USER MANUAL Wearable Health Monitoring System
(MODEL : S-PATCH3) Document No. : S-UM-01 (Rev. 0.2) Samsung Electronics Co., Ltd. 1-1, Samsungjeonja-ro, Hwaseong-si, Gyeonggi-do, South Korea Samsung Confidential
[1]
Revision History Revision Number Rev.0.0 Rev.0.1 Rev.0.2 Issue Date Contents of revision 2016.12.28 Established by Manufacturers User Manual Policy 2017.1.5 2017.2.9 5.1 Label Update 5.1 Label Update Rev.0.2 2017.2.9 4.2.1 Spec Update
:
Samsung Confidential
[2]
Contents 1. Product Introduction .................................................................................................................................... 5 1.1 Introducing S-PATCH3 ......................................................................................................................... 5 1.2 Indication for Use .................................................................................................................................... 5 2. Cautions ............................................................................................................................................................. 5 2.1 General ........................................................................................................................................................ 5 2.2 Safety ............................................................................................................................................................ 5 2.3 Contra-Indication .................................................................................................................................... 6 2.4 Usage and Storage Conditions ............................................................................................................. 7 2.4.1 Conditions for Usage ........................................................................................................................ 7 2.4.2 Conditions for Storage ...................................................................................................................... 7 2.4.3 Cleaning Condition ........................................................................................................................... 7 3. Components and Installation of S-PATCH3 .......................................................................................... 8 3.1 Main Device Components ..................................................................................................................... 8 3.2 Accessory Components .......................................................................................................................... 9 3.3 Display viewer ........................................................................................................................................ 10 3.4 How to Install S-PATCH3 .................................................................................................................. 18 3.4.1 Installing S-PATCH ....................................................................................................................... 18 3.5 Precautions .............................................................................................................................................. 18 4. Using and Operating S-PATCH3 ............................................................................................................ 20 4.1 How to Use S-PATCH3 ....................................................................................................................... 20 4.2 Specifications .......................................................................................................................................... 25 4.2.1 S-PATCH3 Specifications ............................................................................................................ 25 4.3 Error Message and Alarm ................................................................................................................. 26 4.3.1 Device Disconnection .................................................................................................................... 26 Samsung Confidential
[3]
4.3.2 Abnormal signal detect .................................................................................................................. 26 4.3.3 Low Battery mode .......................................................................................................................... 27 4.3.4 Low Remaining Storage ................................................................................................................ 28 5. Labels and Packaging ................................................................................................................................. 30 5.1 Labels ........................................................................................................................................................ 30 5.1.1 Label for Packaging ....................................................................................................................... 30 5.1.2 Label for module ............................................................................................................................. 30 5.2 Packaging................................................................................................................................................. 31 5.2.1 Package List ..................................................................................................................................... 31 5.2.2 Package Units .................................................................................................................................. 32 6. Information on EMC ................................................................................................................................... 33 6.1 Guidance and Manufacturers Declaration Electromagnetic Emissions ........................ 33 6.2 Guidance and Manufacturers Declaration Electromagnetic Immunity ........................ 33 6.3 Guidance and Manufacturers Declaration Electromagnetic Immunity ........................ 34 6.4 Recommended separation distances between portable and mobile RF communications equipment and the EUT .................................................................................................................................. 35 6.5 Immunity and Compliance Level .................................................................................................... 