monthly and dispose yearly. Dispose between patients. Requirements: replacement filter, water To rinse the disposable inlet filter:
1. Ensure you have a replacement filter nearby. 2. Pinch the filter and pull it out of the filter cover. 3. 4. Visually inspect the filter you just removed from the device. 5. Insert the clean replacement filter into the filter cover. Ensure it is positioned securely. If it is damaged, discard it according to your local regulations. Otherwise, proceed to the next step. 6. Rinse the dirty filter in clear water. Inspect the filter for cleanliness and repeat previous step until the filter is clean. 7. Allow the filter to air dry completely before reinstalling it. Rinsing the Air-Inlet Foam Filter 72 Instructions for Use | Cleaning and Disinfection External Active Exhalation Valve Cleaning and Disinfection Frequency: Clean the active exhalation valve weekly. Clean and disinfect or sterilize the active exhalation valve between patients. Replace the silicon diaphragm each time you disinfect or sterilize the valve. Cleaning the External Active Exhalation Valve Requirements: Lint-free cloth or cotton swab, soft-bristle brush, liquid dishwashing detergent solution, such as Dawn Ultra Dishwashing Liquid: 1 teaspoon detergent to 1 gallon of water Figure 2 Figure 3 Figure 4 1. Remove the valve from Trilogy Evo Universal: Squeeze the two tabs and pull out from the device (Figure 1). 2. Open the valve: Squeeze the two tabs (Figure 2)and pull apart (Figure 3). 3. Detach the silicon diaphragm. (If you are going to disinfect or sterilize the valve, discard the diaphragm according to your local regulations on discarding biohazardous waste.) Submerge the valve parts in warm water with a liquid dishwashing detergent solution. 4. 5. Agitate the parts to ensure all internal and external surfaces are saturated with the solution. 6. Use a lint-free cloth or cotton swab to remove all visible soil. 7. Use a soft-bristle brush to clean any small areas, such as crevices or small openings that are not accessible with the cloth or swab. 8. Rinse thoroughly with clear water to remove all detergent residue. 9. Allow the valve to air dry completely before reinstalling it. 10. Examine the valve. If it is damaged, discard it. If you are going to disinfect the valve, go to Disinfecting the External Active Exhalation Valve. Otherwise, continue to the next step. 11. Replace the diaphragm as shown in the illustration. Ensure you have positioned the diaphragm correctly. 12. Reassemble the valve and replace it as shown in Figures 4 and 5. Figure 5 Figure 6 Disinfecting the External Active Exhalation Valve Requirements:
Replacement silicon diaphragm 2.4% Activated Glutaraldehyde Solution (CIDEX) Distilled water External Active Exhalation Valve Cleaning and Disinfection 73 To disinfect the valve:
Instructions for Use | Cleaning and Disinfection 1. Disassemble the valve, discarding the silicon diaphragm according to your local regulations on discarding biohazardous waste. Then proceed to clean and dry the valve according to the instructions in the previous section, Cleaning the External Active Exhalation Valve. 2. Read and follow the manufacturers instructions on how to prepare, activate, use, and dispose of the Cidex solution. Follow all safety 3. precautions. Submerge the clean and dry valve in Cidex solution. Agitate the parts to ensure all internal and external surfaces are saturated with the solution. Soak the valve in Cidex for 45 minutes at 20C. 4. Remove the valve from the solution and pour excess cleaning agent from the valve. 5. Rinse thoroughly with clear water. 6. Allow the valve to dry entirely. 7. Examine the valve. Replace any part that is damaged (visible cracks, distorted parts, or altered color). 8. 9. Reassemble the valve and replace it as shown in Figures 4 and 5 in the previous section. Insert a new diaphragm as shown in the illustration. Ensure you have positioned the diaphragm correctly. Sterilizing the External Active Exhalation Valve You can sterilize the external active exhalation valve using a hydrogen peroxide gas plasma sterilizer, such as Sterrad. Requirements: Replacement silicon diaphragm To sterilize the valve:
1. Disassemble the valve, discarding the silicon diaphragm according to your local regulations on discarding biohazardous waste. Then proceed to clean and dry the valve according to the instructions in the previous section, Cleaning the External Active Exhalation Valve. 2. Read and follow the instructions of your sterilization device. Follow all safety precautions. 1. Allow the valve to cool and dry entirely in a clean and dry environment. Caution: Allow any liquid to evaporate entirely before reconnecting the valve. 2. Examine the valve. Replace any part that is damaged (visible cracks, distorted parts, or altered color). 3. 4. Reassemble the valve and replace it as shown in Figures 4 and 5 in the previous section. Insert a new diaphragm as shown in the illustration. Ensure you have positioned the diaphragm correctly. External Active Exhalation Valve Cleaning and Disinfection 74 Instructions for Use | Service and Maintenance 9. Service and Maintenance Service Repairs and adjustments must be performed by service personnel only. Unauthorized service could cause death or injury, invalidate the warranty, or result in costly device damage. Service technicians can request circuit diagrams, component part lists, descriptions, and calibration instructions to help service the device. See the service manual. The device should be serviced for preventive maintenance every 24 months or 10,000 blower hours. To see the blower hours: in the menu bar, tap the Device Options icon, and then in the Options window, tap Information. Trilogy Evo Universals expected service life is 10 years. Disposal Separate collection for electrical and electronic equipment per EC Directive 2012/19/EU. Dispose according to local regulations. If you need help, contact Philips Respironics. Routine Maintenance Conduct basic maintenance according to the table below. Daily Monthly Every six months Yearly Visually inspect accessories for damage or signs of wear. Discontinue use and replace if damaged. When using oxygen, to maintain accuracy, calibrate the oxygen sensor daily. See "O2 Sensor Calibration" in the "Device Options" chapter. Replace the particulate filter (when in use) For invasive ventilation, dispose of the air-inlet foam filter. Ensure the internal battery has been charged at least once in the previous six months. The internal battery charges when connected to AC power. For non-invasive ventilation, dispose of the air-inlet foam filter. Replacing the Air-Inlet Foam Filter The air-inlet foam filter is the gray foam located on the back panel. It protects Trilogy Evo Universal from dirt and dust. This filter is for single patient use. For invasive ventilation, dispose monthly. For non-invasive ventilation, dispose yearly. Only use Philips Respironics-supplied filters. Dispose according to local regulations. Ventilation can continue while you are replacing the filter. To replace the disposable inlet filter:
1. Ensure you have a replacement filter nearby. 2. Pinch the filter and pull it out of the filter cover. 3. Insert the clean replacement filter into the filter cover. Ensure it is positioned securely. Replacing the Air-Inlet Foam Filter 75 Instructions for Use | Service and Maintenance Replacing the Particulate Filter The particulate filter is an optional filter that protects Trilogy Evo from dirt and dust. Replace the particulate filter monthly and between patients. Ventilation can continue while you replace the filter. Twist the filter cover counterclockwise a quarter of a turn, and then pull straight out to remove. Twist the filter counterclockwise a quarter of a turn, and then pull straight out to remove. Place a new filter onto the bayonet mount then twist the filter clockwise a quarter of a turn while pressing in to secure. Replace the filter cover and turn clockwise to secure. Preparing the Device for a Use by a Different Patient If you are setting up a prescription for a different patient, before creating a new prescription, in the Home window, tap the New Patient button to reset to the default prescription settings, reset the patient operational hours to zero, and clear all existing patient data, including the following:
event and alarm logs, circuit calibration, and historical data. Additionally, tapping New Patient deletes all Bluetooth settings. If you used an external storage device that contains patient information, such as a USB flash drive, ensure you delete that data by reformatting the storage device. Before using Trilogy Evo Universal with a new patient, perform the following actions. Cleaning and disinfection instructions are in the Cleaning and Disinfection chapter. Replace the circuit, including the bacteria filter. Clean and disinfect the active exhalation valve Clean and disinfect the external flow sensor Clean and disinfect the exterior surface and detachable battery. Replace the air inlet foam filter and particulate filter Clear the old patient data from the system: In the Home window, tap the New Patient button. If you used an external storage device that contains patient information, such as a USB flash drive, ensure you delete that data by reformatting the storage device. Preparing the Device for a Use by a Different Patient 76 Instructions for Use | Trilogy Evo Universal Accessories 10. Trilogy Evo Universal Accessories To prevent adverse performance, use Trilogy Evo Universal only with accessories intended for use with this device, including all patient interfaces, masks, circuits, exhalation ports, and carrying cases. For a list of accessories, see the Trilogy Evo Universal accessories guide at:
https://www.usa.philips.com/healthcare/product/HCDS2110X11B/trilogyevo. You must ensure accessories and parts are compatible before you connect a patient to the device. If you are using a remote alarm or nurse call, ensure you test the accessory before starting ventilation. The accessories described in this chapter are those included in your device package. For all other accessories, see the accessorys instructions. Power Accessories For instructions, see Power Management. Patient Monitors FiO2 sensor An FiO2 sensor is installed in your device. For instructions on calibrating the sensor, see Trilogy Evo Universal Options, O2 Sensor Calibration Intended Use: The Philips Respironics FiO2 Sensor is an oxygen sensor device that is used for measuring the fraction of inspired oxygen (oxygen concentration in a percentage) delivered from the oxygen source to the airway of a patient circuit (such as hose and mask). Storage and Handling Avoid rough handling that would damage the sensor. Avoid exposing sensor(s) to rapid changes in pressure. Avoid puncturing or damaging sensor membrane(s). Accidental release measures: The oxygen sensors contain a strong basic solution encapsulated in a plastic housing. Under normal operating conditions, the solution (electrolyte) is never exposed. In case of a leak, please observe the following instructions:
Personal precautions, protective equipment, and emergency procedures Use personal protective equipment. Avoid dust formation. Avoid breathing vapors, mist, or gas. Ensure adequate ventilation. Evacuate personnel to safe areas. Avoid breathing dust. Personal protective equipment Eye/face protection: Safety glasses with side-shields or googles conforming to appropriate government standards such as ANSI (US) or EN 166(EU) Skin protection: Handle with nitrile gloves. Gloves must be inspected prior to use. Use proper glove removal technique (without touching gloves outer surface) to avoid skin contact with this product. Dispose of contaminated gloves after use in accordance with applicable laws and good laboratory practices. Wash and dry hands. Respiratory and body protection: Wear respiratory protection and full protective clothing tested and approved under appropriate government standards such as ANSI (US) or CEN (EU). Environmental precautions Prevent further leakage or spillage if safe to do so. Do not let product enter drains. Discharge into the environment must be avoided. Methods and materials for containment and cleaning up in case of leaking or spilling Contain spillage. Neutralize spill with soda ash or lime. Carefully place material into clean dry container and cover. Flush spill area with water. Avoid creating dust. Disposal Patient Monitors 77 Instructions for Use | Trilogy Evo Universal Accessories Because this sensor contains lead and a corrosive solution, contact a licensed professional waste disposal service for disposal. Specifications:
Range Of Measurement:
0 TO 100%
Output Range:
Zero Offset:
9.0 TO 16.0 mV 0.250 mV when exposed to 100% nitrogen 90% Response Time:
512 seconds Linearity:
Drift:
Accuracy 3% full scale error maximum 1% volume O2 per month in air
(2.5% FiO2+2.5% of actual FiO2) Expected Lifetime:
>1,000,000 %02 hours Temperature Compensation: NTC Operational Temperature Range:
10 TO 40C Storage Temperature Range:
-20 TO 50C Humidity Range:
0% TO 99% RH, non-condensing Ambient Pressure Range:
700-1250mBAR Interference:
Per EN ISO 80601-2-55 Influence Of Humidity:
-0.03%REL. 02 reading per %RH Electrical Interface:
3 pin Molex Shelf life 2 year warranty / >1,000,000 %O2 hours Patient Monitors 78 To replace the sensor:
Tools required:
Instructions for Use | Trilogy Evo Universal Accessories
-
-
T20 Torx screwdriver Scratch awl or similar slender, pointed tool To remove the FiO2 sensor:
1. Using a T20 Torx screwdriver, unscrew the FiO2 sensor cover and set aside. Note that the cover is also the FiO2 sensor driver. 2. Use the awl to pull the tab lock away from the plug to the unlock position. 3. While holding the tab in the unlock position, pull the white connector straight out. Do not use the wires to pull the connector. 4. Place the FiO2 driver on the sensor and twist counterclockwise to remove the sensor from the device. 5. Because this sensor contains lead and a corrosive solution, contact a licensed professional waste disposal service to dispose of this material. To install a new FiO2 sensor:
1. Place the sensor in the FiO2 socket. 2. Place the FiO2 driver on the sensor and twist clockwise to secure the sensor in the receptacle. 3. Push the connector onto the pins on the sensor. Ensure you have aligned the pins with the receptacles on the connector. 4. Replace the FiO2 sensor cover and secure with the screws. Patient Monitors 79 Instructions for Use | Trilogy Evo Universal Accessories External Flow Sensor and Cable Intended Use:
The Philips Respironics External Flow Sensor is intended for use with Trilogy Evo Universal-series ventilators. It measures the gas (air, oxygen or carbon dioxide) flow rate in the airway from the patient circuit to the patients lung during inhalation and exhalation. One side of the sensor device is attached to the patient circuit while the other side is connected to the ventilator. Two sizes of this sensor are included in the case: adult/pediatric, and pediatric/infant. To use the sensor, follow the instructions provided with the sensor. To connect the external flow sensor to Trilogy Evo Universal, see Figure 1 and Figure 2. Figure 8: Attaching the sensor to a circuit Figure 7: Attaching the sensor to Trilogy Evo l Colorimetric CO2 detector Two sizes of this detector are included in the package. To use the detector, see the detectors instructions. External Pulse Oximeter and Sensors Three disposable pulse oximeter sensors are included in the package: adult, pediatric, and infant. Instructions for Use:
To use the oximeter sensors, follow the instructions provided with the sensors. To connect the oximeter to Trilogy Evo Universal, plug the USB connector into the USB port on the Patient Panel Filters Air-inlet foam filter The air-inlet foam filter protects Trilogy Evo Universal from dirt and dust. This filter is for single patient use. For instructions on rinsing the filter, see the Cleaning and Disinfection chapter. For instructions on replacing the filter, see the Service and Maintenance chapter. Particulate filter The particulate filter protects Trilogy Evo Universal from fine particulates. This filter is for single patient use. For instructions on replacing the filter, see the Service and Maintenance chapter. CBRN filter adapter Use the CBRN filter adapter when using a CBRN filter in place of the disposable inlet filter. For instructions on installing the filter adapter, see the Device Setup chapter. Filters 80 Patient Circuits Instructions for Use | Trilogy Evo Universal Accessories For instructions on attaching circuits to the ventilator, see the Device Setup chapter. Circuit Principles Each patient circuit has a different compliance and resistance. During Circuit Calibration, the ventilator verifies that the maximum pressure drop does not exceed the specifications below. For optimal performance, calibrate the circuit before using it with a patient. See Circuit Calibration in the Device Options chapter. When adding any components to the breathing system, the flow resistance and dead space of the added components such as humidifiers, speaking valves, Heat Moisture Exchangers (HMEs) and filters should be carefully considered in relation to the potential for adverse effects on the patients ventilator management and device alarms. If a breathing system filter is exposed to nebulization or humidification, to prevent increased resistance or blockage, the breathing system filter requires replacement more frequently. Circuit Requirements For safe operation, the ventilator requires a patient circuit and filter(s) that meet the following requirements. Compliance: up to 4 ml/cmH2O Inspiratory/Expiratory Resistance Pediatric/adult circuit (20-22 mm): less than 5 cmH2O at 30 l/min Pediatric/adult circuit (19 mm): less than 5 cmH2O at 15 l/min Infant/pediatric circuit (14-16 mm): less than 5 cmH2O at 15 l/min Infant/pediatric circuit (9-13 mm): less than 5 cmH2O at 2.5 l/min Leak Compensation The passive circuit provides leak compensation for inhaled and exhaled tidal volume measurements. This includes compensation for intentional leak in the patient circuit and leaks that occur at the patient interface, such as cuff leak or mask leak. The active flow and dual limb circuits compensate for leak that occurs between the ventilator and the external flow sensor. Leaks that are downstream of the external flow sensor are not compensated for in measurement of inhaled or exhaled tidal volume. Leak compensation is not available in the active PAP circuit. Circuit Accessories HMEs Three Heat and Moisture Exchangers (HMEs) are included in this package: adult, pediatric, and infant. Follow the instructions provided with the sensors. Flexible Trach Adapter Adult Endo-trach adapter with port for suctioning. Follow the instructions provided with the adapter. Circuit Accessories 81 Instructions for Use | Trilogy Evo Universal Accessories Dual Limb Active Exhalation Valve The Philips Respironics Dual Limb Active Exhalation Valve is a dual limb circuit accessory. The valve will close during inspiration and open during exhalation to divert flow into the expiratory limb for venting outside of the breathing circuit. This device is intended for multi-patient use. Clean and disinfect or clean and sterilize this item between uses on different patients. For help setting up, using, or maintaining this device, contact Philips Respironics. Warnings:
To reduce the likelihood of disconnection, and to prevent degraded or adverse ventilator performance, use this accessory only with a compatible ventilator. This device is compatible and validated for use with Philips Respironics Trilogy Evo Universal-series ventilators. Before use, ensure the compatibility of the ventilator and all of the parts, materials, and accessories you plan to use to connect to the patient. Do not modify this equipment. This product may become contaminated by bodily fluids and expired gases. Dispose according to your local regulations. Storage and Handling: Avoid direct sunlight. Disposal: This product may become contaminated by bodily fluids and expired gases. Dispose according to your local regulations. Instructions for Use: Install according to the illustration. For cleaning and disinfection instructions, see the Cleaning and Disinfection chapter. Oxygen Adding Oxygen to the Device Delivered oxygen concentration varies with changes in flow in the circuit. The following may have an impact on oxygen concentration:
Pressure settings Patient Tidal Volume Peak Inspiratory Flow I:E Ratio Respiratory rate Circuit leak rate Oxygen flow rate Low Flow Oxygen Warnings When administering fixed-flow supplemental oxygen, the oxygen concentration may not be constant. The inspired oxygen concentration will vary, depending on the pressures, patient flow, and circuit leak. Substantial leaks may reduce the inspired oxygen concentration to less than the expected value. Use appropriate patient monitoring, such as pulse oximeter with alarm, as medically indicated. This device DOES NOT alarm for loss of the low flow oxygen supply. Do not connect the device to an unregulated or high-pressure oxygen source. The device may result in incorrect flow and tidal volume measurements and improper operation of related alarms if you add low flow oxygen directly into the patient circuit or mask instead of directly adding it into the oxygen inlet on the back of the ventilator. Turn off oxygen when the device is not in use. When the device is not in operation and the oxygen flow remains on, oxygen delivered into the tubing may accumulate within the devices enclosure. Do not operate the ventilator in the presence of flammable gasses. This could cause a fire or explosion. To add oxygen to the circuit, the oxygen supply must comply with the local regulations for medical oxygen. The oxygen flow into the oxygen valve cannot exceed 30 l/min and the pressure cannot exceed 10 psi. For instructions, see the Device Setup chapter. Oxygen 82 Instructions for Use | Trilogy Evo Universal Accessories Portability and Travel Accessories When using mounting hardware, confirm that the ventilator is secured to the wheelchair mount or roll stand mount. See the instructions that accompany the accessory. Remote Alarms If you are using a remote alarm or nurse call, ensure you test the accessory before starting ventilation. Ensure that you can hear the ventilators audible alarms on the remote alarm. Ensure the remote alarm signals when the remote alarm cable is disconnected. To use the alarm, see the alarms instructions. Use the RJ9 connection on the Utility Panel to connect to this ventilator. Remote Alarms 83 Instructions for Use | Power Management 11. Power Management Power Sources Trilogy Evo Universal can operate on AC (wall outlet) or DC (battery) power from several sources listed in descending priority below. 1. AC Power 2. External 24- or 12-volt battery (DC) such as a vehicle battery requires an external battery cable 3. Detachable battery (DC) 4. Internal battery (DC) Battery Power Indicator The battery power indicator appears in the status bar at the bottom of the window. The indicator shows the status of batteries in use. Battery operating time depends on the characteristics of the battery and usage of the device. The capacity of the battery shown on the Battery Power Indicator is only an estimate. You can also view battery status in the following locations:
The battery status Monitoring Window see Monitoring Window in the About Trilogy Evo Universal chapter. The device information window: In the menu bar, tap the Options icon. In the Options window, tap Information. Battery Status Indicator To view the battery status indicator, tap the battery power indicator in the Status Bar:
This indicator shows:
-
-
Battery status Estimated battery time remaining (for internal and detachable batteries) Estimated remaining battery power shown as a percent of total capacity (for external battery) Battery status
-
AC Power AC power has the highest priority. When AC power is present, Trilogy Evo Universal uses that power to run the device and to charge the detachable and internal batteries. To use AC power:
1. Plug the socket end of the AC power cord into Trilogy Evo Universals power inlet. 2. Plug the pronged end of the power cord into an electrical outlet that is not controlled by a wall switch. 3. 4. Verify Trilogy Evo Universal is using AC power. You should see the green LED light next to the power button. If you do not see this light, Secure the cord to the device using the retention clip. contact Philips customer service. Periodically inspect the power cord for damage or signs of wear. Discontinue use and replace if damaged. AC Power 84 Instructions for Use | Power Management To remove AC power, disconnect the power supply cord from the electrical outlet. External Battery You can use an external DC power source, such as a vehicle battery, to power Trilogy Evo. You must use the Philips Respironics External Battery Cable to connect the battery to the device. Do not use any other cable or improper operation of the device may occur. This cable is pre-wired and properly terminated to ensure safe connection of an external battery to the ventilator. See the instructions included with the battery cable. The external battery will power the device when Trilogy Evo is disconnected from AC power. Battery operating time depends on the characteristics of the battery and usage of the device. The capacity of the external battery shown on the Battery Power Indicator is only an estimate of the actual remaining capacity. The external battery is installed correctly when the external battery icon appears in the Battery Power Indicator on the Status Bar. Detachable Battery Trilogy Evo Universal detachable battery pack will power the ventilator, if attached, when Trilogy Evo Universal is disconnected from AC or external battery power. The detachable battery automatically charges when plugged into AC power. Battery operating time depends on the characteristics of the battery and usage of the device. The capacity of the external battery shown on the Battery Power Indicator is only an estimate of the actual remaining capacity. Refer to the Technical Specifications included in this manual for more information on the detachable battery. The battery includes an LED charge indicator. To view the percentage of charge, press the battery button. Green lights appear to indicate how much charge remains in the battery. To remove the detachable battery: open the detachable battery access door. Lift the battery handle and pull the battery to remove it from the battery bay. To replace the detachable battery: open the detachable battery access door. Slide the battery into the battery bay until you hear a click. The detachable battery is installed correctly when the detachable battery icon appears in the Battery Power Indicator at the bottom of the screen. Detachable Battery 85 Internal Battery Internal battery power is the lowest priority power source. Instructions for Use | Power Management The internal battery is functioning correctly when the internal battery icon appears in the Battery Power Indicator at the bottom of the screen in a fully charged state. For help, see the following section titled, Battery Power Indicator. The detachable battery automatically charges when plugged into AC power. The expected life of the internal battery is two years. Replace the internal battery every two years or when there is a noticeable reduction in usage time when fully charged. When storing the device, ensure that internal battery is recharged once every six months. See the Technical Specifications included in this manual for internal battery specifications. Power Loss Several alarms are related to power management and power loss. For information about these alarms and how to test them, see the Alarms chapter. If all power sources are lost while Trilogy Evo Universal is delivering therapy, as soon as a power source is connected, the device will begin to deliver therapy again. All alarm settings are retained and restored. Power Indicator Icons Internal Battery Status Charging Charging Charging Charging Charging Charging Charging Charging Capacity remaining 0% capacity Nearly depleted Low power remaining 01-20% capacity 21-40% capacity 41-60% capacity 61-80% capacity 81-100% capacity Discharging 81-100% capacity Discharging 61-80% capacity Discharging 41-60% capacity Discharging 21-40% capacity Discharging 01-20% capacity Discharging Low power remaining Discharging Depleted, not charging Nearly depleted 0% capacity or failed Detachable Battery Status Capacity remaining Charging 0% capacity Charging Nearly depleted Charging Low power remaining Charging Charging Charging Charging Charging 01-20% capacity 21-40% capacity 41-60% capacity 61-80% capacity 81-100% capacity Discharging 81-100% capacity Discharging 61-80% capacity Discharging 41-60% capacity Discharging 21-40% capacity Discharging 01-20% capacity Discharging Low power remaining Power Indicator Icons 86 Discharging Nearly depleted Depleted, not charging 0% capacity or failed Status Capacity remaining Discharging 81-100% capacity Discharging 61-80% capacity Discharging 41-60% capacity External Battery Instructions for Use | Power Management Discharging 21-40% capacity Discharging 01-20% capacity Discharging Low power remaining Discharging Nearly depleted Discharging 0% capacity Power Indicator Icons 87 Instructions for Use | Technical Specifications 12. Technical Specifications Specifications Ventilation types and modes A/C-PC: Assist control (pressure control) A/C-VC: Assist control (volume control) CPAP: Continuous positive airway pressure PSV: Pressure support ventilation S/T: Spontaneous/timed ventilation SIMV-PC: Synchronized intermittent mandatory ventilation (pressure control) - mandatory breaths are synchronized to a patients inspiratory efforts SIMV-VC: Synchronized intermittent mandatory ventilation (volume control) - mandatory breaths are synchronized to a patients inspiratory efforts Controls Delivered tidal volume Breath rate PEEP EPAP/CPAP IPAP Pressure support/pressure control Inspiratory time Rise time Triggering and cycling Flow trigger sensitivity Flow cycle Flow pattern FiO2 Measured and Displayed Patient Parameters Tidal volume (Vti or Vte) Minute ventilation (MinVent) Leak Respiratory rate (RR) Peak inspiratory flow Peak inspiratory pressure (PIP) Mean airway pressure Percentage spontaneous triggered breaths I:E ratio Dynamic compliance Dynamic resistance Dynamic plateau pressure Auto-PEEP FiO2 SpO2 with pulse oximeter accessory Pulse rate with pulse oximeter accessory etCO2 with end tidal CO2 accessory 35 1,200 ml 0 80 BPM with a resolution of 1 BPM 0 35 cm H20 for active exhaust circuits 3 25 cm H20 for passive circuits 3 25 cm H20 3 60 cm H20 0 60 cm H20, Patient pressure limited to 60cmH2O 0.3 5.0s, constrained to prohibit an inverse I:E ratio 0, 1, 2, 3, 4, 5, 6 AutoTrak, Sensitive AutoTrak, and Flow Trigger 0.5 9 l/min 10% 90% of peak flow Square, Ramp 21% 100% when available 0 to 2000 ml with a resolution of 1 ml 0 to 30 l/min with a resolution of 0.1 l/min 0 to 200 l/min with a resolution of 0.1 l/min 0 to 90 BPM with a resolution of 1 BPM 0 to 200 l/min with a resolution of 0.1 l/min 0 to 90 cmH2O with a resolution of 0.1 cmH2O 0 to 90 cmH2O with a resolution of 0.1 cmH2O 0 to 100% with a resolution of 1%
