FCC ID: 2BMCJGMIX-TRI-03 Geri Mixer

by: Suzhou Basecare Medical Device Co., Ltd. latest update: 2026-03-23 model: GMIX-TRI-03

Two grants have been issued under this FCC ID on 03/23/2026 (this is one of five releases in 2026 for this grantee). Public details available include manuals, internal & external photos, and frequency information. some products also feature user submitted or linked instructions, drivers, troubleshooting guides, warrantee terms, & password defaults.
permalink for this FCC identifier
|
Get Slack Alerts
|
|   
eg for Slack usage
manual
app s submitted / available
1 2 Users Manual pdf 1.10 MiB March 23 2026
1 2 Attestation Statements pdf March 23 2026
1 2 Attestation Statements pdf March 23 2026
1 2 Attestation Statements pdf March 23 2026
1 2 Test Report pdf March 23 2026
1 2 Test Report pdf March 23 2026
1 2 Cover Letter(s) pdf March 23 2026
1 2 Test Report pdf March 23 2026
1 2 Test Report pdf March 23 2026
1 2 Cover Letter(s) pdf March 23 2026
1 2 BLOCK Block Diagram pdf March 23 2026 confidential
1 2 External Photos pdf March 23 2026
1 2 Cover Letter(s) pdf March 23 2026
1 2 Cover Letter(s) pdf March 23 2026
1 2 Internal Photos pdf March 23 2026
1 2 ID Label/Location Info pdf March 23 2026
1 2 RF Exposure Info pdf March 23 2026
1 2 OP Operational Description pdf March 23 2026 confidential
1 2 SCH Schematics pdf March 23 2026 confidential
1 2 SOFTWARESECURITYREQUIREMENTSFORU-NIIDEVICES SDR Software/Security Info. pdf March 23 2026 confidential
1 2 Test Setup Photos pdf March 23 2026
1 2 Attestation Statements pdf March 23 2026
1 2 Cover Letter(s) pdf March 23 2026
1 2 Block Diagram Block Diagram pdf March 23 2026 confidential
1 2 External Photos pdf March 23 2026
1 2 ID Label/Location Info pdf March 23 2026
1 2 Test Report pdf March 23 2026
1 2 Test Setup Photos pdf March 23 2026

 

Geri Mixer GMIX-TRI-03 User Manual Genea Biomedx Pty Ltd i GENERAL INFORMATION Copyright This user manual and all contents contained herein are subject to copyright. All rights reserved. No part of this user manual may be reproduced, copied, translated or transmitted in any form or by any means without the prior written approval of Genea Biomedx. Technical Support Manufacturer If any serious incident has occurred in relation to this device, it should be reported directly to the manufacturer and the competent authority of the Member State in which the user and/or patient is established. Genea Biomedx Pty Ltd Level 2, 321 Kent Street Sydney, New South Wales, 2000, Australia Country of origin: China Email: [email protected] Web: www.geneabiomedx.com ii List of accessories No. 1 2 3 4 5 6 Content Main Device Power Cord Gas Connecting Pipe Fuse Gas Pipe Connector Supporting Documents GMIX-TRI-03 1 1 1 2 1 1 In case of any missing accessories, please promptly contact your local Genea Biomedx representative. This product does not contain any parts that are prone to wear and tear or loss. The accessories require replacement only in the event of damage.

*Note: In some cases, the User Manual can be obtained digitally as an eIFU; please contact your local Genea Biomedx representative to get the latest version of the User Manual. Copyright description This manual contains proprietary information protected by copyright law. All rights reserved. Reproduction, photography, or translation of any part of the Manual into other languages is prohibited without the express written consent of the manufacturer. See the nameplate for the date of manufacture for the device. iii Table of contents 1. Safety &Precautions ................................................................................................... 1 1.1 1.2 Important Notice ................................................................................................ 1 Safety Tips ........................................................................................................... 1 1.2.1 General Security .................................................................................... 1 1.2.2 Electrical safety ..................................................................................... 2 1.2.3 Accessory safety .................................................................................... 2 1.3 Definition of icons, symbols and abbreviations ................................................ 2 2. Instrument Overview ................................................................................................. 4 2.1 Intended Use ...................................................................................................... 4 2.2 Working Principle ............................................................................................... 4 2.3 System composition ........................................................................................... 4 3. Instrument Installation ............................................................................................... 7 3.1 3.2 3.3 3.4 Normal operating conditions ............................................................................. 7 Check the instrument ......................................................................................... 8 3.2.1 Access Rights......................................................................................... 8 Prepare the power supply .................................................................................. 9 Instrument Installation ....................................................................................... 9 4. Instrument Operation .............................................................................................. 10 4.1 4.2 4.3 4.4 Equipment Installation ..................................................................................... 10 Power on and off .............................................................................................. 10 Start and stop ................................................................................................... 10 Operation mode switching ............................................................................... 10 4.5 Mute Alarm ....................................................................................................... 11 5. Maintenance and care ............................................................................................. 11 5.1 Main device maintenance ................................................................................ 11 5.2 5.3 5.4 Power supply maintenance .............................................................................. 11 Software Maintenance ..................................................................................... 11 Transportation and storage .............................................................................. 12 6. Warnings & Faults: Troubleshooting ........................................................................ 12 iv 1. Safety &Precautions This chapter introduces the precautions for using the instrument. Before using, please read and understand the contents carefully to ensure correct use and operator safety. 1.1 Important Notice In this manual, some words are preceded by the following symbols for warnings and special reminders. The specific usage is as follows:

Item Description Prompts the operator to follow the instructions, otherwise personal Warning injury may result. The operator is reminded to follow the instructions, otherwise the Caution instrument may malfunction or be damaged. 1.2 Safety Tips 1.2.1 General Security Warning Do not perform any operation on the instrument before reading the user manual to avoid damage to the instrument or personal injury. This instrument must be operated by trained doctors or laboratory technicians. Do not place the instrument in a humid place or near water to avoid leakage or electric shock caused by low insulation. Do not use this instrument in flammable or explosive environments, otherwise it may cause personal injury. When the power of the instrument is on, unauthorized personnel are not allowed to dismantle the instrument casing to avoid personal injury. Caution 1 This instrument can only be disassembled, debugged and repaired by our authorized and qualified technical support personnel. Improper operation may affect the stability of the instrument or damage the instrument. This instrument is not waterproof, please do not spray liquid on it. Otherwise, it may cause damage to the parts. If liquid enters the instrument, please turn off the power immediately and contact the technician in time. Do not use accessories and consumables other than those specified. 1.2.2 Electrical safety Warning It is prohibited to use this instrument near strong radiation sources (such as unshielded high-power radio frequency sources). Otherwise, it may interfere with the normal operation of the instrument. Do not remove the instrument cover to expose the internal components. Otherwise, there may be a risk of electric shock. If the power cord is damaged, frayed, cracked, or broken, replace it immediately. Do not use an underrated power cord instead of the power cord for this instrument. Otherwise, it may overheat and cause a fire. Caution Please make sure that the input voltage meets the requirements of this instrument. Please use the power cord provided with this instrument to connect to the power supply. 1.2.3 Accessory safety Replace it with a fuse of the specifications specified by Genea Biomedx (250V, F6AL). Please contact after-sales for specific operation methods. 1.3 Definition of icons, symbols and abbreviations 2 Definition of icons, symbols and abbreviations used on the labels of the product is as follows:

Sign Description Indicates the device manufacturer Indicates the date when the device was manufactured Indicates the manufacturers catalogue number Indicates the manufacturers serial number so that a specific device can be identified Indicates the need for the user to consult the user manual Contents of the distribution packages are fragile therefore it shall be handled with care This is the correct upright position of the distribution packages for transport and/or storage Distribution packages shall not be rolled or turned over Distribution packages shall be kept away from rain and be kept in dry conditions Distribution packages shall not be exposed to sunlight Can only be placed in a single layer during transportation Temperature limit(s) Humidity limit(s) For connecting to WIFI 3 2. Instrument Overview This chapter introduces the intended use, working principle and system composition of the instrument. Geri Mixer (GMIX-TRI-03) is a high-precision, high-reliability gas mixing device developed by Genea Biomedx. The device uses advanced gas sensing technology and microprocessor control to accurately control and mix CO2, O2, and N2 to ensure the stability and accuracy of the mixed gas. The output mixed gas has a CO2 concentration of 6%, an O2 concentration of 5%, and the rest is N2. 2.1 Intended Use The three gases CO2, O2 and N2 are precisely mixed in a specific proportion to prepare a uniform mixed gas. 2.2 Working Principle The Geri Mixer blends CO, O, and N to a default composition of 6% CO, 5% O, with the balance N. The mixed gas is first delivered into a 5 L gas tank and then supplied to external lines at a default outlet pressure of 0.15 MPa. Tank pressure is monitored in real time: when it falls below 0.25 MPa, the high-precision gas mixer replenishes the tank; when the pressure reaches 0.35 MPa, refilling stops automatically. 2.3 System composition The system is composed of high-precision gas distribution system, gas storage tank, pipeline system, electronic control system and casing. Composition Function High-precision gas distribution system By default, the gas mixture is set to 6% CO, 5% O, and the remainder N. Gas storage tank Cache the prepared gas mixture. Pipeline system Connect the gas distribution system with the gas tank and realizes external gas supply. Electronic control system Control high-precision gas distributors. 4 Casing Used to cover the surrounding of the instrument and protect the internal parts of the instrument. 2.4 Basic components No. Name Function 1 2 3 Cooling holes Used for instrument heat dissipation. Display Casters Used to display the current status of the instrument. Used to move and support the instrument. 5 No. Name Function 1 2 3 4 5 6 7 8 To turn the instrument power on or off:

Power switch Switch to position to turn on the power. Switch to position to turn off the power. USB interface Connects to an external PC for monitoring and modifying the control program. Nitrogen interface Used to connect 99.999% high purity N2 gas cylinders. Oxygen interface Used to connect 99.999% high-purity O2 gas cylinders. Carbon Dioxide interface Used to connect 99.95% high purity CO2 cylinders. Premixed gas interface Used to connect to 6% CO2 +5% O2 +89% N2 premixed gas cylinder. Sampling port For gas sampling. Drain port For emptying gas tanks. 6 9 10 11 12 13 14 15 Air outlet For access to embryo incubator. Pressure reducing valve Used for outlet pressure regulation, the default value is 0.15MPa. Operation mode switch knob Used to switch the working state of high-

precision gas distributor. Device switch Used to control the device to start/stop working Alarm interface For external alarm Mute button Used to eliminate alarm sound. Thumb screw Used to fix the housing on both sides. 3. Instrument Installation Instrument installation requires advance preparation, including laboratory preparation before instrument installation and preparation before instrument use. 3.1 Normal operating conditions Power supply: 100 V240 V AC, 50 Hz/60 Hz, power consumption: 150 VA Ambient temperature: 10 C 50 C Relative humidity: 80 %

Place the instrument on a stable, level surface away from heaters; avoid moisture and direct sunlight. Keep the instrument away from flammable or corrosive gases. Maintain at least 100 mm of clear space on all sides to ensure adequate ventilation. Gas source requirements:

No. 1 2 3 Gas Source CO2 O2 N2 Requirements 99.95 % purity, moisture-free. Cylinder must be fitted with a pressure regulator set to 0.4 MPa0.6 MPa. 99.999 % purity, moisture-free. Cylinder must be fitted with a pressure regulator set to 0.4 MPa0.6 MPa. 99.999 % purity, moisture-free. Cylinder must be fitted with a pressure regulator set to 0.4 MPa0.6 MPa. 7 4 Premixed gas

(spare) Moisture-free blend containing 6% 0.5% CO and 5% 0.5% O, balance N. Cylinder must be fitted with a pressure regulator set to 0.4 MPa0.6 MPa. 3.2 Check the instrument This instrument can only be unpacked by after-sales personnel. After receiving the instrument, please contact after-sales personnel for unpacking and installation. After receiving the instrument, please check whether the outer packaging is intact and whether the anti-shock and anti-tilt labels are abnormal. If there are any abnormalities, please contact the after-sales staff in time. This instrument contains precision components. In order to monitor the impact on the instrument during transportation, anti-vibration labels and anti-tilt labels are affixed to the outer packaging box. The meaning of the transistor status in the label is as follows. Label transistor status explanation Label Transistor status explanation No color change: The instrument is intact and there is no strong collision during transportation or the collision degree does not exceed the required range. Turns red: Indicates that the instrument has been severely impacted during transportation and exceeds the required range. No color change: The instrument is intact and has not tilted during transportation or the tilt degree does not exceed the required range. Turns red: Indicates that the instrument is tilted during transportation and exceeds the required range. Shockproof label Anti-tilt label 3.2.1 Access Rights This instrument operates as a standalone system; users cannot modify the control program. For any required changes, please contact the authorized and certified technical support personnel. 8 3.3 Prepare the power supply In order to ensure the safe and stable operation of this instrument, it is recommended to equip it with a UPS power supply (please consult the relevant manufacturer for installation methods). UPS power requirement 3kVA (including server power supply), please connect the instrument to the UPS power output terminal. Before using the instrument, please use a voltmeter or multimeter to measure the voltage of the AC power socket or UPS output terminal to ensure that the voltage is within the required range, otherwise the performance of the electronic components of the instrument may be reduced. Such as directly connecting the instrument to the mains power supply. The power requirements are as follows:

Wall socket power supply voltage: 100-240V~, 50/60Hz, power supply voltage fluctuation should not exceed 10% of the nominal voltage. Rated power: 150VA, supply current not less than 10A. Transient overvoltage category:

3.4 Instrument Installation 1. Upon receiving the instrument, first confirm that the anti-tilt and anti-shock indicators on the wooden crate are intact. Only then should you dismantle the crate and position the instrument in the prepared location. 2. Gas supply requirements:

No. 1 2 3 4 Gas Source CO2 O2 N2 Premixed gas

(spare) Requirements 99.95 % purity, moisture-free. Cylinder must be fitted with a pressure regulator set to 0.4 MPa0.6 MPa. 99.999 % purity, moisture-free. Cylinder must be fitted with a pressure regulator set to 0.4 MPa0.6 MPa. 99.999 % purity, moisture-free. Cylinder must be fitted with a pressure regulator set to 0.4 MPa0.6 MPa. Moisture-free blend containing 6% 0.5% CO and 5% 0.5% O, balance N. Cylinder must be fitted with a pressure regulator set to 0.4 MPa0.6 MPa. 3. The simplified diagram of equipment gas pipeline connection is as follows:

9 Note: The equipment in the dotted box is provided by the user. The gas concentration tester shall be connected and tested by the company's authorized and qualified technical support personnel when necessary. 4. Connect the power supply. If the instrument can start normally after power-on, the installation is complete. 4. Instrument Operation This chapter introduces how to use the instrument. To ensure the normal operation of the instrument, please operate it as required. 4.1 Equipment Installation Connect the pipes and power supply to the instrument according to 3.4. (The gas concentration tester may not be connected) 4.2 Power on and off The instrument is powered on/off by the power switch. Switch to position to turn on the power. Switch to position to turn off the power. 4.3 Start and stop Use the device switch to control the instrument to start/stop working. 4.4 Operation mode switching Use the operating mode switching knob to switch the working state of the high-precision gas 10 distributor: AUTO mode; SG mode. AUTO mode: The instrument mixes gases to a fixed ratio of 6% CO, 5% O, and the balance N. After blending, the mixture is stored in a gas tank and delivered from there. SG mode: Directly supply gas to the outside from premixed gas (spare). 4.5 Mute Alarm Press the Mute Button to mute the device alarm sound. The mute time is 3 minutes. If the alarm is not cleared after 3 minutes, the alarm sound will sound again. If the alarm is cleared within 3 minutes and there is no new alarm, the alarm sound will not sound again after 3 minutes. 5. Maintenance and care 5.1 Main device maintenance With the power off, perform the following maintenance:

Once a month, it is recommended to wipe the instrument casing (including the display) with a dust-free cloth;

Use a small brush to remove dust from the small holes on the side of the instrument. If the instrument is not used for more than 7 days, the external gas source should be disconnected. 5.2 Power supply maintenance Warning If the instrument is not used for more than 7 days, the power should be turned off. Before each use, check that the "power switch" is in the off position. Before each use, check that the cables are properly connected and intact. If damaged, replace them promptly. 5.3 Software Maintenance The maintenance and update of the built-in software of this instrument must be performed by professional technicians or under the guidance of technicians. If the software runs abnormally, 11 please contact the technicians in time. 5.4 Transportation and storage Packaging Fully enclosed wooden crate. Storage Environment Transport Other Temperature: 20C to +65 C. Humidity: 80 % RH at 25 C. Store indoors in an environment free of acidic, alkaline, or any other corrosive gases. May be shipped by road, air, sea, or rail. Protect from direct rain or snow during transit. No stacking or excessive pressure. Keep upright. Avoid impact. 6. Warnings & Faults: Troubleshooting Problem Cause Solution Unable to turn on The alarm sounds and the CO2 indicator turns red on the display. The alarm sounds and O2 turns red on the display. The alarm sounds and N2 turns red on the display. The alarm sounds and SG on the display turns red. Power connected cord not The power switch is not turned on Connect the power cord and turn on the power switch 1. Verify that the CO cylinder output pressure Insufficient CO2 pressure is set to 0.40.6 MPa. 2. Replacement of CO2 cylinder 1. Verify that the O cylinder output pressure is Insufficient O2 pressure set to 0.40.6 MPa. 2. Replace O2 cylinder 1. Verify that the N cylinder output pressure is Insufficient N2 pressure set to 0.40.6 MPa. 2. Replace N2 cylinder 1. Verify that the Span Gas cylinder output Insufficient SG pressure pressure is set to 0.40.6 MPa. 2. Replace Span Gas Cylinders 12 FCC Warning This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Any Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

-Reorient or relocate the receiving antenna.