36 6.6 Guidance and Manufacturers Declaration Electromagnetic Immunity ........................ 37 7. FCC Compliance Statement ..................................................................................................................... 38 8. Product Life time and Warranty ............................................................................................................ 38 8.1 Product Life time .................................................................................................................................. 38 8.2 Product Warranty ................................................................................................................................ 38 9. Company Introduction ............................................................................................................................... 39 9.1 Company Name and Address ........................................................................................................... 39 Samsung Confidential
[4]
1. Product Introduction 1.1 Introducing S-PATCH3 This product is intended to measure, transfer and record electrocardiogram signal. The viewer on mobile device may display ECG wave form, Heart Rate, Respiration Rate, Heart Rate Variability and other information. The product consists of S-PATCH3 body signal sensing module, ECG electrodes and mobile app viewer to monitor the signals. Product Type Ambulatory electrocardiographic systems Name S-PATCH3 Wearable Health Monitoring System Serial Number SP3-1612-0041 ManufacturerSAMSUNG ELECTRONICS Co., Ltd. 1-1, Samsungjeonja-ro, Hwaseong-si, Gyeonggi-do, 18448 Korea 1.2 Indication for Use The S-PATCH3 Wearable Health Monitoring System is intended to record, transfer and store single channel electrocardiogram signals. The S-PATCH3 signal sensing module transmits data through wireless communication to mobile device. The product is not intended for pediatric use. The S-PATCH3 wearable health monitoring system monitors and displays:
Heart Rate ECG signal Heart Rate Variability Respiration Rate 2. Cautions 2.1 General 1) DO NOT store in extremely hot, cold, humid, or wet conditions. 2) Since this product is a medical device, instructions are required to properly dispose of it. Contact the deputy or manufacturer for such instructions. Improper disposing of this device can lead to legal consequences. 3) Contact the manufacturer if the product functions abnormally, problematically, or not at all. Any attempt to repair without the manufacturers guide is not recommended. 4) Reuse of the electrodes is prohibited in any circumstances due to the infection. 5) Use of electrodes sticker may cause a skin irritation or reaction. 6) DO NOT expose to strong electromagnetic fields. 7) Too much body hair may cause an unsuccessful recording. 8) DO NOT use to diagnose heart related conditions. 9) No warranty for any data or information that is collected erroneously by the device, or misuse or malfunction as a result of abuse, accidents, alteration, misuse, neglect, or failure to maintain the products as instructed. 2.2 Safety 1) Before treatment, the doctor must explain cautions to the patient. 2) Operate the product in the correct order as described in this manual. 3) Use this device under doctors prescription. 4) Beware of the polarity of the Coin battery when exchange and insert the battery. 5) DO NOT use during magnetic resonance imaging (MRI) or external defibrillation procedures. Samsung Confidential
[5]
6) DO NOT drop or bump with excessive force. 7) DO keep components out of reach of children. 8) DO NOT swallow the device or wind the cable around the neck. 2.3 Contra-Indication 1) Patients with artificial cardiac pacemaker, cardioverter defibrillator, or other implantable electric devices. 2) Pregnant or breast-feeding mothers 3) A current sign or medical history of skin cancer, rash, skin disorder, keloid, and/or any injury. Samsung Confidential
[6]
2.4 Usage and Storage Conditions 2.4.1 Conditions for Usage 1) Temperature: 0-45(32 to 113) 2) Relative humidity: 10%-95%(non-condensing) 3) Atmospheric pressure: 700hPa-1060hPa 2.4.2 Conditions for Storage 1) Temperature: 20-70(-4 to 158) 2) Relative humidity: 10%-95%( non-condensing) 3) Atmospheric pressure: 400hPa-1500hPa 4) Keep the device in the case when it doesn't use. 2.4.3 Cleaning Condition 1) Clean the device with soft, dry cloth 2) Equipment failure may occur due to dust and debris during long-term use of the equipment. Samsung Confidential
[7]
3. Components and Installation of S-PATCH3 3.1 Main Device Components
<S-PATCH3>
Label Name Description 1 Main Body1 Coin battery inset module 2 Main Body2 Main PCB module 3 4 5 6 Connect Cable Cable between two main bodies Power Button Power On/Off button LED LED Lamp to indicate the states Electrode connect hole Holes for ECG electrodes connecting Samsung Confidential
[8]
3.2 Accessory Components
<Electrode>
<Battery>
Label Name Description Electrode Multi-purpose monitoring electrodes with sticky gel feature (high performance adhesive and foam backing) Note: Use the FDA registered Electrode. Battery Power supply for S-PATCH3. DC 3V Coin Battery. 1 2 The Applied Parts are main device and electrode Samsung Confidential
[9]
3.3 Display viewer 1) Main Display of S-PATCH3 App
: : () , ()
, 11 pt Samsung Confidential
[10]
A. STATUS Bar A.1. Elapsed Time: The elapsed time from the first signal receiving time A.2. Device ID: Connected S-PATCH Devices ID A.3. BLE Connection Indicator: Bluetooth Low Energy Connection State A.4. Device Battery Indicator: Remained battery state of the connected device In the case of the ratio is lower than 25%, Battery is too low Notification will appear. Samsung Confidential
[11]
: : () Times New Roman, ()
: : () Times New Roman, () In the case of the ratio is lower than 10%, Change Devices Battery! Notification will appear. B. ECG View Double Tapping on ECG View will toggle MONITORING and PAUSE mode. In PAUSE mode, its enabled that Scrolling by swipe gesture on ECG Graph. Scaling
(magnification ratio changing 50 ~ 150%, by 10%) by zooming gestures is able for both modes. B.1. Measured ECG Signal: Measured ECG signals B.2. ECG Paper Scale: Scale of ECG Paper. Its same as actual size in case of the magnification ratio is 100%
B.3. Heart Rate: Current Measured Heart Rate and Maximum/Minimum HR B.4. Respiratory Rate: Measured Respiratory Rate. The unit of the respiratory rate is BrPM
(Breath per minute). Actual range of the respiratory rate is from 9 BrPM to 30 BrPM. The resolution is 0.96 BrPM.