9.9:1 to 1:9.9 0.5 to 200 ml/cmH2O with a resolution of 0.1 ml/cmH2O 2 to 200 cmH2O/l/sec with a resolution of 0.1 cmH2O/l/sec 0 to 90 cmH2O with a resolution of 0.1 cmH2O 0 to 90 cmH2O with a resolution of 0.1 cmH2O 21% to 100% with a resolution of 1%
0 to 100% with a resolution of 1%
25 to 240 beats per minute with a resolution of 1 beat per minute 0 to 150 mmHg with a resolution of 1 mmHg Specifications 88 Environmental Operating Transient operating temperature, excluding high pressure oxygen blending Storage temperature Physical Weight Size Screen dimensions Ingress protection IEC 60601-1 classification Composition Expected service life Internal memory capacity Mode of operation Maximum limited pressure Electrical AC input voltage DC input voltage Internal and detachable Li-ion batteries Degree of protection against electric shock Audio Alarm sound pressure level Sound pressure level - ventilator, circuit and exhalation device Sound power level -ventilator, circuit and exhalation device Oxygen Low flow High pressure Ventilator response to a 21 -90%
increase in oxygen concentration Control Accuracy Pressure Tidal volume FiO2 Instructions for Use | Technical Specifications Temperature: 0C to 40C Relative humidity: 5% to 90% RH, non-condensing Atmospheric pressure: 62 to 106 kPa Altitude: -1261 to 12,971 feet Battery charging temperature: 5C to 40C
-20C to 50C
-25C to 70C 5% to 93% RH, non-condensing 6.3 Kg (13.9 lbs) 19.3 cm D x 28.6 cm W x 24.5 cm H 7.6 D x 11.25 W x 9.65 H 8, 20.32 cm IP22: protection against finger-sized objects and protected against dripping water when tilted up to 15 degrees. Type of protection against electric shock: Class II equipment Degree of protection against electric shock type BF applied part This device does not contain natural latex rubber or dry natural rubber. 10 years 2Gb Continuous 90 cmH2O 100V 240V - 50/60 Hz 1.7A 12/24V 6.5A 7.5 hours nominal total run time per method in IEC 80601-2-72 (each battery) Charge time for detachable and internal battery from 0% to 80%: 2.5 hours from 0% to 100%: 3.5 hours Type BF Applied Part Lowest setting: 53.9 dBA Highest setting: 85.5 dBA 43.7 dBA as measured according to ISO 80601-2-12 51.6 dBA as measured according to ISO 80601-2-12 0 to 30 l/min; maximum 10 psi 280 to 600 kPa (41 to 87 psi)
< 30 seconds
(2 cmH2O + 4% of the setting)
(4 ml + 15% of setting) 5 % FiO2 Specifications 89 Instructions for Use | Technical Specifications
(2 cmH2O + 4% of actual)
(4 ml + 15% of actual) for volumes 35 ml 10 ml for volumes < 35 ml
(2.5% FiO2 + 2.5% of actual reading) within a 24-hour 2-point calibration period, or a change in altitude
(volume fraction of 0.43% + 8% of gas level) See the sensor manufacturers operating instruction sheet. Pediatric/adult circuit (20-22 mm): less than 5 cmH2O at 30 l/min Pediatric/adult circuit (19 mm): less than 5 cmH2O at 15 l/min Infant/pediatric circuit (14-16 mm): less than 5 cmH2O at 15 l/min Infant/pediatric circuit (9-13 mm): less than 5 cmH2O at 2.5 l/min Up to 4 ml/cmH2O Monitored Parameter Accuracy Airway pressure Tidal volume FiO2 EtCO2 (with Mainstream CO2 sensor) SpO2 and pulse rate Operational range of the Patient Circuit and Circuit accessories Inspiratory/expiratory resistance Compliance Wireless Bluetooth Operating frequency range Channel bandwidth Maximum output power Modulation Near-field communication (NFC) Operating frequency Receiving section bandwidth Maximum output power Modulation Wi-Fi 2402-2480 MHz 1 MHz/2 MHz 12.0 dBm GFSK, Pi/4 DQPSK, 8DQPSK 13.56 MHz 1.4 MHz 23.0 dBm ASK, OOK Operating frequency range Channel bandwidth Maximum output power Modulation Security All flows and volumes are expressed in BTPS. 2402-2480 MHz 20 MHz/40MHz 17 dBm DSSS, OFDM, DBPSK, DQPSK, CCK, 16-QAM Wpa2 Measurement Uncertainty for Control and Performance Specifications The stated tolerances account for the measurement uncertainty of the test equipment used to verify performance:
Pressure: 0.75% cmH2O Tidal Volume: 2 ml Oxygen: 1% FiO2 Specifications 90 Instructions for Use | Technical Specifications Standards Compliance This device conforms to the following standards:
Standards General Collateral Particular Wireless communication IEC 60601-1-1 Medical electrical equipment. Part 1-1: General requirements for safety. Collateral standard: Safety requirements for medical electrical systems IEC 60601-1-11 Home Health Care Environment according to transit-operable usage Device essential performance is specified in each of the following standards:
ISO 80601-2-12: Medical electrical equipment. Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators ISO 80601-2-61 Medical electrical equipment. Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment ISO 80601-2-55 Medical electrical equipment. Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors Bluetooth Core Specification version 4.1 ISO/IEC 18092:2013: Information technology. Telecommunications and information exchange between systems. Near Field Communication. Interface and Protocol (NFCIP-1) ISO IEC 21481 Ed 2.0: Information technology. Telecommunications and information exchange between systems. Near Field Communication Interface and Protocol -2 (NFCIP-2) ISO/IEC 14443 Ed 2.0: Identification cards. Contactless integrated circuit cards. Proximity cards. WLAN Standard: IEEE 802.11 (2012) b/g/n: Information technology. Telecommunications and information exchange between systems. Local and metropolitan area networks. Specific requirements. Part 11: Wireless LAN Medium Access Control (MAC) and Physical Layer (PHY) Specifications Hereby, Respironics Inc. declares that this class 1 radio equipment is in compliance with Directive 2014/53/EU. The full text of the EU declaration of conformity is available at the following internet address: http://incenter.medical.philips.com/PMSPublic Standards Compliance 91 Instructions for Use | Technical Specifications Pneumatic Diagram Low Flow O2 Filter Air Inlet Inlet Flow Path OBM High Flow O2 O2 Flow Sensor O2 Press FiO2 Sensor Blower Machine Flow Sensor /
Element Outlet Flow Path AECM Dual Limb /
PAP System PCA Monitor Pressure Outlet Pressure EMC Information This ventilator generates audible and visual alarms to alert you if it is not able to provide ventilation or if external monitoring is lost during an EMC disturbance. Warnings Avoid using this equipment adjacent to or stacked with other equipment because it could result in improper operation. Although the other equipment may comply with EMC standard requirements, interference can occur. If such use is necessary, observe this equipment and the other equipment to verify that both are operating normally. This device is unsuitable for use in a magnetic resonance imaging (MRI) environment. The use of accessories, transducers and/or cables other than those specified, with the exception of those sold by the manufacturer as replacement parts for internal components, may result in increased emissions or decreased immunity of the equipment or system. EMC Information 92 Instructions for Use | Technical Specifications Guidance and Manufacturers Declaration - Electromagnetic Emissions This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. Emissions Test RF emissions CISPR 11 Industrial, scientific and medical equipment. Radio-frequency disturbance characteristics. Limits and methods of measurement RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Electromagnetic compatibility
