TheBPM usesRF energyonly for its internal Declaraon Electromagnec Emissions and Immunity For equipment and systems that are not life-supporng and are specified for use only in a shielded locaon Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operaon. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operang normally. Portable RF communicaons equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the unit, including cables specified by the manufacturer. Otherwise, degradaon of the performance of this equipment could result. The BPM is intended for use in the electromagnec environment specified below. The customer or the user of the BPM should assure that it is used in such an environment. Immunity test Electrostac discharge (ESD) IEC 61000-4-2 Electrical fast transient/ burst IEC 61000-4-4 Surge IEC 61000-4-5 Voltage dips, short interrupons, and voltage variaons on power supply input lines IEC 61000-4-11 IEC 60601 test level Contact: 8 kV Air: 2 kV, 4 kV, 8 kV, 15 kV Compliance level Contact: 8 kV Air: 2 kV, 4 kV, 8 kV, 15 kV 2 kV for power supply lines 1 kV for input/
output lines
+ 0.5kV, +1kV line(s) to line(s)
+ 0.5kV, +1kV,+ 2kV line(s) to earth Voltage dips:
0 % UT; 0,5 cycle 0 % UT; 1 cycle 70 % UT; 25/30 cycles Voltage interrupons:
N/A N/A Voltage dips: N/A 0 % UT; 250/300 cycle 30 A/m 50 Hz or 60 Hz Power frequency
(50/60 Hz) magnec field IEC 61000-4-8 Note: NA: not applicable. The BPM is battery-powered, not AC-powered 30 A/m 50 Hz and 60 Hz Electromagnec environment guidance Floors should be wood, concrete or ceramic le. If floors are covered with synthec material, the relave humidity should be at least 30 %. Mains power quality should be that of a typical commercial or hospital environment. A Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. If the user of the BPM requires connued operaon during power mains interrupons, it is recommended that the BPM be powered from an uninterrupble power supply or a baery. The BPM Power frequency magnec fields should be at levels characterisc of a typical locaon in a typical commercial or hospital environment. The BPM is intended for use in the electromagnec environment specified below. The customer or the user of the BPM should assure that it is used in such an environment. Immunity test Conducted RF IEC 61000-4-6 IEC 60601 test level 3 Vrms 150 kHz to 80 MHz Compliance level 6 Vrms:
in ISM and amateur radio bands between 0,15 MHz and 80 MHz N/A Radiated RF IEC 61000-4-3 80 % AM at 1 kHz 10 V/m 80 MHz 2,7 GHz 80 % AM at 1 kHz 10 V/m 80 MHz2,7 GHz 80 % AM at 1 kHz Electromagnec environment guidance Portable and mobile RF communicaons equipment should be used no closer to any part of the BPM including cables, than the recommended separaon distance calculated from the equation applicable to the frequency of the transmier. Recommended separaon distance:
d = 1,2 P d = 1,2 P 80MHz to 800 MHz d = 2,3 P 800MHz to 2,7 GHz Where P is the maximum output power rang of the transmier in was (W) according to the transmier manufacturer and d is the recommended separaon distance in meters (m). Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE2 These guidelines may not apply in all situaons. Electromagnec propagaon is affected by absorpon and reflecon from structures, objects, and people. Note3: "NA" means not applicable. The EUT is operang by baery, no supply for AC mains. Recommended separaon distance between portable and mobile RF communicaons equipment and the BPM The BPM is intended for use in an electromagnetic environment (for home healthcare) in which radiated RF disturbances are controlled. The customer or the user of the BPM can help prevent electromagnec interference by maintaining a minimum distance between portable and mobile RF communicaons equipment (transmiers) and the BPM as recommended below, according to the maximum output power of the communicaons equipment. Rated maximum output power of transmier W 0,01 0,1 1 10 100 Separaon distance according to frequency of transmier m 80 MHz to 800 MHz d =1,2P 0,12 0,38 1,2 3,8 12 800 MHz to 2,7 GHz d =2,3P 0,23 0,73 2,3 7,3 23 150 kHz to 80 MHz d =1,2P N/A N/A N/A N/A N/A For transmiers rated at a maximum output power not listed above, the recommended separaon distance d in meters (m) can be esmated using the equaon applicable to the frequency of the transmier, where p is the maximum output power rang of the transmier in watts (W) according to the transmier manufacturer. NOTE1 At 80 MHz and 800 MHz, the separaon distance for the higher frequency range applies. NOTE2 These guidelines may not apply in all situaons. Electromagnec propagaon is affected by absorpon and reflecon from structures, objects and people. Manufacturers declaraon-electromagnec immunity Test specificaons for ENCLOSURE PORT IMMUNITY to RF wireless communicaons equipment The BPM is intended for use in the electromagnec environment (for home healthcare) specified below. The customer or the user of the BPM should assure that it is used in such an environment. Test frequency
(MHz) Band a)
(MHz) Service a) Modulaon b) Maximum power
(W) Distance
(m) IMMUNITY TEST LEVEL
(V/m) Compliance LEVEL
(V/m)
(for home healthcare) 385 380 390 TETRA 400 450 430 470 GMRS 460, FRS 460 710 745 780 810 870 930 1 720 1 845 1 970 2 450 5 240 5 500 5 785 704 787 LTE Band 13,17 800 960 1 700 1 990 2 400 2 570 GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5 GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3, 4, 25; UMTS Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7 5 100 5 800 WLAN 802.11 a/n Pulse modulaon b) 18 Hz FM c) 5 kHz deviaon 1 kHz sine Pulse modulaon b) 217 Hz Pulse modulaon b) 18 Hz Pulse modulaon b) 217 Hz Pulse modulaon b) 217 Hz Pulse modulaon b) 217 Hz 1,8 0,3 2 0,3 0,2 0,3 27 28 9 2 2 2 0,3 28 0,3 28 0,3 28 0,2 0,3 9 27 28 9 28 28 28 9 NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transming antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permied by IEC 61000-4-3. a) For some services, only the uplink frequencies are included. b) The carrier shall be modulated using a 50 % duty cycle square wave signal. c) As an alternave to FM modulaon, 50 % pulse modulaon at 18 Hz may be used because while it does not represent actual modulaon, it would be worst case. Maintenance 1 No Firmware update is needed during the service life of this product. 2 On very rare occasions, the device might need reinstalling the firmware due to hardware malfuncon, in this case, the intended use remains unchanged. 3 Any modificaon or change to the firmware is prohibited. Contact your local dealer far repair service or contact the Ya Horng customer service at support.yahorng.com.