** During measure the Respiratory Rate, minimize move. B.5. PAUSE Indicator: PAUSE mode indicator. Even in PAUSE mode, received signal will be stored. B.6. Left-side Time: The elapsed time of the left-side signal B.7. Right-side Time: The elapsed time of the right-side signal B.8. Magnification Ratio: The magnification ratio of the ECG signal and ECG Paper. Its invisible in case of the ratio is 100%
C. Heart Rate Trend View: Heart rate trend graph for 1 Hour. D. Menu Button: Its translucent. Touching the button makes it opaque and show menu buttons. Nothing selected until 3 seconds then it backs to translucent and menu buttons will be disappeared. D.1. Change Device: Device Select View will appear D.2. Set MARK: add User event mark D.3. Show/Hide STATUS Bar: show/hide STATUS Bar Samsung Confidential
[12]
D.4. ECG Views Background:
BLACK: black background, green signal PAPER: ECG Paper background, black signal D.5. Import from file: read signal data from stored file. S-PATCH format(*.sp) and MIT BIH format(.hea/.dat, 16 format - only for exported by this app) can be loaded. If there is no folder stored signal files, check your system options like below. In case of Android OS, Internal Storage would be hidden. Samsung Confidential
[13]
D.6. Export to file: Received Signal data automatically stored to devices storage as formatted .sp (S-PATCH format). This button makes MIT BIH format file. The file name is automatically generated by based on the Device Number and the first connection time. D.7. Show HRV View: shows HRV information view D.8. Show/Hide HR Trend View: show/hide HR Trend View HR Trend View shows HR trend and minimum/maximum HR for 1 Hour. Samsung Confidential
[14]
E. HRV View: shows HRV information E.1. Time Domain HRV E.1.1. AVNN: mean of all the intervals in ms E.1.2. SDNN: standard deviation all RRIs over the complete test duration in ms E.1.3. SDANN: standard deviation of the 5 min means in ms E.1.4. ASDNN: mean of the 5 min standard deviation in ms E.1.5. NN50: the number of RRI difference of successive RRIs greater than 50ms E.1.6. PNN50: NN50 as a percentage of all allowed RRIs E.1.7. RMSSD: root mean square of successive differences in ms E.1.8. TINN: triangular index interval is the baseline width of the distribution measured as E.2. Frequency Domain HRV E.2.1. VLF: very low frequency power (0.00333 Hz to 0.040 Hz) E.2.2. LF: low frequency power (0.040 Hz to 0.150 Hz) E.2.3. HF: high frequency power (0.150 Hz to 0.400 Hz) a base of a triangle approximating the RRI distribution
** During measure the HRV, minimize move. Samsung Confidential
[15]
F. Device Select View: scanning S-PATCH devices nearby and show IDs. F.1. Devices: Scanned S-PATCH devices are shown. One of them is selected then connection will be tried. The number in bracket after S-PATCH name is the Device ID. RSSI is Received signal strength indication in a wireless environment, in arbitrary units. It could be used to calculate distance of the S-PATCH and the host device. F.2. Nothing available: This will disconnect connected device and stay without any F.3. Cancel: This will back without any changes F.4. App version: This shows Apps version connection Samsung Confidential
[16]
G. Signal Viewer G.1. File Name: Loaded File Name G.2. Total Time Range: Stored total signals range G.3. Left-side time of signal: Signal Viewers left side time G.4. Right-side time of signal: Signal Viewers right side time G.5. Near Signal Navigation: This shows signals of larger range than Signal Viewer. Swipe gesture will scroll the time range. G.6. Small Time Range Navigation: This supports specific time selection in small time range by touch. If total time range is too small, this will disappear. G.7. Total Time Range Navigation: This supports the selection of the Small Time Range Navigations range on the total time range. H. Viewer Menu Button BACK button supports return of previous window as like as Android Devices Cancel button. Other buttons are same as S-PATCH real-time signal display state. Samsung Confidential
[17]