(EMC). Part 3-2: Limits. Limits for harmonic current emissions (equipment input current smaller than or equal to 16 A per phase) Voltage fluctuations/Flicker emissions IEC 61000-3-3 Electromagnetic compatibility
(EMC) - Part 3-3: Limits - Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current 16 A per phase and not subject to conditional connection Emission of Radio Frequency Energy RTCA/DO-160G Section 21 Electromagnetic Environment - Guidance The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network. Compliance Group 1 Class B Class A Complies Category M This device is suitable for use onboard commercial airplanes inside passenger cabin. EMC Information 93 Instructions for Use | Technical Specifications Guidance and Manufacturers Declaration - Electromagnetic Immunity This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. IEC 60601 Test Level 8 kV contact 2 kV, 4 kV, 8 kV, and 15 kV air Compliance Level 8 kV contact 2 kV, 4 kV, 8 kV, and 15 kV air Electromagnetic Environment - Guidance Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. 2 kV for power supply lines 1 kV for input-output lines 2 kV for supply mains 1 kV for input/output lines Mains power quality should be that of a typical home or hospital environment. 1 kV line to ground 2 kV line to ground 1 kV line to line N/A - this Class II device does not connect to earth ground Mains power quality should be that of a typical home or hospital environment. O% UT 0.5 cycle at 45 degree increments 0% U T 1 cycle 70% U T 25 cycles
(30 cycles if US) 0% U T 5 sec O% U T 0.5 cycle at 45 degree increments 0% U T 1 cycle 70% U T 25 cycles
(30 cycles if US) 0% U T 5 sec Mains power quality should be that of a typical home or hospital environment. If the user of the device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery. Immunity Test IEC 61000-4-2 Electromagnetic compatibility
(EMC). Part 4-2:
Testing and measurement techniques. Electrostatic discharge immunity test IEC 61000-4-4 Electromagnetic compatibility
(EMC). Part 4-4:
Testing and measurement techniques. Electrical fast transient/burst immunity test IEC 61000-4-5 Electromagnetic compatibility
(EMC) - Part 4-5 Testing and measurement techniques. Surge immunity test IEC 61000-4-11 Electromagnetic compatibility
(EMC). Part 4-11:
Testing and measurement techniques. Voltage dips, short interruptions and voltage variations immunity tests EMC Information 94 Instructions for Use | Technical Specifications Electromagnetic Environment - Guidance Power frequency magnetic fields should be at levels characteristic of a typical location in a typical hospital or home environment. Compliance Level 30 A/m IEC 60601 Test Level 30 A/m Immunity Test IEC 61000-4-8 Electromagnetic compatibility
(EMC). Part 4-8:
Testing and measurement techniques. Power frequency (50/60 Hz) magnetic field immunity test NOTE: UT is the AC mains voltage prior to application of the test level. Guidance and Manufacturers Declaration - Electromagnetic Immunity This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. Immunity Test Conducted RF IEC 61000-
4-6 Electromagnetic compatibility (EMC). Part 4-6: Testing and measurement techniques. Immunity to conducted disturbances, induced by radio-frequency fields. Radiated RF IEC 61000-4-3 Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques
- Radiated, radio-
frequency, electromagnetic field immunity test IEC 60601 Test Level Compliance Level 3 Vrms 150 kHz to 80 MHz 6 Vrms Amateur Radio &
ISM Bands between 150 kHz and 80 MHz 3 Vrms 150 kHz to 80 MHz 6 Vrms Amateur Radio & ISM Bands between 150 kHz and 80 MHz Electromagnetic Environment - Guidance Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended 30 cm separation distance. 10 V/m 10 V/m 80 MHz to 2.7 GHz Up to 28V/m in telecommunication bands as specified in clause 8.10 of IEC 60601-1-2 Up to 28V/m in telecommunication bands as specified in clause 8.10 of IEC 60601-1-2 EMC Information 95 Instructions for Use | Wireless Connectivity 13. Wireless Connectivity This device has Bluetooth SmartReady wireless technology, which includes Bluetooth Classic and Low Energy Bluetooth. Bluetooth allows Trilogy Evo to communicate with a compatible Bluetooth device approved by Philips. Bluetooth functionality may not be present in all models. Warning: Other equipment may interfere with this device, even if the other equipment complies with CISPR8 emission requirements. The Health Industry Manufacturers Association recommends that a minimum separation of six inches be maintained between a wireless phone and a pacemaker to avoid potential interference with the pacemaker. The Trilogy Evo on-board Bluetooth communication should be considered a wireless phone in this regard. Caution: When traveling by air, inform the airline of the presence of wireless technology in this device. Ensure such devices are permitted. Notices: The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Philips Respironics is under license. Other trademarks and trade names are those of their respective owners. The Trilogy Evo device transmits data between the therapy device and a mobile device, but it does not store any of your personal data. This connection between the therapy device and a mobile device is encrypted. This device contains a certified Bluetooth/Wi-Fi radio FCC ID: 2AN9Z-1127941BT, IC ID: 3234B-1127941BT. Nearfield communication (NFC) is certified under FCC ID: 2AN9Z-1127941, IC ID: 3234B-1127941. Use of non-original manufacturer-approved accessories may violate your local RF exposure guidelines and should be avoided. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio, TV reception, or other devices which can be determined by turning the equipment on and off, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna (on the radio, TV, or other device). Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer of the device for help. Bluetooth Actions To access the Bluetooth window:
1. 2. 3. In the menu bar, tap the Options icon. In the Options window, tap Device Settings. In the Device Settings window, tap Bluetooth. To enable Bluetooth connections, In the Bluetooth window, tap On. To view connection status, in the Bluetooth window view the connection status indicator, including the MAC address of the current or most recently connected device. Bluetooth Actions 96 Instructions for Use | Wireless Connectivity NFC Actions To access the NFC window:
1. 2. 3. In the menu bar, tap the Options icon. In the Options window, tap Device Settings. In the Device Settings window, tap Bluetooth. To enable NFC connections, in the Near-Field Communication window, tap On. To disable NFC connections, in the Near-Field Communication window, tap Off. Troubleshooting If the devices display is erratic:
Relocate the device to an area away from electronic equipment (such as cellular phones, cordless phones, computers, TVs, electronic games, or hair dryers) Troubleshooting 97 Instructions for Use | Glossary 14. Glossary The following terms and acronyms appear throughout this manual. Definition Circuit that includes an active exhalation device Pressure measured at the patient connection port. Temporary cessation of breathing. The total number of hours that the blower has been on over the life of the device. This value helps determine when the ventilator needs to be serviced. You cannot reset this value. It can only be reset by a service center. Breaths per minute or beats per minute Body Temperature and Pressure Saturated; A standardization for lung volumes and flows to barometric pressure at sea level, body temperature, and saturated with water vapor reflecting the condition of air in the lung. Continuous Positive Airway Pressure Expiratory positive airway pressure Electrostatic Discharge End tidal carbon dioxide. The amount of carbon dioxide at the end of exhalation. Fractionally Inspired Oxygen (the percentage of oxygen in the air inhaled) The ratio of inspiratory time to expiratory time. Full term newborn up to one month in age with mass that is greater than or equal to 2.5 kg. Inspiratory Positive Airway Pressure Liters Per Minute Mandatory Breath is completely controlled by the ventilator. The volume of gas that moves in and out of the lungs in one minute. It is calculated by multiplying the tidal volume by the respiratory rate. Consists of the tubing, filtration, exhaust valves (passive or active), and flow sensor external to the ventilator. Maximum flow rate (in liters per minute) reached during a breath. Highest pressure reached during inspiration. Term Active Circuit Airway Pressure Apnea Blower Hours BPM BTPS CPAP EPAP ESD etCO2 FiO2 I:E Ratio Infant IPAP l/min Mandatory Breath Minute Ventilation