-Increase the separation between the equipment and receiver.

-Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.

-Consult the dealer or an experienced radio/TV technician for help. This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance 20cm between the radiator & your body. 13

 

1 2 Users Manual pdf 1.10 MiB March 23 2026
frequency equipment class purpose
1 2026-03-23 5745 ~ 5825 NII - Unlicensed National Information Infrastructure TX Original Equipment
2 2412 ~ 2462 DTS - Digital Transmission System

Application Forms

app s Applicant Information
1 2 Effective 2026-03-23
1 2 Applicant's complete, legal business name Suzhou Basecare Medical Device Co., Ltd.
1 2 FCC Registration Number (FRN) 0036309508
1 2 Physical Address 77 Jingu Road, Industrial Park
1 2 Suzhou, Jiangsu Province,, N/A 215000
1 2 China
app s TCB Information
1 2 TCB Application Email Address c******@micomlabs.com
1 2 TCB Scope A4: UNII devices & low power transmitters using spread spectrum techniques
app s FCC ID
1 2 Grantee Code 2BMCJ
1 2 Equipment Product Code GMIX-TRI-03
app s Person at the applicant's address to receive grant or for contact
1 2 Name L****** h****
1 2 Title Manager
1 2 Telephone Number 18550********
1 2 Fax Number 18550********
1 2 E-mail p******@basecare.cn
app s Technical Contact
1 2 Firm Name Suzhou Basecare Medical Device Co., Ltd.
1 2 Physical Address China
1 2 E-mail p******@basecare.cn
app s Non Technical Contact
1 2 Firm Name AOGES CO., LIMITED
1 2 Name X**** L******
1 2 Physical Address 1624 Market St., Ste 226, #30327 Denver CO, 80202
1 2 Colorado 80202
1 2 United States
1 2 Telephone Number 303 6********
1 2 E-mail l******@aoges-cl.com
app s Confidentiality (long or short term)
1 2 Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?:Yes
1 2 Long-Term Confidentiality Does this application include a request for confidentiality for any portion(s) of the data contained in this application pursuant to 47 CFR § 0.459 of the Commission Rules?:No
if no date is supplied, the release date will be set to 45 calendar days past the date of grant.
app s Cognitive Radio & Software Defined Radio, Class, etc
1 2 Is this application for software defined/cognitive radio authorization?No
1 2 Equipment ClassNII - Unlicensed National Information Infrastructure TX
1 2 DTS - Digital Transmission System
1 2 Description of product as it is marketed: (NOTE: This text will appear below the equipment class on the grant)Geri Mixer
1 2 Related OET KnowledgeDataBase Inquiry: Is there a KDB inquiry associated with this application?No
1 2 Modular Equipment TypeDoes not apply
1 2 Purpose / Application is forOriginal Equipment
1 2 Composite Equipment: Is the equipment in this application a composite device subject to an additional equipment authorization?Yes
1 2 Related Equipment: Is the equipment in this application part of a system that operates with, or is marketed with, another device that requires an equipment authorization?No
1 2 Grant CommentsOutput power listed is maximum combined conducted power. This device contains 20 and 40 MHz signal bandwidth. The antennas used with this transmitter must be installed to provide a minimum separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter, except in accordance with FCC multi- transmitter product procedures. End-users must be provided with operating procedures for satisfying RF exposure compliance.
1 2 Output power listed is conducted power. This device contains 20MHz signal bandwidth. The antenna used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter, except in accordance with FCC multi- transmitter product procedures. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance.
1 2 Is there an equipment authorization waiver associated with this application?No
1 2 If there is an equipment authorization waiver associated with this application, has the associated waiver been approved and all information uploaded?No
app s Test Firm Name and Contact Information
1 2 Firm Name SHENZHEN STS TEST SERVICES CO.,LTD.
1 2 Name B**** Y****
1 2 Telephone Number 86-75********
1 2 E-mail B******@stsapp.com
Equipment Specifications
Line Rule Parts Grant Notes Lower Frequency Upper Frequency Power Output Tolerance Emission Designator Microprocessor Number
1 1 15E CC 5180 5240 0.0394
1 2 15E CC 5745 5825 0.0479
Line Rule Parts Grant Notes Lower Frequency Upper Frequency Power Output Tolerance Emission Designator Microprocessor Number
2 1 15C CC 2412.00000000 2462.00000000 0.0462000

 

some individual PII (Personally Identifiable Information) available on the public forms may be redacted, original source may include additional details

 

  This product uses the FCC Data API but is not endorsed or certified by the FCC