The app shows as Landscape orientation as following as devices orientation. 3.4 How to Install S-PATCH3 3.4.1 Installing S-PATCH 1) Download the S-PATCH App from Google Play Store (Android KitKat or later). The device must support Bluetooth Low Energy. 2) Install the downloaded file on users mobile device. 3) Insert battery to main body 1 of S-PATCH3. 4) Connect Electrodes to electrode connect hole in each side of S-PATCH3. 3.5 Precautions 1) The electrodes and coin battery are disposable. Please observe local laws for disposal of electrodes and coin battery. 2) Please ensure users hands are clean and dry when handle the S-PATCH3 3) The body hair possibly causes contact problem which results a fault detection of physiological data. 4) After using the device, wipe dust and other foreign substances using a dry cloth. 5) This product must not be disposed of with your other household waste. Instead, it is your responsibility to dispose of your waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic Samsung Confidential
[18]
equipment. The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment. For more information about where you can drop off your waste equipment for recycling, please contact your local city office, your household waste disposal service or the shop where you purchased the product. Samsung Confidential
[19]
4. Using and Operating S-PATCH3 4.1 How to Use S-PATCH3 1) After connect the electrodes with S-PATCH3, peel the plastic on the backside of the electrodes. 2) Attach the S-PATCH3 on recommended position as drawn. 3) Push the power button longer than 3 seconds. 4) Confirm that the LED is ON and OFF 3 times. Samsung Confidential
[20]
< ON >
< OFF >
5) Put the Bluetooth mode to ON mode of users mobile device. 6) Please execute S-PATCH App. 7) Please touch the Menu Button located right bottom side like as below. 8) Menu Buttons will appear then touch DEVICE button Buttons will disappear if there is no selection for 3 seconds. 9) Connectable S-PATCH Devices will be added to the list. Samsung Confidential
[21]
If the app has not yet proper permission to access the host devices BLE, permission requesting messages will be shown like below. S-PATCH app uses BLE and storage of the host device. And BLE needs Location Permission, so, BLE, Location and Storage permissions are requested. If its denied one of them, the app could not receive and store signal data. 10) Select the device to be connected. 11) After pairing S-PATCH device, top-right sides connection indicator will be updated, and ECG Signals will be shown as following as BLE transmission.
[22]
Samsung Confidential Received signal data is automatically stored to the host devices storage as file under
/SPATCH folder. The file and folder name is generated by S-PATCH device number and connection time. Above example means S-PATCH (3105) devices signals are received since 2015 year January 1st day PM 1:30:47. Same names folder has detail data. Algorithm Result (Heart rate, Respiratory Rate and Heart rate variability) will be updated after valid signals are received for required time range for each algorithm.
[23]
Samsung Confidential 12) Select N/A to finish recording on Device Selection 13) To load a stored signal, select IMPORT button through Menu-Button and choose the file. Then it will be loaded like below. Bottom navigation and swiping on signal graphs are able to move on the showing time. And HR, RR and HRV values are updated as following as the time. If the S-PATCH device is connected, received signal will be stored in background. Samsung Confidential
[24]
BACK button of Menu Buttons will back. 14) To finish S-PATCH app, push BACK button of Android device. To prevent unintended finish, below message is shown. If BACK button is pressed again during 1 second, one more confirmation is required. YES will terminate the app. NO will not. In case of using iOS, just push HOME button. 4.2 Specifications 4.2.1 S-PATCH3 Specifications Classification Performance Circuitry Name Type Channel Communication Mobile device DC Offset Tolerance ADC Resolution ADC Sampling Rate Description CF-Type Single channel with Bluetooth
+/- 300mV 12 bits 1000 Samples/Second Input Impedance
>100 M Power Requirements Software Physical Characteristics Power Supply Battery Life F/W version CPU BLE App(Android) Weight (Exc. Battery) Dimension(Main Body1,2) Dimension(Connect Cable) DC 3V , Coin Battery ( CR2032 )
>48 Hr. V1.00 S1SBP6A (Cortex-M4F microprocessor) Dialog DA14583(ARM Cortex-M0) Kitkat(API19) or later 6g 38*27*5 (mm), 38*27*3 (mm)[W*D*H]
110 5 (mm) Samsung Confidential
[25]
4.3 Error Message and Alarm 4.3.1 Device Disconnection If the BLE is disconnected on receiving signal, BLE connection indicator is updated. And notification will appear with alarm or vibration as user s setting.