(MinVent) Patient Circuit Peak Flow Peak Inspiratory Pressure (PIP) PEEP Prescription Pressure Control PS Ramp Positive End Expiratory Pressure is the pressure control setting in expiration. A given set of therapy mode control settings, alarm settings, and patient circuit type. Ventilation in which breaths are controlled by operator-defined pressure, inspiratory time, and rise time. Pressure Support The ramp feature reduces pressure and then gradually increases the pressure to the prescription setting. The time it takes the ventilator to change from expiration to inspiration. Respiratory Rate (the number of breaths per minute). Delivers a periodic, larger volume breath. Settings adjust the frequency and volume. Synchronous intermittent mandatory ventilation Saturation of peripheral oxygen Rise Time RR Sigh SIMV SpO2 Spontaneous Breath Breath type in which the breath is patient-triggered. Spontaneous/Timed
(S/T) Mode Square Therapy mode that is similar to S mode, except that it can also deliver a mandatory breath if the patient does not spontaneously breathe within a set time. With a square wave pattern, airflow is generally constant throughout inspiration of the breath. The amount of air passing in and out of the lungs for each breath. Troubleshooting 98 Tidal Volume Instructions for Use | Glossary Term VBS Volume Control Vte Vti Definition Ventilator breathing system; inspiratory or expiratory pathways through which gas flows at respiratory pressures and bounded by the port through which fresh gas enters, the patient connection port and the exhaust port. Ventilation in which breaths are controlled by an operator-defined volume, flow pattern, breath rate, and inspiratory time. Exhaled Tidal Volume Inhaled Tidal Volume Troubleshooting 99 Instructions for Use | Warranty 15. Warranty WARRANTY Respironics, Inc. (Philips Respironics) warrants that defects in the Product due to faulty materials and workmanship will be repaired or replaced at Respironics expense if convincing proof of purchase within the Warranty Period (as defined herein) is provided. Philips Respironics warrants that the Product will be free from defects in material and workmanship under normal and proper use in accordance with applicable instructions for a period of ninety (90) days (Warranty Period) from the date of shipment by Philips Respironics to the original purchaser. Accessories and replacement parts are not covered under this warranty. However, Philips Respironics warrants that the internal battery in the Product will be free from defects in material and workmanship, under normal and proper use and when correctly maintained, for a period of 90 days from the date of shipment by Philips Respironics to the original purchaser. Contact our Consumer Care Center at 1-800-682-7664 (North America), outside North America contact your local Respironics, Inc. Consumer Care Center. Internet information: www.philips.com\healthcare (North America) or www.philips.com (outside North America). LIMITATIONS If any Product purchased from Philips Respironics fails to conform to the warranties set forth herein during the Warranty Period as determined by Philips Respironics, in its sole discretion, Philips Respironics may discharge its warranty obligation by choosing to repair or replace the Product. This may be accomplished by installing new or remanufactured assemblies or components, or by other repairs deemed appropriate in the sole discretion of Philips Respironics. The choice of repair or replacement by Philips Respironics will be the sole and exclusive remedy of the original purchaser. Philips Respironics reserves the right, in its sole discretion, to refund the purchase price in lieu of repair or replacement of the Product. In no event will Philips Respironics maximum liability under these warranties exceed the price paid to Philips Respironics by the original purchaser for the Product. CONDITIONS This warranty does not cover damage or injury whether to the Product or to personal property or persons caused by accident, misuse, abuse, negligence, failure to install in accordance with Philips Respironics installation instructions, failure to operate under conditions of normal use and in accordance with the terms of the operating manual and instructions, failure to maintain in accordance with the applicable service manuals, or alteration or any defects not related to materials or workmanship of the Product. This warranty does not cover damage which may occur in shipment and does not apply to the device if dropped, misused, altered or otherwise damaged after shipment. This warranty does not apply to any Product or individual part of a Product that may have been repaired or altered by anyone other than Philips Respironics or an authorized Philips Respironics service center. This warranty does not apply to any Product which is not purchased new. DISCLAIMERS EXCEPT AS SET FORTH IN THIS LIMITED WARRANTY, PHILIPS RESPIRONICS MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, STATUTORY OR OTHERWISE, REGARDING THE PRODUCT, ITS QUALITY OR PERFORMANCE. PHILIPS RESPIRONICS SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTY OF MERCHANTABILITY AND THE IMPLIED WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT WILL PHILIPS RESPIRONICS BE LIABLE FOR THE COST OF PROCUREMENT OF SUBSTITUTE PRODUCTS OR FOR ANY INDIRECT, SPECIAL, CONSEQUENTIAL, PUNITIVE, EXEMPLARY OR INCIDENTAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, COMMERCIAL LOSS OR LOST REVENUES), FOR ANY CAUSE OF ACTION, WHETHER IN CONTRACT OR TORT, AND WHETHER OR NOT PHILIPS RESPIRONICS WAS AWARE OR SHOULD HAVE BEEN AWARE OF THE POSSIBILITY OF THESE DAMAGES. EXCEPT FOR THE OBLIGATIONS UNDER THIS LIMITED WARRANTY, PHILIPS RESPIRONICS WILL NOT HAVE ANY OBLIGATION OR LIABILITY FOR ANY OTHER LOSS, DAMAGE OR INJURY RESULTING DIRECTLY OR INDIRECTLY FROM THE PRODUCT. PURCHASERS SOLE AND EXCLUSIVE REMEDY FOR BREACH OF THE WARRANTIES SET FORTH IN THIS DOCUMENT WILL BE AS PROVIDED IN THE PARAGRAPH DESCRIBING LIMITATIONS. Purchaser is cautioned that no person or entity is authorized to make any warranties on behalf of Philips Respironics and any such alleged warranties are hereby disclaimed by Philips Respironics. Troubleshooting 100
1 | Label and location info | ID Label/Location Info | 290.39 KiB |
External Trilogy Evo Label and Location FCC-ID: 2AN9Z-1127941 Figure 1: Label for DS2000X11B Figure 2: Label for DS2100X11B Respironics Inc. (Philips-Respironics) 1001 Murry Ridge Lane Murrysville, PA 15668 Figure 3: Label for DS2110X11B Figure 4: Label for DS2200X11B Respironics Inc. (Philips-Respironics) 1001 Murry Ridge Lane Murrysville, PA 15668 Figure 5: Label Location Respironics Inc. (Philips-Respironics) 1001 Murry Ridge Lane Murrysville, PA 15668
1 | Confidentiality letter | Cover Letter(s) | 150.43 KiB |