(Mute setting woulddoes not make any alarm sound and vibration) 4.3.2 Abnormal signal detect In case of algorithm detects invalid signal, assume LEAD OFF state and show below.
[26]
Samsung Confidential And alarm like below. All algorithms are not available. After valid signal is detected, its changed to LEAD ON state and algorithms are restarted. 4.3.3 Low Battery mode Battery state is updated on top-right of the STATUS bar. Battery Icons meaning is like below. Samsung Confidential
[27]
In the case of the ratio is lower than 25%, Battery is too low Notification will appear. In the case of the ratio is lower than 10%, Change Devices Battery! Notification will appear. 4.3.4 Low Remaining Storage If the remaining storage is under 10MB, caution will appear. Then make some more remaining storage. If not, signal would be lost. Samsung Confidential
[28]
4.4 Maintenance For cleaning, gently wipe with a soft dry cloth after using the S-PATCH 3. Please attempt to keep S-PATCH3 dust free. S-PATCH is waterproof. But it should be kept dry. This device does not have serviceable components. Do not disassemble, crush, puncture, short external contacts or circuits, dispose of in fire or water. Samsung Confidential
[29]
5. Labels and Packaging 5.1 Labels 5.1.1 Label for Packaging 5.1.2 Label for module Front- side 2 Labels Rear-side 2 Labels Samsung Confidential
[30]
- Descriptions on the visual symbols of the label Label Symbol Description 1 2 3 4 5 6 7 8 9 The serial number that identifies the object. Date of manufacture Manufacturer Caution / Warning Instruction for User Manual Type of applied part Prescription only WEEE Mark Temperature limitation 5.2 Packaging 5.2.1 Package List Classifications Components Quantity Main Box S-PATCH3 module Electrodes Battery Manual 1 4 1 1 Samsung Confidential
[31]
5.2.2 Package Units 1) S-PATCH3 module Main module. 2) Electrodes 4 Electrodes are in the 1 plastic bag. 3) Battery 1 Coin Cell battery is shipped at package. 4) Manual Samsung Confidential
[32]
6. Information on EMC 6.1 Guidance and Manufacturers Declaration Electromagnetic Emissions
- The EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT should assure that it is used in such an environment. Immunity test Compliance Electromagnetic environment - Guidance RF Emissions CISPR 11 Group 1 The EUT uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class B The EUT is suitable for use in ail establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000-3-2 Voltage fluctuations /
Flicker emissions IEC 61000-3-3 Class A Complies 6.2 Guidance and Manufacturers Declaration Electromagnetic Immunity
- The EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT should assure that it is used in such an environment. Immunity test IEC 60601 Test level Compliance level Electromagnetic environment -
Guidance Electrostatic discharge (ESD) IEC 61000-4-2 Electrical fast transient/burst IEC 61000-4-4 Surge IEC 61000-4-5 6kV Contact 8kV air 6kV Contact 8kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. 2kV for power supply lines 1kV for input/output lines 2kV for power supply lines 1kV for input/output lines 1kV differential mode 2kV common mode 1kV differential mode 2kV common mode Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. Samsung Confidential
[33]
<5% U
(>95% dip in U) for 0.5cycle 40% U
(60% dip in U ) for 5 cycle 70% U
(30% dip in U) for 25 cycle
<5% U
(<95% dip in U ) for 5 s
<5% U
(>95% dip in U) for 0.5cycle 40% U
(60% dip in U) for 5 cycle 70% U
(30% dip in U) for 25 cycle
<5% U
(<95% dip in U) for 5 s 3 A/m 3 A/m Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 Power frequency
(50/60Hz) magnetic field IEC 61000-4-8 Mains power quality should be that of a typical commercial or hospital environment. If the user of the EUT image intensifier requires continued operation during power mains interruptions, it is recommended that the EUT image intensifier be powered from an uninterruptible power supply or a battery. Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE U is the A.C. Mains voltage prior to application of the test level. 6.3 Guidance and Manufacturers Declaration Electromagnetic Immunity
- The EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT should assure that it is used in such an environment. Immunity test IEC 60601 Test level Compliance level Electromagnetic environment - Guidance Portable and mobile RF communications equipment should be used no closer to any part of the EUT, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3 3 Vrms 150 kHz to 80MHz 3 V/m 80 MHz to 2.5GHz 3 Vrms 150 kHz to 80MHz 3 V/m 80MHz to 2.5GHz Samsung Confidential
[34]
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as deter-mined by an electromagnetic site a survey, should be less than the compliance level in b each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol :
NOTE 1) At 80MHz and 800MHz, the higher frequency range applies. NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the EUT is used exceeds the applicable RF compliance level above, the EUT should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the EUT. b Over the frequency range 150kHz to 80MHz, field strengths should be less than [V 1
] V/m. 6.4 Recommended separation distances between portable and mobile RF communications equipment and the EUT
- There is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the EUT can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the EUT as recommended below, according to the maximum output power of the communications equipment. Samsung Confidential
[35]
Separation distance according to frequency of transmitter [m]
Rated maximum output 150kHz to 80MHz 80MHz to 800MHz 800MHz to 2.5GHz power of transmitter
[W]
0.01 0.1 1 10 100 V1=3Vrms
=3V/m E 1 E
=3V/m 1 0.116 0.368 1.166 3.687 11.660 0.1166 0.3687 1.1660 3.6872 11.6600 0.2333 0.7378 2.3333 7.3785 23.333 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies. NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. 6.5 Immunity and Compliance Level Immunity test IEC 60601 Test Level Actual Immunity Level Compliance Level Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3 3Vrms 150kHz to 80MHz 3Vrms 80MHz to 2.5GHz 3Vrms 3Vrms 3V/m 3V/m Samsung Confidential
[36]
6.6 Guidance and Manufacturers Declaration Electromagnetic Immunity
- The EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3 3 Vrms 150 kHz to 80MHz 3 V/m 80 MHz to 2.5GHz 3 Vrms 150 kHz to 80MHz 3 V/m 80MHz to 2.5GHz Electromagnetic environment - Guidance The EUT must be used only in a shielded location with a minimum RF shielding effectiveness and, for each cable that enters the shielded location with a minimum RF shielding effectiveness and, for each cable that enters the shielded location Field strengths outside the shielded location from fixed RF transmitters, as determined by an electromagnetic site survey, should be less a than 3V/m. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. NOTE 2) It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to assure that they meet the minimum specification. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength outside the shielded location in which the EUT is used exceeds 3V/m, the EUT should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as relocating the EUT or using a shielded location with a higher RF shielding effectiveness and filter attenuation. Samsung Confidential
[37]
This device may not cause harmful interference, and This device must accept any interference received, including interference that may Reorient or relocate the receiving antennae Increase the separation between the equipment and the receiver Connect the equipment into an outlet on a circuit different from that to which the 7. FCC Compliance Statement This device complies with part 15 of the FCC rules. Operation is subject to the following two conditions:
(1)
(2) cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment on and off, the user is encouraged to try to correct the interference by one or more of the following measures:
receiver is connected. FCC RF Exposure Statement This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. End users must follow the specific operating instructions for satisfying RF exposure compliance. The antenna used for this transmitter must not transmit simultaneously with any other antenna or transmitter, except in accordance with FCC multi-transmitter product procedures. FCC Caution Any changes or modifications to the equipment not expressly approved by the party responsible for compliance could void users authority to operate the equipment. 8. Product Life time and Warranty Consult the dealer or an experienced radio/TV technician for help. 8.1 Product Life time