03/26/18 Attention: Application Examiner RE: Request for Confidentiality Applicant: Respironics, Inc. FCC ID: 2AN9Z-1127941 To Whom It May Concern:
Request is hereby submitted by Respironics Inc. to withhold permanently from public review certain portions of the application for equipment certification for the referenced FCC identifiers. This request for confidentiality is made pursuant to 47 CFR 0.457(d) and 0.459 of the FCC Rules. In particular, the following sections of the application are to be kept permanently confidential:
Schematics Block diagrams Operational/Functional Description Request is hereby submitted by Respironics Inc. to withhold from public review for a period of 180 days from the date of the Grant of Equipment Authorization and prior to marketing, certain portions of the application for equipment certification for the referenced FCC identifiers. This request for confidentiality is made pursuant to 47 CFR 0.457(d) and 0.459 of the FCC Rules. In particular, the following sections of the application are to be kept confidential for a period of 180 days from the date of Authorization:
Test setup photos External photos Internal photos User manual Rationale for request for confidentiality:
Respironics has invested considerable time and materials in research and development to produce the referenced product. Disclosure of the permanently confidential portions of this application to competitors would not only give them significant competitive advantages in developing similar products, but would also disclose successful implementation of unpublished, leading edge technology developed by us. Disclosure of the short-term confidential portions of this application during the period of importation and/or distribution would reveal key aspects of Respironics Inc. (Philips-Respironics) 1001 Murry Ridge Lane Murrysville, PA 15668 proprietary technology to competitors and diminish the value of our investment in research and development. The fee for confidentiality has been submitted along with the fee for certification. If you have questions or need further information, please contact the undersigned. Best Regards, On behalf of Respironics, Inc. Dave Brooking Senior Staff Engineer Respironics Inc. (Philips-Respironics) 1001 Murry Ridge Lane Murrysville, PA 15668 Dave Brooking
1 | Granteee Contact Signature Authorization | Cover Letter(s) | 33.75 KiB |
I, Joe Olsavsky, as the Grantee Contact, of Respironics Inc., hereby authorize Dave Brooking of Respironics Inc., to sign documents related to the pertaining to the Federal Communications Commission filing. PHILIPS January 25, 2018 Federal Communications Commission 445 1 2th Street SW Washington, DC 20554 RE: FCC Grantee Contact Signature Authority TO WHOM IT MAY CONCERN:
Best regards, Printed Name: Joe Olsavsky Title: Director of Regulatory Affairs Respironics, inc. (Philips-Respironics) 1001 Murry Ridge Lane Murrysville, PA 15668
1 | Letter of agency | Cover Letter(s) | 193.15 KiB |
03/26/18 Attention: Application Examiner Federal Communications Commission 445 12th Street SW Washington, DC 20554 Subject:
Limited Agency Agreement Respironics, Inc. FCC ID: 2AN9Z-1127941 To Whom It May Concern:
We, Respironics Inc. hereby authorized Intertek Testing Services to act as our Agent for the purpose of preparing application for FCC ID number 2AN9Z-1127941 under all applicable parts of the FCC rules and regulations. The effective date of this limited agency agreement is (03/26/2018). The Limited Agency Agreement expires on (03/26/2019), unless sooner terminated or extended by written notice to Intertek Testing Services and the Federal Communications Commission. This is to advise that we are in full compliance with the Anti-Drug Abuse Act. The applicant is not subject to a denial of federal benefits pursuant to Section 5301 of the Anti-Drug Act of 1988, 21 U.S.C. 862, and no party to the application is subject to a denial of federal benefits pursuant to that section. If you have any questions or comments, please do not hesitate to contact me. Sincerely, Dave Brooking Senior Staff Engineer Respironics Inc. (Philips-Respironics) 1001 Murry Ridge Lane Murrysville, PA 15668 Dave Brooking
frequency | equipment class | purpose | ||
---|---|---|---|---|
1 | 2018-05-15 | 13.56 ~ 13.56 | DXX - Part 15 Low Power Communication Device Transmitter | Original Equipment |
app s | Applicant Information | |||||
---|---|---|---|---|---|---|
1 | Effective |
2018-05-15
|
||||
1 | Applicant's complete, legal business name |
Respironics, Inc.
|
||||
1 | FCC Registration Number (FRN) |
0026975250
|
||||
1 | Physical Address |
1001 Murry Ridge Lane
|
||||
1 |
Murrsyville
|
|||||
1 |
United States
|
|||||
app s | TCB Information | |||||
1 | TCB Application Email Address |
T******@intertek.com
|
||||
1 | TCB Scope |
A1: Low Power Transmitters below 1 GHz (except Spread Spectrum), Unintentional Radiators, EAS (Part 11) & Consumer ISM devices
|
||||
app s | FCC ID | |||||
1 | Grantee Code |
2AN9Z
|
||||
1 | Equipment Product Code |
1127941
|
||||
app s | Person at the applicant's address to receive grant or for contact | |||||
1 | Name |
J******** E**** O****
|
||||
1 | Title |
Director - Regulatory Affairs
|
||||
1 | Telephone Number |
724-3********
|
||||
1 | Fax Number |
724-3********
|
||||
1 |
j******@philips.com
|
|||||
app s | Technical Contact | |||||
1 | Firm Name |
Philips-Respironics
|
||||
1 | Name |
D****** B********
|
||||
1 | Physical Address |
1740 Golden Mile Hwy
|
||||
1 |
Monroeville, 15146
|
|||||
1 |
United States
|
|||||
1 | Telephone Number |
724-3********
|
||||
1 |
d******@philips.com
|
|||||
app s | Non Technical Contact | |||||
n/a | ||||||
app s | Confidentiality (long or short term) | |||||
1 | Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | Yes | ||||
1 | Long-Term Confidentiality Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | Yes | ||||
1 | If so, specify the short-term confidentiality release date (MM/DD/YYYY format) | 11/11/2018 | ||||
if no date is supplied, the release date will be set to 45 calendar days past the date of grant. | ||||||
app s | Cognitive Radio & Software Defined Radio, Class, etc | |||||
1 | Is this application for software defined/cognitive radio authorization? | No | ||||
1 | Equipment Class | DXX - Part 15 Low Power Communication Device Transmitter | ||||
1 | Description of product as it is marketed: (NOTE: This text will appear below the equipment class on the grant) | Ventilator | ||||
1 | Related OET KnowledgeDataBase Inquiry: Is there a KDB inquiry associated with this application? | No | ||||
1 | Modular Equipment Type | Does not apply | ||||
1 | Purpose / Application is for | Original Equipment | ||||
1 | Composite Equipment: Is the equipment in this application a composite device subject to an additional equipment authorization? | No | ||||
1 | Related Equipment: Is the equipment in this application part of a system that operates with, or is marketed with, another device that requires an equipment authorization? | No | ||||
1 | Is there an equipment authorization waiver associated with this application? | No | ||||
1 | If there is an equipment authorization waiver associated with this application, has the associated waiver been approved and all information uploaded? | No | ||||
app s | Test Firm Name and Contact Information | |||||
1 | Firm Name |
Intertek Testing Services NA, Inc.
|
||||
1 | Name |
G******** L********
|
||||
1 | Telephone Number |
949-4********
|
||||
1 | Fax Number |
949 4********
|
||||
1 |
g******@intertek.com
|
|||||
Equipment Specifications | |||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Line | Rule Parts | Grant Notes | Lower Frequency | Upper Frequency | Power Output | Tolerance | Emission Designator | Microprocessor Number | |||||||||||||||||||||||||||||||||
1 | 1 | 15C | 13.56000000 | 13.56000000 |