- S-PATCH3 is guaranteed to last for a minimum of two years. 8.2 Product Warranty
- The warranty covers a year. However, the conditions excluded by the warranty are as follows. Natural aging of the product from daily usage Samsung Confidential
[38]
Product damage due to improper storage Product damage due to improper usage 9. Company Introduction 9.1 Company Name and Address Company name: Samsung Electronics Co., Ltd. Address: 1-1, Samsungjeonja-ro, Hwaseong-si, Gyeonggi-do, 18448 Korea Tel : +(82) 31-8037-3533 e-mail: s.patch@samsung.com Samsung Confidential
[39]
frequency | equipment class | purpose | ||
---|---|---|---|---|
1 | 2017-06-06 | 2402 ~ 2480 | DTS - Digital Transmission System | Original Equipment |
app s | Applicant Information | |||||
---|---|---|---|---|---|---|
1 | Effective |
2017-06-06
|
||||
1 | Applicant's complete, legal business name |
Samsung Electronics Co Ltd
|
||||
1 | FCC Registration Number (FRN) |
0027908797
|
||||
1 | Physical Address |
19 Chapin Rd., Building D
|
||||
1 |
Pine Brook, New Jersey 07058
|
|||||
1 |
United States
|
|||||
app s | TCB Information | |||||
1 | TCB Application Email Address |
t******@siemic.com
|
||||
1 | TCB Scope |
A4: UNII devices & low power transmitters using spread spectrum techniques
|
||||
app s | FCC ID | |||||
1 | Grantee Code |
A3L
|
||||
1 | Equipment Product Code |
SPATCH3
|
||||
app s | Person at the applicant's address to receive grant or for contact | |||||
1 | Name |
J******** C******
|
||||
1 | Title |
General Manager
|
||||
1 | Telephone Number |
973-8********
|
||||
1 | Fax Number |
973-8********
|
||||
1 |
j******@samsung.com
|
|||||
app s | Technical Contact | |||||
n/a | ||||||
app s | Non Technical Contact | |||||
n/a | ||||||
app s | Confidentiality (long or short term) | |||||
1 | Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | Yes | ||||
1 | Long-Term Confidentiality Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?: | Yes | ||||
1 | If so, specify the short-term confidentiality release date (MM/DD/YYYY format) | 12/03/2017 | ||||
if no date is supplied, the release date will be set to 45 calendar days past the date of grant. | ||||||
app s | Cognitive Radio & Software Defined Radio, Class, etc | |||||
1 | Is this application for software defined/cognitive radio authorization? | No | ||||
1 | Equipment Class | DTS - Digital Transmission System | ||||
1 | Description of product as it is marketed: (NOTE: This text will appear below the equipment class on the grant) | Wearable Health Monitoring System | ||||
1 | Related OET KnowledgeDataBase Inquiry: Is there a KDB inquiry associated with this application? | No | ||||
1 | Modular Equipment Type | Does not apply | ||||
1 | Purpose / Application is for | Original Equipment | ||||
1 | Composite Equipment: Is the equipment in this application a composite device subject to an additional equipment authorization? | No | ||||
1 | Related Equipment: Is the equipment in this application part of a system that operates with, or is marketed with, another device that requires an equipment authorization? | No | ||||
1 | Grant Comments | Power listed is the maximum conducted output power. | ||||
1 | Is there an equipment authorization waiver associated with this application? | No | ||||
1 | If there is an equipment authorization waiver associated with this application, has the associated waiver been approved and all information uploaded? | No | ||||
app s | Test Firm Name and Contact Information | |||||
1 | Firm Name |
DT&C Co., Ltd.
|
||||
1 | Name |
W****** L******
|
||||
1 | Telephone Number |
82-31********
|
||||
1 | Fax Number |
82-31********
|
||||
1 |
l******@dtnc.net
|
|||||
Equipment Specifications | |||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Line | Rule Parts | Grant Notes | Lower Frequency | Upper Frequency | Power Output | Tolerance | Emission Designator | Microprocessor Number | |||||||||||||||||||||||||||||||||
1 | 1 | 15C | 2402.00000000 | 2480.00000000 | 0.0004000 |
some individual PII (Personally Identifiable Information) available on the public forms may be redacted, original source may include additional details
This product uses the FCC Data API but is not endorsed or certified by the FCC